Abstract
Objectives
The bony shoulder stability ratio (BSSR) allows for quantification of the bony stabilisers in vivo. We aimed to biomechanically validate the BSSR, determine whether joint incongruence affects the stability ratio (SR) of a shoulder model, and determine the correct parameters (glenoid concavity versus humeral head radius) for calculation of the BSSR in vivo.
Methods
Four polyethylene balls (radii: 19.1 mm to 38.1 mm) were used to mould four fitting sockets in four different depths (3.2 mm to 19.1mm). The SR was measured in biomechanical congruent and incongruent experimental series. The experimental SR of a congruent system was compared with the calculated SR based on the BSSR approach. Differences in SR between congruent and incongruent experimental conditions were quantified. Finally, the experimental SR was compared with either calculated SR based on the socket concavity or plastic ball radius.
Results
The experimental SR is comparable with the calculated SR (mean difference 10%, sd 8%; relative values). The experimental incongruence study observed almost no differences (2%, sd 2%). The calculated SR on the basis of the socket concavity radius is superior in predicting the experimental SR (mean difference 10%, sd 9%) compared with the calculated SR based on the plastic ball radius (mean difference 42%, sd 55%).
Conclusion
The present biomechanical investigation confirmed the validity of the BSSR. Incongruence has no significant effect on the SR of a shoulder model. In the event of an incongruent system, the calculation of the BSSR on the basis of the glenoid concavity radius is recommended.
Cite this article: L. Ernstbrunner, J-D. Werthel, T. Hatta, A. R. Thoreson, H. Resch, K-N. An, P. Moroder. Biomechanical analysis of the effect of congruence, depth and radius on the stability ratio of a simplistic ‘ball-and-socket’ joint model. Bone Joint Res 2016;5:453–460. DOI: 10.1302/2046-3758.510.BJR-2016-0078.R1.
Article focus
-
Is the bony shoulder stability ratio (BSSR) valid?
-
Does incongruence affect the stability ratio (SR) of a shoulder model?
-
Is it valid to use the radius of the glenoid concavity of an incongruent shoulder joint in order to calculate the BSSR in vivo?
Key messages
-
The BSSR was proved valid for congruent and incongruent systems.
-
For the calculation of the BSSR in vivo, the use of the radius of the glenoid concavity is recommended over the use of the humeral head radius.
-
Our study findings support the use of the BSSR for in vivo CT-based determination of bony shoulder stability.
Strengths and limitations
-
First biomechanical study analysing the effect of joint incongruence on the SR.
-
This study determines the correct parameters (glenoid concavity radius versus humeral head radius) for calculation of the BSSR in vivo.
-
Material properties and friction coefficients are limitations of the present study.
Introduction
In shoulder instability research, the ratio of the peak translational force causing dislocation to the compressive load that provides stability is called the stability ratio (SR).1,2 The ratio is determined by the morphology of the articulating surfaces,1,3 and in clinical practice largely depends on the integrity and geometry of the glenoid. It is known that patients with a glenoid defect are at high risk of recurrent instability,4,5 which necessitates glenoid reconstruction surgery by means of coracoid transfers, iliac crest bone graft transfers, or allografting.6-11 On the other side, atraumatic and traumatic shoulder instability, without any sign of glenoid bone loss, is associated with an inherent flattening of the bony glenoid concavity.12
Thus far, glenohumeral stability in experimental studies has been described either by the SR1,3,5,13 or by displacement after applying a predetermined translational force.14-18 Oosterom et al19 aimed to predict stability by means of the “translational stiffness” of the shoulder after joint replacement. Based on their mathematical model for circular shapes, translational force over the “distance to dislocation” can be calculated. Most recently, Willemot et al20 derived a mathematical model based on machine design principles. The humeral head as a follower translates over the cam-like glenoid surface, an applicable concept for non-circular shapes like the native glenohumeral anatomy. However, none of the reported studies described a method allowing for calculation of glenohumeral stability in daily clinical practice.
Lazarus et al1 aimed to approximate the SR based on mathematical considerations. The relationship between the effective glenoid concavity depth and the SR was described by an empirical equation. This equation was based on a linear relationship between the two factors. Moroder et al12 have presented a theoretical model to calculate the bony shoulder stability ratio (BSSR). In this CT-based study, the BSSR was calculated based on the patients’ glenoid morphology, allowing for quantification of the passive bony stabilisers of a human shoulder. It was shown that the relationship between the glenoid concavity depth and the SR is non-linear.
The BSSR is based on a theoretical congruent joint, the humeral head and the glenoid concavity having the same radius of curvature. Calculations of the BSSR in vivo were therefore conducted with both glenoid and humeral head radii.12 However, several studies have shown that the bony structures of the glenohumeral joint are incongruent, with the glenoid concavity having a larger radius than the humeral head.21-24 Yet it is not clear which of the radii can be used for valid in vivo calculations of the bony shoulder stability.
The first aim of the present study was to validate the BSSR with a biomechanically relevant model. Secondly, we aimed to analyse the effect of joint incongruence on a shoulder model and subsequently on a mathematical equation based on a theoretical congruent joint. Finally, we aimed to analyse which of the radii, glenoid concavity or humeral head, should be used for valid calculations of the BSSR in vivo. We hypothesised that the congruent experimental SR would not be significantly different from the calculated SR based on the BSSR approach, that incongruence has no effect on the SR of a simplistic ball-and-socket joint, and that the radius of the glenoid concavity, rather than of the humeral head, should be used for calculation of the BSSR in vivo.
Materials and Methods
Model preparation
A total of four high density polyethylene (HDPE) plastic balls with four different radii of curvature were used in the present study. Based on morphological analyses,21,25 radii were selected to be 19.1 mm (Ba), 25.4 mm (Bb), 31.8 mm (Bc) and 38.1 mm (Bd). Plastic balls were mounted on an aluminium rod, forming an analogue to a human proximal humerus. Similar plastic balls were used as a pattern to mould four fitting sockets from resin (Smooth-Cast 65D, Smooth-On, Macungie, Pennsylvania) with matching radius of curvature (Sa, Sb, Sc, Sd). Furthermore, each socket was moulded in four different concavity depths: 3.2 mm (I), 6.4 mm (II), 12.7 mm (III) and 19.1 mm (IV). In total, 16 different sockets were created for the experiment. In order to minimise friction, the surfaces of all moulded sockets were smoothed with 240-grit sandpaper.
Testing apparatus
Components were evaluated using a custom multi-axis electromechanical testing machine described in previous studies of shoulder stability.13,26 Briefly, the system consists of a six-axis load cell (model 45E15A-E24ES-A; JR3, Inc., Woodland, California) fitted on a motorised XY table (DCI Design Components, Franklin, Massachusetts), the movement of which simulates anterior-posterior or superior-inferior motion. Sockets were mounted to the XY table (Fig. 1). The z-axis is pneumatically controlled and compressive (medial-lateral) load is applied to the humerus analogue with motion measured with a linear potentiometer (TR-50; Novotechnik, Stuttgart, Germany). According to the manufacturer, the capacity and resolution of the load cell are 100 N and 1.3 N, respectively (Fig. 1). Following previous studies, measuring the point where the ball was seated at the deepest part of the socket concavity identified the neutral position.13,26 A 50-N concavity compression force, selected based on previous studies,1,13,27 was applied to the plastic ball by the pneumatic cylinder. All analyses were based on a normalised displacement, where each head size was displaced one diametrical distance anteriorly at 2 mm/s. The translational force was defined as the maximum force observed as the ball translated across its normalised distance.
Fig. 1
Test conditions
Each experiment and condition was repeated three times to evaluate repeatability of the results.
The peak translational force was measured in two different ways: with congruent radii of curvature between the ball and socket (Fig. 2a) and with a series of incongruent radii of curvature between the ball and socket (Fig. 2b).
Fig.
In the validation series, each ball (Ba to Bd) was translated against a mating socket (Sa to Sd) of each of the different concavity depths (I to IV) (Table I). The ratio of the peak translational force causing dislocation and the constant concavity compression force was defined as the SR for each condition.1,3
Table I.
Test conditions | |||||
---|---|---|---|---|---|
Congruent System | |||||
Ba : Sa (I) | Bb : Sb (I) | Bc : Sc (I) | Bd : Sd (I) | ||
Ba : Sa (II) | Bb : Sb (II) | Bc : Sc (II) | Bd : Sd (II) | ||
Ba : Sa (III) | Bb : Sb (III) | Bc : Sc (III) | Bd : Sd (III) | ||
Ba : Sa (IV) | Bb : Sb (IV) | Bc : Sc (IV) | Bd : Sd (IV) | ||
Incongruent System | |||||
.75 | .80 | .83 | |||
Ba : Sb (I) | Bb : Sc (I) | Bc : Sd (I) | |||
Ba : Sb (II) | Bb : Sc (II) | Bc : Sd (II) | |||
Ba : Sb (III) | Bb : Sc (III) | Bc : Sd (III) | |||
Ba : Sb (IV) | Bb : Sc (IV) | Bc : Sd (IV) |
-
0.75, 0.80 and 0.83 indicate the ratio of the mismatched radii
-
Ba to Bd, radii of curvature of the balls; Sa to Sd, radii of curvature of the sockets; I to IV, concavity depths of the sockets
The second testing series evaluated incongruent radii of curvature. As previously reported, mismatch was defined as the ratio of the ball radius to the radius of curvature of the socket.19,21 Our mismatched conditions had a ratio of 0.75, 0.80 and 0.83. In other words, we articulated the 19.1 mm ball (Ba) with the 25.4 mm socket (Sb) (ratio = 0.75), the 25.4 mm ball (Bb) with the 31.8 mm socket (Sc) (ratio = 0.80) and the 31.8 mm ball (Bc) with the 38.1 mm socket (Sd) (ratio = 0.83). Each ball was translated against the incongruent sockets of four different depths (I to IV) (Table I).
Statistical analysis
The BSSR (i.e. calculated SR) for each ball-and-socket configuration was calculated as described by Moroder et al.12 The ratio between translational force (T) and compressive force (C) (Equation 1) was expressed in terms of the radius of curvature (r) and concavity depth (d) based on Pythagorean trigonometric identities (Equation 2). The full derivation of Equations 1 and 2 are given in the supplementary material.
BSSRs calculated from this approach were validated against SRs derived from experimental data of a congruent system. To analyse the influence of incongruence, experiment-derived SRs of congruent conditions were compared with their incongruent counterparts. In order to determine the correct parameters (glenoid concavity radius versus humeral head radius) for calculation of the BSSR in vivo, the experimental SR was compared with the calculated SR based on the socket concavity or plastic ball radius.
In the validation study, data were analysed by comparing SRs with the same radii of curvature of the ball and socket but different depths (e.g. experimental Ba : Sa (I to IV) versus calculated Ba : Sa (I to IV)) and with same depth but different radii of curvature (e.g. experimental Ba-d : Sa-d (I) versus calculated Ba-d : Sa-d (I)).
In order to prove the influence of incongruence on the SR of a simplistic ball-and-socket joint, data were analysed by comparing only experimental SRs with the same radii of curvature (same mismatched radii) of the ball and socket, but different depths (e.g. 0.75 mismatch; Bb : Sb (I to IV) versus Ba : Sb (I to IV)).
Finally, the experimental SR was compared with the calculated SR based on the radius of the socket concavity and the calculated SR based on the radius of the plastic ball. Data were analysed by comparing the experimental SRs with each of the corresponding calculated SR based on the socket concavity or plastic ball radius (e.g. experimental Ba : Sb (I to IV) versus calculated Sb (I to IV); experimental Ba : Sb (I to IV) versus calculated Ba (I to IV)).
Paired sample t-tests or Wilcoxon signed-rank tests were applied using SPSS Statistics (Version 22, IBM Corporation, Chicago, Illinois). In addition, differences between experimental versus calculated SR and congruent versus incongruent SR were plotted on a graph. In order to evaluate repeatability of the results, test-retest reliability was proved. A p-value less than 0.05 indicated statistical significance.
Results
Experimental versus calculated SR
In the test condition Ba (IV) :): Sa (IV), the measured translational force exceeded the load cell capacity. These results were excluded.
The mean difference in this series was 10% (sd 8%) (relative value) (Fig. 3). Comparing the SRs of the simplistic ball-and-socket joints, no significant differences were observed, except in the comparisons Bb : Sb (I to IV) (p = 0.010;) and Bd : Sd (I to IV) (p = 0.049), as well as in Ba-d : Sa-d (I) (p = 0.029) and Ba-d : Sa-d (III) (p = 0.026). Of these significant comparisons, the mean difference in Bb : Sb (I to IV) was 11% (sd 5%) and in Bd : Sd (I to IV) 3% (sd 3%) (relative value). On the other side, the mean difference in Ba-d : Sa-d (I) was 8% (sd 5%) and in Ba-d : Sa-d (III) 12% (sd 8%) (relative value).
Fig. 3
The test-retest reliability of the congruent experimental SRs was proved excellent (R = 0.995).
Congruent versus incongruent SR
In this experimental study, the mean difference was 2% (sd 2%) (relative value) (Fig. 4). SRs of congruent compared with incongruent simplistic joints revealed no significant differences, except for the comparison Bd : Sd (I to IV) versus Bc : Sd (I to IV) (0.83 mismatch). In this significant comparison, the mean difference gave a result of 1% (sd 1%) (relative value).
Fig. 4
Repeatability of the incongruent experimental SRs observed an excellent test-retest reliability (R = 0.995).
Experimental versus calculated SR based on socket concavity radius and experimental versus calculated SR based on plastic ball radius
Comparing the experimental SR of an incongruent system with the corresponding calculated SR based on the radius of the socket concavity, a mean difference of 10% (sd 9%) (relative value) was observed. When comparing the experimental SR with the corresponding calculated SR based on the radius of the plastic ball, a mean difference of 42% (sd 55%) (relative value) was proved.
Discussion
The biomechanical relationship between the SR and the glenoid concavity was explained in previous studies.1,3,13,28 Lippitt et al3 showed that the greater the glenoid concavity, the greater the SR. Yamamoto et al5,28 observed that loss of glenoid concavity due to glenoid bone loss results in a significant decrease of SR of the shoulder,5 which cannot be overcome by mere capsulolabral repair but necessitates glenoid reconstruction surgery.28 The measurement of the SR is, however, only possible by means of in vitro biomechanical testing. Until recently, there was no technique available to estimate the SR of a glenohumeral joint in vivo. Moroder et al12 presented a CT-based technique, which allows for prediction of the SR of the glenohumeral joint. It was named the bony shoulder stability ratio (BSSR) since it only takes into consideration the bony glenoid morphology visible on CT scans and neglects the stability-contributing effects of soft-tissue structures. The BSSR represents an easy method to calculate the stability ratio created by the bony glenoid concavity and thus offers the possibility to calculate the actual biomechanical effect of glenoid concavity alterations or glenoid reconstruction surgery in vivo. However, in the reported study it could not be clarified whether the mathematical equation based on a theoretical congruent joint is also valid for in vivo measurements on an incongruent glenohumeral joint with mismatching radii of curvature.
In the present study we aimed to biomechanically validate the BSSR by comparing the experimental SR of a congruent system with the calculated SR based on the BSSR approach. According to our results, almost all experimental setups yielded the same results as calculated with the theoretical model. The mean relative difference in this series was 10%. The comparisons with statistically significant differences showed mean relative differences of not more than 12%. Although acceptably small, the presence of differences between the experimental and calculated SRs is evident. Willemot et al20 observed in his cam-follower study on mid-range instability a similar tendency, whereas experimental SRs were constantly smaller than the calculated SRs and relied on elasticity and friction in the experimental model. The validity of the BSSR has previously been confirmed by Finite Element Analysis,12 which revealed only negligible discrepancies between the computer simulation and the BSSR. Accordingly, our results also prove the validity of the BSSR.
The observed experimental SRs were consistently slightly smaller than the calculated SRs. According to Moroder et al,12 increasing the radius of curvature is indirectly related to the SR. We assume that due to the constant concavity load, local deformation of the material occurs, leading to an increased local radius of curvature of the socket and resulting in a smaller experimental SR compared with the calculated SR.
Previous studies described the bony morphology of the glenohumeral joint as a “golf ball on a tee” phenomenon.12,29 A completely flat “tee” would result in a theoretical BSSR of 0%.12 On the other side, increasing the concavity depth results in a steep increase of the BSSR. Increasing the concavity depth up to its radius of curvature, the BSSR would increase exponentially and the ball would be secured against any translational force. It was described as “placing the golf ball in an egg cup”.12 The same finding was observed in this study. A radius of curvature of the ball of 19.1 mm (Ba IV) and a concavity depth of 19.1 mm (Sa IV) led to translational forces exceeding the load cell capacity, and these results had to be excluded.
The BSSR is based on a congruent articulating system with the same radii of curvature for both the glenoid concavity and the humeral head.12 However, it is known that the cartilage of the humeral head and the glenoid are better conforming than the underlying bone, which feature a mismatch regarding their radii.21 According to our study analysing the effect of incongruence on a shoulder model, with a mismatch of 0.75 and 0.80, no significant differences between the congruent and incongruent experimental SRs could be noted. Only one experimental setup resulted in a significant difference between the congruent and incongruent experimental system, however, the mean relative difference was 1% (Fig. 4), and almost negligible. Although Hopkins et al30 observed that congruence in anatomic total shoulder arthroplasty is linearly related to humeral head translations, they also concluded that the force ratio (i.e. the SR) required to destabilise the shoulder joint is not affected by congruence. Furthermore, Karduna et al31 described similar patterns of humeral head translation in native and reconstructed shoulder joints. These findings support our second conclusion, that incongruence has no effect on the SR of a simplistic ball-and-socket joint.
The BSSR was calculated using not only the radius of the glenoid concavity, but also the humeral head radius.12 A constantly increased BSSR of 10% was observed. It remained unclear which of the radii can be used for valid in vivo calculations of the bony shoulder stability. It was postulated that calculating the BSSR using the glenoid concavity radius is key in order to estimate the bony shoulder stability based on the mathematical equation for congruent radii. In the present study, comparing the experimental SR of an incongruent system with the corresponding calculated SR based on the radius of the socket concavity observed a mean relative difference of 10%. In comparison, using the radius of the plastic ball for the corresponding calculated SR, the mean relative difference was over 40%. Our results conform to recently published studies describing that, independently of the humeral head radius, the glenoid concavity depth and radius are the determining variables of the SR.12,19 Oosterom et al19 showed that in shoulder joint arthroplasty, the radius of curvature of the glenoid cavity is, besides the concavity depth, the SR-determining factor. It was observed that varying the humeral head radii results in similar maximum allowable subluxation forces, which are only dependent on the radius of the glenoid concavity.
According to our results, as long as the humeral head radius is no bigger than the radius of the glenoid concavity, calculating the BSSR in vivo based on the radius of the glenoid concavity is valid. Although the BSSR relies on a theoretical congruent glenohumeral joint, the present study proved that this method is a valid clinical measure for patients suffering from recurrent glenohumeral instability. Despite this, the mathematical equation is still a simplified method, and does not consider additional factors such as the glenoid cartilage, the labrum with its concavity-increasing effect,3 or the negative intra-articular pressure.32
Currently, the necessity for bony glenoid reconstruction surgery in patients with anterior shoulder instability is determined by measuring the lack of glenoid surface area.13,33-35 Considering glenoid bone loss in its 2D extent, however, seems to be overly simplistic and might underestimate the effect of glenoid concavity alterations on stability. This underestimation might be a possible explanation for the high long-term failure rate after soft-tissue-based stabilisation procedures.36-39 The recently introduced BSSR makes it possible to perform an in vivo quantification of the stability ratio created by the glenoid concavity based on CT measurements.12 This additional information could be useful to determine critical glenoid defects more precisely. Our results prove the validity of the BSSR formula and provide a reasonable explanation as to which variable should be used in order to calculate the patient-specific BSSR in daily clinical practice.
There are limitations in this study, namely the material properties and friction coefficients of the ball-and-socket models. For the validation of the BSSR, a mathematical equation, this systematic error is negligible. Another limitation is the size effect of the incongruent experimental setup with increasingly mismatched radii of curvature. The size of the model could have led to a scaling effect on the force data, thus influencing the results of the incongruent experimental setup.
In conclusion, biomechanical testing confirmed the validity of the BSSR. Incongruence has no effect on the SR of a simplistic ball-and-socket joint. For the calculation of the BSSR in vivo, the use of the radius of the glenoid concavity is recommended over the use of the humeral head radius. Finally, our study findings support the use of the BSSR for in vivo CT-based determination of bony shoulder stability.
Supplementary material
A description of the relation of the translational (T) and compressive (C) force vectors to the geometric parameters radius (r) and concavity depth (d) of the ball-and-socket configuration can be found alongside this paper at http://www.bjr.boneandjoint.org.uk/
Funding Statement
This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.
ICMJE conflict of interest
The authors confirm that there are no conflicts of interest associated with this publication. The authors, their immediate family, and any research foundation with which they are affiliated did not receive any financial payments or other benefits from any commercial entity related to the subject of this article. There was no outside funding or grants received that assisted in this study.
References
1 Lazarus MD , SidlesJA, HarrymanDTII, MatsenFAIII. Effect of a chondral-labral defect on glenoid concavity and glenohumeral stability. A cadaveric model. J Bone Joint Surg [Am]1996;78-A:94-102.CrossrefPubMed Google Scholar
2 Fukuda K , ChenCM, CofieldRH, ChaoEY. Biomechanical analysis of stability and fixation strength of total shoulder prostheses. Orthopedics1988;11:141-149.CrossrefPubMed Google Scholar
3 Lippitt SB , VanderhooftJE, HarrisSL, et al.. Glenohumeral stability from concavity-compression: A quantitative analysis. J Shoulder Elbow Surg1993;2:27-35.CrossrefPubMed Google Scholar
4 Burkhart SS , De BeerJF. Traumatic glenohumeral bone defects and their relationship to failure of arthroscopic Bankart repairs: significance of the inverted-pear glenoid and the humeral engaging Hill-Sachs lesion. Arthroscopy2000;16:677-694.CrossrefPubMed Google Scholar
5 Yamamoto N , ItoiE, AbeH, et al.. Effect of an anterior glenoid defect on anterior shoulder stability: a cadaveric study. Am J Sports Med2009;37:949-954.CrossrefPubMed Google Scholar
6 Auffarth A , SchauerJ, MatisN, et al.. The J-bone graft for anatomical glenoid reconstruction in recurrent posttraumatic anterior shoulder dislocation. Am J Sports Med2008;36:638-647.CrossrefPubMed Google Scholar
7 Latarjet M . A propos du traitement des luxations récidivantes de l’épaule. Lyon Chir1954;49:994-997. (In French)[[bibmisc]] Google Scholar
8 Provencher MT , GhodadraN, LeClereL, SolomonDJ, RomeoAA. Anatomic osteochondral glenoid reconstruction for recurrent glenohumeral instability with glenoid deficiency using a distal tibia allograft. Arthroscopy2009;25:446-452.CrossrefPubMed Google Scholar
9 Weng PW , ShenHC, LeeHH, WuSS, LeeCH. Open reconstruction of large bony glenoid erosion with allogeneic bone graft for recurrent anterior shoulder dislocation. Am J Sports Med2009;37:1792-1797.CrossrefPubMed Google Scholar
10 Zhao J , HuangfuX, YangX, XieG, XuC. Arthroscopic glenoid bone grafting with nonrigid fixation for anterior shoulder instability: 52 patients with 2- to 5-year follow-up. Am J Sports Med2014;42:831-839.CrossrefPubMed Google Scholar
11 Petrera M , VeilletteCJ, TaylorDW, ParkSS, TheodoropoulosJS. Use of fresh osteochondral glenoid allograft to treat posteroinferior bone loss in chronic posterior shoulder instability. Am J Orthop (Belle Mead NJ)2013;42:78-82.PubMed Google Scholar
12 Moroder P , ErnstbrunnerL, PomwengerW, et al.. Anterior Shoulder Instability Is Associated With an Underlying Deficiency of the Bony Glenoid Concavity. Arthroscopy2015;31:1223-1231.CrossrefPubMed Google Scholar
13 Itoi E , LeeSB, BerglundLJ, BergeLL, AnKN. The effect of a glenoid defect on anteroinferior stability of the shoulder after Bankart repair: a cadaveric study. J Bone Joint Surg [Am]2000;82-A:35-46.CrossrefPubMed Google Scholar
14 Bryce CD , DavisonAC, OkitaN, et al.. A biomechanical study of posterior glenoid bone loss and humeral head translation. J Shoulder Elbow Surg2010;19:994-1002.CrossrefPubMed Google Scholar
15 Elkinson I , GilesJW, FaberKJ, et al.. The effect of the remplissage procedure on shoulder stability and range of motion: an in vitro biomechanical assessment. J Bone Joint Surg [Am]2012;94-A:1003-1012.CrossrefPubMed Google Scholar
16 Peltier KE , McGarryMH, TiboneJE, LeeTQ. Effects of combined anterior and posterior plication of the glenohumeral ligament complex for the repair of anterior glenohumeral instability: a biomechanical study. J Shoulder Elbow Surg2012;21:902-909.CrossrefPubMed Google Scholar
17 Schulze-Borges J , AgneskirchnerJD, BobrowitschE, et al.. Biomechanical comparison of open and arthroscopic Latarjet procedures. Arthroscopy2013;29:630-637.CrossrefPubMed Google Scholar
18 Sekiya JK , JollyJ, DebskiRE. The effect of a Hill-Sachs defect on glenohumeral translations, in situ capsular forces, and bony contact forces. Am J Sports Med2012;40:388-394.CrossrefPubMed Google Scholar
19 Oosterom R , HerderJL, van der HelmFC, SwieszkowskiW, BerseeHE. Translational stiffness of the replaced shoulder joint. J Biomech2003;36:1897-1907.CrossrefPubMed Google Scholar
20 Willemot L , ThoresonA, BreighnerRyan, et al.. Mid-range shoulder instability modeled as a cam-follower mechanism. J Biomech2015;48:2227-2231.CrossrefPubMed Google Scholar
21 Zumstein V , Kraljevic’M, HoechelS, et al.. The glenohumeral joint - a mismatching system? A morphological analysis of the cartilaginous and osseous curvature of the humeral head and the glenoid cavity. J Orthop Surg Res2014;9:34.CrossrefPubMed Google Scholar
22 Iannotti JP , GabrielJP, SchneckSL, EvansBG, MisraS. The normal glenohumeral relationships. An anatomical study of one hundred and forty shoulders. J Bone Joint Surg [Am]1992;74-A:491-500.PubMed Google Scholar
23 Soslowsky LJ , FlatowEL, BiglianiLU, MowVC. Articular geometry of the glenohumeral joint. Clin Orthop Relat Res1992;285:181-190.PubMed Google Scholar
24 Walch G , EdwardsTB, BoulahiaA, et al.. The influence of glenohumeral prosthetic mismatch on glenoid radiolucent lines: results of a multicenter study. J Bone Joint Surg [Am]2002;84-A:2186-2191.CrossrefPubMed Google Scholar
25 Walch G , MesihaM, BoileauP, et al.. Three-dimensional assessment of the dimensions of the osteoarthritic glenoid. Bone Joint J2013;95-B:1377-1382.CrossrefPubMed Google Scholar
26 Tammachote N , SperlingJW, BerglundLJ, et al.. The effect of glenoid component size on the stability of total shoulder arthroplasty. J Shoulder Elbow Surg2007;16(3 Suppl):S102-106.CrossrefPubMed Google Scholar
27 Lippitt S , MatsenF. Mechanisms of glenohumeral joint stability. Clin Orthop Relat Res1993;291:20-28.PubMed Google Scholar
28 Yamamoto N , MurakiT, SperlingJW, et al.. Stabilizing mechanism in bone-grafting of a large glenoid defect. J Bone Joint Surg [Am]2010;92-A:2059-2066.CrossrefPubMed Google Scholar
29 O’Brien SJ , VoosJE, NeviaserAS, DrakosMC. Developmental anatomy of the shoulder and anatomy of the glenohumeral joint. In: RockwoodCA, MatsenFA, eds. The Shoulder. Fourth ed.Philadelphia: WB Saunders Company, 1990:1-31. Google Scholar
30 Hopkins AR , HansenUN, AmisAA, TaylorM, EmeryRJ. Glenohumeral kinematics following total shoulder arthroplasty: a finite element investigation. J Orthop Res2007;25:108-115.CrossrefPubMed Google Scholar
31 Karduna AR , WilliamsGR, WilliamsJL, IannottiJP. Glenohumeral joint translations before and after total shoulder arthroplasty. A study in cadavera. J Bone Joint Surg [Am]1997;79-A:1166-1174.CrossrefPubMed Google Scholar
32 Habermeyer P , SchullerU, WiedemannE. The intra-articular pressure of the shoulder: an experimental study on the role of the glenoid labrum in stabilizing the joint. Arthroscopy1992;8:166-172.CrossrefPubMed Google Scholar
33 Bushnell BD , CreightonRA, HerringMM. Bony instability of the shoulder. Arthroscopy2008;24:1061-1073.CrossrefPubMed Google Scholar
34 Sugaya H , MoriishiJ, DohiM, KonY, TsuchiyaA. Glenoid rim morphology in recurrent anterior glenohumeral instability. J Bone Joint Surg [Am]2003;85-A:878-884.CrossrefPubMed Google Scholar
35 Schrumpf MA , MaakTG, DelosD, et al.. The management of anterior glenohumeral instability with and without bone loss: AAOS exhibit selection. J Bone Joint Surg [Am]2014;96:e12.CrossrefPubMed Google Scholar
36 Castagna A , MarkopoulosN, ContiM, et al.. Arthroscopic bankart suture-anchor repair: radiological and clinical outcome at minimum 10 years of follow-up. Am J Sports Med2010;38:2012-2016.CrossrefPubMed Google Scholar
37 Cheung EV , SperlingJW, HattrupSJ, CofieldRH. Long-term outcome of anterior stabilization of the shoulder. J Shoulder Elbow Surg2008;17:265-270.CrossrefPubMed Google Scholar
38 Fabre T , Abi-ChahlaML, BillaudA, GenesteM, DurandeauA. Long-term results with Bankart procedure: a 26-year follow-up study of 50 cases. J Shoulder Elbow Surg2010;19:318-323.CrossrefPubMed Google Scholar
39 Moroder P , OdorizziM, PizzininiS, et al.. Open Bankart Repair for the Treatment of Anterior Shoulder Instability without Substantial Osseous Glenoid Defects: Results After a Minimum Follow-up of Twenty Years. J Bone Joint Surg [Am]2015;97-A:1398-1405.CrossrefPubMed Google Scholar