Glenoid bone loss can be a challenging problem when revising
a shoulder arthroplasty. Precise pre-operative planning based on
plain radiographs or CT scans is essential. We have investigated
a new radiological classification system to describe the degree
of medialisation of the bony glenoid and that will indicate the
amount of bone potentially available for supporting a glenoid component.
It depends on the relationship between the most medial part of the
articular surface of the glenoid with the base of the coracoid process
and the spinoglenoid notch: it classifies the degree of bone loss
into three types. It also attempts to predict the type of glenoid reconstruction
that may be possible (impaction bone grafting, structural grafting
or simple non-augmented arthroplasty) and gives guidance about whether
a pre-operative CT scan is indicated. Inter-method reliability between plain radiographs and CT scans
was assessed retrospectively by three independent observers using
data from 39 randomly selected patients. Inter-observer reliability and test-retest reliability was tested
on the same cohort using Cohen's kappa statistics. Correlation of
the type of glenoid with the Constant score and its pain component
was analysed using the Kruskal-Wallis method on data from 128 patients.
Anatomical studies of the scapula were reviewed to explain the findings.Aims
Patients and Methods
Dislocation of the acromioclavicular joint is
a relatively common injury and a number of surgical interventions
have been described for its treatment. Recently, a synthetic ligament
device has become available and been successfully used, however,
like other non-native solutions, a compromise must be reached when
choosing non-anatomical locations for their placement. This cadaveric
study aimed to assess the effect of different clavicular anchorage points
for the Lockdown device on the reduction of acromioclavicular joint
dislocations, and suggest an optimal location. We also assessed
whether further stability is provided using a coracoacromial ligament
transfer (a modified Neviaser technique). The acromioclavicular
joint was exposed on seven fresh-frozen cadaveric shoulders. The
joint was reconstructed using the Lockdown implant using four different
clavicular anchorage points and reduction was measured. The coracoacromial
ligament was then transferred to the lateral end of the clavicle,
and the joint re-assessed. If the Lockdown ligament was secured
at the level of the conoid tubercle, the acromioclavicular joint
could be reduced anatomically in all cases. If placed medial or
2 cm lateral, the joint was irreducible. If the Lockdown was placed
1 cm lateral to the conoid tubercle, the joint could be reduced
with difficulty in four cases. Correct placement of the Lockdown
device is crucial to allow anatomical joint reduction. Even when the
Lockdown was placed over the conoid tubercle, anterior clavicle
displacement remained but this could be controlled using a coracoacromial
ligament transfer. Cite this article:
The LockDown device (previously called Surgilig)
is a braided polyester mesh which is mostly used to reconstruct the
dislocated acromioclavicular joint. More than 11 000 have been implanted
worldwide. Little is known about the tissue reaction to the device
nor to its wear products when implanted in an extra-articular site
in humans. This is of importance as an adverse immunological reaction
could result in osteolysis or damage to the local tissues, thereby affecting
the longevity of the implant. We analysed the histology of five LockDown implants retrieved
from five patients over the last seven years by one of the senior
authors. Routine analysis was carried out in all five cases and
immunohistochemistry in one. The LockDown device acts as a scaffold for connective tissue
which forms an investing fibrous pseudoligament. The immunological
response at the histological level seems favourable with a limited
histiocytic and giant cell response to micron-sized wear particles.
The connective tissue envelope around the implant is less organised
than a native ligament. Cite this article:
In the UK we have many surgeon inventors – surgeons who innovate and create new ways of doing things, who invent operations, who design new instruments to facilitate surgery or design new implants for using in patients. However truly successful surgeon inventors are a rare breed and they need to develop additional knowledge and skills during their career in order to push forward their devices and innovations. This article reviews my own experiences as a surgeon inventor and the highs and lows over the whole of my surgical career.
We describe a surgical procedure used to achieve glenothoracic fusion after one-stage radical, near total scapulectomy for primary or secondary malignant lesions in six patients. The aim was complete excision of the lesion and preservation of the glenoid to provide a fulcrum for movement of the arm. Post-operative assessment involved determination of the range of movement and the application of the Musculoskeletal Tumour Society score and the Constant score to evaluate function. The results showed that a functional range of movement was attainable despite such radical surgery, although there was a considerable reduction in power and ability to lift.
We have studied aseptic loosening of the femoral component in 76 patients with primary total hip replacement using the Capital prosthesis. The mean follow-up was 26 months (10 to 37). Twelve femoral components (16%) were definitely and eight (10%) were possibly loose. They were characterised by a thin cement mantle (p <
0.001) and excessive residual cancellous bone in the proximomedial region (p <
0.01). We recommend that the cement mantle around the prosthesis should be 2 to 3 mm and that further long-term studies are needed to evaluate the wear properties of titanium-nitride-coated titanium femoral heads.
The digital nerves to the thumb and index finger have been studied by dissecting twenty-five embalmed upper limbs. The palmar digital nerves to the thumb were constant in position and course, with a short lateral cutaneous branch from the radial palmar digital nerve in 30 per cent of cases. The palmar digital nerves to the index finger had a variable pattern, the commonest arrangement, well described in