Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Trauma

ACCURACY OF INTRAMEDULLARY TEMPLATES IN FEMORAL AND TIBIAL RADIOGRAPHS



Download PDF

Abstract

The accuracy of templates used for the preoperative planning of the fixation of intramedullary fractures depends on radiological magnification. To study the accuracy of these templates, we randomly selected 100 femoral and 100 tibial radiographs taken after stabilisation by an intramedullary nail using a standard technique. We then compared the known nail length with the corresponding measurements on the radiographs.

The mean magnification factor for the femur was 9% and for the tibia 7%; these differ considerably from the range of magnification of the manufacturers’ templates (femur, 15% to 17%; tibia 10% to 15%). We conclude that templates are unreliable for the selection of implant length and that this should be done by intraoperative measurements.


Correspondence should be sent to Professor C. Krettek.

For access options please click here