Advertisement for orthosearch.org.uk
Results 1 - 32 of 32
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1271 - 1276
1 Sep 2012
Luyckx T Peeters T Vandenneucker H Victor J Bellemans J

Obtaining a balanced flexion gap with correct femoral component rotation is one of the prerequisites for a successful outcome after total knee replacement (TKR). Different techniques for achieving this have been described. In this study we prospectively compared gap-balancing versus measured resection in terms of reliability and accuracy for femoral component rotation in 96 primary TKRs performed in 96 patients using the Journey system. In 48 patients (18 men and 30 women) with a mean age of 65 years (45 to 85) a tensor device was used to determine rotation. In the second group of 48 patients (14 men and 34 women) with a mean age of 64 years (41 to 86), an ‘adapted’ measured resection technique was used, taking into account the native rotational geometry of the femur as measured on a pre-operative CT scan. Both groups systematically reproduced a similar external rotation of the femoral component relative to the surgical transepicondylar axis: 2.4° . (sd. 2.5) in the gap-balancing group and 1.7° (. sd. 2.1) in the measured resection group (p = 0.134). Both gap-balancing and adapted measured resection techniques proved equally reliable and accurate in determining femoral component rotation after TKR. There was a tendency towards more external rotation in the gap-balancing group, but this difference was not statistically significant (p = 0.134). The number of outliers for our ‘adapted’ measured resection technique was much lower than reported in the literature


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1248 - 1255
1 Oct 2019
Pineda A Pabbruwe MB Kop AM Vlaskovsky P Hurworth M

Aims. The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Materials and Methods. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed. Results. In 104 implants, a decrease in the odds of loosening and an increase in metallosis and tissue staining in the Duofix group relative to Porocoat group was detected (p = 0.028). There was an increased presence of embedded metallic debris in the Duofix group (p < 0.001). Decreased tissue ingrowth was associated with the Duofix surface (p < 0.001). The attached beads had reduced microhardness, indicative of adverse thermal processing, which resulted in bead shedding, particulate debris, and metallosis. Conclusion. Hydroxyapatite coating of the LCS femoral component produced unexpected results and led to its recall. The root cause was likely a combination of retained alumina grit and a reduction in bead microhardness (mechanical strength) resulting in increased particle debris, metallosis, and early revision. The Duofix LCS femoral component was not equivalent to the Porocoat version despite its approval through the Food and Drug Administration (FDA) 510(k) equivalance approval process. Regulation of the introduction of modified existing devices needs to be improved and the Duofix LCS should have been considered to be a new device for which equivalence had not been demonstrated at the point of introduction. Cite this article: Bone Joint J 2019;101-B:1248–1255


Bone & Joint Open
Vol. 5, Issue 10 | Pages 832 - 836
4 Oct 2024
Kayani B Mancino F Baawa-Ameyaw J Roussot MA Haddad FS

Aims

The outcomes of patients with unexpected positive cultures (UPCs) during revision total hip arthroplasty (THA) and total knee arthroplasty (TKA) remain unknown. The objectives of this study were to establish the prevalence and infection-free implant survival in UPCs during presumed aseptic single-stage revision THA and TKA at mid-term follow-up.

Methods

This study included 297 patients undergoing presumed aseptic single-stage revision THA or TKA at a single treatment centre. All patients with at least three UPCs obtained during revision surgery were treated with minimum three months of oral antibiotics following revision surgery. The prevalence of UPCs and causative microorganisms, the recurrence of periprosthetic joint infections (PJIs), and the infection-free implant survival were established at minimum five years’ follow-up (5.1 to 12.3).


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 900 - 907
1 Nov 1997
McLaughlin JR Lee KR

We followed 138 patients (145 hips) who had had uncemented total hip arthroplasty using the Taperloc femoral component for a mean of ten years (8 to 12.5). No patient was lost to follow-up; 31 (31 hips) died before the minimum time of eight years for inclusion in the study, and 30 of these still had their femoral component in place. One well-fixed prosthesis had been exchanged at the time of acetabular revision. Of the remaining 114 hips, one femoral component required revision for aseptic loosening and one for sepsis. Three other well-fixed femoral components were removed during acetabular revision. Complete clinical and radiological follow-up was obtained in the 109 hips which had not had revision. Clinically, 94 (87%) were rated good or excellent, eight (7%) fair and seven (6%) poor. The average Harris hip score increased from 48 before operation to 88 at the time of the last follow-up. Radiologically, 103 hips (94%) had fixation by bone ingrowth, three (3%) showed stable fibrous ingrowth and three (3%) were unstable. Osteolysis of the femoral cortex was seen in seven hips (6%), with major lysis in only one. At a mean follow-up of ten years, the results of the Taperloc femoral component are comparable with those of modern techniques of cementing in primary total hip arthroplasty


The Bone & Joint Journal
Vol. 104-B, Issue 8 | Pages 929 - 937
1 Aug 2022
Gurung B Liu P Harris PDR Sagi A Field RE Sochart DH Tucker K Asopa V

Aims

Total hip arthroplasty (THA) and total knee arthroplasty (TKA) are common orthopaedic procedures requiring postoperative radiographs to confirm implant positioning and identify complications. Artificial intelligence (AI)-based image analysis has the potential to automate this postoperative surveillance. The aim of this study was to prepare a scoping review to investigate how AI is being used in the analysis of radiographs following THA and TKA, and how accurate these tools are.

Methods

The Embase, MEDLINE, and PubMed libraries were systematically searched to identify relevant articles. The Preferred Reporting Items for Systematic Reviews and Meta-Analyses extension for scoping reviews and Arksey and O’Malley framework were followed. Study quality was assessed using a modified Methodological Index for Non-Randomized Studies tool. AI performance was reported using either the area under the curve (AUC) or accuracy.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 122 - 129
1 Apr 2016
Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives. Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Methods. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques. Results. Rotational malalignment between femoral and tibial components generated 40% less overall tibial tray micromotion in RP designs than in standard fixed bearing tibial trays. RP trays reduced micromotion by up to 172 µm in axial compression and 84 µm in rotational malalignment models. Conclusions. Reduced torque transfer at the tibiofemoral interface in RP tibial trays reduces relative component micromotion and may aid long-term stability in cases of revision TKA or poor bone quality. Cite this article: Mr S. R. Small. Micromotion at the tibial plateau in primary and revision total knee arthroplasty: fixed versus rotating platform designs. Bone Joint Res 2016;5:122–129. DOI: 10.1302/2046-3758.54.2000481


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 959 - 964
1 Jul 2020
Malik AT Li M Khan SN Alexander JH Li D Scharschmidt TJ

Aims

Currently, the US Center for Medicaid and Medicare Services (CMS) has been testing bundled payments for revision total joint arthroplasty (TJA) through the Bundled Payment for Care Improvement (BPCI) programme. Under the BPCI, bundled payments for revision TJAs are defined on the basis of diagnosis-related groups (DRGs). However, these DRG-based bundled payment models may not be adequate to account appropriately for the varying case-complexity seen in revision TJAs.

Methods

The 2008-2014 Medicare 5% Standard Analytical Files (SAF5) were used to identify patients undergoing revision TJA under DRG codes 466, 467, or 468. Generalized linear regression models were built to assess the independent marginal cost-impact of patient, procedural, and geographic characteristics on 90-day costs.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 5 | Pages 792 - 795
1 Sep 1997
Lehtimäki MY Lehto MUK Kautiainen H Savolainen HA Hämäläinen MMJ

Between 1971 and 1991 we performed Charnley low-friction arthroplasty (LFA) on 116 patients (186 hips) with juvenile chronic arthritis (JCA). We have now carried out a survival study, taking endpoints as revision, death or the end of the year 1993. Overall survival was 91.9% at ten years and 83.0% at 15 years. That of the femoral component was 95.6% at ten years and 91.9% at 15 years and of the acetabulum 95.0% and 87.8%, respectively. Only the use of steroids significantly impaired the survival. We therefore recommend the use of Charnley LFA for young patients with JCA requiring total hip replacement


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 1 | Pages 103 - 107
1 Jan 2000
Pekkarinen J Alho A Lepistö J Ylikoski M Ylinen P Paavilainen T

We have reviewed retrospectively 68 revisions of the femoral component in arthroplasties of the hip in 65 patients, using impaction bone grafting, at a median of three years (1 month to 6 years). We employed the cemented Exeter X-Change technique in 36 patients and the uncemented Bi-Metric allografting method in 32. The 37 bone defects were grade 3 or grade 4 on the Endo-Klinik classification. The Mayo hip score improved from a mean of 32 (. sd. ± 18) to 62 (. sd. ± 15). Most (25) of the 34 complications occurred in grade-3 and grade-4 defects; nine were intraoperative diaphyseal fractures and eight fractures of the greater trochanter. All the fractures united. The risk of intraoperative fracture was prevented by supporting the bone with wires in 16 hips, with reinforcement mesh in 18 and by a plate in six. Early migration of the stem of more than 10 mm during the first year indicated rotational instability; it occurred in three cases. In difficult revision cases with large defects of the femoral bone, bone-impaction techniques carry a high risk of complications


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 960 - 969
1 Aug 2019
Odgaard A Laursen MB Gromov K Troelsen A Kristensen PW Schrøder H Madsen F Overgaard S

Aims

The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check.

Materials and Methods

Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 951 - 959
1 Aug 2019
Preston N McHugh GA Hensor EMA Grainger AJ O’Connor PJ Conaghan PG Stone MH Kingsbury SR

Aims

This study aimed to develop a virtual clinic for the purpose of reducing face-to-face orthopaedic consultations.

Patients and Methods

Anonymized experts (hip and knee arthroplasty patients, surgeons, physiotherapists, radiologists, and arthroplasty practitioners) gave feedback via a Delphi Consensus Technique. This consisted of an iterative sequence of online surveys, during which virtual documents, made up of a patient-reported questionnaire, standardized radiology report, and decision-guiding algorithm, were modified until consensus was achieved. We tested the patient-reported questionnaire on seven patients in orthopaedic clinics using a ‘think-aloud’ process to capture difficulties with its completion.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1592 - 1599
1 Dec 2018
Galea VP Connelly JW Shareghi B Kärrholm J Sköldenberg O Salemyr M Laursen MB Muratoglu O Bragdon C Malchau H

Aims

The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure.

Patients and Methods

A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 60 - 64
1 Jan 2017
Lange J Haas SB

Valgus knee deformity can present a number of unique surgical challenges for the total knee arthroplasty (TKA) surgeon. Understanding the typical patterns of bone and soft-tissue pathology in the valgus arthritic knee is critical for appropriate surgical planning. This review aims to provide the knee arthroplasty surgeon with an understanding of surgical management strategies for the treatment of valgus knee arthritis.

Lateral femoral and tibial deficiencies, contracted lateral soft tissues, attenuated medial soft tissues, and multiplanar deformities may all be present in the valgus arthritic knee. A number of classifications have been reported in order to guide surgical management, and a variety of surgical strategies have been described with satisfactory clinical results. Depending on the severity of the deformity, a variety of TKA implant designs may be appropriate for use.

Regardless of an operating surgeon’s preferred surgical strategy, adherence to a step-wise approach to deformity correction is advised.

Cite this article: Bone Joint J 2017;99-B(1 Supple A):60–4.


The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1482 - 1489
1 Nov 2017
Jacobs AME Bénard M Meis JF van Hellemondt G Goosen JHM

Aims

Positive cultures are not uncommon in cases of revision total knee and hip arthroplasty (TKA and THA) for presumed aseptic causes. The purpose of this study was to assess the incidence of positive intra-operative cultures in presumed aseptic revision of TKA and THA, and to determine whether the presence of intra-operative positive cultures results in inferior survival in such cases.

Patients and Methods

A retrospective cohort study was assembled with 679 patients undergoing revision knee (340 cases) or hip arthroplasty (339 cases) for presumed aseptic causes. For all patients three or more separate intra-operative cultures were obtained. Patients were diagnosed with a previously unsuspected prosthetic joint infection (PJI) if two or more cultures were positive with the same organism. Records were reviewed for demographic details, pre-operative laboratory results and culture results. The primary outcome measure was infection-free implant survival at two years.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 349 - 358
1 Mar 2016
Akiyama K Nakata K Kitada M Yamamura M Ohori T Owaki H Fuji T

Aims

We investigated changes in the axial alignment of the ipsilateral hip and knee after total hip arthroplasty (THA).

Patients and Methods

We reviewed 152 patients undergoing primary THA (163 hips; 22 hips in men, 141 hips in women) without a pre-operative flexion contracture. The mean age was 64 years (30 to 88). The diagnosis was osteoarthritis (OA) in 151 hips (primary in 18 hips, and secondary to dysplasia in 133) and non-OA in 12 hips. A posterolateral approach with repair of the external rotators was used in 134 hips and an anterior approach in 29 hips. We measured changes in leg length and offset on radiographs, and femoral anteversion, internal rotation of the hip and lateral patellar tilt on CT scans, pre- and post-operatively.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 101 - 104
1 Jan 2016
Mullaji AB Shetty GM

Collateral ligament release is advocated in total knee arthroplasty (TKA) to deal with significant coronal plane deformities, but is also associated with significant disadvantages.

We describe steps to avoid release of the collateral (superficial medial and lateral collateral) ligaments during TKA in severely deformed knees, while correcting deformity and balancing the knee.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):101–4.


Bone & Joint Research
Vol. 3, Issue 4 | Pages 95 - 100
1 Apr 2014
Kaneyama R Otsuka M Shiratsuchi H Oinuma K Miura Y Tamaki T

Objectives

Because posterior cruciate ligament (PCL) resection makes flexion gaps wider in total knee replacement (TKR), preserving or sacrificing a PCL affects the gap equivalence; however, there are no criteria for the PCL resection that consider gap situations of each knee. This study aims to investigate gap characteristics of knees and to consider the criteria for PCL resection.

Methods

The extension and flexion gaps were measured, first with the PCL preserved and subsequently with the PCL removed (in cases in which posterior substitute components were selected). The PCL preservation or sacrifice was solely determined by the gap measurement results, without considering other functions of the PCL such as ‘roll back.’


Bone & Joint Research
Vol. 5, Issue 7 | Pages 307 - 313
1 Jul 2016
Sandgren B Skorpil M Nowik P Olivecrona H Crafoord J Weidenhielm L Persson A

Objectives

Computed tomography (CT) plays an important role in evaluating wear and periacetabular osteolysis (PAO) in total hip replacements. One concern with CT is the high radiation exposure since standard pelvic CT provides approximately 3.5 millisieverts (mSv) of radiation exposure, whereas a planar radiographic examination with three projections totals approximately 0.5 mSv. The objective of this study was to evaluate the lowest acceptable radiation dose for dual-energy CT (DECT) images when measuring wear and periacetabular osteolysis in uncemented metal components.

Materials and Methods

A porcine pelvis with bilateral uncemented hip prostheses and with known linear wear and acetabular bone defects was examined in a third-generation multidetector DECT scanner. The examinations were performed with four different radiation levels both with and without iterative reconstruction techniques. From the high and low peak kilo voltage acquisitions, polychrmoatic images were created together with virtual monochromatic images of energies 100 kiloelectron volts (keV) and 150 keV.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1669 - 1673
1 Dec 2014
Van der Merwe JM Haddad FS Duncan CP

The Unified Classification System (UCS) was introduced because of a growing need to have a standardised universal classification system of periprosthetic fractures. It combines and simplifies many existing classification systems, and can be applied to any fracture around any partial or total joint replacement occurring during or after operation. Our goal was to assess the inter- and intra-observer reliability of the UCS in association with knee replacement when classifying fractures affecting one or more of the femur, tibia or patella.

We used an international panel of ten orthopaedic surgeons with subspecialty fellowship training and expertise in adult hip and knee reconstruction (‘experts’) and ten residents of orthopaedic surgery in the last two years of training (‘pre-experts’). They each received 15 radiographs for evaluation. After six weeks they evaluated the same radiographs again but in a different order.

The reliability was assessed using the Kappa and weighted Kappa values.

The Kappa values for inter-observer reliability for the experts and the pre-experts were 0.741 (95% confidence interval (CI) 0.707 to 0.774) and 0.765 (95% CI 0.733 to 0.797), respectively. The weighted Kappa values for intra-observer reliability for the experts and pre-experts were 0.898 (95% CI 0.846 to 0.950) and 0.878 (95% CI 0.815 to 0.942) respectively.

The UCS has substantial inter-observer reliability and ‘near perfect’ intra-observer reliability when used for periprosthetic fractures in association with knee replacement in the hands of experienced and inexperienced users.

Cite this article: Bone Joint J 2014;96-B:1669–73.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 492 - 496
1 Apr 2014
Klatte TO Kendoff D Kamath AF Jonen V Rueger JM Frommelt L Gebauer M Gehrke T

Fungal peri-prosthetic infections of the knee and hip are rare but likely to result in devastating complications. In this study we evaluated the results of their management using a single-stage exchange technique. Between 2001 and 2011, 14 patients (ten hips, four knees) were treated for a peri-prosthetic fungal infection. One patient was excluded because revision surgery was not possible owing to a large acetabular defect. One patient developed a further infection two months post-operatively and was excluded from the analysis. Two patients died of unrelated causes.

After a mean of seven years (3 to 11) a total of ten patients were available for follow-up. One patient, undergoing revision replacement of the hip, had a post-operative dislocation. Another patient, undergoing revision replacement of the knee, developed a wound infection and required revision 29 months post-operatively following a peri-prosthetic femoral fracture.

The mean Harris hip score increased to 74 points (63 to 84; p < 0.02) in those undergoing revision replacement of the hip, and the mean Hospital for Special Surgery knee score increased to 75 points (70 to 80; p < 0.01) in those undergoing revision replacement of the knee.

A single-stage revision following fungal peri-prosthetic infection is feasible, with an acceptable rate of a satisfactory outcome.

Cite this article: Bone Joint J 2014;96-B:492–6.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 263 - 269
1 Feb 2014
Batta V Coathup MJ Parratt MT Pollock RC Aston WJ Cannon SR Skinner JA Briggs TW Blunn GW

We reviewed the outcome of 69 uncemented, custom-made, distal femoral endoprosthetic replacements performed in 69 patients between 1994 and 2006. There were 31 women and 38 men with a mean age at implantation of 16.5 years (5 to 37). All procedures were performed for primary malignant bone tumours of the distal femur. At a mean follow-up of 124.2 months (4 to 212), 53 patients were alive, with one patient lost to follow-up. All nine implants (13.0%) were revised due to aseptic loosening at a mean of 52 months (8 to 91); three implants (4.3%) were revised due to fracture of the shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic fracture. Bone remodelling associated with periosteal cortical thinning adjacent to the uncemented intramedullary stem was seen in 24 patients but this did not predispose to failure. All aseptically loose implants in this series were diagnosed to be loose within the first five years.

The results from this study suggest that custom-made uncemented distal femur replacements have a higher rate of aseptic loosening compared to published results for this design when used with cemented fixation. Loosening of uncemented replacements occurs early indicating that initial fixation of the implant is crucial.

Cite this article: Bone Joint J 2014;96-B:263–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 856 - 861
1 Jun 2010
Emms NW Stockley I Hamer AJ Wilkinson JM

Between 1988 and 1998 we implanted 318 total hip replacements (THRs) in 287 patients using the Plasmacup (B. Braun Ltd, Sheffield, United Kingdom) and a conventional metal-on-polyethylene articulation. The main indications for THR were primary or secondary osteoarthritis.

At follow-up after a mean 11.6 years (7.6 to 18.4) 17 patients had died and 20 could not be traced leaving a final series of 280 THRs in 250 patients. There were 62 revisions (22.1%) in 59 patients. A total of 43 acetabular shells (15.4%) had been revised and 13 (4.6%) had undergone exchange of the liner. The most frequent indications for revision were osteolysis and aseptic loosening, followed by polyethylene wear. The mean Kaplan-Meier survival of the Plasmacup was 91% at ten years and 58% at 14 years. Osteolysis was found around 36 (17.1%) of the 211 surviving shells. The median annual rate of linear wear in the surviving shells was 0.12 mm/year and 0.25 mm/year in those which had been revised (p < 0.001). Polyethylene wear was a strong independent risk factor for osteolysis and aseptic loosening. The percentage of patients with osteolysis increased proportionately with each quintile of wear-rate.

There is a high late rate of failure of the Plasmacup. Patients with the combination of this prosthesis and bearing should be closely monitored after ten years.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 229 - 233
1 Feb 2009
Kempshall PJ Metcalfe A Forster MC

As part of the government’s initiative to reduce waiting times for major joint surgery in Wales, the Cardiff and Vale NHS Trust sent 224 patients (258 knees) to the NHS Treatment Centre in Weston-Super-Mare for total knee replacement. The Kinemax total knee replacement system was used in all cases. The cumulative survival rate at three years was 79.2% (95% confidence interval (CI) 69.2 to 86.8) using re-operation for any cause as an endpoint and 85.3% (95% CI 75.9 to 91.8) using aseptic revision as an endpoint. This is significantly worse than that recorded in the published literature. These poor results have resulted in a significant impact on our service.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 136 - 141
1 Jan 2010
Franken M Grimm B Heyligers I

We have investigated the accuracy of the templating of digital radiographs in planning total hip replacement using two common object-based calibration methods with the ball placed laterally (method 1) or medially (method 2) and compared them with two non-object-based methods. The latter comprised the application of a fixed magnification of 121% (method 3) and calculation of magnification based on the object-film-distance (method 4). We studied the post-operative radiographs of 57 patients (19 men, 38 women, mean age 73 years (53 to 89)) using the measured diameter of the prosthetic femoral head and comparing it with the true value.

Both object-based methods (1 and 2) produced large errors (mean/maximum: 2.55%/17.4% and 2.04%/6.46%, respectively). Method 3 applying a fixed magnification and method 4 (object-film-distance) produced smaller errors (mean/maximum 1.42%/5.22% and 1.57%/4.24%, respectively; p < 0.01). The latter results were clinically relevant and acceptable when planning was allowed to within one implant size. Object-based calibration (methods 1 and 2) has fundamental problems with the correct placement of the calibration ball. The accuracy of the fixed magnification (method 3) matched that of object-film-distance (method 4) and was the most reliable and efficient calibration method in digital templating.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 862 - 866
1 Jun 2010
Hay GC Kampshoff J Kuster MS

The lateral subvastus approach combined with an osteotomy of the tibial tubercle is a recognised, but rarely used approach for total knee replacement (TKR). A total of 32 patients undergoing primary TKR was randomised into two groups, in one of which the lateral subvastus approach combined with a tibial tubercle osteotomy and in the other the medial parapatellar approach were used. The patients were assessed radiologically and clinically using measurement of the range of movement, a visual analogue patient satisfaction score, the Western Ontario McMasters University Osteoarthritis Index and the American Knee Society score. Four patients were lost to the complete follow-up at two years.

At two years there were no significant differences between the groups in any of the parameters for clinical outcome. In the lateral approach group there was one complication due to displacement of the tibial tubercle osteotomy and two osteotomies took more than six months to unite. In the medial approach group, one patient had a partial tear of the quadriceps. There was a significantly greater incidence of lateral patellar subluxation in the medial approach group (3 of 12) compared with the lateral approach group (0 of 16) (p = 0.034), but without any apparent clinical detriment.

We conclude that the lateral approach with tibial tubercle osteotomy is a safe technique with an outcome comparable with that of the medial parapatellar approach for TKR, but the increased surgical time and its specific complications do not support its routine use. It would seem to be more appropriate to reserve this technique for patients in whom problems with patellar tracking are anticipated.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 413 - 418
1 Mar 2010
Rothwell AG Hooper GJ Hobbs A Frampton CM

We analysed data from the Oxford hip and knee questionnaires collected by the New Zealand Joint Registry at six months and five years after joint replacement, to determine if there was any relationship between the scores and the risk of early revision. Logistic regression of the six-month scores indicated that for every one-unit decrease in the Oxford score, the risk of revision within two years increased by 9.7% for total hip replacement (THR), 9.9% for total knee replacement (TKR) and 12.0% for unicompartmental knee replacement (UKR). Our findings showed that 70% of the revisions within two years for TKR and 67% for THR and UKR would have been captured by monitoring the lowest 22%, 28% and 28%, respectively, of the Oxford scores. When analysed using the Kalairajah classification a score of < 27 (poor) was associated with a risk of revision within two years of 7.6% for THR, 7.0% for TKR and 24.3% for UKR, compared with risks of 0.7%, 0.7% and 1.8%, respectively, for scores > 34 (good or excellent).

Our study confirms that the Oxford hip and knee scores at six months are useful predictors of early revision after THR and TKR and we recommend their use for the monitoring of the outcome and potential failure in these patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 123 - 129
1 Jan 2010
Jameson SS Bottle A Malviya A Muller SD Reed MR

The National Institute for Clinical Excellence (NICE) produces recommendations on appropriate treatment within the National Health Service (NHS) in England and Wales. The NICE guidelines on prophylaxis for venous thromboembolism in orthopaedic surgery recommend that all patients be offered a low molecular weight heparin (LMWH). The linked hospital episode statistics of 219 602 patients were examined to determine the rates of complications following lower limb arthroplasty for the 12-month periods prior to and following the publication of these guidelines. These were compared with data from the National Joint Registry (England and Wales) regarding the use of LMWH during the same periods. There was a significant increase in the reported use of LMWH (59.5% to 67.6%, p < 0.001) following the publication of the guidelines. However, the 90-day venous thromboembolism events actually increased slightly following total hip replacement (THR, 1.69% to 1.84%, p = 0.06) and remained unchanged following total knee replacement (TKR, 1.99% to 2.04%). Return to theatre in the first 30 days for infection did not show significant changes. There was an increase in the number of patients diagnosed with thrombocytopenia, which was significant following THR (0.11% to 0.16%, p = 0.04). The recommendations from NICE are based on predicted reductions in venous thromboembolism events, reducing morbidity, mortality and costs to the NHS.

The early results in orthopaedic patients do not support these predictions, but do show an increase in complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1360 - 1365
1 Oct 2009
Sewell MD Spiegelberg BGI Hanna SA Aston WJS Meswania JM Blunn GW Henry C Cannon SR Briggs TWR

We describe the application of a non-invasive extendible endoprosthetic replacement in skeletally-mature patients undergoing revision for failed joint replacement with resultant limb-length inequality after malignant or non-malignant disease. This prosthesis was developed for tumour surgery in skeletally-immature patients but has now been adapted for use in revision procedures to reconstruct the joint or facilitate an arthrodesis, replace bony defects and allow limb length to be restored gradually in the post-operative period.

We record the short-term results in nine patients who have had this procedure after multiple previous reconstructive operations. In six, the initial reconstruction had been performed with either allograft or endoprosthetic replacement for neoplastic disease and in three for non-neoplastic disease. The essential components of the prosthesis are a magnetic disc, a gearbox and a drive screw which allows painless lengthening of the prosthesis using the principle of electromagnetic induction. The mean age of the patients was 37 years (18 to 68) with a mean follow-up of 34 months (12 to 62). They had previously undergone a mean of six (2 to 14) open procedures on the affected limb before revision with the non-invasive extendible endoprosthesis.

The mean length gained was 56 mm (19 to 107) requiring a mean of nine (3 to 20) lengthening episodes performed in the outpatient department. There was one case of recurrent infection after revision of a previously infected implant and one fracture of the prosthesis after a fall. No amputations were performed. Planned exchange of the prosthesis was required in three patients after attainment of the maximum lengthening capacity of the implant. There was no failure of the lengthening mechanism. The Mean Musculoskeletal Tumour Society rating score was 22 of 30 available points (18 to 28).

The use of a non-invasive extendible endoprosthesis in this manner provided patients with good functional results and restoration of leg-length equality, without the need for multiple open lengthening procedures.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1217 - 1222
1 Sep 2009
King RJ Makrides P Gill JA Karthikeyan S Krikler SJ Griffin DR

We have developed a novel method of calculating the radiological magnification of the hip using two separate radio-opaque markers. We recruited 74 patients undergoing radiological assessment following total hip replacement. Both the new double marker and a conventional single marker were used by the radiographer at the time of x-ray. The predicted magnification according to each marker was calculated, as was the true radiological magnification of the components. The correlation between true and predicted magnification was good using the double marker (r = 0.90, n = 74, p < 0.001), but only moderate for the single marker (r = 0.50, n = 63, p < 0.001). The median error was significantly less for the double marker than for the single (1.1% vs 4.8%, p < 0.001). The double marker method demonstrated excellent validity (intraclass correlation coefficient = 0.89), in contrast to the single marker (0.32).

The double marker method appears to be superior to the single marker method when used in the clinical environment.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1623 - 1626
1 Dec 2008
Kulkarni A Partington P Kelly D Muller S

Digital radiography is becoming widespread. Accurate pre-operative templating of digital images of the hip traditionally involves positioning a calibration object at its centre. This can be difficult and cause embarrassment. We have devised a method whereby a planar disc placed on the radiographic cassette accounts for the expected magnification. Initial examination of 50 pelvic CT scans showed a mean hip centre distance of 117 mm (79 to 142) above the gluteal skin. Further calculations predicted that a disc of 37.17 mm diameter, placed on the cassette, would appear identical to a 30 mm sphere placed at the level of the centre of the hip as requested by our templating software. We assessed accuracy and reproducibility by ‘reverse calibration’ of 20 radiographs taken three months after hip replacement using simultaneous sphere and disc methods, and a further 20 with a precision disc of accurate size. Even when variations in patient size were ignored, the disc proved more accurate and reliable than the sphere.

The technique is reliable, robust, cost effective and acceptable to patients and radiographers. It can easily be used in any radiography department after a few simple calculations and manufacture of appropriately-sized discs.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 928 - 934
1 Jul 2009
Palan J Gulati A Andrew JG Murray DW Beard DJ

Balancing service provision and surgical training is a challenging issue that affects all healthcare systems. A multicentre prospective study of 1501 total hip replacements was undertaken to investigate whether there is an association between surgical outcome and the grade of the operating surgeon, and whether there is any difference in outcome if surgeons’ assistants assist with the operation, rather than orthopaedic trainees. The primary outcome measure was the change in the Oxford hip score (OHS) at five years. Secondary outcomes included the rate of revision and dislocation, operating time, and length of hospital stay.

There was no significant difference in ΔOHS or complication rates between operations undertaken by trainers and trainees, or those at which surgeons’ assistants and trainees were the assistant. However, there was a significant difference in the duration of surgery, with a mean reduction of 28 minutes in those in which a surgeons’ assistant was the assistant.

This study provides evidence that total hip replacements can be performed safely and effectively by appropriately trained surgeons in training, and that there are potential benefits of using surgeons’ assistants in orthopaedic surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1524 - 1526
1 Nov 2005
Kwong FNK Ibrahim T Power RA

Implantation of allograft bone is an integral part of revision surgery of the hip. One major concern with its use is the risk of transmission of infective agents. There are a number of methods of processing allograft bone in order to reduce this risk. One method requires washing the tissue using pulsed irrigation immediately before implantation. We report the incidence of deep bacterial infection in 138 patients (144 revision hip arthroplasties) who had undergone implantation of allograft bone. The bone used was fresh-frozen, non-irradiated and pulse-washed with normal saline before implantation. The deep infection rate at a minimum follow-up of one year was 0.7%. This method of processing appears to be associated with a very low risk of allograft-related bacterial infection.