Advertisement for orthosearch.org.uk
Results 1181 - 1200 of 1506
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 738 - 745
1 Jun 2011
Davda K Lali FV Sampson B Skinner JA Hart AJ

We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging.

Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1217 - 1222
1 Sep 2009
King RJ Makrides P Gill JA Karthikeyan S Krikler SJ Griffin DR

We have developed a novel method of calculating the radiological magnification of the hip using two separate radio-opaque markers. We recruited 74 patients undergoing radiological assessment following total hip replacement. Both the new double marker and a conventional single marker were used by the radiographer at the time of x-ray. The predicted magnification according to each marker was calculated, as was the true radiological magnification of the components. The correlation between true and predicted magnification was good using the double marker (r = 0.90, n = 74, p < 0.001), but only moderate for the single marker (r = 0.50, n = 63, p < 0.001). The median error was significantly less for the double marker than for the single (1.1% vs 4.8%, p < 0.001). The double marker method demonstrated excellent validity (intraclass correlation coefficient = 0.89), in contrast to the single marker (0.32).

The double marker method appears to be superior to the single marker method when used in the clinical environment.


Bone & Joint Research
Vol. 2, Issue 11 | Pages 238 - 244
1 Nov 2013
Keurentjes JC Fiocco M So-Osman C Onstenk R Koopman-Van Gemert AWMM Pöll RG Nelissen RGHH

Objectives

Electronic forms of data collection have gained interest in recent years. In orthopaedics, little is known about patient preference regarding pen-and-paper or electronic questionnaires. We aimed to determine whether patients undergoing total hip (THR) or total knee replacement (TKR) prefer pen-and-paper or electronic questionnaires and to identify variables that predict preference for electronic questionnaires.

Methods

We asked patients who participated in a multi-centre cohort study investigating improvement in health-related quality of life (HRQoL) after THR and TKR using pen-and-paper questionnaires, which mode of questionnaire they preferred. Patient age, gender, highest completed level of schooling, body mass index (BMI), comorbidities, indication for joint replacement and pre-operative HRQoL were compared between the groups preferring different modes of questionnaire. We then performed logistic regression analyses to investigate which variables independently predicted preference of electronic questionnaires.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 42 - 49
1 Apr 2012
Kwon Y Mellon SJ Monk P Murray DW Gill HS

Objectives

Pseudotumours (abnormal peri-prosthetic soft-tissue reactions) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated metal ion levels, suggesting that excessive wear may occur due to edge-loading of these MoM implants. This study aimed to quantify in vivo edge-loading in MoMHRA patients with and without pseudotumours during functional activities.

Methods

The duration and magnitude of edge-loading in vivo was quantified during functional activities by combining the dynamic hip joint segment contact force calculated from the three-dimensional (3D) motion analysis system with the 3D reconstruction of orientation of the acetabular component and each patient’s specific hip joint centre, based on CT scans.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1610 - 1616
1 Dec 2013
Epinette J Asencio G Essig J Llagonne B Nourissat C

We report a multicentre prospective consecutive study assessing the long-term outcome of the proximally hydroxyapatite (HA)-coated ABG II monobloc femoral component in a series of 1148 hips in 1053 patients with a mean age at surgery of 64.77 years (22 to 80) at a mean follow-up of 10.84 years (10 to 15.25). At latest follow-up, the mean total Harris hip score was 94.7 points (sd; 6.87; 49 to 100), and the mean Merle d’Aubigné–Postel score was 17.6 points (sd 1.12; 7 to 18). The mean total Engh radiological score score was 21.54 (sd 5.77; 3.5 to 27), with 95.81% of ‘confirmed ingrowth’, according to Engh’s classification. With aseptic loosening or pain as endpoints, three AGB II stems (0.26%) failed, giving a 99.7% survival rate (se 0.002; 95% confidence interval (CI) 0.994 to 1) at 14 years’ follow-up. The survival of patients ≤ 50 years of age (99.0% (se 0.011; 95% CI 0.969 to 1)) did not differ significantly from those of patients aged > 50 years (99.8% (se 0.002; 95% CI 0.994 to 1)). This study confirmed the excellent long-term results currently achieved with the ABG II proximally HA-coated monobloc stem.

Cite this article: Bone Joint J 2013;95-B:1610–16.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1280 - 1284
1 Sep 2013
Yamaguchi T Matsumine A Niimi R Nakamura T Matsubara T Asanuma K Hasegawa M Sudo A

The aim of this study was to define the incidence of venous thromboembolism (VTE) and risk factors for the development of deep-vein thrombosis (DVT) after the resection of a musculoskeletal tumour. A total of 94 patients who underwent resection of a musculoskeletal tumour between January 2003 and December 2005 were prospectively studied. There were 42 men and 52 women with a mean age of 54.4 years (18 to 86). All patients wore intermittent pneumatic compression devices and graduated compression stockings. Ultrasound examination of the lower limbs was conducted to screen for DVT between the fifth and ninth post-operative days.

DVT was detected in 21 patients (22%). Of these, two were symptomatic (2%). One patient (1%) had a fatal pulmonary embolism. Patients aged ≥ 70 years had an increased risk of DVT (p = 0.004).

The overall incidence of DVT (both symptomatic and asymptomatic) after resection of a musculoskeletal tumour with mechanical prophylaxis was high. It seems that both mechanical and anticoagulant prophylaxis is needed to prevent VTE in patients who have undergone the resection of a musculoskeletal tumour.

Cite this article: Bone Joint J 2013;95-B:1280–4.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 4 | Pages 449 - 454
1 Apr 2006
Hart AJ Hester T Sinclair K Powell JJ Goodship AE Pele L Fersht NL Skinner J

We have studied the relationship between metal ion levels and lymphocyte counts in patients with metal-on-metal hip resurfacings. Peripheral blood samples were analysed for lymphocyte subtypes and whole blood cobalt and chromium ion levels in 68 patients (34 with metal-on-metal hip resurfacings and 34 with standard metal-on-polyethylene total hip replacements). All hip components were radiologically well-fixed and the patients were asymptomatic. Cobalt and chromium levels were significantly elevated in the patients with metal-on-metal hip resurfacings, compared with the patients with standard metal-on-polyethylene designs (p < 0.0001). There was a statistically significant decrease in the level of CD8+ cells (T-cytotoxic/suppressor) (p = 0.005) in the metal-on-metal hip resurfacing group. A threshold level of blood cobalt and chromium ions was associated with reduced CD8+ T-cell counts. We have no evidence that our patients suffered as a result of this reduced level of CD8+ T-cells.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1617 - 1625
1 Dec 2013
Schmitz MWJL Timmer C Rijnen WHC Gardeniers JWM Schreurs BW

Despite the worldwide usage of the cemented Contemporary acetabular component (Stryker), no published data are available regarding its use in patients aged < 50 years. We undertook a mid- to long-term follow-up study, including all consecutive patients aged < 50 years who underwent a primary total hip replacement using the Contemporary acetabular component with the Exeter cemented stem between January 1999 and January 2006. There were 152 hips in 126 patients, 61 men and 65 women, mean age at surgery 37.6 years (16 to 49 yrs). One patient was lost to follow-up.

Mean clinical follow-up of all implants was 7.6 years (0.9 to 12.0). All clinical questionnaire scores, including Harris hip score, Oxford hip score and several visual analogue scales, were found to have improved. The eight year survivorship of all acetabular components for the endpoints revision for any reason or revision for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2 to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up was complete for 146 implants. The eight year survival for the endpoint radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving implants were considered radiologically loose but were asymptomatic. The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent lines (n = 20, 13.9%) in the 144 surviving cups indicates a need for continued observation in the second decade of follow-up in order to observe their influence on long-term survival.

The clinical and radiological data resulting in a ten-year survival rate > 90% in young patients support the use of the Contemporary acetabular component in this specific patient group.

Cite this article: Bone Joint J 2013;95-B:1617–25.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 307 - 314
1 Mar 2011
Matthies A Underwood R Cann P Ilo K Nawaz Z Skinner J Hart AJ

This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR).

There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared.

Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR.

Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur.

These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 881 - 886
1 Jul 2013
Bjorgul K Novicoff WN Andersen ST Ahlund OR Bunes A Wiig M Brevig K

A total of 397 hips were randomised to receive Metasul metal-on-metal (MoM), metal-on-conventional polyethylene (MoP) or ceramic-on-polyethylene (CoP) bearings using a cemented triple-tapered polished femoral component (MS-30). There were 129 MoM hips in 123 patients (39 male and 84 female, mean age 63.3 years (40.7 to 72.9)), 137 MoP hips in 127 patients (39 male and 88 female, mean age 62.8 years (24.5 to 72.7)) and 131 CoP hips in 124 patients (51 male and 73 female, mean age 63.9 years (30.6 to 73.8)). All acetabular components were cemented Weber polyethylene components with the appropriate inlay for the MoM articulation. Clinical evaluation was undertaken using the Harris hip score (HHS) and radiological assessments were made at two, five and seven years. The HHS and radiological analysis were available for 341 hips after seven years. The MoM group had the lowest mean HHS (p = 0.124), a higher rate of revision (p < 0.001) and a higher incidence of radiolucent lines in unrevised hips (p < 0.001). In all, 12 revisions had been performed in 12 patients: eight in the MoM group (four for infection, four for aseptic loosening, three in the MoP group (one each of infection, dislocation and pain) and one in the CoP group (infection).

Our findings reveal no advantage to the MoM bearing and identified a higher revision rate and a greater incidence of radiolucent lines than with the other articulations. We recommend that patients with a 28 mm Metasul MoM bearing be followed carefully.

Cite this article: Bone Joint J 2013;95-B:881–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 291 - 295
1 Mar 2007
Langdown AJ Pickard RJ Hobbs CM Clarke HJ Dalton DJN Grover ML

We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 412 - 418
1 Mar 2012
Judge A Arden NK Kiran A Price A Javaid MK Beard D Murray D Field RE

We obtained information from the Elective Orthopaedic Centre on 1523 patients with baseline and six-month Oxford hip scores (OHS) after undergoing primary hip replacement (THR) and 1784 patients with Oxford knee scores (OKS) for primary knee replacement (TKR) who completed a six-month satisfaction questionnaire.

Receiver operating characteristic curves identified an absolute change in OHS of 14 points or more as the point that discriminates best between patients’ satisfaction levels and an 11-point change for the OKS. Satisfaction is highest (97.6%) in patients with an absolute change in OHS of 14 points or more, compared with lower levels of satisfaction (81.8%) below this threshold. Similarly, an 11-point absolute change in OKS was associated with 95.4% satisfaction compared with 76.5% below this threshold. For the six-month OHS a score of 35 points or more distinguished patients with the highest satisfaction level, and for the six-month OKS 30 points or more identified the highest level of satisfaction. The thresholds varied according to patients’ pre-operative score, where those with severe pre-operative pain/function required a lower six-month score to achieve the highest levels of satisfaction.

Our data suggest that the choice of a six-month follow-up to assess patient-reported outcomes of THR/TKR is acceptable. The thresholds help to differentiate between patients with different levels of satisfaction, but external validation will be required prior to general implementation in clinical practice.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 12 | Pages 1642 - 1647
1 Dec 2010
deSouza RM Parsons NR Oni T Dalton P Costa M Krikler S

We report serum metal ion level data in patients with unilateral and bilateral hip resurfacing over a ten-year period. In these patients there is an increase in both cobalt and chromium levels above the accepted reference ranges during the first 18 months after operation. Metal ion levels remain elevated, but decline slowly for up to five years. However, the levels then appear to start rising again in some patients up to the ten-year mark. There was no significant difference in cobalt or chromium levels between men and women. These findings appear to differ from much of the current literature.

The clinical significance of a raised metal ion level remains under investigation.


Bone & Joint 360
Vol. 2, Issue 3 | Pages 27 - 29
1 Jun 2013

The June 2013 Shoulder & Elbow Roundup360 looks at: whether suture anchors are still the gold standard; infection and revision elbow arthroplasty; the variable success of elbow replacements; sliding knots; neurologic cuff pain and the suprascapular nerve; lies, damn lies and statistics; osteoarthritis; and one- or two-stage treatment for the infected shoulder revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 571 - 571
1 May 2011
Faux JC

This year is the centenary of the birth of John Charnley. Some of his legacy to Orthopaedic Surgery is described in this editorial.


Bone & Joint 360
Vol. 2, Issue 3 | Pages 35 - 38
1 Jun 2013

The June 2013 Children’s orthopaedics Roundup360 looks at: whether reaching a diagnosis is more difficult than previously thought; adolescent and paediatric DDH; the A-frame orthosis and Legg-Calvé-Perthes’ disease; failure of hip surgery in patients with cerebral palsy; adolescent rotator cuff injuries; paediatric peripheral nerve injuries; predicting residual deformity following Ponseti treatment; and the Dunn procedure.


Bone & Joint Research
Vol. 2, Issue 6 | Pages 96 - 101
1 Jun 2013
Harvie P Whitwell D

Objectives

Guidelines for the management of patients with metastatic bone disease (MBD) have been available to the orthopaedic community for more than a decade, with little improvement in service provision to this increasingly large patient group. Improvements in adjuvant and neo-adjuvant treatments have increased both the number and overall survival of patients living with MBD. As a consequence the incidence of complications of MBD presenting to surgeons has increased and is set to increase further. The British Orthopaedic Oncology Society (BOOS) are to publish more revised detailed guidelines on what represents ‘best practice’ in managing patients with MBD. This article is designed to coincide with and publicise new BOOS guidelines and once again champion the cause of patients with MBD.

Methods

A series of short cases highlight common errors frequently being made in managing patients with MBD despite the availability of guidelines.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 855 - 858
1 Jul 2009
Wroblewski BM Siney PD Fleming PA

The design of the Charnley total hip replacement follows the principle of low frictional torque. It is based on the largest possible difference between the radius of the femoral head and that of the outer aspect of the acetabular component. The aim is to protect the bone-cement interface by movement taking place at the smaller radius, the articulation. This is achieved in clinical practice by a 22.225 mm diameter head articulating with a 40 mm or 43 mm diameter acetabular component of ultra-high molecular weight polyethylene.

We compared the incidence of aseptic loosening of acetabular components with an outer diameter of 40 mm and 43 mm at comparable depths of penetration with a mean follow-up of 17 years (1 to 40).

In cases with no measurable wear none of the acetabular components were loose. With increasing acetabular penetration there was an increased incidence of aseptic loosening which reflected the difference in the external radii, with 1.5% at 1 mm, 8.8% at 2 mm, 9.7% at 3 mm and 9.6% at 4 mm of penetration in favour of the larger 43 mm acetabular component.

Our findings support the Charnley principle of low frictional torque. The level of the benefit is in keeping with the predicted values.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1562 - 1569
1 Dec 2008
Mäkelä K Eskelinen A Pulkkinen P Paavolainen P Remes V

We have analysed from the Finnish Arthroplasty Register the long-term survivorship of the 12 most commonly-used cemented implants between 1980 and 2005 in patients aged 55 years or older with osteoarthritis.

Only two designs of femoral component, the Exeter Universal and the Müller Straight femoral component had a survivorship of over 95% at ten years with revision for aseptic loosening as the endpoint. At 15 years of the femoral and acetabular component combinations, only the Exeter Universal/Exeter All-poly implant had a survival rate of over 90% with revision for aseptic loosening as the endpoint. In the subgroup of patients aged between 55 and 64 years, survivorship overall was less than 90% at ten years.

The variation in the long-term rates of survival of different cemented hip implants was considerable in patients aged 55 years or older. In those aged between 55 and 64 years, none of the cemented prostheses studied yielded excellent long-term survival rates (≥ 90% at 15 years).


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 11 - 15
1 Jan 2008
Jaiswal PK Jagiello J David LA Blunn G Carrington RWJ Skinner JA Cannon SR Briggs TWR

We have managed 27 patients (16 women and 11 men) with a mean age of 68.4 years (50 to 84), with failed total hip replacement and severe proximal femoral bone loss by revision using a distal fix/proximal wrap prosthesis. The mean follow-up was for 55.3 months (25 to 126). The mean number of previous operations was 2.2 (1 to 4). The mean Oxford hip score decreased from 46.2 (38 to 60) to 28.5 (17 to 42) (paired t-test, p < 0.001) and the mean Harris Hip score increased from 30.4 (3 to 57.7) to 71.7 (44 to 99.7) (paired t-test, p < 0.001). There were two dislocations, and in three patients we failed to eradicate previous infection. None required revision of the femoral stem.

This technique allows instant distal fixation while promoting biological integration and restoration of bone stock. In the short term, the functional outcome is encouraging and the complication rates acceptable in this difficult group of patients.