Advertisement for orthosearch.org.uk
Results 1 - 7 of 7
Results per page:
The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1503 - 1509
1 Nov 2014
Ritter MA Davis KE Small SR Merchun JG Farris A

The relationship between post-operative bone density and subsequent failure of total knee replacement (TKR) is not known. This retrospective study aimed to determine the relationship between bone density and failure, both overall and according to failure mechanism. All 54 aseptic failures occurring in 50 patients from 7760 consecutive primary cemented TKRs between 1983 and 2004 were matched with non-failing TKRs, and 47 failures in 44 patients involved tibial failures with the matching characteristics of age (65.1 for failed and 69.8 for non-failed), gender (70.2% female), diagnosis (93.6% OA), date of operation, bilaterality, pre-operative alignment (0.4 and 0.3 respectively), and body mass index (30.2 and 30.0 respectively). In each case, the density of bone beneath the tibial component was assessed at each follow-up interval using standardised, calibrated radiographs. Failing knees were compared with controls both overall and, as a subgroup analysis, by failure mechanism. Knees were compared with controls using univariable linear regression.

Significant and continuous elevation in tibial density was found in knees that eventually failed by medial collapse (p < 0.001) and progressive radiolucency (p < 0.001) compared with controls, particularly in the medial region of the tibia. Knees failing due to ligamentous instability demonstrated an initial decline in density (p = 0.0152) followed by a non-decreasing density over time (p = 0.034 for equivalence). Non-failing knees reported a decline in density similar to that reported previously using dual-energy x-ray absorptiometry (DEXA). Differences between failing and non-failing knees were observable as early as two months following surgery. This tool may be used to identify patients at risk of failure following TKR, but more validation work is needed.

Cite this article: Bone Joint J 2014;96-B:1503–9.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1484 - 1489
1 Nov 2013
Meding JB Ritter MA Davis KE Farris A

The strain on clinic and surgeon resources resulting from a rise in demand for total knee replacement (TKR) requires reconsideration of when and how often patients need to be seen for follow-up. Surgeons will otherwise require increased paramedical staff or need to limit the number of TKRs they undertake. We reviewed the outcome data of 16 414 primary TKRs undertaken at our centre to determine the time to re-operation for any reason and for specific failure mechanisms. Peak risk years for failure were determined by comparing the conditional probability of failure, the number of failures divided by the total number of TKRs cases, for each year. The median times to failure for the most common failure mechanisms were 4.9 years (interquartile range (IQR) 1.7 to 10.7) for femoral and tibial loosening, 1.9 years (IQR 0.8 to 3.9) for infection, 3.1 years (IQR 1.6 to 5.5) for tibial collapse and 5.6 years (IQR 3.4 to 9.3) for instability. The median time to failure for all revisions was 3.3 years (IQR 1.2 to 8.5), with an overall revision rate of 1.7% (n = 282). Results from our patient population suggest that patients be seen for follow-up at six months, one year, three years, eight years, 12 years, and every five years thereafter. Patients with higher pain in the early post-operative period or high body mass index (≥ 41 kg/m2) should be monitored more closely.

Cite this article: Bone Joint J 2013;95-B:1484–9.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 911 - 916
1 Jul 2013
Small SR Ritter MA Merchun JG Davis KE Rogge RD

Stress shielding resulting in diminished bone density following total knee replacement (TKR) may increase the risk of migration and loosening of the prosthesis. This retrospective study was designed to quantify the effects of the method of fixation on peri-prosthetic tibial bone density beneath cemented and uncemented tibial components of similar design and with similar long-term survival rates. Standard radiographs taken between two months and 15 years post-operatively were digitised from a matched group of TKRs using cemented (n = 67) and uncemented (n = 67) AGC tibial prostheses. Digital radiograph densitometry was used to quantify changes in bone density over time. Age, length of follow-up, gender, body mass index and alignment each significantly influenced the long-term pattern of peri-prosthetic bone density. Similar long-term changes in density irrespective of the method of fixation correlated well with the high rate of survival of this TKR at 20 years, and suggest that cemented and uncemented fixation are both equally viable.

Cite this article: Bone Joint J 2013;95-B:911–16.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 64 - 70
1 Apr 2012
Ritter MA Davis KE Meding JB Farris A

Objectives

The purpose of this study was to examine the effect of posterior cruciate ligament (PCL) retention, PCL recession, and PCL excision during cruciate-retaining total knee replacement.

Methods

A total of 3018 anatomic graduated component total knee replacements were examined; 1846 of these retained the PCL, 455 PCLs were partially recessed, and in 717 the PCL was completely excised from the back of the tibia.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 745 - 749
1 Jun 2009
Ritter MA

The long-term success of total knee replacement is multifactorial, including factors relating to the patient, the operation and the implant. The purpose of this study was to examine the 20-year survival of the cemented Anatomical Graduated Component (AGC) total knee replacement. Between 1983 and 2004, 7760 of these were carried out at our institution. Of these, 6726 knees which received the non-modular metal-backed tibial component with compression-moulded polyethylene and had a minimum two-year follow-up were available for study. In all, 36 knees were followed over 20 years with a survival of the tibial and femoral components together of 97.8% (95% confidence interval (CI) 0.9851 to 0.9677), with no implants being revised for polyethylene wear or osteolysis. Age > 70 was associated with increased survival (99.6%, 95% CI 99.0 to 99.8) (p < 0.0001) but pre-operative valgus alignment reduced survival (95.1%, 95% CI 90.0 to 97.6) (p = 0.0056). Age < 55 (p = 0.129), pre-operative varus alignment (p = 0.707), osteonecrosis (p = 0.06), rheumatoid arthritis (p = 0.247), and gender (p = 0.666) were not statistically associated with failure.

We attribute the success of the AGC implant to its relatively unconstrained articular geometry and the durability of a non-modular metal-backed tibial component with compression moulded polyethylene.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 1 | Pages 43 - 47
1 Jan 2004
Ritter MA Faris PM Thong AE Davis KE Meding JB Berend ME

Interest in unicompartmental knee arthroplasty (UKA) for the treatment of medial compartment osteoarthritis has increased in recent years with apparent improvement in the long-term results. This is a result of improved surgical technique, patient selection, and implant design. In an effort further to improve patient selection we analysed the relationship between the pre-operative alignment of the knee and the anatomical findings at the time of surgery. We compared these findings with the indications for UKA.

From 4021 total knee arthroplasties we compared intra-operative observations with the pre-operative clinical data in order to identify knees with isolated, medial, compartment changes, which would have been ideal candidates for UKA.

We found that only 247 of the knees (6.1%) met anatomical qualifications for isolated, medial, unicompartmental osteoarthritis, and of these, only 168 (4.3%) met clinical standards ideal for UKA. Pre-operative alignment showed a significant relationship with patterns of disease. Logistic regression revealed a relationship between pre-operative alignment and intra-operative findings resembling a Gaussian distribution. Patients with a pre-operative varus alignment of 7° were slightly more likely to be selected for UKA. But the further the anatomical alignment in either direction varies from 7° of varus, the more unlikely it is for the knee to exhibit a disease pattern of isolated, medial, unicompartmental osteoarthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 2 | Pages 301 - 303
1 Mar 1999
Ritter MA Koehler M Keating EM Faris PM Meding JB

The postoperative analgesic effects of intra-articular injections of bupivacaine and/or morphine were examined prospectively in 437 patients who had total knee replacement for osteoarthritis. They were divided randomly into four groups. Group I received 10 mg of morphine (1 ml) and 9 ml of saline, group II received 10 ml of bupivacaine (2.5 mg/ml), group III received 10 ml of saline, and group IV received 10 mg of morphine (1 ml) and 9 ml of bupivacaine (2.5 mg/ml). All analgesics administered in the first 24 hours after operation were recorded. The patients rated their pain on the McGill-Melzack scale at 1, 6, 12 and 24 hours. No significant differences were found between any of the groups in the use of Demoral and/or Toradol in 24 hours, the length of stay in hospital or the pain rating at 1, 6, 12 or 24 hours. Patients in groups I and IV, whose injections included morphine, used significantly more morphine in the first 24 postoperative hours than did groups II or III.