Advertisement for orthosearch.org.uk
Results 1 - 16 of 16
Results per page:
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_18 | Pages 6 - 6
1 Nov 2017
Halai M Sharp E Drury C MacLean A
Full Access

Limb length deficiency, secondary to trauma or infection, is a common reason for referral to our tertiary service. After experiencing troubles with the Intramedullary Skeletal Kinetic Distractor (ISKD), we changed implant to the magnet operated Precice nail. We evaluated the safety and reliability of this novel device and compared it to our early ISKD results. To minimise variables, we selected femurs only. In total, we reviewed medical and radiographic records of 20 cases (8 ISKD, 12 Precise) from 2010–2015. At each postoperative visit, the accuracy and precision of distraction and complications were recorded. Accuracy reflected how close the measured lengthening was to the prescribed distraction at each postoperative visit. Precision reflected how close the repeated measurements were to each other over the course of total lengthening period. No patients were lost to follow-up (1–3.5 years). With the Precice nail (2012–2015), 11/12 were male and 10/12 were caused by trauma. The mean age was 34. Mean total lengthening was 38mm (range, 29–53mm), with an accuracy of 98percnt; and precision of 92percnt;. All patients achieved target lengthening at a rate of 1mm lengthening per day. In one case, the Precice nail fractured and this was revised successfully. With the ISKD group (2010–2012), 8/8 had complications (magnet jamming, nail breakage, equinus contractures and claw toes), with 25percnt; achieving accurate lengthening and precision of 38percnt;. Our results so far have justified our change in implant choice and, in our experience, support the Precice nail as being safe and precise


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_6 | Pages 29 - 29
1 May 2021
Rouse B Giles S Fernandes J
Full Access

Introduction

We have previously published limb lengthening using external fixation in pathological bone diseases. We would like to report a case series of femoral lengthening using the PRECICE system in a similar pathological group especially looking at it's feasibility and complications.

Materials and Methods

This is a case series of four patients, two patients with osteogenesis imperfecta and two with Ollier's disease, who underwent femoral lengthening via distraction osteogenesis using the PRECICE intramedullary nail system. It was a retrospective study from a prospective database from clinical records and radiographs.


Bone & Joint Research
Vol. 7, Issue 7 | Pages 476 - 484
1 Jul 2018
Panagiotopoulou VC Davda K Hothi HS Henckel J Cerquiglini A Goodier WD Skinner J Hart A Calder PR

Objectives. The Precice nail is the latest intramedullary lengthening nail with excellent early outcomes. Implant complications have led to modification of the nail design. The aim of this study was to perform a retrieval study of Precice nails following lower-limb lengthening and to assess macroscopical and microscopical changes to the implants and evaluate differences following design modification, with the aim of identifying potential surgical, implant, and patient risk factors. Methods. A total of 15 nails were retrieved from 13 patients following lower-limb lengthening. Macroscopical and microscopical surface damage to the nails were identified. Further analysis included radiology and micro-CT prior to sectioning. The internal mechanism was then analyzed with scanning electron microscopy and energy dispersive x-ray spectroscopy to identify corrosion. Results. Seven male and three female patients underwent 12 femoral lengthenings. Three female patients underwent tibial lengthening. All patients obtained the desired length with no implant failure. Surface degradation was noted on the telescopic part of every nail design, less on the latest implants. Microscopical analysis confirmed fretting and pitting corrosion. Following sectioning, black debris was noted in all implants. The early designs were found to have fractured actuator pins and the pin and bearings showed evidence of corrosive debris. The latest designs showed evidence of biological deposits suggestive of fluid ingress within the nail but no corrosion. Conclusion. This study confirms less internal corrosion following modification, but evidence of titanium debris remains. We recommend no change to current clinical practice. However, potential reuse of the Precice nail, for secondary limb lengthening in the same patient, should be undertaken with caution. Cite this article: V. C. Panagiotopoulou, K. Davda, H. S. Hothi, J. Henckel, A. Cerquiglini, W. D. Goodier, J. Skinner, A. Hart, P. R. Calder. A retrieval analysis of the Precice intramedullary limb lengthening system. Bone Joint Res 2018;7:476–484. DOI: 10.1302/2046-3758.77.BJR-2017-0359.R1


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_10 | Pages 21 - 21
1 Jun 2023
Wade A Devane Y Nowlan N Donnelly T Green C
Full Access

Introduction. There is no doubt that the future of limb lengthening lies with internal lengthening. Complication rates are reduced and patient satisfaction is increased. The evolution of internal lengthening peaked with the dual direction, easily inserted and externally controlled PRECICE Nail. It has excelled in performance in accuracy and satisfaction. Its versatility increased with smaller sizes and increased excursion. A field safety notice was issued was issued in October 2021 by the parent company NuVasive. The advice was monitoring of current cases and a hold on implantation until after a review of process. At the National Orthopaedic Hospital Cappagh we elected to remove all implanted nails and assess the nail integrity and physiological changes associated with implantation. Materials & Methods. All patients in who a retained Precice nail at the time of the field safety notice were identified. Patients and families were contacted to explain the issued safety notice from the company and explain that we would be scheduling them for nail removal. This was part of our standard care but we prioritised this group on our waiting list. Consent was obtained for nail removal but also for histological assessment of canal scrapings, blood ion level analysis and independent assessment of the retrieved nail by our academic collaborators at University College Dublin. Ion levels were then repeated at an interval post removal with consent. Results. We identified 7 Precice nails in 5 patients still implanted. All patients had consolidated after lengthening and were ready for nail removal. Elevated Titanium blood ion levels were identified at the time of retrieval. Histological abnormalities consistent with metalosis were identified. In all cases the nail was grossly intact and examination of the motor showed no evidence of wear or failure. Follow up ion levels obtained post retrieval showed persistent elevation. Conclusions. Our retrieval audit shows persistent elevation of Titanium ion levels and abnormal histology despite apparent nail integrity. These findings require further evaluation in a larger retrieval series to determine if there is a high incidence of this phenomenon and if ion levels have a clinical effect


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_13 | Pages 1 - 1
1 Jun 2017
Panagiotopoulou V Davda K Hothi H Henckel J Cerquiglini A Goodier W Skinner J Hart A Calder P
Full Access

Introduction. The Precice nail is the latest intramedullary lengthening nail with excellent early outcomes. Implant complications have led to modification of the nail design. The aim of this study was to perform a retrieval study of Precice nails following lower limb lengthening. To assess macroscopic and microscopic changes to the implants and assess differences following design modification, with identification of potential surgical, implant and patient risk factors. Method. 15 nails were retrieved from 13 patients following lower limb lengthening. Macroscopic and microscopic surface damage to the nails were identified. Further analysis included radiology and micro-CT prior to sectioning. The internal mechanism was then analysed with Scanning Electron Microscopy and Energy Dispersive X-ray Spectroscopy to identify corrosion. Results. 7 male and 3 females underwent 12 femoral lengthenings, 9 antegrade and 3 retrograde. 3 females underwent tibial lengthening. All patients obtained the desired length with no implant failure and full regenerate consolidation. Surface degradation was noted on the telescopic part of every nail design, less on the latest implants. Microscopic analysis confirmed fretting and pitting corrosion. Following sectioning black debris was noted in all implants. The early designs were found to have fractured actuator pins and the pin and bearings had evidence of corrosive debris. The latest designs had evidence of biological deposits suggestive of fluid ingress within the nail. Conclusion. This study suggests fluid ingress occurs with every generation of Precice nail despite modifications. The presence of biological fluid could be an early warning sign of potential corrosion. This in theory could lead to actuator pin fracture and implant failure. The clinical relevance is the potential re-use of a “dormant” nail in patients requiring secondary limb segment lengthening. Retraction of the nail in-situ and re-use for further lengthening requires careful consent for potential implant failure


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_10 | Pages 28 - 28
1 Jun 2023
Musielak B Green N Giles S Madan S Fernandes J
Full Access

Introduction. Intramedullary lengthening devices have been in use in older children with closed /open growth plates with good success. This study aims to present the early experience of the FITBONE nail since withdrawal of the PRECICE nail. Materials & Methods. Retrospective analysis of both antegrade and retrograde techniques were utilized. Only patients where union was achieved and full weight bearing commenced were included. The complication rate, length gained, distraction index, weight bearing index (WBI) as well as mechanical axes were analysed. Results. 14 (7 males, 7 females) of a total of 16 (7 males, 7 females) patients with a mean age of 16.9 years with varied diagnosis of LLD were analysed. The mean length gained was 38 mm with an average distraction index of 0,74 mm/day. WBI in these patients on average was 59,6 days/cm lengthened. 6 complications were observed, including two nonunions (successfully treated) and a knee subluxation. Mechanical axis deviation improved from 13,3 mm to 6 mm on average. Overall there has been a nonsignificant tendency for WBI to decrease (Spearman's rank correlation coefficient −0.47, p=0.08) with increasing number of cases done, while no correlation between length gained and WBI (−0.01, p=0.96, respectively) was observed. Some nuances will be discussed. Conclusions. Limb lengthening with the FITBONE nail is relatively safe and efficient, however no significant change was seen in the outcome with previous motorized nails


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 24 - 24
23 Apr 2024
Thompson E James L Narayan B Peterson N
Full Access

Introduction. Management of deformity involving limb length discrepancy (LLD) using intramedullary devices offers significant benefits to both patients and clinicians over traditional external fixation. Following the withdrawal of the PRECICE nail, the Fitbone became the primary implant available for intramedullary lengthening and deformity correction within our service. This consecutive series illustrates the advantages and complications associated with the use of this device, and describes a novel technique modification for antegrade intramedullary lengthening nails. Materials & Methods. A retrospective cohort review was performed of patient outcomes after treatment with the Fitbone nail at two tertiary referral limb reconstruction services (one adult, one paediatric) between January 2021 to December 2023. Aetiology, indications, initial and final LLD, use of concomitant rail assisted deformity correction (ORDER), removal time and healing index were assessed. Complications of treatment were evaluated and described in detail, alongside technique modifications to reduce the rate of these complications. Results. 21 nails (18 femoral, 2 tibial, 1 humeral) were inserted in 6 adult and 13 paediatric patients. Post-traumatic and congenital/developmental LLD were the most common indications for surgery in the adult and paediatric cohorts respectively. ORDER was employed in 11 cases (9 femurs and 2 tibias). Treatment goals were achieved in all but one case. Complications included superficial infection, locking bolt migration, periprosthetic fracture and component failure. Seven patients required unplanned returns to theatre. Conclusions. The Fitbone nail is an established option for intramedullary limb lengthening, however its use in the UK has been relatively limited compared to the PRECICE until 2021. Our data helps to define its place for limb lengthening and complex deformity correction in both adult and paediatric patients, including in humeral lengthening and retrograde femoral insertion across an open physis. We have identified important potential risks and novel techniques to simplify surgery and avoid complications


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_23 | Pages 11 - 11
1 May 2013
Calder P McGrath A Chasseaud M Timms A Goodier W
Full Access

We present the results of a new non-invasive lengthening nail enabling accurate control of the lengthening process and joint rehabilitation. Introduction. The use of intramedullary lengthening nails have gained popularity as they reduce common complications associated with external fixators, including infection, joint stiffness, bone regenerate deformity, late fracture and patient implant acceptance. Current nails however are associated with complications including implant breakage, mechanical failure, runaway nail and requiring MUA to restart or obtain segment lengthening. The Precice nail incorporates magnet technology with a hand held device allowing non-invasive lengthening. The nail is also reversible allowing shortening if required. Physiotherapy can continue throughout treatment to maintain joint range of motion without concern of uncontrolled nail runaway. The lengthening is axial reducing shear/torsional forces on the regenerate. Method. The lengthening of 4 femora was undertaken in 3 patients, mean age 34 yrs for post-traumatic shortening and short stature. A standard technique included an Ilizarov corticotomy followed by a 6 day latent period. Patients were mobilised partial weight bearing and knee range of motion maintained. The femora were lengthened one third of a millimetre three times per day. Radiographic and clinical review was performed every 2 weeks. Results. The desired femoral length was obtained in all patients, mean 5.6 cms (4.5–6.5). There were no complications of infection, poor bone regenerate or premature consolidation. One patient undergoing bilateral femoral lengthening underwent surgical release of a tight Tensor Fascia Lata following completion of lengthening due to hip flexion contracture. One patient required slowing of the lengthening rate due to knee flexion contracture. Physiotherapy corrected the deformity and the normal lengthening rate could be resumed. Conclusion. The Precice nail is a new device that offers accurate control of the lengthening process and rehabilitation. This should reduce complications of poor regenerate formation and soft tissue contractures


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_8 | Pages 4 - 4
1 May 2018
Timms A Roskrow T Doyle H Edel P Fugazzotto S Goodier W Calder P
Full Access

Introduction. Muscle stiffness and joint contractures are common complications of limb lengthening. Authors have demonstrated less permanent soft tissue complications with intramedullary lengthening than external fixation. Our aim was to evaluate the joint response following intramedullary femoral lengthening and need for physiotherapy and alteration to rate/rhythm of lengthening. Method. A retrospective review of documentation for all femoral Precice nails in our centre inserted between 2012 and 2017. This involved 98 nails (68 antegrade, 30 retrograde) in 88 patients (59 males, 29 females) with a mean age of 32 years (range 12–69 yrs). We excluded cases where there was no documentation regarding Range of Movement (ROM). Bilateral lengthenings were recorded as separate cases. This left 50 antegrade, 16 retrograde cases with hip ROM data and 55 antegrade, 26 retrograde cases with Knee ROM data. Results. Hip. In the antegrade group 20 cases (39%) developed no stiffness throughout treatment. ROM decreased in 61% (n=30) of cases and the median length at which this occurred was 3cm. Where a retrograde nail was inserted 62.5% of cases (n=10) maintained normal range of motion. The median length at which ROM decreased was 3.25cm, this occurred in 6 cases (37.5%). . Knee. Where the antegrade nail was inserted, 22 cases (40%) did not develop stiffness. There was reduced ROM in 33 cases (60%), occurring at a median distance of 2cm. The median distance at which reduced ROM occurred with the retrograde nails was 1.5cm, this occurred in 23 cases (88.5%). 3 cases did not develop stiffness. All cases regained full ROM. Conclusion. Although there was no permanent loss, Joint Stiffness still occurs with intramedullary limb lengthening and there remains the need for regular physiotherapy. There are no data from other institutions for comparison. Rehabilitation guidelines and a proforma to accurately monitor patients ROM throughout treatment have been developed


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_5 | Pages 40 - 40
1 Apr 2022
Hafez M Nicolaou N Dixon S Obasohan P Giles S Madan S Fernandes J Offiah A
Full Access

Introduction. Motorised intramedullary lengthening nails are considered more expensive than external fixators for limb lengthening. This research aims to compare the cost of femoral lengthening in children using the PRECICE magnetic lengthening nail with external fixation. Materials and Methods. Patients: Retrospective analysis of 50 children who underwent femoral lengthening. One group included patients who were treated with PRECICE lengthening nails, the other group included patients who had lengthening with external fixation. Each group included 25 patients aged between 11–17 years. The patients in both groups were matched for age. Cost analysis was performed following micro-costing and analysis of the used resources during the different phases of the treatments. Results. : Each group's mean patient age was 14.7 years. Lengthening nails were associated with longer operative times compared to external fixators, both for implantation and removal surgery (P-value 0.007 and <0.0001 respectively). Length of stay following the implantation surgery, frequency of radiographs, frequency of outpatient department appointments were all more favourable with lengthening nails. The overall cost of lengthening nails was £1393 more than external fixators, although this difference was not statistically significant (P-value 0.088). Conclusions. The cost of femoral lengthening with lengthening nails was not significantly higher than the external fixators’ cost. Further research to review the effectiveness of the devices and the quality of life during the lengthening process is crucial for robust health economic evaluation


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_6 | Pages 25 - 25
1 May 2021
Stoddart M Elsheikh A Wright J Goodier D Calder P
Full Access

Introduction. Pixel Value Ratio (PVR) is a radiographic measure of the relative density of the regenerate to the adjacent bone. This has been reported as an objective criterion for regenerate healing and a guide for when to allow full weight bearing (FWB) in lengthening with intramedullary telescopic nails. The threshold for which magnitude of PVR is adequate to allow bearing full weight is not yet agreed. The aim of this study was to identify from our cohort of adult limb lengthening patients the time to FWB following lengthening, the PVR at this point, and how this compared with the recommended values in the literature. Materials and Methods. A retrospective database review identified 30 adult patients treated with the PRECICE femoral nail by two senior authors. Time from completion of lengthening to instruction to fully weight bear was noted. The PVR was calculated for each cortex on plain radiographs taken at each postoperative visit following completion of lengthening. Significance was set at p <0.05. Results. The median age was 30.5 years (IQR 22.5 – 42.5), 19 male and 11 female patients were included. The underlying cause of shortening was post traumatic in 12 (40%) with the remaining due to a wide variety of causes including congenital, infective, and idiopathic leg length discrepancies. Twenty-two nails were antegrade and the mean distance lengthened was 42 mm. Median time from completion of lengthening to full weight bearing was 42 days (IQR 28 – 69). The overall mean PVR at FWB was 0.79. Each surgeon had differing protocols for weightbearing, however there was no statistical difference between time to weight bearing, nor the mean PVR at FWB. There were no implant failures, shortening, or regenerate fractures. Conclusions. We report on our cohort of adult patients treated with intramedullary lengthening nails with a mean PVR of 0.79 at time of full weightbearing. This is considerably lower than the values reported in the literature ranging from 0.90 to 0.93. We therefore suggest that weightbearing following lengthening nails can be permitted earlier than previously reported without an increased risk of complication


Bone & Joint Open
Vol. 2, Issue 8 | Pages 599 - 610
1 Aug 2021
Hothi H Bergiers S Henckel J Iliadis AD Goodier WD Wright J Skinner J Calder P Hart AJ

Aims

The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices.

Methods

We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_5 | Pages 1 - 1
1 May 2015
Laubscher M Mitchell C Timms A Goodier D Calder P
Full Access

Background:. External fixators are not as well tolerated around the femur when compared to the tibia. Lengthening with an intramedullary device is therefore attractive. Method:. We reviewed all cases of femoral lengthening performed at our unit from 2007 to 2014. Cases of non-unions, concurrent deformities, congenital limb deficiencies and lengthening with an unstable hip were excluded. This left 33 cases for review. Healing index, implant tolerance and complications were compared. Results and Discussion:. In 20 cases the Precice lengthening nail was used and in 13 cases the LRS external fixator system. The desired length was achieved in all cases in the Precice group and in 12 of 13 cases in the LRS group. The Precice group had a more rapid return to full weight bearing. The mean healing index was 31.3 days/cm in the Precice and 47.1 days/cm in the LRS group. There was an increased incidence of complications with LRS lengthening, including pin site infections and regenerate deformity. Implant tolerance and the patients' perception of the cosmetic result were better with the Precice treatment. Conclusion:. We conclude excellent functional results with fewer complications and greater patient satisfaction in femoral lengthening with a Precice intramedullary nail


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_5 | Pages 4 - 4
1 May 2015
Laubscher M Mitchell C Timms A Goodier D Calder P
Full Access

Background:. Leg length discrepancy related to absence of the femoral head with proximal migration of the femur presents a treatment dilemma. Late sequelae of neonatal hip sepsis and chronic hip dislocation due to developmental dysplasia are the most common causes. Traditional teaching dictates that the hip is stabilised prior to limb lengthening. Reconstructive options alter the shape of the proximal femur which complicates future surgery. Methods:. We retrospectively reviewed 3 cases of femoral lengthening with an ‘unstable’ hip without prior stabilization. The aetiology was neonatal hip sepsis in 2 cases and chronic hip dislocation due to developmental dysplasia in 1 case. Lengthening was performed with the use of a retrograde Precice lengthening nail. Results and Discussion:. The desired length (range 5.5 to 6cm) was achieved in all patients without migration of the proximal femur. In all cases patient satisfaction was high with the lengthening and the ability to abort shoe raises. There was an improvement in gait pattern in all patients. Conclusion:. Femoral lengthening is possible without prior stabilisation of the hip in so called ‘unstable’ hip situations. The proximal femoral anatomy is not further altered and future salvage procedures such as a total hip replacement are not compromised


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims

This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported.

Methods

This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed.


Bone & Joint 360
Vol. 3, Issue 3 | Pages 34 - 37
1 Jun 2014

The June 2014 Children’s orthopaedics Roundup360 looks at: plaster wedging in paediatric forearm fractures; the medial approach for DDH; Ponseti – but not as he knew it?; Salter osteotomy more accurate than Pemberton in DDH; is the open paediatric fracture an emergency?; bang up-to-date with femoral external fixation; indomethacin, heterotopic ossification and cerebral palsy hips; lengthening nails for congenital femoral deformities, and is MRI the answer to imaging of the physis?