Aims. The aim of this study was to estimate the clinical and economic burden of dislocation following primary total hip arthroplasty (THA) in
Aims. This study describes the variation in the annual volumes of revision hip arthroplasty (RHA) undertaken by consultant surgeons nationally, and the rate of accrual of RHA and corresponding primary hip arthroplasty (PHA) volume for new consultants entering practice. Methods. National Joint Registry (NJR) data for
Using the Trent regional arthroplasty register, we analysed the survival at ten years of 1198 consecutive Charnley total hip replacements carried out across a single health region of the United Kingdom in 1990. At ten years, information regarding outcome was available for 1001 hips (83.6%). The crude revision rate was 6.2% (62 of 1001) and the cumulative survival rate with revision of the components as an end-point was 93.1%. At five years, a review of this series of patients identified gross radiological failure in 25 total hip replacements which had previously been unrecognised. At ten years the outcome was known for 18 of these 25 patients (72%), of whom 13 had not undergone revision. This is the first study to assess the survival at ten years for the primary Charnley total hip replacement performed in a broad cross-section of hospitals in the United Kingdom, as opposed to specialist centres. Our results highlight the importance of the arthroplasty register in identifying the long-term outcome of hip prostheses.
Using a regional arthroplasty register we assessed the outcome at five years of 1198 primary Charnley total hip replacements (THRs) carried out in 1152 patients across a single UK health region in 1990. Information regarding outcome was available for 1080 hips (90%) and 499 had an independent clinical and radiological assessment. By five years the known rate of aseptic loosening was 2.3%, of deep infection 1.4%, of dislocation 5.0% and of revision 3.2%. The radiological assessment of 499 THRs revealed gross failure in a further 5.2%, which had been previously unrecognised. The combined rate of failure of nearly 9% is higher than those published from specialist centres and surgeons, but is probably more representative of the norm. Our study supports the need for a national register and surveillance of THRs. It emphasises that all implants should be followed, and suggests that the results of such surgery, when performed in the general setting, may not be as good as expected.
Aims. To develop and externally validate a parsimonious statistical prediction model of 90-day mortality after elective total hip arthroplasty (THA), and to provide a web calculator for clinical usage. Methods. We included 53,099 patients with cemented THA due to osteoarthritis from the Swedish Hip Arthroplasty Registry for model derivation and internal validation, as well as 125,428 patients from
Aims. There are reports of a marked increase in perioperative mortality in patients admitted to hospital with a fractured hip during the COVID-19 pandemic in the UK, USA, Spain, and Italy. Our study aims to describe the risk of mortality among patients with a fractured neck of femur in
Aims. Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA. Methods. We performed an observational study using National Joint Registry (NJR) data. We included adult patients who underwent primary THA for osteoarthritis, recorded in the NJR between 2003 and 2016. Exposures were operating surgeon grade (consultant or trainee) and whether or not trainees were directly supervised by a scrubbed consultant. Outcomes were all-cause revision and the indication for revision up to ten years. We used methods of survival analysis, adjusted for patient, operation, and healthcare setting factors. Results. We included 603,474 THAs, of which 58,137 (9.6%) procedures were performed by a trainee. There was no association between surgeon grade and all-cause revision up to ten years (crude hazard ratio (HR) 1.00 (95% confidence interval (CI) 0.94 to 1.07); p = 0.966), a finding which persisted with adjusted analysis. Fully adjusted analysis demonstrated an association between trainees operating without scrubbed consultant supervision and an increased risk of all-cause revision (HR 1.10 (95% CI 1.00 to 1.21); p = 0.045). There was an association between trainee-performed THA and revision for instability (HR 1.14 (95% CI 1.01 to 1.30); p = 0.039). However, this was not observed in adjusted models, or when trainees were supervised by a scrubbed consultant. Conclusion. Within the current training system in
Aims. This study aims to assess the feasibility of conducting a pragmatic, multicentre randomized controlled trial (RCT) to test the clinical and cost-effectiveness of an informal caregiver training programme to support the recovery of people following hip fracture surgery. Methods. This will be a mixed-methods feasibility RCT, recruiting 60 patients following hip fracture surgery and their informal caregivers. Patients will be randomized to usual NHS care, versus usual NHS care plus a caregiver-patient dyad training programme (HIP HELPER). This programme will comprise of three, one-hour, one-to-one training sessions for the patient and caregiver, delivered by a nurse, physiotherapist, or occupational therapist. Training will be delivered in the hospital setting pre-patient discharge. It will include practical skills for rehabilitation such as: transfers and walking; recovery goal setting and expectations; pacing and stress management techniques; and introduction to the HIP HELPER Caregiver Workbook, which provides information on recovery, exercises, worksheets, and goal-setting plans to facilitate a ‘good’ recovery. After discharge, patients and caregivers will be supported in delivering rehabilitation through three telephone coaching sessions. Data, collected at baseline and four months post-randomization, will include: screening logs, intervention logs, fidelity checklists, quality assurance monitoring visit data, and clinical outcomes assessing quality of life, physical, emotional, adverse events, and resource use outcomes. The acceptability of the study intervention and RCT design will be explored through qualitative methods with 20 participants (patients and informal caregivers) and 12 health professionals. Discussion. A multicentre recruitment approach will provide greater external validity across population characteristics in
Aims. “Get It Right First Time” (GIRFT) and NHS
Aims. The aim of this study was to investigate whether the use of antibiotic-loaded bone cement influenced the risk of revision surgery after primary total hip arthroplasty (THA) for osteoarthritis. Methods. The study involved data collected by the National Joint Registry (NJR) for
Aims. To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). Methods. In this registry study, data from the National Joint Registry of
Aims. The rate of dislocation when traditional single bearing implants are used in revision total hip arthroplasty (THA) has been reported to be between 8% and 10%. The use of dual mobility bearings can reduce this risk to between 0.5% and 2%. Dual mobility bearings are more expensive, and it is not clear if the additional clinical benefits constitute value for money for the payers. We aimed to estimate the cost-effectiveness of dual mobility compared with single bearings for patients undergoing revision THA. Methods. We developed a Markov model to estimate the expected cost and benefits of dual mobility compared with single bearing implants in patients undergoing revision THA. The rates of revision and further revision were calculated from the National Joint Registry of
Aims. The aim of this study was to describe temporal trends and survivorship of total hip arthroplasty (THA) in very young patients, aged ≤ 20 years. Patients and Methods. A descriptive observational study was undertaken using data from the National Joint Registry (NJR) for
The popularity of cementless total hip replacement
(THR) has surpassed cemented THR in
Objectives. The National Hip Fracture Database (NHFD) publishes hospital-level risk-adjusted mortality rates following hip fracture surgery in
The Exeter V40 femoral stem is the most implanted stem in the National Joint Registry (NJR) for primary total hip arthroplasty (THA). In 2004, the 44/00/125 stem was released for use in ‘cement-in-cement’ revision cases. It has, however, been used ‘off-label’ as a primary stem when patient anatomy requires a smaller stem with a 44 mm offset. We aimed to investigate survival of this implant in comparison to others in the range when used in primary THAs recorded in the NJR. We analyzed 328,737 primary THAs using the Exeter V40 stem, comprising 34.3% of the 958,869 from the start of the NJR to December 2018. Our exposure was the stem, and the outcome was all-cause construct revision. We stratified analyses into four groups: constructs using the 44/00/125 stem, those using the 44/0/150 stem, those including a 35.5/125 stem, and constructs using any other Exeter V40 stem.Aims
Methods
Aims. Periprosthetic fracture (PF) after primary total hip arthroplasty
(THA) is an uncommon but potentially devastating complication. This
study aims to investigate the influence of cemented stem designs
on the risk of needing a revision for a PF. Patients and Methods. We analysed data on 257 202 primary THAs with cemented stems
and 390 linked first revisions for PF recorded in the National Joint
Registry (NJR) of
Cemented hemiarthroplasty is an effective form of treatment for most patients with an intracapsular fracture of the hip. However, it remains unclear whether there are subgroups of patients who may benefit from the alternative operation of a modern uncemented hemiarthroplasty – the aim of this study was to investigate this issue. Knowledge about the heterogeneity of treatment effects is important for surgeons in order to target operations towards specific subgroups who would benefit the most. We used causal forest analysis to compare subgroup- and individual-level treatment effects between cemented and modern uncemented hemiarthroplasty in patients aged > 60 years with an intracapsular fracture of the hip, using data from the World Hip Trauma Evaluation 5 (WHiTE 5) multicentre randomized clinical trial. EuroQol five-dimension index scores were used to measure health-related quality of life at one, four, and 12 months postoperatively.Aims
Methods
The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.Aims
Methods
Several different designs of hemiarthroplasty are used to treat intracapsular fractures of the proximal femur, with large variations in costs. No clinical benefit of modular over monoblock designs has been reported in the literature. Long-term data are lacking. The aim of this study was to report the ten-year implant survival of commonly used designs of hemiarthroplasty. Patients recorded by the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) between 1 September 1999 and 31 December 2020 who underwent hemiarthroplasty for the treatment of a hip fracture with the following implants were included: a cemented monoblock Exeter Trauma Stem (ETS), cemented Exeter V40 with a bipolar head, a monoblock Thompsons prosthesis (Cobalt/Chromium or Titanium), and an Exeter V40 with a Unitrax head. Overall and age-defined cumulative revision rates were compared over the ten years following surgery.Aims
Methods
The burden of revision total hip arthroplasty (rTHA) continues to grow. The surgery is complex and associated with significant costs. Regional rTHA networks have been proposed to improve outcomes and to reduce re-revisions, and therefore costs. The aim of this study was to accurately quantify the cost and reimbursement for a rTHA service, and to assess the financial impact of case complexity at a tertiary referral centre within the NHS. A retrospective analysis of all revision hip procedures was performed at this centre over two consecutive financial years (2018 to 2020). Cases were classified according to the Revision Hip Complexity Classification (RHCC) and whether they were infected or non-infected. Patients with an American Society of Anesthesiologists (ASA) grade ≥ III or BMI ≥ 40 kg/m2 are considered “high risk” by the RHCC. Costs were calculated using the Patient Level Information and Costing System (PLICS), and remuneration based on Healthcare Resource Groups (HRG) data. The primary outcome was the financial difference between tariff and cost per patient episode.Aims
Methods
This study compares the re-revision rate and mortality following septic and aseptic revision hip arthroplasty (rTHA) in registry data, and compares the outcomes to previously reported data. This is an observational cohort study using data from the German Arthroplasty Registry (EPRD). A total of 17,842 rTHAs were included, and the rates and cumulative incidence of hip re-revision and mortality following septic and aseptic rTHA were analyzed with seven-year follow-up. The Kaplan-Meier estimates were used to determine the re-revision rate and cumulative probability of mortality following rTHA.Aims
Methods
This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.Aims
Methods
Aims. Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA)
are commonly used, but concerns exist regarding ceramic fracture.
This study aims to report the risk of revision for fracture of modern
CoC bearings and identify factors that might influence this risk,
using data from the National Joint Registry (NJR) for
Periprosthetic proximal femoral fractures (PFFs) are a major complication after total hip arthroplasty (THA). Health status after PFF is not specifically investigated. The aim of this study is to evaluate the health status pattern over two years after sustaining a PFF. A cohort of patients with PFF after THA was derived from the Brabant Injury Outcomes Surveillance (BIOS) study. The BIOS study, a prospective, observational, multicentre follow-up cohort study, was conducted to obtain data by questionnaires pre-injury and at one week, and one, three, six, 12, and 24 months after trauma. Primary outcome measures were the EuroQol five-dimension three-level questionnaire (EQ-5D-3L), the Health Utility Index 2 (HUI2), and the Health Utility Index 3 (HUI3). Secondary outcome measures were general measurements such as duration of hospital stay and mortality.Aims
Methods
The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
Young adults undergoing total hip arthroplasty (THA) largely have different indications for surgery, preoperative function, and postoperative goals compared to a standard patient group. The aim of our study was to describe young adult THA preoperative function and quality of life, and to assess postoperative satisfaction and compare this with functional outcome measures. A retrospective cohort analysis of young adults (aged < 50 years) undergoing THA between May 2018 and May 2023 in a single tertiary centre was undertaken. Median follow-up was 31 months (12 to 61). Oxford Hip Score (OHS) and focus group-designed questionnaires were distributed. Searches identified 244 cases in 225 patients. Those aged aged under 30 years represented 22.7% of the cohort. Developmental dysplasia of the hip (50; 45.5%) and Perthes’ disease (15; 13.6%) were the commonest indications for THA.Aims
Methods
Delirium is associated with adverse outcomes following hip fracture, but the prevalence and significance of delirium for the prognosis and ongoing rehabilitation needs of patients admitted from home is less well studied. Here, we analyzed relationships between delirium in patients admitted from home with 1) mortality; 2) total length of hospital stay; 3) need for post-acute inpatient rehabilitation; and 4) hospital readmission within 180 days. This observational study used routine clinical data in a consecutive sample of hip fracture patients aged ≥ 50 years admitted to a single large trauma centre during the COVID-19 pandemic between 1 March 2020 and 30 November 2021. Delirium was prospectively assessed as part of routine care by the 4 A’s Test (4AT), with most assessments performed in the emergency department. Associations were determined using logistic regression adjusted for age, sex, Scottish Index of Multiple Deprivation quintile, COVID-19 infection within 30 days, and American Society of Anesthesiologists grade.Aims
Methods
The aim of this study was to compare the cost-effectiveness of cemented hemiarthroplasty (HA) versus hydroxyapatite-coated uncemented HA for the treatment of displaced intracapsular hip fractures in older adults. A within-trial economic evaluation was conducted based on data collected from the World Hip Trauma Evaluation 5 (WHiTE 5) multicentre randomized controlled trial in the UK. Resource use was measured over 12 months post-randomization using trial case report forms and participant-completed questionnaires. Cost-effectiveness was reported in terms of incremental cost per quality-adjusted life year (QALY) gained from the NHS and personal social service perspective. Methodological uncertainty was addressed using sensitivity analysis, while decision uncertainty was represented graphically using confidence ellipses and cost-effectiveness acceptability curves.Aims
Methods
The aim of this study was to explore the functional results in a fitter subgroup of participants in the Hip Fracture Evaluation with Alternatives of Total Hip Arthroplasty versus Hemiarthroplasty (HEALTH) trial to determine whether there was an advantage of total hip arthroplasty (THA) versus hemiarthroplasty (HA) in this population. We performed a post hoc exploratory analysis of a fitter cohort of patients from the HEALTH trial. Participants were aged over 50 years and had sustained a low-energy displaced femoral neck fracture (FNF). The fittest participant cohort was defined as participants aged 70 years or younger, classified as American Society of Anesthesiologists grade I or II, independent walkers prior to fracture, and living at home prior to fracture. Multilevel models were used to estimate the effect of THA versus HA on functional outcomes. In addition, a sensitivity analysis of the definition of the fittest participant cohort was performed.Aims
Methods
The aim of this modified Delphi process was to create a structured Revision Hip Complexity Classification (RHCC) which can be used as a tool to help direct multidisciplinary team (MDT) discussions of complex cases in local or regional revision networks. The RHCC was developed with the help of a steering group and an invitation through the British Hip Society (BHS) to members to apply, forming an expert panel of 35. We ran a mixed-method modified Delphi process (three rounds of questionnaires and one virtual meeting). Round 1 consisted of identifying the factors that govern the decision-making and complexities, with weighting given to factors considered most important by experts. Participants were asked to identify classification systems where relevant. Rounds 2 and 3 focused on grouping each factor into H1, H2, or H3, creating a hierarchy of complexity. This was followed by a virtual meeting in an attempt to achieve consensus on the factors which had not achieved consensus in preceding rounds.Aims
Methods
Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.Aims
Methods
Objectives. To investigate psychosocial and biomedical outcomes following
total hip replacement (THR) and to identify predictors of recovery
from THR. Methods. Patients with osteoarthritis (OA) on the waiting list for primary
THR in North West
Perthes’ disease is an uncommon hip disorder with limited data on the long-term outcomes in adulthood. We partnered with community-based foundations and utilized web-based survey methodology to develop the Adult Perthes Survey, which includes demographics, childhood and adult Perthes’ disease history, the University of California Los Angeles (UCLA) Activity Scale item, Short Form-36, the Hip disability and Osteoarthritis Outcome Score, and a body pain diagram. Here we investigate the following questions: 1) what is the feasibility of obtaining > 1,000 survey responses from adults who had Perthes’ disease using a web-based platform?; and 2) what are the baseline characteristics and demographic composition of our sample? The survey link was available publicly for 15 months and advertised among support groups. Of 1,505 participants who attempted the Adult Perthes survey, 1,182 completed it with a median timeframe of 11 minutes (IQR 8.633 to 14.72). Participants who dropped out were similar to those who completed the survey on several fixed variables. Participants represented 45 countries including the USA (n = 570; 48%), UK (n = 295; 25%), Australia (n = 133; 11%), and Canada (n = 46; 4%). Of the 1,182 respondents, 58% were female and the mean age was 39 years (SD 12.6).Aims
Methods
Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost. In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA.Aims
Methods
Arthroplasty registries are important for the
surveillance of joint replacements and the evaluation of outcome. Independent
validation of registry data ensures high quality. The ability for
orthopaedic implant retrieval centres to validate registry data
is not known. We analysed data from the National Joint Registry
for
The aim of this study was to assess medium-term improvements following total hip arthroplasty (THA), and to evaluate what effect different preoperative Oxford Hip Score (OHS) thresholds for treatment may have on patients’ access to THA and outcomes. Patients undergoing primary THA at our institution with an OHS both preoperatively and at least four years postoperatively were included. Rationing thresholds were explored to identify possible deprivation of OHS improvement.Aims
Methods
A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined.Aims
Methods
Golf is a popular pursuit among those requiring total hip arthroplasty (THA). The aim of this study was to determine if participating in golf is associated with greater functional outcomes, satisfaction, or improvement in quality of life (QoL) compared to non-golfers. All patients undergoing primary THA over a one-year period at a single institution were included with one-year postoperative outcomes. Patients were retrospectively followed up to assess if they had been golfers at the time of their surgery. Multivariate linear regression analysis was performed to assess the independent association of preoperative golfing status on outcomes.Aims
Methods
The Patient-Reported Outcomes Measurement Information System (PROMIS) has demonstrated faster administration, lower burden of data capture and reduced floor and ceiling effects compared to traditional Patient Reported Outcomes Measurements (PROMs). We investigated the suitability of PROMIS Mobility score in assessing physical function in the sequelae of childhood hip disease. In all, 266 adolscents (aged ≥ 12 years) and adults were identified with a prior diagnosis of childhood hip disease (either Perthes’ disease (n = 232 (87.2%)) or Slipped Capital Femoral Epiphysis (n = 34 (12.8%)) with a mean age of 27.73 years (SD 12.24). Participants completed the PROMIS Mobility Computer Adaptive Test, the Non-Arthritic Hip Score (NAHS), EuroQol five-dimension five-level questionnaire, and the Numeric Pain Rating Scale. We investigated the correlation between the PROMIS Mobility and other tools to assess use in this population and any clustering of outcome scores.Aims
Methods
Increased femoral head size may reduce dislocation rates following total hip replacement. The National Joint Registry for England and Wales has highlighted a statistically significant increase in the use of femoral heads ≥ 36 mm in diameter from 5% in 2005 to 26% in 2009, together with an increase in the use of the posterior approach. The aim of this study was to determine whether rates of dislocation have fallen over the same period. National data for
The primary aim was to determine the influence of COVID-19 on 30-day mortality following hip fracture. Secondary aims were to determine predictors of COVID-19 status on presentation and later in the admission; the rate of hospital acquired COVID-19; and the predictive value of negative swabs on admission. A nationwide multicentre retrospective cohort study was conducted of all patients presenting with a hip fracture to 17 Scottish centres in March and April 2020. Demographics, presentation blood tests, COVID-19 status, Nottingham Hip Fracture Score, management, length of stay, and 30-day mortality were recorded.Aims
Methods
The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA. A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.Aims
Methods
Postoperative delirium (POD) and postoperative cognitive decline (POCD) are common surgical complications. In the UK, the Best Practice Tariff incentivizes the screening of delirium in patients with hip fracture. Further, a National Hip Fracture Database (NHFD) performance indicator is the reduction in the incidence of POD. To aid in its recognition, we sought to determine factors associated with POD and POCD in patients with hip fractures. We interrogated the NHFD data on patients presenting with hip fractures to our institution from 2016 to 2018. POD was determined using the 4AT score, as recommended by the NHFD and UK Department of Health. POCD was defined as a decline in Abbreviated Mental Test Score (AMTS) of two or greater. Using logistic regression, we adjusted for covariates to identify factors associated with POD and POCD.Aims
Methods
Hip fracture is a common condition of the older, frailer person. This population is also at risk from SARS-CoV-2 infection. It is important to understand the impact of coexistent hip fracture and SARS-CoV-2 for informed decision-making at patient and service levels. We undertook a systematic review and meta-analysis of observational studies of older (> 60 years) people with fragility hip fractures and outcomes with and without SARS-CoV-2 infection during the first wave of the COVID-19 pandemic. The primary outcome was early (30-day or in-hospital) mortality. Secondary outcomes included length of hospital stay and key clinical characteristics known to be associated with outcomes after hip fracture.Aims
Methods
Using a regional arthroplasty register, we assessed the outcome, at five years, of 1198 primary Charnley total hip replacements (THRs) undertaken across a single health region in
Our rural orthopaedic service has undergone service restructure during the COVID-19 pandemic in order to sustain hip fracture care. All adult trauma care has been centralised to the Royal Shrewsbury Hospital for assessment and medical input, before transferring those requiring operative intervention to the Robert Jones and Agnes Hunt Orthopaedic Hospital. We aim to review the impact of COVID-19 on hip fracture workload and service changes upon management of hip fractures. We reviewed our prospectively maintained trust database and National Hip Fracture Database records for the months of March and April between the years 2016 and 2020. Our assessment included fracture pattern (intrascapular vs extracapsular hip fracture), treatment intervention, length of stay and mortality.Aims
Methods
Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.Aims
Methods
To assess the variation in pre-fracture quality of life (QoL) within the UK hip fracture population, and quantify the nature and strength of associations between QoL and other routinely collected patient characteristics and treatment choices. The World Hip Trauma Evaluation (WHiTE) study, an observational cohort study of UK hip fracture patients, collects a range of routine data and a health-related QoL score (EuroQol five-dimension questionnaire (EQ-5D)). Pre-fracture QoL data are summarized and statistical models fitted to understand associations between QoL, patient characteristics, fracture types, and operations.Aims
Methods
Despite advances in the treatment of paediatric hip disease, adolescent and young adult patients can develop early onset end-stage osteoarthritis. The aims of this study were to address the indications and medium-term outcomes for total hip arthroplasty (THA) with ceramic bearings for teenage patients. Surgery was performed by a single surgeon working in the paediatric orthopaedic unit of a tertiary referral hospital. Databases were interrogated from 2003 to 2017 for all teenage patients undergoing THA with a minimum 2.3 year follow-up. Data capture included patient demographics, the underlying hip pathology, number of previous surgeries, and THA prostheses used. Institutional ethical approval was granted to contact patients for prospective clinical outcomes and obtain up-to-date radiographs. In total, 60 primary hips were implanted in 51 patients (35 female, 16 male) with nine bilateral cases. The mean age was 16.7 years (12 to 19) and mean follow-up was 9.3 years (2.3 to 16.8).Aims
Methods
The coronavirus disease (COVID)-19 pandemic forced an unprecedented period of challenge to the NHS in the UK where hip fractures in the elderly population are a major public health concern. There are approximately 76,000 hip fractures in the UK each year which make up a substantial proportion of the trauma workload of an average orthopaedic unit. This study aims to assess the impact of the COVID-19 pandemic on hip fracture care service and the emerging lessons to withstand any future outbreaks. Data were collected retrospectively on 157 hip fractures admitted from March to May 2019 and 2020. The 2020 group was further subdivided into COVID-positive and COVID-negative. Data including the four-hour target, timing to imaging, hours to operation, anaesthetic and operative details, intraoperative complications, postoperative reviews, COVID status, Key Performance Indicators (KPIs), length of stay, postoperative complications, and the 30-day mortality were compiled from computer records and our local National Hip Fracture Database (NHFD) export data.Aims
Methods
To compare long-term survival of all-cemented and hybrid total hip arthroplasty (THA) using the Exeter Universal stem. Details of 1,086 THAs performed between 1999 and 2005 using the Exeter stem and either a cemented (632) or uncemented acetabular component (454) were collected from local records and the New Zealand Joint Registry. A competing risks regression survival analysis was performed with death as the competing risk with adjustments made for age, sex, approach, and bearing.Aims
Methods
We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice. Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.Aims
Methods
The primary aim of this paper was to outline the processes involved in building the Partners Arthroplasty Registry (PAR), established in April 2016 to capture baseline and outcome data for patients undergoing arthroplasty in a regional healthcare system. A secondary aim was to determine the quality of PAR’s data. A tertiary aim was to report preliminary findings from the registry and contributions to quality improvement initiatives and research up to March 2019. Structured Query Language was used to obtain data relating to patients who underwent total hip or knee arthroplasty (THA and TKA) from the hospital network’s electronic medical record (EMR) system to be included in the PAR. Data were stored in a secure database and visualized in dashboards. Quality assurance of PAR data was performed by review of the medical records. Capture rate was determined by comparing two months of PAR data with operating room schedules. Linear and binary logistic regression models were constructed to determine if length of stay (LOS), discharge to a care home, and readmission rates improved between 2016 and 2019.Aims
Methods
Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.Aims
Methods
The aim of this study was to identify modifiable risk factors associated with mortality in patients requiring revision total hip arthroplasty (THA) for periprosthetic hip fracture. The electronic records of consecutive patients undergoing revision THA for periprosthetic hip fracture between December 2011 and October 2018 were reviewed. The data which were collected included age, sex, body mass index (BMI), American Society of Anesthesiologists (ASA) classification, the preoperative serum level of haemoglobin, time to surgery, operating time, blood transfusion, length of hospital stay, and postoperative surgical and medical complications. Univariate and multivariate logistic regression analyses were used to determine independent modifiable factors associated with mortality at 90 days and one year postoperatively.Aims
Methods
Dislocation remains a significant complication after total hip arthroplasty (THA), being the third leading indication for revision. We present a series of acetabular revision using a dual mobility cup (DMC) and compare this with our previous series using the posterior lip augmentation device (PLAD). A retrospective review of patients treated with either a DMC or PLAD for dislocation in patients with a Charnley THA was performed. They were identified using electronic patient records (EPR). EPR data and radiographs were evaluated to determine operating time, length of stay, and the incidence of complications and recurrent dislocation postoperatively.Aims
Methods
This feasibility study investigates the utilization and cost of health resources related to formal and informal care, home adaptations, and physiotherapy among patients aged 60 years and above after hip fracture from a multicentre cohort study (World Hip Trauma Evaluation (WHiTE)) in the UK. A questionnaire containing health resource use was completed at baseline and four months post-injury by patients or their carer. Completion rate and mean cost of each health resource item were assessed and sensitivity analysis was performed to derive a conservative estimate of the informal care cost. All costs are presented in 2017/18 pound sterling.Aims
Methods
We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls. This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a national joint registry. Implant and patient survival rates were compared between groups using Cox regression.Aims
Patients and Methods
Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (Objectives
Methods
The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom National Joint Registry (NJR) were included in a multivariable survival and regression analysis to identify the adjusted hazard of revision for periprosthetic fracture following primary THA using a cementless stem. The effect of a collar in cementless THA on this risk was evaluated in an Aims
Materials and Methods
The aim of this randomized trial was to compare the functional outcome of two different surgical approaches to the hip in patients with a femoral neck fracture treated with a hemiarthroplasty. A total of 150 patients who were treated between February 2014 and July 2017 were included. Patients were allocated to undergo hemiarthroplasty using either an anterolateral or a direct lateral approach, and were followed for 12 months. The mean age of the patients was 81 years (69 to 90), and 109 were women (73%). Functional outcome measures, assessed by a physiotherapist blinded to allocation, and patient-reported outcome measures (PROMs) were collected postoperatively at three and 12 months.Aims
Patients and Methods
The hypothesis of this study was that thigh circumference, distinct from body mass index (BMI), may be associated with the positioning of components when undertaking total hip arthroplasty (THA) using the direct anterior approach (DAA), and that an increased circumference might increase the technical difficulty. We performed a retrospective review of prospectively collected data involving 155 consecutive THAs among 148 patients undertaken using the DAA at an academic medical centre by a single fellowship-trained surgeon. Preoperatively, thigh circumference was measured at 10 cm, 20 cm, and 30 cm distal to the anterior superior iliac spine, in quartiles. Two blinded reviewers assessed the inclination and anteversion of the acetabular component, radiological leg-length discrepancy, and femoral offset. The radiological outcomes were considered as continuous and binary outcome variables based on Lewinnek’s ‘safe zone’.Aims
Patients and Methods
There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs. We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.Aims
Patients and Methods
The primary objective of this study was to compare accuracy in restoring the native centre of hip rotation in patients undergoing conventional manual total hip arthroplasty (THA) This prospective cohort study included 50 patients undergoing conventional manual THA and 25 patients receiving robotic-arm assisted THA. Patients undergoing conventional manual THA and robotic-arm assisted THA were well matched for age (mean age, 69.4 years (Objectives
Materials and Methods
Aims
Patients and Methods
Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event. A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The ‘pre-2007’ group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the ‘2007 onwards’ group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors.Objectives
Methods
The aim of this study was to determine the stability of a new
short femoral stem compared with a conventional femoral stem in
patients undergoing cementless total hip arthroplasty (THA), in
a prospective randomized controlled trial using radiostereometric
analysis (RSA). A total of 53 patients were randomized to receive cementless
THA with either a short femoral stem (MiniHip, 26 patients, mean
age: 52 years, nine male) or a conventional length femoral stem
(MetaFix, 23 patients, mean age: 53 years, 11 male). All patients
received the same cementless acetabular component. Two-year follow-up
was available on 38 patients. Stability was assessed through migration
and dynamically inducible micromotion. Radiographs for RSA were
taken postoperatively and at three, six, 12, 18, and 24 months.Aims
Patients and Methods
This study aimed to evaluate implant survival of reverse hybrid
total hip arthroplasty (THA) at medium-term follow-up. A consecutive series of 1082 THAs in 982 patients with mean follow-up
of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21
to 94). Of these, 194 (17.9%) were in patients under 60 years, 663
(61.3%) in female patients and 348 (32.2%) performed by a trainee.
Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%).
Survival analysis was performed and subgroups compared using log
rank tests.Aims
Patients and Methods
We used the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man (NJR) to investigate the risk of revision due to prosthetic joint infection (PJI) for patients undergoing primary and revision hip arthroplasty, the changes in risk over time, and the overall burden created by PJI. We analysed revision total hip arthroplasties (THAs) performed due to a diagnosis of PJI and the linked index procedures recorded in the NJR between 2003 and 2014. The cohort analysed consisted of 623 253 index primary hip arthroplasties, 63 222 index revision hip arthroplasties and 7585 revision THAs performed due to a diagnosis of PJI. The prevalence, cumulative incidence functions and the burden of PJI (total procedures) were calculated. Overall linear trends were investigated with log-linear regression.Objectives
Methods
Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision. We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.Objectives
Methods
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods
This study investigates the reporting of health-related quality of life (HRQoL) in patients following hip fracture. We compare the relative merits and make recommendations for the use for two methods of measuring HRQoL; (i) including patients who died during follow-up and (ii) including survivors only. The World Hip Trauma Evaluation has previously reported changes in HRQoL using EuroQol-5D for patients with hip fractures. We performed additional analysis to investigate the effect of including or excluding those patients who died during the first four months of the follow-up period.Objectives
Methods
The aim of this study was to determine whether patients with
metal-on-metal (MoM) arthroplasties of the hip have an increased
risk of cardiac failure compared with those with alternative types
of arthroplasties (non-MoM). A linkage study between the National Joint Registry, Hospital
Episodes Statistics and records of the Office for National Statistics
on deaths was undertaken. Patients who underwent elective total
hip arthroplasty between January 2003 and December 2014 with no
past history of cardiac failure were included and stratified as
having either a MoM (n = 53 529) or a non-MoM (n = 482 247) arthroplasty.
The primary outcome measure was the time to an admission to hospital
for cardiac failure or death. Analysis was carried out using data
from all patients and from those matched by propensity score.Aims
Patients and Methods
The aim of this study was to determine whether the rates of revision
for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle
components varied according to the year of the initial operation,
and compare these with the rates of revision for other designs of
MoM THA. Data from the National Joint Registry for England and Wales included
36 mm MoM THAs with Pinnacle acetabular components which were undertaken
between 2003 and 2012 with follow-up for at least five years (n
= 10 776) and a control group of other MoM THAs (n = 13 817). The
effect of the year of the primary operation on all-cause rates of revision
was assessed using Cox regression and interrupted time-series analysis.Aims
Patients and Methods
The number of patients undergoing arthroscopic surgery of the
hip has increased significantly during the past decade. It has now
become an established technique for the treatment of many intra-
and extra-articular conditions affecting the hip. However, it has
a steep learning curve and is not without the risk of complications.
The purpose of this systematic review was to determine the prevalence
of complications during and following this procedure. Preferred Reporting Items for Systematic Reviews and Meta-Analyses
guidelines were used in designing this study. Two reviewers systematically
searched the literature for complications related to arthroscopy
of the hip. The research question and eligibility criteria were
established Aims
Materials and Methods
The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty
(THA). Our one-year results demonstrated elevation in metal ion
levels above baseline with no significant difference between the
CoM and MoM groups. This paper reviews the five-year data. The implants used in each patient differed only in respect to
the type of femoral head (ceramic or metal). At five-year follow-up
of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available
for comparison.Aims
Patients and Methods
The aims of the study were to review and analyse the reported
series of debridement, antibiotics and implant retention (DAIR)
in the management of infected total hip arthroplasties (THAs) to
establish the overall success and the influencing factors. Using a standardised recognised study protocol, meta-analysis
of observational studies in epidemiology guidelines, a comprehensive
review and analysis of the literature was performed. The primary
outcome measure was the success of treatment. The search strategy
and inclusion criteria which involved an assessment of quality yielded
39 articles for analysis, which included 1296 patients.Aims
Patients and methods
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
The most effective surgical approach for total hip arthroplasty
(THA) remains controversial. The direct anterior approach may be
associated with a reduced risk of dislocation, faster recovery,
reduced pain and fewer surgical complications. This systematic review
aims to evaluate the current evidence for the use of this approach
in THA. Following the Cochrane collaboration, an extensive literature
search of PubMed, Medline, Embase and OvidSP was conducted. Randomised
controlled trials, comparative studies, and cohort studies were
included. Outcomes included the length of the incision, blood loss,
operating time, length of stay, complications, and gait analysis.Aims
Materials and Methods
The aim of this study was to investigate the outcomes of Vancouver
type B2 and B3 fractures by performing a systematic review of the
methods of surgical treatment which have been reported. A systematic search was performed in Ovid MEDLINE, Embase and
the Cochrane Central Register of Controlled Trials. For inclusion,
studies required a minimum of ten patients with a Vancouver type
B2 and/or ten patients with a Vancouver type B3 fracture, a minimum
mean follow-up of two years and outcomes which were matched to the type
of fracture. Studies were also required to report the rate of re-operation
as an outcome measure. The protocol was registered in the PROSPERO
database. Aims
Materials and Methods
The aim of this study was to compare early functional and health
related quality of life outcomes (HRQoL) in patients who have undergone
total hip arthroplasty (THA) using a bone conserving short stem
femoral component and those in whom a conventional length uncemented
component was used. Outcome was assessed using a validated performance
based outcome instrument as well as patient reported outcome measures
(PROMs). We prospectively analysed 33 patients whose THA involved a contemporary
proximally porous coated tapered short stem femoral component and
53 patients with a standard conventional femoral component, at a
minimum follow-up of two years. The mean follow-up was 31.4 months
(24 to 39). Patients with poor proximal femoral bone quality were
excluded. The mean age of the patients was 66.6 years (59 to 77)
and the mean body mass index was 30.2 kg/m2 (24.1 to
41.0). Outcome was assessed using the Oxford Hip Score (OHS) and
the University College Hospital (UCH) hip score which is a validated
performance based instrument. HRQoL was assessed using the EuroQol
5D (EQ-5D).Aims
Patients and Methods
To determine ten-year failure rates following 36 mm metal-on-metal
(MoM) Pinnacle total hip arthroplasty (THA), and identify predictors
of failure. We retrospectively assessed a single-centre cohort of 569 primary
36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012
according to Medicines and Healthcare Products Regulation Agency
recommendations. All-cause failure rates (all-cause revision, and
non-revised cross-sectional imaging failures) were calculated, with predictors
for failure identified using multivariable Cox regression.Aims
Patients and Methods
There is little evidence on the cost effectiveness
of different brands of hip prostheses. We compared lifetime cost effectiveness
of frequently used brands within types of prosthesis including cemented
(Exeter V40 Contemporary, Exeter V40 Duration and Exeter V40 Elite
Plus Ogee), cementless (Corail Pinnacle, Accolade Trident, and Taperloc Exceed)
and hybrid (Exeter V40 Trilogy, Exeter V40 Trident, and CPT Trilogy).
We used data from three linked English national databases to estimate
the lifetime risk of revision, quality-adjusted life years (QALYs)
and cost. For women with osteoarthritis aged 70 years, the Exeter V40 Elite
Plus Ogee had the lowest risk of revision (5.9% revision risk, 9.0
QALYs) and the CPT Trilogy had the highest QALYs (10.9% revision
risk, 9.3 QALYs). Compared with the Corail Pinnacle (9.3% revision
risk, 9.22 QALYs), the most commonly used brand, and assuming a
willingness-to-pay of £20 000 per QALY gain, the CPT Trilogy is
most cost effective, with an incremental net monetary benefit of £876.
Differences in cost effectiveness between the hybrid CPT Trilogy
and Exeter V40 Trident and the cementless Corail Pinnacle and Taperloc
Exceed were small, and a cautious interpretation is required, given
the limitations of the available information. However, it is unlikely that cemented brands are among the most
cost effective. Similar patterns of results were observed for men
and other ages. The gain in quality of life after total hip arthroplasty,
rather than the risk of revision, was the main driver of cost effectiveness. Cite this article:
The National Joint Registry for England, Wales and Northern Ireland
(NJR) has extended its scope to report on hospital, surgeon and
implant performance. Data linkage of the NJR to the London Implant
Retrieval Centre (LIRC) has previously evaluated data quality for
hip primary procedures, but did not assess revision records. We analysed metal-on-metal hip revision procedures performed
between 2003 and 2013. A total of 69 929 revision procedures from
the NJR and 929 revised pairs of components from the LIRC were included.Aims
Methods
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
Metal-on-metal resurfacing of the hip (MoMHR)
has enjoyed a resurgence in the last decade, but is now again in question
as a routine option for osteoarthritis of the hip. Proponents of
hip resurfacing suggest that its survival is superior to that of
conventional hip replacement (THR), and that hip resurfacing is
less invasive, is easier to revise than THR, and provides superior
functional outcomes. Our argument serves to illustrate that none
of these proposed advantages have been realised and new and unanticipated
serious complications, such as pseudotumors, have been associated
with the procedure. As such, we feel that the routine use of MoMHR
is not justified. Cite this article:
Total hip replacement causes a short-term increase
in the risk of mortality. It is important to quantify this and to identify
modifiable risk factors so that the risk of post-operative mortality
can be minimised. We performed a systematic review and critical
evaluation of the current literature on the topic. We identified
32 studies published over the last 10 years which provide either
30-day or 90-day mortality data. We estimate the pooled incidence
of mortality during the first 30 and 90 days following hip replacement
to be 0.30% (95% CI 0.22 to 0.38) and 0.65% (95% CI 0.50 to 0.81),
respectively. We found strong evidence of a temporal trend towards
reducing mortality rates despite increasingly co-morbid patients.
The risk factors for early mortality most commonly identified are
increasing age, male gender and co-morbid conditions, particularly
cardiovascular disease. Cardiovascular complications appear to have
overtaken fatal pulmonary emboli as the leading cause of death after
hip replacement. Cite this article:
We report on the outcome of the Exeter Contemporary flanged cemented
all-polyethylene acetabular component with a mean follow-up of 12
years (10 to 13.9). This study reviewed 203 hips in 194 patients.
129 hips in 122 patients are still A retrospective review was undertaken of a consecutive series
of 203 routine primary cemented total hip arthroplasties (THA) in
194 patients.Aims
Patients and Methods
Many different designs of total hip arthroplasty
(THA) with varying performance and cost are available. The identification
of those which are the most cost-effective could allow significant
cost-savings. We used an established Markov model to examine the
cost effectiveness of five frequently used categories of THA which differed
according to bearing surface and mode of fixation, using data from
the National Joint Registry for England and Wales. Kaplan–Meier
analyses of rates of revision for men and women were modelled with
parametric distributions. Costs of devices were provided by the
NHS Supply Chain and associated costs were taken from existing studies.
Lifetime costs, lifetime quality-adjusted-life-years (QALYs) and
the probability of a device being cost effective at a willingness
to pay £20 000/QALY were included in the models. The differences in QALYs between different categories of implant
were extremely small (<
0.0039 QALYs for men or women over the
patient’s lifetime) and differences in cost were also marginal (£2500
to £3000 in the same time period). As a result, the probability
of any particular device being the most cost effective was very
sensitive to small, plausible changes in quality of life estimates
and cost. Our results suggest that available evidence does not support
recommending a particular device on cost effectiveness grounds alone.
We would recommend that the choice of prosthesis should be determined
by the rate of revision, local costs and the preferences of the
surgeon and patient. Cite this article:
The highly cross-linked polyethylene Exeter RimFit flanged cemented
acetabular component was introduced in the United Kingdom in 2010.
This study aimed to examine the rates of emergence of radiolucent
lines observed when the Rimfit acetabular component was implanted
at total hip arthroplasty (THA) using two different techniques: firstly,
the ‘rimcutter’ technique in which the flange sits on a pre-prepared
acetabular rim; and secondly, the ‘trimmed flange’ technique in
which the flange is trimmed and the acetabular component is seated
inside the rim of the acetabulum. The radiographs of 150 THAs (75 ‘rimcutter’, 75 ‘trimmed flange’)
involving this component were evaluated to assess for radiolucencies
at the cement/bone interface by three observers. Aims
Patients and Methods
The optimal management of intracapsular fractures of the femoral
neck in independently mobile patients remains open to debate. Successful
fixation obviates the limitations of arthroplasty for this group
of patients. However, with fixation failure rates as high as 30%,
the outcome of revision surgery to salvage total hip arthroplasty
(THA) must be considered. We carried out a systematic review to
compare the outcomes of salvage THA and primary THA for intracapsular
fractures of the femoral neck. We performed a Preferred Reporting Items for Systematic Reviews
and Meta-Analysis (PRISMA) compliant systematic review, using the
PubMed, EMBASE and Cochrane libraries databases. A meta-analysis
was performed where possible, and a narrative synthesis when a meta-analysis
was not possible.Aims
Patients and Methods
The purpose of this study was to validate the diagnosis of periprosthetic
joint infection (PJI) in the Danish Hip Arthroplasty Register (DHR). We identified a cohort of patients from the DHR who had undergone
primary total hip arthroplasty (THA) since 1 January 2005 and followed
them until first-time revision, death, emigration or until 31 December
2012. Revision for PJI, as registered in the DHR, was validated against
a benchmark which included information from microbiology databases,
prescription registers, clinical biochemistry registers and clinical
records. We estimated the sensitivity, specificity, positive predictive
value (PPV) and negative predictive value (NPV) for PJI in the DHR
alone and in the DHR when combined with microbiology databases.Aims
Patients and Methods
Femoroacetabular Junction Impingement (FAI) describes abnormalities
in the shape of the femoral head–neck junction, or abnormalities
in the orientation of the acetabulum. In the short term, FAI can
give rise to pain and disability, and in the long-term it significantly increases
the risk of developing osteoarthritis. The Femoroacetabular Impingement
Trial (FAIT) aims to determine whether operative or non-operative
intervention is more effective at improving symptoms and preventing
the development and progression of osteoarthritis. FAIT is a multicentre superiority parallel two-arm randomised
controlled trial comparing physiotherapy and activity modification
with arthroscopic surgery for the treatment of symptomatic FAI.
Patients aged 18 to 60 with clinical and radiological evidence of
FAI are eligible. Principal exclusion criteria include previous
surgery to the index hip, established osteoarthritis (Kellgren–Lawrence
≥ 2), hip dysplasia (centre-edge angle <
20°), and completion
of a physiotherapy programme targeting FAI within the previous 12
months. Recruitment will take place over 24 months and 120 patients
will be randomised in a 1:1 ratio and followed up for three years.
The two primary outcome measures are change in hip outcome score
eight months post-randomisation (approximately six-months post-intervention
initiation) and change in radiographic minimum joint space width
38 months post-randomisation. ClinicalTrials.gov: NCT01893034. Cite this article: Aims
Methods
Despite a lack of long-term follow-up, there
is an increasing trend towards using femoral heads of large diameter
in total hip replacement (THR), partly because of the perceived
advantage of lower rates of dislocation. However, increasing the
size of the femoral head is not the only way to reduce the rate
of dislocation; optimal alignment of the components and repair of
the posterior capsule could achieve a similar effect. In this prospective study of 512 cemented unilateral THRs (Male:Female
230:282) performed between 2004 and 2011, we aimed to determine
the rate of dislocation in patients who received a 22 mm head on
a 9/10 Morse taper through a posterior approach with capsular repair
and using the transverse acetabular ligament (TAL) as a guide for the
alignment of the acetabular component. The mean age of the patients
at operation was 67 years (35 to 89). The mean follow-up was 2.8
years (0.5 to 6.6). Pre- and post-operative assessment included
Oxford hip, Short Form-12 and modified University of California
Los Angeles and Merle D’Aubigne scores. The angles of inclination
and anteversion of the acetabular components were measured using
radiological software. There were four dislocations (0.78%), all
of which were anterior. In conclusion, THR with a 22 mm diameter head performed through
a posterior approach with capsular repair and using the TAL as a
guide for the alignment of the acetabular component was associated
with a low rate of dislocation. Cite this article:
In 2005, we demonstrated that the polished triple-tapered
C-stem at two years had migrated distally and rotated internally.
From that series, 33 patients have now been followed radiologically,
clinically and by radiostereometric analysis (RSA) for up to ten
years. The distal migration within the cement mantle had continued
and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation,
also within the cement mantle, was a mean 3.8° (external 1.6° to
internal 6.6°) The cement mantle did not show any sign of migration
or loosening in relation to the femoral bone. There were no clinical
or radiological signs indicating that the migration or rotation
within the cement mantle had had any adverse effects for the patients. Cite this article:
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article:
Despite excellent results, the use of cemented
total hip replacement (THR) is declining. This retrospective cohort study
records survival time to revision following primary cemented THR
using the most common combination of components that accounted for
almost a quarter of all cemented THRs, exploring risk factors independently associated
with failure. All patients with osteoarthritis who had an Exeter
V40/Contemporary THR (Stryker) implanted before 31 December 2010
and recorded in the National Joint Registry for England and Wales
were included in the analysis. Cox’s proportional hazard models
were used to analyse the extent to which risk of revision was related
to patient, surgeon and implant covariates, with a significance
threshold of p <
0.01. A total of 34 721 THRs were included in
the study. The overall seven-year rate of revision for any reason
was 1.70% (99% confidence interval (CI) 1.28 to 2.12). In the final
adjusted model the risk of revision was significantly higher in
THRs with the Contemporary hooded component (hazard ratio (HR) 1.88,
p <
0.001) than with the flanged version, and in smaller head
sizes (<
28 mm) compared with 28 mm diameter heads (HR 1.50,
p = 0.005). The seven-year revision rate was 1.16% (99% CI 0.69
to 1.63) with a 28 mm diameter head and flanged component. The overall
risk of revision was independent of age, gender, American Society
of Anesthesiologists grade, body mass index, surgeon volume, surgical
approach, brand of cement/presence of antibiotic, femoral head material
(stainless steel/alumina) and stem taper size/offset. However, the
risk of revision for dislocation was significantly higher with a
‘plus’ offset head (HR 2.05, p = 0.003) and a hooded acetabular component
(HR 2.34, p <
0.001). In summary, we found that there were significant differences
in failure between different designs of acetabular component and
sizes of femoral head after adjustment for a range of covariates.
Concerns have been raised that deformation of
acetabular shells may disrupt the assembly process of modular prostheses.
In this study we aimed to examine the effect that the strength of
bone has on the amount of deformation of the acetabular shell. The
hypothesis was that stronger bone would result in greater deformation.
A total of 17 acetabular shells were inserted into the acetabula
of eight cadavers, and deformation was measured using an optical
measuring system. Cores of bone from the femoral head were taken
from each cadaver and compressed using a materials testing machine.
The highest peak modulus and yield stress for each cadaver were used
to represent the strength of the bone and compared with the values
for the deformation and the surgeon’s subjective assessment of the
hardness of the bone. The mean deformation of the shell was 129
µm (3 to 340). No correlation was found between deformation and
either the maximum peak modulus (r² = 0.011, t = 0.426, p = 0.676) or
the yield stress (r² = 0.024, t = 0.614, p = 0.549) of the bone.
Although no correlation was found between the strength of the bone
and deformation, the values for the deformation observed could be
sufficient to disrupt the assembly process of modular acetabular
components. Cite this article: