Introduction.
Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels. All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the hip prosthesis and whole blood cobalt ion measurements. Overall, 94 patients (95 hips) fulfilled our requirements for a minimum ten-year follow-up.Aims
Methods
Recent case reports have described V40 taper failure with clinically relevant adverse reaction to metal debris (ARMD). The real incidence of V40 taper damage and potential consequences are currently unclear, however. Aim of this study is therefore, to evaluate the long-term incidence of
Introduction. Modular neck (MN) components in total hip arthroplasty (THA) offer intraoperative flexibility, but adverse local tissue reactions (ALTR) due to tribocorrosion at modular junctions are a potential complication of such designs. Serum ion levels and metal artifact reduction sequence (MARS) MRI are used to assess ALTR following modular THA. This study investigates serum ion levels and MARS MRI findings in a series of hips with MN components and differing articulating surfaces. Methods. We retrospectively evaluated a cohort of 184 primary THAs in 159 patients implanted with a dual modular femoral stem by one surgeon from 2005–2013. 121 THAs had a cobalt-chromium neck component and non-metal-on-metal articulation, while 63 THAs had a titanium neck component and metal-on-metal (MoM) articulation. Serum ion levels were recorded for all patients. MARS MRI scans were read by musculoskeletal-trained radiologists.
Purpose. Our primary purpose was to study the rate of occurrence and the natural course of
Introduction. Pseudotumors have been reported following metal-on-metal total hip arthroplasty (THA); however, the natural history and longitudinal imaging findings of
High complication rates and poor outcomes have been widely reported in patients undergoing revision of large head metal-on-metal arthroplasty. A previous study from our centre showed high rates of dislocation, nerve injury, early cup loosening and
Introduction. The use of metal-on-metal (MOM) and modular total hip arthroplasty (THA) is associated with potentially serious complications including elevated serum metal ion levels,
Metal on metal total hip arthroplasty (MoM THA) provides the potential improvement in articular wear. However, several adverse events including
Background. High complication rates and poor outcomes have been widely reported in patients undergoing revision of large head metal-on-metal arthroplasty. A previous study from our center showed high rates of dislocation, nerve injury, early cup loosening and
Background.
We previously reported the five to ten-year results of the Birmingham Hip Resurfacing (BHR) implant. The purpose of this study was to evaluate the survivorship, radiographic results, and clinical outcomes of the BHR at long-term follow-up. We retrospectively reviewed 250 patients from the original cohort of 324 BHRs performed from 2006 to 2013 who met contemporary BHR indications. Of these, 4 patients died and 4 withdrew. From the 242 patients, 224 patients (93%) were available for analysis. Modified Harris hip score (mHHS) and University of California, Los Angeles (UCLA) scores were collected and compared to a matched total hip arthroplasty (THA) cohort. Mean follow-up was 14 years. Survivorship free of aseptic revision was 97.4% and survivorship free of any revision was 96.0% at 15 years. Revisions included 3 periprosthetic joint infections, 2 for elevated metal ions and symptomatic
Increased femoral head size reduces the rate of dislocation after total hip arthroplasty (THA). With the introduction of highly crosslinked polyethylene (HXLPE) liners in THA there has been a trend towards using larger size femoral heads in relatively smaller cup sizes, theoretically increasing the risk of liner fracture, wear, or aseptic loosening. Short to medium follow-up studies have not demonstrated a negative effect of using thinner HXLPE liners. However, there is concern that these thinner liners may prematurely fail in the long-term, especially in those with thinner liners. The aim of this study was to evaluate the long-term survival and revision rates of HXLPE liners in primary THA, as well as the effect of liner thickness on these outcomes. We hypothesized that there would be no significant differences between the different liner thicknesses. We performed a retrospective database analysis from a single center of all primary total hip replacements using HXLPE liners from 2010 and earlier, including all femoral head sizes. All procedures were performed by fellowship trained arthroplasty surgeons. Patient characteristics, implant details including liner thickness, death, and revisions (all causes) were recorded. Patients were grouped for analysis for each millimeter of PE thickness (e.g. 4.0-4.9mm, 5.0-5.9mm). Kaplan-Meier survival estimates were estimated with all-cause and aseptic revisions as the endpoints. A total of 2354 patients (2584 hips) were included (mean age 64.3 years, min-max 19-96). Mean BMI was 29.0 and 47.6% was female. Mean follow-up was 13.2 years (range 11.0-18.8). Liner thickness varied from 4.9 to 12.7 mm. Seven patients had a liner thickness <5.0mm and 859 had a liner thickness of <6.0mm. Head sizes were 28mm (n=85, 3.3%), 32mm (n=1214, 47.0%), 36mm (n=1176, 45.5%), and 40mm (n=109, 4.2%), and 98.4% were metal heads. There were 101 revisions, and in 78 of these cases the liner was revised. Reason for revision was instability/dislocation (n=34),
(Introduction). Metal on Metal (MoM) THA bearing as the second generation Total Hip Arthroplasty, widely came into use in Japan since 2000s. And it was expected to be able to ensure long-term use, the range of motion in the joints after surgeries, and the resistance to the dislocation, because MoM THAhad been considered to be able to use the large diameter head for its lower wear. However, there are some reports that the metal wear debris from MoM bearing surface generates the elevated serum ion level in the blood and the incidence of
Summary. The M2a-38. tm. metal on metal total hip arthroplasty showed a high incidence of
Introduction:. To date, there have been few reports of the results of revisions for failed metal-on-metal hip arthroplasties (MoM HA's). These series have included relatively modest numbers, and classification of the severity of adverse local tissue reaction (ALTR) has been under-reported. In this study, early outcomes and complications are analyzed as a function of pre-operative MRI grade and intra-operative ALTR severity to determine their prognostic value. Methods:. This is a retrospective review of revisions of 121 failed MoM HA's performed between 2008 and 2012. Indications for revision include ALTR (n = 97), loose cup (n = 13), and combined loose cup and ALTR (n = 11). When pre-operative MRI's were available, these were graded according to Anderson's system by 2 radiologists who were blinded to the clinical results. Intra-operative findings were graded for severity according to an ALTR classification. Post-operative complications were recorded. Correlations between pre-operative MRI grade, intra-operative ALTR classification and post-operative complications were analyzed using Spearman's rank correlation. ALTR Classification:. Type 0: No abnormal fluid or
Introduction. Metal on metal bearings are used especially in hip resurfacing. On the one hand, small bone preserving implants can be used. On the other hand recent studies found a variety of local and systemic side effects, for instance the appearance of
There have been numerous reports regarding “pseudotumor” associated with hip arthroplasty. We present two reports in which main etiology in the
The biological reaction in metallosis and
Metal-on-metal bearing surfaces were used frequently until recently because of the potential for decreased wear and the ability to use large femoral heads which can reduce instability. However, data reported in the Australian registry over the past 5 years demonstrated an increase in failure rates compared to metal-on-polyethylene bearings. In addition, adverse local tissue reactions (ALTR) associated with