Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 316 - 316
1 Mar 2013
Russell N Rives A Bruce W Pelletier M Walsh W
Full Access

Introduction

Gamma Irradiation is often considered the gold standard for sterilizing bone allograft. However, a dose dependant decrease in the static mechanical properties of gamma irradiated bone has been well established. Supercritical Fluid Sterilization (SCF) using carbon dioxide represents a potential alternate method to sterilize allografts. This study aimed to evaluate the effect of SCF on the static and dynamic (fatigue) properties of cortical bone in 3-point bending.

Methods

Eighty paired 18-month old rabbit humeri were randomized to 4 treatments: Gamma Irradiation at 10 kGy or 25 kGy, SCF Control and SCF with Peracetic Acid (Figure 1) (n=20 pairs per group). One side was treated while the other acted as a control. Ten pairs in each group were tested statically at 5 mm/min; while ten were tested dynamically between 15–150 N at 4 Hz. Samples were fatigued to failure or 50000 cycles (run-out). All testing was performed at room temperature in a saline bath. A 2-tailed t-test was used to test for significance within pairs and a one-way ANOVA with Games-Howell post-hoc test was used to test between groups.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 294 - 294
1 Mar 2013
Oliver R Brinkman M Christou C Bruce W Walsh W
Full Access

Introduction

The reduction of intraoperative blood loss during total knee arthroplasty (TKA) and total hip arthroplasty (THA) and even organ resection is an important factor for surgeons as well as the patient. In order to cauterize blood vessels to stop bleeding diathermy is commonly used and involves the use of high frequency and induces localized tissue damage and burning. Saline-coupled bipolar sealing RFE technology however has been shown to reduce tissue carbonization, however the dosage effects of RFE are not well known for both bone and soft tissue. This study examined sealing progression of blood vessels using a range of energy levels of saline-coupled bipolar RFE on bone and various soft tissues in a non-survival animal study.

Materials and Methods

Following institutional ethical approval, three mature sheep were used to examine the cancellous bone of the femoral trochlear groove and soft tissue (liver, kidney, lung, pancreas and mesentry peritoneum) subjected to the following treatment regime varying by watts and time: (1) untreated control, (2) 50 W for 1 sec, 2 sec, 3 sec and 5 sec, (3) 140 W for 1 sec, 2 sec, 3 sec and 5 sec and (4) 170 W for 1 sec, 2 sec, 3 sec and 5 sec. The Aquamantys™ System Generator and hand piece (Salient Surgical Technologies, Inc, Portsmouth, NH) coupled to a saline (0.9% NaCl) drip was used to apply RFE to the various tissues. Two clinical diathermy settings were used as controls. Tissues were immediately harvested, fixed in 10% buffered formalin and prepared for routine paraffin histology. Stained sections were evaluated in a blinded fashion for the acute in vivo response.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 59 - 59
1 Mar 2013
Esposito C Roques A Tuke M Zicat B Walter WK Walsh W Walter WL
Full Access

Introduction

Edge loading commonly occurs in all bearings in hip arthroplasty. Edge loading wear can occur in these bearings when the biomechanical loading axis reaches the edge and the femoral head loads the edge of the cup producing wear damage on both the head and cup edge. When the biomechanical loading axis passes through the polished articulating surface of the acetabular component and does not reach the edge, the center of the head and the center of the cup are concentric. The resulting wear known as concentric wear is low in metal-on-metal (MOM) bearings, and is negligible in ceramic-on-ceramic (COC) bearings. Edge loading is well defined in COC hip bearings. However, edge loading is difficult to identify in MOM bearings, since the metal bearing surfaces do not show wear patterns macroscopically. The aims of this study are to compare edge loading wear rates in COC and MOM bearings, and to relate edge loading to clinical complications.

Materials and Methods

Twenty-nine failed large diameter metal-on-metal hip bearings (17 total hips, 12 resurfacings) were compared to 54 failed alumina-on-alumina bearings collected from 1998 to 2011. Most COC bearings were revised for aseptic loosening or periprosthetic bone fracture, while most MOM bearings were revised for pain, soft tissue reactions or impingement. The median time to revision was 3.2 years for the metal hip bearings and 3.5 years for alumina hip bearings. The surface topography of the femoral heads was measured using a RedLux AHP (Artificial Hip Profiler, RedLux Ltd, Southampton, UK).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 60 - 60
1 Mar 2013
Esposito C Oliver R Campbell P Walter WK Walter WL Walsh W
Full Access

In patients with conventional metal-on-Polyethylene (MoP) hip replacements, osteolysis can occur in response to wear debris. During revision hip surgery, surgeons usually remove the source of osteolysis (polyethylene) but cannot always remove all of the inflammatory granulomatous tissues in the joint. We used a human/rat xenograft model to evaluate the effects of polyethylene granuloma tissues on bone healing. Human osteoarthritic and periprosthetic tissues collected during primary and revision hip arthroplasty surgeries were transplanted into the distal femora of athymic (nude) rats. The tissues were assessed before and after implantation and the bone response to the tissues was evaluated after 1 week and 3 weeks using micro-computed tomography, histology, and immunohistochemistry. After 3 weeks, the majority (70%) of defects filled with osteoarthritic tissues healed, while only 21% of defects with polyethylene granuloma tissues healed. Polyethylene granuloma tissues in trabecular bone defects inhibited bone healing. Surgeons should remove polyethylene granuloma tissues during revision surgery when possible, since these tissues may slow bone healing around a newly implanted prosthesis. This model provides a method for delivering clinically relevant sized particles into an in vivo model for investigation.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 359 - 359
1 Mar 2013
Walsh W Christou C Low A Yu Y Oliver R Bertollo N Schlossberg B Lloyd W Ahn E
Full Access

Introduction

The need for regeneration and repair of bone presents itself in a variety of clinical situations. The current gold standard of treatment is autograft harvested from the iliac crest or local bone. Inherent disadvantages associated with the use of autogenous bone include limited supply, increased operating time and donor site morbidity. This study utilized a challenging model of posterolateral fusion to evaluate the in vivo response of an engineered collagen carrier combined with nano-structured hydroxyapatite (NanOss Bioactive 3D, Pioneer Surgical) compared to a collagen porous beta-tricalcium phosphate bone void filler (Vitoss BA, Orthovita).

Materials and Methods

A single level posterolateral fusion was performed in 72 adult rabbits at 6, 12 and 26 weeks (8 per group per time point). Group 1: nanOss Bioactive 3D + bone marrow aspirate (BMA) + autograft, Group 2: Vitoss BA + BMA and Group 3: Autograft + BMA were compared were compared using radiographic (X-ray and Micro-computed tomography (μCT), biomechanics (manual palpation and tensile testing at 12 and 26 weeks) and histology.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 16 - 16
1 Mar 2013
Bertollo N Sorial R Low A Walsh W
Full Access

Introduction

The marriage of rapid prototyping technologies with Arthroplasty has resulted in the fabrication and use of cutting jigs and guides which are tailored to a patients' individual anatomy. These disposable cutting blocks are designed based on input parameters obtained from pre-operative CT and MRI scans and manufactured using 3-D printers. Indirect benefits include a reduction in inventory and a decrease in the burden for central sterilising units. This approach is advantageous for the surgeon in the attainment of ideal mechanical alignment, which is known to be associated with an improved clinical outcome and implant longevity. This study evaluated the postoperative alignment parameters from a single surgeon series of patients following TKA with the Signature (Biomet) system.

Methods and Materials

The postoperative alignment of a single surgeon series of 60 consecutive patients receiving a Vanguard cruciate retaining TKR (Biomet) using the Signature patient-specific surgical positioning guides was performed. Postoperative CT and preoperative templating MRI scans were imported into Mimics 14.0 (Materialise, Belgium) where specific bony landmarks were identified in both data sets. A subset of these points was used to transform the MRI data into the CT coordinate frame to enable the computation of femoral mechanical alignment in the absence of a full-length lower limb CT scan. CT and transformed MRI landmarks were then imported into ProEngineer (PTC, MA) where angular measurements were made by projecting axes onto anotomical planes. Flexion, rotation, valgus/varus of the femoral component and posterior slope, rotation and valgus/varus of the tibial component were computed. Femoral rotation was referenced to the trans-epicondylar axis as opposed to Whiteside's line. Overall limb alignment was determined based on individual component position.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 63 - 63
1 Sep 2012
Wong M Yu Y Yang J Walsh W
Full Access

The purpose of this study was to understand the effects of terminal sterilisation and residual calcium on human demineralised bone matrix (DBM) in ectopic bone formation in nude rat.

The intramuscular implantation of human DBM prepared by the Queensland Bone Bank (QBB) from four donors into eight male athymic rats was used to assess osteoinductivity. The DBM contained different levels of residual calcium and treated with or without gamma-irradiation at 11kGy. At 6 weeks post-implantation, calcium deposition was assessed by manual palpitation and radiological imaging. Tissue morphology and cellular interactions was analysed using various histological staining methods whilst protein expression of anabolic and catabolic biomarkers were examined through immunohistochemistry. All results were then analysed in qualitative, semi-quantitative and quantitative manners and tested for statistical significance.

Bone formation was observed in all specimens at the gross level. This was confirmed by histology which revealed bony capsules surrounded by soft tissue in the muscle pockets and differences in tissue components. On a cellular level, variations in osteoclast expression were found between the two groups as well as amongst individual donors through statistical analysis which resulted in an imbalance of the expression of anabolic and catabolic markers. Furthermore, a positive relationship between residual calcium and new bone formation in gamma irradiated DBM samples was found. To date, no studies have compared the effect of calcium in gamma irradiated DBM.

Our results suggest that gamma irradiation even at low doses and residual calcium may affect new bone formation. Taken together, this study stresses the importance of selecting ideal conditions for graft processing and the need to identify an optimal level of irradiation and remaining calcium levels that confers a balance between osteoinductivity and sterility.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 131 - 131
1 Sep 2012
Walter W Esposito C Roques A Zicat B Walter W Walsh W
Full Access

Edge loading commonly occurs in all bearings in hip arthroplasty. The aim of this study compares metal bearings with edge loading to alumina bearings with edge loading and to metal bearings without edge loading.

Seventeen failed large diameter metal-on-metal hip bearings (8 total hips, 9 resurfacings) were compared to 55 failed alumina-on-alumina bearings collected from 1998 to 2010. The surface topography of the femoral heads was measured using a chromatically encoded confocal measurement machine (Artificial Hip Profiler, RedLux Ltd.).

The median time to revision for the metal hip bearings and the alumina hip bearings was 2.7 years. Forty-six out of 55 (84%) alumina bearings and 9 out 17 (53%) metal bearings had edge loading wear (p<0.01). The average volumetric wear rate for metal femoral heads was 7.87 mm3/yr (median 0.25 mm3/yr) and for alumina heads was 0.78 mm3/yr (median 0.18 mm3/yr) (p=0.02).

The average volumetric wear rate for metal heads with edge loading was 16.51 mm3/yr (median 1.77 mm3/yr) and for metal heads without edge loading was 0.19 mm3/yr (median 0 mm3/yr) (p=0.1). There was a significant difference in gender, with a higher ratio of females in the alumina group than the metal group (p=0.02).

Large diameter metal femoral heads with edge loading have a higher wear rate than smaller alumina heads with edge loading. Metal-on-metal bearings have low wear when edge loading does not occur.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 92 - 92
1 Sep 2012
Bertollo N Crook T Hope B Scougall P Lunz D Walsh W
Full Access

Shape memory staples have several uses in both hand and foot and ankle surgery. There is relatively little data available regarding the biomechanical properties of staples, in terms of both the compression achieved and potential decay of mechanical advantage with time. An understanding of these properties is therefore important for the surgeon.

Two blocks of synthetic polyurethane mimicking properties of cancellous bone were fixed in jigs to both the actuator and 6 degree-of-freedom load cell of an MTS servohydraulic testing machine. With the displacement between the blocks held constant the peak value and subsequent decay in compressive force applied by both the smooth and barbed version of the nitinol OSStaple (Biomedical Enterprises), Easyclip (LMT), Herbert Bone Screws (Martin) and the Headless Compression Screw (Synthes) was measured. Nitinol staples were energised once only. A second experiment was conducted to assess the effects of repeated energisation on these parameters.

The Easyclip staples achieved a mean peak force of 5.2N, whilst the smooth and barbed OSStaples achieved values of 9.3N and 5.7N, respectively. The Herbert screws achieved a mean peak force of 9N and the headless compression screws 23.9N. The mean peak force achieved with 2 Easyclip staples in parallel was 8.1N. Following the application of a single energisation the OSStaples exhibited a significant reduction in compressive load, losing up to approximately 70% of the peak value attained. The repeated energisation of these nitinol staples produced progressive increases in both peak and trough loads, the positive effects exhibited a plateau with time.

Performance of both OSStaples was comparable to the Herbert screw with regard to reduction load applied across a simulated fracture plane. The maximum load applied by the OSStaples diminished with time. Staples provide fixation without violating the fracture plane which has the potential to offer some benefits from a healing perspective.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 32 - 32
1 Sep 2012
Yu Y Luk F Yang J Walsh W
Full Access

To set up an osteosarcoma mouse model with spontaneous lung metastasis and to identify a marker of osteosarcoma metastasis and to inhibit the marker against the invasive ability of an osteosarcoma cell line.

A human osteosarcoma orthotopic mouse model was set up by injecting 143B human osteosarcoma cells into mouse tibia. Type I insulin-like growth factor receptor (IGF-1R) and its downstream signalling factors were measured in samples from the primary tumor and the lung secondaries by immunohistochemistry. Human Alu mRNA expression was tested using in situ hybridization assay. A Matrigel assay was used to assess cell invasion ability under the interference of a MEK/ERK pathway specific inhibitor, U0126.

All fifteen mice showed tumour mass at the left tibia and lung metastasis. Human Alu expression in the primary and secondary tumours confirmed human origin of the tumour cells. Total IGF-1R, MEK, Akt, p38 and phosphorylated MEK (p-MEK), but not p-Akt and p-p38, were positive in both local tumours and lung secondaries. Leiomyosarcoma controls expressed p-Akt and p-MEK, but not p-p38. The 143B cells treated with U0126 had significantly lower in vitro invasion ability compared with controls.

The IGF-1R-MEK signalling pathway, particularly Ras/Raf/MEK/ERK, may play an important role in osteosarcoma lung metastasis, and the targeting MEK/ERK by its specific inhibitor may have a potential use in the effective treatment of osteosarcoma.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 8 - 8
1 Sep 2012
Lovric V Ledger M Goldberg J Harper W Yu Y Walsh W
Full Access

Animal studies examining tendon-bone healing have demonstrated that the overall structure, composition, and organization of direct type entheses are not regenerated following repair. We examined the effect of Low-Intensity Pulsed Ultrasound (LIPUS) on tendon-bone healing. LIPUS may accelerate and augment the tendon-bone healing process through alteration of critical molecular expressions.

Eight skeletally mature wethers, randomly allocated to either control group (n=4) or LIPUS group (n=4), underwent rotator cuff surgery following injury to the infraspinatus tendon. All animals were sacrificed 28 days post surgery to allow examination of early effects of LIPUS. Humeral head – infraspinatus tendon constructs were harvested and processed for histology and immunohistochemical staining for BMP2, Smad4, VEGF and RUNX2. All the growth factors were semiquantitative evaluated. T-tests were used to examine differences which were considered significant at p < 0.05. Levene's Test (p < 0.05) was used to confirm variance homogeneity of the populations.

The surgery and LIPUS treatment were well tolerated by all animals. Placement of LIPUS sensor did not unsettle the animals. Histologic appearance at the tendon-bone interface in LIPUS treated group demonstrated general improvement in appearance compared to controls. Generally a thicker region of newly formed woven bone, morphologically resembling trabecular bone, was noted at the tendon-bone interface in the LIPUS-treated group compared to the controls. Structurally, treatment group also showed evidence of a mature interface between tendon and bone as indicated by alignment of collagen fibres as visualized under polarized light. Immunohistochemistry revealed an increase in the protein expression patterns of VEGF (p = 0.038), RUNX2 (p = 0.02) and Smad4 (p = 0.05) in the treatment group. There was no statistical difference found in the expression patterns of BMP2. VEGF was positively stained within osteoblasts in newly formed bone, endothelial cells and some fibroblasts at the interface and focally within fibroblasts around the newly formed vessels. Expression patterns of RUNX2 were similar to that of BMP-2; the staining was noted in active fibroblasts found at the interface as well as in osteoblast-like cells and osteoprogenitor cells. Immunostaining of Smad4 was present in all cell types at the healing interface.

The results of this study indicate that LIPUS may aid in tendon to bone healing process in patients who have undergone rotator cuff repair. This treatment may also be beneficial following other types of reconstructive surgeries involving the tendon-bone interface.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 150 - 150
1 Sep 2012
Christou C MacDonald M Walsh W
Full Access

Treatment of large segmental defects in the extremities is challenging. A segmental tibial defect model in a large animal can provide a basis through which in vivo testing of materials and techniques for use in non-unions and severe trauma cases can be examined.

This study reports such a model.

Six aged ewes (> 5 years) were used following ethical approval. A 5cm piece of the mid diaphysis of the left tibia was removed including its associated periosteum. The tibia was stabilized with an 8mm stainless steel cross locked intramedullary nail and all tissues closed in their respective layers. Animals were euthanised at 12 weeks following surgery and evaluated using radiographic, micro-computed tomography (CT), soft tissue and hard tissue histology techniques.

Three weeks post operatively one of the intramedullary nails failed through the first of the distal two cross locking screw holes, the sheep was euthanised and the tibia was harvested. Early signs of callus formation were evident at the osteotomy edges originating from the periosteal surface; the defect space was bridged by fibrous scar tissue.

The remaining 5 sheep were taken out to the 12 week time point then all relevant tissues were harvested. Gross dissection revealed a lack of bony union in the defect site and no evidence of infection. X-rays and CT showed a lack of hard tissue callus bridging in the defect region at 12 weeks. Histological sections of the bridging tissues revealed, callus originating from both the periosteal and endosteal surfaces, with fibrous tissue completing the bridging in all instances. One case had cartilaginous tissue developing; however this was incomplete at 12 weeks.

As none of the 12 week time point sheep achieved clinical union; this model may be effective as a basis for the investigation of healing adjuncts to be used in non-union cases, where severe traumatic injury has lead to significant bone loss such as blast injuries or following large tumour removal.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 143 - 143
1 Sep 2012
Esposito C Roques A Tuke M Walter W Walsh W
Full Access

Two types of ceramic materials currently used in total hip replacements are third generation hot isostatic pressed (HIPed) alumina ceramic (commercially known as BIOLOX®forte, CeramTec) and an alumina matrix composite material consisting of 75% alumina, 24% zirconia, and 1% mixed oxides (BIOLOX®delta, CeramTec). The aim of this study is to compare BIOLOX delta femoral heads to BIOLOX forte femoral heads revised within 2 years in vivo.

Ceramic bearings revised at one center from 1998 to 2010 were collected (61 bearings). BIOLOX delta heads (n=11) revised between 1–33 months were compared to BIOLOX forte femoral heads with less than 24 months in vivo (n=20). The surface topography of the femoral heads was measured using a chromatically encoded confocal measurement machine (Artificial Hip Profiler, RedLux Ltd.).

The median time to revision for BIOLOX delta femoral heads was 12 months, compared to 13 months for BIOLOX forte femoral heads. Sixteen out of 20 BIOLOX forte femoral heads and 6 out of 11 BIOLOX delta femoral heads had edge loading wear. The average volumetric wear rate for BIOLOX forte was 0.96 mm3/yr (median 0.13 mm3/yr), and 0.06 mm3/yr (median 0.01 mm3/yr) for BIOLOX delta (p=0.03). There was no significant difference (p>0.05) in age, gender, time to revision or femoral head diameter between the two groups.

Early results suggest less volumetric wear with BIOLOX delta femoral heads in comparison to BIOLOX forte femoral heads.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 124 - 124
1 Sep 2012
Lovric V Chen D Oliver R Yu Y Genin F Walsh W
Full Access

Fibrocartilaginous entheses are formed through endochondral ossification and characterized by four zones morphologically separated into tendon, uncalcified fibrocartilage, calcified fibrocartilage and bone [1]. These zones are not successfully regenerated following surgical repair. Demineralized Bone (DBM) presented at the tendon bone interface may improve healing between tendon and bone.

Fifty six female nude rats were randomly allocated into either a control reconstruction or treatment group (DBM at the tendon-bone healing site). A modified rodent model of anterior cruciate ligament reconstruction was adopted [2]. Animals were sacrificed at 2, 4 and 6 weeks following surgery. Four rats per group were prepared for histology at each time point while eight rats were culled for biomechanical testing at 4 and 6 week time points. ANOVA and post hoc tests were used to examine differences which were considered significant at p < 0.05.

The surgical procedure was well tolerated. Macroscopic dissection did not reveal any infection and all joint surfaces appeared normal. An intra-articular graft between the femur and tibia was present in all specimens. Mechanical differences were noted between groups. Peak loads were significantly higher in treatment group at 4 and 6 weeks (6.0 ± 3.6N and 9.1 ± 2.6 N, respectively) compared to controls (2.9 ± 1.9 N and 5.8 ± 2.7 N). No statistical differences were found in graft stiffness between the groups at 4 or 6 week time points. Histology showed an initial influx of inflammatory cells coupled with formation of a loose disorganized fibrovascular interface layer between tendon and bone in both groups. By the 6 weeks the interface layer in the DBM group fused into the newly formed bone to create a continuum between the tendon and bone, in an interdigitated fashion, containing Sharpy's like fibres. In the control group the continuum was less apparent with evidence of large areas of discontinuity between the two zones. A thicker region of newly formed woven bone with increased osteoblast activity along the bone tunnel was evident in the DBM group.

DBM has the potential to increase the quality of repair following surgical procedures involving reattachment of tendon to bone.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 153 - 153
1 May 2012
Goldberg J Walsh W Chen D
Full Access

The diagnosis and treatment of disorders of the long head of the biceps tendon remains controversial. There is uncertainty as to the role of the long head of biceps and it can be difficult to determine whether the patient's pathology is coming from the biceps or other adjacent structures. In addition, the appropriate type of treatment remains controversial.

We retrospectively reviewed the files of the senior author's experience in over 4000 arthroscopic shoulder procedures. We examined cases involving isolated biceps pathology, excluding those patients with rotator cuff tears and labral pathology, involving 92 biceps tenotomies and 103 biceps tenodeses.

Our analysis supports the benefit of clinical examination over all types of radiological investigations. The benefits and technique of biceps tenodesis is described including surgical technique. Irritation by PLA interference screw is examined. A paradigm is put forward to help in diagnosis and management of these lesions.

Long head of biceps pathology is a significant cause of shoulder pain in association with other shoulder problems and in isolation. Biceps tenodesis and tenotomy is an efficacious way of dealing with this pathology.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 64 - 64
1 May 2012
McMahon S Hawdon G Bare J Sim Y Bertollo N Walsh W
Full Access

Thermal damage to bone related to the exothermic polymerisation of bone cement (PMMA) remains a concern. A series of studies were conducted to examine PMMA bone interface during cemented arthroplasty.

In vitro and in vivo temperature distributions were performed in the laboratory and human and animal surgery. In vivo (10 patients) measurements of cement temperature during cementing of BHR femoral prosthesis using thermocouples. Intra-operative measurement of cement temperature in BHR in the presence of femoral head cysts was examined in patients. The BHR femoral heads were sectioned to assess cement mantle as well as position of thermocouples. An additional study was performed in sheep with PMMA implanted into cancellous defects. Thermocouples were used to monitor temperature in the cement as well as adjacent bone. Histology and CT was used to assess any thermal damage.

The exothermic reaction of PMMA during polymerization does indeed result in an increase in temperature at the interface with bone. The in vivo study recorded a maximum temperature of 49.12C for approximately three minutes in the cancellous bone underneath the BHR prosthesis. This exposure is probably not sufficient to cause significant injury to the femoral head. The maximum temperature of the cement on the surface of the bone was 54.12C, whereas the maximum recorded in the cement in the mixing bowl was 110.2C.

In the presence of artificial cysts within the bone, however, temperatures generated within the larger cysts, and even at the bone-cement interface of these cysts, reached levels greater than those previously shown to be harmful to bone. This occurred in one case even in the 1 cc cyst.

Routine histology revealed a fibrous layer at the cement bone interface in the sheep study. Fluorescent microscopy demonstrated bone label uptake adjacent to the defect site. Histology did not reveal thermal necrosis in the defects in terms of bony necrosis. CT data was used to measure the amount of PMMA placed into each defect. This analysis revealed a range of volumes that did not seem to influence the histology.

The heat of cement polymerisation in resurfacing as performed in our study is not sufficient to cause necrosis. This may reflect the ability of the body to rapidly conduct heat away by acting as a heat sink. The temperature-conducting properties of the metal prosthesis are also likely to be important.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 164 - 164
1 May 2012
Pak P Oliver R Bell D Yu Y Bellemore J Walsh W
Full Access

Posterolateral spinal fusion using autograft in adult rabbits has been reported by many groups using the Boden model. Age in general has an adverse effect on skeletal healing; although, its role in posterolateral fusion is not well understood. This study examined the influence of animal age on spinal fusion using a standard model and experimental endpoints. We hypothesised that fusion quality and quantity would be less with increasing age.

A single level posterolateral fusion between the fifth and sixth lumbar segments were performed in six-month and two-year-old New Zealand white rabbits (n=6 per group) using morcelized iliac crest autograft. All animals were sacrificed at 12 weeks following surgery. Posteroanterior Faxitron radiographs and CT scans were taken and DICOM data was analysed (MIMICS Version 12, Materialise, Belgium). Axial, sagittal, coronal and three-dimensional models were created to visualise the fusion masses. Bone mineral density (BMD) of the fusion mass was measured using a Lunar DPXL Dexa machine. An MTS Bionix testing machine was then used to assess peak load and stiffness. Sagittal and coronal plane histology was evaluated in a blinded fashion using H&E, Tetrachrome and Pentachrome stains. Assessment included overall bony response on and between the transverse processes. Radiographs and CT confirmed a more robust healing response in younger animals. Radiographic union rates decreased from 83% to 50% in the aged animals. A neo- cortex surrounding the fusion mass was observed in the younger group but absent in the aged animals. Fusion mass BMD and that of the vertebral body was decreased in the older animals (P<0.05). Tensile mechanical data revealed a 30% reduction in peak load (P=0.024) and 34% reduction in stiffness (P=0.073) in the two-year-old animals compared with the six-month-old animals. Histological evaluation demonstrated a reduction in overall biological activity in the two-year-old animals. This reduction in activity was observed in the more challenging intertransverse space as well as adjacent to the transverse processes and vertebral bodies at the decortication sites. Numerous sites of new bone formation was present in the middle of the fusion mass in the six-month-old animals while the bone graft in the two-year- old animals were less viable.

Skeletal healing is complex and mediated by both local and systemic factors. This study demonstrated that ageing leads to an impaired and delayed skeletal repair.

Where autograft is utilised, diminished graft osteoinductivity and reduced levels of growth factors and nutritional supply in the surrounding milieu explains our observations. The aged rabbit posterolateral spinal fusion model has not been previously described but would be a useful to evaluate new treatment modalities in a more challenging host environment.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 126 - 126
1 May 2012
Russell N Rives A Pelletier M Hoffman M Bruce W Walsh W
Full Access

Cortical bone is a complex composite material composed of an inorganic mineral phase and organic matrix of type I collagen and various non-collagenous proteins. The hierarchical organisation of bone results in a transversely isotropic material with the mechanical properties in the long-axis (z) being superior to the radial and circumferential axes which are equivalent. This directional dependence of bone has been well reported, whilst the mechanisms/anisotropy are more difficult to study. This study examined the anistropic nature of cortical bone and the influence of different sterilisation procedures.

Ninety cortical bone cubes were prepared using established techniques (Walsh and Guzelsu) and randomly allocated to three treatments; control, 15 KGy, Super Critical Fluid (SCF) (n=30 per group). The ultrasonic moduli was examined using longitudinal sound waves at 5 MHz using a pulse receive technique. Unconfined compression was performed non-destructively in longitudinal (z), circumferential (ï±) and radial orientations (r). Samples were tested to failure in the z axis. A two-way analysis of variance (treatment and time) followed by a Games Howell post hoc test and covariate analysis was performed using SPSS for Windows.

Data from this study revealed some interesting and intriguing results with respect to the effects of gamma irradiation and dense gas technology on the properties of cortical bone and load transmission. A statistical decrease in the compressive stiffness and strength was noted with 15 KGy of whilst SCF treatment did not alter the properties in the r or ï orientations. Similar results were found with respect to the ultrasonic moduli (data not shown). The pilot data confirmed the adverse effects of bone in compression following gamma irradiation as we found in our recently presented ORS work. However, the study in compression demonstrated that the directional dependence that makes cortical bone a transversely isotropic material is removed following gamma irradiation with SCF did not appear to have this effect.

The effects of gamma irradiation on the mechanical performance of allografts in the long bone axis may play a role in their in vivo performance. The removal of the anisotropy following gamma irradiation provides insight into the relationship(s) between the mineral and organic constituents, which requires further study.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIII | Pages 105 - 105
1 May 2012
Pinczewski L Miller C Salmon L Williams H Walsh W
Full Access

The aim of this study was to compare the outcome of cemented TKR using either oxidized zirconium (oxinium) or cobalt chrome (CoCr) femoral components in patients undergoing simultaneous bilateral TKR. Patients involved in the study received one of each prosthesis, thereby acting as their own control. The hypothesis was that there would be no difference in the clinical and radiographic outcome between the two prosthetic materials.

Forty consecutive patients who were undergoing bilateral Genesis ll TKR consented to participate in the study. Patients were assessed preoperatively, at five days, six weeks and one, two and five years, postoperatively. The outcome measures included the KOOS, Knee Society Score, BOA Patient Satisfaction Scale, and radiographs at six weeks and one, two and five years. In two patients polyethlylene exchange was performed at 56 months from surgery during patellofemoral resurfacing. The four retrieved polyethylene liners were studied for wear with the aid of a stereo zoom microscope and an environmental scanning electron microscope (ESEM). Both the patients and the all examiners were blinded as to the prosthesis type throughout the study.

Forty patients (80 knees) were included in the study. At five years, three patients were deceased and two had developed senile dementia. No patients were lost to follow up. At five years from surgery the CoCr knee was preferred by 41% of patients compared to 13% who preferred the Oxinium knee (p=0.009). There was no significant difference in range of motion between the two prosthesis at five days, six weeks or one, two and five years. There were also no significant differences between the two prostheses in any of the other variables assessed. The four retrieved polyethylene inserts showed similar patterns of wear in terms of both wear types and patterns under examination with both the stereo zoom and scanning electron microscope with no clear differences between CoCr and Oxinium bearing against the polyethlylene. There was no difference in the grade or incidence of radiographic lucencies between the two prosthesis at five years.

At five years after surgery the only significant difference between the Genesis II Oxinium prosthesis and the CoCr prosthesis was a subjective preference for the CoCr prosthesis by a higher proportion of patients. There were no unexpected complications associated with the use the Oxinium femoral implants. In the four retrieved polyethylene liners, no significant differences were identified between the two prosthesis materials in terms of detectable wear type and patterns. Continued follow up of this cohort is planned to establish whether Oxinium femoral implants have an improved survivorship compared to CoCr femoral component in total knee replacement to warrant the additional cost.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IV | Pages 112 - 112
1 Mar 2012
Biant L Bruce W Walker P Herrmann S Walsh W
Full Access

‘High flexion’ polyethylene tibial tray inserts are available from total knee replacement (TKR) manufacturers. There is currently no published data available that examines how much extra knee flexion these new implants give or if there are any wear consequences for the change in design. The high flexion inserts are narrower posteriorly than standard inserts and have chamfers anteriorly and on the post in cruciate sacrificing designs.

This prospective randomised controlled trial of 100 patients undergoing posterior stabilised TKR compared knee flexion, measured intra-operatively by a computer navigation system, of the standard and high flexion trial inserts in the same knee. Patients were then randomised to receive either a standard or ‘high flexion’ definitive component and the stability assessed. The post-operative knee flexion of all patients was measured at six months.

High flexion inserts did not give significantly more knee flexion than standard inserts either per-operatively at the trial insert stage, or at six months post-op and resulted in marginally more anterior draw. The average per-operative difference in flexion between standard and high flex inserts measured in the same knee was 3.2° (range -4-18°) The average knee flexion at 6 months post op was 106° for both groups. The average change in knee flexion comparing pre and post op was 2.3° for the high flex group and 0.6° for the standard insert group.

Laboratory Tek scan contact pressure analysis at the surface of the standard and high flexion designs was not significantly different, but the thinner polyethylene of the high flexion design raises questions about wear characteristics. High flexion polyethylene inserts are probably not justified in terms of improved knee flexion, but may be a useful option in certain technical circumstances during TKR such as patella baja or if the patella impinges on the post in deep flexion.