Two mechanical devices for prophylaxis of thromboembolism after total knee arthroplasty

A PROSPECTIVE, RANDOMISED STUDY



Download PDF

Abstract

The optimal characteristics of pneumatic compression for mechanical prophylaxis of thromboembolism after total knee arthroplasty (TKA) are not known. Our study compared two methods of calf compression, with the hypothesis that the device which provided a larger increase in peak venous velocity would produce a lower rate of thromboembolism. We performed a prospective, randomised study on 423 patients (472 knees). Duplex ultrasonography was carried out by experienced technicians who were blinded to the device used. Overall, 206 patients (232 knees) used a rapid inflation, asymmetrical compression (RIAC) device and 217 (240 knees) a sequential circumferential compression device (SCD).

The rate of venous thromboembolism was 6.9% with the RIAC device compared with 15% for the SCD device (p = 0.007). The incidence of thrombi with unilateral primary TKA was 8.4% for the RIAC compared with 16.8% for the SCD device (p = 0.03). In 47 patients with a bilateral TKA, the incidence of thrombi was 4% for the RIAC compared with 22.7% for the SCD device (p = 0.05 per knee). There was a low rate of mortality and pulmonary embolism when using mechanical prophylaxis for thromboembolism after TKA. Our findings show that the use of rapid inflation, asymmetrical calf compression gave a significantly lower rate of thromboembolism.


Correspondence should be sent to Professor P. F. Lachiewicz.

For access options please click here