Advertisement for orthosearch.org.uk
Results 61 - 80 of 121
Results per page:
Bone & Joint Research
Vol. 2, Issue 5 | Pages 84 - 95
1 May 2013
Sidaginamale RP Joyce TJ Lord JK Jefferson R Blain PG Nargol AVF Langton DJ

Objectives

The aims of this piece of work were to: 1) record the background concentrations of blood chromium (Cr) and cobalt (Co) concentrations in a large group of subjects; 2) to compare blood/serum Cr and Co concentrations with retrieved metal-on-metal (MoM) hip resurfacings; 3) to examine the distribution of Co and Cr in the serum and whole blood of patients with MoM hip arthroplasties; and 4) to further understand the partitioning of metal ions between the serum and whole blood fractions.

Methods

A total of 3042 blood samples donated to the local transfusion centre were analysed to record Co and Cr concentrations. Also, 91 hip resurfacing devices from patients who had given pre-revision blood/serum samples for metal ion analysis underwent volumetric wear assessment using a coordinate measuring machine. Linear regression analysis was carried out and receiver operating characteristic curves were constructed to assess the reliability of metal ions to identify abnormally wearing implants. The relationship between serum and whole blood concentrations of Cr and Co in 1048 patients was analysed using Bland-Altman charts. This relationship was further investigated in an in vitro study during which human blood was spiked with trivalent and hexavalent Cr, the serum then separated and the fractions analysed.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 40 - 48
1 Jan 2016
Matharu GS Mansour R Dada O Ostlere S Pandit HG Murray DW

Aims

The aims of this study were to compare the diagnostic test characteristics of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI) alone, and ultrasound combined with MARS-MRI for identifying intra-operative pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients undergoing revision surgery.

Methods

This retrospective diagnostic accuracy study involved 39 patients (40 MoMHRs). The time between imaging modalities was a mean of 14.6 days (0 to 90), with imaging performed at a mean of 5.3 months (0.06 to 12) before revision. The prevalence of intra-operative pseudotumours was 82.5% (n = 33).


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1617 - 1625
1 Dec 2013
Schmitz MWJL Timmer C Rijnen WHC Gardeniers JWM Schreurs BW

Despite the worldwide usage of the cemented Contemporary acetabular component (Stryker), no published data are available regarding its use in patients aged < 50 years. We undertook a mid- to long-term follow-up study, including all consecutive patients aged < 50 years who underwent a primary total hip replacement using the Contemporary acetabular component with the Exeter cemented stem between January 1999 and January 2006. There were 152 hips in 126 patients, 61 men and 65 women, mean age at surgery 37.6 years (16 to 49 yrs). One patient was lost to follow-up.

Mean clinical follow-up of all implants was 7.6 years (0.9 to 12.0). All clinical questionnaire scores, including Harris hip score, Oxford hip score and several visual analogue scales, were found to have improved. The eight year survivorship of all acetabular components for the endpoints revision for any reason or revision for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2 to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up was complete for 146 implants. The eight year survival for the endpoint radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving implants were considered radiologically loose but were asymptomatic. The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent lines (n = 20, 13.9%) in the 144 surviving cups indicates a need for continued observation in the second decade of follow-up in order to observe their influence on long-term survival.

The clinical and radiological data resulting in a ten-year survival rate > 90% in young patients support the use of the Contemporary acetabular component in this specific patient group.

Cite this article: Bone Joint J 2013;95-B:1617–25.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1045 - 1051
1 Aug 2013
Arthroplasty Society C

The purpose of this study was twofold: first, to determine whether the five-year results of hip resurfacing arthroplasty (HRA) in Canada justified the continued use of HRA; and second, to identify whether greater refinement of patient selection was warranted.

This was a retrospective cohort study that involved a review of 2773 HRAs performed between January 2001 and December 2008 at 11 Canadian centres. Cox’s proportional hazards models were used to analyse the predictors of failure of HRA. Kaplan–Meier survival analysis was performed to predict the cumulative survival rate at five years. The factors analysed included age, gender, body mass index, pre-operative hip pathology, surgeon’s experience, surgical approach, implant sizes and implant types. The most common modes of failure were also analysed.

The 2773 HRAs were undertaken in 2450 patients: 2127 in men and 646 in women. The mean age at operation was 50.5 years (sd 8.72; 18 to 82) and mean follow-up was 3.4 years (sd 2.1; 2.0 to 10.1). At the last follow-up a total of 101 HRAs (3.6%) required revision. Using revision for all causes of failure as the endpoint, Kaplan–Meier survival analysis showed a cumulative survival of 96.4% (95% confidence interval (CI) 96.1 to 96.9) at five years. With regard to gender, the five-year overall survival was 97.4% in men (95% CI 97.1 to 97.7) and 93.6% in women (95% CI 92.6 to 94.6). Female gender, smaller femoral components, specific implant types and a diagnosis of childhood hip problems were associated with higher rates of failure. The most common cause of failure was fracture of the femoral neck, followed by loosening of the femoral component.

The failure rates of HRA at five years justify the ongoing use of this technique in men. Female gender is an independent predictor of failure, and a higher failure rate at five years in women leads the authors to recommend this technique only in exceptional circumstances for women.

Cite this article: Bone Joint J 2013;95-B:1044–50.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1100 - 1102
1 Aug 2009
Kutty S Pettit P Powell JN

Fracture of the neck of the femur after resurfacing arthroplasty usually leads to failure and conversion to a total hip replacement. We describe an intracapsular fracture of the femoral neck sustained after hip resurfacing which was treated by cannulated screws, resulting in union and retention of the resurfacing implant. The result at follow-up three years later was very satisfactory with a Harris hip score of 99.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 617 - 622
1 May 2015
Haddad FS Konan S Tahmassebi J

The aim of this study was to evaluate the ten-year clinical and functional outcome of hip resurfacing and to compare it with that of cementless hip arthroplasty in patients under the age of 55 years.

Between 1999 and 2002, 80 patients were enrolled into the study: 24 were randomised (11 to hip resurfacing, 13 to total hip arthroplasty), 18 refused hip resurfacing and chose cementless total hip arthroplasty with a 32 mm bearing, and 38 insisted on resurfacing. The mean follow-up for all patients was 12.1 years (10 to 14).

Patients were assessed clinically and radiologically at one year, five years and ten years. Outcome measures included EuroQol EQ5D, Oxford, Harris hip, University of California Los Angeles and University College Hospital functional scores.

No differences were seen between the two groups in the Oxford or Harris hip scores or in the quality of life scores. Despite a similar aspiration to activity pre-operatively, a higher proportion of patients with a hip resurfacing were running and involved in sport and heavy manual labour after ten years.

We found significantly higher function scores in patients who had undergone hip resurfacing than in those with a cementless hip arthroplasty at ten years. This suggests a functional advantage for hip resurfacing. There were no other attendant problems.

Cite this article: Bone Joint J 2015; 97-B:617–22.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 738 - 745
1 Jun 2011
Davda K Lali FV Sampson B Skinner JA Hart AJ

We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging.

Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1072 - 1078
1 Aug 2010
Grammatopoulos G Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gill HS Murray DW

Pseudotumours are a rare complication of hip resurfacing. They are thought to be a response to metal debris which may be caused by edge loading due to poor orientation of the acetabular component. Our aim was to determine the optimal acetabular orientation to minimise the risk of pseudotumour formation.

We matched 31 hip resurfacings revised for pseudotumour formation with 58 controls who had a satisfactory outcome from this procedure. The radiographic inclination and anteversion angles of the acetabular component were measured on anteroposterior radiographs of the pelvis using Einzel-Bild-Roentgen-Analyse software. The mean inclination angle (47°, 10° to 81°) and anteversion angle (14°, 4° to 34°) of the pseudotumour cases were the same (p = 0.8, p = 0.2) as the controls, 46° (29° to 60°) and 16° (4° to 30°) respectively, but the variation was greater. Assuming an accuracy of implantation of ± 10° about a target position, the optimal radiographic position was found to be approximately 45° of inclination and 20° of anteversion. The incidence of pseudotumours inside the zone was four times lower (p = 0.007) than outside the zone.

In order to minimise the risk of pseudotumour formation we recommend that surgeons implant the acetabular component at an inclination of 45° (± 10) and anteversion of 20° (± 10) on post-operative radiographs. Because of differences between the radiographic and the operative angles, this may be best achieved by aiming for an inclination of 40° and an anteversion of 25°.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 42 - 49
1 Apr 2012
Kwon Y Mellon SJ Monk P Murray DW Gill HS

Objectives

Pseudotumours (abnormal peri-prosthetic soft-tissue reactions) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated metal ion levels, suggesting that excessive wear may occur due to edge-loading of these MoM implants. This study aimed to quantify in vivo edge-loading in MoMHRA patients with and without pseudotumours during functional activities.

Methods

The duration and magnitude of edge-loading in vivo was quantified during functional activities by combining the dynamic hip joint segment contact force calculated from the three-dimensional (3D) motion analysis system with the 3D reconstruction of orientation of the acetabular component and each patient’s specific hip joint centre, based on CT scans.


Bone & Joint 360
Vol. 4, Issue 1 | Pages 6 - 11
1 Feb 2015
Manktelow A Bloch B

This review examines the future of total hip arthroplasty, aiming to avoid past mistakes


The Bone & Joint Journal
Vol. 96-B, Issue 9 | Pages 1172 - 1177
1 Sep 2014
Lawrence H Deehan D Holland J Kirby J Tyson-Capper A

Abnormal wear of cobalt-containing metal-on-metal joints is associated with inflammatory pseudotumours. Cobalt ions activate human toll-like receptor 4 (TLR4), which normally responds to bacterial lipopolysaccharide (LPS) in sepsis. Activation of TLR4 by LPS increases the expression of chemokines IL-8 and CXCL10, which recruit leukocytes and activated T-cells, respectively. This study was designed to determine whether cobalt induces a similar inflammatory response to LPS by promoting the expression of IL-8 and CXCL10. A human monocytic cell line, derived from acute monocytic leukaemia, was treated with cobalt ions and expression of IL-8 and CXCL10 measured at mRNA and protein levels. Cobalt-treated macrophages showed a 60-fold increase in IL-8 mRNA, and an eightfold increase in production of the mature chemokine (both p < 0.001); expression of the CXCL10 gene and protein was also significantly increased by cobalt (both p < 0.001). Experiments were also performed in the presence of CLI-095, a TLR4-specific antagonist which abrogated the cobalt-mediated increase in IL-8 and CXCL10 expression.

These findings suggest that cobalt ions induce inflammation similar to that observed during sepsis by the simultaneous activation of two TLR4-mediated signalling pathways. These pathways result in increased production of IL-8 and CXCL10, and may be implicated in pseudotumour formation following metal-on-metal replacement.

Cite this article: Bone Joint J 2014; 96-B:1172–7.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1404 - 1410
1 Oct 2014
Wafa H Grimer RJ Jeys L Abudu AT Carter SR Tillman RM

The aim of this study was to evaluate the functional and oncological outcome of extracorporeally irradiated autografts used to reconstruct the pelvis after a P1/2 internal hemipelvectomy.

The study included 18 patients with a primary malignant bone tumour of the pelvis. There were 13 males and five females with a mean age of 24.8 years (8 to 62). Of these, seven had an osteogenic sarcoma, six a Ewing’s sarcoma, and five a chondrosarcoma. At a mean follow-up of 51.6 months (4 to 185), nine patients had died with metastatic disease while nine were free from disease. Local recurrence occurred in three patients all of whom eventually died of their disease. Deep infection occurred in three patients and required removal of their graft in two while the third underwent a hindquarter amputation for extensive flap necrosis.

The mean Musculoskeletal Tumor Society functional score of the 16 patients who could be followed-up for at least 12 months was 77% (50 to 90). Those 15 patients who completed the Toronto Extremity Salvage Score questionnaire had a mean score of 71% (53 to 85).

Extracorporeal irradiation and re-implantation of bone is a valid method of reconstruction after an internal hemipelvectomy. It has an acceptable morbidity and a functional outcome that compares favourably with other available reconstructive techniques.

Cite this article: Bone Joint J 2014;96-B:1404–10


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1308 - 1313
1 Oct 2011
Hart AJ Sabah SA Bandi AS Maggiore P Tarassoli P Sampson B A. Skinner J

Blood metal ions have been widely used to investigate metal-on-metal hip replacements, but their ability to discriminate between well-functioning and failed hips is not known. The Medicines and Healthcare products Regulatory Agency (MHRA) has suggested a cut-off level of 7 parts per billion (ppb).

We performed a pair-matched, case-control study to investigate the sensitivity and specificity of blood metal ion levels for diagnosing failure in 176 patients with a unilateral metal-on-metal hip replacement. We recruited 88 cases with a pre-revision, unexplained failed hip and an equal number of matching controls with a well-functioning hip. We investigated the 7 ppb cut-off level for the maximum of cobalt or chromium and determined optimal mathematical cut-off levels from receiver-operating characteristic curves.

The 7 ppb cut-off level for the maximum of cobalt or chromium had a specificity of 89% and sensitivity 52% for detecting a pre-operative unexplained failed metal on metal hip replacement. The optimal cut-off level for the maximum of cobalt or chromium was 4.97 ppb and had sensitivity 63% and specificity 86%.

Blood metal ions had good discriminant ability to separate failed from well-functioning hip replacements. The MHRA cut-off level of 7 ppb provides a specific test but has poor sensitivity.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 562 - 569
1 May 2008
Daniel J Pradhan A Pradhan C Ziaee H Moss M Freeman J McMinn DJW

We report a retrospective review of the incidence of venous thromboembolism in 463 consecutive patients who underwent primary total hip arthroplasty (487 procedures). Treatment included both total hip replacement and hip resurfacing, and the patients were managed without anticoagulants. The thromboprophylaxis regimen included an antiplatelet agent, generally aspirin, hypotensive epidural anaesthesia, elastic compression stockings and early mobilisation. In 258 of these procedures (244 patients) performed in 2005 (cohort A) mechanical compression devices were not used, whereas in 229 (219 patients) performed during 2006 (cohort B) bilateral intermittent pneumatic calf compression was used.

All operations were performed through a posterior mini-incision approach. Patients who required anticoagulation for pre-existing medical problems and those undergoing revision arthroplasty were excluded. Doppler ultrasonographic screening for deep-vein thrombosis was performed in all patients between the fourth and sixth post-operative days. All patients were reviewed at a follow-up clinic six to ten weeks after the operation. In addition, reponse to a questionnaire was obtained at the end of 12 weeks post-operatively.

No symptomatic calf or above-knee deep-vein thrombosis or pulmonary embolism occurred. In 25 patients in cohort A (10.2%) and in ten patients in cohort B (4.6%) asymptomatic calf deep-vein thromboses were detected ultrasonographically. This difference was statistically significant (p = 0.03). The regimen followed by cohort B offers the prospect of a low incidence of venous thromboembolism without subjecting patients to the higher risk of bleeding associated with anticoagulant use.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 572 - 579
1 May 2011
Haddad FS Thakrar RR Hart AJ Skinner JA Nargol AVF Nolan JF Gill HS Murray DW Blom AW Case CP

Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 356 - 361
1 Mar 2010
Kwon Y- Glyn-Jones S Simpson DJ Kamali A McLardy-Smith P Gill HS Murray DW

The presence of pseudotumours, which are soft-tissue masses relating to the hip, after metal-on-metal hip resurfacing arthroplasty has been associated with elevated levels of metal ions in serum, suggesting that pseudotumours occur when there is increased wear. We aimed to quantify the wear in vivo of implants revised for pseudotumours (eight) and of a control group of implants (22) revised for other reasons of failure.

We found that the implant group with pseudotumours had a significantly higher rate of median linear wear of the femoral component at 8.1 μm/year (2.75 to 25.4) than the 1.79 μm/year (0.82 to 4.15; p = 0.002) of the non-pseudotumour group. For the acetabular component a significantly higher rate of median linear wear of 7.36 μm/year (1.61 to 24.9) was observed in the pseudotumour group compared with 1.28 μm/year (0.81 to 3.33, p = 0.001) in the other group. Wear of the acetabular component in the pseudotumour group always involved the edge of the implant, indicating that edge-loading had occurred.

Our findings are the first direct evidence that pseudotumour is associated with increased wear at the metal-on-metal articulation. Furthermore, edge-loading with the loss of fluid-film lubrication may be an important mechanism of generation of wear in patients with a pseudotumour.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 989 - 989
1 Jul 2007
VENDITTOLI P GANAPATHI M LAVIGNE M


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1589 - 1590
1 Dec 2012
Scott J


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1603 - 1609
1 Dec 2013
Chana R Facek M Tilley S Walter WK Zicat B Walter WL

We report the clinical and radiological outcomes of a series of contemporary cementless ceramic-on-ceramic total hip replacements (THRs) at ten years in patients aged ≤ 55 years of age. Pre- and post-operative activity levels are described. A total of 120 consecutive ceramic cementless THRs were performed at a single centre in 110 patients from 1997 to 1999. The mean age of the patients at operation was 45 years (20 to 55). At ten years, four patients had died and six were lost to follow-up, comprising ten hips. The mean post-operative Harris hip score was 94.7 (55 to 100). Radiological analysis was undertaken in 90 available THRs of the surviving 106 hips at final review: all had evidence of stable bony ingrowth, with no cases of osteolysis. Wear was undetectable. There were four revisions. The survival for both components with revision for any cause as an endpoint was 96.5% (95% confidence interval 94.5 to 98.7). The mean modified University of California, Los Angeles activity level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6 to 10) at the ten-year post-operative period.

Alumina ceramic-on-ceramic bearings in cementless primary THR in this series have resulted in good clinical and radiological outcomes with undetectable rates of wear and excellent function in the demanding younger patient group at ten years.

Cite this article: Bone Joint J 2013;95-B:1603–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 847 - 851
1 Jul 2008
Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gibbons CLM Ostlere S Athanasou N Gill HS Murray DW

We report 17 patients (20 hips) in whom metal-on-metal resurfacing had been performed and who presented with various symptoms and a soft-tissue mass which we termed a pseudotumour. Each patient underwent plain radiography and in some, CT, MRI and ultrasonography were also performed. In addition, histological examination of available samples was undertaken.

All the patients were women and their presentation was variable. The most common symptom was discomfort in the region of the hip. Other symptoms included spontaneous dislocation, nerve palsy, a noticeable mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. To date, 13 of the 20 hips have required revision to a conventional hip replacement. Two are awaiting revision.

We estimate that approximately 1% of patients who have a metal-on-metal resurfacing develop a pseudotumour within five years. The cause is unknown and is probably multifactorial. There may be a toxic reaction to an excess of particulate metal wear debris or a hypersensitivity reaction to a normal amount of metal debris. We are concerned that with time the incidence of these pseudotumours may increase. Further investigation is required to define their cause.