The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 459 - 465
1 Apr 2012
Nikolaou VS Edwards MR Bogoch E Schemitsch EH Waddell JP

The ideal bearing surface for young patients undergoing total hip replacement (THR) remains controversial. We report the five-year results of a randomised controlled trial comparing the clinical and radiological outcomes of 102 THRs in 91 patients who were <  65 years of age. These patients were randomised to receive a cobalt–chrome on ultra-high-molecular-weight polyethylene, cobalt–chrome on highly cross-linked polyethylene, or a ceramic-on-ceramic bearing. In all, 97 hip replacements in 87 patients were available for review at five years. Two hips had been revised, one for infection and one for peri-prosthetic fracture.

At the final follow-up there were no significant differences between the groups for the mean Western Ontario and McMaster Universities osteoarthritis index (pain, p = 0.543; function, p = 0.10; stiffness, p = 0.99), Short Form-12 (physical component, p = 0.878; mental component, p = 0.818) or Harris hip scores (p = 0.22). Radiological outcomes revealed no significant wear in the ceramic group.

Comparison of standard and highly cross-linked polyethylene, however, revealed an almost threefold difference in the mean annual linear wear rates (0.151 mm/year versus 0.059 mm/year, respectively) (p < 0.001).


Bone & Joint Research
Vol. 5, Issue 9 | Pages 379 - 386
1 Sep 2016
Pahuta M Smolders JM van Susante JL Peck J Kim PR Beaule PE

Objectives

Alarm over the reported high failure rates for metal-on-metal (MoM) hip implants as well as their potential for locally aggressive Adverse Reactions to Metal Debris (ARMDs) has prompted government agencies, internationally, to recommend the monitoring of patients with MoM hip implants. Some have advised that a blood ion level >7 µg/L indicates potential for ARMDs. We report a systematic review and meta-analysis of the performance of metal ion testing for ARMDs.

Methods

We searched MEDLINE and EMBASE to identify articles from which it was possible to reconstruct a 2 × 2 table. Two readers independently reviewed all articles and extracted data using explicit criteria. We computed a summary receiver operating curve using a Bayesian random-effects hierarchical model.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 56 - 63
1 Apr 2012
Langton DJ Sidaginamale R Lord JK Nargol AVF Joyce TJ

Objectives

An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer.

Methods

Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular bearing surfaces and also the surfaces of the taper junctions. The relationship between taper wear rates and a number of variables, including bearing diameter and orientation of the acetabular component, was investigated.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1475 - 1479
1 Nov 2005
Goosen JHM Verheyen CCPM Tulp NJA

We investigated the rate of polyethylene wear of a cementless acetabular component at different periods of follow-up in order to test the hypothesis than an irrecoverable deformation process (creep) was followed by an initially low, but gradually increasing wear rate. We studied prospectively 93 uncemented total hip arthroplasties in 83 patients (mean age 50 years (22 to 63)) with a mean follow-up of 8.2 years (3 to 12). We measured the penetration of the femoral head from radiographs taken immediately after surgery at three, six and nine years, or at the latest follow-up.

The median wear rate was 0.17 mm per year in the first three years, a finding which we considered to be caused by creep. Thereafter, the rate of wear declined to 0.07 mm per year (four- to six-year period) and then increased to 0.17 mm per year (seven to nine years) and 0.27 mm per year (more than nine years), which we considered to be a reflection of genuine polyethylene wear. After the nine-year follow-up the wear rates were higher in patients with marked osteolysis. We found no relationship between the inclination angle of the acetabular component or femoral head orientation and the rate of wear. No acetabular component required revision.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 162 - 168
1 May 2016
Athanasou NA

Pathological assessment of periprosthetic tissues is important, not only for diagnosis, but also for understanding the pathobiology of implant failure. The host response to wear particle deposition in periprosthetic tissues is characterised by cell and tissue injury, and a reparative and inflammatory response in which there is an innate and adaptive immune response to the material components of implant wear. Physical and chemical characteristics of implant wear influence the nature of the response in periprosthetic tissues and account for the development of particular complications that lead to implant failure, such as osteolysis which leads to aseptic loosening, and soft-tissue necrosis/inflammation, which can result in pseudotumour formation. The innate response involves phagocytosis of implant-derived wear particles by macrophages; this is determined by pattern recognition receptors and results in expression of cytokines, chemokines and growth factors promoting inflammation and osteoclastogenesis; phagocytosed particles can also be cytotoxic and cause cell and tissue necrosis. The adaptive immune response to wear debris is characterised by the presence of lymphoid cells and most likely occurs as a result of a cell-mediated hypersensitivity reaction to cell and tissue components altered by interaction with the material components of particulate wear, particularly metal ions released from cobalt-chrome wear particles.

Cite this article: Professor N. A. Athanasou. The pathobiology and pathology of aseptic implant failure. Bone Joint Res 2016;5:162–168. DOI: 10.1302/2046-3758.55.BJR-2016-0086.


Bone & Joint 360
Vol. 4, Issue 6 | Pages 8 - 10
1 Dec 2015

The December 2015 Hip & Pelvis Roundup360 looks at: Vitamin E infusion helpful in polyethylene; Hip replacement in fracture and arthritis; Non-surgical treatment for arthritis; Cost and approach in hip surgery; Who does well in FAI surgery?; AAOS Thromboembolism guidelines; Thromboprophylaxis and periprosthetic joint infection; Fluid collections not limited to metal-on-metal THR


Bone & Joint 360
Vol. 4, Issue 6 | Pages 10 - 13
1 Dec 2015

The December 2015 Knee Roundup360 looks at: Albumin and complications in knee arthroplasty; Tantalum: a knee fixation for all seasons?; Dynamic knee alignment; Tibial component design in UKA; Managing the tidal wave of revision knee arthroplasty; Scoring pain in TKR; Does anyone have a ‘normal’ tibial slope?; XLPE in TKR? A five-year clinical study; Spacers and infected revision arthroplasties; Dialysis and arthroplasty


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 58 - 63
1 Jan 2009
Hooper G Rothwell A Frampton C

We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time.

At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p < 0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees.

There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p < 0.001).

The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis.


Bone & Joint 360
Vol. 4, Issue 3 | Pages 15 - 16
1 Jun 2015

The June 2015 Foot & Ankle Roundup360 looks at: Syndesmosis and outcomes in ankle fracture; Ankle arthrodesis or arthroplasty: a complications-based analysis; Crosslinked polyethylene and ankle arthroplasty; Reducing screw removal in calcaneal osteotomies; Revisiting infection control policies; Chevron osteotomy: proximal or distal?; Ankle distraction for osteoarthritis


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1134 - 1141
1 Sep 2009
Isaac GH Brockett C Breckon A van der Jagt D Williams S Hardaker C Fisher J Schepers A

This study reports on ceramic-on-metal (CoM) bearings in total hip replacement. Whole blood metal ion levels were measured. The median increase in chromium and cobalt at 12 months was 0.08 μg/1 and 0.22 μg/1, respectively, in CoM bearings. Comparable values for metal-on-metal (MoM) were 0.48 μg/1 and 0.32 μg/1. The chromium levels were significantly lower in CoM than in MoM bearings (p = 0.02). The cobalt levels were lower, but the difference was not significant. Examination of two explanted ceramic heads revealed areas of thin metal transfer. CoM bearings (one explanted head and acetabular component, one explanted head and new acetabular component, and three new heads and acetabular components) were tested in a hip joint simulator. The explanted head and acetabular component had higher bedding-in. However, after one million cycles all the wear rates were the same and an order of magnitude less than that reported for MoM bearings. There were four outliers in each clinical group, primarily related to component malposition.


Bone & Joint 360
Vol. 4, Issue 2 | Pages 10 - 12
1 Apr 2015

The April 2015 Hip & Pelvis Roundup360 looks at: Goal-directed fluid therapy in hip fracture; Liberal blood transfusion no benefit in the longer term; Repeated measures: increased accuracy or compounded errors?; Peri-acetabular osteotomy safer than perhaps thought?; Obesity and peri-acetabular osteotomy: poor bedfellows; Stress fracture in peri-acetabular osteotomy; Infection and tantalum implants; Highly crosslinked polyethylene really does work


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 883 - 889
1 Jul 2015
Jassim SS Patel S Wardle N Tahmassebi J Middleton R Shardlow DL Stephen A Hutchinson J Haddad FS

Oxidised zirconium (OxZi) has been developed as an alternative bearing surface for femoral heads in total hip arthroplasty (THA). This study has investigated polyethylene wear, functional outcomes and complications, comparing OxZi and cobalt–chrome (CoCr) as part of a three-arm, multicentre randomised controlled trial. Patients undergoing THA from four institutions were prospectively randomised into three groups. Group A received a CoCr femoral head and highly cross-linked polyethylene (XLPE) liner; Group B received an OxZi femoral head and XLPE liner; Group C received an OxZi femoral head and ultra-high molecular weight polyethylene (UHMWPE) liner. At five years, 368 patients had no statistically significant differences in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western Ontario and McMaster Universities Osteoarthritis Index (p = 0.847), pain scores (p = 0.458) or complications. The mean rate of linear wear was 0.028 mm/year (standard deviation (sd) 0.010) for Group A, 0.023 mm/year (sd 0.010) for Group B, and 0.09 mm/year (sd 0.045) for Group C. Penetration was significantly higher in the UHMWPE liner group compared with both XLPE liner groups (p < 0.001) but no significant difference was noted between CoCr and OxZi when articulating with XLPE (p = 0.153). In this, the largest randomised study of this bearing surface, it appears that using a XLPE acetabular liner is more important in reducing THA component wear than the choice of femoral head bearing, at mid-term follow-up. There is a non-significant trend towards lower wear, coupling OxZi rather than CoCr with XLPE but long-term analysis is required to see if this observation changes with time and becomes significant.

Cite this article: Bone Joint J 2015;97-B:883–9.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 23 - 26
1 Nov 2014
Cooper HJ Della Valle CJ

Dislocation remains among the most common complications of, and reasons for, revision of both primary and revision total hip replacements (THR). Hence, there is great interest in maximising stability to prevent this complication. Head size has been recognised to have a strong influence on the risk of dislocation post-operatively. As femoral head size increases, stability is augmented, secondary to an increase in impingement-free range of movement. Larger head sizes also greatly increase the ‘jump distance’ required for the head to dislocate in an appropriately positioned cup. Level-one studies support the use of larger diameter heads as they decrease the risk of dislocation following primary and revision THR. Highly cross-linked polyethylene has allowed us to increase femoral head size, without a marked increase in wear. However, the thin polyethylene liners necessary to accommodate larger heads may increase the risk of liner fracture and larger heads have also been implicated in causing soft-tissue impingement resulting in groin pain. Larger diameter heads also impart larger forces on the femoral trunnion, which may contribute to corrosion, metal release, and adverse local tissue reactions. Alternative large bearings including large ceramic heads and dual mobility bearings may mitigate some of these risks, and several of these devices have been used with clinical success.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):23–6.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 911 - 916
1 Jul 2015
Del Balso C Teeter MG Tan SC Lanting BA Howard JL

Tribocorrosion at the head–neck taper interface – so-called ‘taperosis’ – may be a source of metal ions and particulate debris in metal-on-polyethylene total hip arthroplasty (THA).

We examined the effect of femoral head length on fretting and corrosion in retrieved head–neck tapers in vivo for a minimum of two years (mean 8.7 years; 2.6 to 15.9). A total of 56 femoral heads ranging from 28 mm to 3 mm to 28 mm + 8 mm, and 17 femoral stems featuring a single taper design were included in the study. Fretting and corrosion were scored in three horizontally oriented concentric zones of each taper by stereomicroscopy.

Head length was observed to affect fretting (p = 0.03), with 28 mm + 8 mm femoral heads showing greater total fretting scores than all other head lengths. The central zone of the femoral head bore taper was subject to increased fretting damage (p = 0.01), regardless of head length or stem offset. High-offset femoral stems were associated with greater total fretting of the bore taper (p = 0.04).

Increased fretting damage is seen with longer head lengths and high-offset femoral stems, and occurs within a central concentric zone of the femoral head bore taper. Further investigation is required to determine the effect of increased head size, and variations in head–neck taper design.

Cite this article: Bone Joint J 2015; 97-B:911–16.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1183 - 1191
1 Sep 2015
Briggs TWR Hanna SA Kayani B Tai S Pollock RC Cannon SR Blunn GW Carrington RWJ

The long term biological effects of wear products following total hip arthroplasty (THA) are unclear. However, the indications for THA are expanding, with increasingly younger patients undergoing the procedure.

This prospective, randomised study compared two groups of patients undergoing THA after being randomised to receive one of two different bearing surfaces: metal-on-polyethylene (MoP) n = 22 and metal-on-metal (MoM) n = 23. We investigated the relationship between three variables: bearing surface (MoP vs MoM), whole blood levels of chromium (Cr) and cobalt (Co) and chromosomal aberrations in peripheral lymphocyte pre-operatively and at one, two and five years post-surgery.

Our results demonstrated significantly higher mean cobalt and chromium (Co and Cr) blood levels in the MoM group at all follow-up points following surgery (p < 0.01), but there were no significant differences in the chromosomal aberration indices between MoM and MoP at two or five years (two years: p = 0.56, p = 0.08, p = 0.91, p = 0.51 and five years: p = 0.086, p = 0.73, p = 0.06, p = 0.34) for translocations, breaks, loss and gain of chromosomes respectively. Regression analysis showed a strong linear relationship between Cr levels and the total chromosomal aberration indices in the MoM group (R2 = 0.90016), but this was not as strong for Co (R2 = 0.68991). In the MoP group, the analysis revealed a poor relationship between Cr levels and the total chromosomal aberration indices (R2 = 0.23908) but a slightly stronger relationship for Co (R2 = 0.64292). Across both groups, Spearman’s correlation detected no overall association between Co and Cr levels and each of the studied chromosomal aberrations. There remains no clear indication which THA bearing couple is the most biocompatible, especially in young active patients. While THA continues to be very successful at alleviating pain and restoring function, the long-term biological implications of the procedure still require further scrutiny.

Cite this article: Bone Joint J 2015;97-B:1183–91.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 949 - 954
1 Jul 2011
Bisseling P Zeilstra DJ Hol AM van Susante JLC

The purpose of this study was to evaluate whether concerns about the release of metal ions in metal-on-metal total hip replacements (THR) should be extended to patients with metal-bearing total disc replacements (TDR).

Cobalt and chromium levels in whole blood and serum were measured in ten patients with a single-level TDR after a mean follow-up of 34.5 months (13 to 61) using inductively-coupled plasma mass spectrometry. These metal ion levels were compared with pre-operative control levels in 81 patients and with metal ion levels 12 months after metal-on-metal THR (n = 21) and resurfacing hip replacement (n = 36). Flexion-extension radiographs were used to verify movement of the TDR.

Cobalt levels in whole blood and serum were significantly lower in the TDR group than in either the THR (p = 0.007) or the resurfacing group (p < 0.001). Both chromium levels were also significantly lower after TDR versus hip resurfacing (p < 0.001), whereas compared with THR this difference was only significant for serum levels (p = 0.008). All metal ion levels in the THR and resurfacing groups were significantly higher than in the control group (p < 0.001). In the TDR group only cobalt in whole blood appeared to be significantly higher (p < 0.001). The median range of movement of the TDR was 15.5° (10° to 22°).

These results suggest that there is minimal cause for concern about high metal ion concentrations after TDR, as the levels appear to be only moderately elevated. However, spinal surgeons using a metal-on-metal TDR should still be aware of concerns expressed in the hip replacement literature about toxicity from elevated metal ion levels, and inform their patients appropriately.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 608 - 615
1 May 2011
Bolland BJRF Culliford DJ Langton DJ Millington JPS Arden NK Latham JM

This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83).

Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem.

The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4).

This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 469 - 476
1 Apr 2010
Shimmin AJ Walter WL Esposito C

The survivorship of contemporary resurfacing arthroplasty of the hip using metal-on-metal bearings is better than that of first generation designs, but short-term failures still occur. The most common reasons for failure are fracture of the femoral neck, loosening of the component, osteonecrosis of the femoral head, reaction to metal debris and malpositioning of the component. In 2008 the Australian National Joint Registry reported an inverse relationship between the size of the head component and the risk of revision in resurfacing hip arthroplasty. Hips with a femoral component size of ≤ 44 mm have a fivefold increased risk of revision than those with femoral components of ≥ 55 mm irrespective of gender. We have reviewed the literature to explore this observation and to identify possible reasons including the design of the implant, loading of the femoral neck, the orientation of the component, the production of wear debris and the effects of metal ions, penetration of cement and vascularity of the femoral head. Our conclusion is that although multifactorial, the most important contributors to failure in resurfacing arthroplasty of the hip are likely to be the design and geometry of the component and the orientation of the acetabular component.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 63 - 66
1 Nov 2013
Rodriguez JA Cooper HJ

Large ceramic femoral heads offer several advantages that are potentially advantageous to patients undergoing both primary and revision total hip replacement. Many high-quality studies have demonstrated the benefit of large femoral heads in reducing post-operative instability. Ceramic femoral heads may also offer an advantage in reducing polyethylene wear that has been reported in vitro and is starting to become clinically apparent in mid-term clinical outcome studies. Additionally, the risk of taper corrosion at a ceramic femoral head–neck junction is clearly lower than when using a metal femoral head. With improvements in the material properties of both modern ceramic femoral heads and polyethylene acetabular liners that have reduced the risk of mechanical complications, large ceramic heads have gained popularity in recent years.

Cite this article: Bone Joint J 2013;95-B, Supple A:63–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 367 - 373
1 Mar 2010
Kendrick BJL Longino D Pandit H Svard U Gill HS Dodd CAF Murray DW Price AJ

The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (sd 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/− signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p < 0.001).

These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.