header advert
Results 1 - 50 of 164
Results per page:
Bone & Joint Research
Vol. 13, Issue 4 | Pages 193 - 200
23 Apr 2024
Reynolds A Doyle R Boughton O Cobb J Muirhead-Allwood S Jeffers J

Aims

Manual impaction, with a mallet and introducer, remains the standard method of installing cementless acetabular cups during total hip arthroplasty (THA). This study aims to quantify the accuracy and precision of manual impaction strikes during the seating of an acetabular component. This understanding aims to help improve impaction surgical techniques and inform the development of future technologies.

Methods

Posterior approach THAs were carried out on three cadavers by an expert orthopaedic surgeon. An instrumented mallet and introducer were used to insert cementless acetabular cups. The motion of the mallet, relative to the introducer, was analyzed for a total of 110 strikes split into low-, medium-, and high-effort strikes. Three parameters were extracted from these data: strike vector, strike offset, and mallet face alignment.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 78 - 78
2 Jan 2024
Ponniah H Edwards T Lex J Davidson R Al-Zubaidy M Afzal I Field R Liddle A Cobb J Logishetty K
Full Access

Anterior approach total hip arthroplasty (AA-THA) has a steep learning curve, with higher complication rates in initial cases. Proper surgical case selection during the learning curve can reduce early risk. This study aims to identify patient and radiographic factors associated with AA-THA difficulty using Machine Learning (ML).

Consecutive primary AA-THA patients from two centres, operated by two expert surgeons, were enrolled (excluding patients with prior hip surgery and first 100 cases per surgeon). K- means prototype clustering – an unsupervised ML algorithm – was used with two variables - operative duration and surgical complications within 6 weeks - to cluster operations into difficult or standard groups.

Radiographic measurements (neck shaft angle, offset, LCEA, inter-teardrop distance, Tonnis grade) were measured by two independent observers. These factors, alongside patient factors (BMI, age, sex, laterality) were employed in a multivariate logistic regression analysis and used for k-means clustering. Significant continuous variables were investigated for predictive accuracy using Receiver Operator Characteristics (ROC).

Out of 328 THAs analyzed, 130 (40%) were classified as difficult and 198 (60%) as standard. Difficult group had a mean operative time of 106mins (range 99–116) with 2 complications, while standard group had a mean operative time of 77mins (range 69–86) with 0 complications. Decreasing inter-teardrop distance (odds ratio [OR] 0.97, 95% confidence interval [CI] 0.95–0.99, p = 0.03) and right-sided operations (OR 1.73, 95% CI 1.10–2.72, p = 0.02) were associated with operative difficulty. However, ROC analysis showed poor predictive accuracy for these factors alone, with area under the curve of 0.56. Inter-observer reliability was reported as excellent (ICC >0.7).

Right-sided hips (for right-hand dominant surgeons) and decreasing inter-teardrop distance were associated with case difficulty in AA-THA. These data could guide case selection during the learning phase. A larger dataset with more complications may reveal further factors.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 83 - 83
23 Jun 2023
Cobb J
Full Access

The trend towards more minimal access has led to a series of instruments being developed to enable adequate access for Direct Anterior Approach (DAA) for hip arthroplasty. These include longer levers, hooks attached to the operating table and a series of special attachments to the operating table to position the leg and apply traction where necessary. The forces applied in this way may be transmitted locally, damaging muscle used as a fulcrum, or the knee and ankle joints when torque has to be applied to the femur through a boot. The arthroplasty surgeon's aim is to minimise the forces applied to both bone and soft tissue during surgery.

We surmised that the forces needed for adequate access were related to the extent of the capsular and soft tissue releases, and that they could be measured and optimised. with the aim of minimising the forces applied to the tissues around the hip.

Eight fresh frozen specimens from pelvis to mid tibia from four cadavers were approached using the DAA. A 6-axis force/torque sensor and 6-axis motion tracking sensor were attached to a threaded rod securely fastened to the tibial and femoral diaphysis. The torque needed to provide first extension, then external rotation, adequate for hip arthroplasty were measured as the capsular structures were divided sequentially.

The Zona Orbicularis (ZO) and Ischiofemoral Ligament(IFL) contributed most of the resistance to both extension (4.0 and 3.1Nm) and external rotation torque (5.8 and 3.9Nm). The contributions of the conjoint tendon (1.5 and 2.4Nm) and piriformis (1.2 and 2.3Nm) were substantially smaller.

By releasing the Zona Orbicularis and Ischiofemoral Ligament, the torque needed to deliver the femur for hip arthroplasty could be reduced to less than the torque needed to open a jar (2.9–5.5Nm).


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 40 - 40
7 Jun 2023
Edwards T Soussi D Gupta S Khan S Patel A Patil A Badri D Liddle A Cobb J Logishetty K
Full Access

Superior teamwork in the operating theatre is associated with improved technical performance and clinical outcomes. Yet modern rota patterns, workforce shortages, and increasing complexity of surgery, means that there is less familiarity between staff and the required choreography. Immersive Virtual Reality (iVR) can successfully train surgical staff individually, however iVR team training has yet to be investigated. We aimed to design a multiplayer iVR platform for anterior approach total hip arthroplasty (AA-THA) and assess if multiplayer iVR training was superior to single player training for acquisition of both technical and non-technical skills.

An iVR platform with choreographed roles for the surgeon and scrub nurse was developed using Cognitive Task Analysis. Forty participants were randomised to individual or team iVR training. Individually- trained participants practiced alongside virtual avatar counterparts, whilst teams trained live in pairs. Both groups underwent five iVR training sessions over 6-weeks. Subsequently, they underwent a real-life assessment in which they performed AA-THA on a high-fidelity model with real equipment in a simulated theatre. Teams performed together and individually trained participants were randomly paired up. Videos were marked by two blinded assessors recording the NOTSS, NOTECHS II and SPLINTS scores - validated technical and non-technical scores assessing surgeon and scrub nurse skills. Secondary outcomes were procedure time and number of technical errors.

Teams outperformed individually trained participants for non-technical skills in the real-world assessment (NOTSS 13.1 ± 1.5 vs 10.6 ± 1.6, p =0.002, NOTECHS-II score 51.7 ± 5.5 vs 42.3 ± 5.6, p=0.001 and SPLINTS 10 ± 1.2 vs 7.9 ± 1.6, p = 0.004). They completed the assessment 28.1% faster (27.2 minutes ± 5.5 vs 41.8 ±8.9, p<0.001), and made fewer than half the number of technical errors (10.4 ± 6.1 vs 22.6 ± 5.4, p<0.001).

Multiplayer training leads to faster surgery with fewer technical errors and the development of superior non-technical skills for anterior approach total hip arthroplasty. The convention of surgeons and nurses training separately, but undertaking real complex surgery together, can be supplanted by team training, delivered through immersive virtual reality.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 37 - 37
7 Jun 2023
Edwards T Kablean-Howard F Poole I Edwards J Karia M Liddle A Cobb J Logishetty K
Full Access

Superior team performance in surgery leads to fewer technical errors, reduced mortality, and improved patient outcomes. Scrub nurses are a pivotal part of this team, however they have very little structured training, leading to high levels of stress, low confidence, inefficiency, and potential for harm. Immersive virtual reality (iVR) simulation has demonstrated excellent efficacy in training surgeons. We tested the efficacy of an iVR curriculum for training scrub nurses in performing their role in an anterior approach total hip arthroplasty (AA-THA).

Sixty nursing students were included in this study and randomised in a 1:1 ratio to learning the scrub nurse role for an AA-THA using either conventional training or iVR. The training was derived through expert consensus with senior surgeons, scrub nurses and industry reps. Conventional training consisted of a 1-hour seminar and 2 hours of e-learning where participants were taught the equipment and sequence of steps. The iVR training involved 3 separate hour-long sessions where participants performed the scrub nurse role with an avatar surgeon in a virtual operation. The primary outcome was their performance in a physical world practical objective assessment with real equipment. Data were confirmed parametric using the Shapiro-Wilk test and means compared using the independent samples student's t-test.

53 participants successfully completed the study (26 iVR, 27 conventional) with a mean age of 31±9 years. There were no significant differences in baseline characteristics or baseline knowledge test scores between the two groups (p>0.05). The iVR group significantly outperformed the conventionally trained group in the real-world assessment, scoring 66.9±17.9% vs 41.3±16.7%, p<0.0001.

iVR is an easily accessible, low cost training modality which could be integrated into scrub nursing curricula to address the current shortfall in training. Prolonged operating times are strongly associated with an increased risk of developing serious complications. By upskilling scrub nurses, operations may proceed more efficiently which in turn may improve patient safety.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 6 - 6
17 Apr 2023
Maslivec A Leon D Cobb J
Full Access

Reports of improved functional outcome of Metal on Metal Hip Resurfacing Arthroplasty (mHRA) to Total Hip Replacement needs to be balanced with concerns of metal ion release. By removing cobalt-chrome, cHRA reduces these risks. To the author's knowledge, there is no data available on functional outcomes of cHRA, therefore the aim of the study was to compare the function between cHRA patients and mHRA patients.

24 patients received a unilateral cHRA (H1, Embody) and was compared to 24 age and gender matched patients with a unilateral mHRA (BHR, Smith and Nephew). All patients completed the Oxford Hip Score (OHS)[T2] and underwent gait analysis on an instrumented treadmill before and at a mean of 74wks (+/− 10) for mHRA and 53wks (+/− 2) for cHRA post op. Walking trials started at 4km/h and increased in 0.5km/h increments until a top walking speed (TWS) was achieved. Vertical ground reaction forces (GRF) were recorded along with the symmetry index (SI). Spatiotemporal measures of gait were also recorded. Vertical GRF were captured for the entire normalised stance phase using statistical parametric mapping (SPM; CI = 95%).

The gain in OHS was similar: H1 (25-46), BHR(27-47). TWS increased by 19% with H1 (6.02 – 8.0km/hr), and 20% with BHR (6.02 – 7.37km/hr). SPM of the entire gait cycle illustrated the restoration of symmetry in both groups with no difference in GRF across the stance phase between groups at 5km/hr pre-op and post-op. At faster speeds (6.5km/hr), H1 patients had a mid-support GRF slightly closer to normal compared to BHR. Both groups increased step length similar from pre to post op (H1:0.76 – 0.85cm, BHR:0.77-0.86cm).

In this study, subjective and objective functional outcome measures suggest that short term functional outcomes of ceramic resurfacing is not inferior to metal resurfacing.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 13 - 13
11 Apr 2023
Edwards T Gupta S Soussi D Patel A Khan S Liddle A Cobb J Logishetty K
Full Access

Current evidence suggests that superior surgical team performance is linked to fewer intra-operative errors, reductions in mortality and even improved patient outcomes. Virtual reality has demonstrated excellent efficacy in training surgeons and scrub nurses individually, however its impact on training teams is currently unknown. This study aimed to assess if training together (scrub nurse and surgeon) in an innovative multiplayer virtual reality program was superior to single player training for novices learning anterior approach total hip arthroplasty (AA-THA).

40 participants (20 novice surgeons (CT1-ST3 level) and 20 novice scrub nurses) were enrolled in this study and randomised to individual or team virtual reality training. Individually-trained participants played with virtual avatar counterparts, whilst teams trained live in pairs (surgeon and scrub nurse). Both groups underwent 5 VR training sessions over 6 weeks. Subsequently, they underwent a real-life assessment in which they performed AA-THA on a high-fidelity model with real equipment in a simulated operating theatre. Teams performed together and individually-trained participants were randomly paired up with a solo player of the opposite role. Videos of the assessment were marked by two blinded expert assessors. The primary outcome was team performance as graded by the validated NOTECHs II score. Secondary outcomes were procedure time and number of technical errors from an expert pre-defined protocol.

Teams outperformed individually-trained participants for non-technical skills in the real-world assessment (NOTECHS-II score 50.3 ± 6.04 vs 43.90 ± 5.90, p=0.0275). They completed the assessment 28.1% faster (31.22 minutes ±2.02 vs 43.43 ±2.71, p=0.01), and made close to half the number of technical errors when compared to the individual group (12.9 ± 8.3 vs 25.6 ± 6.1, p=0.001).

Multiplayer, team training appears to lead to faster surgery with fewer technical errors and the development of superior non-technical skills.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 104 - 104
4 Apr 2023
Edwards T Khan S Patel A Gupta S Soussi D Liddle A Cobb J Logishetty K
Full Access

Evidence supporting the use of virtual reality (VR) training in orthopaedic procedures is rapidly growing. However, the impact of the timing of delivery of this training is yet to be tested. We aimed to investigate whether spaced VR training is more effective than massed VR training.

24 medical students with no hip arthroplasty experience were randomised to learning the direct anterior approach total hip arthroplasty using the same VR simulation, training either once-weekly or once-daily for four sessions. Participants underwent a baseline physical world assessment on a saw bone pelvis. The VR program recorded procedural errors, time, assistive prompts required and hand path length across four sessions. The VR and physical world assessments were repeated at one-week, one-month, and 3 months after the last training session.

Baseline characteristics between the groups were comparable (p > 0.05). The daily group demonstrated faster skills acquisition, reducing the median ± IQR number of procedural errors from 68 ± 67.05 (session one) to 7 ± 9.75 (session four), compared to the weekly group's improvement from 63 ± 27 (session one) to 13 ± 15.75 (session four), p < 0.001. The weekly group error count plateaued remaining at 14 ± 6.75 at one-week, 16.50 ± 16.25 at one-month and 26.45 ± 22 at 3-months, p < 0.05. However, the daily group showed poorer retention with error counts rising to 16 ± 12.25 at one-week, 17.50 ± 23 at one-month and 41.45 ± 26 at 3-months, p<0.01. A similar effect was noted for the number of assistive prompts required, procedural time and hand path length. In the real-world assessment, both groups significantly improved their acetabular component positioning accuracy, and these improvements were equally maintained (p<0.01).

Daily VR training facilitates faster skills acquisition; however weekly practice has superior skills retention.


Bone & Joint Open
Vol. 4, Issue 3 | Pages 129 - 137
1 Mar 2023
Patel A Edwards TC Jones G Liddle AD Cobb J Garner A

Aims

The metabolic equivalent of task (MET) score examines patient performance in relation to energy expenditure before and after knee arthroplasty. This study assesses its use in a knee arthroplasty population in comparison with the widely used Oxford Knee Score (OKS) and EuroQol five-dimension index (EQ-5D), which are reported to be limited by ceiling effects.

Methods

A total of 116 patients with OKS, EQ-5D, and MET scores before, and at least six months following, unilateral primary knee arthroplasty were identified from a database. Procedures were performed by a single surgeon between 2014 and 2019 consecutively. Scores were analyzed for normality, skewness, kurtosis, and the presence of ceiling/floor effects. Concurrent validity between the MET score, OKS, and EQ-5D was assessed using Spearman’s rank.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 20 - 20
1 Dec 2022
Ng G El Daou H Bankes M Cobb J Beaulé P
Full Access

Femoroacetabular impingement (FAI) – enlarged, aspherical femoral head deformity (cam-type) or retroversion/overcoverage of the acetabulum (pincer-type) – is a leading cause for early hip osteoarthritis. Although anteverting/reverse periacetabular osteotomy (PAO) to address FAI aims to preserve the native hip and restore joint function, it is still unclear how it affects joint mobility and stability. This in vitro cadaveric study examined the effects of surgical anteverting PAO on range of motion and capsular mechanics in hips with acetabular retroversion.

Twelve cadaveric hips (n = 12, m:f = 9:3; age = 41 ± 9 years; BMI = 23 ± 4 kg/m2) were included in this study. Each hip was CT imaged and indicated acetabular retroversion (i.e., crossover sign, posterior wall sign, ischial wall sign, retroversion index > 20%, axial plane acetabular version < 15°); and showed no other abnormalities on CT data. Each hip was denuded to the bone-and-capsule and mounted onto a 6-DOF robot tester (TX90, Stäubli), equipped with a universal force-torque sensor (Omega85, ATI). The robot positioned each hip in five sagittal angles: Extension, Neutral 0°, Flexion 30°, Flexion 60°, Flexion 90°; and performed hip internal-external rotations and abduction-adduction motions to 5 Nm in each position. After the intact stage was tested, each hip underwent an anteverting PAO, anteverting the acetabulum and securing the fragment with long bone screws. The capsular ligaments were preserved during the surgery and each hip was retested postoperatively in the robot. Postoperative CT imaging confirmed that the acetabular fragment was properly positioned with adequate version and head coverage. Paired sample t-tests compared the differences in range of motion before and after PAO (CI = 95%; SPSS v.24, IBM).

Preoperatively, the intact hips with acetabular retroversion demonstrated constrained internal-external rotations and abduction-adduction motions. The PAO reoriented the acetabular fragment and medialized the hip joint centre, which tightened the iliofemoral ligament and slackenend the pubofemoral ligament. Postoperatively, internal rotation increased in the deep hip flexion positions of Flexion 60° (∆IR = +7°, p = 0.001) and Flexion 90° (∆IR = +8°, p = 0.001); while also demonstrating marginal decreases in external rotation in all positions. In addition, adduction increased in the deep flexion positions of Flexion 60° (∆ADD = +11°, p = 0.002) and Flexion 90° (∆ADD = +12°, p = 0.001); but also showed marginal increases in abduction in all positions.

The anteverting PAO restored anterosuperior acetabular clearance and increased internal rotation (28–33%) and adduction motions (29–31%) in deep hip flexion. Restricted movements and positive impingement tests typically experienced in these positions with acetabular retroversion are associated with clinical symptoms of FAI (i.e., FADIR). However, PAO altered capsular tensions by further tightening the anterolateral hip capsule which resulted in a limited external rotation and a stiffer and tighter hip. Capsular tightness may still be secondary to acetabular retroversion, thus capsular management may be warranted for larger corrections or rotational osteotomies. In efforts to optimize surgical management and clinical outcomes, anteverting PAO is a viable option to address FAI due to acetabular retroversion or overcoverage.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_7 | Pages 83 - 83
1 Jul 2022
Dandridge O Garner A Amis A Cobb J Arkel RV
Full Access

Abstract

Patellofemoral Arthroplasty (PFA) is an alternative to TKA for patellofemoral osteoarthritis that preserves tibiofemoral compartments. It is unknown how implant positioning affects biomechanics, especially regarding the patella. This study analysed biomechanical effects of femoral and patellar component position, hypothesising femoral positioning is more important.

Nine cadaveric knees were studied using a repeated-measures protocol. Knees were tested intact, then after PFA implanted in various positions: neutral (as-planned), patellar over/understuffing (±2mm), patellar tilt, patellar flexion, femoral rotation, and femoral tilt (all ±6°). Arthroplasties were implemented with CT-designed patient-specific instrumentation. Anterior femoral cuts referenced Whiteside's line and all femoral positions ensured smooth condyle-to-component transition. Knee extension moments, medial patellofemoral ligament (MPFL) length-change, and tibiofemoral and patellofemoral kinematics were measured under physiological muscle loading. Data were analysed with one-dimensional statistical parametric mapping (Bonferroni-Holm corrected).

PFA changed knee function, altering extension moments (p<0.001) and patellofemoral kinematics (p<0.05), but not tibiofemoral kinematics. Patellar component positioning affected patellofemoral kinematics: over/understuffing influenced patellar anterior translation and the patellar tendon moment arm (p<0.001). Medially tilted patellar cuts produced lateral patellar tilt (p<0.001) and vice versa. A similar inverse effect occurred with extended/flexed patellar cuts, causing patellar flexion and extension (p<0.001), respectively. Of all variants, only extending the patellar cut produced near-native extension moments throughout. Conversely, the only femoral effect was MPFL length change between medially/laterally rotated components.

PFA can restore native knee biomechanics. Provided anterior femoral cuts are controlled and smooth condyle-to-component transition assured, patellar position affects biomechanics more than femoral, contradicting the hypothesis.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 9 - 9
1 Dec 2021
Edwards T Soussi D Gupta S Patel A Liddle A Khan S Cobb J Logishetty K
Full Access

Abstract

Objectives

Non-technical skills including teamwork play a pivotal role in surgical outcomes. Virtual reality is effective at improving technical skills, however there is a paucity of evidence on team-based virtual reality (VR) training. This study aimed to assess if multiplayer virtual reality training was superior to solo training for acquisition of both technical and non-technical skills in learning the complex anterior approach total hip arthroplasty operation.

Methods

10 novice surgeons and 10 novice scrub nurses, were randomised to solo or team virtual reality training to perform anterior approach total hip arthroplasty. Solo participants trained with virtual avatar counterparts, whilst teams trained in pairs (surgeon and scrub nurse). Both groups underwent 5 VR training sessions over 6 weeks. Then, they underwent a real-life assessment in which they performed AA-THA on a high-fidelity model with real equipment in a simulated operating theatre. Teams performed together and solo participants were randomly paired up with a solo player of the opposite role. Videos of the assessment were marked by two blinded expert assessors. Outcomes were procedure time, procedural errors from an expert pre-defined protocol and acetabular component positioning. Non-technical skills were assessed using the NOTECHs II and NOTSS scores.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 32 - 32
1 Dec 2021
Edwards T Khan S Patel A Gupta S Soussi D Liddle A Cobb J Logishetty K
Full Access

Abstract

Objectives

Evidence supporting the use of immersive virtual reality (iVR) training in orthopaedic procedures is rapidly growing. However, the impact of the timing of delivery of this training is yet to be tested. This study investigated whether spaced iVR training is more effective than massed iVR training for novices learning hip arthroplasty.

Methods

24 medical students with no hip arthroplasty experience were randomised to learning total hip arthroplasty using the same iVR simulation training either once-weekly or once-daily for four sessions. Participants underwent a baseline physical world assessment to orientate an acetabular component on a saw bone pelvis, and a baseline knowledge test. In iVR, we recorded procedural errors, time, numbers of prompts required and path lengths of the hands and head across 4 sessions. To assess skill retention, the iVR and baseline physical world assessments were repeated at one-week and one-month.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 54 - 54
1 Dec 2021
Maslivec A Ng KCG Cobb J
Full Access

Abstract

Objectives

Although hip replacement and resurfacing procedures both aim to restore mobility, improve joint function, and relieve pain, it is unclear how each differ in terms of gait mechanics and if they are affected by varying walking speeds. We compared limb symmetry and ground reaction force (GRF) profiles between bilateral total hip arthroplasty patients (THA), bilateral hip resurfacing arthroplasty patients (HRA), and healthy control participants (CON) during level-treadmill walking at different speeds.

Methods

Bilateral THA and bilateral HRA patients (nTHA = 15; nHRA = 15; postoperative 12–18 months), and age-, mass-, and height-matched CON participants (nCON = 20) underwent gait analysis on an instrumented treadmill. Walking trials started at 4 km/h and increased in 0.5 km/h increments until top walking speed (TWS) was achieved. Gait symmetry index (SI = 0% for symmetry) was assessed between limbs during weight-acceptance, mid-stance and push-off phases of gait; and vertical GRFs were captured for the normalised stance phase using statistical parametric mapping (SPM; CI = 95%).


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 75 - 75
1 Dec 2021
Stoddart J Garner A Tuncer M Cobb J van Arkel R
Full Access

Abstract

Objectives

There is renewed interest in bi-unicondylar arthroplasty (Bi-UKA) for patients with medial and lateral tibiofemoral osteoarthritis, but a spared patellofemoral compartment and functional cruciate ligaments. The bone island between the two tibial components may be at risk of tibial eminence avulsion fracture, compromising function. This finite element analysis compared intraoperative tibial strains for Bi-UKA to isolated medial unicompartmental arthroplasty (UKA-M) to assess the risk of avulsion.

Methods

A validated model of a large, high bone-quality tibia was prepared for both UKA-M and Bi-UKA. Load totalling 450N was distributed between the two ACL bundles, implant components and collateral ligaments based on experimental and intraoperative measurements with the knee extended and appropriately sized bearings used. 95th percentile maximum principal elastic strain was predicted in the proximal tibia. The effect of overcuts/positioning for the medial implant were studied; the magnitude of these variations was double the standard deviation associated with conventional technique.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 16 - 16
1 Dec 2021
Munford M Stoddart J Liddle A Cobb J Jeffers J
Full Access

Abstract

Objectives

Unicompartmental and total knee arthroplasty (UKA and TKA) are successful treatments for osteoarthritis, but monolithic implants disrupt the natural homeostasis of bone which leads to bone loss over time. This can cause problems if the implant needs to be revised. This study aimed to demonstrate that tibial implants made from titanium lattice could replace the tibial condyle surface while minimising disruption of the bone's natural mechanical loading environment. A secondary aim was to determine whether implants perform better if they replicate more closely bone's mechanical modulus, anisotropy and spatial heterogeneity. This study was conducted in a human cadaveric model.

Methods

In a cadaveric model, UKA and TKA procedures were performed on 8 fresh-frozen knee specimens by a board-certified consultant orthopaedic surgeon, using tibial implants made from conventional monolithic material and titanium lattice structures. Stress at the bone-implant interfaces was measured with pressure film and compared to the native knee.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 6 - 6
1 Nov 2021
Edwards T Maslivec A Ng G Woringer M Wiik A Cobb J
Full Access

Patients may be able to return to higher level activities following hip arthroplasty with modern techniques and prostheses, but the Oxford hip score, the standard PROM used by the NJS exhibits severe skew and kurtosis. The commonest score is 48/48. Most patients score above 40 preventing any discrimination between approaches or prostheses. We therefore sought both subjective and objective metrics which were relevant and valid without skew or high kurtosis in postoperative patients. The Metabolic Equivalent of Task (MET) reports energy usage in kcal/min burnt across a range of activities, condensed into a score of 0–25. A MET over 8 is considered ‘conditioning exercise’ tethered to life expectancy. A 2 point difference in average MET is considered a clinically relevant difference. Walking speed is a simple valid metric tethered to life expectancy, with a 0.1m/sec difference in walking speed equates to a clinically important difference.

Oxford Hip Score (OHS), and the MET were prospectively recorded in 221 primary hip arthroplasty procedures pre-operatively and at 1-year using a web based application. Pre and postoperative Gait analysis was undertaken on a subgroup of 34 patients, in comparison with age and sex matched controls.

Post-operatively, the OHS demonstrated significant skewed distributions with ceiling effects of 41% scoring 48/48. The MET was normally distributed around a mean of 10.3, with a standard deviation of 3.8 and no ceiling effect. Walking speed was normally distributed around a mean of 1.8m/sec, with a standard deviation was 0.15 m/sec

The MET is a simple patient reported score, which is normally distributed in patients following hip arthroplasty, around a mean of 10.3 with a standard deviation of 3.8. This valid activity metric correlates well with fast walking speed. This is also normally distributed with a standard deviation of over 0.1m/sec confirming low kurtosis. These simple measures have face validity: undertaking less active pastimes and being unable to keep up with other walkers are obviously inadvisable. The normal kurtosis of these metrics suggest that they may able to detect clinically relevant differences in outcome which are undetectable with commonly used PROMs. For surgeons developing less invasive approaches or using novel stems, these measures may detect clinically important improvements undetectable by the Oxford Hip Score.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_10 | Pages 27 - 27
1 Aug 2021
Edwards T Keane B Garner A Logishetty K Liddle A Cobb J
Full Access

This study investigates the use of the Metabolic Equivalent of Task (MET) score in a hip arthroplasty population and its ability to capture additional benefit beyond the maximum Oxford Hip Score (OHS).

OHS, EuroQol-5D index (EQ-5D), and the MET were prospectively recorded in 221 primary hip arthroplasty procedures pre-operatively and at 1-year. The distribution was examined reporting the presence of ceiling & floor effects. Validity was assessed correlating the MET with the other scores using Spearman's rank and determining responsiveness using the standardised response mean (SRM). A subgroup of 93 patients scoring 48/48 on the OHS were analysed by age group, sex, BMI and pre-operative MET using the other two metrics to determine if differences could be established despite all scoring identically on the OHS.

117 total hip and 104 hip resurfacing arthroplasty operations were included. Mean age was 59.4 ± 11.3. Post-operatively the OHS and EQ-5D demonstrate significant negatively skewed distributions with ceiling effects of 41% and 53%, respectively. The MET was normally distributed post-operatively with no ceiling effect. Weak-moderate but statistically significant correlations were found between the MET and the other two metrics both pre & post-operatively. Responsiveness was excellent, SRM for OHS: 2.01, EQ-5D: 1.06 and MET: 1.17. In the 48/48 scoring subgroup, no differences were found comparing groups with the EQ-5D, however significantly higher MET scores were demonstrated for patients aged <60 (12.7 vs 10.6, p=0.008), male patients (12.5 vs 10.8, p=0.024) and those with pre-operative MET scores >6 (12.6 vs 11.0, p=0.040).

The MET is normally distributed in patients following hip arthroplasty, recording levels of activity which are undetectable using the OHS. As a simple, valid activity metric, it should be considered in addition to conventional PROMs in order to capture the entire benefit experienced following hip arthroplasty.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 35 - 35
1 Mar 2021
Ng G Bankes M Daou HE Beaulé P Cobb J Jeffers J
Full Access

Abstract

OBJECTIVES

Although surgical periacetabular osteotomy (PAO) for hip dysplasia aims to optimise acetabular coverage and restore hip function, it is unclear how surgery affects capsular mechanics and joint stability. The purpose was to examine how the reoriented acetabular coverage affects capsular mechanics and joint stability in dysplastic hips.

METHODS

Twelve cadaveric dysplastic hips (n = 12) were denuded to the capsule and mounted onto a robotic tester. The robot positioned each hip in multiple flexion angles (Extension, Neutral 0°, Flexion 30°, Flexion 60°, Flexion 90°) and performed internal-external rotations and abduction-adduction to 5 Nm in each rotational or planar direction. Each hip underwent a PAO, preserving the capsule, and was retested postoperatively in the robot. Paired sample t-tests compared the range of motion before and after PAO surgery (CI = 95%).


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 55 - 55
1 Mar 2021
Dandridge O Garner A van Arkel R Amis A Cobb J
Full Access

Abstract

Objectives

The need for gender specific knee arthroplasty is debated. This research aimed to establish whether gender differences in patellar tendon moment arm (PTMA), a composite measure that characterises function of both the patellofemoral and tibiofemoral joints, are a consequence of knee size or other variation.

Methods

PTMA about the instantaneous helical axis was calculated from positional data acquired using optical tracking. First, data post-processing was optimised, comparing four smoothing techniques (raw, Butterworth filtered, generalised cross-validation cubic spline interpolated and combined filtered/interpolated) using a fabricated knee. Then PTMA was measured during open-chain extension for N=24 (11 female) fresh-frozen cadaveric knees, with physiologically based loading and extension rates (420°/s) applied. Gender differences in PTMA were assessed before and after accounting for knee size with epicondylar width.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 18 - 18
1 Mar 2021
Ng G Bankes M Grammatopoulos G Jeffers J Cobb J
Full Access

Abstract

OBJECTIVES

Cam femoroacetabular impingement (FAI – femoral head-neck deformity) and developmental dysplasia of the hip (DDH – insufficient acetabular coverage) constitute a large portion of adverse hip loading and early degeneration. Spinopelvic anatomy may play a role in hip stability thus we examined which anatomical relationships can best predict range of motion (ROM).

METHODS

Twenty-four cadaveric hips with cam FAI or DDH (12:12) were CT imaged and measured for multiple femoral (alpha angles, head-neck offset, neck angles, version), acetabular (centre-edge angle, inclination, version), and spinopelvic features (pelvic incidence). The hips were denuded to the capsule and mounted onto a robotic tester. The robot positioned each hip in multiple flexion angles (Extension, Neutral 0°, Flexion 30°, Flexion 60°, Flexion 90°); and performed internal-external rotations to 5 Nm in each position. Independent t-tests compared the anatomical parameters and ROM between FAI and DDH (CI = 95%). Multiple linear regressions determined which anatomical parameters could predict ROM.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 27 - 27
1 Mar 2021
Dandridge O Garner A van Arkel R Amis A Cobb J
Full Access

Abstract

Objectives

Unicompartmental (UKA) and bicompartmental (BCA) knee arthroplasty are associated with improved functional outcomes compared to Total Knee Arthroplasty (TKA) in suitable patients, although the reason is poorly understood. The aim of this study was to measure how the different arthroplasties affect knee extensor function.

Methods

Extensor function was measured for sixteen cadaveric knees and then re-tested following the different arthroplasties. Eight knees underwent medial UKA then BCA, then posterior-cruciate retaining TKA, and eight underwent the lateral equivalents then TKA. Extensor efficiency was calculated for ranges of knee flexion associated with common activities of daily living. Data were analyzed with repeated measures analysis of variance (α=0.05).


Bone & Joint Open
Vol. 2, Issue 2 | Pages 134 - 140
24 Feb 2021
Logishetty K Edwards TC Subbiah Ponniah H Ahmed M Liddle AD Cobb J Clark C

Aims

Restarting planned surgery during the COVID-19 pandemic is a clinical and societal priority, but it is unknown whether it can be done safely and include high-risk or complex cases. We developed a Surgical Prioritization and Allocation Guide (SPAG). Here, we validate its effectiveness and safety in COVID-free sites.

Methods

A multidisciplinary surgical prioritization committee developed the SPAG, incorporating procedural urgency, shared decision-making, patient safety, and biopsychosocial factors; and applied it to 1,142 adult patients awaiting orthopaedic surgery. Patients were stratified into four priority groups and underwent surgery at three COVID-free sites, including one with access to a high dependency unit (HDU) or intensive care unit (ICU) and specialist resources. Safety was assessed by the number of patients requiring inpatient postoperative HDU/ICU admission, contracting COVID-19 within 14 days postoperatively, and mortality within 30 days postoperatively.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 21 - 21
1 Feb 2021
Logishetty K Edwards T Liddle A Dean E Cobb J Clark C
Full Access

Background

In the United Kingdom, over 1 million elective surgeries were cancelled due to COVID-19, resulting in over 1.9 million people now waiting more than 4 months for their procedure – 3x the number last year. To address this backlog, the healthcare service has been asked to develop locally-designed ‘COVID-light’ facilities. In our local system, 822 patients awaited orthopaedic surgery when elective surgery was permitted to resume. The phased return of service required a careful and pragmatic prioritisation of patients, to protect resources, patients, and healthcare workers.

Aims

We aim to describe how the COVID-19 Algorithm for Resuming Elective Surgery (CARES) was used to consider 1) Which type of operation and patient should be prioritised? and 2) Which patients are safe to undergo surgery? The central tenets to this were patient safety, predicted efficacy of the surgery, and delivering compassionate care by considering biopsychosocial factors.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 54 - 54
1 Feb 2021
Dandridge O Garner A Amis A Cobb J van Arkel R
Full Access

As treatments of knee osteoarthrosis are continually refined, increasingly sophisticated methods of evaluating their biomechanical function are required. Whilst TKA shows good preoperative pain relief and survivorship, functional outcomes are sub-optimal, and research focus has shifted towards their improvement. Restoration of physiological function is a common design goal that relies on clear, detailed descriptions of native biomechanics. Historical simplifications of true biomechanisms, for example sagittal plane approximation of knee kinematics, are becoming progressively less suitable for evaluation of new technologies. The patellar tendon moment arm (PTMA) is an example of such a metric of knee function that usefully informs design of knee arthroplasty but is not fully understood, in part due to limitations in its measurement. This research optimized PTMA measurement and identified the influence of knee size and sex on its variation.

The PTMA about the instantaneous helical axis was calculated from optical tracked positional data. A fabricated knee model facilitated calculation optimization, comparing four data smoothing techniques (raw, Butterworth filtering, generalized cross-validated cubic spline-interpolation and combined filtering/interpolation). The PTMA was then measured for 24 fresh-frozen cadaveric knees, under physiologically based loading and extension rates. Sex differences in PTMA were assessed before and after size scaling.

Large errors were measured for raw and interpolated-only techniques in the mid-range of extension, whilst both raw and filtered-only methods saw large inaccuracies at terminal extension and flexion. Combined filtering/interpolation enabled sub-mm PTMA calculation accuracy throughout the range of knee flexion, including at terminal extension/flexion (root-mean-squared error 0.2mm, max error 0.5mm) (Figure 1).

Before scaling, mean PTMA throughout flexion was 46mm; mean, peak, and minimum PTMA values were larger in males, as was the PTMA at terminal flexion, the change in PTMA from terminal flexion to peak, and the change from peak to terminal extension (mean differences ranging from 5 to 10mm, p<0.05). Knee size was highly correlated with PTMA magnitude (r>0.8, p<0.001) (Figure 2). Scaling eliminated sex differences in PTMA magnitude, but peak PTMA occurred closer to terminal extension in females (female 15°, male 29°, p=0.01) (Figure 3).

Improved measurement of the PTMA reveals previously undocumented characteristics that may help to improve the functional outcomes of knee arthroplasty. Knee size accounted for two-thirds of the variation in PTMA magnitude, but not the flexion angle at which peak PTMA occurred, which has implications for morphotype-specific arthroplasty and musculoskeletal models. The developed calculation framework is applicable both in vivo and vitro for accurate PTMA measurement and might be used to evaluate the relative performance of emerging technologies.

For any figures or tables, please contact the authors directly.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 44 - 44
1 Feb 2021
Edwards T Patel A Szyszka B Coombs A Kucheria R Cobb J Logishetty K
Full Access

Background

Revision total knee arthroplasty (rTKA) is a high stakes procedure with complex equipment and multiple steps. For rTKA using the ATTUNE system revising femoral and tibial components with sleeves and stems, there are over 240 pieces of equipment that require correct assembly at the appropriate time. Due to changing teams, work rotas, and the infrequency of rTKR, scrub nurses may encounter these operations infrequently and often rely heavily on company representatives to guide them. In turn, this delays and interrupts surgical efficiency and can result in error. This study investigates the impact of a fully immersive virtual reality (VR) curriculum on training scrub nurses in technical skills and knowledge of performing a complex rTKA, to improve efficiency and reduce error.

Method

Ten orthopaedic scrub nurses were recruited and trained in four VR sessions over a 4-week period. Each VR session involved a guided mode, where participants were taught the steps of rTKA surgery by the simulator in a simulated operating theatre. The latter 3 sessions involved a guided mode followed by an unguided VR assessment. Outcome measures in the unguided assessment were related to procedural sequence, duration of surgery and efficiency of movement. Transfer of skills was assessed during a pre-training and post-training assessment, where participants completed multi-step instrument selection and assembly using the real equipment. A pre and post-training questionnaire assessed the participants knowledge, confidence and anxiety.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 53 - 53
1 Feb 2021
Garner A Dandridge O Amis A Cobb J van Arkel R
Full Access

Combined Partial Knee Arthroplasty (CPKA) is a promising alternative to Total Knee Arthroplasty (TKA) for the treatment of multi-compartment arthrosis. Through the simultaneous or staged implantation of multiple Partial Knee Arthroplasties (PKAs), CPKA aims to restore near-normal function of the knee, through retention of the anterior cruciate ligament and native disease-free compartment. Whilst PKA is well established, CPKA is comparatively novel and associated biomechanics are less well understood.

Clinically, PKA and CPKA have been shown to better restore knee function compared to TKA, particularly during fast walking. The biomechanical explanation for this superiority remains unclear but may be due to better preservation of the extensor mechanism. This study sought to assess and compare extensor function after PKA, CPKA, and TKA.

An instrumented knee extension rig facilitated the measurement extension moment of twenty-four cadaveric knees, which were measured in the native state and then following a sequence of arthroplasty procedures. Eight knees underwent medial Unicompartmental Knee Arthroplasty (UKA-M), followed by patellofemoral arthroplasty (PFA) thereby converting to medial Bicompartmental Knee Arthroplasty (BCA-M). In the final round of testing the PKA implants were removed a posterior-cruciate retaining TKA was implanted. The second eight received lateral equivalents (UKA-L then BCA-L) then TKA. The final eight underwent simultaneous Bi-Unicondylar Arthroplasty (Bi-UKA) before TKA. Extensor efficiencies over extension ranges typical of daily tasks were also calculated and differences between arthroplasties were assessed using repeated measures analysis of variance.

For both the medial and lateral groups, UKA demonstrated the same extensor function as the native knee. BCA resulted in a small reduction in extensor moment between 70–90° flexion but, in the context of daily activity, extensor efficiency was largely unaffected and no significant reductions were found. TKA, however, resulted in significantly reduced extensor moments, leading to efficiency deficits ranging from 8% to 43% in flexion ranges associated with downhill walking and the stance phase of gait, respectively.

Comparing the arthroplasties: TKA was significantly less efficient than both UKA-M and BCA-M over ranges representing stair ascent and gait; TKA showed a significant 23% reduction compared to BCA-L in the same range. There were no differences in efficiency between the UKAs and BCAs over any flexion range and TKA efficiency was consistently lower than all other arthroplasties.

Bi-UKA generated the same extensor moment as native knee at flexion angles typical of fast gait (0–30°). Again, TKA displayed significantly reduced extensor moments towards full extension but returned to the normal range in deep flexion. Overall, TKA was significantly less efficient following TKA than Bi-UKA.

Recipients of PKA and CPKA have superior functional outcomes compared to TKA, particularly in relation to fast walking. This in vitro study found that both UKA and CPKA better preserve extensor function compared to TKA, especially when evaluated in the context of daily functional tasks. TKA reduced knee extensor efficiency by over 40% at flexion angles associated with gait, arguably the most important activity to maintain patient satisfaction. These findings go some way to explaining functional deficiencies of TKA compared to CPKA observed clinically.


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 922 - 928
1 Aug 2019
Garner A van Arkel RJ Cobb J

Aims

There has been a recent resurgence in interest in combined partial knee arthroplasty (PKA) as an alternative to total knee arthroplasty (TKA). The varied terminology used to describe these procedures leads to confusion and ambiguity in communication between surgeons, allied health professionals, and patients. A standardized classification system is required for patient safety, accurate clinical record-keeping, clear communication, correct coding for appropriate remuneration, and joint registry data collection.

Materials and Methods

An advanced PubMed search was conducted, using medical subject headings (MeSH) to identify terms and abbreviations used to describe knee arthroplasty procedures. The search related to TKA, unicompartmental (UKA), patellofemoral (PFA), and combined PKA procedures. Surveys were conducted of orthopaedic surgeons, trainees, and biomechanical engineers, who were asked which of the descriptive terms and abbreviations identified from the literature search they found most intuitive and appropriate to describe each procedure. The results were used to determine a popular consensus.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 18 - 18
1 May 2019
Logishetty K Rudran B Gofton W Beaule P Field R Cobb J
Full Access

Background

For total hip arthroplasty (THA), cognitive training prior to performing real surgery may be an effective adjunct alongside simulation to shorten the learning curve. This study sought to create a cognitive training tool to perform direct anterior approach THA, validated by expert surgeons; and test its use as a training tool compared to conventional material.

Methods

We employed a modified Delphi method with four expert surgeons from three international centres of excellence. Surgeons were independently observed performing THA before undergoing semi-structured cognitive task analysis (CTA) before completing successive rounds of electronic surveys until consensus. The agreed CTA was incorporated into a mobile and web-based platform. Forty surgical trainees (CT1-ST4) were randomised to CTA-training or a digital op-tech with surgical videos, before performing a simulated DAA THA in a validated fully-immersive virtual reality simulator.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 27 - 27
1 May 2019
Logishetty K Rudran B Gofton W Beaule P Cobb J
Full Access

Background

Virtual Reality (VR) uses headsets and motion-tracked controllers so surgeons can perform simulated total hip arthroplasty (THA) in a fully-immersive, interactive 3D operating theatre. The aim of this study was to investigate the effect of laboratory-based VR training on the ability of surgical trainees to perform direct anterior approach THA on cadavers.

Methods

Eighteen surgical trainees (CT1-ST4) with no prior experience of direct anterior approach (DAA) THA completed an intensive 1-day course (lectures, dry-bone workshops and technique demonstrations). They were randomised to either a 5-week protocol of VR simulator training or conventional preparation (operation manuals and observation of real surgery). Trainees performed DAA-THA on cadaveric hips, assisted by a passive scrub nurse and surgical assistant. Performance was measured on the Intercollegiate Surgical Curriculum Project (ISCP) procedure-based assessment (PBA), on a 9-point global summary score (Table 1). This was independently assessed by 2 hip surgeons blinded to group allocation. The secondary outcome measure was error in cup orientation from a predefined target (40° inclination and 20° anteversion).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 124 - 124
1 Apr 2019
Karia M Ali A Harris S Abel R Cobb J
Full Access

Background

Defining optimal coronal alignment in Total Knee Replacement (TKR) is a controversial and poorly understood subject. Tibial bone density may affect implant stability and functional outcomes following TKR. Our aim was to compare the bone density profile at the implant-tibia interface following TKR in mechanical versus kinematic alignment.

Methods

Pre-operative CT scans for 10 patients undergoing medial unicompartmental knee arthroplasty were obtained. Using surgical planning software, tibial cuts were made for TKR with 7 degrees posterior slope and either neutral (mechanical) or 3 degrees varus (kinematic) alignment. Signal intensity, in Hounsfield Units (HU), was measured at 25,600 points throughout an axial slice at the implant-tibia interface and density profiles compared along defined radial axes from the centre of the tibia towards the cortices (Hotelling's t-squared and paired t-test).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 108 - 108
1 Apr 2019
Riviere C Maillot C Auvinet E Cobb J
Full Access

Introduction

The objective of our study was to determine the extent to which the quality of the biomechanical reconstruction when performing hip replacement influences gait performances. We aimed to answer the following questions: 1) Does the quality of restoration of hip biomechanics after conventional THR influence gait outcomes? (question 1), and 2) Is HR more beneficial to gait outcomes when compared with THR? (question 2).

Methods

we retrospectively reviewed 52 satisfied unilateral prosthetic hip patients (40 THRs and 12 HRs) who undertook objective gait assessment at a mean follow-up of 14 months. The quality of the prosthetic hip biomechanical restoration was assessed on standing pelvic radiograph by comparison to the healthy contralateral hip.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 123 - 123
1 Apr 2019
Karia M Vishnu-Mohan S Boughton O Auvinet E Wozencroft R Clarke S Halewood C Cobb J
Full Access

Aims

Accurate and precise acetabular reaming is a requirement for the press-fit stability of cementless acetabular hip replacement components. The accuracy of reaming depends on the reamer, the reaming technique and the bone quality. Conventional reamers wear with use resulting in inaccurate reaming diameters, whilst the theoretical beneficial effect of ‘whirlwind’ reaming over straight reaming has not previously been documented. Our aim was to compare the accuracy and precision of single use additively-manufactured reamers with new conventional reamers and to compare the effect of different acetabular reaming techniques.

Materials and Methods

Forty composite bone models, half high-density and half low-density, were reamed with a new 61 mm conventional acetabular reamer using either straight or ‘whirlwind’ reaming techniques. This was repeated with a 61 mm single use additively-manufactured reamer. Reamed cavities were scanned using a 3D laser scanner with mean diameters of reamed cavities compared using the Mann-Whitney U test to determine any statistically significant differences between groups (p<0.05) [Fig. 1).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 75 - 75
1 Apr 2019
Boughton O Uemura K Tamura K Takao M Hamada H Cobb J Sugano N
Full Access

Objectives

For patients with Developmental Dysplasia of the Hip (DDH) who progress to needing total joint arthroplasty it is important to understand the morphology of the femur when planning for and undertaking the surgery, as the surgery is often technically more challenging in patients with DDH on both the femoral and acetabular parts of the procedure1. The largest number of male DDH patients with degenerative joint disease previously assessed in a morphological study was 122. In this computed tomography (CT) based morphological study we aimed to assess whether there were any differences in femoral morphology between male and female patients with developmental dysplasia undergoing total hip arthroplasty (THA) in a cohort of 49 male patients, matched to 49 female patients.

Methods

This was a retrospective study of the pre-operative CT scans of all male patients with DDH who underwent THA at two hospitals in Japan between 2006–2017. Propensity score matching was used to match these patients with female patients in our database who had undergone THA during the same period, resulting in 49 male and 49 female patients being matched on age and Crowe classification. The femoral length, anteversion, neck-shaft angle, offset, canal-calcar ratio, canal flare index, lateral centre-edge angle, alpha angle and pelvic incidence were measured for each patient on their pre-operative CT scans.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_2 | Pages 26 - 26
1 Jan 2019
Choudhury A Auvient E Iranpour F Lambkin R Wiik A Hing C Cobb J
Full Access

Patellofemoral osteoarthritis (PFOA) affects 32% men and 36% women over the age of 60years and is associated with anterior knee pain, stiffness, and poor mobility. Patellofemoral arthroplasty (PFA) is a bone-sparing treatment for isolated PFOA. This study set out to investigate the relationship between patient-related outcome measures (PROMs) and measurements obtained from gait analysis before and after PFA. There are currently no studies relating to gait analysis and PFA available in the literature

A prospective cohort study was conducted of ten patients known to have isolated PFOA who had undergone PFA compared to a gender and age matched control group. The patients were also asked to complete questionnaires (Oxford knee score (OKS), EQ-5D-5L) before surgery and one year after surgery. Gait analysis was done on an instrumented treadmill comparing Ground reaction force parameters between the control and pre and post-operative PFA patients

The average age 60 (49–69) years with a female to male ratio of 9:1. Patient and healthy subjects were matched for age and gender, with no significant difference in BMI. Post-op PFA improvement in gait seen in ground reaction force at 6.5km/h. Base support difference was statistically significant both on the flat P=0.0001 and uphill P=0.429 (5% inclination) and P=0.0062 (10% inclination). PROMS response rate was 70%(7/10) pre-operative and 60%(6/10) post-operative. EQ-5D-5L scores reflected patient health state was better post-operatively.

This study found that gait analysis provides an objective measure of functional gait and reflected by significant quality-of-life improvement of patients post PFA. Literature lacks studies relating to gait-analysis and PFA. Valuable information provided by this study highlights that PFA has a beneficial outcome reflected by PROMs and improvement in vertical ground reaction force and gait

Further research is needed to assess how care-providers may use gait-analysis as part of patient care plans for PFOA patients.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_2 | Pages 2 - 2
1 Jan 2019
Boughton O Yan L Ma S Hansen U Giuliani F Cobb J Marrow J Abel R
Full Access

With information about a patient's bone mechanical properties orthopaedic operations could be optimised to reduce intra- and post-operative complications. However, there is currently no reliable method of measuring a patient's bone mechanical properties in vivo. We have previously investigated microindentation, using a 1.5mm diameter spherical indenter tip, and found no correlation between these measurements and compression testing measurements. We hypothesised that by using a larger diameter indenter tip we would closer match bone millimetre-scale mechanical properties.

20 bone samples were taken from 20 patients undergoing hip replacement surgery. The samples were machined from the femoral neck calcar cortical bone into 6×3×3mm parallelepiped specimens, aligned with the osteons along the long axis. The samples were micro-computed tomography (CT) scanned to calculate porosity. Microindentation was performed using a 6mm diameter, sapphire, spherical indenter tip. 12 indentations were performed in a grid and the reduced moduli were calculated using the Oliver-Pharr method. Compression testing was then performed to failure and the apparent elastic modulus was calculated for each sample.

A moderate correlation was found between the indentation reduced moduli and compression testing elastic moduli (r=0.52, r2=0.275, p=0.018). In addition, a moderate correlation was found between the indentation reduced moduli and CT-measured porosity (r=0.5, r2=0.251, p=0.025) and a strong correlation was found between compression testing moduli and porosity (r=0.75, r2=0.568, p<0.001).

Using large-tip spherical microindentation, indentation reduced moduli correlated significantly with compression testing apparent elastic moduli in these 20 cortical bone specimens. Microindentation using a large, spherical indenter tip may predict the mechanical properties of bone at the millimetre length scale and shows promise as a potential future clinical decision aid in surgery.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_2 | Pages 1 - 1
1 Jan 2019
Logishetty K Van Arkel R Muirhead-Allwood S Ng G Cobb J Jeffers J
Full Access

The hip's capsular ligaments (CL) passively restrain extreme range of motion (ROM) by wrapping around the native femoral head/neck, and protect against impingement and instability. We compared how CL function was affected by device (hip resurfacing arthroplasty, HRA; dual mobility total hip arthroplasty, DM-THA; and conventional THA, C-THA), and surgical approach (anterior and posterior), with and without CL surgical-repair. We hypothesized that CL function would only be preserved when native head-size (HRA/DM-THA) was restored.

CL function was quantified on sixteen cadaveric hips, by measuring ROM by internally (IR) and externally rotating (ER) the hip in six functional positions, ranging from full extension with abduction to full flexion with adduction (squatting). Native ROM was compared to ROM after posterior capsulotomy (right hips) or anterior capsulotomy (left hips), and HRA, and C-THA and DM-THA, before and after CL repair.

Independent of approach, ROM increased most following C-THA (max 62°), then DM-THA (max 40°), then HRA (max 19°), indicating later CL engagement and reduced biomechanical function with smaller head-size. Dislocations also occurred in squatting after C-THA and DM-THA. CL-repair following HRA restored ROM to the native hip (max 8°). CL-repair following DM-THA reduced ROM hypermobility in flexed positions only and prevented dislocation (max 36°). CL-repair following C-THA did not reduce ROM or prevent dislocation.

For HRA and repair, native anatomy was preserved and ligament function was restored. For DM-THA with repair, ligament function depended on the movement of the mobile-bearing, with increased ROM in positions when ligaments could not wrap around head/neck. For C-THA, the reduced head-size resulted in inferior capsular mechanics in all positions as the ligaments remained slack, irrespective of repair.

Choosing devices with anatomic head-sizes (HRA/DM-THA) with capsular repair may have greater effect than surgical approach to protect against instability in the early postoperative period.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_16 | Pages 53 - 53
1 Nov 2018
Karia M Ali A Harris S Abel R Cobb J
Full Access

Tibial bone density may affect implant stability and functional outcomes following total knee replacement (TKR). Our aim was to characterise the bone density profile at the implant-tibia interface following TKR in mechanical versus kinematic alignment. Pre-operative computed tomography scans for 10 patients were obtained. Using surgical planning software, tibial cuts were made for TKR either neutral (mechanical) or 3 degrees varus (kinematic) alignment. Signal intensity, in Hounsfield Units (HU), was measured at 25,600 points throughout an axial slice at the implant-tibia interface and density profiles compared along defined radial axes from the centre of the tibia towards the cortices. From the tibial centre towards the lateral cortex, trabecular bone density for kinematic and mechanical TKR are similar in the inner 50% but differ significantly beyond this (p= 0.012). There were two distinct density peaks, with peak trabecular bone density being higher in kinematic TKR (p<0.001) and peak cortical bone density being higher in mechanical TKR (p<0.01). The difference in peak cortical to peak trabecular signal was 43 HU and 185 HU respectively (p<0.001). On the medial side there was no significant difference in density profile and a linear increase from centre to cortex. In the lateral proximal tibia, peak cortical and peak trabecular bone densities differ between kinematic TKR and mechanical TKR. Laterally, mechanical TKR may be more dependent upon cortical bone for support compared to kinematic TKR, where trabecular bone density is higher. This may have implications for surgical planning and implant design.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 4 - 4
1 Aug 2018
Cobb J Clarke S Halewood C Wozencroft R Jeffers J Logishetty K Keane B Johal H
Full Access

We aimed to demonstrate the clinical safety of a novel anatomic cementless ceramic hip resurfacing device. Concerns around the safety of metal on metal arthroplasty have made resurfacing less attractive, while long term function continues to make the concept appealing. Biolox Delta ceramic is now used in more than 50% of all hip arthroplasties, suggesting that it's safety profile is acceptable. We wondered if a combination of these concepts might work?

The preclinical testing of anatomic hip resurfacing device developed by our group was presented last year. A twenty patient safety study was designed. Patients had to be between the ages of 18 and 70. The initial size range was restricted to femoral heads between 46 and 54, representing the common sizes of hip resurfacing. The primary outcomes were clinical safety, PROMs and radiological control. Secondary outcomes include CTRSA and metal ion levels.

20 patients were recruited, aged 30–69. 7 were women and 13 were men. There were no operative adverse events in their operations undertaken between September 2017 and February 2018. One patient had a short episode of atrial fibrillation on the second postoperative day, and no other complications. At three months the median oxford hip score had risen from 27 (range 14–38) to 46 (31–48). Cobalt and chromium levels were almost undetectable at 3 months. Fixation appeared satisfactory in all patients, with no migration detected in either component. CTRSA is in process.

The initial safety of a novel cementless ceramic resurfacing device is demonstrated by this data. The 10 year, 250 case efficacy study will continue in 5 other European centres.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 11 - 11
1 Aug 2018
Muirhead-Allwood S Logishetty K van Arkel R Ng G Cobb J Jeffers J
Full Access

The hip joint capsular ligaments (CL) passively restrain extreme range of motion (ROM) by wrapping around the native femoral head, and protect against impingement, edge loading wear and dislocation. This study compared how ligament function was affected by device (hip resurfacing arthroplasty, HRA; dual mobility total hip arthroplasty, DM-THA; and conventional THA, C-THA), with and without CL repair. It was hypothesized that ligament function would only be preserved when native anatomy was preserved: with restoration of head-size (HRA or DM-THA) and repair.

Eight normal male cadaveric hips were skeletonised, retaining the hip capsule. CL function was quantified by measuring ROM by internally (IR) and externally rotating (ER) the hip in six functional positions, ranging from full extension with abduction to full flexion with adduction (squatting). Native ROM was compared to ROM after posterior capsulotomy and HRA, and C-THA and DM-THA, before and after surgical CL repair.

ROM increased most following C-THA (max 62°), then DM-THA (max 40°), then HRA (max 19°), indicating later engagement of the capsule and reduced biomechanical function with smaller head-size. Dislocations also occurred in squatting after C-THA and DM-THA. CL-repair following HRA restored ROM to the native hip (max 8°). CL-repair following DM-THA reduced ROM hypermobility in flexed positions only and prevented dislocation (max 36°). CL-repair following C-THA did not reduce ROM or prevent dislocation.

When HRA was combined with repair, native anatomy was preserved and ligament function was restored. For DM-THA with repair, ligament function depended on the movement of the mobile bearing resulting in near-native function in some positions, but increased ROM when ligaments were unable to wrap around the head/neck. Following C-THA, the reduced head-size resulted in inferior capsular mechanics in all positions as the ligaments remained slack, irrespective of repair.

Choosing devices with anatomic head-sizes (resurfacing or dual-mobility) and repairing the capsular ligaments may protect against instability in the early postoperative period.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_4 | Pages 14 - 14
1 Apr 2018
Van Der Straeten C Abdulhussein D Brevadt M Cobb J
Full Access

Background

Hip resurfacing arthroplasty (HRA) and total hip arthroplasty (THA) are treatments of end-stage hip disease. Gait analysis studies comparing HRA and THA have demonstrated HRA results in a more normal gait than THA. The reasons may include the larger, more anatomic head diameter, the preservation of the femoral neck with restoration of the anatomical hip centre position and normal proprioception. This study investigated (1)whether femoral head size diameter affects gait; (2)whether gait still differs between THA and HRA patients even with comparable head diameters.

Methods

We analysed the gait of 33 controls and 50 patients with unilateral hip replacement. Follow-up ranged from 9–68 months. In 27 hips a small femoral head size was used (≤ 36mm); in 23 hips a large head size (>36mm). The small size group consisted of 11 long femoral stem THA and 16 short-stem THA; the large group of 5 long-stem, 8 short-stem THA and 10 HRA patients. There were 14 females/19 males in the control group; 22 females/5 males in the small size group; 13 females/10 males in the large size group.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 4 - 4
1 Apr 2018
Western L Logishetty K Morgan R Cobb J Auvinet E
Full Access

Background

Accurate implant orientation is associated with improved outcomes after artificial joint replacement. We investigated if a novel augmented-reality (AR) platform (with live feedback) could train novice surgeons to orientate an acetabular implant as effectively as conventional training (CT).

Methods

Twenty-four novice surgeons (pre-registration level medical students) voluntarily participated in this trial. Baseline demographics, data on exposure to hip arthroplasty, and baseline performance in orientating an acetabular implant to six patient-specific values on a phantom pelvis, were collected prior to training. Participants were randomised to a training session either using a novel AR headset platform or receiving one-on-one tuition from a hip surgeon (CT). After training, they were asked to perform the six orientation tasks again. The solid-angle error in degrees between the planned and achieved orientations was measured using a head-mounted navigation system.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 32 - 32
1 Apr 2018
Van Der Straeten C Abdulhussein D Brevadt MJ Cobb J
Full Access

Background

Hip resurfacing arthroplasty (HRA) and total hip arthroplasty (THA) are treatments of end-stage hip disease. Gait analysis studies comparing HRA and THA have demonstrated that HRA results in a more normal gait than THA. The reasons may include the larger, more anatomic head diameter or the preservation of the neck of the femur with restoration of the anatomical position of the hip centre and normal proprioception. This study investigated (1) whether femoral head size diameter affects gait; (2) whether gait still differs between THA and HRA patients even with comparable head diameters.

Methods

We retrospectively analysed the gait of 33 controls and 50 patients with a unilateral hip replacement, operated by the same surgeon. Follow-up ranged from 9–68 months. In 27 hips a small femoral head size was used (≤ 36mm); in 23 hips a large head size (>36mm). The small size group consisted of 11 long femoral stem THA and 16 short-stem THA and the large group of 5 long-stem THA, 8 short-stem THA and 10 HRA patients. There were 14 females/19 males in the control group; 22 females/5 males in the small size group; 13 females/10 males in the large size group.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 42 - 42
1 Apr 2018
Western L Logishetty K Morgan R Cobb J Auvinet E
Full Access

Background

Complications such as dislocations, impingement and early wear following total hip arthroplasty (THA) increase with acetabular cup implant malorientation. These errors are more common with low-volume centres or in novice hands. Currently, this skill is most commonly taught during real surgery with an expert trainer, but simulated training may offer a safer and more accessible solution. This study investigated if a novel MicronTracker® enhanced Microsoft HoloLens® augmented reality (EAR) headset was as effective as one-on-one expert surgeon (ES) training for teaching novice surgeons hip cup orientation skill.

Methods

Twenty-four medical students were randomly assigned to EAR or ES training groups. Participants used a modified sawbone/foam pelvis model for hip cup orientation simulation. A validated EAR headset measured the orientation of acetabular cup implants and displayed this in the participant”s field of view. The system calculated the difference between planned and achieved orientation as a solid-angle error.

Six different inclination and anteversion combinations, related to hypothetical patient-specific anatomy, were used as target orientations. Learning curves were measured over four sessions, each one week apart. Error in orientations of non-taught angles and during a concealed pelvic tilt were measured to assess translation of skills. A post-test questionnaire was used for qualitative analysis of procedure understanding and participant experience.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_4 | Pages 18 - 18
1 Apr 2018
Farrokhnik R Wiik A Brevadt M Lambkin R Cobb J
Full Access

The use of hip resurfacing arthroplasty (HRA) has largely regressed due to the fear of metal-on-metal bearings. However committed HRA users continue to assert the functional advantages that a geometry retaining implant would have on a patient”s hip. Currently worldwide, HRA is only recommended to men who demand an active lifestyle. Despite this precarious indication, it is not clear to what extent HRA has on higher activity function. The aim of this study was to determine the functional extent to which could be achieved with HRA. The primary objective is to assess the loading pattern change for patients implanted with HRA at high walking speeds and inclinations. The second objective is to compare their loading features to a healthy group to determine if a normal gait pattern could be achieved.

Between 2012 and 2016, a total of 28 prospective unilateral HRA patients were analysed on an instrumented treadmill from a single centre. All 28 patient patients had a uniform implant type and had no other lower limb operations or disease. Perioperative plain orthogonal radiographs were used to measure hip length and global hip offset change. A healthy control group (n=35) were analysed to compare. All HRA patients gait characteristics were assessed at incrementally higher speeds and inclinations to determine the extent of improvement HRA has on a challenging activity. A Student t-test along with a multivariate analysis was done with significance set at α=0.05. Weight and height variance was accounted with Hof normalisation.

The HRA and control group were reasonably matched for age (57 vs 55yrs), BMI (27 vs 25) and height (175 vs 170cm) respectively. Hip measurements revealed less than 5mm change for all cases. The mean time from initial preoperative gait assessment to postoperative assessment was 30 months (24–48months). The mean top walking speed for controls was 1.97m/s and postoperatively 2.1 m/sec for the HRA group. The significant (p<0.001) loading change during flat walking can be seen with restoration of symmetry. Walking at an inclination demonstrated a marked change during weight acceptance (p<0.001) and a loading pattern returning to near normal.

This prospective study found HRA patients walking faster than age matched controls. They demonstrated a significant change in their loading pattern, by significantly shifting load from the unaffected side to the implanted side. Uphill walking, an activity which requires more hip flexion, demonstrated a change in stance phase which was near normal. This small comparative study confirms near physiological function can be achieved with HRA at higher activity levels.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_4 | Pages 49 - 49
1 Apr 2018
Morgan R Logishetty K Western L Cobb J Auvinet E
Full Access

Background

Trust in the validity of a measurement tool is critical to its function in both clinical and educational settings. Acetabular cup malposition within total hip arthroplasty (THA) can lead to increased dislocation rates, impingement and increased wear as a result of edge loading. We have developed a THA simulator incorporating a foam/Sawbone pelvis model with a modified Microsoft HoloLens® augmented reality (AR) headset. We aimed to measure the trueness, precision, reliability and reproducibility of this platform for translating spatial measurements of acetabular cup orientation to angular values before developing it as a training tool.

Methods

A MicronTracker® stereoscopic camera was integrated onto a HoloLens® AR system. Trueness and precision values were obtained through comparison of the AR system measurements to a gold-standard motion capture system”s (OptiTrack®) measurements for acetabular cup orientation on a benchtop trainer, in six clinically relevant pairs of anteversion and inclination angles. Four surgeons performed these six orientations, and repeated each orientation twice. Pearson”s coefficients and Bland-Altman plots were computed to assess correlation and agreement between the AR and Motion Capture systems. Intraclass correlation coefficients (ICC) were calculated to evaluate the degree of repeatability and reproducibility of the AR system by comparing repeated tasks and between surgeons, respectively.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 29 - 29
1 Jan 2018
Cobb J Clarke S Jeffers J Wozencroft R Halewood C Amis A
Full Access

Hip resurfacing remains a safe and effective option according to registry data. Results in women were less reliable, in part owing to soft tissue impingement. Biolox Delta ceramic bearing couples are now in widespread use with very low complication rates. We set about merging these three elements to develop a novel hip resurfacing arthroplasty.

Contours of both acetabular and femoral components were generated from biometric data, adapted to the constraints of ceramic machining, to ensure that radii blended from the bearing surface avoiding any sharp boundaries. Plasma spray coating with titanium and hydroxyapatite direct onto ceramic was developed and tested using shear, tensile and taber abrasion testing. Wear testing was carried out to 5 million cycles according to the ASTM. Destructive testing was carried out in a variety of test conditions and angles.

Cadaveric testing demonstrated stability using a single use disposable instruments for both conventional and patient specific procedures. Very low dose CT enabled the entire interface to be observed as the Ceramic is radiolucent, enhancing migration analysis, which will be undertaken at 4 intervals to confirm stability. Functional scores and gait analysis will be used in the safety study.

The CE study recruitment is underway, with first in human trials starting in summer 2017. PMA submission will follow the safety study. Commercial release of the device in Europe is unlikely before 2019, and in the USA may not be until 2027. The path to novel device development in 2017 is very costly in time and money.


Bone & Joint Research
Vol. 6, Issue 10 | Pages 602 - 609
1 Oct 2017
Jin A Cobb J Hansen U Bhattacharya R Reinhard C Vo N Atwood R Li J Karunaratne A Wiles C Abel R

Objectives

Bisphosphonates (BP) are the first-line treatment for preventing fragility fractures. However, concern regarding their efficacy is growing because bisphosphonate is associated with over-suppression of remodelling and accumulation of microcracks. While dual-energy X-ray absorptiometry (DXA) scanning may show a gain in bone density, the impact of this class of drug on mechanical properties remains unclear. We therefore sought to quantify the mechanical strength of bone treated with BP (oral alendronate), and correlate data with the microarchitecture and density of microcracks in comparison with untreated controls.

Methods

Trabecular bone from hip fracture patients treated with BP (n = 10) was compared with naïve fractured (n = 14) and non-fractured controls (n = 6). Trabecular cores were synchrotron scanned and micro-CT scanned for microstructural analysis, including quantification of bone volume fraction, microarchitecture and microcracks. The specimens were then mechanically tested in compression.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 116 - 116
1 Mar 2017
Riviere C Lazennec J Muirhead-Allwood S Auvinet E Van Der Straeten C Cobb J
Full Access

The current, most popular recommendation for cup orientation, namely the Lewinnek box, dates back to the 70's, that is to say at the stone age of hip arthroplasty. Although Lewinnek's recommendations have been associated with a reduction of dislocation, some complications, either impingement or edge loading related, have not been eliminated. Early dislocations are becoming very rare and most of them probably occur in “outlier” patients with atypical pelvic/hip kinematics. Because singular problems usually need singular treatments, those patients need a more specific personalised planning of the treatment rather than a basic systematic application of Lewinnek recommendations. We aim in this review to define the potential impacts that the spine-hip relations (SHRs) have on hip arthroplasty. We highlight how recent improvements in hip implants technology and knowledge about SHRs can substantially modify the planning of a THR, and make the «Lewinnek recommendations» not relevant anymore. We propose a new classification of the SHRs with specific treatment recommendations for hip arthroplasty whose goal is to help at establishing a personalized planning of a THR. This new classification (figures 1 and 2) gives a rationale to optimize the short and long-term patient's outcomes by improving stability and reducing edge loading. We believe this new concept could be beneficial for clinical and research purposes.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 52 - 52
1 Mar 2017
Navruzov T Riviere C Van Der Straeten C Harris S Aframian A Iranpour F Cobb J Auvinet E
Full Access

Background

The accurate positioning of the total knee arthroplasty affects the survival of the implants(1). Alignment of the femoral component in relation to the native knee is best determined using pre- and post-operative 3D-CT reconstruction(2). Currently, the scans are visualised on separate displays. There is a high inter- and intra-observer variability in measurements of implant rotation and translation(3). Correct alignment is required to allow a direct comparison of the pre- and post-operative surfaces. This is prevented by the presence of the prostheses, the bone shape alteration around the implant, associated metal artefacts, and possibly a segmentation noise.

Aim

Create a novel method to automatically register pre- and post-operative femora for the direct comparison of the implant and the native bone.