Three cross sections were cut from the metaphyseal femur and surrounding bone proximal to Gruen zones 2 and 6 (regions with HA coating). The three sections were A (proximal), B (mid-part) and C (distal). Sections were prepared using the Donath technique and then paragon stained for quantitative histomorphometry using an Axioskop microscope (Carl Zeiss, Germany) with image analysing (SAMBA, France). For each segment the total implant perimeter, percentage of implant perimeter covered by bone and the total percentage of residual HA coating were measured. Bone implant contact was defined as direct ongrowth of bone to the coating or the titanium surface.
Bone ongrowth ranged between 18%–56% and was independent of the time in-vivo. Bone ongrowth was most strongly correlated to patient age with younger patients having significantly higher bone ongrowth (p=0.001). Bone ongrowth was correlated with HA-resorption only in the most proximal zone A (p=0.001) with lower ongrowth associated with lower levels of residual HA. However, HA resorption was not significantly correlated with patient age. HA resorption was significantly higher most proximally with less residual HA (13.0%) than mid-stem (22.6%, p=0.05) and distal (28.1%, p=0.05). Metaphyseal stem level and bone ongrowth were not significantly correlated in this manner.
Bone ongrowth but not HA resorption was strongly correlated to patient age indicating that the bone remodelling process is more affected by individual bone health than it can be stimulated by HA coating. HA resorption increased significantly from the distal to mid-stem and the most proximal coating level in the same way as stress shielding and thus osteoclastic stimulation goes up.
We have compared four computer-assisted methods to measure penetration of the femoral head into the acetabular component in total hip replacement. These were the Martell Hip Analysis suite 7.14, Rogan HyperOrtho, Rogan View Pro-X and Roman v1.70. The images used for the investigation comprised 24 anteroposterior digital radiographs and 24 conventional acetate radiographs which were scanned to provide digital images. These radiographs were acquired from 24 patients with an uncemented total hip replacement with a follow-up of approximately eight years (mean 8.1; 6.3 to 9.1). Each image was measured twice by two blinded observers. The mean annual rates of penetration of the femoral head measured in the eight-year single image analysis were: Martell, 0.24 (SD 0.19); HyperOrtho, 0.12 (SD 0.08); View Pro-X, 0.12 (SD 0.06); Roman, 0.12 (SD 0.07). In paired analysis of the six-month and eight-year radiographs: Martell, 0.35 (SD 0.22); HyperOrtho, 0.15 (SD 0.13); View Pro-X, 0.11 (SD 0.06); Roman, 0.11 (SD 0.07). The intra- and inter-observer variability for the paired analysis was best for View Pro-X and Roman software, with intraclass correlations of 0.97, 0.87 and 0.96, 0.87, respectively, and worst for HyperOrtho and Martell, with intraclass correlations of 0.46, 0.13 and 0.33, 0.39, respectively. The Roman method proved the most precise and the most easy to use in clinical practice and the software is available free of charge. The Martell method showed the lowest precision, indicating a problem with its edge detection algorithm on digital images.
This study confirms that the Vancouver classification and the modified algorithm for the management of PPF are a simple,reproducible classification system also for the uncemented treatment modality. Conservative treatment is a valid option in case of a stable implant, while in case of a loose implant surgical intervention is mandatory.
For the ABG-I cup the literature review gave survival rates between 59%–97% after 8–10 years. In our own study cup survival was 97.4% at 10 years. Looking at PE-wear, the literature gave average wear rates ranging from 0.24 to 0.32mm/year, values clearly above the wear rate boundary of 0.10–0.15mm/year usually considered as critical. In our own study augmented PE-wear (>
0.15mm/year) was noted in 23.6% of all implants. The majority (77%) of these implants were from patients younger than 70 years although this group only represented 57% of the total. The revision rates at 10 years reflect a similar trend with values much higher for patients below 70 years (2.8%) than above(4.9%).