The advent of computer-assisted knee replacement surgery has focused interest on the alignment of the components. However, there is confusion at times between the alignment of the limb as a whole and that of the components. The interaction between them is discussed in this article. Alignment is expressed relative to some reference axis or plane and measurements will vary depending on what is selected as the reference. The validity of different reference axes is discussed. Varying prosthetic alignment has direct implications for surrounding soft-tissue tension. In this context the interaction between alignment and soft-tissue balance is explored and the current knowledge of the relationship between alignment and outcome is summarised.
Elastic stable intramedullary nailing (ESIN)
is generally acknowledged to be the treatment of choice for displaced diaphyseal
femoral fractures in children over the age of three years, although
complication rates of up to 50% are described. Pre-bending the nails
is recommended, but there are no published data to support this.
Using synthetic bones and a standardised simulated fracture, we
performed biomechanical testing to determine the influence on the
stability of the fracture of pre-bending the nails before implantation.
Standard ESIN was performed on 24 synthetic femoral models with
a spiral fracture. In eight cases the nails were inserted without
any pre-bending, in a further eight cases they were pre-bent to
30° and in the last group of eight cases they were pre-bent to 60°. Mechanical
testing revealed that pre-bending to 60° produced a significant
increase in the stiffness or stability of the fracture. Pre-bending
to 60° showed a significant positive influence on the stiffness
compared with unbent nails. Pre-bending to 30° improved stiffness
only slightly. These findings validate the recommendations for pre-bending,
but the degree of pre-bend should exceed 30°. Adopting higher degrees
of pre-bending should improve stability in spiral fractures and
reduce the complications of varus deformity and shortening.
The Morscher-Spotorno (MS-30) femoral stem is a stainless-steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement. We report the results at ten years of a consecutive series of 124 total hip replacements in 121 patients with the matt-surfaced MS-30 stem and an alumina ceramic head of 28-mm diameter. All the stems were fixed with Palacos bone cement with gentamicin using a modern cementing technique. They were combined with an uncemented, press-fit cup. The mean period of observation was 10.2 years (8.3 to 12.1) and no patient was lost to follow-up. Twenty-seven patients (22%) died with the implant None of the stems or cups had been revised. The Harris hip score was excellent or good in 97% (85 hips) and moderate in 3% (three hips). Radiologically, six hips (6.8%) had osteolysis adjacent to the stem, mostly in Gruen zone 7. Twenty (22.7%) showed one or more radiolucent lines. Twenty-two stems (25%) had subsided by 2 mm to 5 mm. In these cases two showed osteolysis (9.1%) with subsidence and four without (6.1%). Radiolucent lines were seen in seven with migration (31.8%) and in 13 without (19.7%). No infections and no acetabular osteolysis were observed. The clinical results were excellent with survivorship after ten years of 100% and only a slightly statistically non-significant higher rate of osteolysis and radiolucency in cases of subsidence.
A reduced femoral offset in total hip replacement has been thought to be disadvantageous. We reviewed the results of 54 consecutive primary total hip replacements in 49 patients (mean age of 68 years) performed between August 1990 and December 1994, with a mean follow-up of 8.8 years ( At their latest follow-up, surviving patients had a significant improvement in the performance of their hip. Three had undergone revision, one each for deep infection, recurrent dislocation and late pain with subluxation. No hips had been revised, or were at risk of revision, for aseptic loosening. The mean annual linear rate of wear was 0.2 mm (
Allograft bone is widely used in orthopaedic surgery, but peri-operative infection of the graft remains a common and disastrous complication. The efficacy of systemic prophylactic antibiotics is unproven, and since the graft is avascular it is likely that levels of antibiotic in the graft are low. Using an electrical potential to accelerate diffusion of antibiotics into allograft bone, high levels were achieved in specimens of both sheep and human allograft. In human bone these ranged from 187.1 mg/kg in endosteal ( Structural allograft can be supplemented directly with antibiotics using iontophoresis. The technique is simple and inexpensive and offers a potential means of reducing the rate of peri-operative infection in allograft surgery. Iontophoresis into allograft bone may also be applicable to other therapeutic compounds.
It has been suggested that a supramalleolar osteotomy can return the load distribution in the ankle joint to normal. However, due to the lack of biomechanical data, this supposition remains empirical. The purpose of this biomechanical study was to determine the effect of simulated supramalleolar varus and valgus alignment on the tibiotalar joint pressure, in order to investigate its relationship to the development of osteoarthritis. We also wished to establish the rationale behind corrective osteotomy of the distal tibia. We studied 17 cadaveric lower legs and quantified the changes in pressure and force transfer across the tibiotalar joint for various degrees of varus and valgus deformity in the supramalleolar area. We assumed that a supramalleolar osteotomy which created a varus deformity of the ankle would result in medial overload of the tibiotalar joint. Similarly, we thought that creating a supramalleolar valgus deformity would cause a shift in contact towards the lateral side of the tibiotalar joint. The opposite was observed. The restricting role of the fibula was revealed by carrying out an osteotomy directly above the syndesmosis. In end-stage ankle osteoarthritis with either a valgus or varus deformity, the role of the fibula should be appreciated and its effect addressed where appropriate.
This article is a systematic review of the published
literature about the biomechanics, functional outcome and complications
of intramedullary nailing of fractures of the distal radius. We searched the Medline and EMBASE databases and included all
studies which reported the outcome of intramedullary (IM) nailing
of fractures of the distal radius. Data about functional outcome,
range of movement (ROM), strength and complications, were extracted.
The studies included were appraised independently by both authors
using a validated quality assessment scale for non-controlled studies
and the CONSORT statement for randomised controlled trials (RCTs). The search strategy revealed 785 studies, of which 16 were included
for full paper review. These included three biomechanical studies,
eight case series and five randomised controlled trials (RCTs). The biomechanical studies concluded that IM nails were at least
as strong as locking plates. The clinical studies reported that
IM nailing gave a comparable ROM, functional outcome and grip strength
to other fixation techniques. However, the mean complication rate of intramedullary nailing
was 17.6% (0% to 50%). This is higher than the rates reported in
contemporary studies for volar plating. It raises concerns about
the role of intramedullary nailing, particularly when comparative
studies have failed to show that it has any major advantage over
other techniques. Further adequately powered RCTs comparing the
technique to both volar plating and percutaneous wire fixation are needed. Cite this article:
Primary total knee arthroplasty (TKA) is a reliable
procedure with reproducible long-term results. Nevertheless, there
are conditions related to the type of patient or local conditions
of the knee that can make it a difficult procedure. The most common
scenarios that make it difficult are discussed in this review. These
include patients with many previous operations and incisions, and
those with severe coronal deformities, genu recurvatum, a stiff knee,
extra-articular deformities and those who have previously undergone
osteotomy around the knee and those with chronic dislocation of
the patella. Each condition is analysed according to the characteristics of
the patient, the pre-operative planning and the reported outcomes. When approaching the difficult primary TKA surgeons should use
a systematic approach, which begins with the review of the existing
literature for each specific clinical situation. Cite this article:
Antiplatelet agents are widely prescribed for the primary and secondary prevention of cardiovascular events. A common clinical problem facing orthopaedic and trauma surgeons is how to manage patients receiving these agents who require surgery, either electively or following trauma. The dilemma is to balance the risk of increased blood loss if the antiplatelet agents are continued peri-operatively against the risk of coronary artery/stent thrombosis and/or other vascular event if the drugs are stopped. The traditional approach of stopping these medications up to two weeks before surgery appears to pose significant danger to patients and may require review. This paper covers the important aspects regarding the two most commonly prescribed antiplatelet agents, aspirin and clopidogrel.
Metaphyseal bone loss is common with revision
total knee replacement (RTKR). Using the Anderson Orthopaedic Research
Institute (AORI) classification, type 2-B and type 3 defects usually
require large metal blocks, bulk structural allograft or highly
porous metal cones. Tibial and femoral trabecular metal metaphyseal
cones are a unique solution for large bone defects. These cones
substitute for bone loss, improve metaphyseal fixation, help correct
malalignment, restore the joint line and may permit use of a shorter
stem. The technique for insertion involves sculpturing of the remaining
bone with a high speed burr and rasp, followed by press-fit of the
cone into the metaphysis. The fixation and osteoconductive properties
of the porous cone outer surface allow ingrowth and encourage long-term
biological fixation. The revision knee component is then cemented
into the porous cone inner surface, which provides superior fixation
compared with cementing into native but deficient metaphyseal bone.
The advantages of the cone compared with allograft include: technical
ease, biological fixation, no resorption, and possibly a lower risk
of infection. The disadvantages include: difficult extraction and
relatively short-term follow-up. Several studies using cones report
promising short-term results for the reconstruction of large bone
defects in RTKR. Cite this article:
Of the 11 054 Charnley low-frictional torque arthroplasties carried out at our hospital between 1962 and 1977, 110 (94 patients) had a minimum follow-up of 30 years with a mean of 32.3 years (30.0 to 40.5). The mean age of the patients at operation was 43.3 years (17.0 to 65.0) and 75.7 years (51.0 to 97.0) at follow-up. Overall, 90% of hips (99) were free from pain and activity was reported as normal in 58% of the patients. A total of 13 hips (11.8%) were revised at a mean follow-up of 32.3 years (30.0 to 39.5), with wear and loosening of the acetabular component as the main indications. The clinical results did not reflect the mechanical state of the implant. Follow-up with sequential radiographs of good quality is essential. Revision for radiological changes alone must be accepted if gross loss of bone stock is to be avoided. Improvements in the design, materials and operative technique, based on the long-term outcome, are highlighted.
We investigated the effect of pre-heating a femoral component on the porosity and strength of bone cement, with or without vacuum mixing used for total hip replacement. Cement mantles were moulded in a manner simulating clinical practice for cemented hip replacement. During polymerisation, the temperature was monitored. Specimens of cement extracted from the mantles underwent bending or fatigue tests, and were examined for porosity. Pre-heating the stem alone significantly increased the mean temperature values measured within the mantle (+14.2°C) (p <
0.001) and reduced the mean curing time (−1.5 min) (p <
0.001). The addition of vacuum mixing modulated the mean rise in the temperature of polymerisation to 11°C and reduced the mean duration of the process by one minute and 50 seconds (p = 0.01 and p <
0.001, respectively). In all cases, the maximum temperature values measured in the mould simulating the femur were <
50°C. The mixing technique and pre-heating the stem slightly increased the static mechanical strength of bone cement. However, the fatigue life of the cement was improved by both vacuum mixing and pre-heating the stem, but was most marked (+ 280°C) when these methods were combined. Pre-heating the stem appears to be an effective way of improving the quality of the cement mantle, which might enhance the long-term performance of bone cement, especially when combined with vacuum mixing.
We describe the results of Copeland surface replacement shoulder arthroplasty using the mark III prosthesis in patients over 80 years of age. End-stage arthritis of the shoulder is a source of significant pain and debilitating functional loss in the elderly. An arthroplasty offers good relief of pain and may allow the patient to maintain independence. The risk-benefit ratio of shoulder replacement may be felt to be too high in an elderly age group, but there is no published evidence to support this theory. We have assessed whether the procedure was as reliable and safe as previously seen in a younger cohort of patients. Between 1993 and 2003, 213 Copeland surface replacement arthroplasty procedures were performed in our unit, of which 29 (13.6%) were undertaken in patients over the age of 80. This group of patients was followed up for a mean of 4.5 years (2.1 to 9.3). Their mean age was 84.3 years (81 to 93), the mean operating time was 40 minutes (30 to 45) and the mean in-patient stay was five days (2 to 21). There were no peri-operative deaths or significant complications. The mean Constant score adjusted for age and gender, improved from 15.1% to 77%. Copeland surface replacement shoulder arthroplasty may be performed with minimal morbidity and rapid rehabilitation in the elderly.
It is becoming increasingly common for a patient
to have ipsilateral hip and knee replacements. The inter-prosthetic (IP)
distance, the distance between the tips of hip and knee prostheses,
has been thought to be associated with an increased risk of IP fracture.
Small gap distances are generally assumed to act as stress risers,
although there is no real biomechanical evidence to support this. The purpose of this study was to evaluate the influence of IP
distance, cortical thickness and bone mineral density on the likelihood
of an IP femoral fracture. A total of 18 human femur specimens were randomised into three
groups by bone density and cortical thickness. For each group, a
defined IP distance of 35 mm, 80 mm or 160 mm was created by choosing
the appropriate lengths of component. The maximum fracture strength
was determined using a four-point bending test. The fracture force of all three groups was similar (p = 0.498).
There was a highly significant correlation between the cortical
area and the fracture strength (r = 0.804, p <
0.001), whereas
bone density showed no influence. This study suggests that the IP distance has little influence
on fracture strength in IP femoral fractures: the thickness of the
cortex seems to be the decisive factor. Cite this article:
Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998. The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification. The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively. Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p <
0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97).
Cementless femoral stems are currently preferred
for total hip replacement (THR) in the United States. Improvements
in stem design, instrumentation and surgical technique have made
this technology highly successful, reproducible, and applicable
to the vast majority of patients requiring a THR. However, there
are ongoing developments in some aspects of stem design that influence
clinical results, the incidence of complications and their inherent adaptability
in accommodating the needs of individual patients. Here we examine
some of these design features. Cite this article:
In March 2012, an algorithm for the treatment
of intertrochanteric fractures of the hip was introduced in our academic
department of Orthopaedic Surgery. It included the use of specified
implants for particular patterns of fracture. In this cohort study,
102 consecutive patients presenting with an intertrochanteric fracture
were followed prospectively (post-algorithm group). Another 117
consecutive patients who had been treated immediately prior to the
implementation of the algorithm were identified retrospectively
as a control group (pre-algorithm group). The total cost of the
implants prior to implementation of the algorithm was $357 457 (mean:
$3055 (1947 to 4133)); compared with $255 120 (mean: $2501 (1052
to 4133)) after its implementation. There was a trend toward fewer complications
in patients who were treated using the algorithm (33% pre- The implementation of an evidence-based algorithm for the treatment
of intertrochanteric fractures reduced costs while maintaining quality
of care with a lower rate of complications and re-admissions. Cite this article:
We investigated patterns of refracture and their risk factors in patients with congenital pseudarthrosis of the tibia after Ilizarov osteosynthesis. We studied 43 cases in 23 patients. Temporal and spatial patterns of refracture and refracture-free survival were analysed in each case. The refracture-free rate of cumulative survival was 47% at five years and did not change thereafter. Refracture occurred at the previous pseudarthrosis in 16 of 19 cases of refracture. The risk of refracture was significantly higher when osteosynthesis was performed below the age of four years, when the tibial cross-sectional area was narrow, and when associated with persistent fibular pseudarthrosis. Refracture occurs frequently after successful osteosynthesis in these patients. Delaying osteosynthesis, maximising the tibial cross-sectional area and stabilising the fibula may reduce the risk of refracture.
Flexible intramedullary nailing is gaining popularity as an effective method of treating long-bone fractures in children. We retrospectively reviewed the records and radiographs of 56 unstable fractures of the tibia in 54 children treated between March 1997 and May 2005. All were followed up for at least two months after the removal of the nails. Of the 56 tibial fractures, 13 were open. There were no nonunions. The mean time to clinical and radiological union was ten weeks. Complications included residual angulation of the tibia, leg-length discrepancy, deep infection and failures of fixation. All achieved an excellent functional outcome. We conclude that flexible intramedullary fixation is an easy and effective method of management of both open and closed unstable fractures of the tibia in children.