Introduction: The hip resurfacing in younger patients is being performed more frequently in the UK. At the same time conventional Total Hip Replacement (THR) is also being performed.. We carried out a study to compare the patient satisfaction and outcome following Birmingham Hip Resurfacing (BHR) and Total Hip Replacement (THR) in patients below the age of 55 years. Methods: There were 93
Introduction: The Birmingham Hip Resurfacing (BHR) system comprises both a
This study reports the mid-term results of a large bearing uncemented metal on metal total hip replacement (MOMHTHR) matched series using the Synergy stem and Birmingham modular head in 36 hips (mean follow up 61 months). All patients underwent clinical, metal ion and MRI assessment. Wear analysis was performed on retrieved heads using Redlux non-contact optical profilometry. Seven patients (19%) have undergone revision surgery. All revisions had two or more of either symptoms, high metal ions or an MRI suggestive of an adverse reaction to metal debris (ARMD). There was no evidence of component malposition or impingement. Frank staining of tissues together with high volume dark brown fluid collections were found in all cases. All stems and cups were well fixed. In 4 cases pubic and ischial lysis (adjacent to the inferior fins) was observed. All 7 cases had radiological, intraoperative and histological evidence of ARMD (Figure 1). The failure cohort had significantly higher whole blood cobalt ion levels and OHS (p = 0.001), but no significant difference in cup size (p = 0.77), gender predominance, stem offset or cup position (p = 0.12). Sleeves had been used in all revision cases. Wear analysis (n = 4) demonstrated increased wear at the trunnion/sleeve interface in a distribution compatible with micromotion (Figure 2). There was normal wear at the articulating surface. This series further demonstrates unacceptable failure rates in LHMOMTHR in a series where a compatible stem for the
The difference in the mean values regarding inclination was greater than would be expected by chance; there was a statistically significant difference (P = 0,010).
Navigation technique was discussed to equalize the drawback of MIS. However, tools like imageless navigation may further improve the cup position even in traditional approach.
We previously reported the five to ten-year results of the Birmingham Hip Resurfacing (BHR) implant. The purpose of this study was to evaluate the survivorship, radiographic results, and clinical outcomes of the
Reports of improved functional outcome of Metal on Metal Hip Resurfacing Arthroplasty (mHRA) to Total Hip Replacement needs to be balanced with concerns of metal ion release. By removing cobalt-chrome, cHRA reduces these risks. To the author's knowledge, there is no data available on functional outcomes of cHRA, therefore the aim of the study was to compare the function between cHRA patients and mHRA patients. 24 patients received a unilateral cHRA (H1, Embody) and was compared to 24 age and gender matched patients with a unilateral mHRA (BHR, Smith and Nephew). All patients completed the Oxford Hip Score (OHS)[T2] and underwent gait analysis on an instrumented treadmill before and at a mean of 74wks (+/− 10) for mHRA and 53wks (+/− 2) for cHRA post op. Walking trials started at 4km/h and increased in 0.5km/h increments until a top walking speed (TWS) was achieved. Vertical ground reaction forces (GRF) were recorded along with the symmetry index (SI). Spatiotemporal measures of gait were also recorded. Vertical GRF were captured for the entire normalised stance phase using statistical parametric mapping (SPM; CI = 95%). The gain in OHS was similar: H1 (25-46), BHR(27-47). TWS increased by 19% with H1 (6.02 – 8.0km/hr), and 20% with
When the Birmingham Hip Resurfacing (BHR) metal-on-metal implant system was approved by the United States Food and Drug Administration in 2006, a multicenter, prospective, post-approval study (PAS) was required. This study uses data from the PAS to investigate metal level and glomerular filtration rate (GFR) trends over the first decade in vivo. Between October 2006 and March 2011, 290 primary
A concern of metal on metal hip resurfacing arthroplasty is long term exposure to Cobalt (Co) and Chromium (CR) wear debris from the bearing. This study compares whole blood metal ion levels from patients drawn at one-year following Birmingham Hip Resurfacing (BHR) to levels taken at a minimum 10-year follow-up. A retrospective chart review was conducted to identify all patients who underwent a
We present a consecutive case cohort of the first 100 Birmingham Hip Resurfacing (BHR)'s in 90 patients with a minimum follow up of 20 years. All procedures were performed by a single surgeon having commenced the study in 1998. The original cohort included 68 males with 75 hips (7 bilateral) and 22 females with 25 hips (3 bilateral). The mean age at index procedure was 52. Patients were recalled to review in clinic as per Medicines and Healthcare products Regulatory Agency guidelines with x-rays, metal ions and Harris Hip Scores recorded. After a minimum of 20 year follow-up review the known overall revision rate is 11%. 11 have died and 7 have been lost to review. In males the known revision rate is 4/75 (5%), 3 of which were due to early fractures and 1 revision for infection at 5 years. The known revision rate in females is 7/25 (28%) of which 6 were due to adverse reactions to metal debris; 3 at 7–8 yrs, 1 at 15 yrs, 2 at 18 years, and 1 at 2 months for avascular necrosis. The mean 20 year metal ions results were cobalt 36.6 nmol/l (range 7.4–232.1) and chromium 32.1 nmol/l (range 6–120.8). The mean Harris hip score was 88.5 (range 38.6–99.8). This case series with 20 years follow-up demonstrates excellent long term survival of the BHRs further to registry
The use of Birmingham hip resurfacing (BHR) remains controversial due to the increased revision rate in female patients. We compared the outcomes of
Abstract. Introduction. In general the life expectancy of population is improving. This is causing to increase case load of peri-prosthesis fractures after joint replacements. We present our results of peri-prosthesis fracture around hip managed by revision arthroplasty. Methods. A retrospective analysis of 24 consecutive patients of periprosthetic hip fracture treated with a revision arthroplasty at Major Trauma Centre between February 2021 and January 2022. Results. 12 male and 12 female patients, average age 78 years. 3 fractures around
Birmingham Metal-on-metal total hip resurfacing (BHR) is a bone-conserving option for patients with advanced articular damage. While the outcomes of Total Hip Replacement (THR) are well documented, there is a paucity of literature comparing the patient reported outcomes of
The 22 year survivorship of metal on metal hip resurfacing arthroplasty (RSA) is reported to be 94.3% with expert surgeons, in males with head sizes greater than 48mm. The 2023 National Joint Registry (NJR) report estimates survivorship of all RSA at 19 years to be 85%. This estimate includes all designs, head sizes and females. Our aim was to estimate the survivorship of RSA currently available for implantation (males only, head size >48mm, MatOrtho Adept or Smith and Nephew Birmingham Hip Resurfacing (BHR)) in those under 55 years, performed by all surgeons, compared to conventional THR. We performed a retrospective analysis of the NJR. We included all males under 55 years who had undergone
Hip resurfacing may be a useful surgical procedure when patient selection is correct and only implants with superior performance are used. In order to establish a body of evidence in relation to hip resurfacing, pseudotumour formation and its genetic predisposition, we performed a case-control study investigating the role of HLA genotype in the development of pseudotumour around MoM hip resurfacings. All metal-on-metal (MoM) hip resurfacings performed in the history of the institution were assessed. A total of 392 hip resurfacings were performed by 12 surgeons between February 1st 2005 and October 31st 2007. In all cases, pseudotumour was confirmed in the preoperative setting on Metal Artefact Reduction Sequencing (MARS) MRI. Controls were matched by implant (ASR or BHR) and absence of pseudotumour was confirmed on MRI. Blood samples from all cases and controls underwent genetic analysis using Next Generation Sequencing (NGS) assessing for the following alleles of 11 HLA loci (A, B, C, DRB1, DRB3/4/5, DQA1, DQB1, DPB1, DPA1). Statistical significance was determined using a Fisher's exact test or Chi-Squared test given the small sample size to quantify the clinical association between HLA genotype and the need for revision surgery due to pseudotumour. Both groups were matched for implant type (55% ASR, 45%
The Birmingham Hip Resurfacing (BHR) has become increasingly popular for young active patients with severe OA of the hip. Although it has been in use in the UK since 1991, none were performed in NZ until late 2004. We have reviewed our initial results &
have compared them to a similar group of patients who received conventional total hip replacement (THR) with ceramic-on-ceramic bearings. All patients aged 65 years or less undergoing either
Surface replacement arthroplasty (SRA) has been proposed as a viable option for the treatment of osteoarthritis in young, active patients. Positive results of the Birmingham Hip Resurfacing (BHR) in select patient groups have been described in international series and registry data. We report 5–10 year U.S. follow-up for the
Background: There are no large comparative metal ion studies of commercially available hip resurfacing devices which have taken into account the effects of femoral size and cup inclination and anteversion. Patients and methods: Metal ion analysis is carried out routinely at our independent centre. We present the metal ion results of 95 unilateral ASR patients and 70 unilateral
Introduction: We compared the eight- to ten-year clinical and radiological results of the metal-on-metal Birmingham hip resurfacing (BHR) with a hybrid total hip arthroplasty (HYBRID) in two groups of 54 hips, previously matched for gender, age, body mass index and activity level. Method: Patients were followed up in outpatients and function assessed by using the Oxford Hip Score, UCLA activity score and Euroqol score. Radiographs were assessed for osteolysis and wear.
Introduction. Metal on metal hip arthroplasty continues to be controversial. Emerging evidence suggests that there are multiple modes of failure, and that the results of revision surgery are influenced by host and implant factors. Methods. This study compares a single surgeon series of hip resurfacings (Birmingham Hip Resurfacing {BHR}) and large diameter metal on metal total hip replacements (LDMOMTHR). Primary outcome measures included survival rates, failure secondary to histologically identified Adverse Reaction to Metal Debris (ARMD), and patient reported outcome measures (Oxford Hip Score {OHS}) following revision. Between 1999 and 2005, 458
Introduction. Many worldwide regulatory authorities recommend regular surveillance of metal-on-metal hip arthroplasty patients given high failure rates. However concerns have been raised about whether such regular surveillance, which includes asymptomatic patients, is evidence-based and cost-effective. We determined: (1) the cost of implementing the 2015 MHRA surveillance in “at-risk” Birmingham Hip Resurfacing (BHR) patients, and (2) how many asymptomatic hips with adverse reactions to metal debris (ARMD) would have been missed if patients were not recalled. Methods. All