Using human cadaver specimens, we investigated
the role of supplementary fibular plating in the treatment of distal
tibial fractures using an intramedullary nail. Fibular plating is
thought to improve stability in these situations, but has been reported
to have increased soft-tissue complications and to impair union
of the fracture. We proposed that multidirectional locking screws
provide adequate stability, making additional fibular plating unnecessary.
A distal tibiofibular osteotomy model performed on matched fresh-frozen
lower limb specimens was stabilised with reamed nails using conventional
biplanar distal locking (CDL) or multidirectional distal locking
(MDL) options with and without fibular plating. Rotational stiffness
was assessed under a constant axial force of 150 N and a superimposed
torque of ± 5 Nm. Total movement, and neutral zone and fracture
gap movement were analysed. In the CDL group, fibular plating improved stiffness at the tibial
fracture site, albeit to a small degree (p = 0.013). In the MDL
group additional fibular plating did not increase the stiffness.
The MDL nail without fibular plating was significantly more stable
than the CDL nail with an additional fibular plate (p = 0.008). These findings suggest that additional fibular plating does not
improve stability if a multidirectional distal locking intramedullary
nail is used, and is therefore unnecessary if not needed to aid
reduction. Cite this article:
We describe the results after open reduction and internal fixation of 22 consecutive displaced fractures of the glenoid with a mean follow-up of ten years. A posterior approach was used in 16 patients and an anterior in six, the approach being chosen according to the Ideberg classification of the fractures. The fixation failed in two patients, one of whom required a further operation. There were two cases of deep infection. At follow-up the median Constant score was 94% (mean 79%, range 17 to 100). The score was less than 50% in four patients, including the two who became infected. A further two had an associated complete palsy of the brachial plexus.
Rotational deformity following intramedullary nailing may cause symptoms and require surgical correction by osteotomy. Reamed, locked intramedullary nailing may be performed, but concern about cortical blood supply and potential pulmonary dysfunction from reaming have led many surgeons to limit this and use smaller diameter nails. Slotted nails are commonly used but are less stiff in torsion than the newer unslotted nails, particularly at the lower diameters. We report two cased of recurrent femoral rotational deformity after using statically interlocked slotted intramedullary nails to correct existing femoral rotational deformities. These patients show that small diameter statically interlocked femoral nails with diminished bone-nail contact must be stiff enough in rotation to avoid potential recurrence.
The accuracy of templates used for the preoperative planning of the fixation of intramedullary fractures depends on radiological magnification. To study the accuracy of these templates, we randomly selected 100 femoral and 100 tibial radiographs taken after stabilisation by an intramedullary nail using a standard technique. We then compared the known nail length with the corresponding measurements on the radiographs. The mean magnification factor for the femur was 9% and for the tibia 7%; these differ considerably from the range of magnification of the manufacturers’ templates (femur, 15% to 17%; tibia 10% to 15%). We conclude that templates are unreliable for the selection of implant length and that this should be done by intraoperative measurements.