Clinical graphics allows creation of three dimensional simulation based on CT or MRI that allows pre-operative planning. The software reports several hip morphological parameters routinely. Our aim was to validate the measurements of acetabular morphological parameters using CT based clinical graphics in patients presenting with symptomatic hip pain. We reviewed standardised plain radiographs, CT scans and 3D clinical graphics outputs of 42 consecutive hips in 40 patients presenting with symptomatic hip pain. Acetabular index (AI), lateral centre edge angle (LCE), acetabular and femoral version measurements were analysed for the 3D clinical graphics with radiographs and CT as gold standard. Significant differences were noted in measurements of AI, LCE, acetabular version and femoral version using the 3D motion analysis versus conventional measures, with only acetabular version showing comparable measurements. Correlation between 3D clinical graphics and conventional measures of acetabular morphology (AI, LCE) showed only slight agreement (ICC <0.4); while substantial agreement was noted for acetabular and femoral version (IC > 0.5). Acetabular morphological parameters measured by 3D clinical graphics are not reliable or validated. While clinicians may pursue the use of 3D clinical graphics for preoperative non-invasive planning, caution should be exercised when interpreting the reports of hip morphological parameters such as AI and LCE.
One of the key factors responsible for altered kinematics and joint stability following contemporary total knee arthroplasty (TKA) is resection of the anterior cruciate ligament (ACL). Therefore, retaining the ACL is often considered to be the “holy grail” of TKA. However, ACL retention can present several technical challenges, and in some cases may not be viable due to an absent or non-functional ACL. Therefore, the goal of this research was to investigate whether substitution of ACL function through an anterior post mechanism could improve kinematic deficits of contemporary posterior cruciate ligament (PCL) retaining (CR) implants. This was done using KneeSIM, a previously established
Background. Over 10% of total hip arthroplasty (THA) surgeries performed in England and Wales are revision procedures. 1. Malorientation of the acetabular component in THA may contribute to premature failure. Yet with increasingly younger populations receiving THA surgery (through higher incidences of obesity) and longer life expectancy in general, the lifetime of an implant needs to increase to avoid a rapid increase in revision surgery in the future. The Evaluation of X-ray, Acetabular Guides and Computerised Tomography in THA (EXACT) trial is assessing the pelvic tilt of a patient by capturing x-rays from the patient in sitting, standing and step-up positions. It uses this information, along with a CT scan image, to deliver a personalised
Introduction. Improper seating during head/stem assembly can lead to unintended micromotion between the femoral head and stem taper—resulting in fretting corrosion and implant failure. There is no consensus—either by manufacturers or by the surgical community—on what head/stem taper assembly method maximizes modular junction stability in total hip arthroplasty (THA). A 2018 clinical survey found that orthopedic surgeons prefer applying one strike or three, subsequent strikes when assembling head/stem taper. However, it has been suggested that additional strikes may lead to decreased interference. Additionally, the taper surface finish—micro-grooves—has been shown to affect taper interference and may be influenced by assembly method. Objective. The objective of this study was to employ a novel, micro-grooved finite element (FEA) model of the hip taper interface and assess the role of head/stem assembly method—one vs three strikes—on modular taper junction stability. Methods. A two-dimensional, axisymmetric model representative of a CoCrMo femoral head taper and Ti6Al4V stem taper was created using median geometrical measurements taken from over 100 retrieved implants. Surface finish—micro-grooves—of the head/stem taper were modeled using a sinusoidal function with amplitude and period corresponding to median retrieval measurements of micro-groove height and spacing, respectively (“smooth” stem taper: height=2µm, spacing=50µm; “rough” stem taper: height=11µm, spacing=200µm; head taper: height=2µm, spacing=50µm). All models had a 3’ (0.05°), proximal-locked angular mismatch between the tapers. To simulate modular assembly during surgery, multiple dynamic loads (4kN, 8kN, and 12kN) were applied to the femoral head taper as either one or three sequence of strikes. The input load profile (Figure 1) used for both cases was collected from surgeons assembling an experimental setup with a three-dimensional load sensor. Models were assembled and meshed in ABAQUS Standard (v 6.17) using four-node linear hexahedral, reduced integration elements. Friction was modeled between the stem and head taper using surface-to-surface formulation with penalty contact (µ=0.2). A total of 12 implicit,
Introduction. Improper seating during head/stem assembly can lead to unintended micromotion between the femoral head and stem taper—resulting in fretting corrosion and implant failure. 1. There is no consensus—either by manufacturers or by the surgical community—on what head/stem taper assembly method maximizes modular junction stability in total hip arthroplasty (THA). A 2018 clinical survey. 2. found that orthopedic surgeons prefer applying one strike or three, subsequent strikes when assembling head/stem taper. However, it has been suggested that additional strikes may lead to decreased interference strength. Additionally, the taper surface finish—micro-grooves—has been shown to affect taper interference strength and may be influenced by assembly method. The objective of this study was to employ a novel, micro-grooved finite element (FEA) model of the hip taper interface and assess the role of head/stem assembly method—one vs three strikes—on modular taper junction stability. Methods. A two-dimensional, axisymmetric FEA model representative of a CoCrMo femoral head taper and Ti6Al4V stem taper was created using median geometrical measurements taken from over 100 retrieved implants. 3. Surface finish—micro-grooves—of the head/stem taper were modeled using a sinusoidal function with amplitude and period corresponding to retrieval measurements of micro-groove height and spacing, respectively. Two stem taper micro-groove geometries— “rough” and “smooth”—were modeled corresponding to the median and 5. th. percentile height and spacing measurements from retrievals. All models had a 3' (0.05°), proximal-locked angular mismatch between the tapers. To simulate implant assembly during surgery, multiple dynamic loads (4kN, 8kN, and 12kN) were applied to the femoral head taper in a sequence of one or three strikes. The input load profile (Figure 1) used for both cases was collected from surgeons assembling an experimental setup with a three-dimensional load sensor. Models were assembled and meshed in ABAQUS Standard (v 6.17) using four-node linear hexahedral, reduced integration elements. Friction was modeled between the stem and head taper using surface-to-surface formulation with penalty contact (µ=0.2). A total of 12 implicit,
Posterior stabilized (PS) total knee arthroplasty (TKA), wherein mechanical engagement of the femoral cam and tibial post prevents abnormal anterior sliding of the knee, is a proven surgical technique. However, many patients complain about abnormal clicking sensation, and several reports of severe wear and catastrophic failure of the tibial post have been published. In addition to posterior cam-post engagement during flexion, anterior engagement with femoral intercondylar notch can also occur during extension. The goal of this study was to use
Modeling the kinetic effects of the soft tissue structures is a major challenge for
Background. Stemless prostheses are recognized to be an effective solution for anatomic total shoulder arthroplasty (TSA) while providing bone preservation and shortest operating time. Reverse shoulder arthroplasty (RSA) with stemless has not showed the same effectiveness, as clinical and biomechanical performances strongly depend on the design. The main concern is related to stability and bone response due to the changed biomechanical conditions; few studies have analyzed these effects in anatomic designs through Finite Element Analysis (FEA), however there is currently no study analyzing the reverse configuration. Additionally, most of the studies do not consider the effect of changing the neck-shaft angle (NSA) resection of the humerus nor the proper assignment of spatial bone properties to the bone models used in the simulations. The aim of this FEA study is to analyze bone response and primary stability of the SMR Stemless prosthesis in reverse with two different NSA cuts and two different reverse angled liners, in bone models with properties assigned using a quantitative computed tomography (QCT) methodology. Methods. Sixteen fresh-frozen cadaveric humeri were modelled using the QCT-based finite element methodology. The humeri were CT-scanned with a hydroxyapatite phantom to allow spatial bone properties assignment [Fig. 1]. Two implanted SMR stemless reverse configurations were considered for each humerus: a 150°-NSA cut with a 0° liner and a 135°-NSA cut with a 7° sloped liner [Fig. 2]. A 105° abduction loading condition was simulated on both the implanted reverse models and the intact (anatomic) humerus; load components were derived from previous
Introduction. Dual Mobility (DM) implants have gained popularity for the treatment and prevention of hip dislocation, with increased stability provided by a large diameter mobile liner. However, distal regions of the liner can impinge on soft-tissues like hip capsule and iliopsoas, leading to anterior hip pain. Additionally, soft-tissue impingement may trap the mobile liner, leading to excessive loading of the liner rim, from engagement with the femoral stem, and subsequent intra-prosthetic dislocation. The hypothesis of this study was that reducing the liner profile below the equator (contoured design) can mitigate soft-tissue impingement without compromising inner-head pull-out resistance and overall hip joint stability (Fig. 1). Methods. The interaction of conventional and contoured liners with anterior soft-tissues was evaluated in 10 cadaveric hips (5 specimens; 2 male, 3 female; age 65 ± 10 yrs; liner diameter 42–48mm) via visual observation and fluoroscopic imaging. A metal wire was sutured to the deep fibers of the iliopsoas tendon/muscle, and metal wires were embedded in the mobile liners for fluoroscopic visualization (Fig. 2). All soft-tissue except the anterior hip capsule and iliopsoas was removed, and a rope was attached to the iliopsoas to apply tension along its natural orientation. Resistance to inner-head pull-out was evaluated via Finite Element Analysis (FEA) by simulating a full cycle of insertion of the inner head into the mobile liner and subsequent pullout. The femoral head, acetabular shell, and stem were modeled as rigid, while the mobile liner was modeled as plastically deformable. Hip joint stability was evaluated by
Introduction:. Large diameter femoral heads have been used successfully to prevent dislocation after Total Hip Arthroplasty (THA). However, recent studies show that the peripheral region of contemporary femoral heads can directly impinge against the native soft-tissues, particularly the iliopsoas, leading to activity limiting anterior hip pain. This is because the spherical articular surface of contemporary prosthesis overhangs beyond that of the native anatomy (Fig. 1). The goal of this research was to develop an anatomically shaped, soft-tissue friendly large diameter femoral head that retains the benefits of contemporary implants. Methods:. Various Anatomically Contoured femoral Head (ACH) designs were constructed, wherein the articular surface extending from the pole to a theta (θ) angle, matched that of contemporary implants (Fig. 2). However, the articular surface in the peripheral region was moved inward towards the femoral head center, thereby reducing material that could impinge on the soft-tissues (Fig. 1 and Fig. 2). Finite element analysis was used to determine the femoroacetabular contact area under peak in vivo loads during different activities.
INTRODUCTION. Femoral head diameter has a major influence on stability and dislocation resistance of the hip joint after Total Hip Arthroplasty (THA). Dual Mobility (DM) implants can also reduce the risk of dislocation due the large diameter mobile liner which forms the femoroacetbular articulation. However, recent studies have shown that large head prostheses can directly impinge against native soft tissues, particularly the iliopsoas, leading to anterior hip pain. Dual mobility systems have emerged as a revision option in the treatment of failed metal on metal devices because of the high incidence of post revision instability secondary to abductor loss and need for capsulectomy. We hypothesized that an Anatomically Contoured Dual Mobility (ACDM) liner could provide joint stability while better accommodating the soft tissues surrounding the hip joint. METHODS. The dislocation resistance of a 44 mm ACDM implant was compared to that of a 44 mm conventional DM liner. Both implants consisted of a 28 mm inner small diameter head and the liner was abducted to be in the worst case position for dislocation (Fig. 1). The ACDM liner was based on a 44 mm sphere with smaller radii used to contour the peripheral region below the equator of the liner. MSC Adams was used for
INTRODUCTION. Femoral head diameter has a major influence on stability and dislocation resistance after Total Hip Arthroplasty (THA). Although routine use of large heads is common, several recent studies have shown that contemporary large head prostheses can directly impinge against native soft tissues, particularly the iliopsoas which wraps around the femoral head, leading to refractory anterior hip pain. To address this, we developed a novel Anatomically Contoured large diameter femoral Head (ACH). We hypothesized that anatomical contouring of the ACH implant for soft tissue relief would not compromise dislocation resistance, and the ACH implant would provide increased stability compared to small heads. METHODS. In this study the dislocation resistance of a 36 mm ACH was compared to that of 28 mm and 36 mm contemporary heads. The ACH implant was based on a 36 mm sphere with smaller radii used to contour the peripheral region below the equator of the head. MSC Adams was used for
Introduction. Many finite element (FE) studies have been performed in the past to assess the biomechanical performance of TKA and THA components. The boundary conditions have often been simplified to a few peak loads. With the availability of personalized musculoskeletal (MS) models we becomes possible to estimate dynamic muscle and prosthetic forces in a patient specific manner. By combining this knowledge with FE models, truly patient specific failure analyses can be performed. In this study we applied this combined technique to the femoral part of a cementless THR and calculated the cyclic micro-motions of the stem relative to the bone in order to assess the potential for bone ingrowth. Methods. An FE model of a complete femur with a CLS Spotorno stem inserted was generated. An ideal fit between the implant and the bone was modeled proximally, whereas distally an interface gap of 100μm was created to simulate a more realistic interface condition obtained during surgery. Furthermore, a gait analysis was performed on a young subject and fed into the Anybody™ MS modeling system. The anatomical data set (muscle attachment points) used by the Anybody™ system was morphed to the shape of the femoral reconstruction. In this way a set of muscle attachment points was obtained which was consistent with the FE model. The predicted muscle and hip contact forces by the Anybody™ modeling system were dynamic and divided into 37 increments including two stance phases and a swing phase of the right leg. Results. The magnitude and path of interface micromotions was heavily dependent on the location on the implant. In the proximal region, a unidirectional pattern was visible in proximal-distal direction (max. motion was 39μm). Mid stem micromotions were very small (in the order of 4μm), whereas in the distal region, micromotions had a tendency to develop in anterior-posterior and medial-lateral direction (max. motion was 96μm). Hence, in this example, ingrowth is most likely to start in the mid-region. Conclusion. By combining finite element models with musculoskeletal models more realistic,
Currently, standard total knee arthroplasty (TKA) procedures focus on axial and rotational alignment of the prosthesis components and ligament balancing. Even though TKA has been constantly improved, TKA patients still experience a significantly poorer functional outcome than total hip arthroplasty patients. Among others, complications can occur when knee kinematics (active/passive) after TKA do not correspond with the physiological conditions. We hypothesised that the Q-angle has a substantial impact on active joint kinematics and should be taken into account in TKA. The Q-angle can be influenced by the position of the tibial tuberosity (TT). A pathological position of the TT is commonly related to patellofemoral pain and knee instability. A clinically well accepted surgical treatment is the TT medialisation which causes a change in the orientation of the patella tendon and thus alters the biomechanics of the knee. If active and passive knee kinematics differs, this aspect should be considered for implant design and positioning. Therefore we investigated the sensitivity of active knee kinematics related to the position of the TT by using a complex multi-body model with a
INTRODUCTION. Patellofemoral joint (PFJ) replacement is a successful treatment option for isolated patellofemoral osteoarthritis. With this approach only the involved joint compartment is replaced and the femoro-tibial joint remains intact. Minimizing periprosthetic bone loss, which may occur due to the stress shielding effect of the femoral component, is important to insure long-term outcomes. The objective of this study was to investigate, using finite element analyses, the effects of patellofemoral replacement on the expected stress distribution of the distal femur eventually leading to changes in bone density. METHODS. MRI images of a healthy knee were acquired, segmented and reconstructed into a 3D physiological model of the bony and cartilaginous geometries of distal femur and patella with patellar tendon and insertion of the quadriceps tendon. This model was modified to include PFJ replacements with either a Journey PFJ or a Richards II PFJ prosthesis, and a Genesis II TKA (Smith&Nephew, Memphis, TN). The prosthetic components were incorporated in the intact model based on the manufacturer's instructions or previously described surgical techniques (Figure 1). Cortical bone was modeled with orthotropic properties, while homogeneous linear isotropic elasticity was assumed for trabecular bone, cartilage, cement and femoral components materials. The patellar tendon was given Neo-Hookean behavior. UHMWPE patellar buttons for all designs were assigned non-linear elasto-plastic material. The simulated motion consisted of a 10 second loaded squat, starting from 0° until a flexion angle of 120° matching experimental kinematics tests performed in previous in-vitro analysis on physiological cadaveric legs [1-2]. The patella model was constrained fixing the distal part of the patellar ligament and applying a quadriceps force distributed on the quadriceps insertion on the proximal surface of the patella. During the
Background. Accurate acetabular cup positioning is considered to be essential to prevent postoperative dislocation and improve the long-term outcome of total hip arthroplasty (THA). Recently various devices such as navigation systems and patient-specific guides have been used to ensure the accuracy of acetabular cup positioning. Objectives. The present study evaluated the usefulness of CT-based three-dimensional THA preoperative planning for acetabular cup positioning. Methods. This study included 120 hips aged mean 68.3 years, who underwent primary THA using CT-based THA preoperative planning software ZedHip® (LEXI, Tokyo Japan) and postoperative CT imaging (Fig.1). The surgical approach adopted the modified Watson-Jones approach in the lateral decubitus position and Trident HA acetabular cups were used for all cases. Preoperatively the optimum cup size and position in the acetabular were decided using the ZedHip® software, taking into consideration femoral anteversion and to achieve the maximum range of motion in
Introduction:. While survivorship of total knee arthroplasty (TKA) is excellent, up to 25% of patients remain dissatisfied with their outcome [1, 2]. Knee instability, which is common during high demand activities, contributes to patient dissatisfaction [3]. As younger patients undergo TKA, longevity requirements and functional demands will rise [4]. Design factors influence the functional outcome of the procedure [5, 6], although in clinical studies it can be difficult to distinguish joint mechanics differences between designs due to confounding variability in patient-related factors. The objective of the current study was to assess the stability and mechanics of several current TKA designs during high-demand dynamic activities using a computational model of the lower limb. Methods:. Three high-demand dynamic activities (gait, stepdown, squat) were simulated in a previously described lower limb model (Fig. 1) [7]. The model included calibrated tibiofemoral (TF) soft-tissue structures, patellofemoral (PF) ligaments and extensor mechanism [8]. Loading conditions for the simulations were derived from telemetric patient data in order to evaluate TKA designs under physiological kinematic and loading conditions [7, 9]. Four fixed-bearing TKA designs (both cruciate-retaining (CR) and posterior-stabilizing (PS) versions) were virtually implanted into the lower limb model and joint motion, contact mechanics and interface loads were evaluated during simulation of each dynamic activity. Results:. The range of anterior-posterior (A-P) and internal-external (I-E) motion for the least stable design was twice that of the most stable design during dynamic activity (Fig. 2). The increased anterior stability on some components did not translate directly to the largest bone-implant interface shear loading, which was dependent on coronal plane conformity and I-E torque, as component ranking varied throughout activity (Fig. 2, 3). Current designs varied substantially in conformity, resulting in reduced contact area and increased contact pressures with low-conformity articulation (Fig. 2). Discussion:. While contemporary TKA designs all have good survivorship, there exists significant design differences related to the inherent stability of the articulating surfaces which result in kinematic differences during simulated high demand activities. The
INTRODUCTION. Contemporary PCL sacrificing Total Knee Arthroplasty (TKA) implants (CS) consist of symmetric medial and lateral tibial articular surfaces with high anterior lips designed to substitute for the stability of the native PCL. However, designs vary significantly across implant systems in the level of anteroposterior constraint provided. Therefore, the goal of this study was to investigate kinematics of two CS designs with substantially different constraint levels. The hypothesis was that