Accurate placement of the acetabular component during total hip
arthroplasty (THA) is an important factor in the success of the
procedure. However, the reported accuracy varies greatly and is
dependent upon whether free hand or navigated techniques are used.
The aim of this study was to assess the accuracy of an instrument
system that incorporates 3D printed, patient-specific guides designed
to optimise the placement of the acetabular component. A total of 100 consecutive patients were prospectively enrolled
and the accuracy of placement of the acetabular component was measured
using post-operative CT scans.Aims
Patients and Methods
Aims. The direct anterior approach (DAA) for total hip arthroplasty (THA) has potential advantages over other approaches and is most commonly performed with the patient in the supine position. We describe a technique for DAA THA with the patient in the lateral decubitus position and report the early clinical and radiological outcomes, the characteristics of the learning curve, and perioperative complications. Methods. All primary DAA THAs performed in the lateral position by a single surgeon over a four-year period from the surgeon’s first case using the technique were identified from a prospectively collected database. Modified Harris Hip Scores (mHHS) were collected to assess clinical outcome, and routine radiological analysis was performed. Retrospective review of the medical records identified perioperative complications, the characteristics of the learning curve, and revisions. Results. A total of 257 patients were included in the study. Their mean age was 60 years (SD 9.0). A total of 164 (64%) were female. The mean mHHS improved significantly from 52.1 (SD 16.2) preoperatively to 94.4 (SD 11) at a follow-up of one year (p < 0.001), with 212 of 225 patients (94%) achieving a minimal clinically important difference (MCID) (> 8 points). Radiological evaluation showed a mean leg length discrepancy of 2.6 mm (SD 5.9) and a mean difference in femoral offset of 0.2 mm (SD 4.9). A total of 234/243 acetabular components (96.3%) were positioned within Lewinnek’s safe zone. Analysis of operating time, blood loss, the position of the components, and complications did not identify a learning curve. A total of 14 patients (5.4%) had a major perioperative complication and three (1.2%) required revision THA. There were no major neurovascular complications and no dislocations. Conclusion. We have described and analyzed a surgical technique for undertaking DAA THA in the familiar lateral decubitus position using a routine operating table, positioning devices, and
Aims. Prior studies have identified that malseating of a modular dual mobility liner can occur, with previous reported incidences between 5.8% and 16.4%. The aim of this study was to determine the incidence of malseating in dual mobility implants at our institution, assess for risk factors for liner malseating, and investigate whether liner malseating has any impact on clinical outcomes after surgery. Methods. We retrospectively reviewed the radiographs of 239 primary and revision total hip arthroplasties with a modular dual mobility liner. Two independent reviewers assessed radiographs for each patient twice for evidence of malseating, with a third observer acting as a tiebreaker. Univariate analysis was conducted to determine risk factors for malseating with Youden’s index used to identify cut-off points. Cohen’s kappa test was used to measure interobserver and intraobserver reliability. Results. In all, 12 liners (5.0%), including eight Stryker (6.8%) and four Zimmer Biomet (3.3%), had radiological evidence of malseating. Interobserver reliability was found to be 0.453 (95% confidence interval (CI) 0.26 to 0.64), suggesting weak inter-rater agreement, with strong agreement being greater than 0.8. We found component size of 50 mm or less to be associated with liner malseating on univariate analysis (p = 0.031). Patients with malseated liners appeared to have no associated clinical consequences, and none required revision surgery at a mean of 14 months (1.4 to 99.2) postoperatively. Conclusion. The incidence of liner malseating was 5.0%, which is similar to other reports. Component size of 50 mm or smaller was identified as a risk factor for malseating. Surgeons should be aware that malseating can occur and implant design changes or changes in
Achieving accurate implant positioning and restoring native hip biomechanics are key surgeon-controlled technical objectives in total hip arthroplasty (THA). The primary objective of this study was to compare the reproducibility of the planned preoperative centre of hip rotation (COR) in patients undergoing robotic arm-assisted THA versus conventional THA. This prospective randomized controlled trial (RCT) included 60 patients with symptomatic hip osteoarthritis undergoing conventional THA (CO THA) versus robotic arm-assisted THA (RO THA). Patients in both arms underwent pre- and postoperative CT scans, and a patient-specific plan was created using the robotic software. The COR, combined offset, acetabular orientation, and leg length discrepancy were measured on the pre- and postoperative CT scanogram at six weeks following surgery.Aims
Methods
The aim of this study is to report the long-term outcomes of instrumented femoral revisions with impaction allograft bone grafting (IBG) using the X-change femoral revision system at 30 years after introduction of the technique. We updated the outcomes of our previous study, based on 208 consecutive revisions using IBG and the X-change femoral revision system in combination with a cemented polished stem, performed in our tertiary care institute between 1991 and 2007. Kaplan-Meier survival analyses were used to determine the survival rate of the revisions with endpoint revision for any reason and aseptic loosening. Secondary outcomes were radiological loosening and patient-reported outcome measures.Aims
Methods
When performing revision total hip arthroplasty using diaphyseal-engaging titanium tapered stems (TTS), the recommended 3 to 4 cm of stem-cortical diaphyseal contact may not be available. In challenging cases such as these with only 2 cm of contact, can sufficient axial stability be achieved and what is the benefit of a prophylactic cable? This study sought to determine, first, whether a prophylactic cable allows for sufficient axial stability when the contact length is 2 cm, and second, if differing TTS taper angles (2° vs 3.5°) impact these results. A biomechanical matched-pair cadaveric study was designed using six matched pairs of human fresh cadaveric femora prepared so that 2 cm of diaphyseal bone engaged with 2° (right femora) or 3.5° (left femora) TTS. Before impaction, three matched pairs received a single 100 lb-tensioned prophylactic beaded cable; the remaining three matched pairs received no cable adjuncts. Specimens underwent stepwise axial loading to 2600 N or until failure, defined as stem subsidence > 5 mm.Aims
Methods
This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.Aims
Methods
This study aimed to assess the carbon footprint associated with total hip arthroplasty (THA) in a UK hospital setting, considering various components within the operating theatre. The primary objective was to identify actionable areas for reducing carbon emissions and promoting sustainable orthopaedic practices. Using a life-cycle assessment approach, we conducted a prospective study on ten cemented and ten hybrid THA cases, evaluating carbon emissions from anaesthetic room to recovery. Scope 1 and scope 2 emissions were considered, focusing on direct emissions and energy consumption. Data included detailed assessments of consumables, waste generation, and energy use during surgeries.Aims
Methods
The aim of this study was to identify the optimal lip position for total hip arthroplasties (THAs) using a lipped liner. There is a lack of consensus on the optimal position, with substantial variability in surgeon practice. A model of a THA was developed using a 20° lipped liner. Kinematic analyses included a physiological range of motion (ROM) analysis and a provocative dislocation manoeuvre analysis. ROM prior to impingement was calculated and, in impingement scenarios, the travel distance prior to dislocation was assessed. The combinations analyzed included nine cup positions (inclination 30-40-50°, anteversion 5-15-25°), three stem positions (anteversion 0-15-30°), and five lip orientations (right hip 7 to 11 o’clock).Aims
Methods
Aims. Femoral impaction bone grafting was first developed in 1987 using
morselised cancellous bone graft impacted into the femoral canal
in combination with a cemented, tapered, polished stem. We describe
the evolution of this technique and
There is evidence that morbidly obese patients have more intra- and postoperative complications and poorer outcomes when undergoing total hip arthroplasty (THA) with the direct anterior approach (DAA). The aim of this study was to determine the efficacy of DAA for THA, and compare the complications and outcomes of morbidly obese patients with nonobese patients. Morbidly obese patients (n = 86), with BMI ≥ 40 kg/m2 who underwent DAA THA at our institution between September 2010 and December 2017, were matched to 172 patients with BMI < 30 kg/m2. Data regarding demographics, set-up and operating time, blood loss, radiological assessment, Harris Hip Score (HHS), International Hip Outcome Tool (12-items), reoperation rate, and complications at two years postoperatively were retrospectively analyzed.Aims
Methods
Use of the direct anterior approach (DAA) for total hip arthroplasty (THA) has increased in recent years due to proposed benefits, including a lower risk of dislocation and improved early functional recovery. This study investigates the dislocation rate in a non-selective, consecutive cohort undergoing THA via the DAA without any exclusion or bias in patient selection based on habitus, deformity, age, sex, or fixation method. We retrospectively reviewed all patients undergoing THA via the DAA between 2011 and 2017 at our institution. Primary outcome was dislocation at minimum two-year follow-up. Patients were stratified by demographic details and risk factors for dislocation, and an in-depth analysis of dislocations was performed.Aims
Methods
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods
The purpose of this study was to compare the clinical, radiological, and patient-reported outcome measures (PROMs) in the first 100 consecutive patients undergoing total hip arthroplasty (THA) via a direct superior approach (DSA) with a matched group of patients undergoing THA by the same surgeon, using a posterolateral approach (PLA). This was a retrospective single surgeon study comparing the first 100 consecutive DSA THA patients with a matched group of patients using a standard PLA. Case notes were examined for patient demographics, length of hospital stay, operating time, intra- and postoperative complications, pain score, satisfaction score, and Oxford Hip Score (OHS). Leg length discrepancy and component positioning were measured from postoperative plain radiographs.Aims
Methods
To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the cement technique?; and does such a cementing technique reduce the incidence of cementless complications? A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016 and 2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 years (70 to 86) in the cementless group and 76 years (52 to 94) in the cemented group, with 239 (70%) females in the whole series. Femoral complications were compared between the two groups. Mean follow-up was 1.5 years (0.1 to 4.4) for patients in the cementless group and 1.3 years (0.0 to 3.9) for patients in the cemented group.Aims
Methods
Hip arthroscopy is particularly attractive in
children as it confers advantages over arthrotomy or open surgery,
such as shorter recovery time and earlier return to activity. Developments
in surgical technique and arthroscopic
Total hip arthroplasty (THA) patients undergoing or having a prior lumbar spine fusion (LSF) have an increased risk of mechanical complications. The aim of this registry-based, retrospective comparative cohort study is to assess the longer term survival of THA in patients who have undergone a LSF during a 17-year period (2000 to 2017). A registry-based population study was conducted on 679 patients who underwent both THA and LSF surgeries. Patients were identified from the regional arthroplasty data base and cross linked to patients with LSF from the regional hospital discharge database between 2000 and 2017. Demographic data, diagnosis leading to primary THA, primary implant survival, perioperative complications, number and causes of failure, and patients requiring revision arthroplasty were collated and compared. For comparison, data from 67,919 primary THAs performed during the same time time period were also retrieved and analyzed.Aims
Methods
Cementless femoral stems are currently preferred
for total hip replacement (THR) in the United States. Improvements
in stem design,
Arthroplasty skills need to be acquired safely during training, yet operative experience is increasingly hard to acquire by trainees. Virtual reality (VR) training using headsets and motion-tracked controllers can simulate complex open procedures in a fully immersive operating theatre. The present study aimed to determine if trainees trained using VR perform better than those using conventional preparation for performing total hip arthroplasty (THA). A total of 24 surgical trainees (seven female, 17 male; mean age 29 years (28 to 31)) volunteered to participate in this observer-blinded 1:1 randomized controlled trial. They had no prior experience of anterior approach THA. Of these 24 trainees, 12 completed a six-week VR training programme in a simulation laboratory, while the other 12 received only conventional preparatory materials for learning THA. All trainees then performed a cadaveric THA, assessed independently by two hip surgeons. The primary outcome was technical and non-technical surgical performance measured by a THA-specific procedure-based assessment (PBA). Secondary outcomes were step completion measured by a task-specific checklist, error in acetabular component orientation, and procedure duration.Aims
Patients and Methods
Fixation of osteoporotic proximal humerus fractures remains challenging even with state-of-the-art locking plates. Despite the demonstrated biomechanical benefit of screw tip augmentation with bone cement, the clinical findings have remained unclear, potentially as the optimal augmentation combinations are unknown. The aim of this study was to systematically evaluate the biomechanical benefits of the augmentation options in a humeral locking plate using finite element analysis (FEA). A total of 64 cement augmentation configurations were analyzed using six screws of a locking plate to virtually fix unstable three-part fractures in 24 low-density proximal humerus models under three physiological loading cases (4,608 simulations). The biomechanical benefit of augmentation was evaluated through an established FEA methodology using the average peri-screw bone strain as a validated predictor of cyclic cut-out failure.Aims
Methods
Revision arthroplasty of the hip is expensive
owing to the increased cost of pre-operative investigations, surgical implants
and
The aims of this study were to determine the change in the sagittal alignment of the pelvis and the associated impact on acetabular component position at one-year follow-up after total hip arthroplasty (THA). This study represents the one-year follow-up of a previous short-term study at our institution. Using the patient population from our prior study, the radiological pelvic ratio was assessed in 91 patients undergoing THA, of whom 50 were available for follow-up of at least one year (median 1.5; interquartile range (IQR) 1.1 to 2.0). Anteroposterior radiographs of the pelvis were obtained in the standing position preoperatively and at one year postoperatively. Pelvic ratio was defined as the ratio between the vertical distance from the inferior sacroiliac (SI) joints to the superior pubic symphysis and the horizontal distance between the inferior SI joints. Apparent acetabular component position changes were determined from the change in pelvic ratio. A change of at least 5° was considered clinically meaningful.Aims
Methods
Current treatments of prosthetic joint infection (PJI) are minimally effective against The ability of PlySs2 and vancomycin to kill biofilm and colony-forming units (CFUs) on orthopaedic implants were compared using in vitro models. An in vivo murine PJI model of DAIR was used to assess the efficacy of a combination of PlySs2 and vancomycin on periprosthetic bacterial load.Aims
Methods
This paper aims to review the evidence for patient-related factors associated with less favourable outcomes following hip arthroscopy. Literature reporting on preoperative patient-related risk factors and outcomes following hip arthroscopy were systematically identified from a computer-assisted literature search of Pubmed (Medline), Embase, and Cochrane Library using Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines and a scoping review.Aims
Methods
Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (Aims
Patients and Methods
The aim of this study was to assess the functional gain achieved following hip resurfacing arthroplasty (HRA). A total of 28 patients (23 male, five female; mean age, 56 years (25 to 73)) awaiting Birmingham HRA volunteered for this prospective gait study, with an age-matched control group of 26 healthy adults (16 male, ten female; mean age, 56 years (33 to 84)). The Oxford Hip Score (OHS) and gait analysis using an instrumented treadmill were used preoperatively and more than two years postoperatively to measure the functional change attributable to the intervention.Aims
Patients and Methods
Leucocyte esterase (LE) has been shown to be an accurate marker of prosthetic joint infection (PJI), and has been proposed as an alternative to frozen section (FS) histology for intraoperative diagnosis. In this study, the intraoperative assessment of LE was compared with FS histology for the diagnosis of prosthetic hip infection. A total of 119 patients undergoing revision total hip arthroplasty (THA) between June 2015 and December 2017 were included in the study. There were 56 men and 63 women with a mean age of 66.2 years (27 to 88). Synovial fluid was collected before arthrotomy for the assessment of LE using enzymatic colourimetric strips. Between five and six samples were stained with haematoxylin and eosin for FS histology, and considered suggestive of infection when at least five polymorphonuclear leucocytes were found in five high-power fields.Aims
Patients and Methods
The aim of this study was to evaluate the accuracy of implant placement when using robotic assistance during total hip arthroplasty (THA). A total of 20 patients underwent a planned THA using preoperative CT scans and robotic-assisted software. There were nine men and 11 women (n = 20 hips) with a mean age of 60.8 years (Aims
Patients and Methods
The aims of this study were to review the surgical technique for a combined femoral head reduction osteotomy (FHRO) and periacetabular osteotomy (PAO), and to report the short-term clinical and radiological results of a combined FHRO/PAO for the treatment of selected severe femoral head deformities. Between 2011 and 2016, six female patients were treated with a combined FHRO and PAO. The mean patient age was 13.6 years (12.6 to 15.7). Clinical data, including patient demographics and patient-reported outcome scores, were collected prospectively. Radiologicalally, hip morphology was assessed evaluating the Tönnis angle, the lateral centre to edge angle, the medial offset distance, the extrusion index, and the alpha angle.Aims
Patients and Methods
The management of acetabular defects at the time of revision hip arthroplasty surgery is a challenge. This study presents the results of a long-term follow-up study of the use of irradiated allograft bone in acetabular reconstruction. Between 1990 and 2000, 123 hips in 110 patients underwent acetabular reconstruction for aseptic loosening, using impaction bone grafting with frozen, irradiated, and morsellized femoral heads and a cemented acetabular component. A total of 55 men and 55 women with a mean age of 64.3 years (26 to 97) at the time of revision surgery are included in this study.Aims
Patients and Methods
Lumbar fusion is known to reduce the variation in pelvic tilt
between standing and sitting. A flexible lumbo-pelvic unit increases
the stability of total hip arthroplasty (THA) when seated by increasing
anterior clearance and acetabular anteversion, thereby preventing
impingement of the prosthesis. Lumbar fusion may eliminate this protective
pelvic movement. The effect of lumbar fusion on the stability of
total hip arthroplasty has not previously been investigated. The Medicare database was searched for patients who had undergone
THA and spinal fusion between 2005 and 2012. PearlDiver software
was used to query the database by the International Classification
of Diseases, 9th Revision, Clinical Modification (ICD-9-CM) procedural
code for primary THA and lumbar spinal fusion. Patients who had
undergone both lumbar fusion and THA were then divided into three
groups: 1 to 2 levels, 3 to 7 levels and 8+ levels of fusion. The
rate of dislocation in each group was established using ICD-9-CM codes.
Patients who underwent THA without spinal fusion were used as a
control group. Statistical significant difference between groups
was tested using the chi-squared test, and significance set at p
<
0.05.Aims
Patients and Methods
Accurate placement of the acetabular component is essential in
total hip arthroplasty (THA). The purpose of this study was to determine
if the ability to achieve inclination of the acetabular component
within the ‘safe-zone’ of 30° to 50° could be improved with the
use of an inclinometer. We reviewed 167 primary THAs performed by a single surgeon over
a period of 14 months. Procedures were performed at two institutions:
an inpatient hospital, where an inclinometer was used (inclinometer
group); and an ambulatory centre, where an inclinometer was not
used as it could not be adequately sterilized (control group). We excluded
47 patients with a body mass index (BMI) of > 40 kg/m2,
age of > 68 years, or a surgical indication other than osteoarthritis
whose treatment could not be undertaken in the ambulatory centre.
There were thus 120 patients in the study, 68 in the inclinometer
group and 52 in the control group. The inclination angles of the acetabular
component were measured from de-identified plain radiographs by
two blinded investigators who were not involved in the surgery.
The effect of the use of the inclinometer on the inclination angle
was determined using multivariate regression analysis.Aims
Patients and Methods
Osteosynthesis of anterior pubic ramus fractures using one large-diameter screw can be challenging in terms of both surgical procedure and fixation stability. Small-fragment screws have the advantage of following the pelvic cortex and being more flexible. The aim of the present study was to biomechanically compare retrograde intramedullary fixation of the superior pubic ramus using either one large- or two small-diameter screws. A total of 12 human cadaveric hemipelvises were analysed in a matched pair study design. Bone mineral density of the specimens was 68 mgHA/cm3 (standard deviation (Objectives
Materials and Methods
The number of patients undergoing arthroscopic surgery of the
hip has increased significantly during the past decade. It has now
become an established technique for the treatment of many intra-
and extra-articular conditions affecting the hip. However, it has
a steep learning curve and is not without the risk of complications.
The purpose of this systematic review was to determine the prevalence
of complications during and following this procedure. Preferred Reporting Items for Systematic Reviews and Meta-Analyses
guidelines were used in designing this study. Two reviewers systematically
searched the literature for complications related to arthroscopy
of the hip. The research question and eligibility criteria were
established Aims
Materials and Methods
We wished to quantify the extent of soft-tissue damage sustained
during minimally invasive total hip arthroplasty through the direct
anterior (DA) and direct superior (DS) approaches. In eight cadavers, the DA approach was performed on one side,
and the DS approach on the other, a single brand of uncemented hip
prosthesis was implanted by two surgeons, considered expert in their
surgical approaches. Subsequent reflection of the gluteus maximus
allowed the extent of muscle and tendon damage to be measured and
the percentage damage to each anatomical structure to be calculated.Aims
Materials and Methods
This review summarises the technique of impaction
grafting with mesh augmentation for the treatment of uncontained
acetabular defects in revision hip arthroplasty. The ideal acetabular revision should restore bone stock, use
a small socket in the near-anatomic position, and provide durable
fixation. Impaction bone grafting, which has been in use for over
40 years, offers the ability to achieve these goals in uncontained
defects. The precepts of modern, revision impaction grafting are
that the segmental or cavitary defects must be supported with a
mesh; the contained cavity is filled with vigorously impacted morselised
fresh-frozen allograft; and finally, acrylic cement is used to stabilise
the graft and provide rigid, long-lasting fixation of the revised
acetabular component. Favourable results have been published with this technique. While
having its limitations, it is a viable option to address large acetabular
defects in revision arthroplasty. Cite this article:
One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR.Aims
Patients and Methods
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
The primary aim of this study was to analyse the position of
the acetabular and femoral components in total hip arthroplasty
undertaken using an anterior surgical approach. In a prospective, single centre study, we used the EOS imaging
system to analyse the position of components following THA performed
via the anterior approach in 102 patients (103 hips) with a mean
age of 64.7 years (Aims
Patients and Methods
This study investigated whether the use of tranexamic acid (TXA)
decreased blood loss and transfusion related cost following surface
replacement arthroplasty (SRA). A retrospective review of patients treated with TXA during a
SRA, who did not receive autologous blood (TXA group) was performed.
Two comparison groups were established; the first group comprised
of patients who donated their own blood pre-operatively (auto group)
and the second of patients who did not donate blood pre-operatively
(control). Outcomes included transfusions, post-operative haemoglobin
(Hgb), complications, and length of post-operative stay. Aims
Methods
We evaluated the accuracy with which a custom-made
acetabular component could be positioned at revision arthroplasty
of the hip in patients with a Paprosky type 3 acetabular defect. A total of 16 patients with a Paprosky type 3 defect underwent
revision surgery using a custom-made trabecular titanium implant.
There were four men and 12 women with a median age of 67 years (48
to 79). The planned inclination (INCL), anteversion (AV), rotation
and centre of rotation (COR) of the implant were compared with the post-operative
position using CT scans. A total of seven implants were malpositioned in one or more parameters:
one with respect to INCL, three with respect to AV, four with respect
to rotation and five with respect to the COR. To the best of our knowledge, this is the first study in which
CT data acquired for the pre-operative planning of a custom-made
revision acetabular implant have been compared with CT data on the
post-operative position. The results are encouraging. Cite this article:
There is little evidence on the cost effectiveness
of different brands of hip prostheses. We compared lifetime cost effectiveness
of frequently used brands within types of prosthesis including cemented
(Exeter V40 Contemporary, Exeter V40 Duration and Exeter V40 Elite
Plus Ogee), cementless (Corail Pinnacle, Accolade Trident, and Taperloc Exceed)
and hybrid (Exeter V40 Trilogy, Exeter V40 Trident, and CPT Trilogy).
We used data from three linked English national databases to estimate
the lifetime risk of revision, quality-adjusted life years (QALYs)
and cost. For women with osteoarthritis aged 70 years, the Exeter V40 Elite
Plus Ogee had the lowest risk of revision (5.9% revision risk, 9.0
QALYs) and the CPT Trilogy had the highest QALYs (10.9% revision
risk, 9.3 QALYs). Compared with the Corail Pinnacle (9.3% revision
risk, 9.22 QALYs), the most commonly used brand, and assuming a
willingness-to-pay of £20 000 per QALY gain, the CPT Trilogy is
most cost effective, with an incremental net monetary benefit of £876.
Differences in cost effectiveness between the hybrid CPT Trilogy
and Exeter V40 Trident and the cementless Corail Pinnacle and Taperloc
Exceed were small, and a cautious interpretation is required, given
the limitations of the available information. However, it is unlikely that cemented brands are among the most
cost effective. Similar patterns of results were observed for men
and other ages. The gain in quality of life after total hip arthroplasty,
rather than the risk of revision, was the main driver of cost effectiveness. Cite this article:
The technical advances in arthroscopic surgery
of the hip, including the improved ability to manage the capsule
and gain extensile exposure, have been paralleled by a growth in
the number of conditions that can be addressed. This expanding list
includes symptomatic labral tears, chondral lesions, injuries of
the ligamentum teres, femoroacetabular impingement (FAI), capsular
laxity and instability, and various extra-articular disorders, including snapping
hip syndromes. With a careful diagnostic evaluation and technical
execution of well-indicated procedures, arthroscopic surgery of
the hip can achieve successful clinical outcomes, with predictable
improvements in function and pre-injury levels of physical activity
for many patients. This paper reviews the current position in relation to the use
of arthroscopy in the treatment of disorders of the hip. Cite this article:
We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip. The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° ( Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component.
We performed a case–control study to compare
the rates of further surgery, revision and complications, operating time
and survival in patients who were treated with either an uncemented
hydroxyapatite-coated Corail bipolar femoral stem or a cemented
Exeter stem for a displaced intracapsular fracture of the hip. The
mean age of the patients in the uncemented group was 82.5 years
(53 to 97) and in the cemented group was 82.7 years (51 to 99) We used
propensity score matching, adjusting for age, gender and the presence
or absence of dementia and comorbidities, to produce a matched cohort
receiving an Exeter stem (n = 69) with which to compare the outcome of
patients receiving a Corail stem (n = 69). The Corail had a significantly
lower all-cause rate of further surgery (p = 0.016; odds ratio (OR)
0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further
surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating
time was significantly less for the Corail group than for the cemented Exeter
group (59 min [12 to 136] Cite this article:
Conventional uncemented femoral implants provide
dependable long-term fixation in patients with a wide range of functional
requirements. Yet challenges associated with proximal–distal femoral
dimensional mismatch, preservation of bone stock, and minimally
invasive approaches have led to exploration into alternative implant designs.
Short stem designs focusing on a stable metaphyseal fit have emerged
to address these issues in total hip replacement (THR). Uncemented
metaphyseal-engaging short stem implants are stable and are associated
with proximal bone remodeling closer to the metaphysis when compared
with conventional stems and they also have comparable clinical performances.
Short stem metaphyseal-engaging implants can meet the goals of a
successful THR, including tolerating a high level of patient function,
as well as durable fixation. Cite this article:
This conversation represents an attempt by several
arthroplasty surgeons to critique several abstracts presented over
the last year as well as to use them as a jumping off point for trying
to figure out where they fit in into our current understanding of
multiple issues in modern hip and knee arthroplasty.
Over recent years hip arthroscopic surgery has
evolved into one of the most rapidly expanding fields in orthopaedic surgery.
Complications are largely transient and incidences between 0.5%
and 6.4% have been reported. However, major complications can and
do occur. This article analyses the reported complications and makes recommendations
based on the literature review and personal experience on how to
minimise them.
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
The aim of this study was to determine the accuracy
of registration and the precision of the resection volume in navigated
hip arthroscopy for cam-type femoroacetabular impingement, using
imageless and image-based registration. A virtual cam lesion was
defined in 12 paired cadaver hips and randomly assigned to either
imageless or image-based (three-dimensional (3D) fluoroscopy) navigated
arthroscopic head–neck osteochondroplasty. The accuracy of patient–image
registration for both protocols was evaluated and post-operative
imaging was performed to evaluate the accuracy of the surgical resection.
We found that the estimated accuracy of imageless registration in the
arthroscopic setting was poor, with a mean error of 5.6 mm (standard
deviation ( In conclusion, given the limited femoral surface that can be
reached and digitised during arthroscopy of the hip, imageless registration
is inaccurate and does not allow for reliable surgical navigation.
However, image-based registration does acceptably allow for guided
femoral osteochondroplasty in the arthroscopic management of femoroacetabular
impingement.