Aims. The purpose of the present study was to examine the long-term
fixation of a cemented fixed-bearing polished titanium tibial baseplate
(Genesis ll). . Patients and Methods. Patients enrolled in a previous two-year prospective trial (n
= 35) were recalled at ten years. Available patients (n = 15) underwent
radiostereometric analysis (RSA) imaging in a supine position using
a conventional RSA protocol. Migration of the tibial component in
all planes was compared between initial and ten-year follow-up.
Outcome scores including the Knee Society Score, Western Ontario
and McMaster Universities Arthritis Index, 12-item Short Form Health
Survey, Forgotten Joint Score, and University of California, Los
Angeles Activity Score were recorded. Results. At ten years, the mean migration of the tibial component was
less than 0.1 mm and 0.1° in all planes relative to the post-operative
RSA exam. Maximum total point movement increased with time (p =
0.002) from 0.23 mm (. sd. 0.18) at six weeks to 0.42 mm (. sd. 0.20)
at ten years. Conclusion. The low level of
Aims. This multicentre retrospective observational study’s aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether
Aims. The mean age of patients undergoing total knee arthroplasty (TKA) has reduced with time. Younger patients have increased expectations following TKA. Aseptic loosening of the tibial component is the most common cause of failure of TKA in the UK. Interest in cementless TKA has re-emerged due to its encouraging results in the younger patient population. We review a large series of tantalum trabecular metal cementless implants in patients who are at the highest risk of revision surgery. Methods. A total of 454 consecutive patients who underwent cementless TKA between August 2004 and December 2021 were reviewed. The mean follow-up was ten years. Plain radiographs were analyzed for radiolucent lines. Patients who underwent revision TKA were recorded, and the cause for revision was determined. Data from the National Joint Registry for England, Wales, Northern Island, the Isle of Man and the States of Guernsey (NJR) were compared with our series. Results. No patients in our series had evidence of radiolucent lines on their latest radiological assessment. Only eight patients out of 454 required revision arthroplasty, and none of these revisions were indicated for aseptic loosening of the
Aims. Recent total knee arthroplasty (TKA) designs have featured more anatomical morphologies and shorter tibial keels. However, several reports have raised concerns about the impact of these modifications on implant longevity. The aim of this study was to report the early performance of a modern, cemented TKA design. Methods. All patients who received a primary, cemented TKA between 2012 and 2017 with a minimum two-year follow-up were included. The implant investigated features an asymmetrical
Aims. Our intention was to investigate if the highly porous biological fixation surfaces of a new 3D-printed total knee arthroplasty (TKA) achieved adequate fixation of the tibial and patellar components to the underlying bone. Patients and Methods. A total of 29 patients undergoing primary TKA consented to participate in this prospective cohort study. All patients received a highly porous
Aims. Loosening of the tibial component after total knee arthroplasty (TKA) is a common indication for revision. Increasing the strength of the initial tibial implant/cement interface is desirable. There is little information about the surgical techniques that lead to the highest strength. We investigated the effects of eight variables on the strength of the initial
Aims. The material and design of knee components can have a considerable effect on the contact characteristics of the tibial post. This study aimed to analyze the stress distribution on the tibial post when using different grades of polyethylene for the tibial inserts. In addition, the contact properties of fixed-bearing and mobile-bearing inserts were evaluated. Methods. Three different grades of polyethylene were compared in this study; conventional ultra high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (HXLPE), and vitamin E-stabilized polyethylene (VEPE). In addition,
Revision total knee arthroplasty (rTKA) is a technically challenging and costly procedure. It is well-documented that primary TKA (pTKA) have better survivorship than rTKA; however, we were unable to identify any studies explicitly investigating previous rTKA as a risk factor for failure following rTKA. The purpose of this study is to compare the outcomes following rTKA between patients undergoing index rTKA and those who had been previously revised. This retrospective, observational study reviewed patients who underwent unilateral, aseptic rTKA at an academic orthopaedic speciality hospital between June 2011 and April 2020 with > one-year of follow-up. Patients were dichotomized based on whether this was their first revision procedure or not. Patient demographics, surgical factors, postoperative outcomes, and re-revision rates were compared between the groups.Aims
Methods
A novel enhanced cement fixation (EF) tibial implant with deeper cement pockets and a more roughened bonding surface was released to market for an existing total knee arthroplasty (TKA) system.This randomized controlled trial assessed fixation of the both the EF (ATTUNE S+) and standard (Std; ATTUNE S) using radiostereometric analysis. Overall, 50 subjects were randomized (21 EF-TKA and 23 Std-TKA in the final analysis), and had follow-up visits at six weeks, and six, 12, and 24 months to assess migration of the tibial component. Low viscosity bone cement with tobramycin was used in a standardized fashion for all subjects. Patient-reported outcome measure data was captured at preoperative and all postoperative visits.Aims
Methods
Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear. A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay migration, PE wear, tibial component migration, and the anatomical knee axis were evaluated on weightbearing stereoradiographs. PE inlay wear was measured as the deepest penetration of the femoral component into the PE inlay.Aims
Methods
A functional anterior cruciate ligament (ACL) or posterior cruciate ligament (PCL) has been assumed to be required for patients undergoing unicompartmental knee arthroplasty (UKA). However, this assumption has not been thoroughly tested. Therefore, this study aimed to assess the biomechanical effects exerted by cruciate ligament-deficient knees with medial UKAs regarding different posterior tibial slopes. ACL- or PCL-deficient models with posterior tibial slopes of 1°, 3°, 5°, 7°, and 9° were developed and compared to intact models. The kinematics and contact stresses on the tibiofemoral joint were evaluated under gait cycle loading conditions.Aims
Methods
Total knee arthroplasty (TKA) is a common and safe orthopaedic procedure. Zimmer Biomet's NexGen is the second most popular brand of implant used in the UK. The primary cause of revision after the first year is aseptic loosening. We present our experience of using this implant, with significant concerns around its performance with regards early aseptic loosening of the tibial component. A retrospective, single-surgeon review was carried out of all of the NexGen Legacy Posterior Stabilized (LPS) TKAs performed in this institute. The specific model used for the index procedures was the NexGen Complete Knee System (Legacy Knee-Posterior Stabilized LPS-Flex Articular Surface, LPS-Flex Femoral Component Option, and Stemmed Nonaugmentable Tibial Component Option).Aims
Methods
The June 2023 Knee Roundup360 looks at: Cementless total knee arthroplasty is associated with early aseptic loosening in a large national database; Is cementless total knee arthroplasty safe in females aged over 75 years?; Could novel radiological findings help identify aseptic tibial loosening?; The Attune cementless versus LCS arthroplasty at introduction; Return to work following total knee arthroplasty and unicompartmental knee arthroplasty; Complications and downsides of the robotic total knee arthroplasty; Mid-flexion instability in kinematic alignment better with posterior-stabilized and medial-stabilized implants?; Patellar resurfacing does not improve outcomes in modern knees.
Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening. A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.Aims
Methods
The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs). A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.Aims
Methods
We examined 86 polyethylene inserts, retrieved from total and unicompartmental knee prostheses after an average of 39.5 months in situ, grading them from 0 to 3 for seven modes of polyethylene degradation. Severe wear, with delamination or deformation, was observed in 51% of the implants, and was associated with time in situ, lack of congruency, thin polyethylene, third-body wear debris, and heat-pressed polyethylene. Significant under-surface cold flow was identified in some areas of unsupported polyethylene, and was associated with delamination in the load-bearing areas of thin inserts above screw holes in the underlying metal tray. We recommend the use of thicker polyethylene inserts, particularly in young, active patients and in designs with screw holes in the
Leg length discrepancy (LLD) is a common pre- and postoperative issue in total hip arthroplasty (THA) patients. The conventional technique for measuring LLD has historically been on a non-weightbearing anteroposterior pelvic radiograph; however, this does not capture many potential sources of LLD. The aim of this study was to determine if long-limb EOS radiology can provide a more reproducible and holistic measurement of LLD. In all, 93 patients who underwent a THA received a standardized preoperative EOS scan, anteroposterior (AP) radiograph, and clinical LLD assessment. Overall, 13 measurements were taken along both anatomical and functional axes and measured twice by an orthopaedic fellow and surgical planning engineer to calculate intraoperator reproducibility and correlations between measurements.Aims
Methods
This study compared patient-reported outcomes of three total knee arthroplasty (TKA) designs from one manufacturer: one cruciate-retaining (CR) design, and two cruciate-sacrificing designs, anterior-stabilized (AS) and posterior-stabilized (PS). Patients scheduled for primary TKA were included in a single-centre, prospective, three-armed, blinded randomized trial (n = 216; 72 per group). After intraoperative confirmation of posterior cruciate ligament (PCL) integrity, patients were randomly allocated to receive a CR, AS, or PS design from the same TKA system. Insertion of an AS or PS design required PCL resection. The primary outcome was the mean score of all five subscales of the Knee injury and Osteoarthritis Outcome Score (KOOS) at two-year follow-up. Secondary outcomes included all KOOS subscales, Oxford Knee Score, EuroQol five-dimension health questionnaire, EuroQol visual analogue scale, range of motion (ROM), and willingness to undergo the operation again. Patient satisfaction was also assessed.Aims
Methods
The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution. Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6).Aims
Methods
The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs. A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses.Aims
Methods
Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.Aims
Methods
The aim of this study was to compare a bicruciate-retaining (BCR) total knee arthroplasty (TKA) with a posterior cruciate-retaining (CR) TKA design in terms of kinematics, measured using fluoroscopy and stability as micromotion using radiostereometric analysis (RSA). A total of 40 patients with end-stage osteoarthritis were included in this randomized controlled trial. All patients performed a step-up and lunge task in front of a monoplane fluoroscope one year postoperatively. Femorotibial contact point (CP) locations were determined at every flexion angle and compared between the groups. RSA images were taken at baseline, six weeks, three, six, 12, and 24 months postoperatively. Clinical and functional outcomes were compared postoperatively for two years.Aims
Methods
The mid-term results of kinematic alignment (KA) for total knee arthroplasty (TKA) using image derived instrumentation (IDI) have not been reported in detail, and questions remain regarding ligamentous stability and revisions. This paper aims to address the following: 1) what is the distribution of alignment of KA TKAs using IDI; 2) is a TKA alignment category associated with increased risk of failure or poor patient outcomes; 3) does extending limb alignment lead to changes in soft-tissue laxity; and 4) what is the five-year survivorship and outcomes of KA TKA using IDI? A prospective, multicentre, trial enrolled 100 patients undergoing KA TKA using IDI, with follow-up to five years. Alignment measures were conducted pre- and postoperatively to assess constitutional alignment and final implant position. Patient-reported outcome measures (PROMs) of pain and function were also included. The Australian Orthopaedic Association National Joint Arthroplasty Registry was used to assess survivorship.Aims
Methods
The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS). A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively.Aims
Methods
It has previously been shown that higher-volume hospitals have better outcomes following revision total knee arthroplasty (rTKA). We were unable to identify any studies which investigated the effect of surgeon volume on the outcome of rTKA. We sought to investigate whether patients of high-volume (HV) rTKA surgeons have better outcomes following this procedure compared with those of low-volume (LV) surgeons. This retrospective study involved patients who underwent aseptic unilateral rTKA between January 2016 and March 2019, using the database of a large urban academic medical centre. Surgeons who performed ≥ 19 aseptic rTKAs per year during the study period were considered HV and those who performed < 19 per year were considered LV. Demographic characteristics, surgical factors, and postoperative outcomes were compared between the two groups.Aims
Methods
Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.Aims
Methods
To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period. This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.Aims
Methods
Thresholds of acceptable early migration of the components in total knee arthroplasty (TKA) have traditionally ignored the effects of patient and implant factors that may influence migration. The aim of this study was to determine which of these factors are associated with overall longitudinal migration of well-fixed tibial components following TKA. Radiostereometric analysis (RSA) data over a two-year period were available for 419 successful primary TKAs (267 cemented and 152 uncemented in 257 female and 162 male patients). Longitudinal analysis of data using marginal models was performed to examine the associations of patient factors (age, sex, BMI, smoking status) and implant factors (cemented or uncemented, the size of the implant) with maximum total point motion (MTPM) migration. Analyses were also performed on subgroups based on sex and fixation.Aims
Methods
Modern total knee arthroplasty (TKA) prostheses are designed to restore near normal kinematics including high flexion. Kneeling is a high flexion, kinematically demanding activity after TKA. The debate about design choice has not yet been informed by six-degrees-of-freedom in vivo kinematics. This prospective randomized clinical trial compared kneeling kinematics in three TKA designs. In total, 68 patients were randomized to either a posterior stabilized (PS-FB), cruciate-retaining (CR-FB), or rotating platform (CR-RP) design. Of these patients, 64 completed a minimum one year follow-up. Patients completed full-flexion kneeling while being imaged using single-plane fluoroscopy. Kinematics were calculated by registering the 3D implant models onto 2D-dynamic fluoroscopic images and exported for analysis.Aims
Methods
Simultaneous bilateral total knee arthroplasty (TKA) has been used due to its financial advantages, overall resource usage, and convenience for the patient. The training model where a trainee performs the first TKA, followed by the trainer surgeon performing the second TKA, is a unique model to our institution. This study aims to analyze the functional and clinical outcomes of bilateral simultaneous TKA when performed by a trainee or a supervising surgeon, and also to assess these outcomes based on which side was done by the trainee or by the surgeon. This was a retrospective cohort study of all simultaneous bilateral TKAs performed by a single surgeon in an academic institution between May 2003 and November 2017. Exclusion criteria were the use of partial knee arthroplasty procedures, staged bilateral procedures, and procedures not performed by the senior author on one side and the trainee on another. Primary clinical outcomes of interest included revision and re-revision. Primary functional outcomes included the Oxford Knee Score (OKS) and patient satisfaction scores.Aims
Methods
This study aimed to investigate the relationship between changes in patellar height and clinical outcomes at a mean follow-up of 7.7 years (5 to 10) after fixed-bearing posterior-stabilized total knee arthroplasty (PS-TKA). We retrospectively evaluated knee radiographs of 165 knees, which underwent fixed-bearing PS-TKA with patella resurfacing. The incidence of patella baja and changes in patellar height over a minimum of five years of follow-up were determined using Insall-Salvati ratio (ISR) measurement. We examined whether patella baja (ISR < 0.8) at final follow-up affected clinical outcomes, knee joint range of motion (ROM), and Knee Society Score (KSS). We also assessed inter- and intrarater reliability of ISR measurements and focused on the relationship between patellar height reduction beyond measurement error and clinical outcomes.Aims
Methods
Tibial cones are often utilized in revision total knee arthroplasty (TKA) with metaphyseal defects. Because there are few studies evaluating mid-term outcomes with a sufficient cohort, the purpose of this study was to evaluate tibial cone survival and complications in revision TKAs with tibial cones at minimum follow-up of five years. A retrospective review was completed from September 2006 to March 2015, evaluating 67 revision TKAs (64 patients) that received one specific porous tibial cone during revision TKA. The final cohort was composed of 62 knees (59 patients) with five years of clinical follow-up or reoperation. The mean clinical follow-up of the TKAs with minimum five-year clinical follow-up was 7.6 years (5.0 to 13.3). Survivorship analysis was performed with the endpoints of tibial cone revision for aseptic loosening, tibial cone revision for any reason, and reoperation. We also evaluated periprosthetic joint infection (PJI), risk factors for failure, and performed a radiological review.Aims
Methods
Unicompartmental knee arthroplasty (UKA) has become a popular method of treating knee localized osteoarthritis (OA). Additionally, the posterior cruciate ligament (PCL) is essential to maintaining the physiological kinematics and functions of the knee joint. Considering these factors, the purpose of this study was to investigate the biomechanical effects on PCL-deficient knees in medial UKA. Computational simulations of five subject-specific models were performed for intact and PCL-deficient UKA with tibial slopes. Anteroposterior (AP) kinematics and contact stresses of the patellofemoral (PF) joint and the articular cartilage were evaluated under the deep-knee-bend condition.Aims
Methods
Knee arthroplasty surgery is a highly effective treatment for arthritis and disorders of the knee. There are a wide variety of implant brands and types of knee arthroplasty available to surgeons. As a result of a number of highly publicized failures, arthroplasty surgery is highly regulated in the UK and many other countries through national registries, introduced to monitor implant performance, surgeons, and hospitals. With time, the options available within many brand portfolios have grown, with alternative tibial or femoral components, tibial insert materials, or shapes and patella resurfacings. In this study we have investigated the effect of the expansion of implant brand portfolios and where there may be a lack of transparency around a brand name. We also aimed to establish the potential numbers of compatible implant construct combinations. Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated.Aims
Methods
Bi-unicondylar arthroplasty (Bi-UKA) is a bone and anterior cruciate ligament (ACL)-preserving alternative to total knee arthroplasty (TKA) when the patellofemoral joint is preserved. The aim of this study is to investigate the clinical outcomes and biomechanics of Bi-UKA. Bi-UKA subjects (n = 22) were measured on an instrumented treadmill, using standard gait metrics, at top walking speeds. Age-, sex-, and BMI-matched healthy (n = 24) and primary TKA (n = 22) subjects formed control groups. TKA subjects with preoperative patellofemoral or tricompartmental arthritis or ACL dysfunction were excluded. The Oxford Knee Score (OKS) and EuroQol five-dimension questionnaire (EQ-5D) were compared. Bi-UKA, then TKA, were performed on eight fresh frozen cadaveric knees, to investigate knee extensor efficiency under controlled laboratory conditions, using a repeated measures study design.Aims
Methods
Medial pivot (MP) total knee arthroplasties (TKAs) were designed to mimic native knee kinematics with their deep medial congruent fitting of the tibia to the femur almost like a ball-on-socket, and a flat lateral part. GMK Sphere is a novel MP implant. Our primary aim was to study the migration pattern of the tibial tray of this TKA. A total of 31 patients were recruited to this single-group radiostereometric analysis (RSA) study and received a medial pivot GMK Sphere TKA. The distributions of male patients versus female patients and right versus left knees were 21:10 and 17:14, respectively. Mean BMI was 29 kg/m2 (95% confidence interval (CI) 27 to 30) and mean age at surgery was 63 years (95% CI 61 to 66). Maximum total point motions (MTPMs), medial, proximal, and anterior translations and transversal, internal, and varus rotations were calculated at three, 12, and 24 months. Patient-reported outcome measure data were also retrieved.Aims
Methods
Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design. Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.Aims
Methods
Metaphyseal tritanium cones can be used to manage the tibial bone loss commonly encountered at revision total knee arthroplasty (rTKA). Tibial stems provide additional fixation and are generally used in combination with cones. The aim of this study was to examine the role of the stems in the overall stability of tibial implants when metaphyseal cones are used for rTKA. This computational study investigates whether stems are required to augment metaphyseal cones at rTKA. Three cemented stem scenarios (no stem, 50 mm stem, and 100 mm stem) were investigated with 10 mm-deep uncontained posterior and medial tibial defects using four loading scenarios designed to mimic activities of daily living.Aims
Methods
Cementless total knee arthroplasty (TKA) offers the potential for strong biological fixation compared with cemented TKA where fixation is achieved by the mechanical integration of the cement. Few mid-term results are available for newer cementless TKA designs, which have used additive manufacturing (3D printing). The aim of this study was to present mid-term clinical outcomes and implant survivorship of the cementless Stryker Triathlon Tritanium TKA. This was a single institution registry review of prospectively gathered data from 341 cementless Triathlon Tritanium TKAs at four to 6.8 years follow-up. Outcomes were determined by comparing pre- and postoperative Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) scores, and pre- and postoperative 12-item Veterans RAND/Short Form Health Survey (VR/SF-12) scores. Aseptic loosening and revision for any reason were the endpoints which were used to determine survivorship at five years.Aims
Methods
This study aimed to evaluate the association between the sagittal alignment of the femoral component in total knee arthroplasty (TKA) and new Knee Society Score (2011KSS), under the hypothesis that outliers such as the excessive extended or flexed femoral component were related to worse clinical outcomes. A group of 156 knees (134 F:22 M) in 133 patients with a mean age 75.8 years (SD 6.4) who underwent TKA with the cruciate-substituting Bi-Surface Knee prosthesis were retrospectively enrolled. On lateral radiographs, γ angle (the angle between the distal femoral axis and the line perpendicular to the distal rear surface of the femoral component) was measured, and the patients were divided into four groups according to the γ angle. The 2011KSSs among groups were compared using the Kruskal-Wallis test. A secondary regression analysis was used to investigate the association between the 2011KSS and γ angle.Aims
Methods
In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up. A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis.Aims
Methods
Porous metaphyseal cones can be used for fixation in revision total knee arthroplasty (rTKA) and complex TKAs. This metaphyseal fixation has led to some surgeons using shorter cemented stems instead of diaphyseal engaging cementless stems with a potential benefit of ease of obtaining proper alignment without being beholden to the diaphysis. The purpose of this study was to evaluate short term clinical and radiographic outcomes of a series of TKA cases performed using 3D-printed metaphyseal cones. A retrospective review of 86 rTKAs and nine complex primary TKAs, with an average age of 63.2 years (SD 8.2) and BMI of 34.0 kg/m2 (SD 8.7), in which metaphyseal cones were used for both femoral and tibial fixation were compared for their knee alignment based on the type of stem used. Overall, 22 knees had cementless stems on both sides, 52 had cemented stems on both sides, and 15 had mixed stems. Postoperative long-standing radiographs were evaluated for coronal and sagittal plane alignment. Adjusted logistic regression models were run to assess malalignment hip-knee-ankle (HKA) alignment beyond ± 3° and sagittal alignment of the tibial and femoral components ± 3° by stem type.Aims
Methods
The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.Aims
Methods
Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the migration of cementless, 3D-printed TKA to cemented TKA of a similar design up to two years of follow-up using radiostereometric analysis (RSA) known for its ability to predict aseptic loosening. A total of 72 patients were randomized to either cementless 3D-printed or a cemented cruciate retaining TKA. RSA and clinical scores were evaluated at baseline and postoperatively at three, 12, and 24 months. A mixed model was used to analyze the repeated measurements.Aims
Methods
Mobile-bearing unicompartmental knee arthroplasty (UKA) with a flat tibial plateau has not performed well in the lateral compartment, leading to a high rate of dislocation. For this reason, the Domed Lateral UKA with a biconcave bearing was developed. However, medial and lateral tibial plateaus have asymmetric anatomical geometries, with a slightly dished medial and a convex lateral plateau. Therefore, the aim of this study was to evaluate the extent at which the normal knee kinematics were restored with different tibial insert designs using computational simulation. We developed three different tibial inserts having flat, conforming, and anatomy-mimetic superior surfaces, whereas the inferior surface in all was designed to be concave to prevent dislocation. Kinematics from four male subjects and one female subject were compared under deep knee bend activity.Aims
Methods
A retrospective study was conducted to measure short-term Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear.Aims
Patients and Methods
Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty. This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice. Cite this article:
Commonly performed unicompartmental knee arthroplasty (UKA) is not designed for the lateral compartment. Additionally, the anatomical medial and lateral tibial plateaus have asymmetrical geometries, with a slightly dished medial plateau and a convex lateral plateau. Therefore, this study aims to investigate the native knee kinematics with respect to the tibial insert design corresponding to the lateral femoral component. Subject-specific finite element models were developed with tibiofemoral (TF) and patellofemoral joints for one female and four male subjects. Three different TF conformity designs were applied. Flat, convex, and conforming tibial insert designs were applied to the identical femoral component. A deep knee bend was considered as the loading condition, and the kinematic preservation in the native knee was investigated.Aims
Methods
The medially spherical GMK Sphere (Medacta International AG, Castel San Pietro, Switzerland) total knee arthroplasty (TKA) was previously shown to accommodate lateral rollback while pivoting around a stable medial compartment, aiming to replicate native knee kinematics in which some coronal laxity, especially laterally, is also present. We assess coronal plane kinematics of the GMK Sphere and explore the occurrence and pattern of articular separation during static and dynamic activities. Using pulsed fluoroscopy and image matching, the coronal kinematics and articular surface separation of 16 well-functioning TKAs were studied during weight-bearing and non-weight-bearing, static, and dynamic activities. The closest distances between the modelled articular surfaces were examined with respect to knee position, and proportions of joint poses exhibiting separation were computed.Objectives
Methods
Aims
Patients and Methods
Preservation of both anterior and posterior cruciate ligaments in total knee arthroplasty (TKA) can lead to near-normal post-operative joint mechanics and improved knee function. We hypothesised that a patient-specific bicruciate-retaining prosthesis preserves near-normal kinematics better than standard off-the-shelf posterior cruciate-retaining and bicruciate-retaining prostheses in TKA. We developed the validated models to evaluate the post-operative kinematics in patient-specific bicruciate-retaining, standard off-the-shelf bicruciate-retaining and posterior cruciate-retaining TKA under gait and deep knee bend loading conditions using numerical simulation.Objectives
Methods
The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.Aims
Materials and Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs. The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (Aims
Patients and Methods
Throughout the 20th Century, it has been postulated that the knee moves on the basis of a four-bar link mechanism composed of the cruciate ligaments, the femur and the tibia. As a consequence, the femur has been thought to roll back with flexion, and total knee arthroplasty (TKA) prostheses have been designed on this basis. Recent work, however, has proposed that at a position of between 0° and 120° the medial femoral condyle does not move anteroposteriorly whereas the lateral femoral condyle tends, but is not obliged, to roll back – a combination of movements which equates to tibial internal/ femoral external rotation with flexion. The aim of this paper was to assess if the articular geometry of the GMK Sphere TKA could recreate the natural knee movements The pattern of knee movement was studied in 15 patients (six male: nine female; one male with bilateral TKAs) with 16 GMK Sphere implants, at a mean age of 66 years (53 to 76) with a mean BMI of 30 kg/m2 (20 to 35). The motions of all 16 knees were observed using pulsed fluoroscopy during a number of weight-bearing and non-weight-bearing static and dynamic activities.Objectives
Methods
Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method
This study compares the PFC total knee arthroplasty (TKA) system in a prospective randomized control trial (RCT) of the mobile-bearing rotating-platform (RP) TKA against the fixed-bearing (FB) TKA. This is the largest RCT with the longest follow-up where cruciate-retaining PFC total knee arthroplasties are compared in a non-bilateral TKA study. A total of 167 patients (190 knees with 23 bilateral cases), were recruited prospectively and randomly assigned, with 91 knees receiving the RP and 99 knees receiving FB. The mean age was 65.5 years (48 to 82), the mean body mass index (BMI) was 29.7 kg/m2 (20 to 52) and 73 patients were female. The Knee Society Score (KSS), Knee Society Functional Score (KSFS), Oxford Knee Score (OKS), Western Ontario and McMaster Universities Arthritis Index (WOMAC), and 12-Item Short-Form Health Survey Physical and Mental Component Scores (SF-12 PCS, SF-12 MCS) were gathered and recorded preoperatively, at five-years’ follow-up, and at ten years’ follow-up. Additionally, Knee Injury and Osteoarthritis Outcome Scores (KOOS) were collected at five- and ten-year follow-ups. The prevalence of radiolucent lines (RL) on radiographs and implant survival were recorded at five- and ten-year follow-ups.Aims
Patients and Methods
Total ankle arthroplasty (TAA) surgery is complex and attracts a wide variety of complications. The literature lacks consistency in reporting adverse events and complications. The aim of this article is to provide a comprehensive analysis of each of these complications from a literature review, and to compare them with rates from our Unit, to aid clinicians with the process of informed consent. A total of 278 consecutive total ankle arthroplasties (251 patients), performed by four surgeons over a six-year period in Wrightington Hospital (Wigan, United Kingdom) were prospectively reviewed. There were 143 men and 108 women with a mean age of 64 years (41 to 86). The data were recorded on each follow-up visit. Any complications either during initial hospital stay or subsequently reported on follow-ups were recorded, investigated, monitored, and treated as warranted. Literature search included the studies reporting the outcomes and complications of TAA implants.Aims
Patients and Methods
Because there have been no standard methods to determine pre-operatively
the thickness of resection of the proximal tibia in unicompartmental
knee arthroplasty (UKA), information about the relationship between
the change of limb alignment and the joint line elevation would
be useful for pre-operative planning. The purpose of this study
was to clarify the correlation between the change of limb alignment
and the change of joint line height at the medial compartment after
UKA. A consecutive series of 42 medial UKAs was reviewed retrospectively.
These patients were assessed radiographically both pre- and post-operatively
with standing anteroposterior radiographs. The thickness of bone
resection at the proximal tibia and the distal femur was measured
radiographically. The relationship between the change of femorotibial
angle (δFTA) and the change of joint line height, was analysed.Objectives
Methods
Posterior condylar offset (PCO) and posterior tibial slope (PTS) are critical factors in total knee arthroplasty (TKA). A computational simulation was performed to evaluate the biomechanical effect of PCO and PTS on cruciate retaining TKA. We generated a subject-specific computational model followed by the development of ± 1 mm, ± 2 mm and ± 3 mm PCO models in the posterior direction, and -3°, 0°, 3° and 6° PTS models with each of the PCO models. Using a validated finite element (FE) model, we investigated the influence of the changes in PCO and PTS on the contact stress in the patellar button and the forces on the posterior cruciate ligament (PCL), patellar tendon and quadriceps muscles under the deep knee-bend loading conditions.Objectives
Methods
The aims of this study were to determine the proportion of patients
with outlier varus or valgus alignment in kinematically aligned
total knee arthroplasty (TKA), whether those with outlier varus
or valgus alignment have higher forces in the medial or lateral
compartments of the knee than those with in-range alignment and
whether measurements of the alignment of the limb, knee and components
predict compartment forces. The intra-operative forces in the medial and lateral compartments
were measured with an instrumented tibial insert in 67 patients
who underwent a kinematically aligned TKA during passive movement.
The mean of the forces at full extension, 45° and 90° of flexion
determined the force in the medial and lateral compartments. Measurements
of the alignment of the limb and the components included the hip-knee-ankle
(HKA) angle, proximal medial tibial angle (PMTA), and distal lateral
femoral angle (DLFA). Measurements of the alignment of the knee
and the components included the tibiofemoral angle (TFA), tibial
component angle (TCA) and femoral component angle (FCA). Alignment was
measured on post-operative, non-weight-bearing anteroposterior (AP)
scanograms and categorised as varus or valgus outlier or in-range
in relation to mechanically aligned criteria.Aims
Patients and Methods
Advances in polyethylene (PE) in total hip arthroplasty
have led to interest and increased use of highly crosslinked PE
(HXLPE) in total knee arthroplasty (TKA). Biomechanical data suggest
improved wear characteristics for HXLPE inserts over conventional
PE in TKA. Short-term results from registry data and few clinical
trials are promising. Our aim is to present a review of the history
of HXLPEs, the use of HXLPE inserts in TKA, concerns regarding potential mechanical
complications, and a thorough review of the available biomechanical
and clinical data. Cite this article:
Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.Objectives
Methods
This randomised trial evaluated the outcome of
a single design of unicompartmental arthroplasty of the knee (UKA) with
either a cemented all-polyethylene or a metal-backed modular tibial
component. A total of 63 knees in 45 patients (17 male, 28 female)
were included, 27 in the all-polyethylene group and 36 in the metal-backed
group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up
of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were
revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and
two metal-backed components were revised (at one and five years).
One revision in both groups was for unexplained pain, one in the
metal-backed group was for progression of osteoarthritis. The others
in the all-polyethylene group were for aseptic loosening. The survivorship
at seven years calculated by the Kaplan–Meier method for the all-polyethylene
group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for
the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at
risk 16) This difference was statistically significant (p <
0.001).
At the most recent follow-up, significantly better mean Western
Ontario and McMaster Universities Arthritis Index Scores were found
in the all-polyethylene group (13.4 This randomised study demonstrates that all-polyethylene components
in this design of fixed bearing UKA had unsatisfactory results with
significantly higher rates of failure before ten years compared
with the metal-back components. Cite this article:
This study reports on the first 150 consecutive
Oxford cementless unicompartmental knee arthroplasties (UKA) performed
in an independent centre (126 patients). All eligible patients had
functional scores (Oxford knee score and high activity arthroplasty
score) recorded pre-operatively and at two- and five-years of follow-up. Fluoroscopically
aligned radiographs were taken at five years and analysed for any
evidence of radiolucent lines (RLLs), subsidence or loosening. The
mean age of the cohort was 63.6 years (39 to 86) with 81 (53.1%)
males. Excellent functional scores were maintained at five years
and there were no progressive RLLs demonstrated on radiographs.
Two patients underwent revision to a total knee arthroplasty giving
a revision rate of 0.23/100 (95% confidence interval 0.03 to 0.84)
component years with overall component survivorship of 98.7% at
five years. There were a further four patients who underwent further
surgery on the same knee, two underwent bearing exchanges for dislocation
and two underwent lateral UKAs for disease progression. This was
a marked improvement from other UKAs reported in New Zealand Joint
Registry data and supports the designing centre’s early results. Cite this article:
The aim of this prospective multicentre study
was to report the patient satisfaction after total knee replacement (TKR),
undertaken with the aid of intra-operative sensors, and to compare
these results with previous studies. A total of 135 patients undergoing
TKR were included in the study. The soft-tissue balance of each
TKR was quantified intra-operatively by the sensor, and 18 (13%)
were found to be unbalanced. A total of 113 patients (96.7%) in
the balanced group and 15 (82.1%) in the unbalanced group were satisfied
or very satisfied one year post-operatively (p = 0.043). A review of the literature identified no previous study with
a mean level of satisfaction that was greater than the reported
level of satisfaction of the balanced TKR group in this study. Ensuring
soft-tissue balance by using intra-operative sensors during TKR
may improve satisfaction. Cite this article:
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.
We studied the bone mineral density (BMD) and
the bone mineral content (BMC) of the proximal tibia in patients with
a well-functioning uncemented Oxford medial compartment arthroplasty
using the Lunar iDXA bone densitometer. Our hypothesis was that
there would be decreased BMD and BMC adjacent to the tibial base
plate and increased BMD and BMC at the tip of the keel. There were 79 consecutive patients (33 men, 46 women) with a
mean age of 65 years (44 to 84) with a minimum two-year follow-up
(mean 2.6 years (2.0 to 5.0)) after unilateral arthroplasty, who
were scanned using a validated standard protocol where seven regions
of interest (ROI) were examined and compared with the contralateral
normal knee. All had well-functioning knees with a mean Oxford knee
score of 43 (14 to 48) and mean Knee Society function score of 90
(20 to 100), showing a correlation with the increasing scores and
higher BMC and BMD values in ROI 2 in the non-implanted knee relative
to the implanted knee (p = 0.013 and p = 0.015, respectively). The absolute and percentage changes in BMD and BMC were decreased
in all ROIs in the implanted knee compared with the non-implanted
knee, but this did not reach statistical significance. Bone loss
was markedly less than reported losses with total knee replacement. There was no significant association with side, although there
was a tendency for the BMC to decrease with age in men. The BMC
was less in the implanted side relative to the non-implanted side
in men compared with women in ROI 2 (p = 0.027), ROI 3 (p = 0.049)
and ROI 4 (p = 0.029). The uncemented Oxford medial compartment arthroplasty appears
to allow relative preservation of the BMC and BMD of the proximal
tibia, suggesting that the implant acts more physiologically than
total knee replacement. Peri-prosthetic bone loss is an important
factor in assessing long-term implant stability and survival, and
the results of this study are encouraging for the long-term outcome
of this arthroplasty. Cite this article:
We conducted a randomised prospective study to evaluate the clinical and radiological results of a mobile- and fixed-bearing total knee replacement of similar design in 174 patients who had bilateral simultaneous knee replacement. The mean follow-up was for 5.6 years (5.2 to 6.1). The total knee score, pain score, functional score and range of movement were not statistically different (p >
0.05) between the two groups. Osteolysis was not seen in any knee in either group. Two knees (1%) in the mobile-bearing group required revision because of infection; none in the fixed-bearing group needed revision. Excellent results can be achieved with both mobile- and fixed-bearing prostheses of similar design at mid-term follow-up. We could demonstrate no significant clinical advantage for a mobile bearing.
We performed a prospective, randomised trial of 44 patients to compare the functional outcomes of a posterior-cruciate-ligament-retaining and posterior-cruciate-ligament-substituting total knee arthroplasty, and to gain a better understanding of the At follow-up at five years, no statistically significant differences were found in the clinical outcome measurements for either design. The prevalence of radiolucent lines and the survivorship were the same. In a subgroup of 15 knees, additional image-intensifier analysis in the horizontal and sagittal planes was performed during step-up and lunge activity. Our analysis revealed striking differences. Lunge activity showed a mean posterior displacement of both medial and lateral tibiofemoral contact areas (roll-back) which was greater and more consistent in the cruciate-substituting than in the cruciate-retaining group (medial p <
0.0001, lateral p = 0.011). The amount of posterior displacement could predict the maximum flexion which could be achieved (p = 0.018). Forward displacement of the tibiofemoral contact area in flexion during stair activity was seen more in the cruciate-retaining than in the cruciate-substituting group. This was attributed mainly to insufficiency of the posterior cruciate ligament and partially to that of the anterior cruciate ligament. We concluded that, despite similar clinical outcomes, there are significant kinematic differences between cruciate-retaining and cruciate-substituting arthroplasties.
We describe a cohort of patients with a high rate of mid-term failure following Kinemax Plus total knee replacement inserted between 1998 and 2001. This implant has been recorded as having a survival rate of 96% at ten years. However, in our series the survival rate was 75% at nine years. This was also significantly lower than that of subsequent consecutive series of PFC Sigma knee replacements performed by the same surgeon. No differences were found in the clinical and radiological parameters between the two groups. At revision the most striking finding was polyethylene wear. An independent analysis of the polyethylene components was therefore undertaken. Scanning electron microscopy revealed type 2 fusion defects in the ultra-high molecular weight polyethylene (UHMWPE), which indicated incomplete boundary fusion. Other abnormalities consistent with weak UHMWPE particle interface strength were present in both the explanted inserts and in unused inserts from the same period. We consider that these type 2 fusion defects are the cause of the early failure of the Kinemax implants. This may represent a manufacturing defect resulting in a form of programmed polyethylene failure.
The management of bone loss in revision replacement of the knee remains a challenge despite an array of options available to the surgeon. Bone loss may occur as a result of the original disease, the design of the prosthesis, the mechanism of failure or technical error at initial surgery. The aim of revision surgery is to relieve pain and improve function while addressing the mechanism of failure in order to reconstruct a stable platform with transfer of load to the host bone. Methods of reconstruction include the use of cement, modular metal augmentation of prostheses, custom-made, tumour-type or hinged implants and bone grafting. The published results of the surgical techniques are summarised and a guide for the management of bone defects in revision surgery of the knee is presented.
Our aims were to map the tibial footprint of the posterior cruciate ligament (PCL) using MRI in patients undergoing PCL-preserving total knee replacement, and to document the disruption of this footprint as a result of the tibial cut. In 26 consecutive patients plain radiography and MRI of the knee were performed pre-operatively, and plain radiography post-operatively. The lower margin of the PCL footprint was located a mean of 1 mm (−10 to 8) above the upper aspect of the fibular head. The mean surface area was 83 mm2 (49 to 142). One-third of patients (8 of 22) had tibial cuts made below the lowest aspect of the PCL footprint (complete removal) and one-third (9 of 22) had cuts extending into the footprint (partial removal). The remaining patients (5 of 22) had footprints unaffected by the cuts, keeping them intact. Our study highlights the wide variation in the location of the tibial PCL footprint when referenced against the fibula. Proximal tibial cuts using conventional jigs resulted in the removal of a significant portion, if not all of the PCL footprint in most of the patients in our study. Our findings suggest that when performing PCL-retaining total knee replacement the tibial attachment of the PCL is often removed.
Mobile-bearing posterior-stabilised knee replacements have been developed as an alternative to the standard fixed- and mobile-bearing designs. However, little is known about the We conclude that mobile-bearing posterior-stabilised knee replacements reproduce internal rotation of the tibia more closely during flexion than fixed-bearing posterior-stabilised designs. Furthermore, mobile-bearing posterior-stabilised knee replacements demonstrate a unidirectional movement which occurs at the upper and lower sides of the mobile insert. The femur moves in an anteroposterior direction on the upper surface of the insert, whereas the movement at the lower surface is pure rotation. Such unidirectional movement may lead to less wear when compared with the multidirectional movement seen in fixed-bearing posterior-stabilised knee replacements, and should be associated with more evenly applied cam-post stresses.
We compared the results of 146 patients who received an anatomic modular knee fixed-bearing total knee replacement (TKR) in one knee and a low contact stress rotating platform mobile-bearing TKR in the other. There were 138 women and eight men with a mean age of 69.8 years (42 to 80). The mean follow-up was 13.2 years (11.0 to 14.5). The patients were assessed clinically and radiologically using the rating systems of the Hospital for Special Surgery and the Knee Society at three months, six months, one year, and annually thereafter. The assessment scores of both rating systems pre-operatively and at the final review did not show any statistically significant differences between the two designs of implant. In the anatomic modular knee group, one knee was revised because of aseptic loosening of the tibial component and one because of infection. In addition, three knees were revised because of wear of the polyethylene tibial bearing. In the low contact stress group, two knees were revised because of instability requiring exchange of the polyethylene insert and one because of infection. The radiological analysis found no statistical difference in the incidence of radiolucent lines at the final review (Student’s We found no evidence of the superiority of one design over the other at long-term follow-up.