Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Shoulder & Elbow

Patient-specific glenoid guides provide accuracy and reproducibility in total shoulder arthroplasty



Download PDF

Abstract

Aims

Patient-specific glenoid guides (PSGs) claim an improvement in accuracy and reproducibility of the positioning of components in total shoulder arthroplasty (TSA). The results have not yet been confirmed in a prospective clinical trial. Our aim was to assess whether the use of PSGs in patients with osteoarthritis of the shoulder would allow accurate and reliable implantation of the glenoid component.

Patients and Methods

A total of 17 patients (three men and 14 women) with a mean age of 71 years (53 to 81) awaiting TSA were enrolled in the study. Pre- and post-operative version and inclination of the glenoid were measured on CT scans, using 3D planning automatic software. During surgery, a congruent 3D-printed PSG was applied onto the glenoid surface, thus determining the entry point and orientation of the central guide wire used for reaming the glenoid and the introduction of the component. Manual segmentation was performed on post-operative CT scans to compare the planned and the actual position of the entry point (mm) and orientation of the component (°).

Results

The mean error in the accuracy of the entry point was -0.1 mm (standard deviation (sd) 1.4) in the horizontal plane, and 0.8 mm (sd 1.3) in the vertical plane. The mean error in the orientation of the glenoid component was 3.4° (sd 5.1°) for version and 1.8° (sd 5.3°) for inclination.

Conclusion

Pre-operative planning with automatic software and the use of PSGs provides accurate and reproducible positioning and orientation of the glenoid component in anatomical TSA.

Cite this article: Bone Joint J 2016;98-B:1080–5.


Correspondence should be sent to Dr M. O. Gauci; e-mail:

For access options please click here