Introduction: The hip resurfacing in younger patients is being performed more frequently in the UK. At the same time conventional Total Hip Replacement (THR) is also being performed.. We carried out a study to compare the patient satisfaction and outcome following Birmingham Hip Resurfacing (BHR) and Total Hip Replacement (THR) in patients below the age of 55 years. Methods: There were 93
Introduction: The Birmingham Hip Resurfacing (BHR) system comprises both a
Aims. The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Methods. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis. Results. Mean follow-up was 20.9 years (19.3 to 22.4). Registry data revealed that 19 hips (8.1%) had been revised and 26 patients (12%) had died from causes unrelated to the
This study reports the mid-term results of a large bearing uncemented metal on metal total hip replacement (MOMHTHR) matched series using the Synergy stem and Birmingham modular head in 36 hips (mean follow up 61 months). All patients underwent clinical, metal ion and MRI assessment. Wear analysis was performed on retrieved heads using Redlux non-contact optical profilometry. Seven patients (19%) have undergone revision surgery. All revisions had two or more of either symptoms, high metal ions or an MRI suggestive of an adverse reaction to metal debris (ARMD). There was no evidence of component malposition or impingement. Frank staining of tissues together with high volume dark brown fluid collections were found in all cases. All stems and cups were well fixed. In 4 cases pubic and ischial lysis (adjacent to the inferior fins) was observed. All 7 cases had radiological, intraoperative and histological evidence of ARMD (Figure 1). The failure cohort had significantly higher whole blood cobalt ion levels and OHS (p = 0.001), but no significant difference in cup size (p = 0.77), gender predominance, stem offset or cup position (p = 0.12). Sleeves had been used in all revision cases. Wear analysis (n = 4) demonstrated increased wear at the trunnion/sleeve interface in a distribution compatible with micromotion (Figure 2). There was normal wear at the articulating surface. This series further demonstrates unacceptable failure rates in LHMOMTHR in a series where a compatible stem for the
Modern metal-on-metal hip resurfacing has been
widely performed in the United Kingdom for over a decade. However,
the literature reports conflicting views of the benefits: excellent
medium- to long-term results with some brands in specific subgroups,
but high failure rates and local soft-tissue reactions in others.
The National Joint Registry for England and Wales (NJR) has collected
data on all hip resurfacings performed since 2003. This retrospective
cohort study recorded survival time to revision from a resurfacing
procedure, exploring risk factors independently associated with
failure. All patients with a primary diagnosis of osteoarthritis
who underwent resurfacing between 2003 and 2010 were included in
the analyses. Cox’s proportional hazard models were used to analyse
the extent to which the risk of revision was related to patient,
surgeon and implant covariates. A total of 27 971 hip resurfacings were performed during the
study period, of which 1003 (3.59%) underwent revision surgery.
In the final adjusted model, we found that women were at greater
risk of revision than men (hazard ratio (HR) = 1.30, p = 0.007),
but the risk of revision was independent of age. Of the implant-specific
predictors, five brands had a significantly greater risk of revision
than the Birmingham Hip Resurfacing (BHR) (ASR: HR = 2.82, p <
0.001,
Conserve: HR = 2.03, p <
0.001, Cormet: HR = 1.43, p = 0.001,
Durom: HR = 1.67, p <
0.001, Recap: HR = 1.58, p = 0.007). Smaller
femoral head components were also significantly more likely to require
revision (≤ 44 mm: HR = 2.14, p <
0.001, 45 to 47 mm: HR = 1.48,
p = 0.001) than medium or large heads, as were operations performed
by low-volume surgeons (HR = 1.36, p <
0.001). Once these influences
had been removed, in 4873 male patients <
60 years old undergoing
resurfacing with a
The difference in the mean values regarding inclination was greater than would be expected by chance; there was a statistically significant difference (P = 0,010).
Navigation technique was discussed to equalize the drawback of MIS. However, tools like imageless navigation may further improve the cup position even in traditional approach.
Recent events have highlighted the importance
of implant design for survival and wear-related complications following
metal-on-metal hip resurfacing arthroplasty. The mid-term survival
of the most widely used implant, the Birmingham Hip Resurfacing
(BHR), has been described by its designers. The aim of this study
was to report the ten-year survival and patient-reported functional
outcome of the
We previously reported the five to ten-year results of the Birmingham Hip Resurfacing (BHR) implant. The purpose of this study was to evaluate the survivorship, radiographic results, and clinical outcomes of the
Reports of improved functional outcome of Metal on Metal Hip Resurfacing Arthroplasty (mHRA) to Total Hip Replacement needs to be balanced with concerns of metal ion release. By removing cobalt-chrome, cHRA reduces these risks. To the author's knowledge, there is no data available on functional outcomes of cHRA, therefore the aim of the study was to compare the function between cHRA patients and mHRA patients. 24 patients received a unilateral cHRA (H1, Embody) and was compared to 24 age and gender matched patients with a unilateral mHRA (BHR, Smith and Nephew). All patients completed the Oxford Hip Score (OHS)[T2] and underwent gait analysis on an instrumented treadmill before and at a mean of 74wks (+/− 10) for mHRA and 53wks (+/− 2) for cHRA post op. Walking trials started at 4km/h and increased in 0.5km/h increments until a top walking speed (TWS) was achieved. Vertical ground reaction forces (GRF) were recorded along with the symmetry index (SI). Spatiotemporal measures of gait were also recorded. Vertical GRF were captured for the entire normalised stance phase using statistical parametric mapping (SPM; CI = 95%). The gain in OHS was similar: H1 (25-46), BHR(27-47). TWS increased by 19% with H1 (6.02 – 8.0km/hr), and 20% with
Aims. The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. Methods. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken. Results. Overall implant survival was 83.50% (95% confidence interval (CI) 0.79 to 0.90) at 25 years, and the number at risk was 79. Survival in male patients at 25 years was 89.5% (95% CI 0.83 to 0.96) compared to 66.9% for female patients (95% CI 0.51 to 0.83). Ten additional failures occurred in the period of ten to 25 years. These involved an adverse reaction to metal debris in four patients, a periprosthetic femoral neck fracture affecting five patients, and aseptic loosening in one patient. The median chromium levels were 49.50 nmol/l (interquartile range (IQR) 34 to 70), and the median cobalt serum levels were 42 nmol/l (IQR 24.50 to 71.25). The median OHS at last follow-up was 35 (IQR 10 to 48). During the 25-year study period, 29 patients died. Patient survival at 25 years was 75.10% (95% CI 0.67 to 0.83). Conclusion. This study demonstrates that MoM hip resurfacing using the
Aims. The first aim of this study was to evaluate whether preoperative renal function is associated with postoperative changes in whole blood levels of metal ions in patients who have undergone a Birmingham Hip Resurfacing (BHR) arthroplasty with a metal-on-metal bearing. The second aim was to evaluate whether exposure to increased cobalt (Co) and chromium (Cr) levels for ten years adversely affected renal function. Methods. As part of a multicentre, prospective post-approval study, whole blood samples were sent to a single specialized laboratory to determine Co and Cr levels, and the estimated glomerular filtration rate (eGFR). The study included patients with 117 unrevised unilateral BHRs. There were 36 females (31%). The mean age of the patients at the time of surgery of 51.3 years (SD 6.5), and they all had preoperative one-, four-, five-, and ten-year laboratory data. The mean follow-up was 10.1 years (SD 0.2). Results. Median Co levels at one year postoperatively increased significantly compared with the preoperative values, by a factor of 9.7, from 0.13 to 1.26 ppb (p < 0.001), and the median Cr levels increased significantly by a factor of 2.5, from 0.60 to 1.50 ppb (p < 0.001). Lower preoperative eGFRs were associated with significantly larger increases in Co at one year compared with the preoperative levels (ρ = -0.26; p = 0.005), but there was no relationship between preoperative eGFRs and changes in Cr at one year (ρ = -0.13; p = 0.153). Metal levels remained relatively constant with the passage of time, with a median ten-year value of 1.12 ppb for Co and 1.29 ppb for Cr. There was no significant relationship between the Co and Cr levels at ten-year follow-up and the change in eGFR from the preoperative level to that at ten years (ρ = -0.02; p = 0.827 for Co; ρ = -0.008; p = 0.933 for Cr). Conclusion. Although patients with lower preoperative eGFRs tended to have larger increases in Co levels at one year, increased metal levels for patients who underwent unilateral
Aims. The aim of this study was to report the implant survival and patient-reported outcome measures (PROMs) in a consecutive series of patients aged less than 50 years at the time of arthroplasty using the Birmingham Hip Resurfacing system (BHR), with a minimum follow-up of ten years. Patients and Methods. A total of 226 patients with osteoarthritis of the hip, who underwent
When the Birmingham Hip Resurfacing (BHR) metal-on-metal implant system was approved by the United States Food and Drug Administration in 2006, a multicenter, prospective, post-approval study (PAS) was required. This study uses data from the PAS to investigate metal level and glomerular filtration rate (GFR) trends over the first decade in vivo. Between October 2006 and March 2011, 290 primary
A concern of metal on metal hip resurfacing arthroplasty is long term exposure to Cobalt (Co) and Chromium (CR) wear debris from the bearing. This study compares whole blood metal ion levels from patients drawn at one-year following Birmingham Hip Resurfacing (BHR) to levels taken at a minimum 10-year follow-up. A retrospective chart review was conducted to identify all patients who underwent a
This study compared the demographic, clinical and patient-reported outcomes after total hip replacement (THR) and Birmingham Hip Resurfacing (BHR) carried out by a single surgeon. Patients completed a questionnaire that included the WOMAC, SF-36 scores and comorbid medical conditions. Data were collected before operation and one year after. The outcome scores were adjusted for age, gender, comorbid conditions and, at one year, for the pre-operative scores. There were 214 patients with a THR and 132 with a
We present a consecutive case cohort of the first 100 Birmingham Hip Resurfacing (BHR)'s in 90 patients with a minimum follow up of 20 years. All procedures were performed by a single surgeon having commenced the study in 1998. The original cohort included 68 males with 75 hips (7 bilateral) and 22 females with 25 hips (3 bilateral). The mean age at index procedure was 52. Patients were recalled to review in clinic as per Medicines and Healthcare products Regulatory Agency guidelines with x-rays, metal ions and Harris Hip Scores recorded. After a minimum of 20 year follow-up review the known overall revision rate is 11%. 11 have died and 7 have been lost to review. In males the known revision rate is 4/75 (5%), 3 of which were due to early fractures and 1 revision for infection at 5 years. The known revision rate in females is 7/25 (28%) of which 6 were due to adverse reactions to metal debris; 3 at 7–8 yrs, 1 at 15 yrs, 2 at 18 years, and 1 at 2 months for avascular necrosis. The mean 20 year metal ions results were cobalt 36.6 nmol/l (range 7.4–232.1) and chromium 32.1 nmol/l (range 6–120.8). The mean Harris hip score was 88.5 (range 38.6–99.8). This case series with 20 years follow-up demonstrates excellent long term survival of the BHRs further to registry
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64
The use of Birmingham hip resurfacing (BHR) remains controversial due to the increased revision rate in female patients. We compared the outcomes of
There have been no large comparative studies of the blood levels of metal ions after implantation of commercially available hip resurfacing devices which have taken into account the effects of femoral size and inclination and anteversion of the acetabular component. We present the results in 90 patients with unilateral articular surface replacement (ASR) hip resurfacings (mean time to blood sampling 26 months) and 70 patients with unilateral Birmingham Hip Resurfacing (BHR) implants (mean time 47 months). The whole blood and serum chromium (Cr) and cobalt (Co) concentrations were inversely related to the size of the femoral component in both groups (p <
0.05). Cr and Co were more strongly influenced by the position of the acetabular component in the case of the ASR, with an increase in metal ions observed at inclinations >
45° and anteversion angles of <
10° and >
20°. These levels were only increased in the
Abstract. Introduction. In general the life expectancy of population is improving. This is causing to increase case load of peri-prosthesis fractures after joint replacements. We present our results of peri-prosthesis fracture around hip managed by revision arthroplasty. Methods. A retrospective analysis of 24 consecutive patients of periprosthetic hip fracture treated with a revision arthroplasty at Major Trauma Centre between February 2021 and January 2022. Results. 12 male and 12 female patients, average age 78 years. 3 fractures around