Aims. The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes. Methods. A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service.
Aims. The aim of this study was to evaluate medium-term outcomes and complications of the S-ROM NOILES Rotating Hinge Knee System (DePuy, USA) in revision total knee arthroplasty (rTKA) at a tertiary unit. Methods. A retrospective consecutive study of all patients who underwent a rTKA using this implant from January 2005 to December 2018. Outcome measures included reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service.
Aims. Debate continues regarding the optimum management of periprosthetic distal femoral fractures (PDFFs). This study aims to determine which operative treatment is associated with the lowest perioperative morbidity and mortality when treating low (Su type II and III) PDFFs comparing lateral locking plate fixation (LLP-ORIF) or distal femoral arthroplasty (DFA). Methods. This was a retrospective cohort study of 60 consecutive unilateral (PDFFs) of Su types II (40/60) and III (20/60) in patients aged ≥ 60 years: 33 underwent LLP-ORIF (mean age 81.3 years (SD 10.5), BMI 26.7 (SD 5.5); 29/33 female); and 27 underwent DFA (mean age 78.8 years (SD 8.3); BMI 26.7 (SD 6.6); 19/27 female). The primary outcome measure was reoperation. Secondary outcomes included perioperative complications, calculated blood loss, transfusion requirements, functional mobility status, length of acute hospital stay, discharge destination and mortality.
Total knee arthroplasty (TKA) with a highly congruent condylar-stabilized (CS) articulation may be advantageous due to increased stability versus cruciate-retaining (CR) designs, while mitigating the limitations of a posterior-stabilized construct. The aim was to assess ten-year implant survival and functional outcomes of a cemented single-radius TKA with a CS insert, performed without posterior cruciate ligament sacrifice. This retrospective cohort study included consecutive patients undergoing TKA at a specialist centre in the UK between November 2010 and December 2012. Data were collected using a bespoke electronic database and cross-referenced with national arthroplasty audit data, with variables including: preoperative characteristics, intraoperative factors, complications, and mortality status. Patient-reported outcome measures (PROMs) were collected by a specialist research team at ten years post-surgery. There were 536 TKAs, of which 308/536 (57.5%) were in female patients. The mean age was 69.0 years (95% CI 45.0 to 88.0), the mean BMI was 32.2 kg/m2 (95% CI 18.9 to 50.2), and 387/536 (72.2%) survived to ten years. There were four revisions (0.7%): two deep infections (requiring debridement and implant retention), one aseptic loosening, and one haemosiderosis.Aims
Methods
The purpose of this study was to compare the clinical outcomes, mortalities, implant survival rates, and complications of total knee arthroplasty (TKA) in patients with or without hepatitis B virus (HBV) infection over at least ten years of follow-up. From January 2008 to December 2010, 266 TKAs were performed in 169 patients with HBV (HBV group). A total of 169 propensity score–matched patients without HBV were chosen for the control group in a one-to-one ratio. Then, the clinical outcomes, mortalities, implant survival rates, and complications of TKA in the two groups were compared. The mean follow-up periods were 11.7 years (10.5 to 13.4) in the HBV group and 11.8 years (11.5 to 12.4) in the control group.Aims
Methods
Knowledge on total knee arthroplasties (TKAs) in patients with a history of poliomyelitis is limited. This study compared implant survivorship and clinical outcomes among affected and unaffected limbs in patients with sequelae of poliomyelitis undergoing TKAs. A retrospective review of our total joint registry identified 94 patients with post-polio syndrome undergoing 116 primary TKAs between January 2000 and December 2019. The mean age was 70 years (33 to 86) with 56% males (n = 65) and a mean BMI of 31 kg/m2 (18 to 49). Rotating hinge TKAs were used in 14 of 63 affected limbs (22%), but not in any of the 53 unaffected limbs. Kaplan-Meier survivorship analyses were completed. The mean follow-up was eight years (2 to 19).Aims
Methods
The aim of this study was to determine whether obesity had a detrimental effect on the long-term performance and survival of medial unicompartmental knee arthroplasties (UKAs). This study reviewed prospectively collected functional outcome scores and revision rates of all medial UKA patients with recorded BMI performed in Christchurch, New Zealand, from January 2011 to September 2021. Patient-reported outcome measures (PROMs) were the primary outcome of this study, with all-cause revision rate analyzed as a secondary outcome. PROMs were taken preoperatively, at six months, one year, five years, and ten years postoperatively. There were 873 patients who had functional scores recorded at five years and 164 patients had scores recorded at ten years. Further sub-group analysis was performed based on the patient’s BMI. Revision data were available through the New Zealand Joint Registry for 2,323 UKAs performed during this time period.Aims
Methods
This prospective study reports longitudinal, within-patient, patient-reported outcome measures (PROMs) over a 15-year period following cemented single radius total knee arthroplasty (TKA). Secondary aims included reporting PROMs trajectory, 15-year implant survival, and patient attrition from follow-up. From 2006 to 2007, 462 consecutive cemented cruciate-retaining Triathlon TKAs were implanted in 426 patients (mean age 69 years (21 to 89); 290 (62.7%) female). PROMs (12-item Short Form Survey (SF-12), Oxford Knee Score (OKS), and satisfaction) were assessed preoperatively and at one, five, ten, and 15 years. Kaplan-Meier survival and univariate analysis were performed.Aims
Methods
Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR. A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by >
3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening.
Recent total knee arthroplasty (TKA) designs have featured more anatomical morphologies and shorter tibial keels. However, several reports have raised concerns about the impact of these modifications on implant longevity. The aim of this study was to report the early performance of a modern, cemented TKA design. All patients who received a primary, cemented TKA between 2012 and 2017 with a minimum two-year follow-up were included. The implant investigated features an asymmetrical tibial baseplate and shortened keel. Patient demographic details, Knee Society Scores (KSS), component alignment, and the presence of radiolucent lines at final follow-up were recorded. Kaplan-Meier analyses were performed to estimate survivorship.Aims
Methods
Periprosthetic joint infection (PJI) is a devastating complication following total knee arthroplasty (TKA). Two-stage revision has traditionally been considered the gold standard of treatment for established infection, but increasing evidence is emerging in support of one-stage exchange for selected patients. The objective of this study was to determine the outcomes of single-stage revision TKA for PJI, with mid-term follow-up. A total of 84 patients, with a mean age of 68 years (36 to 92), underwent single-stage revision TKA for confirmed PJI at a single institution between 2006 and 2016. In all, 37 patients (44%) were treated for an infected primary TKA, while the majority presented with infected revisions: 31 had undergone one previous revision (36.9%) and 16 had multiple prior revisions (19.1%). Contraindications to single-stage exchange included systemic sepsis, extensive bone or soft-tissue loss, extensor mechanism failure, or if primary wound closure was unlikely to be achievable. Patients were not excluded for culture-negative PJI or the presence of a sinus.Aims
Methods
Tibial cones are often utilized in revision total knee arthroplasty (TKA) with metaphyseal defects. Because there are few studies evaluating mid-term outcomes with a sufficient cohort, the purpose of this study was to evaluate tibial cone survival and complications in revision TKAs with tibial cones at minimum follow-up of five years. A retrospective review was completed from September 2006 to March 2015, evaluating 67 revision TKAs (64 patients) that received one specific porous tibial cone during revision TKA. The final cohort was composed of 62 knees (59 patients) with five years of clinical follow-up or reoperation. The mean clinical follow-up of the TKAs with minimum five-year clinical follow-up was 7.6 years (5.0 to 13.3). Survivorship analysis was performed with the endpoints of tibial cone revision for aseptic loosening, tibial cone revision for any reason, and reoperation. We also evaluated periprosthetic joint infection (PJI), risk factors for failure, and performed a radiological review.Aims
Methods
This study aimed to identify the tibial component and femoral component coronal angles (TCCAs and FCCAs), which concomitantly are associated with the best outcomes and survivorship in a cohort of fixed-bearing, cemented, medial unicompartmental knee arthroplasties (UKAs). We also investigated the potential two-way interactions between the TCCA and FCCA. Prospectively collected registry data involving 264 UKAs from a single institution were analyzed. The TCCAs and FCCAs were measured on postoperative radiographs and absolute angles were analyzed. Clinical assessment at six months, two years, and ten years was undertaken using the Knee Society Knee score (KSKS) and Knee Society Function score (KSFS), the Oxford Knee Score (OKS), the 36-Item Short-Form Health Survey questionnaire (SF-36), and range of motion (ROM). Fulfilment of expectations and satisfaction was also recorded. Implant survivorship was reviewed at a mean follow-up of 14 years (12 to 16). Multivariate regression models included covariates, TCCA, FCCA, and two-way interactions between them. Partial residual graphs were generated to identify angles associated with the best outcomes. Kaplan-Meier analysis was used to compare implant survivorship between groups.Aims
Methods
To report mid- to long-term results of Oxford mobile bearing domed lateral unicompartmental knee arthroplasty (UKA), and determine the effect of potential contraindications on outcome. A total of 325 consecutive domed lateral UKAs undertaken for the recommended indications were included, and their functional and survival outcomes were assessed. The effects of age, weight, activity, and the presence of full-thickness erosions of cartilage in the patellofemoral joint on outcome were evaluated.Aims
Methods
The aim of this independent multicentre study was to assess the
mid-term results of mobile bearing unicondylar knee arthroplasty
(UKA) for isolated lateral osteoarthritis of the knee joint. We retrospectively evaluated 363 consecutive, lateral UKAs (346
patients) performed using the Oxford domed lateral prosthesis undertaken
in three high-volume knee arthroplasty centres between 2006 and
2014. Mean age of the patients at surgery was 65 years (36 to 88)
with a mean final follow-up of 37 months (12 to 93)Aims
Patients and Methods
We describe a cohort of patients with a high rate of mid-term failure following Kinemax Plus total knee replacement inserted between 1998 and 2001. This implant has been recorded as having a survival rate of 96% at ten years. However, in our series the survival rate was 75% at nine years. This was also significantly lower than that of subsequent consecutive series of PFC Sigma knee replacements performed by the same surgeon. No differences were found in the clinical and radiological parameters between the two groups. At revision the most striking finding was polyethylene wear. An independent analysis of the polyethylene components was therefore undertaken. Scanning electron microscopy revealed type 2 fusion defects in the ultra-high molecular weight polyethylene (UHMWPE), which indicated incomplete boundary fusion. Other abnormalities consistent with weak UHMWPE particle interface strength were present in both the explanted inserts and in unused inserts from the same period. We consider that these type 2 fusion defects are the cause of the early failure of the Kinemax implants. This may represent a manufacturing defect resulting in a form of programmed polyethylene failure.
The low contact stress patellofemoral replacement consists of a trochlear component and a modular patellar component which has a metal-backed mobile polyethylene bearing. We present the early results of the use of this prosthesis for established isolated patellofemoral arthritis in 51 consecutive patellofemoral replacements in 35 patients. The mean follow-up was 25 months (5 to 60). The estimated survival rate at three years was 63% (95% confidence interval 47 to 80) with revision as the endpoint and 46% (95% confidence interval 30 to 63) with revision and ongoing moderate or severe pain as the endpoint. The early results of the use of the low contact stress patellofemoral replacement are disappointing with a high rate of revision. We cannot therefore recommend its use.
We describe 119 meniscal allograft transplantations performed concurrently with articular cartilage repair in 115 patients with severe articular cartilage damage. In all, 53 (46.1%) of the patients were over the age of 50 at the time of surgery. The mean follow-up was for 5.8 years (2 months to 12.3 years), with 25 procedures (20.1%) failing at a mean of 4.6 years (2 months to 10.4 years). Of these, 18 progressed to knee replacement at a mean of 5.1 years (1.3 to 10.4). The Kaplan-Meier estimated mean survival time for the whole series was 9.9 years ( The survival of the transplant was not affected by gender, the severity of cartilage damage, axial alignment, the degree of narrowing of the joint space or medial
We evaluated the rates of survival and cause
of revision of seven different brands of cemented primary total
knee replacement (TKR) in the Norwegian Arthroplasty Register during
the years 1994 to 2009. Revision for any cause, including resurfacing
of the patella, was the primary endpoint. Specific causes of revision
were secondary outcomes. Three posterior cruciate-retaining (PCR) fixed modular-bearing
TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing
posterior cruciate-sacrificing TKRs were investigated in a total
of 17 782 primary TKRs. The median follow-up for the implants ranged
from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from
89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative
to the fixed modular-bearing Profix knee (the most frequently used
TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision
for aseptic tibial loosening was higher in the mobile-bearing LCS
Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the
LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing
Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular
bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared
with the Profix. These implants (except AGC Universal) also had
an increased risk of revision for femoral loosening (RR 2.3
(95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95%
CI 1.1 to 11.0), respectively). These results suggest that aseptic
loosening is related to design in TKR. Cite this article:
Despite many claims of good wear properties following
total knee replacement (TKR) with an oxidised zirconium (OxZr) femoral
component, there are conflicting clinical results. We hypothesised
that there would be no difference in either the mid-term clinical
and radiological outcomes or the characteristics of the polyethylene
wear particles (weight, size and shape) in patients using an OxZr
or cobalt-chrome (CoCr) femoral component. In all 331 patients underwent
bilateral TKR, receiving an OxZr femoral component in one knee and
a CoCr femoral component in the other. The mean follow-up was 7.5
years (6 to 8). Following aspiration, polyethylene wear particles
were analysed using thermogravimetric methods and scanning electron
microscopy. At the most recent follow-up, the mean Knee Society
score, Western Ontario and McMaster Universities Osteoarthritis
Index score, range of movement and satisfaction score were not significantly
different in the two groups. The mean weight, size, aspect ratio
and roundness of the aspirated wear particles were similar for each
femoral component. Survivorship of the femoral, tibial and patellar
components was 100% in both groups. In the absence of evidence of an advantage in the medium term
we cannot justify the additional expense of an OxZr femoral component.
We evaluated 535 consecutive primary cementless total knee replacements (TKR). The mean follow-up was 9.2 years (0.3 to 12.9) and information on implant survival was available for all patients. Patients were divided into two groups: 153 obese patients (BMI ≥ 30) and 382 non-obese (BMI <
30). A case-matched study was performed on the clinical and radiological outcome, comparing 50 knees in each group. We found significantly lower mean improvements in the clinical score (p = 0.044) and lower post-operative total clinical scores in the obese group (p = 0.041). There was no difference in the rate of radiological osteolysis or lucent lines, and no difference in alignment. Log rank test for survival showed no significant differences between the groups (p = 0.167), with a ten-year survival rate of 96.4% (95% confidence interval (CI) 92 to 99) in the obese and 98% (95% CI 96 to 99) in the non-obese. The mid-term survival of TKR in the obese and the non-obese are comparable, but obesity appears to have a negative effect on the clinical outcome. However, good results and high patient satisfaction are still to be expected, and it would seem unreasonable to deny patients a TKR simply on the basis of a BMI indicating obesity.
The long-term success of total knee replacement is multifactorial, including factors relating to the patient, the operation and the implant. The purpose of this study was to examine the 20-year survival of the cemented Anatomical Graduated Component (AGC) total knee replacement. Between 1983 and 2004, 7760 of these were carried out at our institution. Of these, 6726 knees which received the non-modular metal-backed tibial component with compression-moulded polyethylene and had a minimum two-year follow-up were available for study. In all, 36 knees were followed over 20 years with a survival of the tibial and femoral components together of 97.8% (95% confidence interval (CI) 0.9851 to 0.9677), with no implants being revised for polyethylene wear or osteolysis. Age >
70 was associated with increased survival (99.6%, 95% CI 99.0 to 99.8) (p <
0.0001) but pre-operative valgus alignment reduced survival (95.1%, 95% CI 90.0 to 97.6) (p = 0.0056). Age <
55 (p = 0.129), pre-operative varus alignment (p = 0.707), osteonecrosis (p = 0.06), rheumatoid arthritis (p = 0.247), and gender (p = 0.666) were not statistically associated with failure. We attribute the success of the AGC implant to its relatively unconstrained articular geometry and the durability of a non-modular metal-backed tibial component with compression moulded polyethylene.
Between 1989 and 1992 we had 102 knees suitable for unicompartmental knee replacement (UKR). They were randomised to receive either a St Georg Sled UKR or a Kinematic modular total knee replacement (TKR). The early results demonstrated that the UKR group had less complications and more rapid rehabilitation than the TKR group. At five years there were an equal number of failures in the two groups but the UKR group had more excellent results and a greater range of movement. The cases were reviewed by a research nurse at 8, 10 and 12 years after operation. We report the outcome at 15 years follow-up. A total of 43 patients (45 knees) died with their prosthetic knees intact. Throughout the review period the Bristol knee scores of the UKR group have been better and at 15 years 15 (71.4%) of the surviving UKRs and 10 (52.6%) of the surviving TKRs had achieved an excellent score. The 15 years survivorship rate based on revision or failure for any reason was 24 (89.8%) for UKR and 19 (78.7%) for TKR. During the 15 years of the review four UKRs and six TKRs failed. The better early results with UKR are maintained at 15 years with no greater failure rate. The median Bristol knee score of the UKR group was 91.1 at five years and 92 at 15 years, suggesting little functional deterioration in either the prosthesis or the remainder of the joint. These results justify the increased use of UKR.
We carried out a prospective study of 132 patients (159 knees) who underwent closed-wedge high tibial osteotomy for severe medial compartment osteoarthritis between 1988 and 1997. A total of 94 patients (118 knees) was available for review at a mean of 16.4 years (16 to 20). Seven patients (7.4%) (11 knees) required conversion to total knee replacement. Kaplan-Meier survival was 97.6% (95% confidence interval 95.0 to 100) at ten years and 90.4% (95% confidence interval 84.1 to 96.7) at 15 years. Excellent and good results as assessed by the Hospital for Special Surgery knee score were achieved in 87 knees (73.7%). A pre-operative body mass index >
27.5 kg/m2 and range of movement <
100° were risk factors predicting early failure. Although our long-term results were satisfactory, strict indications for osteotomy are required if long-term survival is required.
We wished to determine whether simultaneous bilateral sequential total knee replacement (TKR) carried increased rates of mortality and complications compared with unilateral TKR in low- and high-risk patients. Our study included 2385 patients who had undergone bilateral sequential TKR under one anaesthetic and 719 who had unilateral TKR. There were no significant pre-operative differences between the groups in terms of age, gender, height, weight, body mass index, diagnosis, comorbidity and duration of follow-up, which was a mean of 10.2 years (5 to 14) in the bilateral and 10.4 years (5 to 14) in the unilateral group. The peri-operative mortality rate (eight patients, 0.3%) of patients who had bilateral sequential TKR was similar to that (five patients, 0.7%) of those undergoing unilateral TKR. In bilateral cases the peri-operative mortality rate (three patients, 0.4%) of patients at high risk was similar to that (five patients, 0.3%) of patients at low risk as it was also in unilateral cases (two patients, 1.0% vs three patients, 0.6%). There was no significant difference (p = 0.735) in either the overall number of major complications between bilateral and unilateral cases or between low- (p = 0.57) and high-risk (p = 0.61) patients. Also, the overall number of minor complications was not significantly different between the bilateral and unilateral group (p = 0.143). Simultaneous bilateral sequential TKR can be offered to patients at low and high risk and has an expected rate of complications similar to that of unilateral TKR.
We prospectively reviewed 1000 consecutive patients who underwent a cementless, hydroxyapatite-coated, stemless, total knee replacement over a period of nine years. Regular post-operative clinical follow-up was performed using the Knee Society score. The mean pre-operative score was 96, improving to 182 and 180 at five and ten years, respectively. To date, there have been seven (0.5%) cases which required revision, primarily for septic loosening (four cases), with low rates of other post-operative complications. The cumulative survival at ten years with revision as the end-point, was 99.14% (95% confidence interval 92.5 to 99.8). These results support the use of hydroxyapatite in a cementless total knee replacement since it can give reliable fixation with an excellent clinical and functional outcome.