This study aimed to establish the optimal fixation methods for calcaneal tuberosity avulsion fractures with different fragment thicknesses in a porcine model. A total of 36 porcine calcanea were sawed to create simple avulsion fractures with three different fragment thicknesses (5, 10, and 15 mm). They were randomly fixed with either two suture anchors or one headless screw. Load-to-failure and cyclic loading tension tests were performed for the biomechanical analysis.Aims
Methods
Mechanical stimulation is a key factor in the development and healing of tendon-bone insertion. Treadmill training is an important rehabilitation treatment. This study aims to investigate the benefits of treadmill training initiated on postoperative day 7 for tendon-bone insertion healing. A tendon-bone insertion injury healing model was established in 92 C57BL/6 male mice. All mice were divided into control and training groups by random digital table method. The control group mice had full free activity in the cage, and the training group mice started the treadmill training on postoperative day 7. The quality of tendon-bone insertion healing was evaluated by histology, immunohistochemistry, reverse transcription quantitative polymerase chain reaction, Western blotting, micro-CT, micro-MRI, open field tests, and CatWalk gait and biomechanical assessments.Aims
Methods
Aims. To draw a comparison of the
Nearly a quarter of screws cause damage during insertion by stripping the bone, reducing
Many biomechanical studies have shown that the weakest biomechanical point of a rotator cuff repair is the suture-tendon interface at the medial row. We developed a novel double rip-stop (DRS) technique to enhance the strength at the medial row for rotator cuff repair. The objective of this study was to evaluate the biomechanical properties of the DRS technique with the conventional suture-bridge (SB) technique and to evaluate the biomechanical performance of the DRS technique with medial row knots. A total of 24 fresh-frozen porcine shoulders were used. The infraspinatus tendons were sharply dissected and randomly repaired by one of three techniques: SB repair (SB group), DRS repair (DRS group), and DRS with medial row knots repair (DRSK group). Specimens were tested to failure. In addition, 3 mm gap formation was measured and ultimate failure load, stiffness, and failure modes were recorded.Aims
Methods
Introduction. From 2004 to 2015, elective lumbar fusions increased by 62% in the US. The largest increases were for among age 65 or older (139% in volume) and scoliosis (187%) [1]. Age is a well known factor of osteoporosis. The load-sharing may exceed the pedicular screws constructs in aging spine and lead to non-union and re-do. Surgical options may increase the screw purchase (e.g.: augmentation, extensions) at supplementary risks. Pedicular screw are known to cause vascular, nerve root or cord injuries. Facing these pitfalls, the surgeon's experience and rule of thumbs are the most deciding factors for the surgical planning. The aim of this study is to assess the accuracy of a patient specific tool, designed to plan a safe pedicular trajectory and to provide an intraoperative screw
Suture anchor have been used in surgical procedure of tendon or ligament repair. Recently, there has been developed an all suture anchor (soft anchor) which can be used even when the insertion area is narrow. But, the stability of soft anchors due to narrow zone has not been elucidated. This purpose of this study was to investigate stability of soft anchors with respect to their fixation intervals. Polyurethane foams with two different bone densities (10 pcf; 0.16g / cm³, 20 pcf; 0.32g / cm³) were used. All suture anchors and conventional suture anchors were fixed at 10mm, 5mm, and 2.5mm intervals. The failure load was measured using a mechanical testing machine. The average load to failure of conventional suture anchor were 200.4N, 200.2N, 184.7N in the 10mm, 5mm and 2.5mm interval with 10pcf foam bones and 200.4 N, 200.2 N and 184.7 N with the 20 pcf foam bone respectively. Average load to failure load of soft anchor was 97.3N, 93.9N and 76.9N with 10pcf foam bones and 200.4 N, 200.2 N and 184.7 N with 20 pcf foam bone. Suture screw spacing and bone density are important factors in anchor
Introduction. Suture anchor have been used in surgical procedure of tendon or ligament repair. Recently, there has been developed an all suture anchor (soft anchor) which can be used even when the insertion area is narrow. But, the stability of soft anchors due to narrow zone has not been elucidated. This purpose of this study was to investigate stability of soft anchors with respect to their fixation intervals. Methods. Polyurethane foams with two different bone densities (10 pcf; 0.16g / cm³, 20 pcf; 0.32g / cm) were used. All suture anchors and conventional suture anchors were fixed at 10mm, 5mm, and 2.5mm intervals. The failure load was measured using a mechanical testing machine. Results. The average load to failure of conventional suture anchor were 97.3N, 93.9N, and 76.9N in the 10mm, 5mm and 2.5mm interval with 10pcf foam bones and 200.4 N, 200.2 N and 184.7 N with the 20 pcf foam bone respectively. Average load to failure load of soft anchor was 97.3N, 93.9N and 76.9N with 10pcf foam bones and 200.4 N, 200.2 N and 184.7 N with 20 pcf foam bone. Conclusion. Suture screw spacing and bone density are important factors in anchor
Background. While the biomechanical properties of trans-pedicular screws have proven to be superior in the lumbar spine, little is known concerning
Background. The advantages of treatment by open reduction and internal fixation for intertrochanteric fractures of the proximal femur have been well known for several decades. Failure of fixation can result in revision surgery, prolonged inpatient stay and has major socio-economic consequences. There are many new devices on the market to help deal with this problem. Expandable hip screw (EHS) is one such device, which is an expanding bolt that may offer superior fixation in osteoporotic bone compared to the standard dynamic hip screw (DHS) type device. Methods. Static axial compression tests with elastic deformation of the specimens were performed with a crosshead speed of 10 mm/min to determine stiffness of testing was performed with 3 cycles from 0 N to 250 N, 3 cycles from 0 N to 500 N, 3 cycles from 0 N to 750 N and 3 cycles from 0 N to 1000 N with a holding time of 10 s per test cycle. Displacement control was apply the
Good lag screw holding power in trabecular bone of the femoral head is a requisite to achieve stability in the management of proximal femoral fractures. It has been demonstrated that insertion torque and
This study evaluated the mechanical performance, under low-load cyclic loading, of two different knotless suture anchor designs: sutures completely internal to the anchor body (SpeedScrew) and sutures external to the anchor body and adjacent to bone (MultiFIX P). Using standard suture loops pulled in-line with the rotator cuff (approximately 60°), anchors were tested in cadaveric bone and foam blocks representing normal to osteopenic bone. Mechanical testing included preloading to 10 N and cyclic loading for 500 cycles from 10 N to 60 N at 60 mm/min. The parameters evaluated were initial displacement, cyclic displacement and number of cycles and load at 3 mm displacement relative to preload. Video recording throughout testing documented the predominant source of suture displacement and the distance of ‘suture cutting through bone’.Objectives
Methods
INTRODUCTION. In native knees the anterior cruciate ligament (ACL) plays a major role in joint stability and kinematics. Sacrificing the ACL in contemporary total knee arthroplasty (TKA) is known to cause abnormal knee motion, and reduced function. Hence, there is growing interest in the development of ACL retaining TKA implants. Accommodation of ACL insertion around the tibial eminence is a challenge with these designs. Therefore, a reproducible and practical test setup is necessary to characterize the strength of the ACL/bone construct in ACL retaining implants. Seminal work showed importance of loading the ACL along its anatomical orientation. However, prior setups designed for this purpose are complex and difficult to incorporate into a standardized test for wide adoption. The goal of this study was to develop a standardized and anatomically relevant test setup for repeatable strength assessment of ACL construct using basic force-displacement testing equipment. METHODS. Cadaver knees were positioned with the ACL oriented along the loading axis and being the only connection between femur and tibia. 15° knee flexion was selected based on highest ACL tensions reported in literature. Therefore, the fixtures were adjusted accordingly to retain 15° knee flexion when the ACL was tensioned. The test protocol included 10 cycles of preconditioning between 6N and 60N at 1mm/s, followed by continuous distraction at 1mm/s until failure (Fig. 1). Eleven cadaveric knees (4 male, 7 female; 70.9 yrs +/−13.9 yrs) were tested using this setup to characterize a baseline ACL
Summary. A rotational limit for screw insertion may improve screw purchase and plate compression by reducing stripping, as compared to a torque based limit. Introduction. Over-tightening screws results in inadvertent stripping of 20% of cortical bone screws. The current method of “two-fingers tight” to insert screws relies on the surgeon receiving torque feedback. Torque, however, can be affected by screw pitch, bone density and bone-thread friction. An alternative method of tightening screws is the “turn-of-the-nut” model, commonly used in engineering applications. In the “turn-of-the-nut” method, nuts used to fasten a joint are rotated a specific amount in order to achieve a pre-specified bolt tension. When applied to orthopaedics, bone assumes the role of the nut and the screw is the bolt. The screw is turned a set angular rotation that is independent of torque feedback. Potentially the “turn-of-the-nut” method provides an easier way of screw insertion that might lessen inadvertent screw stripping. The purpose of the current study was to use the “turn-of-the-nut” method to determine the angular rotation that results in peak plate compression and peak screw pullout force. Methods. Three pairs of human humeri in each of three groups (osteopenic, osteoporotic, and normal) underwent plate compression and pullout protocols. For plate compression, 3.5-mm screws were tightened into strain gauge instrumented plate until screw stripping occurred. Insertion torque, plate compression, and screw rotation were measured. For pullout, 3.5-mm screws were inserted until the head contacted the plate, additionally rotated (90, 180, 270, or 360 degrees), and then pulled out. A generalised linear and latent mixed model was used to check for significant associations (P < 0.05). Results. Mean (95% CI) peak plate compression occurred at 286 degrees (range, 261 – 311 degrees) beyond screw seating. Plate compression significantly increased at 90 to 135 degrees but not after 180 degrees. At 270 degrees, 39% of the screws had already reached their peak ability to compress. Peak screw torque lagged behind peak plate compression by 31 ± 50 degrees, and in seeking peak screw torque, a loss of 104 ± 115 N in plate compression resulted. Screw pullout force was greatest at 90 degrees, but it was not significantly different from that of the other angle groups. Conclusions. Screw rotation at 180 degrees provides plate compression and
The purpose of this study was to evaluate and
compare the effect of short segment pedicle screw instrumentation and
an intermediate screw (SSPI+IS) on the radiological outcome of type
A thoracolumbar fractures, as judged by the load-sharing classification,
percentage canal area reduction and remodelling. We retrospectively evaluated 39 patients who had undergone hyperlordotic
SSPI+IS for an AO-Magerl Type-A thoracolumbar fracture. Their mean
age was 35.1 (16 to 60) and the mean follow-up was 22.9 months (12
to 36). There were 26 men and 13 women in the study group. In total,
18 patients had a load-sharing classification score of seven and
21 a score of six. All radiographs and CT scans were evaluated for
sagittal index, anterior body height compression (%ABC), spinal
canal area and encroachment. There were no significant differences
between the low and high score groups with respect to age, duration
of follow-up, pre-operative sagittal index or pre-operative anterior
body height compression (p = 0.217, 0.104, 0.104, and 0.109 respectively).
The mean pre-operative sagittal index was 19.6° (12° to 28°) which
was corrected to -1.8° (-5° to 3°) post-operatively and 2.4° (0°
to 8°) at final follow-up (p = 0.835 for sagittal deformity). No
patient needed revision for loss of correction or failure of instrumentation. Hyperlordotic reduction and short segment pedicle screw instrumentation
and an intermediate screw is a safe and effective method of treating
burst fractures of the thoracolumbar spine. It gives excellent radiological
results with a very low rate of failure regardless of whether the
fractures have a high or low load-sharing classification score. Cite this article
Antegrade nailing of proximal humeral fractures
using a straight nail can damage the bony insertion of the supraspinatus
tendon and may lead to varus failure of the construct. In order
to establish the ideal anatomical landmarks for insertion of the
nail and their clinical relevance we analysed CT scans of bilateral
proximal humeri in 200 patients (mean age 45.1 years ( We therefore emphasise the need for ‘fastidious’ pre-operative
planning to minimise this risk. Cite this article:
In Neer type II (Robinson type 3B) fractures
of the distal clavicle the medial fragment is detached from the coracoclavicular
ligaments and displaced upwards, whereas the lateral fragment, which
is usually small, maintains its position. Several fixation techniques
have been suggested to treat this fracture. The aim of this study
was to assess the outcome of patients with type II distal clavicle
fractures treated with coracoclavicular suture fixation using three
loops of Ethibond. This prospective study included 14 patients with
Neer type II fractures treated with open reduction and coracoclavicular
fixation. Ethibond sutures were passed under the coracoid and around
the clavicle (UCAC loop) without making any drill holes in the proximal
or distal fragments. There were 11 men and three women with a mean
age of 34.57 years (29 to 41). Patients were followed for a mean
of 24.64 months (14 to 31) and evaluated radiologically and clinically
using the Constant score. Fracture union was obtained in 13 patients at
a mean of 18.23 weeks (13 to 23) and the mean Constant score was
96.07 (91 to 100). One patient developed an asymptomatic fibrous
nonunion at one year. This study suggests that open reduction and
internal fixation of unstable distal clavicle fractures using UCAC
loops can provide rigid fixation and lead to bony union. This technique avoids
using metal hardware, preserves the acromioclavicular joint and
provides adequate stability with excellent results. Cite this article:
Introduction. Short-segment posterior instrumentation for spine fractures is threatened by unacceptable failure rates. Two important design objectives of pedicle screws, bending and
Introduction. Pedicle screw pullout or loosening is increased in the osteoporotic spine. Recent studies showed a significant increase of pullout forces especially for PMMA-augmentation. With application of conventional viscosity PMMA the risk of cement extravasation is associated. This risk can be reduced by using radiofrequency-responsive, ultrahigh viscosity bone cement. Method. 11 fresh-frozen lumbar vertebral bodies (VB) from 5 cadavers were collected and freed from soft-tissue and ligaments. By DEXA scan (Siemens QDR 2000) 8 VB were identified as severely osteoporotic (BMD 0.8 g/cm3), 3 VB were above this level. Two screws (6×45 mm, WSI-Expertise Inject, Peter Brehm, Weisendorf, Germany) were placed in the pedicles. Through the right screw 3ml of radiofrequency-responsive bone cement (StabiliT® ER2 Bone Cement, DFine, Germany) were injected via hydraulic cement delivery system (StabiliT® Vertebral Augmentation System, DFine, Germany). As control group, left pedicle screws remained uncemented. After potting the whole VB in technical PMMA (Technovit 4004, Heraeus Kulzer, Germany) axial pullout test was performed by a material testing device (Zwick-Roell, Zmart-Pro, Ulm, Germany). Results. The mean BMD of all specimen was 0.771 g/cm3 (min./max. 0.615/1.116, SD ± 0.170). Due to the definition of osteoporosis 8 specimens had a BMD lower than 0.8 g/cm3 (mean 0.677, min./max. 0.615/0,730, SD ±0.045). The non-osteoporotic group consisted of 3 specimens with a mean BMD of 1.020 g/cm3 (min/max 0.928/1.116, SD ±0.094). Overall we observed an increase in the mean axial