Patient-specific glenoid guides (PSGs) claim an improvement in
accuracy and reproducibility of the positioning of components in
total shoulder arthroplasty (TSA). The results have not yet been
confirmed in a prospective clinical trial. Our aim was to assess
whether the use of PSGs in patients with osteoarthritis of the shoulder
would allow accurate and reliable implantation of the glenoid component. A total of 17 patients (three men and 14 women) with a mean age
of 71 years (53 to 81) awaiting TSA were enrolled in the study.
Pre- and post-operative version and inclination of the glenoid were
measured on CT scans, using 3D planning automatic software. During
surgery, a congruent 3D-printed PSG was applied onto the glenoid
surface, thus determining the entry point and orientation of the
central guide wire used for reaming the glenoid and the introduction
of the component. Manual segmentation was performed on post-operative
CT scans to compare the planned and the actual position of the entry
point (mm) and orientation of the component (°).Aims
Patients and Methods