A Case of Pseudotumor Detection Using Contrast-Enhanced Tomosynthesis After Metal-on-Metal Total Hip Arthroplasty

International Society for Technology in Arthroplasty (ISTA) 2012 Annual Congress



Abstract

Introduction

Pseudotumor is a known complication of Metal-on-Metal (MOM) total hip arthroplasty (THA). MRI is usually used to visualize pseudotumor formation. However, small pseudotumors close to the THA components may not be observed using MRI due to image distortion by the interaction between the metallic objects and the magnetic fields. The CT image quality also degrades because MOM THA components can induce X-ray beam-hardening effects. Therefore, we evaluated contrast-enhanced (CE) tomosynthesis. Tomosynthesis is known as an X-ray tomography technique that provides images with fewer metal artifacts and lower X-ray doses for the patients. The aim of this report was to investigate the detectability of pseudotumors by tomosynthesis.

Case Report

A 71-year-old woman had undergone unilateral cementless large-diameter MOM THA using a couple of Conserve Plus acetabular cup and Profemur Z femoral component (Wright Medical, Memphis, Tennessee) for primary arthritis of the left hip at our hospital. She presented with severe hip, groin and buttock pain and swelling at fifteen months after surgery. Therefore, she was examined MRI and tomosynthesis using Sonialvision-Safire X-ray Radiography/Fluoroscopy System (Shimadzu Corporation, Japan). Plain tomosynthesis was obtained before the contrast media injection and followed by CE-tomosynthesis. Then, subtraction tomosynthesis between plain and CE-tomosynthesis were calculated in order to increase the image contrast. The subtraction tomosynthesis image enhanced the pseudotumor visibility, which was considered to be equal to that depicted using MRI (Fig. 1 and 2). However, using MRI, cystic lesions in the pseudotumor appeared as a very high signal in the T2 weighted images, whereas CE-tomosynthesis resulted in no image contrast (Fig. 3). The tomosynthesis image resulted in less image distortion and fewer metal artifacts than MRI, even in the area close to the hip implants.

Discussion

MRI results in a wide variety of soft tissue contrast when imaging pseudotumors. However, MRI produces distorted images near MOM THA components. As such, CE-tomosynthesis is a promising imaging technique for detecting small, early-stage pseudotumors. Tomosynthesis also has the advantage of delivering a lower X-ray dose to the patients and providing a better spatial resolution than CT in the coronal plane.