Bone & Joint Open
Vol. 5, Issue 6 | Pages 514 - 523
24 Jun 2024
Fishley W Nandra R Carluke I Partington PF Reed MR Kramer DJ Wilson MJ Hubble MJW Howell JR Whitehouse SL Petheram TG Kassam AM

Aims. In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component. Methods. Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years). Results. In the DM group, two (3.8%) patients underwent further surgery: one (1.9%) for dislocation and one (1.9%) for infection. In the AR group, four (9.8%) underwent further procedures: two (4.9%) for loosening of the acetabular component and two (4.9%) following dislocations. There were no other dislocations in either group. In the DM group, operating time (68.4 vs 101.5 mins, p < 0.001), postoperative drop in haemoglobin (16.6 vs 27.8 g/L, p < 0.001), and length of stay (1.8 vs 2.4 days, p < 0.001) were significantly lower. There was a significant reduction in serum metal ions postoperatively in both groups (p < 0.001), although there was no difference between groups for this reduction (p = 0.674 (cobalt); p = 0.186 (chromium)). Conclusion. In selected patients with Birmingham MoM hips, where the acetabular component is well-fixed and in a satisfactory position with no surface damage, the metal head can be exchanged for polyethylene ADM/MDM bearings with retention of the acetabular prosthesis. This presents significant benefits, with a shorter procedure and a lower risk of complications. Cite this article: Bone Jt Open 2024;5(6):514–523


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 21 - 21
7 Jun 2023
Nandra R Fishley W Whitehouse S Carluke I Kramer D Partington P Reed M Evans J Panteli M Charity J Wilson M Howell J Hubble M Petheram T Kassam A
Full Access

In metal-on-metal (MoM) hip replacements or resurfacings, mechanical induced corrosion can lead to a local inflammatory response, pseudo tumours and elevated serum metal ions, requiring revision surgery. The size and diametral clearance of Anatomic (ADM) and Modular (MDM) Dual Mobility bearings matches that of certain MOM components. Presenting the opportunity for revision with exchange of the metal head for ADM/MDM bearings without removal of the acetabular component if it is well-fixed and appropriately positioned. Between 2012 and 2020, across two centres, 94 patients underwent revision of a MoM hip replacement or resurfacing. The mean age was 65.5 (33–87) years. In 53 patients (56.4%), the acetabular component was retained, and dual mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). DM was only considered where the acetabular component was satisfactorily positioned and well-integrated into bone, with no surface damage. Patients underwent clinical and radiographic follow-up to at least one-year (mean 42.4 (12–96) months). One (1.1%) patient died before one-year, for reasons unrelated to the surgery. In the DM group, two (3.8%) patients underwent further surgery; one (1.9%) for dislocation and one (1.9%) for infection. In the AR group, four (12.2%) underwent further procedures; two (4.9%) for loosening of the acetabular component and two (4.9%) following dislocations. There were no other dislocations in either group. In the DM group, operative time (68.4 v 101.5 mins, p<0.001), postoperative drop in haemoglobin (16.6 v 27.8 g/L, p<0.001), and length of stay (1.8 v 2.4 days, p<0.001) were significantly lower. There was a significant reduction in serum metal ions postoperatively in both groups (p<0.001 both Cobalt and Chromium) although there was no difference between groups for this reduction (p=0.674 Cobalt; p=0.186 Chromium). In selected patients with MoM hip arthroplasty, where the acetabular component is well-fixed, in a satisfactory position and there is no surface damage, the metal head can be exchanged for ADM/MDM bearings with retention of the acetabular prosthesis. Presenting significant benefits through a less invasive procedure, and a low risk of complications, including dislocation


Bone & Joint Research
Vol. 8, Issue 3 | Pages 146 - 155
1 Mar 2019
Langton DJ Natu S Harrington CF Bowsher JG Nargol AVF

Objectives. We investigated the reliability of the cobalt-chromium (CoCr) synovial joint fluid ratio (JFR) in identifying the presence of a severe aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) response and/or suboptimal taper performance (SOTP) following metal-on-metal (MoM) hip arthroplasty. We then examined the possibility that the CoCr JFR may influence the serum partitioning of Co and Cr. Methods. For part A, we included all revision surgeries carried out at our unit with the relevant data, including volumetric wear analysis, joint fluid (JF) Co and Cr concentrations, and ALVAL grade (n = 315). Receiver operating characteristic curves were constructed to assess the reliability of the CoCr JFR in identifying severe ALVAL and/or SOTP. For part B, we included only patients with unilateral prostheses who had given matched serum and whole blood samples for Co and Cr analysis (n = 155). Multiple regression was used to examine the influence of JF concentrations on the serum partitioning of Co and Cr in the blood. Results. A CoCr JFR > 1 showed a specificity of 83% (77% to 88%) and sensitivity of 63% (55% to 70%) for the detection of severe ALVAL and/or SOTP. In patients with CoCr JFRs > 1, the median blood Cr to serum Cr ratio was 0.99, compared with 0.71 in patients with CoCr JFRs < 1 (p < 0.001). Regression analysis demonstrated that the blood Cr to serum Cr value was positively associated with the JF Co concentration (p = 0.011) and inversely related to the JF Cr concentration (p < 0.001). Conclusion. Elevations in CoCr JFRs are associated with adverse biological (severe ALVAL) or tribocorrosive processes (SOTP). Comparison of serum Cr with blood Cr concentrations may be a useful additional clinical tool to help to identify these conditions. Cite this article: D. J. Langton, S. Natu, C. F. Harrington, J. G. Bowsher, A. V. F. Nargol. Is the synovial fluid cobalt-to-chromium ratio related to the serum partitioning of metal debris following metal-on-metal hip arthroplasty? Bone Joint Res 2019;8:146–155. DOI: 10.1302/2046-3758.83.BJR-2018-0049.R1


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 5 - 5
1 Dec 2013
Angel C Sexton S MacNair R Fuller N
Full Access

Background:. Higher than expected revision rates have been observed in large bearing metal-on-metal total hip replacements. We have introduced a metal on metal hip screening clinic at our unit and report the results. Methods:. All patients who had a metal-on-metal total hip replacement implanted at our unit were recalled to clinic. Screening consisted of clinical examination, Oxford hip score, radiographs, MARS MRI and serum cobalt/chromium ion levels. MRI scans were graded (Norwich Classification) by a single consultant radiologist who was blinded to the history, examination, and metal levels. Patients were sub-classified into 6 groups, which determined further follow up and were either entered into a regular screening programme or referred to a revision arthroplasty surgeon for a decision on whether revision was required. Results:. A total of 278 patients were identified. Four patients had moved away, 2 refused follow-up and 1 patient had died. This left 271 patients in our cohort, having 299 metal-on-metal hip replacements with a M:F ratio of 1.5:1 and a mean age of 61.9 years (29–87). 59 (21.8%) patients complained of a recent deterioration in symptoms. Sixty seven (24.7%) patients had an abnormal xray appearance of their hip replacement (inclination >50 degrees, loosening, or migration of prosthesis). Of the 232 hips which underwent MARS MRI, 118 (50.9%) were group A (normal appearance), 66 (28.4%) C1 (mild MOM disease), 31 (13.4%) C2 (moderate MOM disease), and 17 (7.3%) C3 (severe MOM disease). One hundred and seventeen (38.9%) had a substantial fluid collection around the hip. Serum cobalt/chromium metal ions were raised in 135 patients (49.5%), 46 of which had one (8.8%) or both (8%) above 7 ppb (chromium >134 nmol/L, cobalt > 119 nmol/L). One hundred and forty nine patients (55%) remained under review in the MOM clinic and 122 (45%, 148 hip replacements) were referred to a revision surgeon for specialist opinion. Of these 21 hips (14.2%) have been revised and a further 14 hips (9.5%) are awaiting revision. In total 64 hips (21.3% of all large head metal on metal hips implanted in our unit) have been revised with a further 22 (7.3%) listed for revision. Of these, 13 were histologically proven to be an ALVAL reaction, 19 had the classic clinical findings of ALVAL (but not histologically confirmed), and 1 was a non-specific reaction to metal wear. 53 patients (19.4%, 56 hips) were completely asymptomatic but had evidence of ALVAL soft tissue reaction on MRI and/or rising metal ion levels – 9 of these hips have already been revised for ALVAL and 6 more listed for revision. Conclusions. Our screening programme has identified 53 (19.6%) asymptomatic patients who have evidence of a suspicious reaction to their metal-on-metal hip replacements. 16 (3%) of these have gone on to revision. We would recommend all patients with a metal-on-metal total hip replacement in situ are fully screened for ARMD regardless of symptoms and metal ion levels


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_5 | Pages 21 - 21
1 Jul 2020
Nandra R Ahmed U Berryman F Brash L Dunlop D Matharu G
Full Access

Introduction. Many worldwide regulatory authorities recommend regular surveillance of metal-on-metal hip arthroplasty patients given high failure rates. However concerns have been raised about whether such regular surveillance, which includes asymptomatic patients, is evidence-based and cost-effective. We determined: (1) the cost of implementing the 2015 MHRA surveillance in “at-risk” Birmingham Hip Resurfacing (BHR) patients, and (2) how many asymptomatic hips with adverse reactions to metal debris (ARMD) would have been missed if patients were not recalled. Methods. All BHR patients subject to the 2015 MHRA recall (all females, and males with head sizes 46mm or below, regardless of symptoms) at one specialist centre were invited for review (707 hips). All patients were investigated (Oxford Hip Score, radiographs, blood metal ions, and targeted cross-sectional imaging) and managed accordingly. Surveillance costs were calculated using finance department data, as was the number needed to treat (NNT) to avoid missing one case of asymptomatic ARMD. Results. The overall institutional surveillance cost to investigate all patients once was £105,922 (range £148 to £258 per patient). The most expensive aspects of surveillance were radiographs (£39,598), advanced nurse practitioner assistance (£23,618), cross-sectional imaging (£14,828), and blood metal ions (£14,825). 31 hips had imaging evidence of ARMD (12 revised, with 19 under surveillance). Seven hips were asymptomatic, however all revisions were symptomatic. The NNT to avoid missing one case of asymptomatic ARMD (on imaging and/or requiring revision) was 101 patients, equating to a screening cost of £18,041 to avoid one case of asymptomatic ARMD. Conclusions. Implementing MHRA surveillance for “at-risk” BHR patients was extremely costly, both financially and logistically. As the risk of asymptomatic ARMD was low, our data suggests the 2015 MHRA surveillance is not cost-effective. We therefore have concerns about the increasingly intensive surveillance recommended in the 2017 MHRA guidance for all metal-on-metal hip arthroplasty patients


Bone & Joint Research
Vol. 7, Issue 6 | Pages 388 - 396
1 Jun 2018
Langton DJ Sidaginamale RP Joyce TJ Bowsher JG Holland JP Deehan D Nargol AVF Natu S

Objectives. We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). Patients and Methods. We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties. Results. In the resurfacing cohort (n = 76), the statistical modelling indicated that the presence of severe ALVAL and a large fluid collection were associated with greater joint fluid Co concentrations after adjustment for volumetric wear rates (p = 0.005). These findings were replicated in the mixed implant group (n = 178), where the presence of severe ALVAL and a large fluid collection were significantly associated with greater fluid Co concentrations (p < 0.001). Conclusion. The development of severe ALVAL is associated with elevations in metal ion concentrations far beyond those expected from the volumetric loss from the prosthetic surfaces. This finding may aid the understanding of the sequence of events leading to soft-tissue reactions following MoM hip arthroplasties. Cite this article: D. J. Langton, R. P. Sidaginamale, T. J. Joyce, J. G. Bowsher, J. P. Holland, D. Deehan, A. V. F. Nargol, S. Natu. Aseptic lymphocyte-dominated vasculitis-associated lesions are related to changes in metal ion handling in the joint capsules of metal-on-metal hip arthroplasties. Bone Joint Res 2018;7:388–396. DOI: 10.1302/2046-3758.76.BJR-2018-0037


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_1 | Pages 140 - 140
1 Jan 2013
Matthies A Chan N Jones H Isamailly S Skinner J Noble P Hart A
Full Access

Introduction. It has been speculated that impact deformation of thin 1-piece cups used for modern metal-on-metal hip replacement may contribute to early failure. The purpose of this study was to reproduce typical impact deformation and quantify the effect of this on the frictional torque generated at the hip. Methods. We tested nine hip couples of three designs (the ASR, BHR and Durom) and three sizes (42mm, 46mm and 50mm). A custom compression device was designed to replicate the in vivo forces and impact deformation of 1-piece metal cups reported in the literature. Each cup was mounted in the device, which itself was mounted on a mechanical testing machine. The cups were compressed with incremental loads up to a maximum of 2000N. At each increment we measured cup deformation, and then the head component was seated into the cup. The hip was lubricated and the head component rotated 60 degrees axially within the cup and the axial torque was measured. Results. Maximum deformation occurred at the rim for all cups, and was not associated with cup diameter (p = 0.42). However, at all applied loads the BHR cups were deformed significantly less than both the ASR and Durom cups (all p < 0.05). We observed deformation that exceeded the diametric clearance for both the ASR and Durom. Deformation resulted in increased axial torque up to eight times higher than those measured at zero deformation. The maximum torques generated for the BHR cups were significantly lower than those observed for both the ASR and Durom cups. Discussion. Deformation of 1-piece cups, equatorial bearing and subsequent increased frictional torque is likely to have widespread consequences for the performance of metal-on-metal hip replacement. This may include loosening of the cup, mechanical damage at the taper junction and increased early wear of the bearing surface


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 187 - 187
1 Dec 2013
Van Der Straeten C Van Quickenborne D Pennynck S De Smet K Victor J
Full Access

Background:. Potential systemic toxicity of metal ions from metal-on-metal hip arthroplasties (MoMHA) is concerning. High blood cobalt (Co) levels have been associated with neurological, cardiac and thyroid dysfunctions. Questions/purposes:. The aim of this research was to investigate the prevalence of systemic Co toxicity in a MoMHA population, to identify confounding factors, and to indicate a Co level above which there is a high risk for systemic toxicity. Patients and Methods:. We conducted a cross-sectional study of patients with a MoMHA, using questionnaires validated to detect cobaltism in cobalt industry workers. We retrospectively identified 161 patients with Co levels >4 μg/L and 337 patients with Co levels <4 μg/L and collected 385 completed questionnaires. There were 207 men and 178 women with a mean age of 60 years (24–86). Mean follow-up was 5.7 years (1 to 13). Co levels were subdivided into <4 μg/L, 4–10 μg/L, 10–20 μg/L and >20 μg/L groups. Results:. There were significant correlations (p ≤ 0.003) between increasing Co levels and prevalence of sleeping disorders, cognitive problems, equilibrium disturbances, neuropathic symptoms, fatigue, somatic disorders and mood changes. More frequent and severe symptoms were found with Co > 20 μg/L (p = 0.017). Several years post-revision, some patients still complained of chronic symptoms possibly associated with longer exposure. Female gender and age ≤50 years were confounding factors. Conclusions:. We demonstrated a correlation between increasing Co levels and toxicity symptoms (p ≤ 0.003). As in the published case reports of MoMHA related cobaltism, the highest incidence of symptoms was found with Co > 20 μg/L. Patients with repeated Co measurements >20 μg/L are at risk for systemic toxicity


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_10 | Pages 20 - 20
1 Oct 2015
Mohanlal P Bawale R Samsani S Jain S Joshi A Singh B Prasad R Pillai D
Full Access

Introduction. The MHRA guidelines for metal on metal (MOM) suggest cobalt and chromium levels of more than 7ppb as potential for soft tissue reaction. However, in some patients soft tissue reaction is seen even in the presence of normal serum metal ions levels. Methods. A prospective review of all patients who had metal-on-metal hip arthroplasty was done. Patients who had both serum metal ion levels and MARS MRI scan were included in this study. Results. There were 574 patients with 319 females and 255 males. 240 patients had both MARS MRI scan and serum metal ions. Fifty nine (24.5%) patients with normal metal ions had metallosis/pseudo tumour on MRI scans, and 17 (7%) patients with increased metal ions had normal MRI scans. Conclusion. This study confirms that there is poor correlation between serum metal ions and soft tissue reaction. Nearly a quarter of patients with normal serum metal ions had evidence of metallosis on MRI scan. Clinicians should be aware of the potential for soft tissue reaction even in the presence of normal metal ions. Hence, consideration should be given to further imaging where appropriate, to ensure early metallosis is picked up and further complications of pseudo tumour prevented


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 109 - 109
1 Sep 2012
Masri BA Williams DH Greidanus NV Duncan CP Garbuz DS
Full Access

Purpose. There is a postulated association between increased serum metal ions and pseudotumour formation in patients with metal-on-metal hip replacements. The primary aim of this study was to assess the prevalence of pseudotumour in 31 asymptomatic patients with a large femoral head (LFH) metal-on-metal hip implant. This was compared to the prevalence of pseudotumour in 20 matched asymptomatic patients with a hip resurfacing (HRA) and 24 matched asymptomatic patients with a standard metal-on-polyethylene (MOP) total hip. A secondary objective was to assess possible correlation between increased serum metal ions and pseudotumour formation. Method. Ultrasound examination of the three groups was performed at a minimum follow up of two years. Serum metal ions were measured in the metal-on-metal LFH and HRA groups at a minimum of two years. Results. There were 10 (32%) solid or cystic masses in the LFH group with a mean size of 89.3 (8 to 437) cm3. In the HRA group there were five (25%) masses with a mean size of 41.6 (6 to 119) cm3. In the MOP group there was one cystic mass measuring 9.8 cm3. Median serum cobalt and chromium ion levels in the LFH group were 4.50 gL (interquartile range, 2.38 to 7.47) and 2.82 gL (interquartile range, 1.48 to 3.18) compared to 0.83 gL (interquartile range, 0.65 to 1.12) and 1.08 gL (interquartile range, 0.78 to 1.31) in the HRA group. Conclusion. This study demonstrates a significantly higher prevalence of pseudotumours in patients with large head metal-on-metal total hips. The higher levels of metal ions in this group suggest that elevated metal ions may lead to pseudotumour formation


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 16 - 16
1 May 2019
Matharu G Berryman F Dunlop D Revell M Judge A Murray D Pandit H
Full Access

Introduction. We investigated predictors of poor outcomes following metal-on-metal hip arthroplasty (MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD), to help inform the revision threshold and type of reconstruction. Patients and Methods. A retrospective cohort study was performed involving 346 MoMHAs revised for histologically confirmed ARMD at two specialist centres (245=hip resurfacing, 101=total hip arthroplasty). Numerous preoperative (blood metal ions and imaging) and intraoperative (findings, and components removed/implanted) factors were used to predict poor outcomes. Poor outcomes were postoperative complications (including re-revisions), 90-day mortality, and poor Oxford Hip Scores (<27/48). Multivariable logistic regression models for predicting poor outcomes were developed using stepwise selection methods. Results. Cumulative implant survival rate seven-years after ARMD revision was 87.0% (95% CI=81.0%-91.2%). Poor outcomes occurred in 39% (n=135). Shorter time (under four-years) from primary to revision surgery (odds ratio (OR)=2.12, CI=1.00–4.46) was the only preoperative predictor of poor outcomes. Pre-revision metal ions and imaging did not influence outcomes. Single-component revisions increased the risk of poor outcomes (acetabular or femoral vs. all component revisions; OR=2.99, CI=1.50–5.97). Intraoperative factors reducing the risk of poor outcomes included the posterior approach (OR=0.22, CI=0.10–0.49), revision head sizes ≥36mm (vs. <36mm: OR=0.37, CI=0.18–0.77), ceramic-on-polyethylene (OR vs. ceramic-on-ceramic=0.30, CI=0.14–0.66) and metal-on-polyethylene revision bearings (OR vs. ceramic-on-ceramic=0.37, CI=0.17–0.83). Discussion. This large cohort study demonstrated 39% of patients experience poor outcomes following MoMHA revision for ARMD. This information will help surgeons when counselling patient's pre-revision about the expected prognosis. No threshold exists for recommending ARMD revision, therefore surgeons must make decisions on an individual case basis. However, surgeons can make intraoperative decisions that influence outcomes following ARMD revision. Conclusion. We recommend optimal outcomes following ARMD revision may be achieved if surgeons use the posterior approach, revise all MoMHA components, and use ≥36mm ceramic-on-polyethylene or metal-on-polyethylene articulations


Outcomes following metal-on-metal hip replacement (MoMHR) revision surgery for adverse reactions to metal debris (ARMD) have been poor, and inferior compared with non-ARMD revisions. Subsequently, surgeons and worldwide authorities widely recommended early revision for ARMD, with a lower surgical threshold adopted. However, the impact of early surgery for ARMD is unknown. We compared the rates of adverse outcomes following MoMHR revision surgery in matched ARMD and non-ARMD patients. We performed a retrospective observational study using data from the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man. All MoMHR patients subsequently undergoing revision surgery for any indication between August 2008 and August 2014 were eligible. ARMD and non-ARMD revisions were matched one-to-one for multiple potential confounding factors using propensity scores. Adverse outcomes following revision surgery (intra-operative complications, mortality, re-revision surgery) were compared between matched groups using regression analysis. In 2,576 matched MoMHR revisions (ARMD=1,288 and non-ARMD=1,288), intra-operative complications were similar between ARMD (2.4%) and non-ARMD (2.5%) revisions (odds ratio=0.97, 95% CI=0.59–1.60; p=0.899). All-cause mortality rates were lower following ARMD revision compared with non-ARMD revision (hazard ratio (HR)=0.43, 95% CI=0.22–0.86; p=0.018). All-cause re-revision rates were lower following ARMD revision compared with non-ARMD revision (HR=0.52, 95% CI=0.36–0.75; p<0.001). Compared with ARMD revision (5-years=94.3%), MoMHR revisions for infection (5-years=81.2%) and dislocation/subluxation (5-years=81.9%) had the lowest implant survival rates. Contrary to previous observations, MoMHRs revised for ARMD have approximately half the risk of re-revision and death compared to non-ARMD revisions. We suspect worldwide regulatory authorities have positively influenced outcomes following ARMD revision by widely recommending that surgeons exercise a lower revision threshold. Our findings suggest the threshold for ARMD revision surgery need not be lowered further. The high risk of failure following MoMHR revision for infection and dislocation is concerning


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 47 - 47
1 Apr 2017
Jelsma J Senden R Schotanus M Kort N Grimm B Heyligers I
Full Access

Background

The second generation metal-on-metal (MoM) prosthesis of the hip became a worldwide success in the 90s. However, after the placement of a MoM prosthesis the cobalt ion concentrations raise significantly. This may lead to systemic complaints and even cobalt toxicity.

Methods

Sixty-one patients (26F/35M) with both an unilateral and bilateral resurfacing or large-head MoM (LHMoM) hip prosthesis were included. At last follow-up (5.77 ± 1.57 yrs) cobalt concentrations in the blood were determined by ICP-MS. Based on the known cobalt toxicity symptoms we developed a non-validated questionnaire. Analysis was done on two groups; a low cobalt concentration group and a high cobalt concentration group. We used 170 nmol/L as the upper limit of well functioning prosthesis as defined by the Dutch Orthopaedic Society (NOV). Independent samples T test and Pearson correlation coefficient were done.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 83 - 83
1 Dec 2016
Burnell C Flynn B Gascoyne T Stockwell K Turgeon T
Full Access

Non-large head Metal-on-metal (MoM) hip replacements were seen as a solution to concerns about implant wear in younger patients. Mid-term loosening of once well-fixed hydroxyapatite (HA) coated femoral stems was recently observed in select MoM patients upon revision surgery. Accordingly, an implant retrieval study was undertaken to examine the incidence of aseptic loosening of in HA-coated femoral stems with MoM, ceramic on ceramic (CoC) and metal on polyethylene (MoP) bearing couples.

A single-centre implant retrieval lab reviewed 44 hydroxyapatite (HA)-coated titanium wedge taper stems of the same design retrieved over a period of 9 years. Ten were MoM articulations, 23 MoP and 11 CoC. Head sizes ranged from 28 to 40 with only four 40mm heads, all of which were MoM. Reason for revision, duration of implantation, femoral head size, patient age and body mass index was recorded for each retrieval. Goldberg corrosion scores were determined for the taper surfaces of each retrieval, with ‘0’ indicating no corrosion and ‘3’ indicating severe corrosion. Logistic regression analysis, Wilcoxan Rank Sum and Fischer's exact test were used for statistical analysis.

Aseptic loosening was the listed reason for revision in 18 of 44 cases. MoM bearing was associated with increased probability of aseptic loosening (Odds ratio 7.1 (95%CI 1.1–47.0) p=0.042). Severity of corrosion was also associated with aseptic loosening (Odds ratio 2.75 (95%CI 1.1–6.6) p=0.02). Head size and patient age had no correlation. Median time to revision of implants for aseptic loosening was 4.5 years (range: 4.2–7.0 years) for MoM versus 1.4 years (range: 0.3–3.0) for other bearing couples (p=0.004). Aseptic loosening was categorised as early (<=2 years) or mid-term (>2 years). No MoM hips were revised for aseptic loosening in the first 2 years while 8 of the 11 mid-term revisions had MoM articulations (p=0.004). Taper corrosion was more severe in mid-term aseptic loosing cases (p=0.049).

MoM HA-coated hip replacements appear to be associated with increased mid-term aseptic loosening compared to other bearing couples. Patients with MoM HA-coated hip replacements should be monitored regularly beyond the initial 1 to 2 years following surgery. Future analyses will examine the presence and progression of femoral radiolucency prior to revision surgery to determine an approximate timeline of stem loosening in this patient cohort. This research highlights the importance of implant retrieval programs to assess post-revision implant characteristics for early identification of possible device issues.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 49 - 49
1 Jul 2014
Gwyn R Mahmood S Malik I Maheson M John A Lyons C Jones S
Full Access

Summary

162 patient cohort with serial Metal Artefact Reduction Sequence MRI scans. Patients with normal initial scans can be followed up at 1 year. Those with abnormal scans should be followed up at a shorter interval of 6 months.

Introduction

Cross-sectional imaging is a key investigation in the assessment and surveillance of patients with metal-on-metal (MoM) hip arthroplasty. We present our experience of Metal Artefact Reduction Sequence (MARS) MRI scanning in metal on metal hip arthroplasty. We aimed to investigate the natural history and radiological disease progression from Adverse Reactions to Metallic Debris.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 38 - 38
1 Jan 2018
Matharu G Pandit H Murray D Judge A
Full Access

Pseudotumours have recently been reported in non-metal-on-metal total hip replacements (non-MoM THRs), however the magnitude and risk factors for this complication are unknown, as is the outcome of its treatment.

3340 primary THR undergoing revision for pseudotumour between 2008 and 2015 were identified in the National Joint Registry for England and Wales. 7.5% (n=249) of these pseudotumour revisions, had non-MoM bearing surfaces. The risk of revision for pseudotumour in non-MoM hips was 0.032% (249/789,397; 95% CI 0.028%–0.036%). The risk of pseudotumour revision was 2.35 times (95% CI 1.76–3.11) higher in ceramic-on-ceramic compared with hard-on-soft bearings, and 2.80 times (95% CI 1.74–4.36) higher in 36mm metal-on-polyethylene bearings compared to 28mm and 32mm metal-on-polyethylene bearings.

The outcome of revision for pseudotumour non-MoM hips was studied in 185 hips. 13.5% (n=25) had re-revisions at a mean of 1.2 years (range 0.1–3.1 years). Infection (32%), dislocation/subluxation (24%), and aseptic loosening (24%) were the commonest indications for re-revision. The 4-year survival rate was 83.8% (95% CI=76.7%–88.9%). Multiple revision indications (Hazard Ratio (HR)=2.78; 95% CI=1.03–7.49) and incomplete revision procedures (HR=5.76; 95% CI=1.28–25.9) increased the risk of re-revision

Although the overall risk of revision for pseudotumour in non-MoM THRs is low, the risk is increasing and is significantly higher in ceramic-on-ceramic and large head metal-on-polyethylene THR. These revisions have a high early failure rate.


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1020 - 1027
1 Aug 2017
Matharu GS Judge A Pandit HG Murray DW

Aims

To determine the outcomes following revision surgery of metal-on-metal hip arthroplasties (MoMHA) performed for adverse reactions to metal debris (ARMD), and to identify factors predictive of re-revision.

Patients and Methods

We performed a retrospective observational study using National Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery for ARMD between 2008 and 2014. The outcomes studied following revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using competing-risk regression modelling.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 65 - 65
1 Mar 2006
Witzleb W Hanisch U Guenther K
Full Access

Question: Is the histopathological response of the peri-prosthetic tissue to metal-on-metal bearings comparable to the well studied reactions to polyethylene debris or do specific reactions exist and are theses reactions depending on the implant design?

Methods: Periprosthetic tissue samples from 19 THR and Hip Resurfacings (11 Birmingham Hip Resurfacings, 2 (historical)McMinn Hybrid Hip Resurfacings, 5 MetaSUL THR) with a variety of failure mechanisms were examined histopathologically and immunohistochemically.

Results: Only the samples of the (historical) McMinn Hybrid resurfacings showed a stronger histiocytic foreign body reaction and a higher grade metallosis. In all other cases only a mild if any histiocytic foreign body reaction was found. Additionally a chronic lymphoplasmacellular tissue reaction was present in all cases. 3 cases showed a higher grade chronic lymphoplasmacellular inflammatory tissue response comparable to a specific immune reaction.

Discussion: We found two different response mechanisms of the periprosthetic tissues to metal-on-metal bearings. In addition to the classic histiocytic foreign body reaction which was usual mild and only stronger in cases with a greater amount of metallic debris a lymphoplasmacellular inflammatory reaction usually was present but did not reach the level of inflammatory alterations associated with potential osteolysis and was not related to the implant or amount of wear. In 3 of our 19 cases the lymphoplasmacellular infiltration was stronger and comparable to a specific immune reaction. The significance of the specific immune reaction could not be further explained because a correlation to the failure mechanisms was not detectable


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 24 - 24
1 Nov 2015
Matharu G Mellon S Murray D Pandit H
Full Access

Introduction

This study aimed to: (1) compare published follow-up guidelines for metal-on-metal (MoM) hip patients and analyse protocols in relation to current evidence, and (2) assess the financial implications of these guidelines.

Methods

Follow-up guidance for MoM hips from five national authorities (MHRA in the UK; EFORT; United States FDA; Therapeutic Goods Administration of Australia; Health Canada) were contrasted and critically appraised. Using National Joint Registry (NJR) data (67,363 MoM hips implanted) the cost of annual surveillance for all MoM hips recorded in the NJR was calculated for each protocol.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_10 | Pages 19 - 19
1 Oct 2015
Vasukutty Rajput Shaikh Uzoigwe Howes Minhas
Full Access

Introduction

The Medicines and Healthcare Products Regulatory Agency (MHRA) of the UK have published guidelines for annual follow up of patients with metal on metal hip replacements following widespread concern regarding metallosis

Methods

We followed up 718 total hip replacements (594 stemmed hips and 124 resurfacing) with metal on metal bearing, implanted between April 1999 and August 2010 in dedicated clinics with clinical and radiological assessment along with assessment of serum metal ions. Survival was calculated using Kaplan Meier analysis.