Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_3 | Pages 95 - 95
23 Feb 2023
Grupp T Reyna AP Bader U Pfaff A Mihalko W Fink B
Full Access

ZrN-multilayer coating is clinically well established in total knee arthroplasty [1-3] and has demonstrated significant reduction in polyethylene wear and metal ion release [4,5].

The goal of our study was to analyze the biotribological behaviour of the ZrN-multilayer coating on a polished cobalt-chromium cemented hip stem.

CoCr28Mo6 alloy hip stems with ZrN-multilayer coating (CoreHip®AS) were tested versus an un-coated version. In a worst-case-scenario the stems with ceramic heads have been tested in bovine serum in a severe cement interface debonding condition under a cyclic load of 3,875 N for 15 million cycles. After 1, 3, 5, 10 & 15 million cycles the surface texture was analysed by scanning-electron-microscopy (SEM) and energy-dispersive x-ray (EDX). Metal ion concentration of Co,Cr,Mo was measured by inductively coupled plasma mass spectroscopy (ICP-MS) after each test interval.

Based on SEM/EDX analysis, it has been demonstrated that the ZrN-multilayer coating keeps his integrity over 15 million cycles of severe stem cemented interface debonding without any exposure of the CoCr28Mo6 substrate.

The ZrN-multilayer coated polished cobalt-chromium cemented hip stem has shown a reduction of Co & Cr metal ion release by two orders of a magnitude, even under severe stem debonding and high interface micro-motion conditions.

ZrN-multilayer coating on polished cobalt-chromium cemented hip stems might be a suitable option for further minimisation of Co & Cr metal ion release in total hip arthroplasty. Clinical evidence has to be proven during the next years.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 142 - 142
1 Apr 2019
Higgs G Rimnac C Mihalko W Gilbert J Kurtz S
Full Access

Introduction

Corrosion at modular junctions of total hip replacements has been identified as a potential threat to implant longevity, resulting in efforts to determine appropriate countermeasures. Visual scoring and volumetric material loss measurements have been useful tools to elucidate various clinical and design factors associated with corrosion damage. However, corrosion involves electron exchange that results in chemical changes to biomedical alloys, and electrochemical assessment may therefore be a more appropriate approach to understand the phenomenon. The purpose of this pilot study was to electrochemically distinguish the severity of corrosion in retrieved femoral heads. A secondary goal was to identify the potential of electrochemical impedance spectroscopy (EIS) as a method to identify different forms of corrosion damage.

Methods

Twenty femoral heads were identified from a larger study of total hip replacements, obtained as part of an ongoing multi- center IRB-approved retrieval program. Using a previously established 4-point scoring method, components were binned by taper damage: 10 components were identified as having severe damage, 7 with moderate damage and 2 with mild damage. One (1) unimplanted control was included to represent minimal corrosion damage. All components were then characterized using electrochemical impedance spectroscopy under the frequency domain: a 10 mV sinusoidal voltage, ranging from 20 kHz to 2 mHz, was applied to the taper of a femoral head (working electrode) filled with a 1M solution of PBS, a platinum counter electrode and a chlorided silver reference electrode. Absolute impedance at 2 mHz (|Z0.002|), and max phase angle (θ) were assessed relative to taper damage severity. After least-squares fitting of the EIS data to a Randles circuit with a constant phase element, circuit elements: polarization resistance (Rp), CPE-capacitance, and CPE-exponent were also evaluated. The seven (7) most severely corroded components were further examined with scanning electron microscopy to identify corrosion modes. For all statistical analyses, significance was determined at alpha=0.05.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 12 - 12
1 Mar 2017
Wodowski D Kerkhof A Mihalko W
Full Access

Introduction

Gelsolin (GSN) is a protein whose function in the cytoplasm is to regulate intracellular actin assembly, while circulating plasma GSN has been implicated in the clearance of cellular byproducts via the body's scavenger system. In knee synovial fluid, GSN appears to be protective of inflammation as lower levels have been found in patients with rheumatoid arthritis while higher levels have been demonstrated in hypersensitivity reactions to metallic implants. The purpose of this study is to define the role of GSN in painful total knee arthroplasty (TKA) patients as a biomarker to distinguish septic from aseptic diagnoses.

Methods

After Institutional Review Board (IRB) approval, 50 patients were enrolled in the study into two cohorts. 25 patients presented with a painful TKA and 25 patients had a painful native knee with primary osteoarthritis (OA). Synovial fluid was obtained from each patient's affected knee. Appropriate lab and culture data was also obtained from the painful TKA group. An ELISA was used to determine GSN levels and the groups were compared. Two tailed Student's t tests were used to compare means while Pearson's Correlation Coefficient and linear regression analyses were used to determine association between laboratory findings and GSN levels.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 14 - 14
1 Mar 2017
Mihalko W Jiao Y Kerkhof A Yan J Hallock J Gu W
Full Access

INTRODUCTION

Since the recall of some metal on metal (MoM) THR bearings, surgeons have seen patients with pain, elevated Co and Cr levels and adverse local tissue reactions (ALTR). While many variables may contribute to THR MoM failures, many times these variables are not present in patients who present with symptoms. We investigate the possible genetic predilection of a group of patients who were revised after MoM THR surgery for pain, high Co/Cr levels and ALTR.

METHODS

IRB approval was obtained prior to our study. We have analyzed 19 control (asymptomatic MoM THR patients > 6 years after surgery) and 19 disease (revised MoM THR for high metal ions and ALTR). The 38 sample intensity files were subject to sample Quality Control (QC) using Contrast QC (< 0.4) with an Affymetrix Genotyping Console. The resulting 38 sample files with genotype calls were loaded and further analyzed using the Association Workflow in Partek Genomics Suite 6.6 (Partek, Missouri). Hardy-Weinberg equilibrium test was performed on the single nucleotide polymorphism (SNP) level. The difference between the observed and expected frequencies of each allele at each locus were tested by Fisher's exact test and χ2 test. To get the working SNP list, two filters were used: (1) a SNP no-call rate should be less than 5%, and (2) minor allele frequency of a SNP should be greater than 5%.

After filtering, association analysis of the SNPS with disease was done using Chi2 Test. In this study, χ2 statistic was used to assess the difference in allele frequencies between the control and disease samples. The value of χ2 statistic, degrees of freedom, and the associated p-value for each SNP were calculated. Dot Plot was used to visualize the genotypes of all samples.

To measure the non-random association of alleles at different loci, Linkage Disequilibrium analysis was performed using the neighborhood size of 20 and statistic r2. The resulting correlations show the value of r2 for SNPs. The r2 = 1 means that two SNPs are tightly associated.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 91 - 91
1 Mar 2017
Wells C Feldman J Timmerman I Chow J Mihalko W Neel M Jennings J Haggard W
Full Access

Introduction

Decreasing tissue damage and recovery time, while improving quality of life have been the focus of many approaches to total hip arthroplasty (THA). In this study, we compared two approaches, a tissue-sparing superior capsulotomy percutaneously assisted approach (SP) and the traditional posterior approach (TR), to address the question of whether the novel technique reduces tissue damage. The secondary aim of this study focused on the measurement technologies utilized to quantify the damage resulting from either SP or TR. Image J, BioQuant, and cellSens were the image analysis programs employed. Statistical validation and comparisons of results between all platforms were performed.

Methods

Both hips of freshly frozen cadaveric specimens (n = 8) were surgically prepared for THA with random procedure performed on left or right hip. All selected specimens had no prior implantation of devices to ensure all observed muscle damage occurred from the surgical technique. Surgeons resected tissue and performed necessary procedural steps up to device implantation. No devices were implanted during the study, as the aim was to quantify the damage caused by the incision and resection. After completion of the surgery, an independent surgeon (IS), who was blinded as to which method was performed on the specimen, excised the muscles and inspected areas of interest Assessment of the tissue damage was executed using a midsubstance cross-sectional area technique, validated by prior studies. High-resolution images of demarcated muscles were used for quantitative analysis. Three blinded independent reviewers quantified damaged tissue. The results were used to detect if statistically significant differences were present between the two methods. Furthermore, an independent reviewer using SPSS statistical software also assessed inter-program and inter-rater reliability.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 13 - 13
1 Mar 2017
Mihalko W Lowell J Woodard E Arnholt C MacDonald D Kurtz S
Full Access

Introduction

Total knee arthroplasty (TKA) is a successful procedure for end stage arthritis of the knee that is being performed on an exponential basis year after year. Most surgeons agree that soft tissue balancing of the TKA is a paramount to provide a successful TKA. We utilized a set of retrieved lower extremities with an existing TKA to measure the laxity of the knee in all three planes to see if wear scores of the implants correlated to the laxity measured. This data has never been reported in the literature.

Methods

IRB approval was obtained for the local retrieval program. Each specimen was retrieved after removing the skin, subcutaneous tissue and muscle from mid thigh to mid tibia. The femur, tibia and fibula were then transversely cut to remove the specimen for testing. Each specimen was then imaged using a flouroscopic imaging unit (OEC, Inc) in the AP, Lateral and sunrise views. These images were used to analyze whether there were any signs of osteolysis. Each specimen was mounted into a custom knee testing machine (Little Rock AR). Each specimen then was tested at full extension, 30, 60, and 90 degrees of flexion. At each flexion angle the specimen was subjected to a 10Nm varus and valgus torque, a 1.5Nm internal and external rotational torque and a 35N anterior and posterior directed force. Each specimen's implants were removed to record manufacturer and lot numbers. Polyethylene damage scores (Hood et al. JBMR 1983) were then calculated in the medial, lateral and backside of the polyethylene insert as well as on the medial and lateral femoral condyle. (Figure 1) Correlation coefficients were then calculated to show any relationship with soft tissue balancing in all three planes and wear scores.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 15 - 15
1 Mar 2017
Mihalko W Braman M Lowell J Dopico P Zucker-Levin A
Full Access

Introduction

Early hip OA may be attributed to smaller coverage of the femoral head leading to higher loads per unit area. We hypothesize that tight hamstrings may contribute to increased loads per unit area on the femoral head during gait. When a patient has tight hamstrings they cannot flex their pelvis in a normal fashion which may result in smaller coverage of the femoral head (Figure 1). This study aimed to determine if subjects with tight hamstrings can improve femoral head coverage during gait after a stretching intervention.

Methods

Nine healthy subjects with tight hamstrings (popliteal angle>25°) were recruited and consented for this IRB approved study. Gait analysis with 58 reflective markers were placed by palpation on anatomical landmarks of the torso and lower extremities. Ten optoelectronic cameras (Qualisys, Gothenburg, Sweden) and three force plates (AMTI, Watertown, MA) were used to track marker position and measure foot strike forces. Subjects walked at a self-selected speed across the force plates until ten clean trials were performed and then were scanned with the reflective markers on the spine using an EOS (EOS Imaging, France) bi-planar x-ray system. Following testing participants completed a six week stretching program to increase hamstring length. Pelvic tilt (PT) was measured at heel strike for each trial and averaged.

Using EOS scans the femoral head radius was measured using three points that best fit the load bearing surface on the sagittal view from the anterior acetabular rim to a point on the posterior acetabulum 45 degrees from vertical. The radius of femoral head and angle of acetabular coverage were used to calculate the load bearing surface area of femoral head. Load on the femur was calculated using an Anybody lower body model (Anybody Technology, Aalborg, Denmark) and load per unit area change was compared.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 19 - 19
1 Feb 2017
Arnholt C MacDonald D Lowell J Gilbert J Mihalko W Kurtz S
Full Access

Introduction

Previous studies of CoCr alloy femoral components for total knee arthroplasty (TKA) have identified 3rd body abrasive wear, and apparent inflammatory cell induced corrosion (ICIC) [1] as potential damage mechanisms. The association between observed surface damage on the femoral condyle and metal ion release into the surrounding tissues is currently unclear. The purpose of this study was to investigate the damage on the bearing surface in TKA femoral components recovered at autopsy and compare the damage to the metal ion concentrations in the synovial fluid.

Methods

12 autopsy TKA CoCr femoral components were collected as part of a multi-institutional orthopedic implant retrieval program. The autopsy components included Depuy Synthes Sigma Mobile Bearing (n=1) and PFC (n=1), Stryker Triathlon (n=1) and Scorpio (n=3), and Zimmer Nexgen (n=4) and Natural Knee (n=2). Fluoro scans of all specimens prior to removal was carried out to assure no signs of osteolysis or aseptic loosening were present.

Third-body abrasive wear of CoCr was evaluated using a semi-quantitative scoring method similar to the Hood method [2]. ICIC damage was reported as location of affected area and confirmed using a digital optical microscope with 4000X magnification.

Synovial fluid was aspirated from the joint capsule prior to removal of the TKA device. The synovial fluid was spun at 1600 rpm for 20 minutes in a centrifuge with the cell pellet removed. The supernatant was analyzed in 1 mL quantities for ICP-MS (inductively coupled plasma mass spectrometry) by Huffman Hazen Laboratories. Data was expressed as ppb.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 127 - 127
1 Jan 2016
Woodard E Williams J Mihalko W
Full Access

Introduction

When performing total knee arthroplasty (TKA), surgeons often utilize a posterior-stabilized (PS) design which compensates for the loss of the posterior cruciate ligament (PCL). These designs attempt to replicate normal knee kinematics and loading using a cam and post to provide posterior restraint of the tibia during flexion. However, these designs may not be able to compensate for the increase in flexion space or the inherent loss of coronal stability after PCL release compared to a cruciate retaining (CR) design. This study aimed to compare stability of PS and CR TKA designs by assessing laxity in three planes.

Methods

The specimens utilized in this study were lower extremities from fresh cadavers of donors who had previously undergone a total knee replacement (Medical Education and Research Institute (Memphis, TN) and Restore Life USA (Johnson City, TN)). IRB approval was obtained prior to performing the study. Twenty-three knee specimens (8 left, 15 right) were retrieved and all skin, subcutaneous tissue and muscle was removed. The femur and tibia were cut transversely 180 mm superior and inferior to the knee joint line, respectively, and specimens were mounted in a custom knee testing machine. Specimens were tested with the knee joint at full extension and at 30, 60, and 90 degrees of flexion. Laxity was assessed at 1.5 Nm of internal and external torque and 10 Nm varus and valgus torque, as well as a 35 N anterior and posterior force. Laxity was expressed as degrees of tibial displacement in the coronal plane under a varus/valgus torque and degrees of displacement in the transverse plane under an internal/external torque, as well as mm of anterior or posterior displacement. TKA components were retrieved to determine PS or CR design and grouped accordingly.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 128 - 128
1 Jan 2016
Sanford B Williams J Huffman K Zucker-Levin A Mihalko W
Full Access

Introduction

The sit-to-stand (STS) movement is a physically demanding activity of daily living and is performed more than 50 times per day in healthy adults. Several studies have shown that the normal ‘screw-home’ mechanism is altered after total knee arthroplasty (TKA). However, these studies have been criticized due to their limitations of the movement being non-weight-bearing or atypical daily activity (lunge maneuver). We analyzed TKA subjects during a STS activity to determine if the internal-external rotation of their TKA knees differed from the knees of control subjects.

Materials and Methods

Six TKA subjects (3 M, 3 F) participated following institutional review board approval and informed consent. One subject had bilateral knee replacement. Surgery was performed by the same surgeon using the same type of implant (6 posterior-stabilized, 1 cruciate-retaining). The control group included eight healthy subjects (6 M, 2 F).

Retro-reflective markers were placed over bony landmarks of the torso, pelvis, and lower extremities, and arrays of four markers were attached to the thighs and shanks using elastic wrap. A digitizing pointer was used to create virtual markers at the anterior superior iliac spines. A nine camera video-based opto-electronic system (Qualisys) was used for 3D motion capture. Subjects were barefoot and seated on a 46 cm armless bench with one foot on each force plate (AMTI). Subjects rose from their seated position, paused, and returned to the seated position at a self-selected pace repeatedly for 30 seconds. Subjects did not use their arms to push off the bench. Only the STS portion of the task was analyzed. The start of the STS cycle was defined when the C7 marker began to move forward in the sagittal plane and ended at the point of maximum knee extension. Only the right leg of the control subjects was used for analysis.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 129 - 129
1 Jan 2016
Sanford B Williams J Zucker-Levin A Mihalko W
Full Access

Introduction

In a previous study of subjects with no history of lower extremity injury or disease we found a linear relationship between body weight and peak hip, knee, and ankle joint forces during the stance phase of gait. To investigate the effect of total knee arthroplasty (TKA) on forces in the operated joints as well as the other joints of the lower extremities, we tested TKA subjects during gait and performed inverse dynamics analyses of the results.

Materials and Methods

TKA subjects (3 M, 1 F; 58 ± 5 years; body mass index range (BMI): 26–36 kg/m2) participated in this investigation following institutional review board approval and informed consent. One subject had bilateral knee replacement. Each patient received the same implant design (4 PS, 1 CR). Data from previously tested control subjects (8 M, 4 F; 26 ± 4 years; BMI: 20–36 kg/m2) were used for comparison.

Retro-reflective markers were placed over bony landmarks of each subject. A nine-camera video-based opto-electronic system was used for 3D motion capture as subjects walked barefoot at a self-selected speed on a 10 meter walkway instrumented with three force plates. Data were imported into a 12-body segment multibody dynamics model (AnyBody Technology) to calculate joint forces. Each leg contained 56 muscles whose mechanical effect was modeled by 159 simple muscle slips, each consisting of a contractile element. The models were scaled to match each subject's anthropometry and BMI. For the control subjects, only one limb was used in determining the relationship between body mass and peak joint force at the hip, knee, and ankle. For the TKA subjects, the peak joint forces were calculated for both the TKA limb and the contralateral limb.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 590 - 590
1 Dec 2013
Woodard E Mihalko W Crockarell J Williams J
Full Access

Introduction:

Repair of the arthrotomy is a performed at the end of every total knee arthroplasty (TKA). After the arthrotomy is performed, most surgeons attempt to close the arthrotomy with the medial and lateral edges anatomically approximated. If no landmarks are made prior to performing the arthrotomy however, there is a risk that anatomic approximation may not be obtained. This study looked into the biomechanical changes in stiffness of the knee before and after a medial parapatellar approach repaired with an anatomic, and shifted capsular repair with the medial side of the arthrotomy shifted up or down when repaired to determine if capsular closure may have an effect on the stiffness of the joint.

Methods:

Fourteen cadaveric TKA specimens were retrieved through the Medical Education and Research Institute (Memphis TN). For each specimen tested, the skin and muscle tissue was removed, and the femur and tibia were cut transversely 180 mm from the joint center. Specimens were fixed in extension in a custom knee testing platform (Little Rock AR) and subjected to a 10 Nm varus and valgus torque and a 1.5 Nm internal and external rotational torque. The angle at which these moments occurred was recorded, and each test was repeated for 0, 30, 60, and 90 degrees of flexion. After tests were performed on retrieved TKA specimens, a fellowship trained orthopedic surgeon vented the knee capsule by making an incision with a number 10 scalpel blade in a horizontal nature to provide a landmark for anatomic reapproximation. Tests were repeated as before, after which the surgeon performed a standard arthrotomy and repaired it using #0 suture and a neutral alignment. Sutures were cut and the repair was repeated using upward 5 mm shift and downward 5 mm shift of the medial side of the arthrotomy during the repair. All tests were repeated after each repair technique. Any increase or decrease in laxity after capsule repair was referenced to the TKA laxity tested prior to an arthrotomy being performed.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 589 - 589
1 Dec 2013
Woodard E Mihalko W Williams J Crockarell J
Full Access

Introduction:

Soft tissue balancing is a part of every total knee arthroplasty (TKA) surgery. Traditionally, balancing the varus knee has been approached by releasing portions of the medial soft tissue sleeve in a sub-periosteal nature off of the proximal tibia, but this may lead to undue laxity or residual pain about the area the release was performed. More recently, “pie crusting” of the medial soft tissue sleeve has been used to balance the varus knee without compromising the structural integrity of the ligament. This technique may provide advantages over a sub-periosteal release by targeting only medial tight bands that can be palpated with the capsule distracted in 90 degrees of flexion and full extension. This study aims to biomechanically validate the pie crusting technique of the medial soft tissue sleeve and compare the results to those of standard medial releases that have been previously reported.

Methods

Six cadaveric TKA specimens were retrieved through the Medical Education and Research Institute (Memphis TN). For each specimen tested, the skin and muscle tissue was removed, and the femur and tibia were cut transversely 180 mm from the joint center. Specimens were fixed in extension in a custom knee testing platform (Little Rock AR) and subjected to a 10 Nm varus and valgus torque. The angle at which these moments occurred was recorded, and each test was repeated for 0, 30, 60, and 90 degrees of flexion. After tests were performed on TKA specimens, a fellowship trained orthopedic surgeon performed “pie crusting,” making alternating stab patterns with a number 11 scalpel blade along the anterior half of the superficial medial collateral ligament (SMCL) or posterior half of the SMCL including the posterior oblique ligament (POL). Three specimens had the anterior capsule pie crusted first and three had the posterior pie crusting performed first, followed by complete pie crusting. After two stages of pie crusting, the medial soft tissue sleeve was released off of the proximal tibia in a sub-periosteal fashion for comparison. Laxity was defined as the angles at which valgus torque equaled 10 Nm. Any increase or decrease in laxity was referenced to the normal TKA laxity.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 339 - 339
1 Dec 2013
Hakki S Mihalko W Saleh KJ
Full Access

Introduction

The prevalence of reversing of extension coronal deformity during flexion and how that may change the routine algorithm of soft tissue balancing in total knee arthroplasty (TKA) has not been published. We name this phenomenon, the reversing coronal deformity (RCD). We observed 12% (45 patients) of coronal deformities consistently reverse in flexion in the osteoarthritic knees before surgery. We conclude that RCD phenomena need to be addressed in every TKA and collateral ligament release need to be modified or avoided; otherwise postoperative flexion instability may be inevitable. Femoral rotation adjustment with posterior capsule release has to be attempted first in RCD patients.

Method

We define RCD as the reversing of a coronal extension deformity of more than 2° while the knee reaches 90°of flexion. That is to say a 2° or more varus knee in extension becomes a 2° or more valgus at 90° of flexion or vice versa.

We retrospectively analyzed, in a multicenter study the alignment patterns of 387 (US = 270, UK = 117) consecutive computer navigated TKA subjects (June 2004–May 2008). 364/387 (US = 252, UK = 112) subjects were eligible for analysis (23 subjects had incomplete data: US = 18, UK = 5). The coronal deformity kinematics was observed during the range of motion and the range of medial /lateral deflections were analyzed.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 206 - 206
1 Dec 2013
Woodard E Mihalko W Williams J
Full Access

Soft tissue balancing during Total Knee Arthroplasty (TKA) is a step every surgeon takes during surgery. Coronal and transverse plane mechanical alignment is another parameter that surgeons address during surgery in an attempt to decrease wear and increase longevity. To date, a correlation between laxity, component wear patterns, and alignment of the tibial and femoral implant components has not been established. Theoretically, suboptimal alignment and poor soft tissue balancing should increase polyethylene wear and decrease implant survivorship, contributing to implant loosening and costly revision surgeries. This study utilizes a retrieval program of functioning TKAs obtained at the time of necropsy. By utilizing CT scans, mechanical laxity testing, and polyethylene damage scores, we aimed to determine if any correlation between proper alignment and ligament balancing to polyethylene damage scores exists.

Methods:

Computed Tomography (CT) scans were performed on 17 cadaveric knees containing TKAs obtained from the Medical Education and Research Institute (Memphis TN) using a GE Brightspeed scanning system with a 1.25 mm slice thickness. Transverse slices from these scans were used to calculate the femoral and tibial component rotation for each specimen. Component rotation was determined by utilizing previously published methods (Berger et al), and component mismatch was defined as the difference in rotation angles of the femoral and tibial components. After removal of skin, subcutaneous and muscle tissue, the tibia and femur of each leg was cut transversely, and the specimens were mounted in a custom knee testing machine (Little Rock, AR). Specimens were subjected to a 10 Nm varus and valgus torque and a 1.5 Nm internal and external rotation torque. Data was continuously recorded, and the angle or displacement at each torque or force was noted. Each test was performed at full extension and 30, 60, and 90 degrees of flexion. TKA components were then removed from the cadaveric knees, cleaned of PMMA, and visually inspected for wear using a grading system with 10 wear areas on the articulating surface of the polyethylene tibial insert (Hood et al). Scores were assigned based on severity of 7 different degradation characteristics, and were separated based on medial or lateral compartment. The maximum possible total score was 210 for each knee.

Results:

The average length of TKA implantation was 10 years. The coronal angle at +10 Nm (varus) moment ranged from 5 to 12 degrees, while the angle under a −10 Nm (valgus) moment ranged from 7 to 11 degrees across 10 specimens. The average component rotational mismatch was 20.5 degrees. The average overall medial wear score was 8.8, while the lateral average was 9.6. Wear scores showed a higher correlation to laxity in the medial compartment than the lateral side (Figures 1 and 2).