Advertisement for orthosearch.org.uk
Results 1 - 100 of 121
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 642 - 648
1 May 2015
Hunt NC Ghosh KM Blain AP Rushton SP Longstaff LM Deehan DJ

The aim of this study was to compare the maximum laxity conferred by the cruciate-retaining (CR) and posterior-stabilised (PS) Triathlon single-radius total knee arthroplasty (TKA) for anterior drawer, varus–valgus opening and rotation in eight cadaver knees through a defined arc of flexion (0º to 110º). The null hypothesis was that the limits of laxity of CR- and PS-TKAs are not significantly different. The investigation was undertaken in eight loaded cadaver knees undergoing subjective stress testing using a measurement rig. Firstly the native knee was tested prior to preparation for CR-TKA and subsequently for PS-TKA implantation. Surgical navigation was used to track maximal displacements/rotations at 0º, 30º, 60º, 90º and 110° of flexion. Mixed-effects modelling was used to define the behaviour of the TKAs. . The laxity measured for the CR- and PS-TKAs revealed no statistically significant differences over the studied flexion arc for the two versions of TKA. Compared with the native knee both TKAs exhibited slightly increased anterior drawer and decreased varus-valgus and internal-external roational laxities. We believe further study is required to define the clinical states for which the additional constraint offered by a PS-TKA implant may be beneficial. Cite this article: Bone Joint J 2015; 97-B:642–8


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 24 - 30
1 Jun 2020
Livermore AT Erickson JA Blackburn B Peters CL

Aims. A significant percentage of patients remain dissatisfied after total knee arthroplasty (TKA). The aim of this study was to determine whether the sequential addition of accelerometer-based navigation for femoral component preparation and sensor-guided ligament balancing improved complication rates, radiological alignment, or patient-reported outcomes (PROMs) compared with a historical control group using conventional instrumentation. Methods. This retrospective cohort study included 371 TKAs performed by a single surgeon sequentially. A historical control group, with the use of intramedullary guides for distal femoral resection and surgeon-guided ligament balancing, was compared with a group using accelerometer-based navigation for distal femoral resection and surgeon-guided balancing (group 1), and one using navigated femoral resection and sensor-guided balancing (group 2). Primary outcome measures were Patient-Reported Outcomes Measurement Information System (PROMIS) and Knee injury and Osteoarthritis Outcome (KOOS) scores measured preoperatively and at six weeks and 12 months postoperatively. The position of the components and the mechanical axis of the limb were measured postoperatively. The postoperative range of motion (ROM), haematocrit change, and complications were also recorded. Results. There were 194 patients in the control group, 103 in group 1, and 74 in group 2. There were no significant differences in baseline demographics between the groups. Patients in group 2 had significantly higher baseline mental health subscores than control and group 1 patients (53.2 vs 50.2 vs 50.2, p = 0.041). There were no significant differences in any PROMs at six weeks or 12 months postoperatively (p > 0.05). There was no difference in the rate of manipulation under anaesthesia (MUA), complication rates, postoperative ROM, or blood loss. There were fewer mechanical axis outliers in groups 1 and 2 (25.2%, 14.9% respectively) versus control (28.4%), but this was not statistically significant (p = 0.10). Conclusion. The sequential addition of navigation of the distal femoral cut and sensor-guided ligament balancing did not improve short-term PROMs, radiological outcomes, or complication rates compared with conventional techniques. The costs of these added technologies may not be justified. Cite this article: Bone Joint J 2020;102-B(6 Supple A):24–30


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 601 - 605
1 May 2006
Pitto RP Graydon AJ Bradley L Malak SF Walker CG Anderson IA

The object of this study was to develop a method to assess the accuracy of an image-free total knee replacement navigation system in legs with normal or abnormal mechanical axes. A phantom leg was constructed with simulated hip and knee joints and provided a means to locate the centre of the ankle joint. Additional joints located at the midshaft of the tibia and femur allowed deformation in the flexion/extension, varus/valgus and rotational planes. Using a digital caliper unit to measure the coordinates precisely, a software program was developed to convert these local coordinates into a determination of actual leg alignment. At specific points in the procedure, information was compared between the digital caliper measurements and the image-free navigation system. Repeated serial measurements were undertaken. In the setting of normal alignment the mean error of the system was within 0.5°. In the setting of abnormal plane alignment in both the femur and the tibia, the error was within 1°. This is the first study designed to assess the accuracy of a clinically-validated navigation system. It demonstrates in vitro accuracy of the image-free navigation system in both normal and abnormal leg alignment settings


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 471 - 476
1 Apr 2007
Kim Y Kim J Yoon S

Bilateral sequential total knee replacement was carried out under one anaesthetic in 100 patients. One knee was replaced using a CT-free computer-assisted navigation system and the other conventionally without navigation. The two methods were compared for accuracy of orientation and alignment of the components. There were 85 women and 15 men with a mean age of 67.6 years (54 to 83). Radiological and CT imaging was carried out to determine the alignment of the components. The mean follow-up was 2.3 years (2 to 3). The operating and tourniquet times were significantly longer in the navigation group (p < 0.001). There were no significant pre- or post-operative differences between the knee scores of the two groups (p = 0.288 and p = 0.429, respectively). The results of imaging and the number of outliers for all radiological parameters were not statistically different (p = 0.109 to p = 0.920). In this series computer-assisted navigated total knee replacement did not result in more accurate orientation and alignment of the components than that achieved by conventional total knee replacement


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1025 - 1031
1 Aug 2008
Mizu-uchi H Matsuda S Miura H Okazaki K Akasaki Y Iwamoto Y

We compared the alignment of 39 total knee replacements implanted using the conventional alignment guide system with 37 implanted using a CT-based navigation system, performed by a single surgeon. The knees were evaluated using full-length weight-bearing anteroposterior radiographs, lateral radiographs and CT scans. The mean hip-knee-ankle angle, coronal femoral component angle and coronal tibial component angle were 181.8° (174.2° to 188.3°), 88.5° (84.0° to 91.8°) and 89.7° (86.3° to 95.1°), respectively for the conventional group and 180.8° (178.2° to 185.1°), 89.3° (85.8° to 92.0°) and 89.9° (88.0° to 93.0°), respectively for the navigated group. The mean sagittal femoral component angle was 85.5° (80.6° to 92.8°) for the conventional group and 89.6° (85.5° to 94.0°) for the navigated group. The mean rotational femoral and tibial component angles were −0.7° (−8.8° to 9.8°) and −3.3° (−16.8° to 5.8°) for the conventional group and −0.6° (−3.5° to 3.0°) and 0.3° (−5.3° to 7.7°) for the navigated group. The ideal angles of all alignments in the navigated group were obtained at significantly higher rates than in the conventional group. Our results demonstrated significant improvements in component positioning with a CT-based navigation system, especially with respect to rotational alignment


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 6 | Pages 830 - 835
1 Aug 2003
Sparmann M Wolke B Czupalla H Banzer D Zink A

We conducted this prospective randomised and externally evaluated study to investigate whether the use of a navigation system during total knee arthroplasty leads to significantly better results than the hand-guided technique. A total of 240 patients was included in the study. All patients received a condylar knee prosthesis. Two surgeons performed all the operations using the Stryker knee navigation system. Exclusion criteria included the necessity for the primary use of constrained implants. The results revealed a highly significant difference between the two groups in favour of navigation with regard to the mechanical axis, the frontal and sagittal femoral axis and the frontal tibial axis (p < 0.0001). The use of a navigation system was therefore shown to improve the alignment of the implant


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 151 - 152
1 Nov 2012
Lackey WG Berend ME


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 477 - 480
1 Apr 2007
Spencer JM Chauhan SK Sloan K Taylor A Beaver RJ

We previously compared the component alignment in total knee replacement using a computer-navigated technique with a conventional jig-based method. We randomly allocated 71 patients to undergo either computer-navigated or conventional replacement. An improved alignment was seen in the computer-navigated group.

The patients were then followed up post-operatively for two years, using the Knee Society score, the Short Form-36 health survey, the Western Ontario and McMaster Universities osteoarthritis index, the Bartlett Patellar pain questionnaire and the Oxford knee score, to assess functional outcome.

At two years post-operatively 60 patients were available for assessment, 30 in each group and 62 patients completed a postal survey. No patient in either group had undergone revision. All variables were analysed for differences between the groups either by Student’s t-test or the Mann-Whitney U test. Differences between the two groups did not reach significance for any of the outcome measures at any time point. At two years postoperatively, the frequency of mild to severe anterior pain was not significantly different (p = 0.818), varying between 44% (14) for the computer-navigated group, and 47% (14) for the conventionally-replaced group. The Bartlett Patellar score and the Oxford knee score were also not significantly different (t-test p = 0.161 and p = 0.607, respectively).

The clinical outcome of the patients with a computer-navigated knee replacement appears to be no different to that of a more conventional jig-based technique at two years post-operatively, despite the better alignment achieved with computer-navigated surgery.


Bone & Joint Open
Vol. 3, Issue 10 | Pages 767 - 776
5 Oct 2022
Jang SJ Kunze KN Brilliant ZR Henson M Mayman DJ Jerabek SA Vigdorchik JM Sculco PK

Aims. Accurate identification of the ankle joint centre is critical for estimating tibial coronal alignment in total knee arthroplasty (TKA). The purpose of the current study was to leverage artificial intelligence (AI) to determine the accuracy and effect of using different radiological anatomical landmarks to quantify mechanical alignment in relation to a traditionally defined radiological ankle centre. Methods. Patients with full-limb radiographs from the Osteoarthritis Initiative were included. A sub-cohort of 250 radiographs were annotated for landmarks relevant to knee alignment and used to train a deep learning (U-Net) workflow for angle calculation on the entire database. The radiological ankle centre was defined as the midpoint of the superior talus edge/tibial plafond. Knee alignment (hip-knee-ankle angle) was compared against 1) midpoint of the most prominent malleoli points, 2) midpoint of the soft-tissue overlying malleoli, and 3) midpoint of the soft-tissue sulcus above the malleoli. Results. A total of 932 bilateral full-limb radiographs (1,864 knees) were measured at a rate of 20.63 seconds/image. The knee alignment using the radiological ankle centre was accurate against ground truth radiologist measurements (inter-class correlation coefficient (ICC) = 0.99 (0.98 to 0.99)). Compared to the radiological ankle centre, the mean midpoint of the malleoli was 2.3 mm (SD 1.3) lateral and 5.2 mm (SD 2.4) distal, shifting alignment by 0.34. o. (SD 2.4. o. ) valgus, whereas the midpoint of the soft-tissue sulcus was 4.69 mm (SD 3.55) lateral and 32.4 mm (SD 12.4) proximal, shifting alignment by 0.65. o. (SD 0.55. o. ) valgus. On the intermalleolar line, measuring a point at 46% (SD 2%) of the intermalleolar width from the medial malleoli (2.38 mm medial adjustment from midpoint) resulted in knee alignment identical to using the radiological ankle centre. Conclusion. The current study leveraged AI to create a consistent and objective model that can estimate patient-specific adjustments necessary for optimal landmark usage in extramedullary and computer-guided navigation for tibial coronal alignment to match radiological planning. Cite this article: Bone Jt Open 2022;3(10):767–776


Bone & Joint Open
Vol. 2, Issue 11 | Pages 974 - 980
25 Nov 2021
Allom RJ Wood JA Chen DB MacDessi SJ

Aims. It is unknown whether gap laxities measured in robotic arm-assisted total knee arthroplasty (TKA) correlate to load sensor measurements. The aim of this study was to determine whether symmetry of the maximum medial and lateral gaps in extension and flexion was predictive of knee balance in extension and flexion respectively using different maximum thresholds of intercompartmental load difference (ICLD) to define balance. Methods. A prospective cohort study of 165 patients undergoing functionally-aligned TKA was performed (176 TKAs). With trial components in situ, medial and lateral extension and flexion gaps were measured using robotic navigation while applying valgus and varus forces. The ICLD between medial and lateral compartments was measured in extension and flexion with the load sensor. The null hypothesis was that stressed gap symmetry would not correlate directly with sensor-defined soft tissue balance. Results. In TKAs with a stressed medial-lateral gap difference of ≤1 mm, 147 (89%) had an ICLD of ≤15 lb in extension, and 112 (84%) had an ICLD of ≤ 15 lb in flexion; 157 (95%) had an ICLD ≤ 30 lb in extension, and 126 (94%) had an ICLD ≤ 30 lb in flexion; and 165 (100%) had an ICLD ≤ 60 lb in extension, and 133 (99%) had an ICLD ≤ 60 lb in flexion. With a 0 mm difference between the medial and lateral stressed gaps, 103 (91%) of TKA had an ICLD ≤ 15 lb in extension, decreasing to 155 (88%) when the difference between the medial and lateral stressed extension gaps increased to ± 3 mm. In flexion, 47 (77%) had an ICLD ≤ 15 lb with a medial-lateral gap difference of 0 mm, increasing to 147 (84%) at ± 3 mm. Conclusion. This study found a strong relationship between intercompartmental loads and gap symmetry in extension and flexion measured with prostheses in situ. The results suggest that ICLD and medial-lateral gap difference provide similar assessment of soft-tissue balance in robotic arm-assisted TKA. Cite this article: Bone Jt Open 2021;2(11):974–980


Bone & Joint Research
Vol. 9, Issue 6 | Pages 272 - 278
1 Jun 2020
Tapasvi S Shekhar A Patil S Pandit H

Aims. The mobile bearing Oxford unicompartmental knee arthroplasty (OUKA) is recommended to be performed with the leg in the hanging leg (HL) position, and the thigh placed in a stirrup. This comparative cadaveric study assesses implant positioning and intraoperative kinematics of OUKA implanted either in the HL position or in the supine leg (SL) position. Methods. A total of 16 fresh-frozen knees in eight human cadavers, without macroscopic anatomical defects, were selected. The knees from each cadaver were randomized to have the OUKA implanted in the HL or SL position. Results. Tibial base plate rotation was significantly more variable in the SL group with 75% of tibiae mal-rotated. Multivariate analysis of navigation data found no difference based on all kinematic parameters across the range of motion (ROM). However, area under the curve analysis showed that knees placed in the HL position had much smaller differences between the pre- and post-surgery conditions for kinematics mean values across the entire ROM. Conclusion. The sagittal tibia cut, not dependent on standard instrumentation, determines the tibial component rotation. The HL position improves accuracy of this step compared to the SL position, probably due to better visuospatial orientation of the hip and knee to the surgeon. The HL position is better for replicating native kinematics of the knee as shown by the area under the curve analysis. In the supine knee position, care must be taken during the sagittal tibia cut, while checking flexion balance and when sizing the tibial component


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1164 - 1171
1 Sep 2009
Bae DK Song SJ Yoon KH

We assessed the reliability, accuracy and variability of closed-wedge high tibial osteotomy (HTO) using computer-assisted surgery compared to the conventional technique. A total of 50 closed-wedge HTO procedures were performed using the navigation system, and compared with 50 HTOs that had been performed with the conventional technique. In the navigation group, the mean mechanical axis prior to osteotomy was varus 8.2°, and the mean mechanical axis following fixation was valgus 3.6°. On the radiographs the mean pre-operative mechanical axis was varus 7.3°, and the mean post-operative mechanical axis was valgus 2.1°. There was a positive correlation between the measured data taken under navigation and by radiographs (r > 0.3, p < 0.05). The mean correction angle was significantly more accurate in the navigation group (p < 0.002). The variability of the correction was significantly lower in the navigation group (2.3° vs 3.7°, p = 0,012). We conclude that navigation provides reliable real-time intra-operative information, may increase accuracy, and improves the precision of a closed-wedge HTO


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 67 - 73
1 Jun 2021
Lee G Wakelin E Randall A Plaskos C

Aims. Neither a surgeon’s intraoperative impression nor the parameters of computer navigation have been shown to be predictive of the outcomes following total knee arthroplasty (TKA). The aim of this study was to determine whether a surgeon, with robotic assistance, can predict the outcome as assessed using the Knee Injury and Osteoarthritis Outcome Score (KOOS) for pain (KPS), one year postoperatively, and establish what factors correlate with poor KOOS scores in a well-aligned and balanced TKA. Methods. A total of 134 consecutive patients who underwent TKA using a dynamic ligament tensioning robotic system with a tibia first resection technique and a cruciate sacrificing ultracongruent TKA system were enrolled into a prospective study. Each TKA was graded based on the final mediolateral ligament balance at 10° and 90° of flexion: 1) < 1 mm difference in the thickness of the tibial insert and that which was planned (n = 75); 2) < 1 mm difference (n = 26); 3) between 1 mm to 2 mm difference (n = 26); and 4) > 2 mm difference (n = 7). The mean one-year KPS score for each grade of TKA was compared and the likelihood of achieving an KPS score of > 90 was calculated. Finally, the factors associated with lower KPS despite achieving a high-grade TKA (grade A and B) were analyzed. Results. Patients with a grade of A or B TKA had significantly higher mean one-year KPS scores compared with those with C or D grades (p = 0.031). There was no difference in KPS scores in grade A or B TKAs, but 33% of these patients did not have a KPS score of > 90. While there was no correlation with age, sex, preoperative deformity, and preoperative KOOS and Patient-Reported Outcomes Measurement Information System (PROMIS) physical scores, patients with a KPS score of < 90, despite a grade A or B TKA, had lower PROMIS mental health scores compared with those with KPS scores of > 90 (54.1 vs 50.8; p = 0.043). Patients with grade A and B TKAs with KPS > 90 were significantly more likely to respond with “my expectations were too low”, and with “the knee is performing better than expected” compared with patients with these grades of TKA who had a KPS score of < 90 (40% vs 22%; p = 0.004). Conclusion. A TKA balanced with robotic assistance to within 1 mm of difference between the medial and lateral sides in both flexion and extension had a higher KPS score one year postoperatively. Despite accurate ligament balance information, a robotic system could not guarantee excellent pain relief. Patient expectations and mental status also significantly affected the perceived success of TKA. Cite this article: Bone Joint J 2021;103-B(6 Supple A):67–73


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 49 - 58
1 Jun 2020
Mullaji A

Aims. The aims of this study were to determine the effect of osteophyte excision on deformity correction and soft tissue gap balance in varus knees undergoing computer-assisted total knee arthroplasty (TKA). Methods. A total of 492 consecutive, cemented, cruciate-substituting TKAs performed for varus osteoarthritis were studied. After exposure and excision of both cruciates and menisci, it was noted from operative records the corrective interventions performed in each case. Knees in which no releases after the initial exposure, those which had only osteophyte excision, and those in which further interventions were performed were identified. From recorded navigation data, coronal and sagittal limb alignment, knee flexion range, and medial and lateral gap distances in maximum knee extension and 90° knee flexion with maximal varus and valgus stresses, were established, initially after exposure and excision of both cruciate ligaments, and then also at trialling. Knees were defined as ‘aligned’ if the hip-knee-ankle axis was between 177° and 180°, (0° to 3° varus) and ‘balanced’ if medial and lateral gaps in extension and at 90° flexion were within 2 mm of each other. Results. Of 50 knees (10%) with no soft tissue releases (other than cruciate ligaments), 90% were aligned, 81% were balanced, and 73% were aligned and balanced. In 288 knees (59%) only osteophyte excision was performed by subperiosteally releasing the deep medial collateral ligament. Of these, 98% were aligned, 80% were balanced, and 79% were aligned and balanced. In 154 knees (31%), additional procedures were performed (reduction osteotomy, posterior capsular release, and semimembranosus release). Of these, 89% were aligned, 68% were balanced, and 66% were aligned and balanced. The superficial medial collateral ligament was not released in any case. Conclusion. Two-thirds of all knees could be aligned and balanced with release of the cruciate ligaments alone and excision of osteophytes. Excision of osteophytes can be a useful step towards achieving deformity correction and gap balance without having to resort to soft tissue release in varus knees while maintaining classical coronal and sagittal alignment of components. Cite this article: Bone Joint J 2020;102-B(6 Supple A):49–58


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 81 - 86
1 Jun 2021
Mahfouz MR Abdel Fatah EE Johnson JM Komistek RD

Aims. The objective of this study is to assess the use of ultrasound (US) as a radiation-free imaging modality to reconstruct 3D anatomy of the knee for use in preoperative templating in knee arthroplasty. Methods. Using an US system, which is fitted with an electromagnetic (EM) tracker that is integrated into the US probe, allows 3D tracking of the probe, femur, and tibia. The raw US radiofrequency (RF) signals are acquired and, using real-time signal processing, bone boundaries are extracted. Bone boundaries and the tracking information are fused in a 3D point cloud for the femur and tibia. Using a statistical shaping model, the patient-specific surface is reconstructed by optimizing bone geometry to match the point clouds. An accuracy analysis was conducted for 17 cadavers by comparing the 3D US models with those created using CT. US scans from 15 users were compared in order to examine the effect of operator variability on the output. Results. The results revealed that the US bone models were accurate compared with the CT models (root mean squared error (RM)S: femur, 1.07 mm (SD 0.15); tibia, 1.02 mm (SD 0.13). Additionally, femoral landmarking proved to be accurate (transepicondylar axis: 1.07° (SD 0.65°); posterior condylar axis: 0.73° (SD 0.41°); distal condylar axis: 0.96° (SD 0.89°); medial anteroposterior (AP): 1.22 mm (SD 0.69); lateral AP: 1.21 mm (SD 1.02)). Tibial landmarking errors were slightly higher (posterior slope axis: 1.92° (SD 1.31°); and tubercle axis: 1.91° (SD 1.24°)). For implant sizing, 90% of the femora and 60% of the tibiae were sized correctly, while the remainder were only one size different from the required implant size. No difference was observed between moderate and skilled users. Conclusion. The 3D US bone models were proven to be closely matched compared with CT and suitable for preoperative planning. The 3D US is radiation-free and offers numerous clinical opportunities for bone visualization rapidly during clinic visits, to enable preoperative planning with implant sizing. There is potential to extend its application to 3D dynamic ligament balancing, and intraoperative registration for use with robots and navigation systems. Cite this article: Bone Joint J 2021;103-B(6 Supple A):81–86


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 194 - 199
1 Feb 2012
Hoffart H Langenstein E Vasak N

The aim of this prospective single-centre study was to assess the difference in clinical outcome between total knee replacement (TKR) using computerised navigation and that of conventional TKR. We hypothesised that navigation would give a better result at every stage within the first five years. A total of 195 patients (195 knees) with a mean age of 70.0 years (39 to 89) were allocated alternately into two treatment groups, which used either conventional instrumentation (group A, 97 knees) or a navigation system (group B, 98 knees). After five years, complete clinical scores were available for 121 patients (62%). A total of 18 patients were lost to follow-up. Compared with conventional surgery, navigated TKR resulted in a better mean Knee Society score (p = 0.008). The difference in mean Knee Society scores over time between the two groups was not constant (p = 0.006), which suggests that these groups differed in their response to surgery with time. No significant difference in the frequency of malalignment was seen between the two groups. In summary, computerised navigation resulted in a better functional outcome at five years than conventional techniques. Given the similarity in mechanical alignment between the two groups, rotational alignment may prove to be a better method of identifying differences in clinical outcome after navigated surgery


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 493 - 497
1 Apr 2013
Hino K Ishimaru M Iseki Y Watanabe S Onishi Y Miura H

There are several methods for evaluating stability of the joint during total knee replacement (TKR). Activities of daily living demand mechanical loading to the knee joint, not only in full extension, but also in mid-flexion. The purpose of this study was to compare the varus-valgus stability throughout flexion in knees treated with either cruciate-retaining or posterior-stabilised TKR, using an intra-operative navigation technique. A total of 34 knees underwent TKR with computer navigation, during which the investigator applied a maximum varus-valgus stress to the knee while steadily moving the leg from full extension to flexion both before and after prosthetic implantation. The femorotibial angle was measured simultaneously by the navigation system at every 10° throughout the range of movement. It was found that posterior-stabilised knees had more varus-valgus laxity than cruciate-retaining knees at all angles examined, and the differences were statistically significant at 10° (p = 0.0093), 20° (p = 0.0098) and 30° of flexion (p = 0.0252). Cite this article: Bone Joint J 2013;95-B:493–7


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1178 - 1182
1 Sep 2009
Hakki S Coleman S Saleh K Bilotta VJ Hakki A

The requirement for release of collateral ligaments to achieve a stable, balanced total knee replacement has been reported to arise in about 50% to 100% of procedures. This wide range reflects a lack of standardised quantitative indicators to determine the necessity for a release. Using recent advances in computerised navigation, we describe two navigational predictors which provide quantitative measures that can be used to identify the need for release. The first was the ability to restore the mechanical axis before any bone resection was performed and the second was the discrepancy in the measured medial and lateral joint spaces after the tibial osteotomy, but before any femoral resection. These predictors showed a significant association with the need for collateral ligament release (p < 0.001). The first predictor using the knee stress test in extension showed a sensitivity of 100% and a specificity of 98% and the second, the difference between medial and lateral gaps in millimetres, a sensitivity of 83% and a specificity of 95%. The use of the two navigational predictors meant that only ten of the 93 patients required collateral ligament release to achieve a stable, neutral knee


Bone & Joint Open
Vol. 5, Issue 11 | Pages 992 - 998
6 Nov 2024
Wignadasan W Magan A Kayani B Fontalis A Chambers A Rajput V Haddad FS

Aims

While residual fixed flexion deformity (FFD) in unicompartmental knee arthroplasty (UKA) has been associated with worse functional outcomes, limited evidence exists regarding FFD changes. The objective of this study was to quantify FFD changes in patients with medial unicompartmental knee arthritis undergoing UKA, and investigate any correlation with clinical outcomes.

Methods

This study included 136 patients undergoing robotic arm-assisted medial UKA between January 2018 and December 2022. The study included 75 males (55.1%) and 61 (44.9%) females, with a mean age of 67.1 years (45 to 90). Patients were divided into three study groups based on the degree of preoperative FFD: ≤ 5°, 5° to ≤ 10°, and > 10°. Intraoperative optical motion capture technology was used to assess pre- and postoperative FFD. Clinical FFD was measured pre- and postoperatively at six weeks and one year following surgery. Preoperative and one-year postoperative Oxford Knee Scores (OKS) were collected.


The Bone & Joint Journal
Vol. 106-B, Issue 8 | Pages 808 - 816
1 Aug 2024
Hall AJ Cullinan R Alozie G Chopra S Greig L Clarke J Riches PE Walmsley P Ohly NE Holloway N

Aims

Total knee arthroplasty (TKA) with a highly congruent condylar-stabilized (CS) articulation may be advantageous due to increased stability versus cruciate-retaining (CR) designs, while mitigating the limitations of a posterior-stabilized construct. The aim was to assess ten-year implant survival and functional outcomes of a cemented single-radius TKA with a CS insert, performed without posterior cruciate ligament sacrifice.

Methods

This retrospective cohort study included consecutive patients undergoing TKA at a specialist centre in the UK between November 2010 and December 2012. Data were collected using a bespoke electronic database and cross-referenced with national arthroplasty audit data, with variables including: preoperative characteristics, intraoperative factors, complications, and mortality status. Patient-reported outcome measures (PROMs) were collected by a specialist research team at ten years post-surgery. There were 536 TKAs, of which 308/536 (57.5%) were in female patients. The mean age was 69.0 years (95% CI 45.0 to 88.0), the mean BMI was 32.2 kg/m2 (95% CI 18.9 to 50.2), and 387/536 (72.2%) survived to ten years. There were four revisions (0.7%): two deep infections (requiring debridement and implant retention), one aseptic loosening, and one haemosiderosis.


Bone & Joint Open
Vol. 2, Issue 3 | Pages 191 - 197
1 Mar 2021
Kazarian GS Barrack RL Barrack TN Lawrie CM Nunley RM

Aims. The purpose of this study was to compare the radiological outcomes of manual versus robotic-assisted medial unicompartmental knee arthroplasty (UKA). Methods. Postoperative radiological outcomes from 86 consecutive robotic-assisted UKAs (RAUKA group) from a single academic centre were retrospectively reviewed and compared to 253 manual UKAs (MUKA group) drawn from a prior study at our institution. Femoral coronal and sagittal angles (FCA, FSA), tibial coronal and sagittal angles (TCA, TSA), and implant overhang were radiologically measured to identify outliers. Results. When assessing the accuracy of RAUKAs, 91.6% of all alignment measurements and 99.2% of all overhang measurements were within the target range. All alignment and overhang targets were simultaneously met in 68.6% of RAUKAs. When comparing radiological outcomes between the RAUKA and MUKA groups, statistically significant differences were identified for combined outliers in FCA (2.3% vs 12.6%; p = 0.006), FSA (17.4% vs 50.2%; p < 0.001), TCA (5.8% vs 41.5%; p < 0.001), and TSA (8.1% vs 18.6%; p = 0.023), as well as anterior (0.0% vs 4.7%; p = 0.042), posterior (1.2% vs 13.4%; p = 0.001), and medial (1.2% vs 14.2%; p < 0.001) overhang outliers. Conclusion. Robotic system navigation decreases alignment and overhang outliers compared to manual UKA. Given the association between component placement errors and revision in UKA, this strong significant improvement in accuracy may improve implant survival. Level of Evidence: III. Cite this article: Bone Jt Open 2021;2-3:191–197


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 906 - 910
1 Jul 2013
Lin S Chen C Fu Y Huang P Lu C Su J Chang J Huang H

Minimally invasive total knee replacement (MIS-TKR) has been reported to have better early recovery than conventional TKR. Quadriceps-sparing (QS) TKR is the least invasive MIS procedure, but it is technically demanding with higher reported rates of complications and outliers. This study was designed to compare the early clinical and radiological outcomes of TKR performed by an experienced surgeon using the QS approach with or without navigational assistance (NA), or using a mini-medial parapatellar (MP) approach. In all, 100 patients completed a minimum two-year follow-up: 30 in the NA-QS group, 35 in the QS group, and 35 in the MP group. There were no significant differences in clinical outcome in terms of ability to perform a straight-leg raise at 24 hours (p = 0.700), knee score (p = 0.952), functional score (p = 0.229) and range of movement (p = 0.732) among the groups. The number of outliers for all three radiological parameters of mechanical axis, frontal femoral component alignment and frontal tibial component alignment was significantly lower in the NA-QS group than in the QS group (p = 0.008), but no outlier was found in the MP group. . In conclusion, even after the surgeon completed a substantial number of cases before the commencement of this study, the supplementary intra-operative use of computer-assisted navigation with QS-TKR still gave inferior radiological results and longer operating time, with a similar outcome at two years when compared with a MP approach. Cite this article: Bone Joint J 2013;95-B:906–10


The Bone & Joint Journal
Vol. 106-B, Issue 10 | Pages 1059 - 1066
1 Oct 2024
Konishi T Hamai S Tsushima H Kawahara S Akasaki Y Yamate S Ayukawa S Nakashima Y

Aims

The Coronal Plane Alignment of the Knee (CPAK) classification has been developed to predict individual variations in inherent knee alignment. The impact of preoperative and postoperative CPAK classification phenotype on the postoperative clinical outcomes of total knee arthroplasty (TKA) remains elusive. This study aimed to examine the effect of postoperative CPAK classification phenotypes (I to IX), and their pre- to postoperative changes on patient-reported outcome measures (PROMs).

Methods

A questionnaire was administered to 340 patients (422 knees) who underwent primary TKA for osteoarthritis (OA) between September 2013 and June 2019. A total of 231 patients (284 knees) responded. The ­Knee Society Score 2011 (KSS 2011), Knee injury and Osteoarthritis Outcome Score-12 (KOOS-12), and Forgotten Joint Score-12 (FJS-12) were used to assess clinical outcomes. Using preoperative and postoperative anteroposterior full-leg radiographs, the arithmetic hip-knee-ankle angle (aHKA) and joint line obliquity (JLO) were calculated and classified based on the CPAK classification. To investigate the impact on PROMs, multivariable regression analyses using stepwise selection were conducted, considering factors such as age at surgery, time since surgery, BMI, sex, implant use, postoperative aHKA classification, JLO classification, and changes in aHKA and JLO classifications from preoperative to postoperative.


Bone & Joint Open
Vol. 4, Issue 4 | Pages 262 - 272
11 Apr 2023
Batailler C Naaim A Daxhelet J Lustig S Ollivier M Parratte S

Aims

The impact of a diaphyseal femoral deformity on knee alignment varies according to its severity and localization. The aims of this study were to determine a method of assessing the impact of diaphyseal femoral deformities on knee alignment for the varus knee, and to evaluate the reliability and the reproducibility of this method in a large cohort of osteoarthritic patients.

Methods

All patients who underwent a knee arthroplasty from 2019 to 2021 were included. Exclusion criteria were genu valgus, flexion contracture (> 5°), previous femoral osteotomy or fracture, total hip arthroplasty, and femoral rotational disorder. A total of 205 patients met the inclusion criteria. The mean age was 62.2 years (SD 8.4). The mean BMI was 33.1 kg/m2 (SD 5.5). The radiological measurements were performed twice by two independent reviewers, and included hip knee ankle (HKA) angle, mechanical medial distal femoral angle (mMDFA), anatomical medial distal femoral angle (aMDFA), femoral neck shaft angle (NSA), femoral bowing angle (FBow), the distance between the knee centre and the top of the FBow (DK), and the angle representing the FBow impact on the knee (C’KS angle).


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1047 - 1052
1 Aug 2017
Ikawa T Takemura S Kim M Takaoka K Minoda Y Kadoya Y

Aims. The aim of this study was to evaluate the effects of using a portable, accelerometer-based surgical navigation system (KneeAlign2) in total knee arthroplasty (TKA) on the alignment of the femoral component, and blood loss. Patients and Methods. A total of 241 consecutive patients with primary osteoarthritis of the knee were enrolled in this prospective, randomised controlled study. There were 207 women and 34 men. The mean age of the patients was 74.0 years (57 to 89). The KneeAlign2 system was used for distal femoral resection in 121 patients (KA2 group) and a conventional intramedullary femoral guide was used in 120 patients (IM group). Results. One patient (0.8%) in the KA2 group and 19 in the IM group had an alignment which was > 3° away from the neutral mechanical axis (p < 0.01). The mean deviation from neutral alignment was 1.01° (standard deviation (. sd). 1.0°) in the KA2 group and 1.93° (. sd. 1.7°) in the IM group (p < 0.01). Blood loss was significantly less in the KA2 group compared with the IM group (784 ml (. sd. 357) versus 1071 ml (. sd. 310), p < 0.001). Conclusion. The KneeAlign2 system provides a technically straightforward method for identifying the femoral head and performing an accurate distal femoral resection at TKA with significantly less blood loss compared with a conventional intramedullary guide. Cite this article: Bone Joint J 2017;99-B:1047–52


Bone & Joint Open
Vol. 5, Issue 11 | Pages 1013 - 1019
11 Nov 2024
Clark SC Pan X Saris DBF Taunton MJ Krych AJ Hevesi M

Aims

Distal femoral osteotomies (DFOs) are commonly used for the correction of valgus deformities and lateral compartment osteoarthritis. However, the impact of a DFO on subsequent total knee arthroplasty (TKA) function remains a subject of debate. Therefore, the purpose of this study was to determine the effect of a unilateral DFO on subsequent TKA function in patients with bilateral TKAs, using the contralateral knee as a self-matched control group.

Methods

The inclusion criteria consisted of patients who underwent simultaneous or staged bilateral TKA after prior unilateral DFO between 1972 and 2023. The type of osteotomy performed, osteotomy hardware fixation, implanted TKA components, and revision rates were recorded. Postoperative outcomes including the Forgotten Joint Score-12 (FJS-12), Tegner Activity Scale score, and subjective knee preference were also obtained at final follow-up.


Aims

The aim of this study was to investigate the distribution of phenotypes in Asian patients with end-stage osteoarthritis (OA) and assess whether the phenotype affected the clinical outcome and survival of mechanically aligned total knee arthroplasty (TKA). We also compared the survival of the group in which the phenotype unintentionally remained unchanged with those in which it was corrected to neutral.

Methods

The study involved 945 TKAs, which were performed in 641 patients with primary OA, between January 2000 and January 2009. These were classified into 12 phenotypes based on the combined assessment of four categories of the arithmetic hip-knee-ankle angle and three categories of actual joint line obliquity. The rates of survival were analyzed using Kaplan-Meier methods and the log-rank test. The Hospital for Special Surgery score and survival of each phenotype were compared with those of the reference phenotype with neutral alignment and a parallel joint line. We also compared long-term survival between the unchanged phenotype group and the corrected to neutral alignment-parallel joint line group in patients with Type IV-b (mild to moderate varus alignment-parallel joint line) phenotype.


Bone & Joint Open
Vol. 5, Issue 10 | Pages 937 - 943
22 Oct 2024
Gregor RH Hooper GJ Frampton C

Aims

The aim of this study was to determine whether obesity had a detrimental effect on the long-term performance and survival of medial unicompartmental knee arthroplasties (UKAs).

Methods

This study reviewed prospectively collected functional outcome scores and revision rates of all medial UKA patients with recorded BMI performed in Christchurch, New Zealand, from January 2011 to September 2021. Patient-reported outcome measures (PROMs) were the primary outcome of this study, with all-cause revision rate analyzed as a secondary outcome. PROMs were taken preoperatively, at six months, one year, five years, and ten years postoperatively. There were 873 patients who had functional scores recorded at five years and 164 patients had scores recorded at ten years. Further sub-group analysis was performed based on the patient’s BMI. Revision data were available through the New Zealand Joint Registry for 2,323 UKAs performed during this time period.


Bone & Joint Open
Vol. 5, Issue 2 | Pages 101 - 108
6 Feb 2024
Jang SJ Kunze KN Casey JC Steele JR Mayman DJ Jerabek SA Sculco PK Vigdorchik JM

Aims

Distal femoral resection in conventional total knee arthroplasty (TKA) utilizes an intramedullary guide to determine coronal alignment, commonly planned for 5° of valgus. However, a standard 5° resection angle may contribute to malalignment in patients with variability in the femoral anatomical and mechanical axis angle. The purpose of the study was to leverage deep learning (DL) to measure the femoral mechanical-anatomical axis angle (FMAA) in a heterogeneous cohort.

Methods

Patients with full-limb radiographs from the Osteoarthritis Initiative were included. A DL workflow was created to measure the FMAA and validated against human measurements. To reflect potential intramedullary guide placement during manual TKA, two different FMAAs were calculated either using a line approximating the entire diaphyseal shaft, and a line connecting the apex of the femoral intercondylar sulcus to the centre of the diaphysis. The proportion of FMAAs outside a range of 5.0° (SD 2.0°) was calculated for both definitions, and FMAA was compared using univariate analyses across sex, BMI, knee alignment, and femur length.


Bone & Joint Open
Vol. 5, Issue 8 | Pages 628 - 636
2 Aug 2024
Eachempati KK Parameswaran A Ponnala VK Sunil A Sheth NP

Aims

The aims of this study were: 1) to describe extended restricted kinematic alignment (E-rKA), a novel alignment strategy during robotic-assisted total knee arthroplasty (RA-TKA); 2) to compare residual medial compartment tightness following virtual surgical planning during RA-TKA using mechanical alignment (MA) and E-rKA, in the same set of osteoarthritic varus knees; 3) to assess the requirement of soft-tissue releases during RA-TKA using E-rKA; and 4) to compare the accuracy of surgical plan execution between knees managed with adjustments in component positioning alone, and those which require additional soft-tissue releases.

Methods

Patients who underwent RA-TKA between January and December 2022 for primary varus osteoarthritis were included. Safe boundaries for E-rKA were defined. Residual medial compartment tightness was compared following virtual surgical planning using E-rKA and MA, in the same set of knees. Soft-tissue releases were documented. Errors in postoperative alignment in relation to planned alignment were compared between patients who did (group A) and did not (group B) require soft-tissue releases.


Bone & Joint Open
Vol. 4, Issue 6 | Pages 432 - 441
5 Jun 2023
Kahlenberg CA Berube EE Xiang W Manzi JE Jahandar H Chalmers BP Cross MB Mayman DJ Wright TM Westrich GH Imhauser CW Sculco PK

Aims

Mid-level constraint designs for total knee arthroplasty (TKA) are intended to reduce coronal plane laxity. Our aims were to compare kinematics and ligament forces of the Zimmer Biomet Persona posterior-stabilized (PS) and mid-level designs in the coronal, sagittal, and axial planes under loads simulating clinical exams of the knee in a cadaver model.

Methods

We performed TKA on eight cadaveric knees and loaded them using a robotic manipulator. We tested both PS and mid-level designs under loads simulating clinical exams via applied varus and valgus moments, internal-external (IE) rotation moments, and anteroposterior forces at 0°, 30°, and 90° of flexion. We measured the resulting tibiofemoral angulations and translations. We also quantified the forces carried by the medial and lateral collateral ligaments (MCL/LCL) via serial sectioning of these structures and use of the principle of superposition.


Bone & Joint Open
Vol. 5, Issue 2 | Pages 109 - 116
8 Feb 2024
Corban LE van de Graaf VA Chen DB Wood JA Diwan AD MacDessi SJ

Aims

While mechanical alignment (MA) is the traditional technique in total knee arthroplasty (TKA), its potential for altering constitutional alignment remains poorly understood. This study aimed to quantify unintentional changes to constitutional coronal alignment and joint line obliquity (JLO) resulting from MA.

Methods

A retrospective cohort study was undertaken of 700 primary MA TKAs (643 patients) performed between 2014 and 2017. Lateral distal femoral and medial proximal tibial angles were measured pre- and postoperatively to calculate the arithmetic hip-knee-ankle angle (aHKA), JLO, and Coronal Plane Alignment of the Knee (CPAK) phenotypes. The primary outcome was the magnitude and direction of aHKA, JLO, and CPAK alterations.


Bone & Joint Open
Vol. 4, Issue 11 | Pages 889 - 898
23 Nov 2023
Clement ND Fraser E Gilmour A Doonan J MacLean A Jones BG Blyth MJG

Aims

To perform an incremental cost-utility analysis and assess the impact of differential costs and case volume on the cost-effectiveness of robotic arm-assisted unicompartmental knee arthroplasty (rUKA) compared to manual (mUKA).

Methods

This was a five-year follow-up study of patients who were randomized to rUKA (n = 64) or mUKA (n = 65). Patients completed the EuroQol five-dimension questionnaire (EQ-5D) preoperatively, and at three months and one, two, and five years postoperatively, which was used to calculate quality-adjusted life years (QALYs) gained. Costs for the primary and additional surgery and healthcare costs were calculated.


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1231 - 1239
1 Nov 2024
Tzanetis P Fluit R de Souza K Robertson S Koopman B Verdonschot N

Aims

The surgical target for optimal implant positioning in robotic-assisted total knee arthroplasty remains the subject of ongoing discussion. One of the proposed targets is to recreate the knee’s functional behaviour as per its pre-diseased state. The aim of this study was to optimize implant positioning, starting from mechanical alignment (MA), toward restoring the pre-diseased status, including ligament strain and kinematic patterns, in a patient population.

Methods

We used an active appearance model-based approach to segment the preoperative CT of 21 osteoarthritic patients, which identified the osteophyte-free surfaces and estimated cartilage from the segmented bones; these geometries were used to construct patient-specific musculoskeletal models of the pre-diseased knee. Subsequently, implantations were simulated using the MA method, and a previously developed optimization technique was employed to find the optimal implant position that minimized the root mean square deviation between pre-diseased and postoperative ligament strains and kinematics.


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 450 - 459
1 May 2024
Clement ND Galloway S Baron J Smith K Weir DJ Deehan DJ

Aims

The aim was to assess whether robotic-assisted total knee arthroplasty (rTKA) had greater knee-specific outcomes, improved fulfilment of expectations, health-related quality of life (HRQoL), and patient satisfaction when compared with manual TKA (mTKA).

Methods

A randomized controlled trial was undertaken (May 2019 to December 2021), and patients were allocated to either mTKA or rTKA. A total of 100 patients were randomized, 50 to each group, of whom 43 rTKA and 38 mTKA patients were available for review at 12 months following surgery. There were no statistically significant preoperative differences between the groups. The minimal clinically important difference in the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) pain score was defined as 7.5 points.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 640 - 646
1 May 2017
Matsumoto T Takayama K Ishida K Hayashi S Hashimoto S Kuroda R

Aims. The aim of this study was to compare the post-operative radiographic and clinical outcomes between kinematically and mechanically aligned total knee arthroplasties (TKAs). . Patients and Methods. A total of 60 TKAs (30 kinematically and 30 mechanically aligned) were performed in 60 patients with varus osteoarthritis of the knee using a navigation system. The angles of orientation of the joint line in relation to the floor, the conventional and true mechanical axis (tMA) (the line from the centre of the hip to the lowest point of the calcaneus) were compared, one year post-operatively, on single-leg and double-leg standing long leg radiographs between the groups. The range of movement and 2011 Knee Society Scores were also compared between the groups at that time. Results. The angles of orientation of the joint line in the kinematic group changed from slight varus on double-leg standing to slight valgus with single-leg standing. The mechanical axes in the kinematic group passed through a neutral position of the knee in the true condition when the calcaneus was considered. The post-operative angles of flexion and functional activity scores were significantly better in the kinematic than in the mechanical group (p < 0.003 and 0.03, respectively). Conclusion. A kinematically aligned TKA results in a joint line which has a more parallel orientation in relation to the floor during single- and double-leg standing, and more neutral weight-bearing in tMA than a mechanically aligned TKA. Cite this article: Bone Joint J 2017;99-B:640–6


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 604 - 612
1 May 2022
MacDessi SJ Wood JA Diwan A Harris IA

Aims

Intraoperative pressure sensors allow surgeons to quantify soft-tissue balance during total knee arthroplasty (TKA). The aim of this study was to determine whether using sensors to achieve soft-tissue balance was more effective than manual balancing in improving outcomes in TKA.

Methods

A multicentre randomized trial compared the outcomes of sensor balancing (SB) with manual balancing (MB) in 250 patients (285 TKAs). The primary outcome measure was the mean difference in the four Knee injury and Osteoarthritis Outcome Score subscales (ΔKOOS4) in the two groups, comparing the preoperative and two-year scores. Secondary outcomes included intraoperative balance data, additional patient-reported outcome measures (PROMs), and functional measures.


Aims

Nearly 99,000 total knee arthroplasties (TKAs) are performed in UK annually. Despite plenty of research, the satisfaction rate of this surgery is around 80%. One of the important intraoperative factors affecting the outcome is alignment. The relationship between joint obliquity and functional outcomes is not well understood. Therefore, a study is required to investigate and compare the effects of two types of alignment (mechanical and kinematic) on functional outcomes and range of motion.

Methods

The aim of the study is to compare navigated kinematically aligned TKAs (KA TKAs) with navigated mechanically aligned TKA (MA TKA) in terms of function and ROM. We aim to recruit a total of 96 patients in the trial. The patients will be recruited from clinics of various consultants working in the trust after screening them for eligibility criteria and obtaining their informed consent to participate in this study. Randomization will be done prior to surgery by a software. The primary outcome measure will be the Knee injury and Osteoarthritis Outcome Score The secondary outcome measures include Oxford Knee Score, ROM, EuroQol five-dimension questionnaire, EuroQol visual analogue scale, 12-Item Short-Form Health Survey (SF-12), and Forgotten Joint Score. The scores will be calculated preoperatively and then at six weeks, six months, and one year after surgery. The scores will undergo a statistical analysis.


Bone & Joint Open
Vol. 3, Issue 7 | Pages 589 - 595
1 Jul 2022
Joo PY Chen AF Richards J Law TY Taylor K Marchand K Clark G Collopy D Marchand RC Roche M Mont MA Malkani AL

Aims

The aim of this study was to report patient and clinical outcomes following robotic-assisted total knee arthroplasty (RA-TKA) at multiple institutions with a minimum two-year follow-up.

Methods

This was a multicentre registry study from October 2016 to June 2021 that included 861 primary RA-TKA patients who completed at least one pre- and postoperative patient-reported outcome measure (PROM) questionnaire, including Forgotten Joint Score (FJS), Knee Injury and Osteoarthritis Outcomes Score for Joint Replacement (KOOS JR), and pain out of 100 points. The mean age was 67 years (35 to 86), 452 were male (53%), mean BMI was 31.5 kg/m2 (19 to 58), and 553 (64%) cemented and 308 (36%) cementless implants.


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 498 - 502
1 Apr 2015
Deep K Eachempati KK Apsingi S

The restoration of knee alignment is an important goal during total knee arthroplasty (TKA). In the past surgeons aimed to restore neutral limb alignment during surgery. However, previous studies have demonstrated alignment to be dynamic, varying depending on the position of the limb and the degree of weight-bearing, and between patients. We used a validated computer navigation system to measure the femorotibial mechanical angle (FTMA) in 264 knees in 77 male and 55 female healthy volunteers aged 18 to 35 years (mean 26.2). We found the mean supine alignment to be a varus angle of 1.2° (standard deviation (. sd. ) 4), with few patients having neutral alignment. FTMA differs significantly between males and females (with a mean varus of 1.7° (. sd. 4) and 0.4° (. sd. 3.9), respectively; p = 0.008). It changes significantly with posture, the knee hyperextending by a mean of 5.6°, and coronal plane alignment becoming more varus by 2.2° (. sd. 3.6) on standing compared with supine. Knee alignment is different in different individuals and is dynamic in nature, changing with different postures. This may have implications for the assessment of alignment in TKA, which is achieved in non-weight-bearing conditions and which may not represent the situation observed during weight-bearing. Cite this article: Bone Joint J 2015; 97-B:498–502


The Bone & Joint Journal
Vol. 104-B, Issue 6 | Pages 672 - 679
1 Jun 2022
Tay ML Young SW Frampton CM Hooper GJ

Aims

Unicompartmental knee arthroplasty (UKA) has a higher risk of revision than total knee arthroplasty (TKA), particularly for younger patients. The outcome of knee arthroplasty is typically defined as implant survival or revision incidence after a defined number of years. This can be difficult for patients to conceptualize. We aimed to calculate the ‘lifetime risk’ of revision for UKA as a more meaningful estimate of risk projection over a patient’s remaining lifetime, and to compare this to TKA.

Methods

Incidence of revision and mortality for all primary UKAs performed from 1999 to 2019 (n = 13,481) was obtained from the New Zealand Joint Registry (NZJR). Lifetime risk of revision was calculated for patients and stratified by age, sex, and American Society of Anesthesiologists (ASA) grade.


Bone & Joint Open
Vol. 3, Issue 5 | Pages 390 - 397
1 May 2022
Hiranaka T Suda Y Saitoh A Tanaka A Arimoto A Koide M Fujishiro T Okamoto K

The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA, the tibial cutting surface is decided parallel to the femoral cutting surface (or trial component) with in-line traction. These approaches are categorized as unrestricted KA because there is no consideration of leg alignment or component orientation. Restricted KA is an approach where the periarthritic joint surface is replicated within a safe range, due to concerns about extreme alignments that have been considered ‘alignment outliers’ in the neutral mechanical alignment approach. More recently, functional alignment and inverse kinematic alignment have been advocated, where bone cuts are made following intraoperative planning, using intraoperative measurements acquired with computer assistance to fulfill good coordination of soft-tissue balance and alignment. The KA-TKA approach aims to restore the patients’ own harmony of three knee elements (morphology, soft-tissue balance, and alignment) and eventually the patients’ own kinematics. The respective approaches start from different points corresponding to one of the elements, yet each aim for the same goal, although the existing implants and techniques have not yet perfectly fulfilled that goal.


Bone & Joint Open
Vol. 2, Issue 6 | Pages 397 - 404
1 Jun 2021
Begum FA Kayani B Magan AA Chang JS Haddad FS

Limb alignment in total knee arthroplasty (TKA) influences periarticular soft-tissue tension, biomechanics through knee flexion, and implant survival. Despite this, there is no uniform consensus on the optimal alignment technique for TKA. Neutral mechanical alignment facilitates knee flexion and symmetrical component wear but forces the limb into an unnatural position that alters native knee kinematics through the arc of knee flexion. Kinematic alignment aims to restore native limb alignment, but the safe ranges with this technique remain uncertain and the effects of this alignment technique on component survivorship remain unknown. Anatomical alignment aims to restore predisease limb alignment and knee geometry, but existing studies using this technique are based on cadaveric specimens or clinical trials with limited follow-up times. Functional alignment aims to restore the native plane and obliquity of the joint by manipulating implant positioning while limiting soft tissue releases, but the results of high-quality studies with long-term outcomes are still awaited. The drawbacks of existing studies on alignment include the use of surgical techniques with limited accuracy and reproducibility of achieving the planned alignment, poor correlation of intraoperative data to long-term functional outcomes and implant survivorship, and a paucity of studies on the safe ranges of limb alignment. Further studies on alignment in TKA should use surgical adjuncts (e.g. robotic technology) to help execute the planned alignment with improved accuracy, include intraoperative assessments of knee biomechanics and periarticular soft-tissue tension, and correlate alignment to long-term functional outcomes and survivorship.


Bone & Joint Research
Vol. 2, Issue 11 | Pages 233 - 237
1 Nov 2013
Russell DF Deakin AH Fogg QA Picard F

Objectives. We performed in vitro validation of a non-invasive skin-mounted system that could allow quantification of anteroposterior (AP) laxity in the outpatient setting. Methods. A total of 12 cadaveric lower limbs were tested with a commercial image-free navigation system using trackers secured by bone screws. We then tested a non-invasive fabric-strap system. The lower limb was secured at 10° intervals from 0° to 60° of knee flexion and 100 N of force was applied perpendicular to the tibia. Acceptable coefficient of repeatability (CR) and limits of agreement (LOA) of 3 mm were set based on diagnostic criteria for anterior cruciate ligament (ACL) insufficiency. Results. Reliability and precision within the individual invasive and non-invasive systems was acceptable throughout the range of flexion tested (intra-class correlation coefficient 0.88, CR 1.6 mm). Agreement between the two systems was acceptable measuring AP laxity between full extension and 40° knee flexion (LOA 2.9 mm). Beyond 40° of flexion, agreement between the systems was unacceptable (LOA > 3 mm). Conclusions. These results indicate that from full knee extension to 40° flexion, non-invasive navigation-based quantification of AP tibial translation is as accurate as the standard validated commercial system, particularly in the clinically and functionally important range of 20° to 30° knee flexion. This could be useful in diagnosis and post-operative evaluation of ACL pathology. Cite this article: Bone Joint Res 2013;2:233–7


Aims

The aim of this study was to compare any differences in the primary outcome (biphasic flexion knee moment during gait) of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) at one year post-surgery.

Methods

A total of 76 patients (34 bi-UKA and 42 TKA patients) were analyzed in a prospective, single-centre, randomized controlled trial. Flat ground shod gait analysis was performed preoperatively and one year postoperatively. Knee flexion moment was calculated from motion capture markers and force plates. The same setup determined proprioception outcomes during a joint position sense test and one-leg standing. Surgery allocation, surgeon, and secondary outcomes were analyzed for prediction of the primary outcome from a binary regression model.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 642 - 647
1 May 2012
Mullaji A Lingaraju AP Shetty GM

We retrospectively reviewed the records of 1150 computer-assisted total knee replacements and analysed the clinical and radiological outcomes of 45 knees that had arthritis with a pre-operative recurvatum deformity. The mean pre-operative hyperextension deformity of 11° (6° to 15°), as measured by navigation at the start of the operation, improved to a mean flexion deformity of 3.1° (0° to 7°) post-operatively. A total of 41 knees (91%) were managed using inserts ≤ 12.5 mm thick, and none had mediolateral laxity > 2 mm from a mechanical axis of 0° at the end of the surgery. At a mean follow-up of 26.4 months (13 to 48) there was significant improvement in the mean Knee Society, Oxford knee and Western Ontario and McMaster Universities Osteoarthritis Index scores compared with the pre-operative values. The mean knee flexion improved from 105° (80° to 125°) pre-operatively to 131° (120° to 145°), and none of the limbs had recurrent recurvatum. These early results show that total knee replacement using computer navigation and an algorithmic approach for arthritic knees with a recurvatum deformity can give excellent radiological and functional outcomes without recurrent deformity


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 479 - 483
1 Apr 2011
Chang C Lai K Yang C Lan S

Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design. At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%). In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1039 - 1044
1 Aug 2008
Lützner J Krummenauer F Wolf C Günther K Kirschner S

After obtaining informed consent, 80 patients were randomised to undergo a navigated or conventional total knee replacement. All received a cemented, unconstrained, cruciate-retaining implant with a rotating platform. Full-length standing and lateral radiographs and CT scans of the hip, knee and ankle joint were carried out five to seven days after operation. No notable differences were found between computer-assisted navigation and conventional implantation techniques as regards the rotational alignment of the femoral or tibial components. Although the deviation from the transepicondylar axis was relatively low, there was a considerable range of deviation for the tibial rotational alignment. There was no statistically significant difference regarding the occurrence pattern of outliers in mechanical malalignment but the number of outliers was reduced in the navigated group


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1561 - 1570
1 Oct 2021
Blyth MJG Banger MS Doonan J Jones BG MacLean AD Rowe PJ

Aims

The aim of this study was to compare the clinical outcomes of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) during the first six weeks and at one year postoperatively.

Methods

A per protocol analysis of 76 patients, 43 of whom underwent TKA and 34 of whom underwent bi-UKA, was performed from a prospective, single-centre, randomized controlled trial. Diaries kept by the patients recorded pain, function, and the use of analgesics daily throughout the first week and weekly between the second and sixth weeks. Patient-reported outcome measures (PROMs) were compared preoperatively, and at three months and one year postoperatively. Data were also compared longitudinally and a subgroup analysis was conducted, stratified by preoperative PROM status.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1441 - 1445
1 Nov 2008
Hadjicostas PT Soucacos PN Thielemann FW

We describe the mid-term results of a prospective study of total knee replacement in severe valgus knees using an osteotomy of the lateral femoral condyle and computer navigation. There were 15 knees with a mean valgus deformity of 21° (17° to 27°) and a mean follow-up of 28 months (24 to 60). A cemented, non-constrained fixed bearing, posterior-cruciate-retaining knee prosthesis of the same design was used in all cases (Columbus-B. Braun; Aesculap, Tuttlingen, Germany). All the knees were corrected to a mean of 0.5° of valgus (0° to 2°). Flexion of the knee had been limited to a mean of 85° (75° to 110°) pre-operatively and improved to a mean of 105° (90° to 130°) after operation. The mean Knee Society score improved from 37 (30 to 44) to 90 points (86 to 94). Osteotomy of the lateral femoral condyle combined with computer-assisted surgery gave an excellent mid-term outcome in patients undergoing total knee replacement in the presence of severe valgus deformity


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 2 | Pages 198 - 202
1 Feb 2006
Kalairajah Y Cossey AJ Verrall GM Ludbrook G Spriggins AJ

We undertook a prospective, randomised study using a non-invasive transcranial Doppler device to evaluate cranial embolisation in computer-assisted navigated total knee arthroplasty (n = 14) and compared this with a standard conventional surgical technique using intramedullary alignment guides (n = 10). All patients were selected randomly without the knowledge of the patient, anaesthetists (before the onset of the procedure) and ward staff. The operations were performed by a single surgeon at one hospital using a uniform surgical approach, instrumentation, technique and release sequence. The only variable in the two groups of patients was the use of single tracker pins of the imageless navigation system in the tibia and femur of the navigated group and intramedullary femoral and tibial alignment jigs in the non-navigated group. Acetabular Doppler signals were obtained in 14 patients in the computer-assisted group and nine (90%) in the conventional group, in whom high-intensity signals were detected in seven computer-assisted patients (50%) and in all of the non-navigated patients. In the computer-assisted group no patient had more than two detectable emboli, with a mean of 0.64 (SD 0.74). In the non-navigated group the number of emboli ranged from one to 43 and six patients had more than two detectable emboli, with a mean of 10.7 (. sd. 13.5). The difference between the two groups was highly significant using the Wilcoxon non-parametric test (p = 0.0003). Our findings show that computer-assisted total knee arthroplasty, when compared with conventional jig-based surgery, significantly reduces systemic emboli as detected by transcranial Doppler ultrasonography


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 329 - 337
1 Feb 2021
MacDessi SJ Griffiths-Jones W Harris IA Bellemans J Chen DB

Aims

A comprehensive classification for coronal lower limb alignment with predictive capabilities for knee balance would be beneficial in total knee arthroplasty (TKA). This paper describes the Coronal Plane Alignment of the Knee (CPAK) classification and examines its utility in preoperative soft tissue balance prediction, comparing kinematic alignment (KA) to mechanical alignment (MA).

Methods

A radiological analysis of 500 healthy and 500 osteoarthritic (OA) knees was used to assess the applicability of the CPAK classification. CPAK comprises nine phenotypes based on the arithmetic HKA (aHKA) that estimates constitutional limb alignment and joint line obliquity (JLO). Intraoperative balance was compared within each phenotype in a cohort of 138 computer-assisted TKAs randomized to KA or MA. Primary outcomes included descriptive analyses of healthy and OA groups per CPAK type, and comparison of balance at 10° of flexion within each type. Secondary outcomes assessed balance at 45° and 90° and bone recuts required to achieve final knee balance within each CPAK type.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 59 - 66
1 Jun 2021
Abhari S Hsing TM Malkani MM Smith AF Smith LS Mont MA Malkani AL

Aims

Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment.

Methods

A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1480 - 1482
1 Nov 2005
Kalairajah Y Simpson D Cossey AJ Verrall GM Spriggins AJ

We carried out a prospective randomised study to evaluate the blood loss in 60 patients having a total knee arthroplasty and divided randomly into two equal groups, one having a computer-assisted procedure and the other a standard operation. The surgery was carried out by a single surgeon at one institution using a uniform approach. The only variable in the groups was the use of intramedullary femoral and tibial alignment jigs in the standard group and single tracker pins of the imageless navigation system in the tibia and femur in the navigated group. The mean drainage of blood was 1351 ml (715 to 2890; 95% confidence interval (CI) 1183 to 1518) in the computer-aided group and 1747 ml (1100 to 3030; CI 1581 to 1912) in the conventional group. This difference was statistically significant (p = 0.001). The mean calculated loss of haemoglobin was 36 g/dl in the navigated group versus 53 g/dl in the conventional group; this was significant at p < 0.00001. There was a highly significant reduction in blood drainage and the calculated Hb loss between the computer-assisted and the conventional techniques. This allows the ordering of less blood before the operation, reduces risks at transfusion and gives financial saving. Computer-assisted surgery may also be useful for patients in whom blood products are not acceptable


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 113 - 122
1 Jan 2021
Kayani B Tahmassebi J Ayuob A Konan S Oussedik S Haddad FS

Aims

The primary aim of this study was to compare the postoperative systemic inflammatory response in conventional jig-based total knee arthroplasty (conventional TKA) versus robotic-arm assisted total knee arthroplasty (robotic TKA). Secondary aims were to compare the macroscopic soft tissue injury, femoral and tibial bone trauma, localized thermal response, and the accuracy of component positioning between the two treatment groups.

Methods

This prospective randomized controlled trial included 30 patients with osteoarthritis of the knee undergoing conventional TKA versus robotic TKA. Predefined serum markers of inflammation and localized knee temperature were collected preoperatively and postoperatively at six hours, day 1, day 2, day 7, and day 28 following TKA. Blinded observers used the Macroscopic Soft Tissue Injury (MASTI) classification system to grade intraoperative periarticular soft tissue injury and bone trauma. Plain radiographs were used to assess the accuracy of achieving the planned postioning of the components in both groups.


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 610 - 618
1 Apr 2021
Batailler C Bordes M Lording T Nigues A Servien E Calliess T Lustig S

Aims

Ideal component sizing may be difficult to achieve in unicompartmental knee arthroplasty (UKA). Anatomical variants, incremental implant size, and a reduced surgical exposure may lead to over- or under-sizing of the components. The purpose of this study was to compare the accuracy of UKA sizing with robotic-assisted techniques versus a conventional surgical technique.

Methods

Three groups of 93 medial UKAs were assessed. The first group was performed by a conventional technique, the second group with an image-free robotic-assisted system (Image-Free group), and the last group with an image-based robotic arm-assisted system, using a preoperative CT scan (Image-Based group). There were no demographic differences between groups. We compared six parameters on postoperative radiographs to assess UKA sizing. Incorrect sizing was defined by an over- or under-sizing greater than 3 mm.


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 117 - 124
1 Jan 2020
MacDessi SJ Griffiths-Jones W Chen DB Griffiths-Jones S Wood JA Diwan AD Harris IA

Aims

It is unknown whether kinematic alignment (KA) objectively improves knee balance in total knee arthroplasty (TKA), despite this being the biomechanical rationale for its use. This study aimed to determine whether restoring the constitutional alignment using a restrictive KA protocol resulted in better quantitative knee balance than mechanical alignment (MA).

Methods

We conducted a randomized superiority trial comparing patients undergoing TKA assigned to KA within a restrictive safe zone or MA. Optimal knee balance was defined as an intercompartmental pressure difference (ICPD) of 15 psi or less using a pressure sensor. The primary endpoint was the mean intraoperative ICPD at 10° of flexion prior to knee balancing. Secondary outcomes included balance at 45° and 90°, requirements for balancing procedures, and presence of tibiofemoral lift-off.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 43 - 48
1 Jun 2020
D’Lima DP Huang P Suryanarayan P Rosen A D’Lima DD

Aims

The extensive variation in axial rotation of tibial components can lead to coronal plane malalignment. We analyzed the change in coronal alignment induced by tray malrotation.

Methods

We constructed a computer model of knee arthroplasty and used a virtual cutting guide to cut the tibia at 90° to the coronal plane. The virtual guide was rotated axially (15° medial to 15° lateral) and with posterior slopes (0° to 7°). To assess the effect of axial malrotation, we measured the coronal plane alignment of a tibial tray that was axially rotated (25° internal to 15° external), as viewed on a standard anteroposterior (AP) radiograph.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 338 - 346
1 Feb 2021
Khow YZ Liow MHL Lee M Chen JY Lo NN Yeo SJ

Aims

This study aimed to identify the tibial component and femoral component coronal angles (TCCAs and FCCAs), which concomitantly are associated with the best outcomes and survivorship in a cohort of fixed-bearing, cemented, medial unicompartmental knee arthroplasties (UKAs). We also investigated the potential two-way interactions between the TCCA and FCCA.

Methods

Prospectively collected registry data involving 264 UKAs from a single institution were analyzed. The TCCAs and FCCAs were measured on postoperative radiographs and absolute angles were analyzed. Clinical assessment at six months, two years, and ten years was undertaken using the Knee Society Knee score (KSKS) and Knee Society Function score (KSFS), the Oxford Knee Score (OKS), the 36-Item Short-Form Health Survey questionnaire (SF-36), and range of motion (ROM). Fulfilment of expectations and satisfaction was also recorded. Implant survivorship was reviewed at a mean follow-up of 14 years (12 to 16). Multivariate regression models included covariates, TCCA, FCCA, and two-way interactions between them. Partial residual graphs were generated to identify angles associated with the best outcomes. Kaplan-Meier analysis was used to compare implant survivorship between groups.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 319 - 328
1 Mar 2020
St Mart J de Steiger RN Cuthbert A Donnelly W

Aim

There has been a significant reduction in unicompartmental knee arthroplasty (UKA) procedures recorded in Australia. This follows several national joint registry studies documenting high UKA revision rates when compared to total knee arthroplasty (TKA). With the recent introduction of robotically assisted UKA procedures, it is hoped that outcomes improve. This study examines the cumulative revision rate of UKA procedures implanted with a newly introduced robotic system and compares the results to one of the best performing non-robotically assisted UKA prostheses, as well as all other non-robotically assisted UKA procedures.

Methods

Data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR) for all UKA procedures performed for osteoarthritis (OA) between 2015 and 2018 were analyzed. Procedures using the Restoris MCK UKA prosthesis implanted using the Mako Robotic-Arm Assisted System were compared to non-robotically assisted Zimmer Unicompartmental High Flex Knee System (ZUK) UKA, a commonly used UKA with previously reported good outcomes and to all other non-robotically assisted UKA procedures using Cox proportional hazard ratios (HRs) and Kaplan-Meier estimates of survivorship.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 36 - 42
1 Jun 2020
Nishitani K Kuriyama S Nakamura S Umatani N Ito H Matsuda S

Aims

This study aimed to evaluate the association between the sagittal alignment of the femoral component in total knee arthroplasty (TKA) and new Knee Society Score (2011KSS), under the hypothesis that outliers such as the excessive extended or flexed femoral component were related to worse clinical outcomes.

Methods

A group of 156 knees (134 F:22 M) in 133 patients with a mean age 75.8 years (SD 6.4) who underwent TKA with the cruciate-substituting Bi-Surface Knee prosthesis were retrospectively enrolled. On lateral radiographs, γ angle (the angle between the distal femoral axis and the line perpendicular to the distal rear surface of the femoral component) was measured, and the patients were divided into four groups according to the γ angle. The 2011KSSs among groups were compared using the Kruskal-Wallis test. A secondary regression analysis was used to investigate the association between the 2011KSS and γ angle.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 857 - 862
1 Jul 2014
Abdel MP Oussedik S Parratte S Lustig S Haddad FS

Substantial healthcare resources have been devoted to computer navigation and patient-specific instrumentation systems that improve the reproducibility with which neutral mechanical alignment can be achieved following total knee replacement (TKR). This choice of alignment is based on the long-held tenet that the alignment of the limb post-operatively should be within 3° of a neutral mechanical axis. Several recent studies have demonstrated no significant difference in survivorship when comparing well aligned versus malaligned TKRs. Our aim was to review the anatomical alignment of the knee, the historical and contemporary data on a neutral mechanical axis in TKR, and the feasibility of kinematically-aligned TKRs. . Review of the literature suggests that a neutral mechanical axis remains the optimal guide to alignment. Cite this article: Bone Joint J 2014;96-B:857–62


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1457 - 1461
1 Nov 2012
Krishnan SP Dawood A Richards R Henckel J Hart AJ

Improvements in the surgical technique of total knee replacement (TKR) are continually being sought. There has recently been interest in three-dimensional (3D) pre-operative planning using magnetic resonance imaging (MRI) and CT. The 3D images are increasingly used for the production of patient-specific models, surgical guides and custom-made implants for TKR. The users of patient-specific instrumentation (PSI) claim that they allow the optimum balance of technology and conventional surgery by reducing the complexity of conventional alignment and sizing tools. In this way the advantages of accuracy and precision claimed by computer navigation techniques are achieved without the disadvantages of additional intra-operative inventory, new skills or surgical time. This review describes the terminology used in this area and debates the advantages and disadvantages of PSI


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 838 - 847
1 Jul 2019
Robinson PG Clement ND Hamilton D Blyth MJG Haddad FS Patton JT

Aims

Robotic-assisted unicompartmental knee arthroplasty (UKA) promises accurate implant placement with the potential of improved survival and functional outcomes. The aim of this study was to present the current evidence for robotic-assisted UKA and describe the outcome in terms of implant positioning, range of movement (ROM), function and survival, and the types of robot and implants that are currently used.

Materials and Methods

A search of PubMed and Medline was performed in October 2018 in line with the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “knee”, and “surgery”. The criteria for inclusion was any study describing the use of robotic UKA and reporting implant positioning, ROM, function, and survival for clinical, cadaveric, or dry bone studies.


Aims

Patient-specific instrumentation of total knee arthroplasty (TKA) is a technique permitting the targeting of individual kinematic alignment, but deviation from a neutral mechanical axis may have implications on implant fixation and therefore survivorship. The primary objective of this randomized controlled study was to compare the fixation of tibial components implanted with patient-specific instrumentation targeting kinematic alignment (KA+PSI) versus components placed using computer-assisted surgery targeting neutral mechanical alignment (MA+CAS). Tibial component migration measured by radiostereometric analysis was the primary outcome measure (compared longitudinally between groups and to published acceptable thresholds). Secondary outcome measures were inducible displacement after one year and patient-reported outcome measures (PROMS) over two years. The secondary objective was to assess the relationship between alignment and both tibial component migration and inducible displacement.

Patients and Methods

A total of 47 patients due to undergo TKA were randomized to KA+PSI (n = 24) or MA+CAS (n = 23). In the KA+PSI group, there were 16 female and eight male patients with a mean age of 64 years (sd 8). In the MA+CAS group, there were 17 female and six male patients with a mean age of 63 years (sd 7). Surgery was performed using cemented, cruciate-retaining Triathlon total knees with patellar resurfacing, and patients were followed up for two years. The effect of alignment on tibial component migration and inducible displacement was analyzed irrespective of study group.


The Bone & Joint Journal
Vol. 101-B, Issue 6 | Pages 660 - 666
1 Jun 2019
Chalmers BP Limberg AK Athey AG Perry KI Pagnano MW Abdel MP

Aims

There is little literature about total knee arthroplasty (TKA) after distal femoral osteotomy (DFO). Consequently, the purpose of this study was to analyze the outcomes of TKA after DFO, with particular emphasis on: survivorship free from aseptic loosening, revision, or any re-operation; complications; radiological results; and clinical outcome.

Patients and Methods

We retrospectively reviewed 29 patients (17 women, 12 men) from our total joint registry who had undergone 31 cemented TKAs after a DFO between 2000 and 2012. Their mean age at TKA was 51 years (22 to 76) and their mean body mass index 32 kg/m2 (20 to 45). The mean time between DFO and TKA was ten years (2 to 20). The mean follow-up from TKA was ten years (2 to 16). The prostheses were posterior-stabilized in 77%, varus-valgus constraint (VVC) in 13%, and cruciate-retaining in 10%. While no patient had metaphyseal fixation (e.g. cones or sleeves), 16% needed a femoral stem.


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 331 - 339
1 Mar 2019
McEwen P Balendra G Doma K

Aims

The results of kinematic total knee arthroplasty (KTKA) have been reported in terms of limb and component alignment parameters but not in terms of gap laxities and differentials. In kinematic alignment (KA), balance should reflect the asymmetrical balance of the normal knee, not the classic rectangular flexion and extension gaps sought with gap-balanced mechanical axis total knee arthroplasty (MATKA). This paper aims to address the following questions: 1) what factors determine coronal joint congruence as measured on standing radiographs?; 2) is flexion gap asymmetry produced with KA?; 3) does lateral flexion gap laxity affect outcomes?; 4) is lateral flexion gap laxity associated with lateral extension gap laxity?; and 5) can consistent ligament balance be produced without releases?

Patients and Methods

A total of 192 KTKAs completed by a single surgeon using a computer-assisted technique were followed for a mean of 3.5 years (2 to 5). There were 116 male patients (60%) and 76 female patients (40%) with a mean age of 65 years (48 to 88). Outcome measures included intraoperative gap laxity measurements and component positions, as well as joint angles from postoperative three-foot standing radiographs. Patient-reported outcome measures (PROMs) were analyzed in terms of alignment and balance: EuroQol (EQ)-5D visual analogue scale (VAS), Knee Injury and Osteoarthritis Outcome Score (KOOS), KOOS Joint Replacement (JR), and Oxford Knee Score (OKS).


Bone & Joint Research
Vol. 8, Issue 10 | Pages 438 - 442
1 Oct 2019
Kayani B Haddad FS


Bone & Joint Research
Vol. 8, Issue 3 | Pages 126 - 135
1 Mar 2019
Sekiguchi K Nakamura S Kuriyama S Nishitani K Ito H Tanaka Y Watanabe M Matsuda S

Objectives

Unicompartmental knee arthroplasty (UKA) is one surgical option for treating symptomatic medial osteoarthritis. Clinical studies have shown the functional benefits of UKA; however, the optimal alignment of the tibial component is still debated. The purpose of this study was to evaluate the effects of tibial coronal and sagittal plane alignment in UKA on knee kinematics and cruciate ligament tension, using a musculoskeletal computer simulation.

Methods

The tibial component was first aligned perpendicular to the mechanical axis of the tibia, with a 7° posterior slope (basic model). Subsequently, coronal and sagittal plane alignments were changed in a simulation programme. Kinematics and cruciate ligament tensions were simulated during weight-bearing deep knee bend and gait motions. Translation was defined as the distance between the most medial and the most lateral femoral positions throughout the cycle.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 5 | Pages 682 - 687
1 Jul 2004
Bäthis H Perlick L Tingart M Lüring C Zurakowski D Grifka J

Restoration of neutral alignment of the leg is an important factor affecting the long-term results of total knee arthroplasty (TKA). Recent developments in computer-assisted surgery have focused on systems for improving TKA. In a prospective study two groups of 80 patients undergoing TKA had operations using either a computer-assisted navigation system or a conventional technique. Alignment of the leg and the orientation of components were determined on post-operative long-leg coronal and lateral films. The mechanical axis of the leg was significantly better in the computer-assisted group (96%, within ±3° varus/valgus) compared with the conventional group (78%, within ±3° varus/valgus). The coronal alignment of the femoral component was also more accurate in the computer-assisted group. Computer-assisted TKA gives a better correction of alignment of the leg and orientation of the components compared with the conventional technique. Potential benefits in the long-term outcome and functional improvement require further investigation


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1230 - 1237
1 Oct 2019
Kayani B Konan S Horriat S Ibrahim MS Haddad FS

Aims

The aim of this study was to assess the effect of posterior cruciate ligament (PCL) resection on flexion-extension gaps, mediolateral soft-tissue laxity, fixed flexion deformity (FFD), and limb alignment during posterior-stabilized (PS) total knee arthroplasty (TKA).

Patients and Methods

This prospective study included 110 patients with symptomatic osteoarthritis of the knee undergoing primary robot-assisted PS TKA. All operations were performed by a single surgeon using a standard medial parapatellar approach. Optical motion capture technology with fixed femoral and tibial registration pins was used to assess gaps before and after PCL resection in extension and 90° knee flexion. Measurements were made after excision of the anterior cruciate ligament and prior to bone resection. There were 54 men (49.1%) and 56 women (50.9%) with a mean age of 68 years (sd 6.2) at the time of surgery. The mean preoperative hip-knee-ankle deformity was 4.1° varus (sd 3.4).


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1463 - 1470
1 Nov 2018
Murphy BPD Dowsey MM Spelman T Choong PFM

Aims

As the population ages, there is projected to be an increase in the level of demand for total knee arthroplasty (TKA) in octogenarians. We aimed to explore whether those aged ≥ 80 years achieved similar improvements in physical function to younger patients while also comparing the rates of length of stay (LOS), discharge to rehabilitation, postoperative complications, and mortality following TKA in older and younger patients.

Patients and Methods

Patients from one institution who underwent primary elective TKA between 1 January 2006 and 31 December 2014 were dichotomized into those ≥ 80 years old (n = 359) and those < 80 years old (n = 2479) for comparison. Multivariable regression was used to compare the physical status component of the 12-Item Short-Form Health Survey (SF-12), LOS, discharge to rehabilitation, complications, and mortality between the two groups.


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 552 - 558
1 May 2019
Clark D Stevens JM Tortonese D Whitehouse MR Simpson D Eldridge J

Aims

The aim of this study was to determine and compare the congruency of the articular surface contact area of the patellofemoral joint (PFJ) during both active and passive movement of the knee with the use of an MRI mapping technique in both the stable and unstable PFJ.

Patients and Methods

A prospective case-control MRI imaging study of patients with a history of PFJ instability and a control group of volunteers without knee symptoms was performed. The PFJs were imaged with the use of an MRI scan during both passive and active movement from 0° through to 40° of flexion. The congruency through measurement of the contact surface area was mapped in 5-mm intervals on axial slices. In all, 40 patients were studied. The case group included 31 patients with symptomatic patellofemoral instability and the control group of nine asymptomatic volunteers. The ages were well matched between the case and control groups. The mean age was 25 years (16 to 42; sd 6.9) in the case group and 26 years (19 to 32; sd 5.1) in the control group. There were 19 female and 12 male patients in the case group.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1033 - 1042
1 Aug 2018
Kayani B Konan S Pietrzak JRT Huq SS Tahmassebi J Haddad FS

Aims

The primary aim of this study was to determine the surgical team’s learning curve for introducing robotic-arm assisted unicompartmental knee arthroplasty (UKA) into routine surgical practice. The secondary objective was to compare accuracy of implant positioning in conventional jig-based UKA versus robotic-arm assisted UKA.

Patients and Methods

This prospective single-surgeon cohort study included 60 consecutive conventional jig-based UKAs compared with 60 consecutive robotic-arm assisted UKAs for medial compartment knee osteoarthritis. Patients undergoing conventional UKA and robotic-arm assisted UKA were well-matched for baseline characteristics including a mean age of 65.5 years (sd 6.8) vs 64.1 years (sd 8.7), (p = 0.31); a mean body mass index of 27.2 kg.m2 (sd 2.7) vs 28.1 kg.m2 (sd 4.5), (p = 0.25); and gender (27 males: 33 females vs 26 males: 34 females, p = 0.85). Surrogate measures of the learning curve were prospectively collected. These included operative times, the Spielberger State-Trait Anxiety Inventory (STAI) questionnaire to assess preoperative stress levels amongst the surgical team, accuracy of implant positioning, limb alignment, and postoperative complications.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 151 - 158
1 Feb 2017
Huang T Long Y George D Wang W

Aims

There are two techniques widely used to determine the rotational alignment of the components in total knee arthroplasty (TKA); gap balancing (GB) and measured resection (MR). Which technique is the best remains controversial. We aimed to investigate this in a systematic review and meta-analysis.

Materials and Methods

In accordance with the methods of Cochrane, databases were searched for all randomised controlled trials in the literature between January 1986 and June 2015 comparing radiographic and clinical outcomes between the use of these two tecniques. Meta-analysis involved the use of the Revman5.3 software provided by Cochrane collaboration.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1620 - 1624
1 Dec 2016
Pailhé R Cognault J Massfelder J Sharma A Rouchy R Rubens-Duval B Saragaglia D

Aims

The role of high tibial osteotomy (HTO) is being questioned by the use of unicompartmental knee arthroplasty (UKA) in the treatment of medial compartment femorotibial osteoarthritis. Our aim was to compare the outcomes of revision HTO or UKA to a total knee arthroplasty (TKA) using computer-assisted surgery in matched groups of patients.

Patients and Methods

We conducted a retrospective study to compare the clinical and radiological outcome of patients who underwent revision of a HTO to a TKA (group 1) with those who underwent revision of a medial UKA to a TKA (group 2). All revision procedures were performed using computer-assisted surgery. We extracted these groups of patients from our database. They were matched by age, gender, body mass index, follow-up and pre-operative functional score. The outcomes included the Knee Society Scores (KSS), radiological outcomes and the rate of further revision.


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1596 - 1602
1 Dec 2017
Dunbar MJ Laende EK Collopy D Richardson CG

Aims

Hydroxyapatite coatings for uncemented fixation in total knee arthroplasty can theoretically provide a long-lasting biological interface with the host bone. The objective of this study was to test this hypothesis with propriety hydroxyapatite, peri-apatite, coated tibial components using component migration measured with radiostereometric analysis over two years as an indicator of long-term fixation.

Patients and Methods

A total of 29 patients at two centres received uncemented PA-coated tibial components and were followed for two years with radiostereometric analysis exams to quantify the migration of the component.


The Bone & Joint Journal
Vol. 98-B, Issue 11 | Pages 1471 - 1478
1 Nov 2016
Mooney LT Smith A Sloan K Clark GW

Aims

The aim of this study was to investigate differences in pain, range of movement function and satisfaction at three months and one year after total knee arthroplasty (TKA) in patients with an oblique pattern of kinematic graph of the knee and those with a varus pattern.

Patients and Methods

A total of 91 patients who underwent TKA were included in this retrospective study. Patients (59 women and 32 men with mean age of 68.7 years; 38.6 to 88.4) were grouped according to kinematic graphs which were generated during navigated TKA and the outcomes between the groups were compared.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 50 - 55
1 Jan 2018
Kono K Tomita T Futai K Yamazaki T Tanaka S Yoshikawa H Sugamoto K

Aims

In Asia and the Middle-East, people often flex their knees deeply in order to perform activities of daily living. The purpose of this study was to investigate the 3D kinematics of normal knees during high-flexion activities. Our hypothesis was that the femorotibial rotation, varus-valgus angle, translations, and kinematic pathway of normal knees during high-flexion activities, varied according to activity.

Materials and Methods

We investigated the in vivo kinematics of eight normal knees in four male volunteers (mean age 41.8 years; 37 to 53) using 2D and 3D registration technique, and modelled the knees with a computer aided design program. Each subject squatted, kneeled, and sat cross-legged. We evaluated the femoral rotation and varus-valgus angle relative to the tibia and anteroposterior translation of the medial and lateral side, using the transepicodylar axis as our femoral reference relative to the perpendicular projection on to the tibial plateau. This method evaluates the femur medially from what has elsewhere been described as the extension facet centre, and differs from the method classically applied.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1360 - 1368
1 Oct 2016
Waterson HB Clement ND Eyres KS Mandalia VI Toms AD

Aims

Our aim was to compare kinematic with mechanical alignment in total knee arthroplasty (TKA).

Patients and Methods

We performed a prospective blinded randomised controlled trial to compare the functional outcome of patients undergoing TKA in mechanical alignment (MA) with those in kinematic alignment (KA). A total of 71 patients undergoing TKA were randomised to either kinematic (n = 36) or mechanical alignment (n = 35). Pre- and post-operative hip-knee-ankle radiographs were analysed. The knee injury and osteoarthritis outcome score (KOOS), American Knee Society Score, Short Form-36, Euro-Qol (EQ-5D), range of movement (ROM), two minute walk, and timed up and go tests were assessed pre-operatively and at six weeks, three and six months and one year post-operatively.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1319 - 1328
1 Oct 2017
Shelton TJ Nedopil AJ Howell SM Hull ML

Aims

The aims of this study were to determine the proportion of patients with outlier varus or valgus alignment in kinematically aligned total knee arthroplasty (TKA), whether those with outlier varus or valgus alignment have higher forces in the medial or lateral compartments of the knee than those with in-range alignment and whether measurements of the alignment of the limb, knee and components predict compartment forces.

Patients and Methods

The intra-operative forces in the medial and lateral compartments were measured with an instrumented tibial insert in 67 patients who underwent a kinematically aligned TKA during passive movement. The mean of the forces at full extension, 45° and 90° of flexion determined the force in the medial and lateral compartments. Measurements of the alignment of the limb and the components included the hip-knee-ankle (HKA) angle, proximal medial tibial angle (PMTA), and distal lateral femoral angle (DLFA). Measurements of the alignment of the knee and the components included the tibiofemoral angle (TFA), tibial component angle (TCA) and femoral component angle (FCA). Alignment was measured on post-operative, non-weight-bearing anteroposterior (AP) scanograms and categorised as varus or valgus outlier or in-range in relation to mechanically aligned criteria.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 78 - 80
1 Jan 2016
Lee G

Patient specific instrumentation (PSI) uses advanced imaging of the knee (CT or MRI) to generate individualised cutting blocks aimed to make the procedure of total knee arthroplasty (TKA) more accurate and efficient. However, in this era of healthcare cost consciousness, the value of new technologies needs to be critically evaluated. There have been several comparative studies looking at PSI versus standard instrumentation. Most compare PSI with conventional instrumentation in terms of alignment in the coronal plane, operative time and surgical efficiency, cost effectiveness and short-term outcomes. Several systematic reviews and meta-analyses have also been published. PSI has not been shown to be superior compared with conventional instrumentation in its ability to restore traditional mechanical alignment in primary TKA. Most studies show comparative efficacy and no decrease in the number of outliers in either group. In terms of operative time and efficiency, PSI tended towards decreasing operative time, saving a mean of five minutes per patient (0 to 20). Furthermore, while some cost savings could be realised with less operative time and reduced instrumentation per patient, these savings were overcome by the cost of the CT/MRI and the cutting blocks. Finally, there was no evidence that PSI positively affected clinical outcomes at two days, two months, or two years. Consequently, current evidence does not support routine use of PSI in routine primary TKA.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):78–80.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 334 - 340
1 Mar 2016
Tayton ER Frampton C Hooper GJ Young SW

Aims

The aim of this study was to identify risk factors for prosthetic joint infection (PJI) following total knee arthroplasty (TKA).

Patients and Methods

The New Zealand Joint Registry database was analysed, using revision surgery for PJI at six and 12 months after surgery as primary outcome measures. Statistical associations between revision for infection, with common and definable surgical and patient factors were tested.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1060 - 1064
1 Aug 2011
Zaffagnini S Bonanzinga T Muccioli GMM Giordano G Bruni D Bignozzi S Lopomo N Marcacci M

We have shown in a previous study that patients with combined lesions of the anterior cruciate (ACL) and medial collateral ligaments (MCL) had similar anteroposterior (AP) but greater valgus laxity at 30° after reconstruction of the ACL when compared with patients who had undergone reconstruction of an isolated ACL injury. The present study investigated the same cohort of patients after a minimum of three years to evaluate whether the residual valgus laxity led to a poorer clinical outcome.

Each patient had undergone an arthroscopic double-bundle ACL reconstruction using a semitendinosus-gracilis graft. In the combined ACL/MCL injury group, the grade II medial collateral ligament injury was not treated. At follow-up, AP laxity was measured using a KT-2000 arthrometer, while valgus laxity was evaluated with Telos valgus stress radiographs and compared with the uninjured knee. We evaluated clinical outcome scores, muscle girth and time to return to activities for the two groups.

Valgus stress radiographs showed statistically significant greater mean medial joint opening in the reconstructed compared with the uninjured knees (1.7 mm (sd 0.9) versus 0.9 mm (sd 0.7), respectively, p = 0.013), while no statistically significant difference was found between the AP laxity and the other clinical parameters. Our results show that the residual valgus laxity does not affect AP laxity significantly at a minimum follow up of three years, suggesting that no additional surgical procedure is needed for the medial collateral ligament in combined lesions.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1372 - 1376
1 Oct 2012
Komzák M Hart R Okál F Safi A

The biomechanical function of the anteromedial (AM) and posterolateral (PL) bundles of the anterior cruciate ligament (ACL) remains controversial. Some studies report that the AM bundle stabilises the knee joint in anteroposterior (AP) translation and rotational movement (both internal and external) to the same extent as the PL bundle. Others conclude that the PL bundle is more important than the AM in controlling rotational movement.

The objective of this randomised cohort study involving 60 patients (39 men and 21 women) with a mean age of 32.9 years (18 to 53) was to evaluate the function of the AM and the PL bundles of the ACL in both AP and rotational movements of the knee joint after single-bundle and double-bundle ACL reconstruction using a computer navigation system. In the double-bundle group the patients were also randomised to have the AM or the PL bundle tensioned first, with knee laxity measured after each stage of reconstruction. All patients had isolated complete ACL tears, and the presence of a meniscal injury was the only supplementary pathology permitted for inclusion in the trial. The KT-1000 arthrometer was used to apply a constant load to evaluate the AP translation and the rolimeter was used to apply a constant rotational force. For the single-bundle group deviation was measured before and after ACL reconstruction. In the double-bundle group deviation was measured for the ACL-deficient, AM- or PL-reconstructed first conditions and for the total reconstruction.

We found that the AM bundle in the double-bundle group controlled rotation as much as the single-bundle technique, and to a greater extent than the PL bundle in the double-bundle technique. The double-bundle technique increases AP translation and rotational stability in internal rotation more than the single-bundle technique.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 628 - 633
1 May 2016
Heijens E Kornherr P Meister C

Aims

In patients undergoing medial opening wedge high tibial osteotomy (MOWHTO), soft tissue opening on the medial side of the knee is difficult to predict. When the load bearing axis is corrected beyond a certain point, the knee joint tilts open on the medial side. We therefore hypothesised that there is a tipping point and defined this as the coronal hypomochlion.

Patients and Methods

In this prospective study of 150 navigated MOWHTOs (144 consecutive patients), data were collected before surgery and at three months post-operatively. In order to calculate the hypomochlion, we compared the respective changes to the joint line convergence angle (JLCA) with the post-operative axis of the leg. The change to the medial proximal tibial angle accounts for only about 80% of the change to the femorotibial angle; 20% of the correction can therefore be attributed to non-osseous, soft-tissue changes.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 56 - 63
1 Jan 2015
Abane L Anract P Boisgard S Descamps S Courpied JP Hamadouche M

In this study we randomised 140 patients who were due to undergo primary total knee arthroplasty (TKA) to have the procedure performed using either patient-specific cutting guides (PSCG) or conventional instrumentation (CI).

The primary outcome measure was the mechanical axis, as measured at three months on a standing long-leg radiograph by the hip–knee–ankle (HKA) angle. This was undertaken by an independent observer who was blinded to the instrumentation. Secondary outcome measures were component positioning, operating time, Knee Society and Oxford knee scores, blood loss and length of hospital stay.

A total of 126 patients (67 in the CI group and 59 in the PSCG group) had complete clinical and radiological data. There were 88 females and 52 males with a mean age of 69.3 years (47 to 84) and a mean BMI of 28.6 kg/m2 (20.2 to 40.8). The mean HKA angle was 178.9° (172.5 to 183.4) in the CI group and 178.2° (172.4 to 183.4) in the PSCG group (p = 0.34). Outliers were identified in 22 of 67 knees (32.8%) in the CI group and 19 of 59 knees (32.2%) in the PSCG group (p = 0.99). There was no significant difference in the clinical results (p = 0.95 and 0.59, respectively). Operating time, blood loss and length of hospital stay were not significantly reduced (p = 0.09, 0.58 and 0.50, respectively) when using PSCG.

The use of PSCG in primary TKA did not reduce the proportion of outliers as measured by post-operative coronal alignment.

Cite this article: Bone Joint J 2015;97-B:56–63.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 601 - 607
1 May 2016
McClelland D Barlow D Moores TS Wynn-Jones C Griffiths D Ogrodnik PJ Thomas PBM

In arthritis of the varus knee, a high tibial osteotomy (HTO) redistributes load from the diseased medial compartment to the unaffected lateral compartment.

We report the outcome of 36 patients (33 men and three women) with 42 varus, arthritic knees who underwent HTO and dynamic correction using a Garches external fixator until they felt that normal alignment had been restored. The mean age of the patients was 54.11 years (34 to 68). Normal alignment was achieved at a mean 5.5 weeks (3 to 10) post-operatively. Radiographs, gait analysis and visual analogue scores for pain were measured pre- and post-operatively, at one year and at medium-term follow-up (mean six years; 2 to 10). Failure was defined as conversion to knee arthroplasty.

Pre-operative gait analysis divided the 42 knees into two equal groups with high (17 patients) or low (19 patients) adductor moments. After correction, a statistically significant (p < 0.001, t-test,) change in adductor moment was achieved and maintained in both groups, with a rate of failure of three knees (7.1%), and 89% (95% confidence interval (CI) 84.9 to 94.7) survivorship at medium-term follow-up.

At final follow-up, after a mean of 15.9 years (12 to 20), there was a survivorship of 59% (95% CI 59.6 to 68.9) irrespective of adductor moment group, with a mean time to conversion to knee arthroplasty of 9.5 years (3 to 18; 95% confidence interval ± 2.5).

HTO remains a useful option in the medium-term for the treatment of medial compartment osteoarthritis of the knee but does not last in the long-term.

Cite this article: Bone Joint J 2016;98-B:601–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1217 - 1222
1 Sep 2011
Bonner TJ Eardley WGP Patterson P Gregg PJ

Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR.

A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by > 3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening. Kaplan-Meier survival analysis showed a weak tendency towards improved survival with restoration of a neutral mechanical axis, but this did not reach statistical significance (p = 0.47).

We found that the relationship between survival of a primary TKR and mechanical axis alignment is weaker than that described in a number of previous reports.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 56 - 61
1 Jan 2012
Kawahara S Matsuda S Fukagawa S Mitsuyasu H Nakahara H Higaki H Shimoto T Iwamoto Y

In posterior stabilised total knee replacement (TKR) a larger femoral component is sometimes selected to manage the increased flexion gap caused by resection of the posterior cruciate ligament. However, concerns remain regarding the adverse effect of the increased anteroposterior dimensions of the femoral component on the patellofemoral (PF) joint. Meanwhile, the gender-specific femoral component has a narrower and thinner anterior flange and is expected to reduce the PF contact force. PF contact forces were measured at 90°, 120°, 130° and 140° of flexion using the NexGen Legacy Posterior Stabilized (LPS)-Flex Fixed Bearing Knee system using Standard, Upsized and Gender femoral components during TKR. Increasing the size of the femoral component significantly increased mean PF forces at 120°, 130° and 140° of flexion (p = 0.005, p < 0.001 and p < 0.001, respectively). No difference was found in contact force between the Gender and the Standard components. Among the patients who had overhang of the Standard component, mean contact forces with the Gender component were slightly lower than those of the Standard component, but no statistical difference was found at 90°, 120°, 130° or 140° of flexion (p = 0.689, 0.615, 0.253 and 0.248, respectively).

Upsized femoral components would increase PF forces in deep knee flexion. Gender-specific implants would not reduce PF forces.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 124 - 128
1 Nov 2013
Hofmann AA Shaeffer JF McCandless JB Magee TH

Isolated patellofemoral arthritis is a common condition and there are varying opinions on the most effective treatments. Non-operative and operative treatments have failed to demonstrate effective long-term treatment for those in an advanced stage of the condition. Newer designs and increased technology in patellofemoral replacement (PFR) have produced more consistent outcomes. This has led to a renewed enthusiasm for this procedure. Newer PFR prostheses have addressed the patellar maltracking issues plaguing some of the older designs. Short-term results with contemporary prostheses and new technology are described here.

Cite this article: Bone Joint J 2013;95-B, Supple A:124–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1238 - 1244
1 Sep 2010
Nicoll D Rowley DI

This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003.

The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p < 0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain.

Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 609 - 618
1 May 2014
Gøthesen Ø Espehaug B Havelin LI Petursson G Hallan G Strøm E Dyrhovden G Furnes O

We performed a randomised controlled trial comparing computer-assisted surgery (CAS) with conventional surgery (CONV) in total knee replacement (TKR). Between 2009 and 2011 a total of 192 patients with a mean age of 68 years (55 to 85) with osteoarthritis or arthritic disease of the knee were recruited from four Norwegian hospitals. At three months follow-up, functional results were marginally better for the CAS group. Mean differences (MD) in favour of CAS were found for the Knee Society function score (MD: 5.9, 95% confidence interval (CI) 0.3 to 11.4, p = 0.039), the Knee Injury and Osteoarthritis Outcome Score (KOOS) subscales for ‘pain’ (MD: 7.7, 95% CI 1.7 to 13.6, p = 0.012), ‘sports’ (MD: 13.5, 95% CI 5.6 to 21.4, p = 0.001) and ‘quality of life’ (MD: 7.2, 95% CI 0.1 to 14.3, p = 0.046). At one-year follow-up, differences favouring CAS were found for KOOS ‘sports’ (MD: 11.0, 95% CI 3.0 to 19.0, p = 0.007) and KOOS ‘symptoms’ (MD: 6.7, 95% CI 0.5 to 13.0, p = 0.035). The use of CAS resulted in fewer outliers in frontal alignment (> 3° malalignment), both for the entire TKR (37.9% vs 17.9%, p = 0.042) and for the tibial component separately (28.4% vs 6.3%, p = 0.002). Tibial slope was better achieved with CAS (58.9% vs 26.3%, p < 0.001). Operation time was 20 minutes longer with CAS. In conclusion, functional results were, statistically, marginally in favour of CAS. Also, CAS was more predictable than CONV for mechanical alignment and positioning of the prosthesis. However, the long-term outcomes must be further investigated.

Cite this article: Bone Joint J 2014; 96-B:609–18.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 153 - 158
1 Nov 2013
Victor J Premanathan A

We have investigated the benefits of patient specific instrument guides, applied to osteotomies around the knee. Single, dual and triple planar osteotomies were performed on tibias or femurs in 14 subjects. In all patients, a detailed pre-operative plan was prepared based upon full leg standing radiographic and CT scan information. The planned level of the osteotomy and open wedge resection was relayed to the surgery by virtue of a patient specific guide developed from the images. The mean deviation between the planned wedge angle and the executed wedge angle was 0° (-1 to 1, sd 0.71) in the coronal plane and 0.3° (-0.9 to 3, sd 1.14) in the sagittal plane. The mean deviation between the planned hip, knee, ankle angle (HKA) on full leg standing radiograph and the post-operative HKA was 0.3° (-1 to 2, sd 0.75). It is concluded that this is a feasible and valuable concept from the standpoint of pre-operative software based planning, surgical application and geometrical accuracy of outcome.

Cite this article: Bone Joint J 2013;95-B, Supple A:153–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 345 - 350
1 Mar 2011
Huang T Hsu W Peng K Hsu RW

We conducted a retrospective study to investigate the effect of femoral bowing on the placement of components in total knee replacement (TKR), with regard to its effect on reestablishing the correct mechanical axis, as we hypothesised that computer-assisted total knee replacement (CAS-TKR) would produce more accurate alignment than conventional TKR. Between January 2006 and December 2009, 212 patients (306 knees) underwent TKR. The conventional TKR was compared with CAS-TKR for accuracy of placement of the components and post-operative alignment, as determined by five radiological measurements. There were significant differences in the reconstructed mechanical axes between the bowed and the non-bowed group after conventional TKR (176.2° (sd 3.4) vs 179.3° (sd 2.1), p < 0.001).

For patients with significant femoral bowing, the reconstructed mechanical axes were significantly closer to normal in the CAS group than in the conventional group (179.2° (sd 1.9) vs 176.2° (sd 3.4), p < 0.001). Femoral bowing resulted in inaccuracy when a conventional technique was used. CAS-TKR provides an effective method of restoring the mechanical axis in the presence of significant femoral bowing.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 475 - 480
1 Apr 2009
Matsumoto T Kuroda R Kubo S Muratsu H Mizuno K Kurosaka M

We have developed a new tensor for total knee replacements which is designed to assist with soft-tissue balancing throughout the full range of movement with a reduced patellofemoral joint. Using this tensor in 40 patients with osteoarthritis we compared the intra-operative joint gap in cruciate-retaining and posterior-stabilised total knee replacements at 0°, 10°, 45°, 90° and 135° of flexion, with the patella both everted and reduced.

While the measurement of the joint gap with a reduced patella in posterior-stabilised knees increased from extension to flexion, it remained constant for cruciate-retaining joints throughout a full range of movement. The joint gaps at deep knee flexion were significantly smaller for both types of prosthetic knee when the patellofemoral joint was reduced (p < 0.05).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 95 - 99
1 Nov 2012
Barrack RL Ruh EL Williams BM Ford AD Foreman K Nunley RM

Patient specific cutting guides generated by preoperative Magnetic Resonance Imaging (MRI) of the patient’s extremity have been proposed as a method of improving the consistency of Total Knee Arthroplasty (TKA) alignment and adding efficiency to the operative procedure. The cost of this option was evaluated by quantifying the savings from decreased operative time and instrument processing costs compared to the additional cost of the MRI and the guide. Coronal plane alignment was measured in an unselected consecutive series of 200 TKAs, 100 with standard instrumentation and 100 with custom cutting guides. While the cutting guides had significantly lower total operative time and instrument processing time, the estimated $322 savings was overwhelmed by the $1,500 additional cost of the MRI and the cutting guide. All measures of coronal plane alignment were equivalent between the two groups. The data does not currently support the proposition that patient specific guides add value to TKA.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 897 - 903
1 Jul 2011
Bachhal V Sankhala SS Jindal N Dhillon MS

We report the outcome of 32 patients (37 knees) who underwent hemicallostasis with a dynamic external fixator for osteoarthritis of the medial compartment of the knee. There were 16 men (19 knees) and 16 women (18 knees) with a mean age at operation of 54.6 years (27 to 72). The aim was to achieve a valgus overcorrection of 2° to 8° or mechanical axis at 62.5% (± 12.5%). At a mean follow-up of 62.8 months (51 to 81) there was no change in the mean range of movement, and no statistically significant difference in the Insall-Salvati index or tibial slope (p = 0.11 and p = 0.15, respectively). The mean hip-knee-ankle angle changed from 190.6 (183° to 197°) to 176.0° (171° to 181°), with a mean final position of the mechanical axis of 58.5% (35.1% to 71.2%).

The desired alignment was attained in 31 of 37 (84%) knees. There were 21 excellent, 13 good, two fair and one poor result according to the Oxford knee score with no correlation between age and final score. This score was at its best at one year with a statistically significant deterioration at two years (p = 0.001) followed by a small but not statistically significant deterioration until the final follow-up (p = 0.17). All the knees with Ahlback grade 1 osteoarthritis had excellent or good results. Complications included pin tract infections involving 16.4% of all pins used, delayed union in two, knee stiffness in four, fracture of the lateral cortex in one and ring sequestrum in one.

In conclusion, hemicallostasis provides precision in attaining the desired alignment without interfering with tibial slope or patellar height, and is relatively free of serious complications.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1333 - 1338
1 Oct 2014
Gustke KA Golladay GJ Roche MW Jerry GJ Elson LC Anderson CR

The aim of this prospective multicentre study was to report the patient satisfaction after total knee replacement (TKR), undertaken with the aid of intra-operative sensors, and to compare these results with previous studies. A total of 135 patients undergoing TKR were included in the study. The soft-tissue balance of each TKR was quantified intra-operatively by the sensor, and 18 (13%) were found to be unbalanced. A total of 113 patients (96.7%) in the balanced group and 15 (82.1%) in the unbalanced group were satisfied or very satisfied one year post-operatively (p = 0.043).

A review of the literature identified no previous study with a mean level of satisfaction that was greater than the reported level of satisfaction of the balanced TKR group in this study. Ensuring soft-tissue balance by using intra-operative sensors during TKR may improve satisfaction.

Cite this article: Bone Joint J 2014;96-B:1333–8


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 339 - 344
1 Mar 2014
Saito T Kumagai K Akamatsu Y Kobayashi H Kusayama Y

Between 2003 and 2007, 99 knees in 77 patients underwent opening wedge high tibial osteotomy. We evaluated the effect of initial stable fixation combined with an artificial bone substitute on the mid- to long-term outcome after medial opening-wedge high tibial osteotomy (HTO) for medial compartmental osteoarthritis or spontaneous osteonecrosis of the knee in 78 knees in 64 patients available for review at a minimum of five years (mean age 68 years; 49 to 82). The mean follow-up was 6.5 years (5 to 10). The mean Knee Society knee score and function score improved from 49.6 (sd 11.4, 26 to 72) and 56.6 (sd 15.6, 5 to 100) before surgery to 88.1 (sd 12.5, 14 to 100) and 89.4 (sd 15.6, 5 to 100) at final follow-up (p <  0.001) respectively. There were no significant differences between patients aged ≥ 70 and < 70 years. The mean standing femorotibial angle was corrected significantly from 181.7° (sd 2.7°, 175° to 185°) pre-operatively to 169.7° (sd 2.4°, 164° to 175°) at one year’s follow-up (p < 0.001) and 169.6° (sd 3.0°, 157° to 179°) at the final follow-up (p = 0.69 vs one year).

Opening-wedge HTO using a stable plate fixation system combined with a bone substitute is a reliable procedure that provides excellent results. Although this treatment might seem challenging for older patients, our results strongly suggest that the results are equally good.

Cite this article: Bone Joint J 2014;96-B:339–44.