The processes linking long-term bisphosphonate treatment to atypical fracture remain elusive. To establish a means of exploring this link, we have examined how long-term bisphosphonate treatment with prior ovariectomy modifies femur fracture behaviour and tibia mass and shape in murine bones. Three groups (seven per group) of 12-week-old mice were: 1) ovariectomized and 20 weeks thereafter treated weekly for 24 weeks with 100 μm/kg subcutaneous ibandronate (OVX+IBN); 2) ovariectomized (OVX); or 3) sham-operated (SHAM). Quantitative fracture analysis generated biomechanical properties for the femoral neck. Tibiae were microCT scanned and trabecular (proximal metaphysis) and cortical parameters along almost its whole length measured.Aims
Methods
Concerns have been raised that deformation of
acetabular shells may disrupt the assembly process of modular prostheses.
In this study we aimed to examine the effect that the strength of
bone has on the amount of deformation of the acetabular shell. The
hypothesis was that stronger bone would result in greater deformation.
A total of 17 acetabular shells were inserted into the acetabula
of eight cadavers, and deformation was measured using an optical
measuring system. Cores of bone from the femoral head were taken
from each cadaver and compressed using a materials testing machine.
The highest peak modulus and yield stress for each cadaver were used
to represent the strength of the bone and compared with the values
for the deformation and the surgeon’s subjective assessment of the
hardness of the bone. The mean deformation of the shell was 129
µm (3 to 340). No correlation was found between deformation and
either the maximum peak modulus (r² = 0.011, t = 0.426, p = 0.676) or
the yield stress (r² = 0.024, t = 0.614, p = 0.549) of the bone.
Although no correlation was found between the strength of the bone
and deformation, the values for the deformation observed could be
sufficient to disrupt the assembly process of modular acetabular
components. Cite this article:
Aims. Type 2 diabetes mellitus (T2DM) impairs
The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component. We performed a retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All POPFFs were identified using regional radiograph archiving and electronic care systems.Aims
Methods
Assessment of bone mineral density (BMD) with dual-energy X-ray absorptiometry (DXA) is a well-established clinical technique, but it is not available in the acute trauma setting. Thus, it cannot provide a preoperative estimation of BMD to help guide the technique of fracture fixation. Alternative methods that have been suggested for assessing BMD include: 1) cortical measures, such as cortical ratios and combined cortical scores; and 2) aluminium grading systems from preoperative digital radiographs. However, limited research has been performed in this area to validate the different methods. The aim of this study was to investigate the evaluation of BMD from digital radiographs by comparing various methods against DXA scanning. A total of 54 patients with distal radial fractures were included in the study. Each underwent posteroanterior (PA) and lateral radiographs of the injured wrist with an aluminium step wedge. Overall 27 patients underwent routine DXA scanning of the hip and lumbar spine, with 13 undergoing additional DXA scanning of the uninjured forearm. Analysis of radiographs was performed on ImageJ and Matlab with calculations of cortical measures, cortical indices, combined cortical scores, and aluminium equivalent grading.Aims
Methods
Cementless acetabular components rely on press-fit fixation for initial stability. In certain cases, initial stability is more difficult to obtain (such as during revision). No current study evaluates how a surgeon’s impaction technique (mallet mass, mallet velocity, and number of strikes) may affect component fixation. This study seeks to answer the following research questions: 1) how does impaction technique affect a) bone strain generation and deterioration (and hence implant stability) and b) seating in different density bones?; and 2) can an impaction technique be recommended to minimize risk of implant loosening while ensuring seating of the acetabular component? A custom drop tower was used to simulate surgical strikes seating acetabular components into synthetic bone. Strike velocity and drop mass were varied. Synthetic bone strain was measured using strain gauges and stability was assessed via push-out tests. Polar gap was measured using optical trackers.Aims
Methods
We aimed to assess the influence of ethnicity on the incidence
of heterotopic ossification (HO) after total hip arthroplasty (THA). We studied the six-month post-operative anteroposterior radiographs
of 1449 consecutive primary THAs (1324 patients) and retrospectively
graded them for the presence of HO, using the Brooker Classification. Aims
Patients and Methods
There is an increased risk of fracture following
osteoplasty of the femoral neck for cam-type femoroacetabular impingement
(FAI). Resection of up to 30% of the anterolateral head–neck junction
has previously been considered to be safe, however, iatrogenic fractures
have been reported with resections within these limits. We re-evaluated
the amount of safe resection at the anterolateral femoral head–neck
junction using a biomechanically consistent model. In total, 28 composite bones were studied in four groups: control,
10% resection, 20% resection and 30% resection. An axial load was
applied to the adducted and flexed femur. Peak load, deflection
at time of fracture and energy to fracture were assessed using comparison
groups. There was a marked difference in the mean peak load to fracture
between the control group and the 10% resection group (p <
0.001).
The control group also tolerated significantly more deflection before
failure (p <
0.04). The mean peak load (p = 0.172), deflection
(p = 0.547), and energy to fracture (p = 0.306) did not differ significantly between
the 10%, 20%, and 30% resection groups. Any resection of the anterolateral quadrant of the femoral head–neck
junction for FAI significantly reduces the load-bearing capacity
of the proximal femur. After initial resection of cortical bone,
there is no further relevant loss of stability regardless of the
amount of trabecular bone resected. Based on our findings we recommend any patients who undergo anterolateral
femoral head–neck junction osteoplasty should be advised to modify
their post-operative routine until cortical remodelling occurs to
minimise the subsequent fracture risk. Cite this article:
We matched 78 patients with a loose cemented Charnley Elite Plus total hip replacement (THR) by age, gender, race, prosthesis and time from surgery with 49 patients with a well-fixed stable hip replacement, to determine if poor bone quality predisposes to loosening. Clinical, radiological, biomechanical and bone mineral density indicators of bone quality were assessed. Patients with loose replacements had more pain, were more likely to have presented with atrophic arthritis and to have a history of fragility fracture, narrower femoral cortices and lower peri-prosthetic or lumbar spine bone mineral density (all In this series of cemented hip replacements performed between 1994 and 1998, aseptic loosening was associated with poor bone quality. Patients with a THR should be screened for osteoporosis and have regular radiological surveillance.