Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Results of uncemented porous-coated anatomic total hip replacement



Download PDF

Abstract

We reviewed a consecutive series of 241 uncemented, porous-coated anatomic (PCA) hip replacements at an average follow-up of five years (2 to 9). Of these, 32 had failed (13%), 26 at the acetabular component (11%) and six at the femoral component (2%). Acetabular failure was associated with local osteolysis and excessive polyethylene wear in 20 cases: in these histological examination showed giant macrophages incorporating numerous particles of high-density polyethylene. The femoral failures were related to a poor intramedullary fit with subsequent subsidence. Using the recommendation for revision as the end point, the cumulative survival rate for prostheses was 91% at six years (95% CI +/- 6%), 73% (+/- 11%) at seven years, and 57% (+/- 20%) at eight years. The result of uncemented PCA hip replacement is satisfactory up to six years, but then increasing failure of the acetabular component appears to be due to polyethylene wear, leading to osteolysis, loosening and component migration. At first, failure is often asymptomatic; routine follow-up of uncemented hip replacement is essential, especially after five years.

For access options please click here