Abstract
Background
In recent years, the use of modern cementless implants in total knee arthroplasty has been increasing in popularity. These implants take advantage of new technologies such as additive manufacturing and potentially provide a promising alternative to cemented implant designs. The purpose of this study was to compare implant migration and tibiofemoral contact kinematics of a cementless primary total knee arthroplasty (TKA) implanted using either a gap balancing (GB) or measured resection (MR) surgical technique.
Methods
Thirty-nine patients undergoing unilateral TKA were recruited and assigned based on surgeon referral to an arthroplasty surgeon who utilizes either a GB (n = 19) or a MR (n = 20) surgical technique. All patients received an identical fixed-bearing, cruciate-retaining beaded peri-apatite coated cementless femoral component and a pegged highly porous cementless tibial baseplate with a condylar stabilizing tibial insert. Patients underwent a baseline radiostereometric analysis (RSA) exam at two weeks post-operation, with follow-up visits at six weeks, three months, six months, and one year post-operation. Migration including maximum total point motion (MTPM) of the femoral and tibial components was calculated over time. At the one year visit patients also underwent a kinematic exam using the RSA system.
Results
Mean MTPM of the tibial component at one year post-operation was not different (mean difference = 0.09 mm, p = 0.980) between the GB group (0.85 ± 0.37 mm) and the MR group (0.94 ± 0.41 mm). Femoral component MTPM at one year post-operation was also not different (mean difference = 0.27 mm, p = 0.463) between the GB group (0.62 ± 0.34 mm) and the MR group (0.89 ± 0.44 mm). Both groups displayed a lateral pivot pattern with similar frequencies of condylar separation.
Conclusion
There was no difference in implant migration and kinematics of a single-radius, cruciate retaining cementless TKA performed using a GB or MR surgical technique. The magnitude of migration suggests there is no risk of early loosening. The results provide support for the use of a cementless TKA as a viable alternative to cemented fixation.