The biomechanics of the patellofemoral joint can become disturbed during total knee replacement by alterations induced by the position and shape of the different prosthetic components. The role of the patella and femoral trochlea has been well studied. We have examined the effect of anterior or posterior positioning of the tibial component on the mechanisms of patellofemoral contact in total knee replacement. The hypothesis was that placing the tibial component more posteriorly would reduce patellofemoral contact stress while providing a more efficient lever arm during extension of the knee. We studied five different positions of the tibial component using a six degrees of freedom dynamic knee simulator system based on the Oxford rig, while simulating an active knee squat under physiological loading conditions. The patellofemoral contact force decreased at a mean of 2.2% for every millimetre of posterior translation of the tibial component. Anterior positions of the tibial component were associated with elevation of the patellofemoral joint pressure, which was particularly marked in flexion >
90°. From our results we believe that more posterior positioning of the tibial component in total knee replacement would be beneficial to the patellofemoral joint.
The long-term functional outcome of total hip arthroplasty (THA)
performed by trainees is not known. A multicentre retrospective
study of 879 THAs was undertaken to investigate any differences
in outcome between those performed by trainee surgeons and consultants. A total of 879 patients with a mean age of 69.5 years (37 to
94) were included in the study; 584 THAs (66.4%) were undertaken
by consultants, 138 (15.7%) by junior trainees and 148 (16.8%) by
senior trainees. Patients were scored using the Harris Hip Score
(HHS) pre-operatively and at one, three, five, seven and ten years
post-operatively. Surgical outcome, complications and survival were
compared between groups. The effect of supervision was determined
by comparing supervised and unsupervised trainees. A primary univariate
analysis was used to select variables for inclusion in multivariate
analysis. Aims
Patients and Methods
The objective of this study was to determine if combining variations in mixing technique of antibiotic-impregnated polymethylmethacrylate (PMMA) cement with low frequency ultrasound (LFUS) improves antibiotic elution during the initial high phase (Phase I) and subsequent low phase (Phase II) while not diminishing mechanical strength. Three batches of vancomycin-loaded PMMA were prepared with different mixing techniques: a standard technique; a delayed technique; and a control without antibiotic. Daily elution samples were analysed using flow injection analysis (FIA). Beginning in Phase II, samples from each mix group were selected randomly to undergo either five, 15, 45, or 0 minutes of LFUS treatment. Elution amounts between LFUS treatments were analysed. Following Phase II, compression testing was done to quantify strength. Objectives
Methods
Of the 11 054 Charnley low-frictional torque arthroplasties carried out at our hospital between 1962 and 1977, 110 (94 patients) had a minimum follow-up of 30 years with a mean of 32.3 years (30.0 to 40.5). The mean age of the patients at operation was 43.3 years (17.0 to 65.0) and 75.7 years (51.0 to 97.0) at follow-up. Overall, 90% of hips (99) were free from pain and activity was reported as normal in 58% of the patients. A total of 13 hips (11.8%) were revised at a mean follow-up of 32.3 years (30.0 to 39.5), with wear and loosening of the acetabular component as the main indications. The clinical results did not reflect the mechanical state of the implant. Follow-up with sequential radiographs of good quality is essential. Revision for radiological changes alone must be accepted if gross loss of bone stock is to be avoided. Improvements in the design, materials and operative technique, based on the long-term outcome, are highlighted.
The long-term results of grafting with hydroxyapatite granules for acetabular deficiency in revision total hip replacement are not well known. We have evaluated the results of revision using a modular cup with hydroxyapatite grafting for Paprosky type 2 and 3 acetabular defects at a minimum of ten years’ follow-up. We retrospectively reviewed 49 acetabular revisions at a mean of 135 months (120 to 178). There was one type 2B, ten 2C, 28 3A and ten 3B hips. With loosening as the endpoint, the survival rate was 74.2% (95% confidence interval 58.3 to 90.1). Radiologically, four of the type 3A hips (14%) and six of the type 3B hips (60%) showed aseptic loosening with collapse of the hydroxyapatite layer, whereas no loosening occurred in type 2 hips. There was consolidation of the hydroxyapatite layer in 33 hips (66%). Loosening was detected in nine of 29 hips (31%) without cement and in one of 20 hips (5%) with cement (p = 0.03, Fisher’s exact probability test). The linear wear and annual wear rate did not correlate with loosening. These results suggest that the long-term results of hydroxyapatite grafting with cement for type 2 and 3A hips are encouraging.
We describe two cases of fracture of Corin Taper-Fit stems used for cement-in-cement revision of congenital dysplasia of the hip. Both prostheses were implanted in patients in their 50s, with high offsets (+7.5 mm and +3.5 mm), one with a large diameter (48 mm) head and one with a constrained acetabular component. Fracture of the stems took place at nine months and three years post-operatively following low-demand activity. Both fractures occurred at the most medial of the two stem introducer holes in the neck of the prosthesis, a design feature that is unique to the Taper-Fit stem. We would urge caution in the use of these particular stems for cement-in-cement revisions.
We investigated whether the indentation of bone
cement spacers used in revision of infected joint arthroplasty with a
MacDonald dissector increased the elution of antibiotic The fluid sampled at 72 hours from the indented discs containing
0.17 g gentamicin (0.88% w/w) contained a mean of 113 mcg/ml (90.12
to 143.5) compared with 44.5 mcg/ml (44.02 to 44.90) in the fluid
sampled from the plain discs (p = 0.012). In discs containing 0.33
g gentamicin (1.75% w/w), the concentration eluted from the indented discs
at 72 hours was a mean of 316 mcg/ml (223 to 421) compared with
a mean of 118 mcg/ml (100 to 140) from the plain discs (p <
0.001). At two weeks, these significant differences persisted. At nine
weeks the indented discs eluted a greater concentration for all
gentamicin doses, but the difference was only significant for the
discs containing 0.17 g (0.88% w/w, p = 0.006). However if the area
under the curve is taken as a measure of the total antibiotic eluted,
the indented discs eluted more gentamicin than the plain discs for
the 0.17 g (0.88% w/w, p = 0.031), the 0.25 g (1.41% w/w, p <
0.001) and the 0.33 g (1.75% w/w, p <
0.001) discs. When preparing antibiotic spacers for use in staged revision
arthroplasty surgery we recommend indenting the spacer with a MacDonald
dissector to increase the elution of antibiotic. Cite this article:
The number of arthroplasties being undertaken
is expected to grow year on year, and periprosthetic joint infections will
be an increasing socioeconomic burden. The challenge to prevent
and eradicate these infections has resulted in the emergence of
several new strategies, which are discussed in this review. Cite this article:
Several techniques are available for the treatment of displaced fractures of the neck of the radius in children. We report our experience in 14 children treated by indirect reduction and fixation using an elastic stable intramedullary nail. The patients were followed up for a mean of 28 months. One developed asymptomatic avascular necrosis. The rest had excellent results. One had a neuropraxia of the posterior interosseous nerve, which recovered within six weeks. We advocate elastic stable intramedullary nailing for the closed reduction and fixation of these fractures in children.
We have developed a novel method of calculating the radiological magnification of the hip using two separate radio-opaque markers. We recruited 74 patients undergoing radiological assessment following total hip replacement. Both the new double marker and a conventional single marker were used by the radiographer at the time of x-ray. The predicted magnification according to each marker was calculated, as was the true radiological magnification of the components. The correlation between true and predicted magnification was good using the double marker (r = 0.90, n = 74, p <
0.001), but only moderate for the single marker (r = 0.50, n = 63, p <
0.001). The median error was significantly less for the double marker than for the single (1.1% The double marker method appears to be superior to the single marker method when used in the clinical environment.
Fractures of the proximal interphalangeal joint include a wide spectrum of injuries, from stable avulsion fractures to complex fracture-dislocations. Stability of the joint is paramount in determining the appropriate treatment, which should aim to facilitate early mobilisation and restoration of function.
A total of 20 pairs of fresh-frozen cadaver femurs were assigned to four alignment groups consisting of relative varus (10° and 20°) and relative valgus (10° and 20°), 75 composite femurs of two neck geometries were also used. In both the cadaver and the composite femurs, placing the component in 20° of valgus resulted in a significant increase in load to failure. Placing the component in 10° of valgus had no appreciable effect on increasing the load to failure except in the composite femurs with varus native femoral necks. Specimens in 10° of varus were significantly weaker than the neutrally-aligned specimens. The results suggest that retention of the intact proximal femoral strength occurs at an implant angulation of ≥ 142°. However, the benefit of extreme valgus alignment may be outweighed in clinical practice by the risk of superior femoral neck notching, which was avoided in this study.
The aim of this study was to compare the maximum
laxity conferred by the cruciate-retaining (CR) and posterior-stabilised
(PS) Triathlon single-radius total knee arthroplasty (TKA) for anterior
drawer, varus–valgus opening and rotation in eight cadaver knees
through a defined arc of flexion (0º to 110º). The null hypothesis
was that the limits of laxity of CR- and PS-TKAs are not significantly
different. The investigation was undertaken in eight loaded cadaver knees
undergoing subjective stress testing using a measurement rig. Firstly
the native knee was tested prior to preparation for CR-TKA and subsequently
for PS-TKA implantation. Surgical navigation was used to track maximal
displacements/rotations at 0º, 30º, 60º, 90º and 110° of flexion.
Mixed-effects modelling was used to define the behaviour of the
TKAs. The laxity measured for the CR- and PS-TKAs revealed no statistically
significant differences over the studied flexion arc for the two
versions of TKA. Compared with the native knee both TKAs exhibited
slightly increased anterior drawer and decreased varus-valgus and
internal-external roational laxities. We believe further study is required
to define the clinical states for which the additional constraint
offered by a PS-TKA implant may be beneficial. Cite this article:
The treatment of bony defects of the tibia at the time of revision total knee replacement is controversial. The place of compacted morsellised bone graft is becoming established, particularly in contained defects. It has previously been shown that the initial stability of impaction-grafted trays in the contained defects is equivalent to that of an uncemented primary knee replacement. However, there is little biomechanical evidence on which to base a decision in the treatment of uncontained defects. We undertook a laboratory-based biomechanical study comparing three methods of graft containment in segmental medial tibial defects and compared them with the use of a modular metal augment to bypass the defect. Using resin models of the proximal tibia with medial defects representing either 46% or 65% of the medial cortical rim, repair of the defect was accomplished using mesh, cement or a novel bag technique, after which impaction bone grafting was used to fill the contained defects and a tibial component was cemented in place. As a control, a cemented tibial component with modular metal augments was used in identical defects. All specimens were submitted to cyclical mechanical loading, during which cyclical and permanent tray displacement were determined. The results showed satisfactory stability with all the techniques except the bone bag method. Using metal augments gave the highest initial stability, but obviously lacked any potential for bone restoration.
Resurfacing of the humeral head is commonly used
within the UK to treat osteoarthritis (OA) of the shoulder. We present
the results of a small prospective randomised study of this procedure
using the Global CAP prosthesis with two different coatings, Porocoat
and DuoFix hydroxyapatite (HA). We followed two groups of ten patients
with OA of the shoulder for two years after insertion of the prosthesis
with tantalum marker beads, recording pain, Constant–Murley and
American Shoulder and Elbow Surgeons (ASES) outcome scores, and
using radiostereometric analysis to assess migration. The outcomes
were similar to those of other series, with significant reductions
in pain (p = 0.003) and an improvement in the Constant (p = 0.001)
and ASES scores (p = 0.006). The mean migration of the prosthesis
three months post-operatively was 0.78 mm (0.51 to 1.69) and 0.72
mm (0.33 to 1.45) for the Porocoat and DuoFix groups, respectively.
Analysis of variance indicated that the rate of migration reached
a plateau after three months post-operatively in both groups. At
follow-up of two years the mean migration was 1 mm ( The addition of a coating of HA to the sintered surface does
not improve fixation of this prosthesis. Cite this article:
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.
Conventional cemented acetabular components are
reported to have a high rate of failure when implanted into previously
irradiated bone. We recommend the use of a cemented reconstruction
with the addition of an acetabular reinforcement cross to improve
fixation. We reviewed a cohort of 45 patients (49 hips) who had undergone
irradiation of the pelvis and a cemented total hip arthroplasty
(THA) with an acetabular reinforcement cross. All hips had received
a minimum dose of 30 Gray (Gy) to treat a primary nearby tumour
or metastasis. The median dose of radiation was 50 Gy (Q1 to Q3:
45 to 60; mean: 49.57, 32 to 72). The mean follow-up after THA was 51 months (17 to 137). The cumulative
probability of revision of the acetabular component for a mechanical
reason was 0% (0 to 0%) at 24 months, 2.9% (0.2 to 13.3%) at 60
months and 2.9% (0.2% to 13.3%) at 120 months, respectively. One
hip was revised for mechanical failure and three for infection. Cemented acetabular components with a reinforcement cross provide
good medium-term fixation after pelvic irradiation. These patients
are at a higher risk of developing infection of their THA. Cite this article:
We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up. Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain. Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.
The objective of this study was to compare the elution characteristics,
antimicrobial activity and mechanical properties of antibiotic-loaded
bone cement (ALBC) loaded with powdered antibiotic, powdered antibiotic
with inert filler (xylitol), or liquid antibiotic, particularly focusing
on vancomycin and amphotericin B. Cement specimens loaded with 2 g of vancomycin or amphotericin
B powder (powder group), 2 g of antibiotic powder and 2 g of xylitol
(xylitol group) or 12 ml of antibiotic solution containing 2 g of
antibiotic (liquid group) were tested.Objectives
Methods