The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 7 | Pages 974 - 978
1 Sep 2004
Thomas SR Shukla D Latham PD

Cemented titanium stems in hip arthroplasty are associated with proximal cement-stem debonding and early failure. This was well publicised with the 3M Capital hip. However, corrosion in this setting has been reported with only one stem design and is less widely accepted.

We present a series of 12 cemented titanium Furlong Straight Stems which required revision at a mean of 78 months for thigh pain. At revision the stems were severely corroded in a pattern which was typical of crevice corrosion. Symptoms were eliminated after revision to an all-stainless steel femoral prosthesis of the same design. We discuss the likely causes for the corrosion. The combination of a titanium stem and cement appears to facilitate crevice corrosion.


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 8 | Pages 1206 - 1207
1 Nov 2002
AMSTUTZ HC


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 5 | Pages 869 - 875
1 Sep 1998
Edidin AA Merritt PO Hack BH Manley MT

We describe the development and early clinical application of a ported, proximally-cemented titanium stem for cemented total hip arthroplasty. PMMA bone cement is delivered to the proximal femur under pressure after the stem has been positioned within the femoral canal. A mid-stem cement occluder contains the cement to the proximal stem only. A tapered body is incorporated in the design of the stem to reduce the structural stiffness and hence the degree of stress shielding within the reconstructed joint.

We performed preclinical studies to measure the reduction in porosity and the pressurisation achieved. The porosity, as measured by the void percentage within the cured cement mantle, was reduced by more than 50% and there was an almost threefold increase in the mean pressure. Mechanical testing of the stem, using a three-point bend test, showed that the addition of cement injection ports on the anterior and posterior sides of the body of the proximal stem did not reduce its strength. Finite-element analysis indicated that, compared with a fully-cemented conventional stem, there was no change in the stresses within the cement mantle. In a series of 40 proximally-cemented stems followed for up to six years (mean 51 months) the mean Harris hip score was 91, and 85% of patients had good or excellent results. There was excellent pain relief, an increased level of activity and good patient satisfaction. One mechanical failure of the stem required revision at three years after implantation.

The early results indicate that the clinical performance was equal to that achieved with other modern cemented stems. Radiological evaluation showed excellent results with no evidence of stress shielding. Further follow-up will determine if long-term stress shielding is reduced and if revision is made easier by the absence of a distal cement mantle.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 170 - 175
1 Feb 2009
Young L Duckett S Dunn A

We describe the survivorship of the Exeter femoral component in a District General Hospital. Between 1994 and 1996, 230 Exeter Universal cemented femoral components were implanted in 215 patients who were reviewed at a mean of 11.2 years (10 to 13). We used one acetabular implant, the Elite Ogee component, in 218 of the 230 hips.

During the period of this study 76 patients (79 hips) died. Of the remaining 139 patients (151 hips), 121 were able to attend for radiological analysis at a minimum of ten years. One patient was lost to follow-up.

No femoral component was revised for aseptic loosening. Three hips were revised for deep infection and six acetabular components required revision, four for loosening and two for recurrent dislocation. Taking the ‘worst-case scenario’ including the one patient lost to follow-up, the overall survival rate was 94.4% at 13 years.

Our results confirm excellent medium-term results for the Exeter Universal femoral component, implanted in a general setting. The excellent survival of this femoral component, when used in combination with the Ogee acetabular component, suggests that this is a successful pairing.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 8 | Pages 1118 - 1123
1 Nov 2004
Singh S Trikha SP Edge AJ

We describe the clinical and radiological results of 38 total hip replacements (THR) using the JRI Furlong hydroxyapatite-ceramic (HAC)-coated femoral component in patients younger than 50 years. The mean age at the time of operation was 42 years (22 to 49) and the mean length of follow-up was ten years (63 to 170 months). All patients receiving a Furlong HAC THR were entered into the study regardless of the primary pathology including patients who had undergone previous hip surgery.

The mean Harris hip score improved from 44 before operation to 92 at the last postoperative review. After 12 years the cumulative surivival for the stem was 100% (95% confidence interval 89 to 100). No femoral component was revised. Our results show that the Furlong HAC implant gives excellent long-term results in young patients with high demands.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 1 | Pages 13 - 19
1 Jan 2004
Learmonth ID


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 8 | Pages 1182 - 1188
1 Nov 2000
Barker DS Wang AW Yeo MF Nawana NS Brumby SA Pearcy MJ Howie DW

We studied the effect of the surface finish of the stem on the transfer of load in the proximal femur in a sheep model of cemented hip arthroplasty. Strain-gauge analysis and corresponding finite-element (FE) analysis were performed to assess the effect of friction and creep at the cement-stem interface.

No difference was seen between the matt and polished stems. FE analysis showed that the effects of cement creep and friction at the stem-cement interface on femoral strain were small compared with the effect of inserting a cemented stem.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1574 - 1580
1 Dec 2007
Hallan G Lie SA Furnes O Engesaeter LB Vollset SE Havelin LI

Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data.

With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well.

Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1115 - 1121
1 Aug 2007
Messick KJ Miller MA Damron LA Race A Clarke MT Mann KA

The role of vacuum mixing on the reduction of porosity and on the clinical performance of cemented total hip replacements remains uncertain. We have used paired femoral constructs prepared with either hand-mixed or vacuum-mixed cement in a cadaver model which simulated intra-operative conditions during cementing of the femoral component. After the cement had cured, the distribution of its porosity was determined, as was the strength of the cement-stem and cement-bone interfaces.

The overall fraction of the pore area was similar for both hand-mixed and vacuum-mixed cement (hand 6%; vacuum 5.7%; paired t-test, p = 0.187). The linear pore fractions at the interfaces were also similar for the two techniques. The pore number-density was much higher for the hand-mixed cement (paired t-test, p = 0.0013). The strength of the cement-stem interface was greater with the hand-mixed cement (paired t-test, p = 0.0005), while the strength of the cement-bone interface was not affected by the conditions of mixing (paired t-test, p = 0.275). The reduction in porosity with vacuum mixing did not affect the porosity of the mantle, but the distribution of the porosity can be affected by the technique of mixing used.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 921 - 927
1 Jul 2005
Glyn-Jones S Gill HS Beard DJ McLardy-Smith P Murray DW

Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability.

Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°, sd 0.42°) and internally rotate (mean 1.28°, sd 0.99°). The CPS-Plus showed no significant valgus rotation (mean 0.2°, sd 0.42°) or internal rotation (mean −0.03°, sd 0.75°). A wider, more rectangular cross-section improves rotational stability and may have a better long-term outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 6 | Pages 838 - 842
1 Aug 2001
Chambers IR Fender D McCaskie AW Reeves BC Gregg PJ

The radiological features of the cement mantle around total hip replacements (THRs) have been used to assess aseptic loosening. In this case-control study we investigated the risk of failure of THR as predictable by a range of such features using data from patients recruited to the Trent Regional Arthroplasty Study (TRAS).

An independent radiological assessment was undertaken on Charnley THRs with aseptic loosening within five years of surgery and on a control group from the TRAS database. Chi-squared tests were used to test the probability of obtaining the observed data by chance, and odds ratios were calculated to estimate the strength of association for different features.

Several features were associated with a clinically important increase (> twofold) in the risk of loosening, which was statistically significant for four features (p < 0.01). Inadequate cementation (Barrack C and D grades) was the most significant feature, with an estimated odds ratio of 9.5 (95% confidence interval 3.2 to 28.4, p < 0.0001) for failure.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 2 | Pages 259 - 265
1 Mar 1993
Dall D Learmonth I Solomon M Miles A Davenport J

We report the results of a 4- to 17-year clinical and radiological follow-up of 264 Charnley first-generation stems in comparison with those of 402 second- and subsequent-generation stems. The incidence of fracture was 4.1% in first-generation stems and 0.5% in second- and subsequent-generation stems. The incidence of stem loosening requiring or likely to require revision was 3.1% in first-generation and 11.4% in second-generation stems. We believe that the increased loosening rate in second- and subsequent-generation stems is due to their larger cross-sectional area, which produces an increase in flexural stiffness.


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 6 | Pages 912 - 917
1 Nov 1994
Karrholm J Borssen B Lowenhielm G Snorrason F

Roentgen stereophotogrammetry was used to measure the migration of the centre of the femoral head in 84 cemented Lubinus SP I hip arthroplasties (58 primary operations, 26 revisions). Four to seven years later, seven femoral components had been revised because of painful loosening. These implants showed greater subsidence, medial migration and posterior migration during the first two postoperative years than did the hips which had not been revised. Six months after operation, subsidence of more than 0.33 mm combined with a total migration of more than 0.85 mm predicted an increased risk of subsequent revision; the amount of subsidence at two years was an even better predictor. The probability of revision was greater than 50% if the subsidence at two years was 1.2 mm or more.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1435 - 1441
1 Oct 2010
Bischel OE Böhm PM

Endoprosthetic reconstruction following resection of 31 tumours of the proximal femur in 30 patients was performed using a Wagner SL femoral revision stem. The mean follow-up was 25.6 months (0.6 to 130.0). Of the 28 patients with a metastasis, 27 died within a mean follow-up period of 18.1 months (0.6 to 56.3) after the operation, and the remaining patient was excluded from the study 44.4 months post-operatively when the stem was removed. The two patients with primary bone tumours were still alive at the latest follow-up of 81.0 and 130.0 months, respectively. One stem only was removed for suspected low-grade infection 44.4 months post-operatively. The worst-case survival rate with removal of the stem for any cause and/or loss to follow-up was 80.0% (95% confidence interval 44.9 to 100) at 130.0 months. The mean Karnofsky index increased from 44.2% (20% to 70%) pre-operatively to 59.7% (0% to 100%) post-operatively, and the mean Merle d’Aubigné score improved from 4.5 (0 to 15) to 12.0 (0 to 18). The mean post-operative Musculoskeletal Tumour Society score was 62.4% (3.3% to 100%). The Wagner SL femoral revision stem offers an alternative to special tumour prostheses for the treatment of primary and secondary tumours of the proximal femur. The mid-term results are very promising, but long-term experience is necessary


The Bone & Joint Journal
Vol. 104-B, Issue 11 | Pages 1196 - 1201
1 Nov 2022
Anderson CG Brilliant ZR Jang SJ Sokrab R Mayman DJ Vigdorchik JM Sculco PK Jerabek SA

Aims. Although CT is considered the benchmark to measure femoral version, 3D biplanar radiography (hipEOS) has recently emerged as a possible alternative with reduced exposure to ionizing radiation and shorter examination time. The aim of our study was to evaluate femoral stem version in postoperative total hip arthroplasty (THA) patients and compare the accuracy of hipEOS to CT. We hypothesize that there will be no significant difference in calculated femoral stem version measurements between the two imaging methods. Methods. In this study, 45 patients who underwent THA between February 2016 and February 2020 and had both a postoperative CT and EOS scan were included for evaluation. A fellowship-trained musculoskeletal radiologist and radiological technician measured femoral version for CT and 3D EOS, respectively. Comparison of values for each imaging modality were assessed for statistical significance. Results. Comparison of the mean postoperative femoral stem version measurements between CT and 3D hipEOS showed no significant difference (p = 0.862). In addition, the two version measurements were strongly correlated (r = 0.95; p < 0.001), and the mean paired difference in postoperative femoral version for CT scan and 3D biplanar radiography was -0.09° (95% confidence interval -1.09 to 0.91). Only three stem measurements (6.7%) were considered outliers with a > 5° difference. Conclusion. Our study supports the use of low-dose biplanar radiography for the postoperative assessment of femoral stem version after THA, demonstrating high correlation with CT. We found no significant difference for postoperative femoral version when comparing CT to 3D EOS. We believe 3D EOS is a reliable option to measure postoperative femoral version given its advantages of lower radiation dosage and shorter examination time. Cite this article: Bone Joint J 2022;104-B(11):1196–1201


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 435 - 441
1 May 2024
Angelomenos V Mohaddes M Kärrholm J Malchau H Shareghi B Itayem R

Aims. Refobacin Bone Cement R and Palacos R + G bone cement were introduced to replace the original cement Refobacin Palacos R in 2005. Both cements were assumed to behave in a biomechanically similar fashion to the original cement. The primary aim of this study was to compare the migration of a polished triple-tapered femoral stem fixed with either Refobacin Bone Cement R or Palacos R + G bone cement. Repeated radiostereometric analysis was used to measure migration of the femoral head centre. The secondary aims were evaluation of cement mantle, stem positioning, and patient-reported outcome measures. Methods. Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected. Results. Median distal migration (y-axis) at two years for the Refobacin-Optivac system was -0.79 mm (-2.01 to -0.09), for the Refobacin-Optipac system was -0.75 mm (-2.16 to 0.20), and for the Palacos-Optivac system was -1.01 mm (-4.31 to -0.29). No statistically significant differences were found between the groups. Secondary outcomes did not differ statistically between the groups at the two-year follow-up. Conclusion. At two years, we found no significant differences in distal migration or clinical outcomes between the three groups. Our data indicate that Refobacin Bone Cement R and Palacos R + G are comparable in terms of stable fixation and early clinical outcomes. Cite this article: Bone Joint J 2024;106-B(5):435–441


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 504 - 510
1 May 2023
Evans JT Salar O Whitehouse SL Sayers A Whitehouse MR Wilton T Hubble MJW

Aims. The Exeter V40 femoral stem is the most implanted stem in the National Joint Registry (NJR) for primary total hip arthroplasty (THA). In 2004, the 44/00/125 stem was released for use in ‘cement-in-cement’ revision cases. It has, however, been used ‘off-label’ as a primary stem when patient anatomy requires a smaller stem with a 44 mm offset. We aimed to investigate survival of this implant in comparison to others in the range when used in primary THAs recorded in the NJR. Methods. We analyzed 328,737 primary THAs using the Exeter V40 stem, comprising 34.3% of the 958,869 from the start of the NJR to December 2018. Our exposure was the stem, and the outcome was all-cause construct revision. We stratified analyses into four groups: constructs using the 44/00/125 stem, those using the 44/0/150 stem, those including a 35.5/125 stem, and constructs using any other Exeter V40 stem. Results. In all 328,737 THAs using an Exeter V40 stem, the revision estimate was 2.8% (95% confidence interval (CI) 2.7 to 2.8). The 44/00/125 stem was implanted in 2,158 primary THAs, and the ten-year revision estimate was 4.9% (95% CI 3.6 to 6.8). Controlling for age, sex, year of operation, indication, and American Society of Anesthesiologists grade demonstrated an increased overall hazard of revision for constructs using the 44/00/125 stem compared to constructs using other Exeter V40 femoral stems (hazard ratio 1.8 (95% CI 1.4 to 2.3)). Conclusion. Although the revision estimate is within the National Institute for Health and Care Excellence ten-year benchmark, survivorship of constructs using the 44/00/125 stem appears to be lower than the rest of the range. Adjusted analyses will not take into account ‘confounding by indication’, e.g. patients with complex anatomy who may have a higher risk of revision. Surgeons and patients should be reassured but be aware of the observed increased revision estimate, and only use this stem when other implants are not suitable. Cite this article: Bone Joint J 2023;105-B(5):504–510


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims. Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. Methods. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis. Results. Results from mixed model analyses revealed a lower mean retroversion for line-to-line (0.72° (95% confidence interval (CI) 0.38° to 1.07°; p < 0.001), but no significant differences in subsidence between the techniques (-0.15 mm (95% CI -0.53 to 0.227; p = 0.429) at 24 months. Radiolucent lines measuring < 2 mm wide were found in three and five arthroplasties cemented by the standard and line-to-line method, respectively. Conclusion. The cemented Corail stem with a force-closed design seems to settle earlier and better with the line-to-line cementing method, although for subsidence the difference was not significant. However, the lower rate of migration into retroversion may reduce the wear and cement deformation, contributing to good long-term fixation and implant survival. Cite this article: Bone Joint J 2022;104-B(1):19–26


Bone & Joint 360
Vol. 13, Issue 4 | Pages 13 - 16
2 Aug 2024

The August 2024 Hip & Pelvis Roundup. 360. looks at: Understanding perceived leg length discrepancy post-total hip arthroplasty: the role of pelvic obliquity; Influence of femoral stem design on revision rates in total hip arthroplasty; Outcomes of arthroscopic labral treatment of femoroacetabular impingement in adolescents; Characteristics and quality of online searches for direct anterior versus posterior approach for total hip arthroplasty; Rapid return to braking after anterior and posterior approach total hip arthroplasty; How much protection does a collar provide?; Timing matters: reducing infection risk in total hip arthroplasty with corticosteroid injection intervals; Identifying pain recovery patterns in total hip arthroplasty using PROMIS data


Bone & Joint 360
Vol. 12, Issue 6 | Pages 17 - 20
1 Dec 2023

The December 2023 Hip & Pelvis Roundup. 360. looks at: Early hip fracture surgery is safe for patients on direct oral anticoagulants; Time to return to work by occupational class after total hip or knee arthroplasty; Is there a consensus on air travel following hip and knee arthroplasty?; Predicting whether patients will achieve minimal clinically important differences following hip or knee arthroplasty; High-dose dual-antibiotic-loaded cement for hip hemiarthroplasty in the UK (WHiTE 8): a randomized controlled trial; Vitamin E – a positive thing in your poly?; Hydroxapatite-coated femoral stems: is there a difference in fixation?