We evaluated the accuracy of augmented reality (AR)-based navigation assistance through simulation of bone tumours in a pig femur model. We developed an AR-based navigation system for bone tumour resection, which could be used on a tablet PC. To simulate a bone tumour in the pig femur, a cortical window was made in the diaphysis and bone cement was inserted. A total of 133 pig femurs were used and tumour resection was simulated with AR-assisted resection (164 resection in 82 femurs, half by an orthropaedic oncology expert and half by an orthopaedic resident) and resection with the conventional method (82 resection in 41 femurs). In the conventional group, resection was performed after measuring the distance from the edge of the condyle to the expected resection margin with a ruler as per routine clinical practice.Objectives
Methods
We conducted a retrospective study to investigate the effect of femoral bowing on the placement of components in total knee replacement (TKR), with regard to its effect on reestablishing the correct mechanical axis, as we hypothesised that computer-assisted total knee replacement (CAS-TKR) would produce more accurate alignment than conventional TKR. Between January 2006 and December 2009, 212 patients (306 knees) underwent TKR. The conventional TKR was compared with CAS-TKR for accuracy of placement of the components and post-operative alignment, as determined by five radiological measurements. There were significant differences in the reconstructed mechanical axes between the bowed and the non-bowed group after conventional TKR (176.2° ( For patients with significant femoral bowing, the reconstructed mechanical axes were significantly closer to normal in the CAS group than in the conventional group (179.2° (
Improvements in the surgical technique of total
knee replacement (TKR) are continually being sought. There has recently
been interest in three-dimensional (3D) pre-operative planning using
magnetic resonance imaging (MRI) and CT. The 3D images are increasingly
used for the production of patient-specific models, surgical guides
and custom-made implants for TKR. The users of patient-specific instrumentation (PSI) claim that
they allow the optimum balance of technology and conventional surgery
by reducing the complexity of conventional alignment and sizing
tools. In this way the advantages of accuracy and precision claimed
by computer navigation techniques are achieved without the disadvantages
of additional intra-operative inventory, new skills or surgical
time. This review describes the terminology used in this area and debates
the advantages and disadvantages of PSI.
Primary total knee arthroplasty (TKA) is a reliable
procedure with reproducible long-term results. Nevertheless, there
are conditions related to the type of patient or local conditions
of the knee that can make it a difficult procedure. The most common
scenarios that make it difficult are discussed in this review. These
include patients with many previous operations and incisions, and
those with severe coronal deformities, genu recurvatum, a stiff knee,
extra-articular deformities and those who have previously undergone
osteotomy around the knee and those with chronic dislocation of
the patella. Each condition is analysed according to the characteristics of
the patient, the pre-operative planning and the reported outcomes. When approaching the difficult primary TKA surgeons should use
a systematic approach, which begins with the review of the existing
literature for each specific clinical situation. Cite this article:
We evaluated the oncological and functional outcome
of 18 patients, whose malignant bone tumours were excised with the
assistance of navigation, and who were followed up for more than
three years. There were 11 men and seven women, with a mean age
of 31.8 years (10 to 57). There were ten operations on the pelvic
ring and eight joint-preserving limb salvage procedures. The resection
margins were free of tumour in all specimens. The tumours, which
were stage IIB in all patients, included osteosarcoma, high-grade
chondrosarcoma, Ewing’s sarcoma, malignant fibrous histiocytoma
of bone, and adamantinoma. The overall three-year survival rate
of the 18 patients was 88.9% (95% confidence interval (CI) 75.4
to 100). The three-year survival rate of the patients with pelvic malignancy
was 80.0% (95% CI 55.3 to 100), and of the patients with metaphyseal
malignancy was 100%. The event-free survival was 66.7% (95% CI 44.9
to 88.5). Local recurrence occurred in two patients, both of whom
had a pelvic malignancy. The mean Musculoskeletal Tumor Society
functional score was 26.9 points at a mean follow-up of 48.2 months
(22 to 79). We suggest that navigation can be helpful during surgery for
musculoskeletal tumours; it can maximise the accuracy of resection
and minimise the unnecessary sacrifice of normal tissue by providing
precise intra-operative three-dimensional radiological information.
The August 2014 Knee Roundup360 looks at: re-admission following total knee replacement; out with the old and in with the new? computer navigation revisited; approach less important in knee replacement; is obesity driving a rise in knee replacements?; knee replacement isn’t cheap in the obese; cruciate substitution doesn’t increase knee flexion; and sonication useful diagnostic aid in two-stage revision.
Orientation of the native acetabular plane as defined by the transverse acetabular ligament (TAL) and the posterior labrum was measured intra-operatively using computer-assisted navigation in 39 hips. In order to assess the influence of alignment on impingement, the range of movement was calculated for that defined by the TAL and the posterior labrum and compared with a standard acetabular component position (abduction 45°/anteversion 15°). With respect to the registration of the plane defined by the TAL and the posterior labrum, there was moderate interobserver agreement (r = 0.64, p <
0.001) and intra-observer reproducibility (r = 0.73, p <
0.001). The mean acetabular component orientation achieved was abduction of 41° (32° to 51°) and anteversion of 18° (−1° to 36°). With respect to the Lewinnek safe zone (abduction 40° ±10°, anteversion 15° ±10°), 35 of the 39 acetabular components were within this zone. However, there was no improvement in the range of movement (p = 0.94) and no significant difference in impingement (p = 0.085). Alignment of the acetabular component with the TAL and the posterior labrum might reduce the variability of acetabular component placement in total hip replacement. However, there is only a moderate interobserver agreement and intra-observer reliability in the alignment of the acetabular component using the TAL and the posterior labrum. No reduction in impingement was found when the acetabular component was aligned with the TAL and the posterior labrum, compared with a standard acetabular component position.
We are entering a new era with governmental bodies
taking an increasingly guiding role, gaining control of registries,
demanding direct access with release of open public information
for quality comparisons between hospitals. This review is written
by physicians and scientists who have worked with the Swedish Knee
Arthroplasty Register (SKAR) periodically since it began. It reviews
the history of the register and describes the methods used and lessons
learned. Cite this article:
In a prospective randomised clinical study acetabular components were implanted either freehand (n = 30) or using CT-based (n = 30) or imageless navigation (n = 30). The position of the component was determined post-operatively on CT scans of the pelvis. Following conventional freehand placement of the acetabular component, only 14 of the 30 were within the safe zone as defined by Lewinnek et al (40° inclination Imageless navigation proved as reliable as that using CT in positioning the acetabular component.
The orientation of the acetabular component can
influence both the short- and long-term outcomes of total hip replacement
(THR). We performed a prospective, randomised, controlled trial
of two groups, comprising of 40 patients each, in order to compare
freehand introduction of the component with introduction using the transverse
acetabular ligament (TAL) as a reference for anteversion. Anteversion
and inclination were measured on pelvic radiographs. With respect to anteversion, in the freehand group 22.5% of the
components were outside the safe zone With respect to inclination, in the freehand group 37.5% of the
components were outside the safe zone The transverse acetabular ligament may be used to obtain the
appropriate anteversion when introducing the acetabular component
during THR, but not acetabular component inclination. Cite this article:
We performed A total of 12 cadaveric lower limbs were tested with a commercial
image-free navigation system using trackers secured by bone screws.
We then tested a non-invasive fabric-strap system. The lower limb
was secured at 10° intervals from 0° to 60° of knee flexion and
100 N of force was applied perpendicular to the tibia. Acceptable
coefficient of repeatability (CR) and limits of agreement (LOA)
of 3 mm were set based on diagnostic criteria for anterior cruciate
ligament (ACL) insufficiency.Objectives
Methods
Obtaining a balanced flexion gap with correct
femoral component rotation is one of the prerequisites for a successful
outcome after total knee replacement (TKR). Different techniques
for achieving this have been described. In this study we prospectively
compared gap-balancing Both groups systematically reproduced a similar external rotation
of the femoral component relative to the surgical transepicondylar
axis: 2.4°
The aim of this prospective single-centre study
was to assess the difference in clinical outcome between total knee replacement
(TKR) using computerised navigation and that of conventional TKR.
We hypothesised that navigation would give a better result at every
stage within the first five years. A total of 195 patients (195
knees) with a mean age of 70.0 years (39 to 89) were allocated alternately
into two treatment groups, which used either conventional instrumentation
(group A, 97 knees) or a navigation system (group B, 98 knees).
After five years, complete clinical scores were available for 121
patients (62%). A total of 18 patients were lost to follow-up. Compared
with conventional surgery, navigated TKR resulted in a better mean
Knee Society score (p = 0.008). The difference in mean Knee Society
scores over time between the two groups was not constant (p = 0.006),
which suggests that these groups differed in their response to surgery
with time. No significant difference in the frequency of malalignment
was seen between the two groups. In summary, computerised navigation resulted in a better functional
outcome at five years than conventional techniques. Given the similarity
in mechanical alignment between the two groups, rotational alignment
may prove to be a better method of identifying differences in clinical
outcome after navigated surgery.
Surgeons need to be able to measure angles and distances in three dimensions in the planning and assessment of knee replacement. Computed tomography (CT) offers the accuracy needed but involves greater radiation exposure to patients than traditional long-leg standing radiographs, which give very little information outside the plane of the image. There is considerable variation in CT radiation doses between research centres, scanning protocols and individual scanners, and ethics committees are rightly demanding more consistency in this area. By refining the CT scanning protocol we have reduced the effective radiation dose received by the patient down to the equivalent of one long-leg standing radiograph. Because of this, it will be more acceptable to obtain the three-dimensional data set produced by CT scanning. Surgeons will be able to document the impact of implant position on outcome with greater precision.
The purpose of this study was to define immediate post-operative ‘quality’ in total hip replacements and to study prospectively the occurrence of failure based on these definitions of quality. The evaluation and assessment of failure were based on ten radiological and clinical criteria. The cumulative summation (CUSUM) test was used to study 200 procedures over a one-year period. Technical criteria defined failure in 17 cases (8.5%), those related to the femoral component in nine (4.5%), the acetabular component in 32 (16%) and those relating to discharge from hospital in five (2.5%). Overall, the procedure was considered to have failed in 57 of the 200 total hip replacements (28.5%). The use of a new design of acetabular component was associated with more failures. For the CUSUM test, the level of adequate performance was set at a rate of failure of 20% and the level of inadequate performance set at a failure rate of 40%; no alarm was raised by the test, indicating that there was no evidence of inadequate performance. The use of a continuous monitoring statistical method is useful to ensure that the quality of total hip replacement is maintained, especially as newer implants are introduced.
This annotation considers the place of extra-articular
reconstruction in the treatment of anterior cruciate ligament (ACL)
deficiency. Extra-articular reconstruction has been employed over
the last century to address ACL deficiency. However, the technique
has not gained favour, primarily due to residual instability and
the subsequent development of degenerative changes in the lateral
compartment of the knee. Thus intra-articular reconstruction has
become the technique of choice. However, intra-articular reconstruction
does not restore normal knee kinematics. Some authors have recommended
extra-articular reconstruction in conjunction with an intra-articular
technique. The anatomy and biomechanics of the anterolateral structures
of the knee remain largely undetermined. Further studies to establish
the structure and function of the anterolateral structures may lead
to more anatomical extra-articular reconstruction techniques that
supplement intra-articular reconstruction. This might reduce residual
pivot shift after an intra-articular reconstruction and thus improve
the post-operative kinematics of the knee.