Aims
Giant cell tumour of bone (GCTB) is a locally aggressive lesion that is difficult to treat as salvaging the joint can be associated with a high rate of local recurrence (LR). We evaluated the risk factors for tumour relapse after treatment of a GCTB of the limbs.
Methods
A total of 354 consecutive patients with a GCTB underwent joint salvage by curettage and reconstruction with bone graft and/or cement or en bloc resection. Patient, tumour, and treatment factors were analyzed for their impact on LR. Patients treated with denosumab were excluded.
Aims
The objectives of this study were to investigate the patient characteristics and mortality of Vancouver type B periprosthetic femoral fractures (PFF) subgroups divided into two groups according to femoral component stability and to compare postoperative clinical outcomes according to treatment in Vancouver type B2 and B3 fractures.
Methods
A total of 126 Vancouver type B fractures were analyzed from 2010 to 2019 in 11 associated centres' database (named TRON). We divided the patients into two Vancouver type B subtypes according to implant stability. Patient demographics and functional scores were assessed in the Vancouver type B subtypes. We estimated the mortality according to various patient characteristics and clinical outcomes between the open reduction internal fixation (ORIF) and revision arthroplasty (revision) groups in patients with unstable subtype.
Aims. Orthopaedic and reconstructive surgeons are faced with large defects after the resection of malignant tumours of the sacrum. Spinopelvic reconstruction is advocated for resections above the level of the S1 neural foramina or involving the sacroiliac joint. Fixation may be augmented with either free vascularized fibular flaps (FVFs) or
Aims
Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique.
Patients and Methods
A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).
Aims
This study reviews the use of a titanium mesh cage (TMC) as an adjunct to intramedullary nail or plate reconstruction of an extra-articular segmental long bone defect.
Patients and Methods
A total of 17 patients (aged 17 to 61 years) treated for a segmental long bone defect by nail or plate fixation and an adjunctive TMC were included. The bone defects treated were in the tibia (nine), femur (six), radius (one), and humerus (one). The mean length of the segmental bone defect was 8.4 cm (2.2 to 13); the mean length of the titanium mesh cage was 8.3 cm (2.6 to 13). The clinical and radiological records of the patients were analyzed retrospectively.
Revision of the humeral component in shoulder arthroplasty is frequently necessary during revision surgery. Newer devices have been developed that allow for easy extraction or conversion at the time of revision preserving bone stock and simplifying the procedure. However, early generation anatomic and reverse humeral stems were frequently cemented into place. Monoblock or fixed collar stems make accessing the canal from above challenging. The cortex of the Humerus is far thinner than the femur and stress shielding has commonly led to osteopenia. Many stem designs have fins that project into the tuberosities putting them at risk for fracture on extraction. Extraction starts with an extended deltopectoral incision from the clavicle to the deltoid insertion. The proximal humerus needs to be freed from adhesions of the deltoid and conjoined tendon. The deltopectoral interval is fully developed. Complete subscapularis and anterior capsular release to the level of the latissimus tendon permits full exposure of the humeral head. After head removal the stem can be assessed for loosening and signs of periprosthetic joint infection. The proximal bone around the fin of the implant should be removed from the canal. If possible, the manufacturer's extractor should be utilised. If not, then a blunt impactor can be placed from below against the collar of the stem to assist in extraction. With luck the stem can be extracted from the cement mantle. If there is no concern for infection, the cement-in-cement technique can be used for revision. Otherwise, attempts should be made to extract all the cement and cement restrictor, if present. The small cement removal tools from the hip set can be used and specialised shoulder tools are available. An ultrasound cement removal device can be very helpful. The surgeon must be particularly careful to avoid perforation of the humeral cortex. This is especially important when near the radial nerve as injury can occur. When a well-fixed stem is encountered, an osteotomy of the proximal humerus is necessary. The surgeon can utilise a linear cut with an oscillating saw along the bicipital groove for the length of the implant. An osteotome is used to crack the cement mantle allowing stem extraction. Alternatively, a window can be created to offer additional access to the cement mantle. In the event the surgeon has required an osteotomy or window, cerclage wires, cables or suture will be needed and when the bone is potentially compromised,
Periprosthetic fractures involving the femoral meta/diaphysis can be treated in various fashions. The overall incidence of those fractures after primary total knee arthroplasties (TKA) ranges from 0.3 to 2.5%, however, can increase above 30% in revision TKA, especially in older patients with poorer bone quality. Various classifications suggest treatment algorithms. However, they are not followed consequently. Revision arthroplasty becomes always necessary if the implant becomes loose. Next, it should be considered in case of an unhappy TKA prior to the fracture rather than going for an osteosynthesis. Coverage of the associated segmental bone loss in combination with proximal fixation, can be achieved in either cemented or non-cemented techniques, with or without the combination of osteosynthetic fracture stabilization. Severe destruction of the metaphyseal bone, often does not allow adequate implant fixation for the revision implant and often does not allow proper anatomic alignment. In addition the destruction might include loss of integrity of the collaterals. Consequently standard or even revision implants might not be appropriate. Although first reports about partial distal femoral replacement are available since the 1960´s, larger case series or technical reports are rare within the literature and limited to some specialised centers. Most series are reported by oncologic centers, with necessary larger osseous resections of the distal femur. The implantation of any mega prosthesis system requires meticulous planning, especially to calculate the appropriate leg length of the implant and resulting leg length. After implant and maybe cement removal, non-structural bone might be resected. Trial insertion is important due to the variation of overall muscle tension and recreation of the former joint line. So far very few companies offer yet such a complete, modular system which might also be expanded to a total femur solution. Furthermore it should allow the implantation of either a cemented or uncemented diaphyseal fixation. In general, the fracture should be well bridged with a longer stem in place. At least 3 cm to 5 cm of intact diaphysis away to the fracture site is required for stable fixation for both cemented and cementless stems. Application of
We determined the efficacy of a devitalised autograft (n = 13) and allograft (n = 16) cortical strut bone graft combined with long-stem endoprosthetic reconstruction in the treatment of massive tumours of the lower limb. A total of 29 patients (18 men:11 women, mean age 20.1 years (12 to 45) with a ratio of length of resection to that of the whole prosthesis of > 50% were treated between May 2003 and May 2012. The mean follow-up was 47 months (15 to 132). The stem of the prosthesis was introduced through bone graft struts filled with cement, then cemented into the residual bone. Bone healing was achieved in 23 patients (86%). The mean Musculoskeletal Tumour Society functional score was 85% (57 to 97). The five-year survival rate of the endoprostheses was 81% (95% confidence intervals 67.3 to 92.3). The mean length of devitalised autografts and allografts was 8.6 cm (5 to 15), which increased the ratio of the the length of the stem of the prosthesis to that of the whole length of the prosthesis from a theoretical 35% to an actual 55%.
Cortical strut bone grafting and long-stem endoprosthetic reconstruction is an option for treating massive segmental defects following resection of a tumour in the lower limb. Patients can regain good function with a low incidence of aseptic loosening. The strut graft and the residual bone together serve as a satisfactory bony environment for a revision prosthesis, if required, once union is achieved.
Cite this article:
Objectives
This study tests the biomechanical properties of adjacent locked plate constructs in a femur model using Sawbones. Previous studies have described biomechanical behaviour related to inter-device distances. We hypothesise that a smaller lateral inter-plate distance will result in a biomechanically stronger construct, and that addition of an anterior plate will increase the overall strength of the construct.
Methods
Sawbones were plated laterally with two large-fragment locking compression plates with inter-plate distances of 10 mm or 1 mm. Small-fragment locking compression plates of 7-hole, 9-hole, and 11-hole sizes were placed anteriorly to span the inter-plate distance. Four-point bend loading was applied, and the moment required to displace the constructs by 10 mm was recorded.
We are currently facing an epidemic of periprosthetic fractures around the hip. They may occur either during surgery or post-operatively. Although the acetabulum may be involved, the femur is most commonly affected. We are being presented with new, difficult fracture patterns around cemented and cementless implants, and we face the challenge of an elderly population who may have grossly deficient bone and may struggle to rehabilitate after such injuries. The correct surgical management of these fractures is challenging. This article will review the current choices of implants and techniques available to deal with periprosthetic fractures of the femur.
Cite this article:
We describe the clinical and radiological results of cementless primary total hip replacement (THR) in 25 patients (18 women and seven men; 30 THRs) with severe developmental dysplasia of the hip (DDH). Their mean age at surgery was 47 years (23 to 89). In all, 21 hips had Crowe type III dysplasia and nine had Crowe type IV. Cementless acetabular components with standard polyethylene liners were introduced as close to the level of the true acetabulum as possible. The modular cementless S-ROM femoral component was used with a low resection of the femoral neck.
A total of 21 patients (25 THRs) were available for review at a mean follow-up of 18.7 years (15.8 to 21.8). The mean modified Harris hip score improved from 46 points pre-operatively to 90 at final follow up (p < 0.001).
A total of 15 patients (17 THRs; 57%) underwent revision of the acetabular component at a mean of 14.6 years (7 to 20.8), all for osteolysis. Two patients (two THRs) had symptomatic loosening. No patient underwent femoral revision. Survival with revision of either component for any indication was 81% at 15 years (95% CI 60.1 to 92.3), with 21 patients at risk.
This technique may reduce the need for femoral osteotomy in severe DDH, while providing a good long-term functional result.
Cite this article:
Peri-prosthetic fractures of the femur around a THA remain challenging injuries to treat. The Vancouver Classification helps to guide decision making, and is based on fracture location, implant fixation status, and remaining bone quality. It is critical to determine fixation status of the implant, even if surgical dislocation is necessary. Type A fractures involve the trochanters, and are usually due to osteolysis. Revision of the bearing surface and bone grafting of the lesions can be effective. Type B1 fractures occur around a well fixed stem, typically at the stem tip. Internal fixation with laterally based locked cable plates is effective. Optimising proximal fixation is important, typically with locked screws and cables.
Vancouver A: If minimal displacement and prosthesis stable can treat nonoperatively. If displacement is unacceptable and/or osteolysis is present consider surgery. AL: Rare, avulsions from osteopenia and lysis. If large, displaced and include large portion of calcar-can destabilise stem and prompt femoral revision. AG: More common. Often secondary to lysis. Does not usually affect implant stability. Minimal displacement. Treat closed × 3 months. Revise later is needed to remove the particle generator, debride defects and bone graft. Displaced with good host bone stock. Consider early ORIF and bone grafting. Vancouver B:. B1: Rarely non-operative. ORIF with femoral component retention. Need to carefully identify stem fixation. B2's classified as B1's are doomed to fail. B1's correctly identified treated with plate,
Treatment of an infected total elbow replacement (TER) is often successful in eradicating or suppressing the infection. However, the extensor mechanism may be compromised by both the infection and the surgery. The goal of this study was to assess triceps function in patients treated for deep infection complicating a TER. Between 1976 and 2007 a total of 217 TERs in 207 patients were treated for infection of a TER at our institution. Superficial infections and those that underwent resection arthroplasty were excluded, leaving 93 TERs. Triceps function was assessed by examination and a questionnaire. Outcome was measured using the Mayo Elbow Performance Score (MEPS).
Triceps weakness was identified in 51 TERs (49 patients, 55%). At a mean follow-up of five years (0.8 to 34), the extensor mechanism was intact in 13 patients, with the remaining 38 having bone or soft-tissue loss. The mean MEPS was 70 points (5 to 100), with a mean functional score of 18 (0 to 25) of a possible 25 points.
Infection following TER can often be eradicated; however, triceps weakness occurs in more than half of the patients and may represent a major functional problem.
Cite this article:
Tapered, fluted, modular, titanium stems are increasingly popular in the operative management of Vancouver B2 and selected B3 peri-prosthetic femoral fractures. We have reviewed the results at our institution looking at stem survival and clinical outcomes and compared this with reported outcomes in the literature. Stem survival at a mean of 54 months was 96% in our series and 97% for combined published cases. Review of radiology showed maintenance or improvement of bone stock in 89% of cases with high rates of femoral union. Favourable clinical outcome scores have reported by several authors. No difference in survival or clinical scores was observed between B2 and B3 fractures. Tapered stems are a useful option in revision for femoral fracture across the spectrum of femoral bone deficiency.
Cite this article:
Aim:. Historically, anterior decompression followed by posterior fusion has been the surgical management of choice in spinal tuberculosis. Due to theatre time being at a premium, we have evolved to performing anterior only debridement,
Infected periprosthetic fractures around total hip arthroplasties are increasingly common and extremely challenging problem. The purpose of the study was to review the experience of two tertiary referral units managing infected periprosthetic femoral fractures using interlocking long-stem femoral prostheses either as temporary functional spacers or as definitive implants. Methods. A prospective review of 19 patients managed at two tertiary referral units between 2000 and 2011. Each patient was diagnosed and managed according to similar institutional protocols. Investigation through aspiration and biopsy of periprosthetic tissue supplemented haematological tests to confirm infection. The Cannulock uncoated stem was used in 14 cases, and the Kent hip prosthesis in 5 cases.
Tapered, fluted, modular, titanium stems have a long history in Europe and are increasing in popularity in North America. We have reviewed the results at our institution looking at stem survival and clinical outcomes. Radiological outcomes and quality of life assessments have been performed and compared to cylindrical non-modular cobalt chromium stems. Survival at five years was 94%. This fell to 85% at ten years due to stem breakage with older designs. Review of radiology showed maintenance or improvement of bone stock in 87% of cases. Outcome scores were superior in tapered stems despite worse pre-operative femoral deficiency. Tapered stems have proved to be a useful alternative in revision total hip arthroplasty across the spectrum of femoral bone deficiency.
We evaluated all cases involving the combined use of a subtrochanteric derotational femoral shortening osteotomy with a cemented Exeter stem performed at our institution. With severe developmental dysplasia of the hip an osteotomy is often necessary to achieve shortening and derotation of the proximal femur. Reduction can be maintained with a 3.5 mm compression plate while the implant is cemented into place. Such a plate was used to stabilise the osteotomy in all cases. Intramedullary autograft helps to prevent cement interposition at the osteotomy site and promotes healing. There were 15 female patients (18 hips) with a mean age of 51 years (33 to 75) who had a Crowe IV dysplasia of the hip and were followed up for a mean of 114 months (52 to 168). None was lost to follow-up. All clinical scores were collected prospectively. The Charnley modification of the Merle D’Aubigné-Postel scores for pain, function and range of movement showed a statistically significant improvement from a mean of 2.4 (1 to 4), 2.3 (1 to 4), 3.4 (1 to 6) to 5.2 (3 to 6), 4.4 (3 to 6), 5.2 (4 to 6), respectively. Three acetabular revisions were required for aseptic loosening; one required femoral revision for access. One osteotomy failed to unite at 14 months and was revised successfully. No other case required a femoral revision. No postoperative sciatic nerve palsy was observed.
Cemented Exeter femoral components perform well in the treatment of Crowe IV dysplasia with this procedure.
We report the use of an allograft prosthetic composite for reconstruction of the skeletal defect in complex revision total hip replacement for severe proximal femoral bone loss. Between 1986 and 1999, 72 patients (20 men, 52 women) with a mean age of 59.9 years (38 to 78) underwent reconstruction using this technique.
At a mean follow-up of 12 years (8 to 20) 57 patients were alive, 14 had died and one was lost to follow-up. Further revision was performed in 19 hips at a mean of 44.5 months (11 to 153) post-operatively. Causes of failure were aseptic loosening in four, allograft resorption in three, allograft nonunion in two, allograft fracture in four, fracture of the stem in one, and deep infection in five. The survivorship of the allograft-prosthesis composite at ten years was 69.0% (95% confidence interval 67.7 to 70.3) with 26 patients remaining at risk. Survivorship was statistically significantly affected by the severity of the pre-operative bone loss (Paprosky type IV; p = 0.019), the number of previous hip revisions exceeding two (p = 0.047), and the length of the allograft used (p = 0.005).