header advert
Results 1 - 3 of 3
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 48 - 48
1 Feb 2017
Kang K Trinh T Yoo O Jang Y Lee M Lim D
Full Access

Introduction

The Rotational alignment is an important factor for survival total knee Arthroplasty. Rotational malalignment causes knee pain, global instability, and wear of the polyethylene inlay. Also, the anterior cortex line was reported that more reliable and more easily identifiable landmark for correct tibial component alignment. The aims of the current study is to identify effect of inserting the tibial baseplate of using anterior cortex line landmark of TKA on stress/strain distributions within cortical bone and bone cement. Through the current study, final aim is to suggest an alternative position of tibia baseplate for reduction of TKA failures with surgical convenience.

Materials and Method

A three-dimensional tibia FE model with TKA was generated based on a traditional TKA surgical guideline. Here, a commercialized TKA (LOSPA, Corentc, Korea) was considered corresponded to a patient specific tibia morphology. Tibia baseplate was positioned at anterior cortex line. Alternative two positions were also considered based on tibia tuberosity 1/3 line and tibia tuberosity end line known as a gold standard (Fig. 1-A). Loading and boundary conditions for the FE analysis were determined based on five activities of daily life of persons with TKA (Fig. 1-B). FE model was additionally validated comparing with an actual mechanical test.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 81 - 81
1 May 2016
Kang K Trinh T Jang Y Yoo O Lee M Lim D
Full Access

Introduction

Revision total knee arthroplasy (TKA) has been often used with a metal block augmentation for patients with poor bone quality. However, bone resorption beneath metal block augmentation has been still reported and little information about the reasons of the occurrence of bone resorption is available. The aim of the current study is to identify a possibility of the potential occurrence of bone resorption beneath metal block augmentation, through evaluation of strain distribution beneath metal block augmentation in revision TKA with metal block augmentation, during high deep flexion.

Materials and Method

LOSPA, revision TKA with a metal block augmentation (Baseplate size #5, Spacer size #5, Stem size Φ9, L30, Augment #5 T5) was considered in this study. For the test, the tibia component of LOSPA was implanted to the tibia sawbone (left, #3401, Sawbones EuropeAB, Malmö, Sweden), which was corresponded to a traditional TKR surgical guideline. The femoral component of LOSPA was mounted to a customized jig attached to the Instron 8872 (Instron, Norwood, MA, USA), which was designed specially to represent the angles ranged from 0° to 140° with consideration of a rollback of knee joint (Figure. 1). Here, a compressive load of 1,600N (10N/s) was applied for each angle. Strain distribution was then measured from rossete strain gauge (Half Bridge type, CAS, Seoul, Korea) together (Figure 1).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 84 - 84
1 May 2016
Trinh T Kang K Lim D Yoo O Lee M Jang Y
Full Access

Introduction

Revision total knee arthroplasty (TKA) has been often used with a metal block augmentation for patients with poor bone quality. However, bone defects are frequently detected in revision TKA used with metal block augmentation. This study focused on identification of a potential possibility of the bone defect occurrence through the evaluation of the strain distribution on the cortical bone of the tibia implanted revision TKA with metal block augmentation, during high deep flexion.

Materials and Methods

Composite tibia finite element (FE) model was developed and revision TKA FE model with a metal block augmentation (Baseplate size #5 44AP/67ML, Spacer size #5 44AP/67ML, Stem size Φ9, L30, Augment #5 44AP/67ML thickness 5mm) was integrated with the composite tibia FE model. 0°, 30° 60°, 90°, 120° and 140° flexion positions were then considered with femoral rollback phenomenon [Fig 1.A]. A compressive load of 1,600N through the femoral component was applied to the composite tibia FE model integrated with the tibia component, sharing by the medial and lateral condyles, simulating a stance phase before toe-off [Fig 1.B].