The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component. There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p <
0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the
Most hip fractures treated with modern internal
fixation techniques will heal. However, failures occasionally occur and
require revision procedures. Salvage strategies employed during
revision are based on whether the fixation failure occurs in the
femoral neck, or in the intertrochanteric region. Patient age and
remaining bone stock also influence decision making. For fractures
in young patients, efforts are generally focused on preserving the
native femoral head via osteotomies and repeat internal fixation.
For failures in older patients, some kind of hip replacement is
usually selected. Disuse osteopenia, deformity, bone loss, and stress-risers
from previous internal fixation devices all pose technical challenges
to successful reconstruction. Attention to detail is important in
order to minimise complications. In the majority of cases, good
outcomes have been reported for the various salvage strategies. Cite this article:
Autologous chondrocyte implantation (ACI) is used widely as a treatment for symptomatic chondral and osteochondral defects of the knee. Variations of the original periosteum-cover technique include the use of porcine-derived type I/type III collagen as a cover (ACI-C) and matrix-induced autologous chondrocyte implantation (MACI) using a collagen bilayer seeded with chondrocytes. We have performed a prospective, randomised comparison of ACI-C and MACI for the treatment of symptomatic chondral defects of the knee in 91 patients, of whom 44 received ACI-C and 47 MACI grafts. Both treatments resulted in improvement of the clinical score after one year. The mean modified Cincinnati knee score increased by 17.6 in the ACI-C group and 19.6 in the MACI group (p = 0.32). Arthroscopic assessments performed after one year showed a good to excellent International Cartilage Repair Society score in 79.2% of ACI-C and 66.6% of MACI grafts. Hyaline-like cartilage or hyaline-like cartilage with fibrocartilage was found in the biopsies of 43.9% of the ACI-C and 36.4% of the MACI grafts after one year. The rate of hypertrophy of the graft was 9% (4 of 44) in the ACI-C group and 6% (3 of 47) in the MACI group. The frequency of re-operation was 9% in each group. We conclude that the clinical, arthroscopic and histological outcomes are comparable for both ACI-C and MACI. While MACI is technically attractive, further long-term studies are required before the technique is widely adopted.
The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite coating in a series of 52 men (six bilateral) and 40 women (two bilateral), has been described previously. None required revision for aseptic loosening. We have extended the follow-up to 20 years with a minimum of 17 years. The mean age of the patients at total hip replacement was 58.9 years (19 to 84). Six patients were lost to follow-up, but were included up to their last clinical review. A total of 22 patients (22 hips) had died, all from causes unrelated to their surgery. There have been 43 re-operations for failure of the acetabular component. However, in 38 of these the stem was not revised since it remained stable and there was no associated osteolysis. Two of the revisions were for damage to the trunnion after fracture of a modular ceramic head, and in another two, removal of the femoral component was because of the preference of the surgeon. In all cases the femoral component was well fixed, but could be extracted at the time of acetabular revision. In one case both components were revised for deep infection. There has been one case of aseptic loosening of the stem which occurred at 14 years. This stem had migrated distally by 7.6 mm in ten years and 8.4 mm at the time of revision at which stage it was found to be rotationally loose. With hindsight this component had been undersized at implantation. The survivorship for the stem at 17 years with aseptic loosening as the endpoint was 98.6% (95% confidence interval 95.9 to 100) when 62 hips were at risk. All remaining stems had a satisfactory clinical and radiological outcome. The Freeman proximally hydroxyapatite-coated femoral component is therefore a dependable implant and its continued use can be recommended.
Mobile-bearing unicompartmental knee replacements
(UKRs) with a flat tibial plateau have not performed well in the
lateral compartment, owing to a high dislocation rate. This led
to the development of the Domed Lateral Oxford UKR (Domed OUKR)
with a biconcave bearing. The aim of this study was to assess the
survival and clinical outcomes of the Domed OUKR in a large patient
cohort in the medium term. We prospectively evaluated 265 consecutive knees with isolated
disease of the lateral compartment and a mean age at surgery of
64 years (32 to 90). At a mean follow-up of four years ( The Domed Lateral OUKR gives good clinical outcomes, low re-operation
and revision rates and a low dislocation rate in patients with isolated
lateral compartmental disease, in the hands of the designer surgeons. Cite this article:
We describe the survivorship of the Medial Rotation total knee replacement (TKR) at ten years in 228 cemented primary replacements implanted between October 1994 and October 2006, with their clinical and radiological outcome. This implant has a highly congruent medial compartment, with the femoral component represented by a portion of a sphere which articulates with a matched concave surface on the medial side of the tibial insert. There were 78 men (17 bilateral TKRs) and 111 women (22 bilateral TKRs) with a mean age of 67.9 years (28 to 90). All the patients were assessed clinically and radiologically using the American Knee Society scoring systems. The mean follow-up was for six years (1 to 13) with only two patients lost to follow-up and 34 dying during the period of study, one of whom had required revision for infection. There were 11 revisions performed in total, three for aseptic loosening, six for infection, one for a periprosthetic fracture and one for a painful but well-fixed replacement performed at another centre. With revision for any cause as the endpoint, the survival at ten years was 94.5% (95% CI 85.1 to 100), and with aseptic loosening as the endpoint 98.4% (95% CI 93 to 100). The mean American Knee Society score improved from 47.6 (0 to 88) to 72.2 (26 to 100) and for function from 45.1 (0 to 100) to 93.1 (45 to 100). Radiological review failed to detect migration in any of the surviving knees. The clinical and radiological results of the Medial Rotation TKR are satisfactory at ten years. The increased congruence of the medial compartment has not led to an increased rate of loosening and continued use can be supported.
The strain on clinic and surgeon resources resulting
from a rise in demand for total knee replacement (TKR) requires reconsideration
of when and how often patients need to be seen for follow-up. Surgeons
will otherwise require increased paramedical staff or need to limit
the number of TKRs they undertake. We reviewed the outcome data
of 16 414 primary TKRs undertaken at our centre to determine the
time to re-operation for any reason and for specific failure mechanisms.
Peak risk years for failure were determined by comparing the conditional
probability of failure, the number of failures divided by the total
number of TKRs cases, for each year. The median times to failure
for the most common failure mechanisms were 4.9 years (interquartile
range (IQR) 1.7 to 10.7) for femoral and tibial loosening, 1.9 years
(IQR 0.8 to 3.9) for infection, 3.1 years (IQR 1.6 to 5.5) for tibial
collapse and 5.6 years (IQR 3.4 to 9.3) for instability. The median
time to failure for all revisions was 3.3 years (IQR 1.2 to 8.5),
with an overall revision rate of 1.7% (n = 282). Results from our
patient population suggest that patients be seen for follow-up at
six months, one year, three years, eight years, 12 years, and every
five years thereafter. Patients with higher pain in the early post-operative
period or high body mass index (≥ 41 kg/m2) should be
monitored more closely. Cite this article:
We present six patients with chronic dislocation of the elbow who were treated by primary semiconstrained total elbow arthroplasty. All were women with a mean age of 65 years (51 to 76), the mean interval between dislocation and surgery was 17 weeks (5 to 52) and the mean follow-up 58 months (24 to 123). The most dramatic improvement was in function. The mean American Shoulder and Elbow Surgeon score was 5.2 times better (p <
0.001) and the mean total range of movement increased from 33° to 121° (p <
0.001) after operation. Three patients developed wear of polyethylene. One required revision for a periprosthetic fracture, and another required a bushing exchange. Primary semiconstrained elbow arthroplasty provides significant, predictable functional improvement. Potential solutions for wear of polyethylene include a different operative technique or design of implant. Despite the high incidence of such wear, total elbow arthroplasty should be considered as a viable treatment option for chronic dislocation of the elbow in elderly patients.
The purpose of this study was to compare the
results of proximal and distal chevron osteotomy in patients with moderate
hallux valgus. We retrospectively reviewed 34 proximal chevron osteotomies without
lateral release (PCO group) and 33 distal chevron osteotomies (DCO
group) performed sequentially by a single surgeon. There were no
differences between the groups with regard to age, length of follow-up,
demographic or radiological parameters. The clinical results were
assessed using the American Orthopaedic Foot and Ankle Society (AOFAS)
scoring system and the radiological results were compared between
the groups. At a mean follow-up of 14.6 months (14 to 32) there were no significant
differences in the mean AOFAS scores between the DCO and PCO groups
(93.9 (82 to 100) and 91.8 (77 to 100), respectively; p = 0.176).
The mean hallux valgus angle, intermetatarsal angle and sesamoid
position were the same in both groups. The metatarsal declination
angle decreased significantly in the PCO group (p = 0.005) and the
mean shortening of the first metatarsal was significantly greater
in the DCO group (p <
0.001). We conclude that the clinical and radiological outcome after
a DCO is comparable with that after a PCO; longer follow-up would
be needed to assess the risk of avascular necrosis. Cite this article:
The Oxford unicompartmental knee replacement
(UKR) is an established treatment option in the management of symptomatic
end-stage medial compartmental osteoarthritis (MCOA), which works
well in the young and active patient. However, previous studies
have shown that it is reliable only in the presence of a functionally
intact anterior cruciate ligament (ACL). This review reports the
outcomes, at a mean of five years and a maximum of ten years, of 52
consecutive patients with a mean age of 51 years (36 to 57) who
underwent staged or simultaneous ACL reconstruction and Oxford UKR.
At the last follow-up (with one patient lost to follow-up), the
mean Oxford knee score was 41 ( In summary, ACL reconstruction and Oxford UKR gives good results
in patients with end-stage MCOA secondary to ACL deficiency.
This prospective study describes the complications and survival of the first 688 Phase 3 Oxford medial unicompartmental knee replacements implanted using a minimally-invasive technique by two surgeons and followed up independently. None was lost to follow-up. We had carried out 132 of the procedures more than five years ago. The clinical assessment of 101 of these which were available for review at five years is also presented. Nine of the 688 knees were revised: four for infection, three for dislocation of the bearing and two for unexplained pain. A further seven knees (1%) required other procedures: four had a manipulation under anaesthesia, two an arthroscopy and one a debridement for superficial infection. The survival rate at seven years was 97.3% (95% confidence interval 5.3). At five years, 96% of the patients had a good or excellent American Knee Society score, the mean Oxford knee score was 39 and the mean flexion was 133°. This study demonstrates that the minimally-invasive Oxford unicompartmental knee replacement is a reliable and effective procedure.
The use of pulsed electromagnetic fields (PEMF)
to stimulate bone growth has been recommended as an alternative to
the surgical treatment of ununited scaphoid fractures, but has never
been examined in acute fractures. We hypothesised that the use of
PEMF in acute scaphoid fractures would accelerate the time to union
by 30% in a randomised, double-blind, placebo-controlled, multicentre
trial. A total of 53 patients in three different medical centres
with a unilateral undisplaced acute scaphoid fracture were randomly
assigned to receive either treatment with PEMF (n = 24) or a placebo
(n = 29). The clinical and radiological outcomes were assessed at
four, six, nine, 12, 24 and 52 weeks. A log-rank analysis showed that neither time to clinical and
radiological union nor the functional outcome differed significantly
between the groups. The clinical assessment of union indicated that
at six weeks tenderness in the anatomic snuffbox (p = 0.03) as well
as tenderness on longitudinal compression of the scaphoid (p = 0.008) differed
significantly in favour of the placebo group. We conclude that stimulation of bone growth by PEMF has no additional
value in the conservative treatment of acute scaphoid fractures.
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
The pathogenesis of rotator cuff disease (RCD) is complex and
not fully understood. This systematic review set out to summarise
the histological and molecular changes that occur throughout the
spectrum of RCD. We conducted a systematic review of the scientific literature
with specific inclusion and exclusion criteria.Introduction
Methods
It is probable that both genetic and environmental
factors play some part in the aetiology of most cases of degenerative
hip disease. Geneticists have identified some single gene disorders
of the hip, but have had difficulty in identifying the genetics
of many of the common causes of degenerative hip disease. The heterogeneity
of the phenotypes studied is part of the problem. A detailed classification
of phenotypes is proposed. This study is based on careful documentation
of 2003 consecutive total hip replacements performed by a single
surgeon between 1972 and 2000. The concept that developmental problems
may initiate degenerative hip disease is supported. The influences
of gender, age and body mass index are outlined. Biomechanical explanations
for some of the radiological appearances encountered are suggested.
The body weight lever, which is larger than the abductor lever, causes
the abductor power to be more important than body weight. The possibility
that a deficiency in joint lubrication is a cause of degenerative
hip disease is discussed. Identifying the phenotypes may help geneticists
to identify genes responsible for degenerative hip disease, and
eventually lead to a definitive classification.
We evaluated the potential of a vastus lateralis muscle flap in controlling infection after resection arthroplasty of the hip. We retrospectively reviewed 119 patients with 120 chronic infections after resection arthroplasty treated with this procedure. The flap was fixed with Mitek anchors in the acetabular cavity. The mean duration of infection after resection before the muscle flap procedure was 6.5 months (2 to 13). The patients had previously undergone a mean of 4.9 operations (2 to 25). In all patients the infected cavity was the origin of the persistent infection. The mean follow-up was for 2.6 years (1.0 to 4.7). No patient had recurrent infection post-operatively and all had an improvement in the pain and better quality of life.
The purpose of this study was to evaluate whether concerns about the release of metal ions in metal-on-metal total hip replacements (THR) should be extended to patients with metal-bearing total disc replacements (TDR). Cobalt and chromium levels in whole blood and serum were measured in ten patients with a single-level TDR after a mean follow-up of 34.5 months (13 to 61) using inductively-coupled plasma mass spectrometry. These metal ion levels were compared with pre-operative control levels in 81 patients and with metal ion levels 12 months after metal-on-metal THR (n = 21) and resurfacing hip replacement (n = 36). Flexion-extension radiographs were used to verify movement of the TDR. Cobalt levels in whole blood and serum were significantly lower in the TDR group than in either the THR (p = 0.007) or the resurfacing group (p <
0.001). Both chromium levels were also significantly lower after TDR These results suggest that there is minimal cause for concern about high metal ion concentrations after TDR, as the levels appear to be only moderately elevated. However, spinal surgeons using a metal-on-metal TDR should still be aware of concerns expressed in the hip replacement literature about toxicity from elevated metal ion levels, and inform their patients appropriately.
Between 1996 and 2008, nine patients with severe post-traumatic arthritis underwent revision of a failed interposition arthroplasty of the elbow with a further interposition procedure using an allograft of tendo Achillis at a mean of 5.6 years (0.7 to 13.1) after the initial procedure. There were eight men and one woman with a mean age of 47 years (36 to 56). The mean follow-up was 4.7 years (2 to 8). The mean Mayo Elbow Performance score improved from 49 (15 to 65) pre-operatively to 73 (55 to 95) (p = 0.04). The mean Disability of the Arm, Shoulder and Hand score was 26 (7 to 42). One patient was unavailable for clinical follow-up and one underwent total elbow replacement three months post-operatively. Of the remaining patients, one had an excellent, two had good, three fair and one a poor result. Subjectively, five of the nine patients were satisfied. Four continued manual labour. Revision interposition arthroplasty is an option for young, active patients with severe post-traumatic arthritis who require both mobility and durability of the elbow.
There is little information available at present regarding the mechanisms of failure of modern metallic radial head implants. Between 1998 and 2008, 44 consecutive patients (47 elbows) underwent removal of a failed metallic radial head replacement. In 13 patients (13 elbows) the initial operation had been undertaken within one week of a fracture of the radial head, at one to six weeks in seven patients (seven elbows) and more than six weeks (mean of 2.5 years (2 to 65 months)) in 22 patients (25 elbows). In the remaining two elbows the replacement was inserted for non-traumatic reasons. The most common indication for further surgery was painful loosening (31 elbows). Revision was undertaken for stiffness in 18 elbows, instability in nine, and deep infection in two. There were signs of over-lengthening of the radius in 11 elbows. Degenerative changes were found in all but one. Only three loose implants had been fixed with cement. Instability was not identified in any of the bipolar implants.