We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other polished tapered stem designs.
We report on 397 consecutive revision total hip
replacements in 371 patients with a mean clinical and radiological follow-up
of 12.9 years (10 to 17.7). The mean age at surgery was 69 years
(37 to 93). A total of 28 patients (8%) underwent further revision,
including 16 (4%) femoral components. In all 223 patients (56%,
233 hips) died without further revision and 20 patients (5%, 20
hips) were lost to follow-up. Of the remaining patients, 209 (221
hips) were available for clinical assessment and 194 (205 hips)
for radiological review at mean follow-up of 12.9 years (10 to 17.7). The mean Harris Hip Score improved from 58.7 (11 to 92) points
to 80.7 (21 to 100) (p <
0.001) and the mean Merle d’Aubigné and
Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to
6) and 4.3 (2 to 6), respectively for pain, mobility and function.
Radiographs showed no lucencies around 186 (90.7%) femoral stems
with stable bony ingrowth seen in 199 stems (97%). The survival
of the S-ROM femoral stem at 15 years with revision for any reason as
the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8)
and with revision for aseptic loosening as the endpoint 99.3% (95%
CI 97.2 to 99.8). We have shown excellent long-term survivorship and good clinical
outcome of a cementless hydroxyapatite proximally-coated modular
femoral stem in revision hip surgery. Cite this article:
The purpose of this study was to evaluate A total of 60 Sprague-Dawley rats (125 g to 149 g) were implanted
subcutaneously with SWCNT/PLAGA composites (10 mg SWCNT and 1gm
PLAGA 12 mm diameter two-dimensional disks), and at two, four, eight
and 12 weeks post-implantation were compared with control (Sham)
and PLAGA (five rats per group/point in time). Rats were observed
for signs of morbidity, overt toxicity, weight gain and food consumption,
while haematology, urinalysis and histopathology were completed
when the animals were killed.Objectives
Methods
In this prospective study we studied the effect
of the inclination angle of the acetabular component on polyethylene wear
and component migration in cemented acetabular sockets using radiostereometric
analysis. A total of 120 patients received either a cemented Reflection
All-Poly ultra-high-molecular-weight polyethylene or a cemented
Reflection All-Poly highly cross-linked polyethylene acetabular
component, combined with either cobalt–chrome or Oxinium femoral
heads. Femoral head penetration and migration of the acetabular
component were assessed with repeated radiostereometric analysis
for two years. The inclination angle was measured on a standard
post-operative anteroposterior pelvic radiograph. Linear regression
analysis was used to determine the relationship between the inclination
angle and femoral head penetration and migration of the acetabular component. We found no relationship between the inclination angle and penetration
of the femoral head at two years’ follow-up (p = 0.9). Similarly,
our data failed to reveal any statistically significant correlation
between inclination angle and migration of these cemented acetabular
components (p = 0.07 to p = 0.9).
We hypothesised that a minimally invasive peroneus
brevis tendon transfer would be effective for the management of
a chronic rupture of the Achilles tendon. In 17 patients (three
women, 14 men) who underwent minimally invasive transfer and tenodesis
of the peroneus brevis to the calcaneum, at a mean follow-up of
4.6 years (2 to 7) the modified Achilles tendon total rupture score
(ATRS) was recorded and the maximum circumference of the calf of
the operated and contralateral limbs was measured. The strength
of isometric plantar flexion of the gastrocsoleus complex and of
eversion of the ankle were measured bilaterally. Functional outcomes
were classified according to the four-point Boyden scale. At the latest review, the mean maximum circumference of the calf
of the operated limb was not significantly different from the pre-operative
mean value, (41.4 cm, 32 to 50 This procedure may be safely performed, is minimally invasive,
and allows most patients to return to pre-injury sport and daily
activities. Cite this article:
The February 2015 Foot &
Ankle Roundup360 looks at: Syndesmosis screw removal in randomised controlled trial; Diagnostic value of Hawkins sign; Chevron rules supreme?; Diabetes and ankle replacement; Fixed-bearing ankle replacement; Fusion for osteomyelitis of the ankle; ‘Reformed’ fallers.
The February 2015 Hip &
Pelvis Roundup360 looks at: Hip arthroplasty in Down syndrome; Bulk femoral autograft successful in acetabular reconstruction; Arthroplasty follow-up: is the internet the solution?; Total hip arthroplasty following acetabular fracture; Salvage arthroplasty following failed hip internal fixation; Bone banking sensible financially and clinically; Allogenic blood transfusion in arthroplasty.
The major problem with repair of an articular cartilage injury
is the extensive difference in the structure and function of regenerated,
compared with normal cartilage. Our work investigates the feasibility
of repairing articular osteochondral defects in the canine knee
joint using a composite lamellar scaffold of nano-ß-tricalcium phosphate
(ß-TCP)/collagen (col) I and II with bone marrow stromal stem cells
(BMSCs) and assesses its biological compatibility. The bone–cartilage scaffold was prepared as a laminated composite,
using hydroxyapatite nanoparticles (nano-HAP)/collagen I/copolymer
of polylactic acid–hydroxyacetic acid as the bony scaffold, and
sodium hyaluronate/poly(lactic-co-glycolic acid) as the cartilaginous
scaffold. Ten-to 12-month-old hybrid canines were randomly divided
into an experimental group and a control group. BMSCs were obtained
from the iliac crest of each animal, and only those of the third
generation were used in experiments. An articular osteochondral
defect was created in the right knee of dogs in both groups. Those
in the experimental group were treated by implanting the composites
consisting of the lamellar scaffold of ß-TCP/col I/col II/BMSCs.
Those in the control group were left untreated.Objectives
Methods
Cartilage repair in terms of replacement, or
regeneration of damaged or diseased articular cartilage with functional tissue,
is the ‘holy grail’ of joint surgery. A wide spectrum of strategies
for cartilage repair currently exists and several of these techniques
have been reported to be associated with successful clinical outcomes
for appropriately selected indications. However, based on respective
advantages, disadvantages, and limitations, no single strategy, or
even combination of strategies, provides surgeons with viable options
for attaining successful long-term outcomes in the majority of patients.
As such, development of novel techniques and optimisation of current techniques
need to be, and are, the focus of a great deal of research from
the basic science level to clinical trials. Translational research
that bridges scientific discoveries to clinical application involves
the use of animal models in order to assess safety and efficacy
for regulatory approval for human use. This review article provides
an overview of animal models for cartilage repair. Cite this article:
Femoral stem version has a major influence on
impingement and early post-operative stability after total hip arthroplasty
(THA). The main objective of this study was to evaluate the validity
of a novel radiological method for measuring stem version. Anteroposterior
(AP) radiographs and three-dimensional CT scans were obtained for
115 patients (female/male 63/72, mean age 62.5 years (50 to 75))
who had undergone minimally invasive, cementless THA. Stem version was
calculated from the AP hip radiograph by rotation-based change in
the projected prosthetic neck–shaft (NSA*) angle using the mathematical
formula ST = arcos [tan (NSA*) / tan (135)]. We used two independent
observers who repeated the analysis after a six-week interval. Radiological
measurements were compared with 3D-CT measurements by an independent,
blinded external institute. We found a mean difference of 1.2° ( We found that femoral tilt was associated with the mean radiological
measurement error (r = 0.22, p = 0.02). The projected neck–shaft angle is a reliable method for measuring
stem version on AP radiographs of the hip after a THA. However,
a highly standardised radiological technique is required for its
precise measurement. Cite this article:
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
This systematic review of the literature summarises
the clinical experience with ceramic-on-ceramic hip bearings over
the past 40 years and discusses the concerns that exist in relation
to the bearing combination. Loosening, fracture, liner chipping
on insertion, liner canting and dissociation, edge-loading and squeaking
have all been reported, and the relationship between these issues
and implant design and surgical technique is investigated. New design
concepts are introduced and analysed with respect to previous clinical
experience.
The Motec cementless modular metal-on-metal ball-and-socket
wrist arthroplasty was implanted in 16 wrists with scaphoid nonunion
advanced collapse (SNAC; grades 3 or 4) and 14 wrists with scapholunate
advanced collapse (SLAC) in 30 patients (20 men) with severe (grades
3 or 4) post-traumatic osteoarthritis of the wrist. The mean age of
the patients was 52 years (31 to 71). All prostheses integrated
well radiologically. At a mean follow-up of 3.2 years (1.1 to 6.1)
no luxation or implant breakage occurred. Two wrists were converted
to an arthrodesis for persistent pain. Loosening occurred in one
further wrist at five years post-operatively. The remainder demonstrated close
bone–implant contact. The clinical results were good, with markedly
decreased Disabilities of the Arm Shoulder and Hand (DASH) and pain
scores, and increased movement and grip strength. No patient used
analgesics and most had returned to work. Good short-term function was achieved using this wrist arthroplasty
in a high-demand group of patients with post-traumatic osteoarthritis.
We have assessed the different adhesive properties
of some of the most common bacteria associated with periprosthetic
joint infection on various types of ultra high molecular Weight
Polyethylene (UHMWPE). Quantitative Cite this article:
Current analysis of unicondylar knee replacements
(UKRs) by national registries is based on the pooled results of medial
and lateral implants. Consequently, little is known about the differential
performance of medial and lateral replacements and the influence
of each implant type within these pooled analyses. Using data from
the National Joint Registry for England and Wales (NJR) we aimed
to determine the proportion of UKRs implanted on the lateral side
of the knee, and their survival and reason for failure compared
with medial UKRs. By combining information on the side of operation
with component details held on the NJR, we were able to determine
implant laterality (medial
Recent guidance recommends the use of a well-proven
cemented femoral stem for hemiarthroplasty in the management of
fractures of the femoral neck, and the Exeter Trauma Stem (ETS)
has been suggested as an example of such an implant. The design
of this stem was based on the well-proven Exeter Total Hip Replacement
stem (ETHRS). This study assessed the surface finish of the ETS
in comparison with the ETHRS. Two ETSs and two ETHRSs were examined
using a profilometer with a precision of 1 nm and compared with
an explanted Exeter Matt stem. The mean roughness average (RA) of
the ETSs was approximately ten times higher than that of the ETHRSs (0.235 μm
(0.095 to 0.452) Cite this article:
We hypothesised that the use of tantalum (Ta)
acetabular components in revision total hip arthroplasty (THA) was protective
against subsequent failure due to infection. We identified 966 patients
(421 men, 545 women and 990 hips) who had undergone revision THA
between 2000 and 2013. The mean follow up was 40.2 months (3 months
to 13.1 years). The mean age of the men and women was 62.3 years
(31 to 90) and 65.1 years (25 to 92), respectively. Titanium (Ti) acetabular components were used in 536 hips while
Ta components were used in 454 hips. In total, 73 (7.3%) hips experienced
subsequent acetabular failure. The incidence of failure was lower
in the Ta group at 4.4% (20/454) compared with 9.9% (53/536) in
the Ti group (p <
0.001, odds ratio 2.38; 95% CI 1.37 to 4.27).
Among the 144 hips (64 Ta, 80 Ti) for which revision had been performed
because of infection, failure due to a subsequent infection was
lower in the Ta group at 3.1% (2/64) compared with 17.5% (14/80)
for the Ti group (p = 0.006). Thus, the use of Ta acetabular components during revision THA
was associated with a lower incidence of failure from all causes
and Ta components were associated with a lower incidence of subsequent
infection when used in patients with periprosthetic joint infection. Cite this article:
In 2012 we reviewed a consecutive series of 92
uncemented THRs performed between 1986 and 1991 at our institution
using the CLS Spotorno stem, in order to assess clinical outcome
and radiographic data at a minimum of 21 years. The series comprised
92 patients with a mean age at surgery of 59.6 years (39 to 77)
(M:F 43;49). At the time of this review, seven (7.6%) patients had died and
two (2.2%) were lost to follow-up. The 23-year Kaplan–Meier survival
rates were 91.5% (95% confidence intervals (CI) 85.4% to 97.6%;
55 hips at risk) and 80.3% (95% CI, 71.8% to 88.7%; 48 hips at risk)
respectively, with revision of the femoral stem or of any component
as endpoints. At the time of this review, 76 patients without stem
revision were assessed clinically and radiologically (mean follow-up
24.0 years (21.5 to 26.5)). For the 76 unrevised hips the mean Harris
hip score was 87.1 (65 to 97). Femoral osteolysis was detected in
five hips (6.6%) only in Gruen zone 7. Undersized stems were at
higher risk of revision owing to aseptic loosening (p = 0.0003).
Patients implanted with the stem in a varus position were at higher risk
of femoral cortical hypertrophy and thigh pain (p = 0.0006 and p
= 0.0007, respectively). In our study, survival, clinical outcome and radiographic data
remained excellent in the third decade after implantation. Nonetheless,
undersized stems were at higher risk of revision owing to aseptic
loosening. Cite this article: