When fracture of an extensively porous-coated
femoral component occurs, its removal at revision total hip arthroplasty
(THA) may require a femoral osteotomy and the use of a trephine.
The remaining cortical bone after using the trephine may develop
thermally induced necrosis. A retrospective review identified 11
fractured, well-fixed, uncemented, extensively porous-coated femoral
components requiring removal using a trephine with a minimum of
two years of follow-up. The mean time to failure was 4.6 years (1.7 to 9.1, standard
deviation ( A total of four patients (36.4%) required further revision: three
for instability and one for fracture of the revision component.
There was no statistically significant difference in the mean Harris
hip score before implant fracture (82.4; These findings suggest that removal of a fractured, well-fixed,
uncemented, extensively porous-coated femoral component using a
trephine does not compromise subsequent fixation at revision THA
and the patient’s pre-operative level of function can be restored.
However, the loss of proximal bone stock before revision may be associated
with a high rate of dislocation post-operatively. Cite this article: