header advert
Results 1 - 20 of 116
Results per page:
The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 11 - 19
1 Jan 2018
Darrith B Courtney PM Della Valle CJ

Aims. Instability remains a challenging problem in both primary and revision total hip arthroplasty (THA). Dual mobility components confer increased stability, but there are concerns about the unique complications associated with these designs, as well as the long-term survivorship. Materials and Methods. We performed a systematic review of all English language articles dealing with dual mobility THAs published between 2007 and 2016 in the MEDLINE and Embase electronic databases. A total of 54 articles met inclusion criteria for the final analysis of primary and revision dual mobility THAs and dual mobility THAs used in the treatment of fractures of the femoral neck. We analysed the survivorship and rates of aseptic loosening and of intraprosthetic and extra-articular dislocation. Results. For the 10 783 primary dual mobility THAs, the incidence of aseptic loosening was 1.3% (142 hips); the rate of intraprosthetic dislocation was 1.1% (122 hips) and the incidence of extra-articular dislocation was 0.46% (41 hips). The overall survivorship of the acetabular component and the dual mobility components was 98.0%, with all-cause revision as the endpoint at a mean follow-up of 8.5 years (2 to 16.5). For the 3008 revision dual mobility THAs, the rate of aseptic acetabular loosening was 1.4% (29 hips); the rate of intraprosthetic dislocation was 0.3% (eight hips) and the rate of extra-articular dislocation was 2.2% (67 hips). The survivorship of the acatabular and dual mobility components was 96.6% at a mean of 5.4 years (2 to 8). For the 554 dual mobility THAs which were undertaken in patients with a fracture of the femoral neck, the rate of intraprosthetic dislocation was 0.18% (one hip), the rate of extra-articular dislocation was 2.3% (13 hips) and there was one aseptic loosening. The survivorship was 97.8% at a mean of 1.3 years (0.75 to 2). Conclusion. Dual mobility articulations are a viable alternative to traditional bearing surfaces, with low rates of instability and good overall survivorship in primary and revision THAs, and in those undertaken in patients with a fracture of the femoral neck. The incidence of intraprosthetic dislocation is low and limited mainly to earlier designs. High-quality, prospective, comparative studies are needed to evaluate further the use of dual mobility components in THA. Cite this article: Bone Joint J 2018;100-B:11–19


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 27 - 32
1 Jul 2020
Heckmann N Weitzman DS Jaffri H Berry DJ Springer BD Lieberman JR

Aims. Dual mobility (DM) bearings are an attractive treatment option to obtain hip stability during challenging primary and revision total hip arthroplasty (THA) cases. The purpose of this study was to analyze data submitted to the American Joint Replacement Registry (AJRR) to characterize utilization trends of DM bearings in the USA. Methods. All primary and revision THA procedures reported to AJRR from 2012 to 2018 were analyzed. Patients of all ages were included and subdivided into DM and traditional bearing surface cohorts. Patient demographics, geographical region, hospital size, and teaching affiliation were assessed. Associations were determined by chi-squared analysis and logistic regression was performed to assess outcome variables. Results. A total of 406,900 primary and 34,745 revision THAs were identified, of which 35,455 (8.7%) and 8,031 (23.1%) received DM implants respectively. For primary THA, DM usage increased from 6.7% in 2012 to 12.0% in 2018. Among revision THA, DM use increased from 19.5% in 2012 to 30.6% in 2018. Patients < 50 years of age had the highest rates of DM implantation in every year examined. For each year of increase in age, there was a 0.4% decrease in the rate of DM utilization (odds ratio (OR) 0.996 (95% confidence interval (CI) 0.995 to 0.997); p < 0.001). Females were more likely to receive a DM implant compared to males (OR 1.077 (95% CI 1.054 to 1.100); p < 0.001). Major teaching institutions and smaller hospitals were associated with higher rates of utilization. DM articulations were used more commonly for dysplasia compared with osteoarthritis (OR 2.448 (95% CI 2.032 to 2.949); p < 0.001) during primary THA and for instability (OR 3.130 (95% CI 2.751 to 3.562) vs poly-wear; p < 0.001) in the revision setting. Conclusion. DM articulations showed a marked increase in utilization during the period examined. Younger patient age, female sex, and hospital characteristics such as teaching status, smaller size, and geographical location were associated with increased utilization. DM articulations were used more frequently for primary THA in patients with dysplasia and for revision THA in patients being treated for instability. Cite this article: Bone Joint J 2020;102-B(7 Supple B):27–32


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 20 - 26
1 Jul 2020
Romero J Wach A Silberberg S Chiu Y Westrich G Wright TM Padgett DE

Aims. This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber. Methods. For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current. Results. The radiological review identified that 32 of 551 MDM liners (5.8%) were malseated. Malseating was noted in all of the three different cup designs. The incidence of malseating was significantly higher in low-volume MDM surgeons than high-volume MDM surgeons (p < 0.001). Pristine well-seated liners showed significantly lower fretting current values at all peak loads greater than 800 N (p < 0.044). Malseated liner-shell couples had lower fretting onset loads at 2,400 N. Conclusion. MDM malseating remains an issue that can occur in at least one in 20 patients at a high-volume arthroplasty centre. The onset of fretting and increased fretting current throughout loading cycles suggests susceptibility to corrosion when this occurs. These results support the hypothesis that malseated liners may be at risk for fretting corrosion. Clinicians should be aware of this phenomenon. Cite this article: Bone Joint J 2020;102-B(7 Supple B):20–26


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 423 - 425
1 Apr 2020
Hoggett L Cross C Helm A

Aims. Dislocation remains a significant complication after total hip arthroplasty (THA), being the third leading indication for revision. We present a series of acetabular revision using a dual mobility cup (DMC) and compare this with our previous series using the posterior lip augmentation device (PLAD). Methods. A retrospective review of patients treated with either a DMC or PLAD for dislocation in patients with a Charnley THA was performed. They were identified using electronic patient records (EPR). EPR data and radiographs were evaluated to determine operating time, length of stay, and the incidence of complications and recurrent dislocation postoperatively. Results. A total of 28 patients underwent revision using a DMC for dislocation following Charnley THA between 2013 and 2017. The rate of recurrent dislocation and overall complications were compared with those of a previous series of 54 patients who underwent revision for dislocation using a PLAD, between 2007 and 2013. There was no statistically significant difference in the mean distribution of sex or age between the groups. The mean operating time was 71 mins (45 to 113) for DMCs and 43 mins (21 to 84) for PLADs (p = 0.001). There were no redislocations or revisions in the DMC group at a mean follow-up of 55 months (21 to 76), compared with our previous series of PLAD which had a redislocation rate of 16% (n = 9) and an overall revision rate of 25% (n = 14, p = 0.001) at a mean follow-up of 86 months (45 to 128). Conclusion. These results indicate that DMC outperforms PLAD as a treatment for dislocation in patients with a Charnley THA. This should therefore be the preferred form of treatment for these patients despite a slightly longer operating time. Work is currently ongoing to review outcomes of DMC over a longer follow-up period. PLAD should be used with caution in this patient group with preference given to acetabular revision to DMC. Cite this article: Bone Joint J 2020;102-B(4):423–425


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 18 - 24
1 Jan 2017
De Martino I D’Apolito R Soranoglou VG Poultsides LA Sculco PK Sculco TP

Aims. The aim of this systematic review was to report the rate of dislocation following the use of dual mobility (DM) acetabular components in primary and revision total hip arthroplasty (THA). Materials and Methods. A systematic review of the literature according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses guidelines was performed. A comprehensive search of Pubmed/Medline, Cochrane Library and Embase (Scopus) was conducted for English articles between January 1974 and March 2016 using various combinations of the keywords “dual mobility”, “dual-mobility”, “tripolar”, “double-mobility”, “double mobility”, “hip”, “cup”, “socket”. The following data were extracted by two investigators independently: demographics, whether the operation was a primary or revision THA, length of follow-up, the design of the components, diameter of the femoral head, and type of fixation of the acetabular component. Results. In all, 59 articles met our inclusion criteria. These included a total of 17 908 THAs which were divided into two groups: studies dealing with DM components in primary THA and those dealing with these components in revision THA. The mean rate of dislocation was 0.9% in the primary THA group, and 3.0% in the revision THA group. The mean rate of intraprosthetic dislocation was 0.7% in primary and 1.3% in revision THAs. Conclusion. Based on the current data, the use of DM acetabular components are effective in minimising the risk of instability after both primary and revision THA. This benefit must be balanced against continuing concerns about the additional modularity, and the new mode of failure of intraprosthetic dislocation. Longer term studies are needed to assess the function of these newer materials compared with previous generations. . Cite this article: Bone Joint J 2017;99-B(1 Supple A):18–24


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 365 - 371
1 Apr 2019
Nam D Salih R Nahhas CR Barrack RL Nunley RM

Aims. Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design. Patients and Methods. This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m. 2. and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (. sd. 6.5)) were enrolled. All patients had a minimum of two years’ clinical follow-up. Patient-reported outcome measures, whole blood metal ion levels (ug/l), and periprosthetic femoral BMD were measured at baseline, as well as at one and two years postoperatively. Power analysis indicated 40 patients necessary to demonstrate a five-fold increase in cobalt levels from baseline (alpha = 0.05, beta = 0.80). A mixed model with repeated measures was used for statistical analysis. Results. Mean Harris Hip Scores improved from 54.1 (. sd. 20.5) to 91.2 (. sd. 10.8) at two years postoperatively (p < 0.001). All patients had radiologically well-fixed components, no patients experienced any instability, and no patients required any further intervention. Mean cobalt levels increased from 0.065 ug/l (. sd. 0.03) preoperatively to 0.30 ug/l (. sd. 0.51) at one year postoperatively (p = 0.01) but decreased at two years postoperatively to 0.16 ug/l (. sd. 0.23; p = 0.2). Four patients (9.3%) had a cobalt level outside the reference range (0.03 ug/l to 0.29 ug/l) at two years postoperatively, with values from 0.32 ug/l to 0.94 ug/l. The mean femoral BMD ratio was maintained in Gruen zones 2 to 7 at both one and two years postoperatively using this stem design. At two years postoperatively, mean BMD in the medial calcar was 101.5% of the baseline value. Conclusion. Use of a modular DM prosthesis and cementless, tapered femoral stem has shown encouraging results in young, active patients undergoing primary THA. Elevation in mean cobalt levels and the presence of four patients outside the reference range at two years postoperatively demonstrates the necessity of continued surveillance in this cohort. Cite this article: Bone Joint J 2019;101-B:365–371


Aims. The optimal treatment for independent patients with a displaced intracapsular fracture of the hip remains controversial. The recognised alternatives are hemiarthroplasty and total hip arthroplasty. At present there is no established standard of care, with both types of arthroplasty being used in many centres. Patients and Methods. We conducted a feasibility study comparing the clinical effectiveness of a dual mobility acetabular component compared with standard polyethylene component in total hip arthroplasty for independent patients with a displaced intracapsular fracture of the hip, for a 12-month period beginning in June 2013. The primary outcome was the risk of dislocation one year post-operatively. Secondary outcome measures were EuroQol 5 Dimensions, ICEpop CAPability measure for Older people, Oxford hip score, mortality and re-operation. Results. Only 20 patients were recruited during this time. The baseline demographics were similar in the two groups and no patient suffered a dislocation. Differences in secondary outcomes were not analysed due to the small sample. Conclusion. This feasibility study suggests that any trial investigating the effectiveness of total hip arthroplasty for fracture of the hip might not be deliverable within the constraints of current systems of care in the United Kingdom. Cite this article: Bone Joint J 2016;98-B:1431–5


Bone & Joint Open
Vol. 2, Issue 10 | Pages 858 - 864
18 Oct 2021
Guntin J Plummer D Della Valle C DeBenedetti A Nam D

Aims. Prior studies have identified that malseating of a modular dual mobility liner can occur, with previous reported incidences between 5.8% and 16.4%. The aim of this study was to determine the incidence of malseating in dual mobility implants at our institution, assess for risk factors for liner malseating, and investigate whether liner malseating has any impact on clinical outcomes after surgery. Methods. We retrospectively reviewed the radiographs of 239 primary and revision total hip arthroplasties with a modular dual mobility liner. Two independent reviewers assessed radiographs for each patient twice for evidence of malseating, with a third observer acting as a tiebreaker. Univariate analysis was conducted to determine risk factors for malseating with Youden’s index used to identify cut-off points. Cohen’s kappa test was used to measure interobserver and intraobserver reliability. Results. In all, 12 liners (5.0%), including eight Stryker (6.8%) and four Zimmer Biomet (3.3%), had radiological evidence of malseating. Interobserver reliability was found to be 0.453 (95% confidence interval (CI) 0.26 to 0.64), suggesting weak inter-rater agreement, with strong agreement being greater than 0.8. We found component size of 50 mm or less to be associated with liner malseating on univariate analysis (p = 0.031). Patients with malseated liners appeared to have no associated clinical consequences, and none required revision surgery at a mean of 14 months (1.4 to 99.2) postoperatively. Conclusion. The incidence of liner malseating was 5.0%, which is similar to other reports. Component size of 50 mm or smaller was identified as a risk factor for malseating. Surgeons should be aware that malseating can occur and implant design changes or changes in instrumentation should be considered to lower the risk of malseating. Although further follow-up is needed, it remains to be seen if malseating is associated with any clinical consequences. Cite this article: Bone Jt Open 2021;2(10):858–864


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 60 - 63
1 Jan 2016
Ko LM Hozack WJ

Dual mobility cups have two points of articulation, one between the shell and the polyethylene (external bearing) and one between the polyethylene and the femoral head (internal bearing). Movement occurs at the inner bearing; the outer bearing only moves at extremes of movement. Dislocation after total hip arthroplasty (THA) is a cause of much morbidity and its treatment has significant cost implications. Dual mobility cups provide an increased range of movement and a may reduce the risk of dislocation. . This paper reviews the use of these cups in THA, particularly where stability is an issue. Dual mobility cups may be of benefit in primary THA in patients at a high risk of dislocation, such as those who are older with increased comorbidities and a higher American Association of Anesthesiology grade and those with a neuromuscular disease. They may be used at revision surgery where the risk of dislocation is high, such as in patients with many prior dislocations, or those with abductor deficiency. They may also be used in THA for displaced fractures of the femoral neck, which has a notoriously high rate of dislocation. Cite this article: Bone Joint J 2016;98-B(1 Suppl A):60–3


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1238 - 1246
1 Jul 2021
Hemmerling KJ Weitzler L Bauer TW Padgett DE Wright TM

Aims. Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. Methods. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy. Results. Fretting was present on 88% (53/60) of the retrieved liners, and corrosion was present on 97% (58/60). Fretting was most often found on the lip of the taper at the transition between the lip and the dome regions. Macrophages and particles reflecting an innate inflammatory reaction to corrosion debris were noted in six of the 48 cases for which periprosthetic tissues were examined, and all were associated with retrieved components that had high corrosion scores. Conclusion. Our results show that corrosion occurs at the interface between MDM liners and shells and that it can be associated with reactions in the local tissues, suggesting continued concern that this problem may become clinically important with longer-term use of these implants. Cite this article: Bone Joint J 2021;103-B(7):1238–1246


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 811 - 821
1 Jul 2020
You D Sepehri A Kooner S Krzyzaniak H Johal H Duffy P Schneider P Powell J

Aims. Dislocation is the most common indication for further surgery following total hip arthroplasty (THA) when undertaken in patients with a femoral neck fracture. This study aimed to assess the complication rates of THA with dual mobility components (THA-DMC) following a femoral neck fracture and to compare outcomes between THA-DMC, conventional THA, and hemiarthroplasty (HA). Methods. We performed a systematic review of all English language articles on THA-DMC published between 2010 and 2019 in the MEDLINE, EMBASE, and Cochrane databases. After the application of rigorous inclusion and exclusion criteria, 23 studies dealing with patients who underwent treatment for a femoral neck fracture using THA-DMC were analyzed for the rate of dislocation. Secondary outcomes included reoperation, periprosthetic fracture, infection, mortality, and functional outcome. The review included 7,189 patients with a mean age of 77.8 years (66.4 to 87.6) and a mean follow-up of 30.9 months (9.0 to 68.0). Results. THA-DMC was associated with a significantly lower dislocation rate compared with both THA (OR 0.26; 95% CI 0.08 to 0.79) and HA (odds ratio (OR) 0.27; 95% confidence interval (CI) 0.15 to 0.50). The rate of large articulations and of intraprosthetic dislocation was 1.5% (n = 105) and 0.04% (n = 3) respectively. Conclusion. THA-DMC when used in patients with a femoral neck fracture is associated with a lower dislocation rate compared with conventional arthroplasty options. There was no increase in the rates of other complication when THA-DMC was used. Future cost analysis and prospective, comparative studies are required to assess the potential benefit of using THA-DMC in these patients. Cite this article: Bone Joint J 2020;102-B(7):811–821


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 66 - 72
1 Jul 2021
Hernandez NM Hinton ZW Wu CJ Lachiewicz PF Ryan SP Wellman SS

Aims. Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years. Methods. Using the database of a single academic centre, 126 revision THAs in 117 patients using a single design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome. Results. The overall rate of dislocation was 11%, with a six-year survival of 91%. Reoperation for dislocation was performed in seven patients (7%), with a six-year survival of 94%. The dislocations were early (at a mean of 33 days) in six patients, and late (at a mean of 4.3 years) in four patients. There were three intraprosthetic dissociations. An outer head diameter of ≥ 48 mm was associated with a lower risk of dislocation (p = 0.013). Lumbrosacral fusion was associated with increased dislocation (p = 0.004). Four revision THAs (4%) were further revised for aseptic acetabular loosening, and severe bone loss (Paprosky III) at the time of the initial revision was significantly associated with further revision for aseptic acetabular loosening (p = 0.008). Fourteen acetabular components (15%) were re-revised for infection, and a pre-revision diagnosis of reimplantation after periprosthetic joint infection (PJI) was associated with subsequent PJI (p < 0.001). Two THAs had visible metallic changes on the backside of the cobalt chromium liner. Conclusion. When using this MDM component in revision THA, at a mean follow-up of 5.5 years, there was a higher rate of dislocation (11%) than previously reported. The size of the outer bearing was related to the risk of dislocation. There was a low rate of aseptic acetabular loosening. Longer follow-up of this MDM component and evaluation of other designs are warranted. Cite this article: Bone Joint J 2021;103-B(7 Supple B):66–72


Bone & Joint Research
Vol. 7, Issue 8 | Pages 508 - 510
1 Aug 2018
Horriat S Haddad FS


Objectives. The annual incidence of hip fracture is 620 000 in the European Union. The cost of this clinical problem has been estimated at 1.75 million disability-adjusted life years lost, equating to 1.4% of the total healthcare burden in established market economies. Recent guidance from The National Institute for Health and Clinical Excellence (NICE) states that research into the clinical and cost effectiveness of total hip arthroplasty (THA) as a treatment for hip fracture is a priority. We asked the question: can a trial investigating THA for hip fracture currently be delivered in the NHS?. Methods. We performed a contemporaneous process evaluation that provides a context for the interpretation of the findings of WHiTE Two – a randomised study of THA for hip fracture. We developed a mixed methods approach to situate the trial centre within the context of wider United Kingdom clinical practice. We focused on fidelity, implementation, acceptability and feasibility of both the trial processes and interventions to stakeholder groups, such as healthcare providers and patients. Results. We have shown that patients are willing to participate in this type of research and that surgeons value being part of a team that has a strong research ethos. However, surgical practice does not currently reflect NICE guidance. Current models of service delivery for hip fractures are unlikely to be able to provide timely total hip arthroplasty for suitable patients. Conclusions. Further observational research should be conducted to define the population of interest before future interventional studies are performed. Cite this article: C. Huxley, J. Achten, M. L. Costa, F. Griffiths, X. L. Griffin. A process evaluation of the WHiTE Two trial comparing total hip arthroplasty with and without dual mobility component in the treatment of displaced intracapsular fractures of the proximal femur: Can a trial investigating total hip arthroplasty for hip fracture be delivered in the NHS? Bone Joint Res 2016;5:444–452. DOI: 10.1302/2046-3758.510.BJR-2015-0008.R1


Bone & Joint 360
Vol. 10, Issue 1 | Pages 4 - 9
1 Feb 2021
White JJE Manktelow ARJ


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1297 - 1302
1 Oct 2018
Elbuluk AM Slover J Anoushiravani AA Schwarzkopf R Eftekhary N Vigdorchik JM

Aims

The routine use of dual-mobility (DM) acetabular components in total hip arthroplasty (THA) may not be cost-effective, but an increasing number of patients undergoing THA have a coexisting spinal disorder, which increases the risk of postoperative instability, and these patients may benefit from DM articulations. This study seeks to examine the cost-effectiveness of DM components as an alternative to standard articulations in these patients.

Patients and Methods

A decision analysis model was used to evaluate the cost-effectiveness of using DM components in patients who would be at high risk for dislocation within one year of THA. Direct and indirect costs of dislocation, incremental costs of using DM components, quality-adjusted life-year (QALY) values, and the probabilities of dislocation were derived from published data. The incremental cost-effectiveness ratio (ICER) was established with a willingness-to-pay threshold of $100 000/QALY. Sensitivity analysis was used to examine the impact of variation.


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 41 - 45
1 Jan 2019
Jones CW De Martino I D’Apolito R Nocon AA Sculco PK Sculco TP

Aims. Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years. Materials and Methods. We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum. Results. There were 151 patients who met the classification of high-risk according to the inclusion criteria and received DM THA during the study period. Mean age was 82 years old (73 to 95) and 114 patients (77.5%) were female. Mean follow-up was 3.6 years (1.9 to 6.1), with five patients lost to follow-up and one patient who died (for a reason unrelated to the index procedure). One patient (0.66%) sustained an intraprosthetic dislocation; there were no other dislocations. Conclusion. At mid-term follow-up, the use of a DM bearing for primary THA in patients at high risk of dislocation provided a stable reconstruction option with excellent radiographic results. Longer follow-up is needed to confirm the durability of these reconstructions


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 17 - 24
1 Jul 2021
Vigdorchik JM Sharma AK Buckland AJ Elbuluk AM Eftekhary N Mayman DJ Carroll KM Jerabek SA

Aims. Patients with spinal pathology who undergo total hip arthroplasty (THA) have an increased risk of dislocation and revision. The aim of this study was to determine if the use of the Hip-Spine Classification system in these patients would result in a decreased rate of postoperative dislocation in patients with spinal pathology. Methods. This prospective, multicentre study evaluated 3,777 consecutive patients undergoing THA by three surgeons, between January 2014 and December 2019. They were categorized using The Hip-Spine Classification system: group 1 with normal spinal alignment; group 2 with a flatback deformity, group 2A with normal spinal mobility, and group 2B with a stiff spine. Flatback deformity was defined by a pelvic incidence minus lumbar lordosis of > 10°, and spinal stiffness was defined by < 10° change in sacral slope from standing to seated. Each category determined a patient-specific component positioning. Survivorship free of dislocation was recorded and spinopelvic measurements were compared for reliability using intraclass correlation coefficient. Results. A total of 2,081 patients met the inclusion criteria. There were 987 group 1A, 232 group 1B, 715 group 2A, and 147 group 2B patients. A total of 70 patients had a lumbar fusion, most had L4-5 (16; 23%) or L4-S1 (12; 17%) fusions; 51 patients (73%) had one or two levels fused, and 19 (27%) had > three levels fused. Dual mobility (DM) components were used in 166 patients (8%), including all of those in group 2B and with > three level fusions. Survivorship free of dislocation at five years was 99.2% with a 0.8% dislocation rate. The correlation coefficient was 0.83 (95% confidence interval 0.89 to 0.91). Conclusion. This is the largest series in the literature evaluating the relationship between hip-spine pathology and dislocation after THA, and guiding appropriate treatment. The Hip-Spine Classification system allows surgeons to make appropriate evaluations preoperatively, and it guides the use of DM components in patients with spinopelvic pathology in order to reduce the risk of dislocation in these high-risk patients. Cite this article: Bone Joint J 2021;103-B(7 Supple B):17–24


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 73 - 77
1 Jul 2021
Lawrie CM Barrack RL Nunley RM

Aims. Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA. Methods. This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m. 2. , and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless titanium femoral stem. Patient-reported outcome measures, whole blood metal ion levels (μg/l), and periprosthetic femoral BMD were measured at baseline and at one, two, and five years postoperatively. The results two years postoperatively for this cohort have been previously reported. Results. A total of 43 patients were enrolled. At minimum follow-up of five years, 23 (53.4%) returned for clinical and radiological review, 25 (58.1%) had metal ion analysis performed, 19 (44.2%) underwent dual energy x-ray absorptiometry scans, and 25 (58%) completed a pain-drawing questionnaire. The mean modified Harris Hip Scores improved significantly from 54.8 (SD 19) preoperatively to 93.08 (SD 10.5) five years postoperatively (p < 0.001). One patient was revised for aseptic acetabular loosening. The mean cobalt levels increased from 0.065 μg/l (SD 0.03) to 0.08 (SD 0.05) and the mean titanium levels increased from 0.35 (SD 0.13) to 0.78 (SD 0.29). The femoral BMD ratio decreased in Gruen Zone 1 (91.9%) at five years postoperatively compared with the baseline scores at six weeks potoperatively. The femoral BMD ratio was maintained in Gruen zones 2 to 7. Conclusion. The use of a modular DM component and a cementless, tapered femoral stem shows excellent mid-term survivorship with minimal concerns for corrosion and metal ion release in a cohort of young, active patients undergoing primary THA. Cite this article: Bone Joint J 2021;103-B(7 Supple B):73–77


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 89 - 96
1 Mar 2024
Heckmann ND Chung BC Liu KC Chen XT Lovro LR Kistler NM White E Christ AB Longjohn DB Oakes DA Lieberman JR

Aims

Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations.

Methods

A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.