Aims. Magnetically controlled growing rods (MCGR) have been gaining popularity in the management of early-onset scoliosis (EOS) over the past decade. We present our experience with the first 44
Magnetically controlled growing rod (MCGR) systems use non-invasive
spinal lengthening for the surgical treatment of early-onset scoliosis
(EOS). The primary aim of this study was to evaluate the performance
of these devices in the prevention of progression of the deformity.
A secondary aim was to record the rate of complications. An observational study of 31 consecutive children with EOS, of
whom 15 were male, who were treated between December 2011 and October
2017 was undertaken. Their mean age was 7.7 years (2 to 14). The
mean follow-up was 47 months (24 to 69). Distractions were completed
using the tailgating technique. The primary outcome measure was
correction of the radiographic deformity. Secondary outcomes were
growth, functional outcomes and complication rates.Aims
Patients and Methods
MAGnetic Expansion Control (MAGEC) rods are used in the surgical treatment of children with early onset scoliosis. The magnetically controlled lengthening mechanism enables rod distractions without the need for repeated invasive surgery. The CE certification of these devices was suspended in March 2021 due, primarily, to performance evidence gaps in the documents provided by the manufacturer to regulators and notified bodies. MAGEC rods are therefore not permitted for use in countries requiring CE marking. This was a survey of 18 MAGEC rod surgeons in the UK about their perception of the impact of the CE suspension on the clinical management of their patients. Unsurprisingly, virtually all perceived a negative impact, reflecting the complexity of this patient group. Reassuringly, these surgeons are highly experienced in alternative treatment methods. Cite this article: Abstract
Aims. The aim of this study was to compare the clinical and radiological outcomes of patients with early-onset scoliosis (EOS), who had undergone spinal fusion after distraction-based spinal growth modulation using either traditional growing rods (TGRs) or magnetically controlled growing rods (MCGRs). Methods. We undertook a retrospective review of skeletally mature patients who had undergone fusion for an EOS, which had been previously treated using either TGRs or MCGRs. Measured outcomes included sequential coronal T1 to S1 height and major curve (Cobb) angle on plain radiographs and any complications requiring unplanned surgery before final fusion. Results. We reviewed 43 patients (63% female) with a mean age of 6.4 years (SD 2.6) at the index procedure, and 12.2 years (SD 2.2) at final fusion. Their mean follow-up was 8.1 years (SD 3.4). A total of 16 patients were treated with MCGRs and 27 with TGRs. The mean number of distractions was 7.5 in the
Aims. To investigate metallosis in patients with magnetically controlled growing rods (MCGRs) and characterize the metal particle profile of the tissues surrounding the rod. Methods. This was a prospective observational study of patients with early onset scoliosis (EOS) treated with MCGRs and undergoing rod exchange who were consecutively recruited between February 2019 and January 2020. Ten patients were recruited (mean age 12 years (SD 1.3); 2 M:8 F). The configurations of the
Aims. We present a case series of five patients who had revision surgery
following magnetic controlled growing rods (MGCR) for early onset
scoliosis. Metallosis was found during revision in four out of five
patients and we postulated a mechanism for rod failure based on
retrieval analysis. Patients and Methods. Retrieval analysis was performed on the seven explanted rods.
The mean duration of
Aims. Magnetically controlled growing rods (MCGRs) allow non-invasive
correction of the spinal deformity in the treatment of early-onset
scoliosis. Conventional growing rod systems (CGRS) need repeated
surgical distractions: these are associated with the effect of the
‘law of diminishing returns’. The primary aim of this study was to quantify this effect in
MCGRs over sequential distractions. . Patients and Methods. A total of 35 patients with a maximum follow-up of 57 months
were included in the study. There were 17 boys and 18 girls with
a mean age of 7.4 years (2 to 14). True Distraction (TD) was determined
by measuring the expansion gap on fluoroscopy. This was compared
with Intended Distraction (ID) and expressed as the ‘T/I’ ratio.
The T/I ratio and the Cobb angle were calculated at several time
points during follow-up. Results. The mean follow-up was 30 months (6 to 57). There was a significant
decrease in the mean T/I ratio over time (convex rod at 3 months
0.81, . sd. 0.58 vs 51 months 0.17, . sd . 0.16,
p = 0.0001; concave rod at 3 months 0.93, . sd. 0.67 vs 51
months 0.18, . sd. 0.15, p = 0.0001). A linear decline of
the mean T/I ratios was noted for both convex rods (r. 2. =
0.90, p = 0.004) and concave rods (r. 2. = 0.81, p = 0.015)
over 51 months. At the 24-month follow-up stage, there was a significant
negative correlation between the mean T/I ratio of the concave rod
with weight (r = -0.59, p = 0.01), age (r = -0.59, p = 0.01), and
BMI of the child (r = -0.54,
p = 0.01). Conclusions. The ‘law of diminishing returns’ is also seen after serial distraction
using
Conventional growing rods are the most commonly
used distraction-based devices in the treatment of progressive early-onset
scoliosis. This technique requires repeated lengthenings with the
patient anaesthetised in the operating theatre. We describe the
outcomes and complications of using a non-invasive magnetically
controlled growing rod (MCGR) in children with early-onset scoliosis.
Lengthening is performed on an outpatient basis using an external remote
control with the patient awake. Between November 2009 and March 2011, 34 children with a mean
age of eight years (5 to 12) underwent treatment. The mean length
of follow-up was 15 months (12 to 18). In total, 22 children were
treated with dual rod constructs and 12 with a single rod. The mean
number of distractions per patient was 4.8 (3 to 6). The mean pre-operative
Cobb angle was 69° (46° to 108°); this was corrected to a mean 47°
(28° to 91°) post-operatively. The mean Cobb angle at final review
was 41° (27° to 86°). The mean pre-operative distance from T1 to
S1 was 304 mm (243 to 380) and increased to 335 mm (253 to 400)
in the immediate post-operative period. At final review the mean distance
from T1 to S1 had increased to 348 mm (260 to 420). Two patients developed a superficial wound infection and a further
two patients in the single rod group developed a loss of distraction.
In the dual rod group, one patient had pull-out of a hook and one
developed prominent metalwork. Two patients had a rod breakage;
one patient in the single rod group and one patient in the dual
rod group. Our early results show that the
The aim of this study was to compare outcomes after growth-friendly treatment for early-onset scoliosis (EOS) between patients with skeletal dysplasias versus those with other syndromes. We retrospectively identified 20 patients with skeletal dysplasias and 292 with other syndromes (control group) who had completed surgical growth-friendly EOS treatment between 1 January 2000 and 31 December 2018. We compared radiological parameters, complications, and health-related quality of life (HRQoL) at mean follow-up of 8.6 years (SD 3.3) in the dysplasia group and 6.6 years (SD 2.6) in the control group.Aims
Methods
To report the mid-term results of a modified self-growing rod (SGR) technique for the treatment of idiopathic and neuromuscular early-onset scoliosis (EOS). We carried out a retrospective analysis of 16 consecutive patients with EOS treated with an SGR construct at a single hospital between September 2008 and December 2014. General demographics and deformity variables (i.e. major Cobb angle, T1 to T12 length, T1 to S1 length, pelvic obliquity, shoulder obliquity, and C7 plumb line) were recorded preoperatively, and postoperatively at yearly follow-up. Complications and revision procedures were also recorded. Only patients with a minimum follow-up of five years after surgery were included.Aims
Methods
We undertook a prospective non-randomised radiological study
to evaluate the preliminary results of using magnetically-controlled
growing rods (MAGEC System, Ellipse technology) to treat children
with early-onset scoliosis. Between January 2011 and January 2015, 19 children were treated
with magnetically-controlled growing rods (MCGRs) and underwent
distraction at three-monthly intervals. The mean age of our cohort
was 9.1 years (4 to 14) and the mean follow-up 22.4 months (5.1
to 35.2). Of the 19 children, eight underwent conversion from traditional growing
rods. Whole spine radiographs were carried out pre- and post-operatively:
image intensification was used during each lengthening in the outpatient
department. The measurements evaluated were Cobb angle, thoracic kyphosis,
proximal junctional kyphosis and spinal growth from T1 to S1.Aims
Patients and Methods
The primary aim of this study was to evaluate the performance
and safety of magnetically controlled growth rods in the treatment
of early onset scoliosis. Secondary aims were to evaluate the clinical
outcome, the rate of further surgery, the rate of complications,
and the durability of correction. We undertook an observational prospective cohort study of children
with early onset scoliosis, who were recruited over a one-year period
and followed up for a minimum of two years. Magnetically controlled
rods were introduced in a standardized manner with distractions
performed three-monthly thereafter. Adverse events which were both related
and unrelated to the device were recorded. Ten children, for whom
relevant key data points (such as demographic information, growth
parameters, Cobb angles, and functional outcomes) were available,
were recruited and followed up over the period of the study. There
were five boys and five girls. Their mean age was 6.2 years (2.5
to 10).Aims
Patients and Methods