A retrospective review was performed of patients
undergoing primary cementless total knee replacement (TKR) using
porous tantalum performed by a group of surgical trainees. Clinical
and radiological follow-up involved 79 females and 26 males encompassing
115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up
was 7 years (2 to 11). Tibial and patellar components were porous
tantalum monoblock implants, and femoral components were posterior
stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological
assessments were made for implant positioning, alignment, radiolucencies,
lysis, and loosening. There was 95.7% survival of implants. There
was no radiological evidence of loosening and no osteolysis found.
No revisions were performed for aseptic loosening. Average tibial
component alignment was 1.4° of varus (4°of valgus to 9° varus),
and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean
femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral
alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar
tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient
satisfaction with improvement in pain was 91%. Cementless TKR incorporating
porous tantalum yielded good clinical and radiological outcomes
at a mean of follow-up of seven-years. Cite this article:
Aims. The processes linking long-term bisphosphonate treatment to atypical fracture remain elusive. To establish a means of exploring this link, we have examined how long-term bisphosphonate treatment with prior ovariectomy modifies femur fracture behaviour and
Aims. The prevalence of combined abnormalities of femoral torsion (FT) and tibial torsion (TT) is unknown in patients with femoroacetabular impingement (FAI) and hip dysplasia. This study aimed to determine the prevalence of combined abnormalities of FT and TT, and which subgroups are associated with combined abnormalities of FT and TT. Methods. We retrospectively evaluated symptomatic patients with FAI or hip dysplasia with CT scans performed between September 2011 and September 2016. A total of 261 hips (174 patients) had a measurement of FT and TT. Their mean age was 31 years (SD 9), and 63% were female (165 hips). Patients were compared to an asymptomatic control group (48 hips, 27 patients) who had CT scans including femur and
Leg length discrepancy (LLD) is a common pre- and postoperative issue in total hip arthroplasty (THA) patients. The conventional technique for measuring LLD has historically been on a non-weightbearing anteroposterior pelvic radiograph; however, this does not capture many potential sources of LLD. The aim of this study was to determine if long-limb EOS radiology can provide a more reproducible and holistic measurement of LLD. In all, 93 patients who underwent a THA received a standardized preoperative EOS scan, anteroposterior (AP) radiograph, and clinical LLD assessment. Overall, 13 measurements were taken along both anatomical and functional axes and measured twice by an orthopaedic fellow and surgical planning engineer to calculate intraoperator reproducibility and correlations between measurements.Aims
Methods
The aim of this study was to evaluate the suitability of the tapered cone stem in total hip arthroplasty (THA) in patients with excessive femoral anteversion and after femoral osteotomy. We included patients who underwent THA using Wagner Cone due to proximal femur anatomical abnormalities between August 2014 and January 2019 at a single institution. We investigated implant survival time using the endpoint of dislocation and revision, and compared the prevalence of prosthetic impingements between the Wagner Cone, a tapered cone stem, and the Taperloc, a tapered wedge stem, through simulation. We also collected Oxford Hip Score (OHS), visual analogue scale (VAS) satisfaction, and VAS pain by postal survey in August 2023 and explored variables associated with those scores.Aims
Methods
This study aimed to investigate the incidence of ≥ 5 mm asymmetry in lower and whole leg lengths (LLs) in patients with unilateral osteoarthritis (OA) secondary to developmental dysplasia of the hip (DDH-OA) and primary hip osteoarthritis (PHOA), and the relationship between lower and whole LL asymmetries and femoral length asymmetry. In total, 116 patients who underwent unilateral total hip arthroplasty were included in this study. Of these, 93 had DDH-OA and 23 had PHOA. Patients with DDH-OA were categorized into three groups: Crowe grade I, II/III, and IV. Anatomical femoral length, femoral length greater trochanter (GT), femoral length lesser trochanter (LT), tibial length, foot height, lower LL, and whole LL were evaluated using preoperative CT data of the whole leg in the supine position. Asymmetry was evaluated in the Crowe I, II/III, IV, and PHOA groups.Aims
Methods
Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Aims
Methods
Natural Language Processing (NLP) offers an automated method to extract data from unstructured free text fields for arthroplasty registry participation. Our objective was to investigate how accurately NLP can be used to extract structured clinical data from unstructured clinical notes when compared with manual data extraction. A group of 1,000 randomly selected clinical and hospital notes from eight different surgeons were collected for patients undergoing primary arthroplasty between 2012 and 2018. In all, 19 preoperative, 17 operative, and two postoperative variables of interest were manually extracted from these notes. A NLP algorithm was created to automatically extract these variables from a training sample of these notes, and the algorithm was tested on a random test sample of notes. Performance of the NLP algorithm was measured in Statistical Analysis System (SAS) by calculating the accuracy of the variables collected, the ability of the algorithm to collect the correct information when it was indeed in the note (sensitivity), and the ability of the algorithm to not collect a certain data element when it was not in the note (specificity).Aims
Methods
Current treatments of prosthetic joint infection (PJI) are minimally effective against The ability of PlySs2 and vancomycin to kill biofilm and colony-forming units (CFUs) on orthopaedic implants were compared using in vitro models. An in vivo murine PJI model of DAIR was used to assess the efficacy of a combination of PlySs2 and vancomycin on periprosthetic bacterial load.Aims
Methods
This retrospective study describes the long-term results of core decompression and placement of a non-vascularised bone graft in the management of avascular necrosis of the femoral head. We treated 80 hips in 65 patients, 18 by a cortical tibial autograft and 62 by a fibular allograft. The mean age of the patients was 36 years (. sd. 13.2). A total of 78 hips were available for evaluation of which pre-operatively six were Ficat-Arlet stage 0, three stage I, 31 stage IIA, 16 stage IIB, 13 stage III and nine stage IV. A total of 34 hips (44%) were revised at a mean of four years (. sd. 3.8). Survivorship analysis using a clinical end-point showed a survival rate of 59% five years after surgery. We found a significant difference (p = 0.002) in survivorship, when using a clinical and radiological end-point, between the two grafts, in favour of the tibial autograft. We considered this difference to be the result of the better quality and increased volume of
Abnormal femoral torsion (FT) is increasingly recognized as an additional cause for femoroacetabular impingement (FAI). It is unknown if in-toeing of the foot is a specific diagnostic sign for increased FT in patients with symptomatic FAI. The aims of this study were to determine: 1) the prevalence and diagnostic accuracy of in-toeing to detect increased FT; 2) if foot progression angle (FPA) and tibial torsion (TT) are different among patients with abnormal FT; and 3) if FPA correlates with FT. A retrospective, institutional review board (IRB)-approved, controlled study of 85 symptomatic patients (148 hips) with FAI or hip dysplasia was performed in the gait laboratory. All patients had a measurement of FT (pelvic CT scan), TT (CT scan), and FPA (optical motion capture system). We allocated all patients to three groups with decreased FT (< 10°, 37 hips), increased FT (> 25°, 61 hips), and normal FT (10° to 25°, 50 hips). Cluster analysis was performed.Aims
Patients and Methods
This study aimed to explore whether intraoperative nerve monitoring can identify risk factors and reduce the incidence of nerve injury in patients with high-riding developmental dysplasia. We conducted a historical controlled study of patients with unilateral Crowe IV developmental dysplasia of the hip (DDH). Between October 2016 and October 2017, intraoperative nerve monitoring of the femoral and sciatic nerves was applied in total hip arthroplasty (THA). A neuromonitoring technician was employed to monitor nerve function and inform the surgeon of ongoing changes in a timely manner. Patients who did not have intraoperative nerve monitoring between September 2015 and October 2016 were selected as the control group. All the surgeries were performed by one surgeon. Demographics and clinical data were analyzed. A total of 35 patients in the monitoring group (ten male, 25 female; mean age 37.1 years (20 to 46)) and 56 patients in the control group (13 male, 43 female; mean age 37.9 years (23 to 52)) were enrolled. The mean follow-up of all patients was 13.1 months (10 to 15).Aims
Patients and Methods
The injured anterior cruciate ligament (ACL) is thought to exhibit an impaired healing response, and attempts at surgical repair have not been successful. Connective tissue growth factor (CTGF) is reported to be associated with wound healing, probably through transforming growth factor beta 1 (TGF-β1). A rabbit ACL injury model was used to study the effect of CTGF on ligament recovery. Quantitative real-time PCR (qRT-PCR) was performed for detection of changes in RNA levels of TGF-β1, type 1 collagen (COL1), type 2 collagen (COL2), SRY-related high mobility group-box gene9 (SOX9), tissue inhibitor of metalloproteinase-1 (TIMP-1) and matrix metallopeptidase 13 (MMP-13). Expression of related proteins was detected by Western blotting.Objectives
Methods
Periprosthetic fractures are an increasingly
common complication following joint replacement. The principles
which underpin their evaluation and treatment are common across
the musculoskeletal system. The Unified Classification System proposes
a rational approach to treatment, regardless of the bone that is
broken or the joint involved. Cite this article:
This conversation represents an attempt by several
arthroplasty surgeons to critique several abstracts presented over
the last year as well as to use them as a jumping off point for trying
to figure out where they fit in into our current understanding of
multiple issues in modern hip and knee arthroplasty.
The Unified Classification System (UCS) emphasises
the key principles in the assessment and management of peri-prosthetic
fractures complicating partial or total joint replacement. We tested the inter- and intra-observer agreement for the UCS
as applied to the pelvis and femur using 20 examples of peri-prosthetic
fracture in 17 patients. Each subtype of the UCS was represented
by at least one case. Specialist orthopaedic surgeons (experts)
and orthopaedic residents (pre-experts) assessed reliability on
two separate occasions. For the pelvis, the UCS showed inter-observer agreement of 0.837
(95% confidence intervals (CI) 0.798 to 0.876) for the experts and
0.728 (95% CI 0.689 to 0.767) for the pre-experts. The intra-observer
agreement for the experts was 0.861 (95% CI 0.760 to 0.963) and
0.803 (95% 0.688 to 0.918) for the pre-experts. For the femur, the
UCS showed an inter-observer kappa value of 0.805 (95% CI 0.765
to 0.845) for the experts and a value of 0.732 (95% CI 0.690 to 0.773)
for the pre-experts. The intra-observer agreement was 0.920 (95%
CI 0.867 to 0.973) for the experts, and 0.772 (95% CI 0.652 to 0.892)
for the pre-experts. This corresponds to a substantial and ‘almost
perfect’ inter- and intra-observer agreement for the UCS for peri-prosthetic
fractures of the pelvis and femur. We hope that unifying the terminology of these injuries will
assist in their assessment, treatment and outcome. Cite this article:
Despite a lack of long-term follow-up, there
is an increasing trend towards using femoral heads of large diameter
in total hip replacement (THR), partly because of the perceived
advantage of lower rates of dislocation. However, increasing the
size of the femoral head is not the only way to reduce the rate
of dislocation; optimal alignment of the components and repair of
the posterior capsule could achieve a similar effect. In this prospective study of 512 cemented unilateral THRs (Male:Female
230:282) performed between 2004 and 2011, we aimed to determine
the rate of dislocation in patients who received a 22 mm head on
a 9/10 Morse taper through a posterior approach with capsular repair
and using the transverse acetabular ligament (TAL) as a guide for the
alignment of the acetabular component. The mean age of the patients
at operation was 67 years (35 to 89). The mean follow-up was 2.8
years (0.5 to 6.6). Pre- and post-operative assessment included
Oxford hip, Short Form-12 and modified University of California
Los Angeles and Merle D’Aubigne scores. The angles of inclination
and anteversion of the acetabular components were measured using
radiological software. There were four dislocations (0.78%), all
of which were anterior. In conclusion, THR with a 22 mm diameter head performed through
a posterior approach with capsular repair and using the TAL as a
guide for the alignment of the acetabular component was associated
with a low rate of dislocation. Cite this article:
Version of the femoral stem is an important factor
influencing the risk of dislocation after total hip replacement (THR)
as well as the position of the acetabular component. However, there
is no radiological method of measuring stem anteversion described
in the literature. We propose a radiological method to measure stem
version and have assessed its reliability and validity. In 36 patients
who underwent THR, a hip radiograph and CT scan were taken to measure
stem anteversion. The radiograph was a modified Budin view. This
is taken as a posteroanterior radiograph in the sitting position
with 90° hip flexion and 90° knee flexion and 30° hip abduction.
The angle between the stem-neck axis and the posterior intercondylar
line was measured by three independent examiners. The intra- and
interobserver reliabilities of each measurement were examined. The
radiological measurements were compared with the CT measurements
to evaluate their validity. The mean radiological measurement was
13.36° ( Cite this article:
Multiple drilling is reported to be an effective treatment for osteonecrosis of the head of femur, but its effect on intra-osseous pressure has not been described. We undertook multiple drilling and recorded the intra-osseous pressure in 75 osteonecrotic hips in 60 patients with a mean age of 42 years (19 to 67). At a mean follow-up of 37.1 months (24 to 60), 42 hips (56%) had a clinically successful outcome. The procedure was effective in reducing the mean intra-osseous pressure from 57 mmHg (SD 22) to 16 mmHg (SD 9). Hips with a successful outcome had a mean pressure of 26 mmHg (SD 19). It was less effective in preventing progression of osteonecrosis in hips with considerable involvement and in those with a high intra-osseous pressure in the intertrochanteric region (mean 45 mmHg (SD 25)). This study is not able to answer whether a return of the intra-osseous pressure to normal levels is required for satisfactory healing.
Iontophoresis is a novel technique which may be used to facilitate the movement of antibiotics into the substance of bone using an electrical potential applied externally. We have examined the rate of early infection in allografts following application of this technique in clinical practice. A total of 31 patients undergoing revision arthroplasty or surgery for limb salvage received 34 iontophoresed sequential allografts, of which 26 survived for a minimum of two years. The mean serum antibiotic levels after operation were low (gentamicin 0.37 mg/l (0.2 to 0.5); flucloxacillin 1 mg/l (0 to 1) and the levels in the drains were high (gentamicin 40 mg/l (2.5 to 131); flucloxacillin 17 mg/l (1 to 43). There were no early deep infections. Two late infections were presumed to be haemotogenous; 28 of the 34 allografts were retained. In 12 patients with pre-existing proven infection further infection has not occurred at a mean follow-up of 51 months (24 to 82).
The incidence and long-term outcome of undisplaced fatigue fractures of the femoral neck treated conservatively were examined in Finnish military conscripts between 1970 and 1990. From 106 cases identified, 66 patients with 70 fractures were followed for a mean of 18.3 years (11 to 32). The original medical records and radiographs were studied and physical and radiological follow-up data analysed for evidence of risk factors for this injury. The development of avascular necrosis and osteoarthritis was determined from the follow-up radiographs and MR scans. The impact of new military instructions on the management of hip-related pain was assessed following their introduction in 1986. The preventive regimen (1986) improved awareness and increased the detected incidence from 13.2 per 100 000 service-years (1970 to 1986) to 53.2 per 100 000 (1987 to 1990). No patient developed displacement of the fracture or avascular necrosis of the femoral head, or suffered from adverse complications. No differences were found in MRI-measured hip joint spaces at final follow-up. The mean Harris Hip Score was 97 (70 to 100) and the Visual Analogue Scale 5.85 mm (0 to 44). Non-operative treatment, including avoidance of or reduced weight-bearing, gave favourable short- and long-term outcomes. Undisplaced fatigue fractures of the femoral neck neither predispose to avascular necrosis nor the subsequent development of osteoarthritis of the hip.
Peri- and sub-prosthetic fractures, or pathological fractures below an existing well-fixed femoral component, with or without an ipsilateral knee replacement, present a difficult surgical challenge. We describe a simple solution, in which a custom-made prosthesis with a cylindrical design is cemented proximally to the stem of an existing, well-fixed femoral component. This effectively treats the fracture without sacrificing the good hip. We describe five patients with a mean age of 73 years (60 to 81) and a mean follow-up of 47 months (6 to 108). The mean overlap of the prosthesis over the femoral component was 7.5 cm (5.5 to 10). There have been no mechanical failures, no new infections and no re-operations. We suggest that in highly selected cases, in which conventional fixation is not feasible, this technique offers a durable option and avoids the morbidity of a total femoral replacement.
Revision arthroplasty after infection can often be complicated by both extensive bone loss and a relatively high rate of re-infection. Using allograft to address the bone loss in such patients is controversial because of the perceived risk of bacterial infection from the use of avascular graft material. We describe 12 two-stage revisions for infection in which segmental allografts were loaded with antibiotics using iontophoresis, a technique using an electrical potential to drive ionised antibiotics into cortical bone. Iontophoresis produced high levels of antibiotic in the allograft, which eluted into the surrounding tissues. We postulate that this offers protection from infection in the high-risk peri-operative period. None of the 12 patients who had two-stage revision with iontophoresed allografts had further infection after a mean period of 47 months (14 to 78).