Advertisement for orthosearch.org.uk
Results 1 - 50 of 59
Results per page:
Bone & Joint Open
Vol. 3, Issue 6 | Pages 475 - 484
13 Jun 2022
Jang SJ Vigdorchik JM Windsor EW Schwarzkopf R Mayman DJ Sculco PK

Aims. Navigation devices are designed to improve a surgeon’s accuracy in positioning the acetabular and femoral components in total hip arthroplasty (THA). The purpose of this study was to both evaluate the accuracy of an optical computer-assisted surgery (CAS) navigation system and determine whether preoperative spinopelvic mobility (categorized as hypermobile, normal, or stiff) increased the risk of acetabular component placement error. Methods. A total of 356 patients undergoing primary THA were prospectively enrolled from November 2016 to March 2018. Clinically relevant error using the CAS system was defined as a difference of > 5° between CAS and 3D radiological reconstruction measurements for acetabular component inclination and anteversion. Univariate and multiple logistic regression analyses were conducted to determine whether hypermobile (. Δ. sacral slope(SS). stand-sit. > 30°), or stiff (. ∆. SS. stand-sit. < 10°) spinopelvic mobility contributed to increased error rates. Results. The paired absolute difference between CAS and postoperative imaging measurements was 2.3° (standard deviation (SD) 2.6°) for inclination and 3.1° (SD 4.2°) for anteversion. Using a target zone of 40° (± 10°) (inclination) and 20° (± 10°) (anteversion), postoperative standing radiographs measured 96% of acetabular components within the target zone for both inclination and anteversion. Multiple logistic regression analysis controlling for BMI and sex revealed that hypermobile spinopelvic mobility significantly increased error rates for anteversion (odds ratio (OR) 2.48, p = 0.009) and inclination (OR 2.44, p = 0.016), whereas stiff spinopelvic mobility increased error rates for anteversion (OR 1.97, p = 0.028). There were no dislocations at a minimum three-year follow-up. Conclusion. Despite high reliability in acetabular positioning for inclination in a large patient cohort using an optical CAS system, hypermobile and stiff spinopelvic mobility significantly increased the risk of clinically relevant errors. In patients with abnormal spinopelvic mobility, CAS systems should be adjusted for use to avoid acetabular component misalignment and subsequent risk for long-term dislocation. Cite this article: Bone Jt Open 2022;3(6):475–484


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 163 - 167
1 Feb 2006
Kalteis T Handel M Bäthis H Perlick L Tingart M Grifka J

In a prospective randomised clinical study acetabular components were implanted either freehand (n = 30) or using CT-based (n = 30) or imageless navigation (n = 30). The position of the component was determined post-operatively on CT scans of the pelvis. Following conventional freehand placement of the acetabular component, only 14 of the 30 were within the safe zone as defined by Lewinnek et al (40° inclination . sd. 10°; 15° anteversion . sd. 10°). After computer-assisted navigation 25 of 30 acetabular components (CT-based) and 28 of 30 components (imageless) were positioned within this limit (overall p < 0.001). No significant differences were observed between CT-based and imageless navigation (p = 0.23); both showed a significant reduction in variation of the position of the acetabular component compared with conventional freehand arthroplasty (p < 0.001). The duration of the operation was increased by eight minutes with imageless and by 17 minutes with CT-based navigation. Imageless navigation proved as reliable as that using CT in positioning the acetabular component


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 455 - 460
1 Apr 2007
Sugano N Nishii T Miki H Yoshikawa H Sato Y Tamura S

We have developed a CT-based navigation system using infrared light-emitting diode markers and an optical camera. We used this system to perform cementless total hip replacement using a ceramic-on-ceramic bearing couple in 53 patients (60 hips) between 1998 and 2001. We reviewed 52 patients (59 hips) at a mean of six years (5 to 8) postoperatively. The mid-term results of total hip replacement using navigation were compared with those of 91 patients (111 hips) who underwent this procedure using the same implants, during the same period, without navigation. There were no significant differences in age, gender, diagnosis, height, weight, body mass index, or pre-operative clinical score between the two groups. The operation time was significantly longer where navigation was used, but there was no significant difference in blood loss or navigation-related complications. With navigation, the acetabular components were placed within the safe zone defined by Lewinnek, while without, 31 of the 111 components were placed outside this zone. There was no significant difference in the Merle d’Aubigne and Postel hip score at the final follow-up. However, hips treated without navigation had a higher rate of dislocation. Revision was performed in two cases undertaken without navigation, one for aseptic acetabular loosening and one for fracture of a ceramic liner, both of which showed evidence of neck impingement on the liner. A further five cases undertaken without navigation showed erosion of the posterior aspect of the neck of the femoral component on the lateral radiographs. These seven impingement-related mechanical problems correlated with malorientation of the acetabular component. There were no such mechanical problems in the navigated group. We conclude that CT-based navigation increased the precision of orientation of the acetabular component and control of limb length in total hip replacement, without navigation-related complications. It also reduced the rate of dislocation and mechanical problems related to impingement


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 310 - 315
1 Mar 2009
Olsen M Davis ET Waddell JP Schemitsch EH

We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip. The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° (. sd. 2.6). The stem-shaft angle after operation differed from that planned by a mean of 2.8° (. sd. 2.0) and in 86% of cases the final angle measured within ± 5° of that planned. We had no instances of notching of the neck or varus alignment of the implant in our series. A learning curve was observed in the time taken for navigation, but not for accurate placement of the implant. Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1281 - 1286
1 Oct 2009
Olsen M Davis ET Chiu M Gamble P Tumia N Boyle RA Schemitsch EH

The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle. Prudent pre-operative planning is advisable for use in conjunction with imageless navigation since misleading information may be registered intraoperatively, which could lead to inappropriate sizing and positioning of the femoral component in hip resurfacing


Bone & Joint Open
Vol. 5, Issue 4 | Pages 260 - 268
1 Apr 2024
Broekhuis D Meurs WMH Kaptein BL Karunaratne S Carey Smith RL Sommerville S Boyle R Nelissen RGHH

Aims. Custom triflange acetabular components (CTACs) play an important role in reconstructive orthopaedic surgery, particularly in revision total hip arthroplasty (rTHA) and pelvic tumour resection procedures. Accurate CTAC positioning is essential to successful surgical outcomes. While prior studies have explored CTAC positioning in rTHA, research focusing on tumour cases and implant flange positioning precision remains limited. Additionally, the impact of intraoperative navigation on positioning accuracy warrants further investigation. This study assesses CTAC positioning accuracy in tumour resection and rTHA cases, focusing on the differences between preoperative planning and postoperative implant positions. Methods. A multicentre observational cohort study in Australia between February 2017 and March 2021 included consecutive patients undergoing acetabular reconstruction with CTACs in rTHA (Paprosky 3A/3B defects) or tumour resection (including Enneking P2 peri-acetabular area). Of 103 eligible patients (104 hips), 34 patients (35 hips) were analyzed. Results. CTAC positioning was generally accurate, with minor deviations in cup inclination (mean 2.7°; SD 2.84°), anteversion (mean 3.6°; SD 5.04°), and rotation (mean 2.1°; SD 2.47°). Deviation of the hip centre of rotation (COR) showed a mean vector length of 5.9 mm (SD 7.24). Flange positions showed small deviations, with the ischial flange exhibiting the largest deviation (mean vector length of 7.0 mm; SD 8.65). Overall, 83% of the implants were accurately positioned, with 17% exceeding malpositioning thresholds. CTACs used in tumour resections exhibited higher positioning accuracy than rTHA cases, with significant differences in inclination (1.5° for tumour vs 3.4° for rTHA) and rotation (1.3° for tumour vs 2.4° for rTHA). The use of intraoperative navigation appeared to enhance positioning accuracy, but this did not reach statistical significance. Conclusion. This study demonstrates favourable CTAC positioning accuracy, with potential for improved accuracy through intraoperative navigation. Further research is needed to understand the implications of positioning accuracy on implant performance and long-term survival. Cite this article: Bone Jt Open 2024;5(4):260–268


Bone & Joint Open
Vol. 5, Issue 8 | Pages 715 - 720
23 Aug 2024
Shen TS Cheng R Chiu Y McLawhorn AS Figgie MP Westrich GH

Aims. Implant waste during total hip arthroplasty (THA) represents a significant cost to the USA healthcare system. While studies have explored methods to improve THA cost-effectiveness, the literature comparing the proportions of implant waste by intraoperative technology used during THA is limited. The aims of this study were to: 1) examine whether the use of enabling technologies during THA results in a smaller proportion of wasted implants compared to navigation-guided and conventional manual THA; 2) determine the proportion of wasted implants by implant type; and 3) examine the effects of surgeon experience on rates of implant waste by technology used. Methods. We identified 104,420 implants either implanted or wasted during 18,329 primary THAs performed on 16,724 patients between January 2018 and June 2022 at our institution. THAs were separated by technology used: robotic-assisted (n = 4,171), imageless navigation (n = 6,887), and manual (n = 7,721). The primary outcome of interest was the rate of implant waste during primary THA. Results. Robotic-assisted THA resulted in a lower proportion (1.5%) of implant waste compared to navigation-guided THA (2.0%) and manual THA (1.9%) (all p < 0.001). Both navigated and manual THA were more likely to waste acetabular shells (odds ratio (OR) 4.5 vs 3.1) and polyethylene liners (OR 2.2 vs 2.0) compared to robotic-assisted THA after adjusting for demographic and perioperative factors, such as surgeon experience (p < 0.001). While implant waste decreased with increasing experience for procedures performed manually (p < 0.001) or with navigation (p < 0.001), waste rates for robotic-assisted THA did not differ based on surgical experience. Conclusion. Robotic-assisted THAs wasted a smaller proportion of acetabular shells and polyethylene liners than navigation-guided and manual THAs. Individual implant waste rates vary depending on the type of technology used intraoperatively. Future studies on implant waste during THA should examine reasons for non-implantation in order to better understand and develop methods for cost-saving. Cite this article: Bone Jt Open 2024;5(8):715–720


Bone & Joint Research
Vol. 10, Issue 10 | Pages 629 - 638
20 Oct 2021
Hayashi S Hashimoto S Kuroda Y Nakano N Matsumoto T Ishida K Shibanuma N Kuroda R

Aims. This study aimed to evaluate the accuracy of implant placement with robotic-arm assisted total hip arthroplasty (THA) in patients with developmental dysplasia of the hip (DDH). Methods. The study analyzed a consecutive series of 69 patients who underwent robotic-arm assisted THA between September 2018 and December 2019. Of these, 30 patients had DDH and were classified according to the Crowe type. Acetabular component alignment and 3D positions were measured using pre- and postoperative CT data. The absolute differences of cup alignment and 3D position were compared between DDH and non-DDH patients. Moreover, these differences were analyzed in relation to the severity of DDH. The discrepancy of leg length and combined offset compared with contralateral hip were measured. Results. The mean values of absolute differences (postoperative CT-preoperative plan) were 1.7° (standard deviation (SD) 2.0) (inclination) and 2.5° (SD 2.1°) (anteversion) in DDH patients, and no significant differences were found between non-DDH and DDH patients. The mean absolute differences for 3D cup position were 1.1 mm (SD 1.0) (coronal plane) and 1.2 mm (SD 2.1) (axial plane) in DDH patients, and no significant differences were found between two groups. No significant difference was found either in cup alignment between postoperative CT and navigation record after cup screws or in the severity of DDH. Excellent restoration of leg length and combined offset were achieved in both groups. Conclusion. We demonstrated that robotic-assisted THA may achieve precise cup positioning in DDH patients, and may be useful in those with severe DDH. Cite this article: Bone Joint Res 2021;10(10):629–638


Bone & Joint Open
Vol. 5, Issue 8 | Pages 671 - 680
14 Aug 2024
Fontalis A Zhao B Putzeys P Mancino F Zhang S Vanspauwen T Glod F Plastow R Mazomenos E Haddad FS

Aims

Precise implant positioning, tailored to individual spinopelvic biomechanics and phenotype, is paramount for stability in total hip arthroplasty (THA). Despite a few studies on instability prediction, there is a notable gap in research utilizing artificial intelligence (AI). The objective of our pilot study was to evaluate the feasibility of developing an AI algorithm tailored to individual spinopelvic mechanics and patient phenotype for predicting impingement.

Methods

This international, multicentre prospective cohort study across two centres encompassed 157 adults undergoing primary robotic arm-assisted THA. Impingement during specific flexion and extension stances was identified using the virtual range of motion (ROM) tool of the robotic software. The primary AI model, the Light Gradient-Boosting Machine (LGBM), used tabular data to predict impingement presence, direction (flexion or extension), and type. A secondary model integrating tabular data with plain anteroposterior pelvis radiographs was evaluated to assess for any potential enhancement in prediction accuracy.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 624 - 629
1 May 2012
Audenaert E Smet B Pattyn C Khanduja V

The aim of this study was to determine the accuracy of registration and the precision of the resection volume in navigated hip arthroscopy for cam-type femoroacetabular impingement, using imageless and image-based registration. A virtual cam lesion was defined in 12 paired cadaver hips and randomly assigned to either imageless or image-based (three-dimensional (3D) fluoroscopy) navigated arthroscopic head–neck osteochondroplasty. The accuracy of patient–image registration for both protocols was evaluated and post-operative imaging was performed to evaluate the accuracy of the surgical resection. We found that the estimated accuracy of imageless registration in the arthroscopic setting was poor, with a mean error of 5.6 mm (standard deviation (. sd. ) 4.08; 95% confidence interval (CI) 4.14 to 7.19). Because of the significant mismatch between the actual position of the probe during surgery and the position of that probe as displayed on the navigation platform screen, navigated femoral osteochondroplasty was physically impossible. The estimated accuracy of image-based registration by means of 3D fluoroscopy had a mean error of 0.8 mm (. sd. 0.51; 95% CI 0.56 to 0.94). In terms of the volume of bony resection, a mean of 17% (. sd. 11; -6% to 28%) more bone was resected than with the virtual plan (p = 0.02). The resection was a mean of 1 mm deeper (. sd. 0.7; -0.3 to 1.6) larger than on the original virtual plan (p = 0.02). In conclusion, given the limited femoral surface that can be reached and digitised during arthroscopy of the hip, imageless registration is inaccurate and does not allow for reliable surgical navigation. However, image-based registration does acceptably allow for guided femoral osteochondroplasty in the arthroscopic management of femoroacetabular impingement


Bone & Joint Open
Vol. 4, Issue 6 | Pages 416 - 423
2 Jun 2023
Tung WS Donnelley C Eslam Pour A Tommasini S Wiznia D

Aims

Computer-assisted 3D preoperative planning software has the potential to improve postoperative stability in total hip arthroplasty (THA). Commonly, preoperative protocols simulate two functional positions (standing and relaxed sitting) but do not consider other common positions that may increase postoperative impingement and possible dislocation. This study investigates the feasibility of simulating commonly encountered positions, and positions with an increased risk of impingement, to lower postoperative impingement risk in a CT-based 3D model.

Methods

A robotic arm-assisted arthroplasty planning platform was used to investigate 11 patient positions. Data from 43 primary THAs were used for simulation. Sacral slope was retrieved from patient preoperative imaging, while angles of hip flexion/extension, hip external/internal rotation, and hip abduction/adduction for tested positions were derived from literature or estimated with a biomechanical model. The hip was placed in the described positions, and if impingement was detected by the software, inspection of the impingement type was performed.


Bone & Joint Open
Vol. 5, Issue 1 | Pages 28 - 36
18 Jan 2024
Selmene MA Moreau PE Zaraa M Upex P Jouffroy P Riouallon G

Aims

Post-traumatic periprosthetic acetabular fractures are rare but serious. Few studies carried out on small cohorts have reported them in the literature. The aim of this work is to describe the specific characteristics of post-traumatic periprosthetic acetabular fractures, and the outcome of their surgical treatment in terms of function and complications.

Methods

Patients with this type of fracture were identified retrospectively over a period of six years (January 2016 to December 2021). The following data were collected: demographic characteristics, date of insertion of the prosthesis, details of the intervention, date of the trauma, characteristics of the fracture, and type of treatment. Functional results were assessed with the Harris Hip Score (HHS). Data concerning complications of treatment were collected.


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 735 - 742
1 Jul 2023
Andronic O Germann C Jud L Zingg PO

Aims

This study reports mid-term outcomes after periacetabular osteotomy (PAO) exclusively in a borderline hip dysplasia (BHD) population to provide a contrast to published outcomes for arthroscopic surgery of the hip in BHD.

Methods

We identified 42 hips in 40 patients treated between January 2009 and January 2016 with BHD defined as a lateral centre-edge angle (LCEA) of ≥ 18° but < 25°. A minimum five-year follow-up was available. Patient-reported outcomes (PROMs) including Tegner score, subjective hip value (SHV), modified Harris Hip Score (mHHS), and Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) were assessed. The following morphological parameters were evaluated: LCEA, acetabular index (AI), α angle, Tönnis staging, acetabular retroversion, femoral version, femoroepiphyseal acetabular roof index (FEAR), iliocapsularis to rectus femoris ratio (IC/RF), and labral and ligamentum teres (LT) pathology.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 136 - 143
1 Feb 2024
van der Lelij TJN Marang-van de Mheen PJ Kaptein BL Koster LA Ljung P Nelissen RGHH Toksvig-Larsen S

Aims

The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface.

Methods

Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.


Bone & Joint Open
Vol. 5, Issue 2 | Pages 79 - 86
1 Feb 2024
Sato R Hamada H Uemura K Takashima K Ando W Takao M Saito M Sugano N

Aims

This study aimed to investigate the incidence of ≥ 5 mm asymmetry in lower and whole leg lengths (LLs) in patients with unilateral osteoarthritis (OA) secondary to developmental dysplasia of the hip (DDH-OA) and primary hip osteoarthritis (PHOA), and the relationship between lower and whole LL asymmetries and femoral length asymmetry.

Methods

In total, 116 patients who underwent unilateral total hip arthroplasty were included in this study. Of these, 93 had DDH-OA and 23 had PHOA. Patients with DDH-OA were categorized into three groups: Crowe grade I, II/III, and IV. Anatomical femoral length, femoral length greater trochanter (GT), femoral length lesser trochanter (LT), tibial length, foot height, lower LL, and whole LL were evaluated using preoperative CT data of the whole leg in the supine position. Asymmetry was evaluated in the Crowe I, II/III, IV, and PHOA groups.


Bone & Joint Research
Vol. 12, Issue 9 | Pages 590 - 597
20 Sep 2023
Uemura K Otake Y Takashima K Hamada H Imagama T Takao M Sakai T Sato Y Okada S Sugano N

Aims

This study aimed to develop and validate a fully automated system that quantifies proximal femoral bone mineral density (BMD) from CT images.

Methods

The study analyzed 978 pairs of hip CT and dual-energy X-ray absorptiometry (DXA) measurements of the proximal femur (DXA-BMD) collected from three institutions. From the CT images, the femur and a calibration phantom were automatically segmented using previously trained deep-learning models. The Hounsfield units of each voxel were converted into density (mg/cm3). Then, a deep-learning model trained by manual landmark selection of 315 cases was developed to select the landmarks at the proximal femur to rotate the CT volume to the neutral position. Finally, the CT volume of the femur was projected onto the coronal plane, and the areal BMD of the proximal femur (CT-aBMD) was quantified. CT-aBMD correlated to DXA-BMD, and a receiver operating characteristic (ROC) analysis quantified the accuracy in diagnosing osteoporosis.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 324 - 335
1 Apr 2024
Fontalis A Kayani B Plastow R Giebaly DE Tahmassebi J Haddad IC Chambers A Mancino F Konan S Haddad FS

Aims

Achieving accurate implant positioning and restoring native hip biomechanics are key surgeon-controlled technical objectives in total hip arthroplasty (THA). The primary objective of this study was to compare the reproducibility of the planned preoperative centre of hip rotation (COR) in patients undergoing robotic arm-assisted THA versus conventional THA.

Methods

This prospective randomized controlled trial (RCT) included 60 patients with symptomatic hip osteoarthritis undergoing conventional THA (CO THA) versus robotic arm-assisted THA (RO THA). Patients in both arms underwent pre- and postoperative CT scans, and a patient-specific plan was created using the robotic software. The COR, combined offset, acetabular orientation, and leg length discrepancy were measured on the pre- and postoperative CT scanogram at six weeks following surgery.


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1766 - 1773
1 Dec 2021
Sculco PK Windsor EN Jerabek SA Mayman DJ Elbuluk A Buckland AJ Vigdorchik JM

Aims

Spinopelvic mobility plays an important role in functional acetabular component position following total hip arthroplasty (THA). The primary aim of this study was to determine if spinopelvic hypermobility persists or resolves following THA. Our second aim was to identify patient demographic or radiological factors associated with hypermobility and resolution of hypermobility after THA.

Methods

This study investigated patients with preoperative posterior hypermobility, defined as a change in sacral slope (SS) from standing to sitting (ΔSSstand-sit) ≥ 30°. Radiological spinopelvic parameters, including SS, pelvic incidence (PI), lumbar lordosis (LL), PI-LL mismatch, anterior pelvic plane tilt (APPt), and spinopelvic tilt (SPT), were measured on preoperative imaging, and at six weeks and a minimum of one year postoperatively. The severity of bilateral hip osteoarthritis (OA) was graded using Kellgren-Lawrence criteria.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 17 - 24
1 Jul 2021
Vigdorchik JM Sharma AK Buckland AJ Elbuluk AM Eftekhary N Mayman DJ Carroll KM Jerabek SA

Aims

Patients with spinal pathology who undergo total hip arthroplasty (THA) have an increased risk of dislocation and revision. The aim of this study was to determine if the use of the Hip-Spine Classification system in these patients would result in a decreased rate of postoperative dislocation in patients with spinal pathology.

Methods

This prospective, multicentre study evaluated 3,777 consecutive patients undergoing THA by three surgeons, between January 2014 and December 2019. They were categorized using The Hip-Spine Classification system: group 1 with normal spinal alignment; group 2 with a flatback deformity, group 2A with normal spinal mobility, and group 2B with a stiff spine. Flatback deformity was defined by a pelvic incidence minus lumbar lordosis of > 10°, and spinal stiffness was defined by < 10° change in sacral slope from standing to seated. Each category determined a patient-specific component positioning. Survivorship free of dislocation was recorded and spinopelvic measurements were compared for reliability using intraclass correlation coefficient.


Bone & Joint Open
Vol. 2, Issue 7 | Pages 476 - 485
8 Jul 2021
Scheerlinck T De Winter E Sas A Kolk S Van Gompel G Vandemeulebroucke J

Aims

Hip arthroplasty does not always restore normal anatomy. This is due to inaccurate surgery or lack of stem sizes. We evaluated the aptitude of four total hip arthroplasty systems to restore an anatomical and medialized hip rotation centre.

Methods

Using 3D templating software in 49 CT scans of non-deformed femora, we virtually implanted: 1) small uncemented calcar-guided stems with two offset options (Optimys, Mathys), 2) uncemented straight stems with two offset options (Summit, DePuy Synthes), 3) cemented undersized stems (Exeter philosophy) with three offset options (CPT, ZimmerBiomet), and 4) cemented line-to-line stems (Kerboul philosophy) with proportional offsets (Centris, Mathys). We measured the distance between the templated and the anatomical and 5 mm medialized hip rotation centre.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 306 - 311
1 Mar 2014
Fujita K Kabata T Maeda T Kajino Y Iwai S Kuroda K Hasegawa K Tsuchiya H

It has recently been reported that the transverse acetabular ligament (TAL) is helpful in determining the position of the acetabular component in total hip replacement (THR). In this study we used a computer-assisted navigation system to determine whether the TAL is useful as a landmark in THR. The study was carried out in 121 consecutive patients undergoing primary THR (134 hips), including 67 dysplastic hips (50%). There were 26 men (29 hips) and 95 women (105 hips) with a mean age of 60.2 years (17 to 82) at the time of operation. After identification of the TAL, its anteversion was measured intra-operatively by aligning the inferomedial rim of the trial acetabular component with the TAL using computer-assisted navigation. The TAL was identified in 112 hips (83.6%). Intra-observer reproducibility in the measurement of anteversion of the TAL was high, but inter-observer reproducibility was moderate. . Each surgeon was able to align the trial component according to the target value of the angle of anteversion of the TAL, but it was clear that methods may differ among surgeons. Of the measurements of the angle of anteversion of the TAL, 5.4% (6 of 112 hips) were outliers from the safe zone. In summary, we found that the TAL is useful as a landmark when implanting the acetabular component within the safe zone in almost all hips, and to prevent it being implanted in retroversion in all hips, including dysplastic hips. However, as anteversion of the TAL may be excessive in a few hips, it is advisable to pay attention to individual variations, particularly in those with severe posterior pelvic tilt. Cite this article: Bone Joint J 2014;96-B:306–11


Bone & Joint Open
Vol. 2, Issue 6 | Pages 365 - 370
1 Jun 2021
Kolodychuk N Su E Alexiades MM Ren R Ojard C Waddell BS

Aims

Traditionally, acetabular component insertion during total hip arthroplasty (THA) is visually assisted in the posterior approach and fluoroscopically assisted in the anterior approach. The present study examined the accuracy of a new surgeon during anterior (NSA) and posterior (NSP) THA using robotic arm-assisted technology compared to two experienced surgeons using traditional methods.

Methods

Prospectively collected data was reviewed for 120 patients at two institutions. Data were collected on the first 30 anterior approach and the first 30 posterior approach surgeries performed by a newly graduated arthroplasty surgeon (all using robotic arm-assisted technology) and was compared to standard THA by an experienced anterior (SSA) and posterior surgeon (SSP). Acetabular component inclination, version, and leg length were calculated postoperatively and differences calculated based on postoperative film measurement.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 129 - 134
1 Jul 2021
Ayekoloye CI Abu Qa'oud M Radi M Leon SA Kuzyk P Safir O Gross AE

Aims

Improvements in functional results and long-term survival are variable following conversion of hip fusion to total hip arthroplasty (THA) and complications are high. The aim of the study was to analyze the clinical and functional results in patients who underwent conversion of hip fusion to THA using a consistent technique and uncemented implants.

Methods

A total of 39 hip fusion conversions to THA were undertaken in 38 patients by a single surgeon employing a consistent surgical technique and uncemented implants. Parameters assessed included Harris Hip Score (HHS) for function, range of motion (ROM), leg length discrepancy (LLD), satisfaction, and use of walking aid. Radiographs were reviewed for loosening, subsidence, and heterotopic ossification (HO). Postoperative complications and implant survival were assessed.


Bone & Joint Open
Vol. 2, Issue 10 | Pages 834 - 841
11 Oct 2021
O'Connor PB Thompson MT Esposito CI Poli N McGree J Donnelly T Donnelly W

Aims

Pelvic tilt (PT) can significantly change the functional orientation of the acetabular component and may differ markedly between patients undergoing total hip arthroplasty (THA). Patients with stiff spines who have little change in PT are considered at high risk for instability following THA. Femoral component position also contributes to the limits of impingement-free range of motion (ROM), but has been less studied. Little is known about the impact of combined anteversion on risk of impingement with changing pelvic position.

Methods

We used a virtual hip ROM (vROM) tool to investigate whether there is an ideal functional combined anteversion for reduced risk of hip impingement. We collected PT information from functional lateral radiographs (standing and sitting) and a supine CT scan, which was then input into the vROM tool. We developed a novel vROM scoring system, considering both seated flexion and standing extension manoeuvres, to quantify whether hips had limited ROM and then correlated the vROM score to component position.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1021 - 1026
1 Aug 2011
Kalteis T Sendtner E Beverland D Archbold PA Hube R Schuster T Renkawitz T Grifka J

Orientation of the native acetabular plane as defined by the transverse acetabular ligament (TAL) and the posterior labrum was measured intra-operatively using computer-assisted navigation in 39 hips. In order to assess the influence of alignment on impingement, the range of movement was calculated for that defined by the TAL and the posterior labrum and compared with a standard acetabular component position (abduction 45°/anteversion 15°). With respect to the registration of the plane defined by the TAL and the posterior labrum, there was moderate interobserver agreement (r = 0.64, p < 0.001) and intra-observer reproducibility (r = 0.73, p < 0.001). The mean acetabular component orientation achieved was abduction of 41° (32° to 51°) and anteversion of 18° (−1° to 36°). With respect to the Lewinnek safe zone (abduction 40° ±10°, anteversion 15° ±10°), 35 of the 39 acetabular components were within this zone. However, there was no improvement in the range of movement (p = 0.94) and no significant difference in impingement (p = 0.085). Alignment of the acetabular component with the TAL and the posterior labrum might reduce the variability of acetabular component placement in total hip replacement. However, there is only a moderate interobserver agreement and intra-observer reliability in the alignment of the acetabular component using the TAL and the posterior labrum. No reduction in impingement was found when the acetabular component was aligned with the TAL and the posterior labrum, compared with a standard acetabular component position


Bone & Joint Research
Vol. 10, Issue 10 | Pages 639 - 649
19 Oct 2021
Bergiers S Hothi H Henckel J Di Laura A Belzunce M Skinner J Hart A

Aims

Acetabular edge-loading was a cause of increased wear rates in metal-on-metal hip arthroplasties, ultimately contributing to their failure. Although such wear patterns have been regularly reported in retrieval analyses, this study aimed to determine their in vivo location and investigate their relationship with acetabular component positioning.

Methods

3D CT imaging was combined with a recently validated method of mapping bearing surface wear in retrieved hip implants. The asymmetrical stabilizing fins of Birmingham hip replacements (BHRs) allowed the co-registration of their acetabular wear maps and their computational models, segmented from CT scans. The in vivo location of edge-wear was measured within a standardized coordinate system, defined using the anterior pelvic plane.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 168 - 172
1 Feb 2006
Mayr E de la Barrera JM Eller G Bach C Nogler M

In navigated total hip arthroplasty, the pelvis and the femur are tracked by means of rigid bodies fixed directly to the bones. Exact tracking throughout the procedure requires that the connection between the marker and bone remains stable in terms of translation and rotation. We carried out a cadaver study to compare the intra-operative stability of markers consisting of an anchoring screw with a rotational stabiliser and of pairs of pins and wires of different diameters connected with clamps. These devices were tested at different locations in the femur. Three human cadavers were placed supine on an operating table, with a reference marker positioned in the area of the greater trochanter. K-wires (3.2 mm), Steinman pins (3 and 4 mm), Apex pins (3 and 4 mm), and a standard screw were used as fixation devices. They were positioned medially in the proximal third of the femur, ventrally in the middle third and laterally in the distal portion. In six different positions of the leg, the spatial positions were recorded with a navigation system. Compared with the standard single screw, with the exception of the 3 mm Apex pins, the two-pin systems were associated with less movement of the marker and could be inserted less invasively. With the knee flexed to 90° and the dislocated hip rotated externally until the lower leg was parallel to the table (figure-four position), all the anchoring devices showed substantial deflection of 1.5° to 2.5°. The most secure area for anchoring markers was the lateral aspect of the femur


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 665 - 671
1 Apr 2021
Osawa Y Seki T Okura T Takegami Y Ishiguro N Hasegawa Y

Aims

We compared the clinical outcomes of curved intertrochanteric varus osteotomy (CVO) with bone impaction grafting (BIG) with CVO alone for the treatment of osteonecrosis of the femoral head (ONFH).

Methods

This retrospective comparative study included 81 patients with ONFH; 37 patients (40 hips) underwent CVO with BIG (BIG group) and 44 patients (47 hips) underwent CVO alone (CVO group). Patients in the BIG group were followed-up for a mean of 12.2 years (10.0 to 16.5). Patients in the CVO group were followed-up for a mean of 14.5 years (10.0 to 21.0). Assessment parameters included the Harris Hip Score (HHS), Oxford Hip Score (OHS), Japanese Orthopaedic Association Hip-Disease Evaluation Questionnaire (JHEQ), complication rates, and survival rates, with conversion to total hip arthroplasty (THA) and radiological failure as the endpoints.


Bone & Joint Research
Vol. 9, Issue 7 | Pages 360 - 367
1 Jul 2020
Kawahara S Hara T Sato T Kitade K Shimoto T Nakamura T Mawatari T Higaki H Nakashima Y

Aims

Appropriate acetabular component placement has been proposed for prevention of postoperative dislocation in total hip arthroplasty (THA). Manual placements often cause outliers in spite of attempts to insert the component within the intended safe zone; therefore, some surgeons routinely evaluate intraoperative pelvic radiographs to exclude excessive acetabular component malposition. However, their evaluation is often ambiguous in case of the tilted or rotated pelvic position. The purpose of this study was to develop the computational analysis to digitalize the acetabular component orientation regardless of the pelvic tilt or rotation.

Methods

Intraoperative pelvic radiographs of 50 patients who underwent THA were collected retrospectively. The 3D pelvic bone model and the acetabular component were image-matched to the intraoperative pelvic radiograph. The radiological anteversion (RA) and radiological inclination (RI) of the acetabular component were calculated and those measurement errors from the postoperative CT data were compared relative to those of the 2D measurements. In addition, the intra- and interobserver differences of the image-matching analysis were evaluated.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1311 - 1318
3 Oct 2020
Huang Y Gao Y Li Y Ding L Liu J Qi X

Aims

Morphological abnormalities are present in patients with developmental dysplasia of the hip (DDH). We studied and compared the pelvic anatomy and morphology between the affected hemipelvis with the unaffected side in patients with unilateral Crowe type IV DDH using 3D imaging and analysis.

Methods

A total of 20 patients with unilateral Crowe-IV DDH were included in the study. The contralateral side was considered normal in all patients. A coordinate system based on the sacral base (SB) in a reconstructed pelvic model was established. The pelvic orientations (tilt, rotation, and obliquity) of the affected side were assessed by establishing a virtual anterior pelvic plane (APP). The bilateral coordinates of the anterior superior iliac spine (ASIS) and the centres of hip rotation were established, and parameters concerning size and volume were compared for both sides of the pelvis.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 47 - 51
1 Jul 2020
Kazarian GS Schloemann DT Barrack TN Lawrie CM Barrack RL

Aims

The aims of this study were to determine the change in the sagittal alignment of the pelvis and the associated impact on acetabular component position at one-year follow-up after total hip arthroplasty (THA).

Methods

This study represents the one-year follow-up of a previous short-term study at our institution. Using the patient population from our prior study, the radiological pelvic ratio was assessed in 91 patients undergoing THA, of whom 50 were available for follow-up of at least one year (median 1.5; interquartile range (IQR) 1.1 to 2.0). Anteroposterior radiographs of the pelvis were obtained in the standing position preoperatively and at one year postoperatively. Pelvic ratio was defined as the ratio between the vertical distance from the inferior sacroiliac (SI) joints to the superior pubic symphysis and the horizontal distance between the inferior SI joints. Apparent acetabular component position changes were determined from the change in pelvic ratio. A change of at least 5° was considered clinically meaningful.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 97 - 103
1 Jun 2019
Novikov D Mercuri JJ Schwarzkopf R Long WJ Bosco III JA Vigdorchik JM

Aims

Studying the indications for revision total hip arthroplasty (THA) may enable surgeons to change their practice during the initial procedure, thereby reducing the need for revision surgery. The aim of this study was to identify and describe the potentially avoidable indications for revision THA within five years of the initial procedure.

Patients and Methods

A retrospective review of 117 patients (73 women, 44 men; mean age 61.5 years (27 to 88)) who met the inclusion criteria was conducted. Three adult reconstruction surgeons independently reviewed the radiographs and medical records, and they classified the revision THAs into two categories: potentially avoidable and unavoidable. Baseline demographics, perioperative details, and quality outcomes up to the last follow-up were recorded.


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 189 - 197
1 Feb 2019
Yoshitani J Kabata T Kajino Y Ueno T Ueoka K Nakamura T Tsuchiya H

Aims

We analyzed the acetabular morphology of Crowe type IV hips using CT data to identify a landmark for the ideal placement of the centre of the acetabular component, as assessed by morphometric geometrical analysis, and its reliability.

Patients and Methods

A total of 52 Crowe IV hips (42 patients; seven male, 35 female; mean age 68.5 years (32 to 82)) and 50 normal hips (50 patients; eight male, 42 female; mean age 60.7 years (34 to 86)) undergoing total hip arthroplasty were retrospectively identified. In this CT-based simulation study, the acetabular component was positioned at the true acetabulum with a radiological inclination of 40° and anteversion of 20°. Acetabular shape and the position of the centre of the acetabular component were analyzed by morphometric geometrical analysis using the generalized Procrustes analysis.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1042 - 1049
1 Sep 2019
Murphy MP Killen CJ Ralles SJ Brown NM Hopkinson WJ Wu K

Aims

Several radiological methods of measuring anteversion of the acetabular component after total hip arthroplasty (THA) have been described. These are limited by low reproducibility, are less accurate than CT 3D reconstruction, and are cumbersome to use. These methods also partly rely on the identification of obscured radiological borders of the component. We propose two novel methods, the Area and Orthogonal methods, which have been designed to maximize use of readily identifiable points while maintaining the same trigonometric principles.

Patients and Methods

A retrospective study of plain radiographs was conducted on 160 hips of 141 patients who had undergone primary THA. We compared the reliability and accuracy of the Area and Orthogonal methods with two of the current leading methods: those of Widmer and Lewinnek, respectively.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 45 - 50
1 Jun 2019
Schloemann DT Edelstein AI Barrack RL

Aims

The aims of this study were to determine the change in pelvic sagittal alignment before, during, and after total hip arthroplasty (THA) undertaken with the patient in the lateral decubitus position, and to determine the impact of these changes on acetabular component position.

Patients and Methods

We retrospectively compared the radiological pelvic ratio among 91 patients undergoing THA. In total, 41 patients (46%) were female. The mean age was 61.6 years (sd 10.7) and the mean body mass index (BMI) was 20.0 kg/m2 (sd 5.5). Anteroposterior radiographs were obtained: in the standing position preoperatively and at six weeks postoperatively; in the lateral decubitus position after trial reduction intraoperatively; and in the supine position in the post-anaesthesia care unit. Pelvic ratio was defined as the ratio between the vertical distance from the inferior aspect of the sacroiliac (SI) joints to the superior pubic symphysis and the horizontal distance between the inferior aspect of the SI joints. Changes in the apparent component position based on changes in pelvic ratio were determined, with a change of > 5° considered clinically significant. Analyses were performed using Wilcoxon’s signed-rank test, with p < 0.05 considered significant.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1303 - 1309
1 Oct 2018
Nodzo SR Chang C Carroll KM Barlow BT Banks SA Padgett DE Mayman DJ Jerabek SA

Aims

The aim of this study was to evaluate the accuracy of implant placement when using robotic assistance during total hip arthroplasty (THA).

Patients and Methods

A total of 20 patients underwent a planned THA using preoperative CT scans and robotic-assisted software. There were nine men and 11 women (n = 20 hips) with a mean age of 60.8 years (sd 6.0). Pelvic and femoral bone models were constructed by segmenting both preoperative and postoperative CT scan images. The preoperative anatomical landmarks using the robotic-assisted system were matched to the postoperative 3D reconstructions of the pelvis. Acetabular and femoral component positions as measured intraoperatively and postoperatively were evaluated and compared.


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 297 - 302
1 Mar 2019
Tamura K Takao M Hamada H Ando W Sakai T Sugano N

Aims

The aim of this study was to examine whether hips with unilateral osteoarthritis (OA) secondary to developmental dysplasia of the hip (DDH) have significant asymmetry in femoral length, and to determine potential related factors.

Patients and Methods

We enrolled 90 patients (82 female, eight male) with DDH showing unilateral OA changes, and 43 healthy volunteers (26 female, 17 male) as controls. The mean age was 61.8 years (39 to 93) for the DDH groups, and 71.2 years (57 to 84) for the control group. Using a CT-based coordinate measurement system, we evaluated the following vertical distances: top of the greater trochanter to the knee centre (femoral length GT), most medial prominence of the lesser trochanter to the knee centre (femoral length LT), and top of the greater trochanter to the medial prominence of the lesser trochanter (intertrochanteric distance), along with assessments of femoral neck anteversion and neck shaft angle.


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 36 - 43
1 Jan 2018
Hambright D Hellman M Barrack R

Aims

The aims of this study were to examine the rate at which the positioning of the acetabular component, leg length discrepancy and femoral offset are outside an acceptable range in total hip arthroplasties (THAs) which either do or do not involve the use of intra-operative digital imaging.

Patients and Methods

A retrospective case-control study was undertaken with 50 patients before and 50 patients after the integration of an intra-operative digital imaging system in THA. The demographics of the two groups were comparable for body mass index, age, laterality and the indication for surgery. The digital imaging group had more men than the group without. Surgical data and radiographic parameters, including the inclination and anteversion of the acetabular component, leg length discrepancy, and the difference in femoral offset compared with the contralateral hip were collected and compared, as well as the incidence of altering the position of a component based on the intra-operative image.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 862 - 866
1 Jul 2018
Darrith B Bell JA Culvern C Della Valle CJ

Aims

Accurate placement of the acetabular component is essential in total hip arthroplasty (THA). The purpose of this study was to determine if the ability to achieve inclination of the acetabular component within the ‘safe-zone’ of 30° to 50° could be improved with the use of an inclinometer.

Patients and Methods

We reviewed 167 primary THAs performed by a single surgeon over a period of 14 months. Procedures were performed at two institutions: an inpatient hospital, where an inclinometer was used (inclinometer group); and an ambulatory centre, where an inclinometer was not used as it could not be adequately sterilized (control group). We excluded 47 patients with a body mass index (BMI) of > 40 kg/m2, age of > 68 years, or a surgical indication other than osteoarthritis whose treatment could not be undertaken in the ambulatory centre. There were thus 120 patients in the study, 68 in the inclinometer group and 52 in the control group. The inclination angles of the acetabular component were measured from de-identified plain radiographs by two blinded investigators who were not involved in the surgery. The effect of the use of the inclinometer on the inclination angle was determined using multivariate regression analysis.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 37 - 45
1 Jan 2017
Stefl M Lundergan W Heckmann N McKnight B Ike H Murgai R Dorr LD

Aims

Posterior tilt of the pelvis with sitting provides biological acetabular opening. Our goal was to study the post-operative interaction of skeletal mobility and sagittal acetabular component position.

Materials and Methods

This was a radiographic study of 160 hips (151 patients) who prospectively had lateral spinopelvic hip radiographs for skeletal and implant measurements. Intra-operative acetabular component position was determined according to the pre-operative spinal mobility. Sagittal implant measurements of ante-inclination and sacral acetabular angle were used as surrogate measurements for the risk of impingement, and intra-operative acetabular component angles were compared with these.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1342 - 1346
1 Oct 2016
Spencer-Gardner L Pierrepont J Topham M Baré J McMahon S Shimmin AJ

Aims

Accurate placement of the acetabular component during total hip arthroplasty (THA) is an important factor in the success of the procedure. However, the reported accuracy varies greatly and is dependent upon whether free hand or navigated techniques are used. The aim of this study was to assess the accuracy of an instrument system that incorporates 3D printed, patient-specific guides designed to optimise the placement of the acetabular component.

Patients and Methods

A total of 100 consecutive patients were prospectively enrolled and the accuracy of placement of the acetabular component was measured using post-operative CT scans.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 890 - 898
1 Jul 2015
Renkawitz T Weber M Springorum H Sendtner E Woerner M Ulm K Weber T Grifka J

We report the kinematic and early clinical results of a patient- and observer-blinded randomised controlled trial in which CT scans were used to compare potential impingement-free range of movement (ROM) and acetabular component cover between patients treated with either the navigated ‘femur-first’ total hip arthroplasty (THA) method (n = 66; male/female 29/37, mean age 62.5 years; 50 to 74) or conventional THA (n = 69; male/female 35/34, mean age 62.9 years; 50 to 75). The Hip Osteoarthritis Outcome Score, the Harris hip score, the Euro-Qol-5D and the Mancuso THA patient expectations score were assessed at six weeks, six months and one year after surgery. A total of 48 of the patients (84%) in the navigated ‘femur-first’ group and 43 (65%) in the conventional group reached all the desirable potential ROM boundaries without prosthetic impingement for activities of daily living (ADL) in flexion, extension, abduction, adduction and rotation (p = 0.016). Acetabular component cover and surface contact with the host bone were > 87% in both groups. There was a significant difference between the navigated and the conventional groups’ Harris hip scores six weeks after surgery (p = 0.010). There were no significant differences with respect to any clinical outcome at six months and one year of follow-up. The navigated ‘femur-first’ technique improves the potential ROM for ADL without prosthetic impingement, although there was no observed clinical difference between the two treatment groups.

Cite this article: Bone Joint J 2015; 97-B:890–8.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1290 - 1297
1 Oct 2017
Devane PA Horne JG Foley G Stanley J

Aims

This paper describes the methodology, validation and reliability of a new computer-assisted method which uses models of the patient’s bones and the components to measure their migration and polyethylene wear from radiographs after total hip arthroplasty (THA).

Materials and Methods

Models of the patient’s acetabular and femoral component obtained from the manufacturer and models of the patient’s pelvis and femur built from a single computed tomography (CT) scan, are used by a computer program to measure the migration of the components and the penetration of the femoral head from anteroposterior and lateral radiographs taken at follow-up visits. The program simulates the radiographic setup and matches the position and orientation of the models to outlines of the pelvis, the acetabular and femoral component, and femur on radiographs. Changes in position and orientation reflect the migration of the components and the penetration of the femoral head. Validation was performed using radiographs of phantoms simulating known migration and penetration, and the clinical feasibility of measuring migration was assessed in two patients.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 44 - 49
1 Jan 2016
Sheth NP Melnic CM Paprosky WG

Given the increasing number of total hip arthroplasty procedures being performed annually, it is imperative that orthopaedic surgeons understand factors responsible for instability. In order to treat this potentially complex problem, we recommend correctly classifying the type of instability present based on component position, abductor function, impingement, and polyethylene wear. Correct classification allows the treating surgeon to choose the appropriate revision option that ultimately will allow for the best potential outcome.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):44–9.


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1615 - 1622
1 Dec 2015
Müller M Abdel MP Wassilew GI Duda G Perka C

The accurate reconstruction of hip anatomy and biomechanics is thought to be important in achieveing good clinical outcomes following total hip arthroplasty (THA). To this end some newer hip designs have introduced further modularity into the design of the femoral component such that neck­shaft angle and anteversion, which can be adjusted intra-operatively. The clinical effect of this increased modularity is unknown. We have investigated the changes in these anatomical parameters following conventional THA with a prosthesis of predetermined neck–shaft angle and assessed the effect of changes in the hip anatomy on clinical outcomes.

In total, 44 patients (mean age 65.3 years (standard deviation (sd) 7); 17 male/27 female; mean body mass index 26.9 (kg/m²) (sd 3.1)) underwent a pre- and post-operative three-dimensional CT scanning of the hip. The pre- and post-operative neck–shaft angle, offset, hip centre of rotation, femoral anteversion, and stem alignment were measured. Additionally, a functional assessment and pain score were evaluated before surgery and at one year post-operatively and related to the post-operative anatomical changes.

The mean pre-operative neck–shaft angle was significantly increased by 2.8° from 128° (sd 6.2; 119° to 147°) to 131° (sd 2.1; 127° to 136°) (p = 0.009). The mean pre-operative anteversion was 24.9° (sd 8; 7.9 to 39.1) and reduced to 7.4° (sd 7.3; -11.6° to 25.9°) post-operatively (p < 0.001). The post-operative changes had no influence on function and pain. Using a standard uncemented femoral component, high pre- and post-operative variability of femoral anteversion and neck–shaft angles was found with a significant decrease of the post-operative anteversion and slight increase of the neck–shaft angles, but without any impact on clinical outcome.

Cite this article: Bone Joint J 2015;97-B:1615–22.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 312 - 318
1 Mar 2014
Meermans G Van Doorn WJ Koenraadt K Kats J

The orientation of the acetabular component can influence both the short- and long-term outcomes of total hip replacement (THR). We performed a prospective, randomised, controlled trial of two groups, comprising of 40 patients each, in order to compare freehand introduction of the component with introduction using the transverse acetabular ligament (TAL) as a reference for anteversion. Anteversion and inclination were measured on pelvic radiographs.

With respect to anteversion, in the freehand group 22.5% of the components were outside the safe zone versus 0% in the transverse acetabular ligament group (p = 0.002). The mean angle of anteversion in the freehand group was 21° (2° to 35°) which was significantly higher compared with 17° (2° to 25°) in the TAL group (p = 0.004). There was a significant difference comparing the variations of both groups (p = 0.008).

With respect to inclination, in the freehand group 37.5% of the components were outside the safe zone versus 20% in the TAL group (p = 0.14). There was no significant difference regarding the accuracy or variation of the angle of inclination when comparing the two groups.

The transverse acetabular ligament may be used to obtain the appropriate anteversion when introducing the acetabular component during THR, but not acetabular component inclination.

Cite this article: Bone Joint J 2014;96-B:312–18.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1445 - 1449
1 Nov 2013
Sonnadara R McQueen S Mironova P Safir O Nousiainen M Ferguson P Alman B Kraemer W Reznick R

Valid and reliable techniques for assessing performance are essential to surgical education, especially with the emergence of competency-based frameworks. Despite this, there is a paucity of adequate tools for the evaluation of skills required during joint replacement surgery. In this scoping review, we examine current methods for assessing surgeons’ competency in joint replacement procedures in both simulated and clinical environments. The ability of many of the tools currently in use to make valid, reliable and comprehensive assessments of performance is unclear. Furthermore, many simulation-based assessments have been criticised for a lack of transferability to the clinical setting. It is imperative that more effective methods of assessment are developed and implemented in order to improve our ability to evaluate the performance of skills relating to total joint replacement. This will enable educators to provide formative feedback to learners throughout the training process to ensure that they have attained core competencies upon completion of their training. This should help ensure positive patient outcomes as the surgical trainees enter independent practice.

Cite this article: Bone Joint J 2013;95-B:1445–9.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1326 - 1331
1 Oct 2013
Eilander W Harris SJ Henkus HE Cobb JP Hogervorst T

Orientation of the acetabular component influences wear, range of movement and the incidence of dislocation after total hip replacement (THR). During surgery, such orientation is often referenced to the anterior pelvic plane (APP), but APP inclination relative to the coronal plane (pelvic tilt) varies substantially between individuals. In contrast, the change in pelvic tilt from supine to standing (dPT) is small for nearly all individuals. Therefore, in THR performed with the patient supine and the patient’s coronal plane parallel to the operating table, we propose that freehand placement of the acetabular component placement is reliable and reflects standing (functional) cup position. We examined this hypothesis in 56 hips in 56 patients (19 men) with a mean age of 61 years (29 to 80) using three-dimensional CT pelvic reconstructions and standing lateral pelvic radiographs. We found a low variability of acetabular component placement, with 46 implants (82%) placed within a combined range of 30° to 50° inclination and 5° to 25° anteversion. Changing from the supine to the standing position (analysed in 47 patients) was associated with an anteversion change < 10° in 45 patients (96%). dPT was < 10° in 41 patients (87%). In conclusion, supine THR appears to provide reliable freehand acetabular component placement. In most patients a small reclination of the pelvis going from supine to standing causes a small increase in anteversion of the acetabular component.

Cite this article: Bone Joint J 2013;95-B:1326–31.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1209 - 1214
1 Sep 2010
Hill JC Gibson DP Pagoti R Beverland DE

The angle of inclination of the acetabular component in total hip replacement is a recognised contributing factor in dislocation and early wear. During non-navigated surgery, insertion of the acetabular component has traditionally been performed at an angle of 45° relative to the sagittal plane as judged by the surgeon’s eye, the operative inclination. Typically, the method used to assess inclination is the measurement made on the postoperative anteroposterior radiograph, the radiological inclination.

The aim of this study was to measure the intra-operative angle of inclination of the acetabular component on 60 consecutive patients in the lateral decubitus position when using a posterior approach during total hip replacement. This was achieved by taking intra-operative photographs of the acetabular inserter, representing the acetabular axis, and a horizontal reference. The results were compared with the post-operative radiological inclination.

The mean post-operative radiological inclination was 13° greater than the photographed operative inclination, which was unexpectedly high. It appears that in the lateral decubitus position with a posterior approach, the uppermost hemipelvis adducts, thus reducing the apparent operative inclination. Surgeons using the posterior approach in lateral decubitus need to aim for a lower operative inclination than when operating with the patient supine in order to achieve an acceptable radiological inclination.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1458 - 1463
1 Nov 2013
Won S Lee Y Ha Y Suh Y Koo K

Pre-operative planning for total hip replacement (THR) is challenging in hips with severe acetabular deformities, including those with a hypoplastic acetabulum or severe defects and in the presence of arthrodesis or ankylosis. We evaluated whether a Rapid Prototype (RP) model, which is a life-sized reproduction based on three-dimensional CT scans, can determine the feasibility of THR and provide information about the size and position of the acetabular component in severe acetabular deformities. THR was planned using an RP model in 21 complex hips in five men (five hips) and 16 women (16 hips) with a mean age of 47.7 years (24 to 70) at operation. An acetabular component was implanted successfully and THR completed in all hips. The acetabular component used was within 2 mm of the predicted size in 17 hips (80.9%). All of the acetabular components and femoral stems had radiological evidence of bone ingrowth and stability at the final follow-up, without any detectable wear or peri-prosthetic osteolysis. The RP model allowed a simulated procedure pre-operatively and was helpful in determining the feasibility of THR pre-operatively, and to decide on implant type, size and position in complex THRs.

Cite this article: Bone Joint J 2013;95-B:1458–63.