Femoral stem version has a major influence on
impingement and early post-operative stability after total hip arthroplasty
(THA). The main objective of this study was to evaluate the validity
of a novel radiological method for measuring stem version. Anteroposterior
(AP) radiographs and three-dimensional CT scans were obtained for
115 patients (female/male 63/72, mean age 62.5 years (50 to 75))
who had undergone minimally invasive, cementless THA. Stem version was
calculated from the AP hip radiograph by rotation-based change in
the projected prosthetic neck–shaft (NSA*) angle using the mathematical
formula ST = arcos [tan (NSA*) / tan (135)]. We used two independent
observers who repeated the analysis after a six-week interval. Radiological
measurements were compared with 3D-CT measurements by an independent,
blinded external institute. We found a mean difference of 1.2° ( We found that femoral tilt was associated with the mean radiological
measurement error (r = 0.22, p = 0.02). The projected neck–shaft angle is a reliable method for measuring
stem version on AP radiographs of the hip after a THA. However,
a highly standardised radiological technique is required for its
precise measurement. Cite this article:
High-intensity narrow-spectrum (HINS) light is
a novel violet-blue light inactivation technology which kills bacteria through
a photodynamic process, and has been shown to have bactericidal
activity against a wide range of species. Specimens from patients
with infected hip and knee arthroplasties were collected over a
one-year period (1 May 2009 to 30 April 2010). A range of these
microbial isolates were tested for sensitivity to HINS-light. During
testing, suspensions of the pathogens were exposed to increasing
doses of HINS-light (of 123mW/cm2 irradiance). Non-light exposed
control samples were also used. The samples were then plated onto
agar plates and incubated at 37°C for 24 hours before enumeration.
Complete inactivation (greater than 4-log10 reduction)
was achieved for all of the isolates. The typical inactivation curve
showed a slow initial reaction followed by a rapid period of inactivation.
The doses of HINS-light required ranged between 118 and 2214 J/cm2.
Gram-positive bacteria were generally found to be more susceptible
than Gram-negative. As HINS-light uses visible wavelengths, it can be safely used
in the presence of patients and staff. This unique feature could
lead to its possible use in the prevention of infection during surgery
and post-operative dressing changes. Cite this article:
Drug therapy forms an integral part of the management
of many orthopaedic conditions. However, many medicines can produce
serious adverse reactions if prescribed inappropriately, either
alone or in combination with other drugs. Often these hazards are
not appreciated. In response to this, the European Union recently
issued legislation regarding safety measures which member states
must adopt to minimise the risk of errors of medication. In March 2014 the Medicines and Healthcare products Regulatory
Agency and NHS England released a Patient Safety Alert initiative
focussed on errors of medication. There have been similar initiatives
in the United States under the auspices of The National Coordinating
Council for Medication Error and The Joint Commission on the Accreditation
of Healthcare Organizations. These initiatives have highlighted
the importance of informing and educating clinicians. Here, we discuss common drug interactions and contra-indications
in orthopaedic practice. This is germane to safe and effective clinical
care. Cite this article:
Pathological assessment of periprosthetic tissues is important, not only for diagnosis, but also for understanding the pathobiology of implant failure. The host response to wear particle deposition in periprosthetic tissues is characterised by cell and tissue injury, and a reparative and inflammatory response in which there is an innate and adaptive immune response to the material components of implant wear. Physical and chemical characteristics of implant wear influence the nature of the response in periprosthetic tissues and account for the development of particular complications that lead to implant failure, such as osteolysis which leads to aseptic loosening, and soft-tissue necrosis/inflammation, which can result in pseudotumour formation. The innate response involves phagocytosis of implant-derived wear particles by macrophages; this is determined by pattern recognition receptors and results in expression of cytokines, chemokines and growth factors promoting inflammation and osteoclastogenesis; phagocytosed particles can also be cytotoxic and cause cell and tissue necrosis. The adaptive immune response to wear debris is characterised by the presence of lymphoid cells and most likely occurs as a result of a cell-mediated hypersensitivity reaction to cell and tissue components altered by interaction with the material components of particulate wear, particularly metal ions released from cobalt-chrome wear particles. Cite this article: Professor N. A. Athanasou. The pathobiology and pathology of aseptic implant failure.
Direct anterior approaches to the hip have gained
popularity as a minimally invasive method when performing primary
total hip replacement (THR). A retrospective review of a single
institution joint registry was performed in order to compare patient
outcomes after THR using the Anterior Supine Intermuscular (ASI)
approach Cite this article
Advances in the treatment of periprosthetic joint
infections of the hip have once more pushed prosthesis preserving techniques
into the limelight. At the same time, the common infecting organisms
are evolving to become more resistant to conventional antimicrobial
agents. Whilst the epidemiology of resistant staphylococci is changing,
a number of recent reports have advocated the use of irrigation
and debridement and one-stage revision for the treatment of periprosthetic
joint infections due to resistant organisms. This review presents
the available evidence for the treatment of periprosthetic joint
infections of the hip, concentrating in particular on methicillin
resistant staphylococci. Cite this article:
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on failure of TKR contain patients
with unexplained pain who not infrequently have unmet expectations.
Surgeons must work to achieve realistic patient expectations pre-operatively,
and therefore, improve patient satisfaction post-operatively. Cite this article:
The aims of this retrospective study were to
compare the mid-term outcomes following revision total knee replacement
(TKR) in 76 patients (81 knees) <
55 years of age with those
of a matched group of primary TKRs based on age, BMI, gender and
comorbid conditions. We report the activity levels, functional scores,
rates of revision and complications. Compared with patients undergoing
primary TKR, those undergoing revision TKR had less improvement
in the mean Knee Society function scores (8.14 (–55 to +60) Young patients undergoing revision TKR should be counselled that
they can expect somewhat less improvement and a higher risk of complications
than occur after primary TKR. Cite this article:
Whether patients with asymptomatic bacteriuria
should be investigated and treated before elective hip and knee replacement
is controversial, although it is a widespread practice. We conducted
a prospective observational cohort study with urine analyses before
surgery and three days post-operatively. Patients with symptomatic
urinary infections or an indwelling catheter were excluded. Post-discharge
surveillance included questionnaires to patients and general practitioners
at three months. Among 510 patients (309 women and 201 men), with
a median age of 69 years (16 to 97) undergoing lower limb joint
replacements (290 hips and 220 knees), 182 (36%) had pre-operative asymptomatic
bacteriuria, mostly due to We conclude that testing and treating asymptomatic urinary tract
colonisation before joint replacement is unnecessary. Cite this article:
A 66-year-old man presented with an infected resurfacing hip replacement in the immediate post-operative period following an uneventful transurethral resection of the prostate. Prophylactic antibiotics had been administered on induction of anaesthesia. The hip prosthesis had been inserted seven years previously and had been hitherto asymptomatic. The hip was washed out and microbiological examination identified Despite current recommendations, clinicians undertaking invasive procedures which can lead to bacteraemia in patients with prosthetic joint replacements should be aware of the risk of haematogenous seeding in such prostheses, which although low, has potentially disastrous consequences.
A 70-year-old man with an uncemented metal-on-polyethylene
total hip prosthesis underwent revision arthroplasty 33 months later
because of pain, swelling and recurrent dislocation. There appeared
to be corrosion and metal release from the prosthetic head, resulting
in pseudotumour formation and severe local soft-tissue destruction.
The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron
taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven.
Over recent years hip arthroscopic surgery has
evolved into one of the most rapidly expanding fields in orthopaedic surgery.
Complications are largely transient and incidences between 0.5%
and 6.4% have been reported. However, major complications can and
do occur. This article analyses the reported complications and makes recommendations
based on the literature review and personal experience on how to
minimise them.
Arthroscopy of the native hip is an established diagnostic and therapeutic procedure. Its application in the symptomatic replaced hip is still being explored. We describe the use of arthroscopy of the hip in 24 symptomatic patients following total hip replacement, resurfacing arthroplasty of the hip and partial resurfacing (study group), and compared it with arthroscopy of the native hip in 24 patients (control group). A diagnosis was made or confirmed at arthroscopy in 23 of the study group and a therapeutic arthroscopic intervention resulted in relief of symptoms in ten of these. In a further seven patients it led to revision hip replacement. In contrast, arthroscopy in the control group was diagnostic in all 24 patients and the resulting arthroscopic therapeutic intervention provided symptomatic relief in 21. The mean operative time in the study group (59.7 minutes (35 to 93)) was less than in the control group (71 minutes (40 to 100), p = 0.04) but the arthroscopic approach was more difficult in the arthroplasty group. We suggest that arthroscopy has a role in the management of patients with a symptomatic arthroplasty when other investigations have failed to provide a diagnosis.
The credibility and creativity of an author may be gauged by the number of scientific papers he or she has published, as well as the frequency of citations of a particular paper reflecting the impact of the data on the area of practice. The object of this study was to identify and analyse the qualities of the top 100 cited papers in orthopaedic surgery. The database of the Science Citation Index of the Institute for Scientific Information (1945 to 2008) was used. A total of 1490 papers were cited more than 100 times, with the top 100 being subjected to further analysis. The majority originated in the United States, followed by the United Kingdom. The top 100 papers were published in seven specific orthopaedic journals. Analysis of the most-cited orthopaedic papers allows us a unique insight into the qualitites, characteristics and clinical innovations required for a paper to attain ‘classic’ status.
Between 1996 and 2008, nine patients with severe post-traumatic arthritis underwent revision of a failed interposition arthroplasty of the elbow with a further interposition procedure using an allograft of tendo Achillis at a mean of 5.6 years (0.7 to 13.1) after the initial procedure. There were eight men and one woman with a mean age of 47 years (36 to 56). The mean follow-up was 4.7 years (2 to 8). The mean Mayo Elbow Performance score improved from 49 (15 to 65) pre-operatively to 73 (55 to 95) (p = 0.04). The mean Disability of the Arm, Shoulder and Hand score was 26 (7 to 42). One patient was unavailable for clinical follow-up and one underwent total elbow replacement three months post-operatively. Of the remaining patients, one had an excellent, two had good, three fair and one a poor result. Subjectively, five of the nine patients were satisfied. Four continued manual labour. Revision interposition arthroplasty is an option for young, active patients with severe post-traumatic arthritis who require both mobility and durability of the elbow.
Antibiotic prophylaxis is routinely administered during joint replacement surgery and may predispose patients to
We report the outcome of total hip replacement in 29 failed metal-on-metal resurfacing hip replacements in which the primary surgery was performed between August 1995 and February 2005. The mean length of follow-up was five years (1.7 to 11.7). Of the 29 hip resurfacings, 19 acetabular components and all the femoral components were revised (28 uncemented stems and one cemented stem). There were no deaths and none of the patients was lost to follow-up. None of the hips underwent any further revision. The results of the revision resurfacing group were compared with those of a control group of age-matched patients. In the latter group there were 236 primary total hip replacements and 523 resurfacings performed during the same period by the same surgeons. The outcome of the revision resurfacing group was comparable with that of the stemmed primary hip replacement group but was less good than that of the primary hip resurfacing group. Long-term follow-up is advocated to monitor the outcome of these cases.
Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p <
0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement.
We have reviewed 54 patients who had undergone 61 total hip replacements using bulk femoral autografts to augment a congenitally dysplastic acetabulum. There were 52 women and two men with a mean age of 42.4 years (29 to 76) at the time of the index operation. A variety of different prostheses was used: 28 (45.9%) were cemented and 33 (54.1%) uncemented. The graft technique remained unchanged throughout the series. Follow-up was at a mean of 8.3 years (3 to 20). The Hospital for Special Surgery hip score improved from a mean of 10.7 (4 to 18) pre-operatively to a mean of 35 (28 to 38) at follow-up. The position of the acetabular component was anatomical in 37 hips (60.7%), displaced less than 1 cm in 20 (32.7%) and displaced more than 1 cm in four (6.6%). Its cover was between 50% and 75% in 34 hips (55.7%) and less than 50% in 25 (41%). In two cases (3.3%), it was more than 75%. There was no graft resorption in 36 hips (59%), mild resorption in 21 (34%) and severe resorption in four (6%). Six hips (9.8%) were revised for aseptic loosening. The overall rate of loosening and revision was 14.8%. Overall survival at 8.3 years was 93.4%. The only significant factor which predicted failure was the implantation of the acetabular component more than 1 cm from the anatomical centre of rotation of the hip.
Evaluation of patients with painful total knee replacement requires a thorough clinical examination and relevant investigations in order to reach a diagnosis. Awareness of the common and uncommon problems leading to painful total knee replacement is useful in the diagnostic approach. This review article aims to act as a guide to the evaluation of patients with painful total knee replacement.