Bone & Joint Research
Vol. 2, Issue 3 | Pages 58 - 65
1 Mar 2013
Johnson R Jameson SS Sanders RD Sargant NJ Muller SD Meek RMD Reed MR

Objectives

To review the current best surgical practice and detail a multi-disciplinary approach that could further reduce joint replacement infection.

Methods

Review of relevant literature indexed in PubMed.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 8 | Pages 1214 - 1219
1 Nov 2004
Jafri AA Green SM Partington PF McCaskie AW Muller SD

Fatigue fractures which originate at stress-concentrating voids located at the implant-cement interface are a potential cause of septic loosening of cemented femoral components. Heating of the component to 44°C is known to reduce the porosity of the cement-prosthesis interface.

The temperature of the cement-bone interface was recorded intra-operatively as 32.3°C. A simulated femoral model was devised to study the effect of heating of the component on the implant-cement interface.

Heating of the implant and vacuum mixing have a synergistic effect on the porosity of the implant-cement interface, and heating also reverses the gradients of microhardness in the mantle.

Heating of the implant also reduces porosity at the interface depending on the temperature. A minimum difference in temperature between the implant and the bone of 3°C was required to produce this effect. The optimal difference was 7°C, representing a balance between maximal reduction of porosity and an increased risk of thermal injury. Using contemporary cementing techniques, heating the implant to 40°C is recommended to produce an optimum effect.