Advertisement for orthosearch.org.uk
Results 141 - 160 of 381
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1011 - 1021
1 Aug 2013
Krishnan H Krishnan SP Blunn G Skinner JA Hart AJ

Following the recall of modular neck hip stems in July 2012, research into femoral modularity will intensify over the next few years. This review aims to provide surgeons with an up-to-date summary of the clinically relevant evidence. The development of femoral modularity, and a classification system, is described. The theoretical rationale for modularity is summarised and the clinical outcomes are explored. The review also examines the clinically relevant problems reported following the use of femoral stems with a modular neck.

Joint replacement registries in the United Kingdom and Australia have provided data on the failure rates of modular devices but cannot identify the mechanism of failure. This information is needed to determine whether modular neck femoral stems will be used in the future, and how we should monitor patients who already have them implanted.

Cite this article: Bone Joint J 2013;95-B:1011–21.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 158 - 162
1 Feb 2012
Sternheim A Backstein D Kuzyk PRT Goshua G Berkovich Y Safir O Gross AE

We report the use of porous metal acetabular revision shells in the treatment of contained bone loss. The outcomes of 53 patients with 50% acetabular bleeding host bone contact were compared with a control group of 49 patients with > 50% to 85% bleeding host bone contact. All patients were treated with the same type of trabecular metal acetabular revision shell. The mean age at revision was 62.4 years (42 to 80) and the mean follow-up of both groups was 72.4 months (60 to 102). Clinical, radiological and functional outcomes were assessed. There were four (7.5%) mechanical failures in the 50% host bone contact group and no failures in the > 50% host bone contact group (p = 0.068). Out of both groups combined there were four infections (3.9%) and five recurrent dislocations (4.9%) with a stable acetabular component construct that were revised to a constrained liner. Given the complexity of the reconstructive challenge, porous metal revision acetabular shells show acceptable failure rates at five to ten years’ follow-up in the setting of significant contained bone defects. This favourable outcome might be due to the improved initial stability achieved by a high coefficient of friction between the acetabular implant and the host bone, and the high porosity, which affords good bone ingrowth.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 70 - 73
1 Nov 2013
Lanting BA MacDonald SJ

Total hip replacement (THR) is a very common procedure undertaken in up to 285 000 Americans each year. Patient satisfaction with THR is very high, with improvements in general health, quality of life, and function while at the same time very cost effective. Although the majority of patients have a high degree of satisfaction with their THR, 27% experience some discomfort, and up to 6% experience severe chronic pain. Although it can be difficult to diagnose the cause of the pain in these patients, this clinical issue should be approached systematically and thoroughly. A detailed history and clinical examination can often provide the correct diagnosis and guide the appropriate selection of investigations, which will then serve to confirm the clinical diagnosis made.

Cite this article: Bone Joint J 2013;95-B, Supple A:70–3.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 80 - 85
1 Jan 2012
Malone AA Sanchez JS Adams R Morrey B

We report the effectiveness of revision of total elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows in 52 patients were treated with re-cementing of a total elbow replacement into part or all of the existing cement mantle or into the debrided host-bone interface, without the use of structural bone augmentation or a custom prosthesis. The original implant revision was still in situ and functional in 42 of 53 elbows (79%) at a mean of 94.5 months (26 to 266) after surgery. In 31 of these 42 elbows (74%) the Mayo Elbow Performance Score was good or excellent. Overall, of the 53 elbows, 18 (34%) required re-operation, ten (19%) for loosening. A classification system was developed to identify those not suitable for revision by this technique, and using this we have showed that successful re-implantation is statistically correlated to properly addressing the bone deficiency for both the humeral (p = 0.005) and the ulnar (p = 0.039) components.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 57 - 62
1 Nov 2013
Stulberg SD Patel RM

Conventional uncemented femoral implants provide dependable long-term fixation in patients with a wide range of functional requirements. Yet challenges associated with proximal–distal femoral dimensional mismatch, preservation of bone stock, and minimally invasive approaches have led to exploration into alternative implant designs. Short stem designs focusing on a stable metaphyseal fit have emerged to address these issues in total hip replacement (THR). Uncemented metaphyseal-engaging short stem implants are stable and are associated with proximal bone remodeling closer to the metaphysis when compared with conventional stems and they also have comparable clinical performances. Short stem metaphyseal-engaging implants can meet the goals of a successful THR, including tolerating a high level of patient function, as well as durable fixation.

Cite this article: Bone Joint J 2013;95-B, Supple A:57–62.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1681 - 1686
1 Dec 2013
Peach CA Nicoletti S Lawrence TM Stanley D

We report our experience of staged revision surgery for the treatment of infected total elbow arthroplasty (TEA). Between 1998 and 2010 a consecutive series of 33 patients (34 TEAs) underwent a first-stage procedure with the intention to proceed to second-stage procedure when the infection had been controlled. A single first-stage procedure with removal of the components and cement was undertaken for 29 TEAs (85%), followed by the insertion of antibiotic-impregnated cement beads, and five (15%) required two or more first-stage procedures. The most common organism isolated was coagulase-negative Staphylococcus in 21 TEAs (62%).

A second-stage procedure was performed for 26 TEAs (76%); seven patients (seven TEAs, 21%) had a functional resection arthroplasty with antibiotic beads in situ and had no further surgery, one had a persistent discharge preventing further surgery.

There were three recurrent infections (11.5%) in those patients who underwent a second-stage procedure. The infection presented at a mean of eight months (5 to 10) post-operatively. The mean Mayo Elbow Performance Score (MEPS) in those who underwent a second stage revision without recurrent infection was 81.1 (65 to 95).

Staged revision surgery is successful in the treatment of patients with an infected TEA and is associated with a low rate of recurrent infection. However, when infection does occur, this study would suggest that it becomes apparent within ten months of the second stage procedure.

Cite this article: Bone Joint J 2013;95-B:1681–6.


The Bone & Joint Journal
Vol. 95-B, Issue 6 | Pages 747 - 757
1 Jun 2013
Jameson SS Baker PN Mason J Rymaszewska M Gregg PJ Deehan DJ Reed MR

The popularity of cementless total hip replacement (THR) has surpassed cemented THR in England and Wales. This retrospective cohort study records survival time to revision following primary cementless THR with the most common combination (accounting for almost a third of all cementless THRs), and explores risk factors independently associated with failure, using data from the National Joint Registry for England and Wales. Patients with osteoarthritis who had a DePuy Corail/Pinnacle THR implanted between the establishment of the registry in 2003 and 31 December 2010 were included within analyses. There were 35 386 procedures. Cox proportional hazard models were used to analyse the extent to which the risk of revision was related to patient, surgeon and implant covariates. The overall rate of revision at five years was 2.4% (99% confidence interval 2.02 to 2.79). In the final adjusted model, we found that the risk of revision was significantly higher in patients receiving metal-on-metal (MoM: hazard ratio (HR) 1.93, p < 0.001) and ceramic-on-ceramic bearings (CoC: HR 1.55, p = 0.003) compared with the best performing bearing (metal-on-polyethylene). The risk of revision was also greater for smaller femoral stems (sizes 8 to 10: HR 1.82, p < 0.001) compared with mid-range sizes. In a secondary analysis of only patients where body mass index (BMI) data were available (n = 17 166), BMI ≥ 30 kg/m2 significantly increased the risk of revision (HR 1.55, p = 0.002). The influence of the bearing on the risk of revision remained significant (MoM: HR 2.19, p < 0.001; CoC: HR 2.09, p = 0.001). The risk of revision was independent of age, gender, head size and offset, shell, liner and stem type, and surgeon characteristics.

We found significant differences in failure between bearing surfaces and femoral stem size after adjustment for a range of covariates in a large cohort of single-brand cementless THRs. In this study of procedures performed since 2003, hard bearings had significantly higher rates of revision, but we found no evidence that head size had an effect. Patient characteristics, such as BMI and American Society of Anesthesiologists grade, also influence the survival of cementless components.

Cite this article: Bone Joint J 2013;95-B:747–57.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 3 - 6
1 Nov 2013
Wassef AJ Schmalzried TP

A modular femoral head–neck junction has practical advantages in total hip replacement. Taper fretting and corrosion have so far been an infrequent cause of revision. The role of design and manufacturing variables continues to be debated. Over the past decade several changes in technology and clinical practice might result in an increase in clinically significant taper fretting and corrosion. Those factors include an increased usage of large diameter (36 mm) heads, reduced femoral neck and taper dimensions, greater variability in taper assembly with smaller incision surgery, and higher taper stresses due to increased patient weight and/or physical activity. Additional studies are needed to determine the role of taper assembly compared with design, manufacturing and other implant variables.

Cite this article: Bone Joint J 2013;95-B, Supple A:3–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 4 | Pages 455 - 459
1 Apr 2006
Shetty NR Hamer AJ Kerry RM Stockley I Eastell R Wilkinson JM

The aims of this study were to examine the repeatability of measurements of bone mineral density (BMD) around a cemented polyethylene Charnley acetabular component using dual-energy x-ray absorptiometry and to determine the longitudinal pattern of change in BMD during the first 24 months after surgery.

The precision of measurements of BMD in 19 subjects ranged from 7.7% to 10.8% between regions, using a four-region-of-interest model. A longitudinal study of 27 patients demonstrated a transient decrease in net pelvic BMD during the first 12 months, which recovered to baseline at 24 months. The BMD in the region medial to the dome of the component reduced by between 7% and 10% during the first three months, but recovered to approximately baseline values by two years.

Changes in BMD in the pelvis around cemented acetabular components may be measured using dual-energy x-ray absorptiometry. Bone loss after insertion of a cemented Charnley acetabular component is small, transient and occurs mainly at the medial wall of the acetabulum. After two years, bone mass returns to baseline values, with a pattern suggesting a uniform transmission of load to the acetabulum.


Bone & Joint 360
Vol. 4, Issue 4 | Pages 2 - 7
1 Aug 2015
Nicol S Jackson M Monsell F

This review explores recent advances in fixator design and used in contemporary orthopaedic practice including the management of bone loss, complex deformity and severe isolated limb injury.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1366 - 1371
1 Oct 2013
Muir D Aoina J Hong T Mason R

We performed a retrospective review of a consecutive series of 178 Mobility total ankle replacements (TARs) performed by three surgeons between January 2004 and June 2009, and analysed radiological parameters and clinical outcomes in a subgroup of 129 patients. The mean follow-up was 4 years (2 to 6.3). A total of ten revision procedures (5.6%) were undertaken. The mean Ankle Osteoarthritis Scale (AOS) pain score was 17 (0 to 88) and 86% of patients were clinically improved at follow-up. However, 18 patients (18 TARs, 14%) had a poor outcome with an AOS pain score of > 30. A worse outcome was associated with a pre-operative diagnosis of post-traumatic degenerative arthritis. However, no pre- or post-operative radiological parameters were significantly associated with a poor outcome. Of the patients with persistent pain, eight had predominantly medial-sided pain. Thirty TARs (29%) had a radiolucency in at least one zone.

The outcome of the Mobility TAR at a mean of four years is satisfactory in > 85% of patients. However, there is a significant incidence of persistent pain, particularly on the medial side, for which we were unable to establish a cause.

Cite this article: Bone Joint J 2013;95-B:1366–71.


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1493 - 1499
1 Nov 2015
Pandit H Hamilton TW Jenkins C Mellon SJ Dodd CAF Murray DW

This prospective study reports the 15-year survival and ten-year functional outcome of a consecutive series of 1000 minimally invasive Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women, 52%, mean age 66 years; 32 to 88). These were implanted by two surgeons involved with the design of the prosthesis to treat anteromedial osteoarthritis and spontaneous osteonecrosis of the knee, which are recommended indications. Patients were prospectively identified and followed up independently for a mean of 10.3 years (5.3 to 16.6).

At ten years, the mean Oxford Knee Score was 40 (standard deviation (sd) 9; 2 to 48): 79% of knees (349) had an excellent or good outcome. There were 52 implant-related re-operations at a mean of 5.5 years (0.2 to 14.7). The most common reasons for re-operation were arthritis in the lateral compartment (2.5%, 25 knees), bearing dislocation (0.7%, seven knees) and unexplained pain (0.7%, seven knees). When all implant-related re-operations were considered as failures, the ten-year rate of survival was 94% (95% confidence interval (CI) 92 to 96) and the 15-year survival rate 91% (CI 83 to 98). When failure of the implant was the endpoint the 15-year survival was 99% (CI 96 to 100).

This is the only large series of minimally invasive UKAs with 15-year survival data. The results support the continued use of minimally invasive UKA for the recommended indications.

Cite this article: Bone Joint J 2015;97-B:1493–99.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1045 - 1050
1 Aug 2012
Malek IA King A Sharma H Malek S Lyons K Jones S John A

Plasma levels of cobalt and chromium ions and Metal Artefact Reduction Sequence (MARS)-MRI scans were performed on patients with 209 consecutive, unilateral, symptomatic metal-on-metal (MoM) hip arthroplasties. There was wide variation in plasma cobalt and chromium levels, and MARS-MRI scans were positive for adverse reaction to metal debris (ARMD) in 84 hips (40%). There was a significant difference in the median plasma cobalt and chromium levels between those with positive and negative MARS-MRI scans (p < 0.001). Compared with MARS-MRI as the potential reference standard for the diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt or chromium with a cut-off of > 7 µg/l was 57%. The specificity was 65%, positive predictive value was 52% and the negative predictive value was 69% in symptomatic patients. A lowered threshold of > 3.5 µg/l for cobalt and chromium ion levels improved the sensitivity and negative predictive value to 86% and 74% but at the expense of specificity (27%) and positive predictive value (44%).

Metal ion analysis is not recommended as a sole indirect screening test in the surveillance of symptomatic patients with a MoM arthroplasty. The investigating clinicians should have a low threshold for obtaining cross-sectional imaging in these patients, even in the presence of low plasma metal ion levels.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 297 - 301
1 Mar 2012
Haddad FS Konan S

An international faculty of orthopaedic surgeons presented their work on the current challenges in hip surgery at the London Hip Meeting which was attended by over 400 delegates. The topics covered included femoroacetabular impingement, thromboembolic phenomena associated with hip surgery, bearing surfaces (including metal-on-metal articulations), outcomes of hip replacement surgery and revision hip replacement. We present a concise report of the current opinions on hip surgery from this meeting with appropriate references to the current literature.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1106 - 1113
1 Aug 2008
Richards L Brown C Stone MH Fisher J Ingham E Tipper JL

Nanometre-sized particles of ultra-high molecular weight polyethylene have been identified in the lubricants retrieved from hip simulators. Tissue samples were taken from seven failed Charnley total hip replacements, digested using strong alkali and analysed using high-resolution field emission gun-scanning electron microscopy to determine whether nanometre-sized particles of polyethylene debris were generated in vivo. A randomised method of analysis was used to quantify and characterise all the polyethylene particles isolated.

We isolated nanometre-sized particles from the retrieved tissue samples. The smallest identified was 30 nm and the majority were in the 0.1 μm to 0.99 μm size range. Particles in the 1.0 μm to 9.99 μm size range represented the highest proportion of the wear volume of the tissue samples, with 35% to 98% of the total wear volume comprised of particles of this size. The number of nanometre-sized particles isolated from the tissues accounted for only a small proportion of the total wear volume. Further work is required to assess the biological response to nanometre-sized polyethylene particles.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 809 - 813
1 Jun 2015
Butt DA Hester T Bilal A Edmonds M Kavarthapu V

Charcot neuro-osteoarthropathy (CN) of the midfoot presents a major reconstructive challenge for the foot and ankle surgeon. The Synthes 6 mm Midfoot Fusion Bolt is both designed and recommended for patients who have a deformity of the medial column of the foot due to CN. We present the results from the first nine patients (ten feet) on which we attempted to perform fusion of the medial column using this bolt. Six feet had concurrent hindfoot fusion using a retrograde nail. Satisfactory correction of deformity of the medial column was achieved in all patients. The mean correction of calcaneal pitch was from 6° (-15° to +18°) pre-operatively to 16° (7° to 23°) post-operatively; the mean Meary angle from 26° (3° to 46°) to 1° (1° to 2°); and the mean talometatarsal angle on dorsoplantar radiographs from 27° (1° to 48°) to 1° (1° to 3°).

However, in all but two feet, at least one joint failed to fuse. The bolt migrated in six feet, all of which showed progressive radiographic osteolysis, which was considered to indicate loosening. Four of these feet have undergone a revision procedure, with good radiological evidence of fusion. The medial column bolt provided satisfactory correction of the deformity but failed to provide adequate fixation for fusion in CN deformities in the foot.

In its present form, we cannot recommend the routine use of this bolt.

Cite this article: Bone Joint J 2015; 97-B:809–13


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1348 - 1353
1 Oct 2013
Valenzuela GA Jacobson NA Buzas D Korecki TD Valenzuela RG Teitge RA

The outcome of high tibial osteotomy (HTO) deteriorates with time, and additional procedures may be required. The aim of this study was to compare the clinical and radiological outcomes between unicompartmental knee replacement (UKR) and total knee replacement (TKR) after HTO as well as after primary UKR. A total of 63 patients (63 knees) were studied retrospectively and divided into three groups: UKR after HTO (group A; n = 22), TKR after HTO (group B; n = 18) and primary UKR (group C; n = 22). The Oxford knee score (OKS), Knee Society score (KSS), hip–knee–ankle angles, mechanical axis and patellar height were evaluated pre- and post-operatively. At a mean of 64 months (19 to 180) post-operatively the mean OKS was 43.8 (33 to 49), 43.3 (30 to 48) and 42.5 (29 to 48) for groups A, B and C, respectively (p = 0.73). The mean KSS knee score was 88.8 (54 to 100), 88.11 (51 to 100) and 85.3 (45 to 100) for groups A, B and C, respectively (p = 0.65), and the mean KSS function score was 85.0 (50 to 100) in group A, 85.8 (20 to 100) in group B and 79.3 (50 to 100) in group C (p = 0.48). Radiologically the results were comparable for all groups except for patellar height, with a higher incidence of patella infra following a previous HTO (p = 0.02).

Cite this article: Bone Joint J 2013;95-B:1348–53.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1183 - 1191
1 Sep 2015
Briggs TWR Hanna SA Kayani B Tai S Pollock RC Cannon SR Blunn GW Carrington RWJ

The long term biological effects of wear products following total hip arthroplasty (THA) are unclear. However, the indications for THA are expanding, with increasingly younger patients undergoing the procedure.

This prospective, randomised study compared two groups of patients undergoing THA after being randomised to receive one of two different bearing surfaces: metal-on-polyethylene (MoP) n = 22 and metal-on-metal (MoM) n = 23. We investigated the relationship between three variables: bearing surface (MoP vs MoM), whole blood levels of chromium (Cr) and cobalt (Co) and chromosomal aberrations in peripheral lymphocyte pre-operatively and at one, two and five years post-surgery.

Our results demonstrated significantly higher mean cobalt and chromium (Co and Cr) blood levels in the MoM group at all follow-up points following surgery (p < 0.01), but there were no significant differences in the chromosomal aberration indices between MoM and MoP at two or five years (two years: p = 0.56, p = 0.08, p = 0.91, p = 0.51 and five years: p = 0.086, p = 0.73, p = 0.06, p = 0.34) for translocations, breaks, loss and gain of chromosomes respectively. Regression analysis showed a strong linear relationship between Cr levels and the total chromosomal aberration indices in the MoM group (R2 = 0.90016), but this was not as strong for Co (R2 = 0.68991). In the MoP group, the analysis revealed a poor relationship between Cr levels and the total chromosomal aberration indices (R2 = 0.23908) but a slightly stronger relationship for Co (R2 = 0.64292). Across both groups, Spearman’s correlation detected no overall association between Co and Cr levels and each of the studied chromosomal aberrations. There remains no clear indication which THA bearing couple is the most biocompatible, especially in young active patients. While THA continues to be very successful at alleviating pain and restoring function, the long-term biological implications of the procedure still require further scrutiny.

Cite this article: Bone Joint J 2015;97-B:1183–91.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 905 - 910
1 Jul 2015
Hsu C Lin P Kuo F Wang J

Tranexamic acid (TXA), an inhibitor of fibrinolysis, reduces blood loss after total knee arthroplasty. However, its effect on minimally invasive total hip arthroplasty (THA) is not clear. We performed a prospective, randomised double-blind study to evaluate the effect of two intravenous injections of TXA on blood loss in patients undergoing minimally invasive THA.

In total, 60 patients (35 women and 25 men with a mean age of 58.1 years; 17 to 84) who underwent unilateral minimally invasive uncemented THA were randomly divided into the study group (30 patients, 20 women and ten men with a mean age of 56.5 years; 17 to 79) that received two intravenous injections 1 g of TXA pre- and post-operatively (TXA group), and a placebo group (30 patients, 15 women and 15 men with a mean age of 59.5 years; 23 to 84). We compared the peri-operative blood loss of the two groups. Actual blood loss was calculated from the maximum reduction in the level of haemoglobin. All patients were followed clinically for the presence of venous thromboembolism.

The TXA group had a lower mean intra-operative blood loss of 441 ml (150 to 800) versus 615 ml (50 to 1580) in the placebo (p = 0.044), lower mean post-operative blood loss (285 ml (120 to 570) versus 392 ml (126 to 660) (p = 0.002), lower mean total blood loss (1070 ml (688 to 1478) versus 1337 ml (495 to 2238) (p = 0.004) and lower requirement for transfusion (p = 0.021). No patients in either group had symptoms of venous thromboembolism or wound complications.

This prospective, randomised controlled study showed that a regimen of two intravenous injections of 1 g TXA is effective for blood conservation after minimally invasive THA.

Cite this article: Bone Joint J 2015;97-B:905–10.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 249 - 253
1 Feb 2012
Sorlí L Puig L Torres-Claramunt R González A Alier A Knobel H Salvadó M Horcajada JP

Patients with infected arthroplasties are normally treated with a two-stage exchange procedure using polymethylmethacrylate bone cement spacers impregnated with antibiotics. However, spacers may act as a foreign body to which micro-organisms may adhere and grow. In this study it was hypothesised that subclinical infection may be diagnosed with sonication of the surface biofilm of the spacer. The aims were to assess the presence of subclinical infection through sonication of the spacer at the time of a second-stage procedure, and to determine the relationship between subclinical infection and the clinical outcome. Of 55 patients studied, 11 (20%) were diagnosed with subclinical infection. At a mean follow-up of 12 months (interquartile range 6 to 18), clinical failure was found in 18 (32.7%) patients. Of the patients previously diagnosed with subclinical infection, 63% (7 of 11) had failed compared with 25% (11 of 44) of those without subclinical infection (odds ratio 5.25, 95% confidence interval 1.29 to 21.4, p = 0.021). Sonication of the biofilm of the surface of the spacer is useful in order to exclude subclinical infection and therefore contributes to improving the outcome after two-stage procedures.