Aims. There are concerns regarding initial stability and early periprosthetic fractures in
We studied the pattern of . 99m. Tc-methylene diphosphonate uptake around uncemented femoral components in 44 asymptomatic hip arthroplasties, performing isotope scans at intervals from 4 to 48 months after operation. We used phase-III images obtained with a high-resolution gamma camera and measured the activity in various zones using a specially designed computer program. The components studied at 4, 6, 9 and 12 months were coated with hydroxyapatite (HA) and those studied at 18, 24, 36 and 48 months were not coated. We found a statistically significant fall in activity between four and six months around HA-coated prostheses in all five femoral periprosthetic zones. After six months activity was relatively uniform, but remained higher than that in normal femoral bone at 48 months in non-coated prostheses. We discuss the application of these patterns in the evaluation of painful
We have analysed the clinical results of a
We present the case of a 65-year-old man who developed a malignant fibrous histiocytoma at the lesser trochanter five months after a
Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (Aims
Patients and Methods
Since 1964, 3573 uncemented Ring total hip replacements of all types have been performed; 194 (5.4%) have required revision for infection, loosening or breakage. Of these, 102 were available for review; 81 had satisfactory results five years after operation. These results are better than many of those reported after revision of cemented hip replacements. It is suggested that the absence of cement makes revision easier because bone erosion rarely occurs.
Delayed sciatic nerve palsy is uncommon after primary hip replacement. Two kinds of sciatic palsy have been reported with regard to the time of onset: early palsy related to wound haematoma or lumbosacral nerve elongation which occurs between surgery and 18 days, is more frequent than delayed palsy, occurring between 10 and 32 months, which is usually caused by cement extrusion or heat produced by cement polymerisation. We present two cases of delayed, transient sciatic nerve palsy arising at three weeks and four months after primary cementless arthroplasty, respectively, without haematoma and with a normal lumbar spine. These palsies were possibly caused by excessive tension from minor limb lengthening of 2 cm to 4 cm required to achieve leg-length equality. As the initial symptoms were limited to calf pain and mild numbness in the foot, surgeons should be aware of this mode of onset, particularly when it is delayed after hip replacement. Both patients recovered fully by 12 months after surgery so we did not undertake surgical exploration of the nerve in either patient.
We have studied the role of computerised axial tomography in the assessment of the acetabulum before an uncemented Mittelmeier arthroplasty was performed in 20 patients. In the absence of gross anatomical abnormality, computerised tomography did not provide additional information, but in eight patients it was useful in determining adequacy of acetabular depth, anterior acetabular cover, and thickness of the anterior and posterior walls. In these patients, the CT scans provided information which was not available from plain radiographs and greatly assisted in pre-operative planning.
The popularity of cementless total hip replacement
(THR) has surpassed cemented THR in England and Wales. This retrospective
cohort study records survival time to revision following primary
cementless THR with the most common combination (accounting for
almost a third of all cementless THRs), and explores risk factors independently
associated with failure, using data from the National Joint Registry
for England and Wales. Patients with osteoarthritis who had a DePuy
Corail/Pinnacle THR implanted between the establishment of the registry
in 2003 and 31 December 2010 were included within analyses. There
were 35 386 procedures. Cox proportional hazard models were used
to analyse the extent to which the risk of revision was related
to patient, surgeon and implant covariates. The overall rate of
revision at five years was 2.4% (99% confidence interval 2.02 to
2.79). In the final adjusted model, we found that the risk of revision
was significantly higher in patients receiving metal-on-metal (MoM:
hazard ratio (HR) 1.93, p <
0.001) and ceramic-on-ceramic bearings
(CoC: HR 1.55, p = 0.003) compared with the best performing bearing
(metal-on-polyethylene). The risk of revision was also greater for
smaller femoral stems (sizes 8 to 10: HR 1.82, p <
0.001) compared
with mid-range sizes. In a secondary analysis of only patients where body
mass index (BMI) data were available (n = 17 166), BMI ≥ 30 kg/m2 significantly
increased the risk of revision (HR 1.55, p = 0.002). The influence
of the bearing on the risk of revision remained significant (MoM:
HR 2.19, p <
0.001; CoC: HR 2.09,
p = 0.001). The risk of revision was independent of age, gender,
head size and offset, shell, liner and stem type, and surgeon characteristics. We found significant differences in failure between bearing surfaces
and femoral stem size after adjustment for a range of covariates
in a large cohort of single-brand cementless THRs. In this study
of procedures performed since 2003, hard bearings had significantly
higher rates of revision, but we found no evidence that head size
had an effect. Patient characteristics, such as BMI and American
Society of Anesthesiologists grade, also influence the survival
of cementless components. Cite this article:
A retrospective review of 148 consecutive porous-coated hip arthroplasties (PCA) showed an incidence of thigh pain of 13% one year after surgery, and 22% at two years. Positive correlations were made with femoral stem subsidence (greater than 2 mm) and with distal periosteal and endosteal bone formation. No positive correlations were made with parameters of bone quality or component fit. Resolution of pain occurred in one-third and an anti-inflammatory agent produced partial relief in two-thirds of the patients. We conclude that thigh pain is secondary to stem instability with distal stress transfer in the absence of stable proximal fixation.
Aims. Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane. Methods. Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation. Results. Overall, 26/40 hips had a centre of rotation displaced posteriorly compared to the contralateral hip, increasing to 33/40 once corrected for sagittal tilt, with a mean posterior displacement of 7 mm. Linear regression analysis indicated that stem anteversion needed to be increased by 10.8° to recreate the head centre in the AP plane. Merely matching the native version would result in a 12 mm posterior displacement. Conclusion. This study demonstrates the significant incidence of posterior displacement of the head centre in
Bilateral,
This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.Aims
Methods
This study aimed to develop and validate a fully automated system that quantifies proximal femoral bone mineral density (BMD) from CT images. The study analyzed 978 pairs of hip CT and dual-energy X-ray absorptiometry (DXA) measurements of the proximal femur (DXA-BMD) collected from three institutions. From the CT images, the femur and a calibration phantom were automatically segmented using previously trained deep-learning models. The Hounsfield units of each voxel were converted into density (mg/cm3). Then, a deep-learning model trained by manual landmark selection of 315 cases was developed to select the landmarks at the proximal femur to rotate the CT volume to the neutral position. Finally, the CT volume of the femur was projected onto the coronal plane, and the areal BMD of the proximal femur (CT-aBMD) was quantified. CT-aBMD correlated to DXA-BMD, and a receiver operating characteristic (ROC) analysis quantified the accuracy in diagnosing osteoporosis.Aims
Methods
The aim of this study was to evaluate the ten-year
clinical and functional outcome of hip resurfacing and to compare it
with that of
In a prospective study, 93 unselected consecutive
A leading problem with
We reviewed a consecutive series of 241 uncemented, porous-coated anatomic (PCA) hip replacements at an average follow-up of five years (2 to 9). Of these, 32 had failed (13%), 26 at the acetabular component (11%) and six at the femoral component (2%). Acetabular failure was associated with local osteolysis and excessive polyethylene wear in 20 cases: in these histological examination showed giant macrophages incorporating numerous particles of high-density polyethylene. The femoral failures were related to a poor intramedullary fit with subsequent subsidence. Using the recommendation for revision as the end point, the cumulative survival rate for prostheses was 91% at six years (95% CI +/- 6%), 73% (+/- 11%) at seven years, and 57% (+/- 20%) at eight years. The result of uncemented PCA hip replacement is satisfactory up to six years, but then increasing failure of the acetabular component appears to be due to polyethylene wear, leading to osteolysis, loosening and component migration. At first, failure is often asymptomatic; routine follow-up of