Advertisement for orthosearch.org.uk
Results 1 - 13 of 13
Results per page:
The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 88 - 96
1 Jan 2023
Vogt B Rupp C Gosheger G Eveslage M Laufer A Toporowski G Roedl R Frommer A

Aims. Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. Methods. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages. Results. Osteolysis next to the telescopic junction was observed in 31/48 segments (65%) lengthened with the STRYDE nail before implant removal compared to 1/91 segment (1%) in the PRECICE cohort. In the STRYDE cohort, osteolysis initially increased, but decreased or resolved in almost all lengthened segments (86%) after implant removal. Implant failure was observed in 9/48 STRYDE (19%) and in 8/92 PRECICE nails (9%). Breakage of the distal locking bolts was found in 5/48 STRYDE nails (10%) compared to none in the PRECICE cohort. Treatment-associated pain was generally recorded as mild and found in 30/48 patients (63%) and 39/92 (42%) in the STRYDE and PRECICE cohorts, respectively. Temporary range of motion (ROM) limitations under distraction were registered in 17/48 (35%) segments treated with the STRYDE and 35/92 segments (38%) treated with the PRECICE nail. Conclusion. Osteolysis and periosteal reaction, implant breakage, and pain during lengthening and consolidation is more likely in patients treated with the STRYDE nail compared to the PRECICE nail. Temporary ROM limitations during lengthening occurred independent of the applied device. Implant-related osseous alterations seem to remodel after implant removal. Cite this article: Bone Joint J 2023;105-B(1):88–96


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1731 - 1735
1 Nov 2021
Iobst CA Frost MW Rölfing JD Rahbek O Bafor A Duncan M Kold S

Aims. Limb-lengthening nails have largely replaced external fixation in limb-lengthening and reconstructive surgery. However, the adverse events and high prevalence of radiological changes recently noted with the STRYDE lengthening nail have raised concerns about the use of internal lengthening nails. The aim of this study was to compare the prevalence of radiological bone abnormalities between STRYDE, PRECICE, and FITBONE nails prior to nail removal. Methods. This was a retrospective case series from three centres. Patients were included if they had either of the three limb-lengthening nails (STYDE, PRECICE, or FITBONE) removed. Standard orthogonal radiographs immediately prior to nail removal were examined for bone abnormalities at the junction of the telescoping nail parts. Results. In total, 306 patients (168 male, 138 female) had 366 limb-lengthening nails removed. The mean time from nail insertion to radiological evaluation was 434 days (36 to 3,015). Overall, 77% of STRYDE nails (20/26) had bone abnormalities at the interface compared with only 2% of FITBONE (4/242) and 1% of PRECICE nails (1/98; p < 0.001). Focal osteolysis in conjunction with periosteal reaction at the telescoping interface was only observed in STRYDE nails. Conclusion. Bone abnormalities at the interface of telescoping nail parts were seen in the majority of STRYDE nails, but only very rarely with FITBONE or PRECICE nails. We conclude that the low prevalence of radiological changes at the junctional interface of 242 FITBONE and 98 PRECICE nails at the time of nail removal does not warrant clinical concerns. Cite this article: Bone Joint J 2021;103-B(11):1731–1735


Bone & Joint Research
Vol. 7, Issue 7 | Pages 476 - 484
1 Jul 2018
Panagiotopoulou VC Davda K Hothi HS Henckel J Cerquiglini A Goodier WD Skinner J Hart A Calder PR

Objectives. The Precice nail is the latest intramedullary lengthening nail with excellent early outcomes. Implant complications have led to modification of the nail design. The aim of this study was to perform a retrieval study of Precice nails following lower-limb lengthening and to assess macroscopical and microscopical changes to the implants and evaluate differences following design modification, with the aim of identifying potential surgical, implant, and patient risk factors. Methods. A total of 15 nails were retrieved from 13 patients following lower-limb lengthening. Macroscopical and microscopical surface damage to the nails were identified. Further analysis included radiology and micro-CT prior to sectioning. The internal mechanism was then analyzed with scanning electron microscopy and energy dispersive x-ray spectroscopy to identify corrosion. Results. Seven male and three female patients underwent 12 femoral lengthenings. Three female patients underwent tibial lengthening. All patients obtained the desired length with no implant failure. Surface degradation was noted on the telescopic part of every nail design, less on the latest implants. Microscopical analysis confirmed fretting and pitting corrosion. Following sectioning, black debris was noted in all implants. The early designs were found to have fractured actuator pins and the pin and bearings showed evidence of corrosive debris. The latest designs showed evidence of biological deposits suggestive of fluid ingress within the nail but no corrosion. Conclusion. This study confirms less internal corrosion following modification, but evidence of titanium debris remains. We recommend no change to current clinical practice. However, potential reuse of the Precice nail, for secondary limb lengthening in the same patient, should be undertaken with caution. Cite this article: V. C. Panagiotopoulou, K. Davda, H. S. Hothi, J. Henckel, A. Cerquiglini, W. D. Goodier, J. Skinner, A. Hart, P. R. Calder. A retrieval analysis of the Precice intramedullary limb lengthening system. Bone Joint Res 2018;7:476–484. DOI: 10.1302/2046-3758.77.BJR-2017-0359.R1


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1168 - 1176
1 Sep 2019
Calder PR McKay JE Timms AJ Roskrow T Fugazzotto S Edel P Goodier WD

Aims. The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients. Patients and Methods. A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications. Results. The mean lengthening was 4.65 cm (1.5 to 8) in the antegrade group and 4.64 cm (1.6 to 8) in the retrograde group. Of the 107 lengthenings, 100 had sufficient datapoints to calculate the mean HI. This was 31.6 days/cm (15 to 108). There was a trend toward a lower (better) HI with an antegrade nail and better outcomes in adolescent patients, but these were not statistically significant. Hip and knee ROM was maintained and/or improved following commencement of femoral lengthening in 44 patients (60%) of antegrade nails and 13 patients (38%) of retrograde nails. In female patients, loss of movement occurred both earlier and following less total length achieved. Minor implant complications included locking bolt migration and in one patient deformity of the nail, but no implant failed to lengthen and there were no deep infections. Three patients had delayed union, five patients required surgical intervention for joint contracture. Conclusion. This study confirms excellent results in femoral lengthening with antegrade and retrograde Precice nails. There is a trend for better healing and less restriction in hip and knee movement following antegrade nails. There are clinical scenarios, that mandate the use of a retrograde nail. However, when these are not present, we recommend the use of antegrade nailing. Cite this article: Bone Joint J 2019;101-B:1168–1176


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1382 - 1388
1 Oct 2016
Laubscher M Mitchell C Timms A Goodier D Calder P

Aims. Patients undergoing femoral lengthening by external fixation tolerate treatment less well when compared to tibial lengthening. Lengthening of the femur with an intramedullary device may have advantages. Patients and Methods. We reviewed all cases of simple femoral lengthening performed at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities, congenital limb deficiencies and lengthening with an unstable hip were excluded, leaving 33 cases (in 22 patients; 11 patients had bilateral procedures) for review. Healing index, implant tolerance and complications were compared. Results. In 20 cases (15 patients) the Precice lengthening nail was used and in 13 cases (seven patients) the LRS external fixator system. The desired length was achieved in all cases in the Precice group and in 12 of 13 cases in the LRS group. The mean healing index was 31.3 days/cm in the Precice and 47.1 days/cm in the LRS group (p < 0.001). This was associated with an earlier ability to bear full weight without aids in the Precice group. There were more complications with LRS lengthening, including pin site infections and regenerate deformity. Implant tolerance and the patients’ perception of the cosmetic result were better with the Precice treatment. Conclusion. Femoral lengthening with the Precice femoral nail achieved excellent functional results with fewer complications and greater patient satisfaction when compared with the LRS system in our patients. Cite this article: Bone Joint J 2016;98-B:1382–8


Bone & Joint 360
Vol. 11, Issue 5 | Pages 6 - 8
1 Oct 2022
Jamal B Calder P


Bone & Joint 360
Vol. 11, Issue 3 | Pages 40 - 43
1 Jun 2022


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1168 - 1172
1 Jun 2021
Iliadis AD Wright J Stoddart MT Goodier WD Calder P

Aims

The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device.

Methods

This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies.


Bone & Joint Open
Vol. 2, Issue 8 | Pages 599 - 610
1 Aug 2021
Hothi H Bergiers S Henckel J Iliadis AD Goodier WD Wright J Skinner J Calder P Hart AJ

Aims

The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices.

Methods

We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions.


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims

This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported.

Methods

This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed.


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 178 - 188
1 Feb 2019
Chaudhary MM Lakhani PH

Aims

Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator.

Patients and Methods

A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score.


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1005 - 1006
1 Jun 2021
Haddad FS


Bone & Joint 360
Vol. 3, Issue 3 | Pages 34 - 37
1 Jun 2014

The June 2014 Children’s orthopaedics Roundup360 looks at: plaster wedging in paediatric forearm fractures; the medial approach for DDH; Ponseti – but not as he knew it?; Salter osteotomy more accurate than Pemberton in DDH; is the open paediatric fracture an emergency?; bang up-to-date with femoral external fixation; indomethacin, heterotopic ossification and cerebral palsy hips; lengthening nails for congenital femoral deformities, and is MRI the answer to imaging of the physis?